Professional Documents
Culture Documents
Application Note
Energy & Chemicals - Petrochemicals
Author Abstract
Edgar Naegele, This Application Note shows the determination of aromatic hydrocarbons under
Agilent Technologies, Inc. normal phase conditions with the Agilent 1260 Infinity Binary LC system with RID
Waldbronn, Germany detection. The obtained results are comparable to the IP391(2000)/ASTM D6591
methods.
Nagesh Babu and
Sumitra Sankar
Gulf Bio Analytical,
United Arab Emirates, Dubai
2
LC method
RID
Peak width > 0.2 minutes (4 second response time) (2.28 Hz)
Temperature 35.00 °C
Column compartment
Initial valve position Port 1 & 6
Temperature 30.00 °C
Change valve position 7.48 minutes: Port 1 & 2
Sampler
Injection mode Standard injection
Injection volume 20.00 μL
Needle wash in vial with heptane
Binary pump
Flow 1.000 mL/min
Stop time 30.00 min
Solvent A Heptane, isocratic
3
Sample preparation RID, Refractive index signal
×10 5 Mono-aromatics
Place between 0.9 g to 1.1 g of sample, 4.197
to the nearest 0.001 g, into a 10-mL 6 Di-aromatics
5.183
volumetric flask and fill to the mark with
n-heptane. Shake thoroughly to mix, and 5
Non-aromatics
filter the sample to remove the particulate 3.460
matter. 4
nRIU
Results and Discussion 3
Area
Area
30
and DAH, and 0.1 to 4 mg/mL for TRI+ 10 3
20 3
aromatics (Figure 3). All calibration curves 5 10 2
for the aromatic hydrocarbons showed 0
1
2 Correlation: 0.99756
1
Correlation: 0.99987
0
linearity coefficients better than 0.9998. 0 20,000 40,000 0 10,000 20,000 30,000
To determine the limit-of-detection (LOD), Amount (µg/mL) Amount (µg/mL)
the signals and the noise were measured C D
for a 1/10 dilution of Standard 4. The ×10 5 Di-aromatics, RID1 A ×10 5 Tri+-aromatics, RID1 A
Area = 245.02653*Amt + 20896.64 Area = 333.880221*Amt + 16078.198
typical LOD for aromatic compounds 100 14
4 12 4
is 1 µg/mL (calculated for a signal-to- 80
10
noise ratio of 3). Statistical evaluation of 60 8 3
Area
Area
Table 1. Statistical evaluation of retention time and area for the aromatic compounds by multiple injection
(n = 5) of calibration standard level 3.
4
Figure 4 displays a RI chromatogram RID, Refractive index signal
×10 5
of a real sample. This sample shows Non-aromatics
approximately 4.3 mg/mL of tri+aromatic
compounds. 6
Mono-aromatics
Conclusion 5
nRIU
LC in combination with the Agilent 1260 3
Di-aromatics
Infinity RI Detector for the determination
2
mono-aromatic (MAH), di-aromatic (DAH),
and tri+-aromatic (TRI+) contaminants Tri-aromatics
1
in petroleum middle distillates and diesel
fuels comparable to IP 391(2000)/ASTM 0
D6591. The calibration of the RID showed
2 4 6 8 10 12 14 16
excellent linearity for MAH, DAH, and Time (min)
TRI+ aromatics. The retention time RSD is
typically below 0.06 % and the area RSD Figure 4. Real diesel oil sample showing aromatic compounds.
is typically below 0.16 %.
References
1. IP391(2000)/ASTM D6591–11,
Standard Test Method for
Determination of Aromatic
Hydrocarbon Types in Middle
Distillates—High Performance
Liquid Chromatography Method with
Refractive Index Detection.
2. http://www.astm.org/Standards/
D6591.htm
5
www.agilent.com/chem
This information is subject to change without notice.