Scale parameters and functions

6.14 DR 13 RS485 parameters

Variable Note Type Length RW Default Min. Max. Modbus


(bytes) registers
Byte swap for "0: Normal BIT 0 rw 0 0 1 1564.8
text fields 1: The two characters in
a Modbus register are
transferred in the wrong
order"
Byte swap for "0: big endian (MSB first) BIT 0 rw 0 0 1 1564.7
16-bit values 1: little endian (LSB
first)"
Byte swap for "0: big endian (MSB first) BIT 0 rw 0 0 1 1564.6
the two halves 1: little endian (LSB
of a 32-bit first)"
value
(INT32/FLOAT)
Word swap for "0: big endian (MSW BIT 0 rw 0 0 1 1564.5
32-bit integers first)
(interchanging 1: little endian (LSW
the two Mod- first)"
bus registers)
Bit 12 Reserve BIT 0 rw 0 0 1 1564.4
Bit 13 Reserve BIT 0 rw 0 0 1 1564.3
Bit 14 Reserve BIT 0 rw 0 0 1 1564.2
Bit 15 Reserve BIT 2 rw 0 0 1 1564.1
RS485 Modbus MODBUS address for USHORT 2 rw 20 1 255 1565
address Vito module
(Page 103)
Decimal place Decimal place for SHORT 2 rw 0 - - 1566
for Siebert Siebert display
indicator
(Page 103)
MODBUS RTU Delay time for response USHORT 2 rw 0 - - 1567
frame delay with MODBUS RTU in
ms (RS485)
Reserve 3 Reserve FLOAT 4 rw 0 - - 1568

6.14.2 RS485 protocol


This parameter defines the protocol for communication via the RS485 interface.

Value Protocol
0 No communication/protocol
1 Modbus RTU
2 SIEBERT display

SIWAREX WP521/WP522
102 Manual, 04/2016, A5E36695225A