Low Level Oxygenates Analyser: Astm D7423, Astm D7754, Uop 960 Robust Highest Uptime 99% Optional Ms Detection

You might also like

You are on page 1of 4

Low level oxygenates analyser

• astM D7423, astM D7754, uoP 960


• robust
• Highest uptime > 99%
• optional Ms detection

an 218Wa0420e

Global analyser solutions


GAS offers custom configured GC The low level oxygenates analyser from GAS
analysers for many application fields for analyses oxygen containing components like
over 40 years. GAS analysers are designed alcohols and ethers at (sub) ppm level. The
to meet many standardised methods from presence of these components in
GPA, ASTM, UOP, ISO, EN and others. The hydrocarbon feedstock reduces catalyst
efficient configurations are based on lifetime, even at low ppm level. Analysis at
proven GC technology, resulting in low levels is also demanded by
robust instruments with an optimal environmental requirements.
return on investment.

ASTM D7423, UOP 960 Figure 1. Diagram ASTM


Methods ASTM D4815 and EN 13132 (see application notes) cover the D7423 analyser with SSL
analysis of oxygenated components in finished products at % level, injection.
while below described methods ASTM D7423 and UOP 960 cover 0.1-
1000 ppm range. Oxygen containing compounds with boiling points up
to 100 oC are analysed in hydrocarbon streams with final boiling point
below 250 oC.
The instrument diagram is shown in figure 1. A highly selective
capillary column (Lowox) separates the components of interest from
the hydrocarbon matrix. Heavier components are backflushed using a
Deans column switch configuration. The diagram shows injection using
SSL (liquid sample injector). GSV (Gas Sampling Valve) and LSV
(Liquid Sampling Valve) with Pressure Facility for analysing LPG
samples are available as well (figure 9). Figure 2 and 3 show
chromatograms of 100 ppm calibration standard and real sample.
Figure 8 demonstrates good repeatability.

LOWOX CALIBRATION STANDARD


(100 ppm for each component)

Dimethylether
Acetaldehyde
Methanol
Ethanol
Propylether
Propionaldehyde
iso-propanol
T-Butanol
Propanol
Methyl Ethyl Ketone
Diethylether
Iso-Butanol
N-Butanol
Iso-Butanal
Buteraldehyde
Ethyl Tert Butyl Ether
Di-Isopropylether
Methyl Tert Butyl Ether
Tert Amyl Methyl Ether TAME
Acetone
Valeraldehyde
Allyl Alcohol

Figure 2. Lowox calibration standard 100 ppm

Global analyser solutions


Figure 3. Chromatogram: Pygas sample
Figure 5. Trace 1310 GC with InstantConnect
injector and detector technology
ASTM D7754
The GAS low level oxygenates analyser covers ASTM D7754 as well, for samples
containing 1-15% Ethanol.

Trace GC 1300
The low level oxygenates analyser is based on Thermo Trace GC 1300 with modular
injector and detector technology, providing low maintenance costs and high uptime
(figure 5, 6).

Lowox analysis using GC-MS


High yield catalyst are extremely prone to poisoning by feedstock impurities like Figure 6. Swift exchange of injector and detector
oxygenated components. Therefore a solution based on GC-MS was developed for
substantial increase in identification and sensitivity compared to GC-FID analyses.
Figure 4 demonstrates sensitivity gain factors of individual oxygenated components in
naphta. The gain is particularly significant when SIM mode is used. The ISQ mass
spectrometer (figure 7) conveniently combines full scan and SIM in a single run for both
increased sensitivity and reliable identification.

Figure 4. Average sensitivity gain factors in different MS detection modes. FS=full scan; Figure 7. Trace 1310 GC with ISQ Mass
EIC=extracted ion chromatogram; SIM= selected ion monitoring. %RSD at 10 ppm level (n=6). Spectrometer and RSH autosampler

Global analyser solutions


Specifications

standardised method: astM D7423, uoP 960


astM D7754 (for samples containing 1-15% v/v ethanol)
application: Analysis of oxygenated components (alcohols, ethers, ketones) in liquids, liquefied gases and gases
Configuration: single channel instrument based on thermo trace 1300 GC series using FiD detection.
Injection: SSL (Liquid Injection), GSV (Gas Sampling Valve) or LSV (Liquid Sampling Valve);
all three injection techniques can be combined
optional: Automated liquid sample injector
Pressure Facility for highly quantitative injection of liquefied gas samples using LSV
Mass Spectrometer for enhanced identification and sensitivity gain factor up to 177 (see page 3)
tubing: Sulfinert ® tubing for inert sample path
analytes: see table (page 2).
Calibration standards: 1, 10, and 100 ppm per individual component in n-Hexane
Dynamic range: 0.1-1000ppm (FiD)
sample requirements: Liquid, liquefied gas or gas sample
analysis time: 25 minutes
Minimum Detectability: 0.1 ppm for all individual components.
repeatability: 2% RSD (n=10)
Data systems Chromeleon, openlab

Figure 8. repeatability lowox


analysis (GC-FiD, 100pmm
calibration standard)

Figure 9. low level oxygenates analyser with optional


Pressure Facility for injection of Liquefied Gas
(using Liquid Sampling Valve)

M a De
in
Ho ll a nD
G. a .s is an
Gas is an
interscience company

Global analyser solutions WWW.Gassite.CoM

You might also like