The following information is displayed in the debug mode during normal operation of
Fabius: system data, measured values, error log. One debug mode page is available.
System data cannot be changed.
An additional screen page is displayed. Fabius remains in the normal mode and can
be operated as desired.
3-8
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Fabius® Repair Instructions
LOG
XX XX XX XX XX XX XX XX XX
3-9
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
®
Fabius
Repair Instructions
The following information is displayed in the debug mode during normal operation of
Fabius: system data, measured values, error log. Two pages are available in this
debug mode. System data cannot be changed.
An additional screen page is displayed. Fabius remains in the normal operating mode
and can be operated as desired.
The selected numeric code is displayed on black background. Copyright 2000 Draeger Medical, Inc.
3-10
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Fabius® Repair Instructions
Rev. A
3-11
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
®Fabius
Repair Instructions
Vtix XXXX mL
Expiratory peak flow in L/min
(normal operating range: 0 to 150 L/min)
Vtev XXXX mL
Plateau pressure in hPa
(normal operating range 0 to 99 hPa)
Plat XX hPa
3-12
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Fabius® Repair Instructions
Overview: VEN
Rev. A
3-13
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
®Fabius
Repair Instructions
Overview: MON
O2 concentration or O2 concentration
not corresponding to calibration mode
(21%/100%).
0010 FiO2/relative error between FiO2 O2 sensor needs replaced.
sensor channels > 25%.
0020 FiO2 calibration constant from O2 sensor not yet calibrated.
EEPROM incorrect. Data loss in the EEPROM.
0040 FiO2/coding for calibration. Current error occurred during
calibration.
3-14
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Fabius® Repair Instructions
3-15
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
® Fabius
Repair Instructions
Overview: EEP
3-16
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Fabius® Repair Instructions
3-17
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
®
Fabius
Repair Instructions
11 Error Log
Rev. A
3-18
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Fabius® Repair Instructions
3-19
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
® Fabius
Repair Instructions
3-20
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Fabius® Repair Instructions
The DrägerService Mode can only be accessed while Fabius is in the standby mode.
• Press the “Menu” key.
• Select “keycode” by turning rotary knob.
• Confirm by pressing rotary knob.
3-21
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
®
Fabius
Repair Instructions
SERVICE
Calibration pressure
Calibration peep valve DISABLED
Reset logbook EMPTY
Reset operating hours LOST XXXXXXXXX
Function MON. + VENT.
Init system *
* Initialization of Fabius
3-22
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Fabius® Repair Instructions
The DrägerService Mode can only be accessed while Fabius is in the standby mode.
• Press the “Menu” key.
• Select “keycode” by turning rotary knob.
• Confirm by pressing rotary knob.
3-23
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
®
Fabius
Repair Instructions
SERVICE
* Initialization of Fabius
** P (pressure) corresponds to PWM (pulse width modulation) ratio
3-24
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Fabius® Repair Instructions
3-25