You are on page 1of 6

Determination of Aromatic

Hydrocarbons in Petroleum Middle


Distillates with the Agilent 1260
Infinity Binary HPLC System with
RID Detection According to
IP391(2000)/ASTM D6591

Application Note
Energy & Chemicals - Petrochemicals

Author Abstract
Edgar Naegele, This Application Note shows the determination of aromatic hydrocarbons under
Agilent Technologies, Inc. normal phase conditions with the Agilent 1260 Infinity Binary LC system with RID
Waldbronn, Germany detection. The obtained results are comparable to the IP391(2000)/ASTM D6591
methods.
Nagesh Babu and
Sumitra Sankar
Gulf Bio Analytical,
United Arab Emirates, Dubai

Verified for Agilent


1260 Infinityty II LC
Introduction Experimental Columns
The IP391(2000)/ASTM D6591 • Agilent ZORBAX NH2,
Instruments
methods cover the determination of 4.6 × 250 mm, 5 µm
mono‑aromatic (MAH), di-aromatic Agilent 1260 Infinity LC System (p/n 880952-708)
(DAH), and poly-aromatic hydrocarbon • Agilent 1260 Infinity Binary Pump
(TRI+) contaminants in petroleum middle • Agilent ZORBAX NH2, Guard
(G1312B) with Agilent Infinity
distillates and diesel fuels boiling in the Cartridges, 4.6 × 12.5 mm, 5 µm
range from 150 °C to 400 °C, for example Standard Degasser (G1322A)
(p/n 820950-908)
gas oil1.
• Agilent 1260 Infinity Standard
The aromatic hydrocarbon content of Autosampler (G1329B) with • Agilent ZORBAX High Performance
motor diesel fuel is a factor that can Agilent 1290 Infinity Thermostat Guard Fittings Kit (p/n 820888-901)
affect exhaust emissions and fuel (G1330B)
combustion characteristics, as measured Reagents
by the cetane number2.
• Agilent 1260 Infinity Mobile Phase: Heptane, HPLC Grade
Using this instrumental setup, it is Thermostatted Column
possible to determine mono and Compartment (G1316A) with Standards
di‑aromatic hydrocarbons according Agilent 1200 Infinity Series • System Performance Standard
to IP436/ASTM D63793. It is also Quick‑Change 2-postition/6-port (SPS) - ASTM‑P‑0135: cyclohexane
possible to determine the total aromatic valve (G 4231A). (10 mg/mL), o-xylene (5 mg/mL,
hydrocarbons and total saturates
according to ASTM D74194. MAH), dibenzothiophene (5 mg/
• Agilent 1260 Infinity Refractive mL, DAH), 9-methylanthracene (5
This Application Note shows the Index Detector (G1362A) mg/mL, TRI+).
possibility to determine aromatic
hydrocarbons under normal phase Software • Calibration Standard:
conditions with RID detection by Agilent OpenLAB CDS ChemStation ASTM-P-0136-SET (Table 1)
using the Agilent 1260 Infinity binary Edition for LC & LC/MS Systems, Rev.
LC system. The obtained results are C.01.04
comparable to the IP391(2000)/ASTM
D6591 methods.

Table 1. Concentration of MAH, DAH, and TRI+ in ASTM calibration standards.

Std. 1 Std. 2 Std. 3 Std. 4


(mg/mL) (mg/mL) (mg/mL) (mg/mL)
Cyclohexane 50 20 5 1
o-Xylene (MAH) 40 10 2.5 0.5
Debenzothiophene (DAH) 40 10 2.5 0.2
9-Methylantracene (TRI+) 4 2 0.5 0.1

2
LC method
RID
Peak width > 0.2 minutes (4 second response time) (2.28 Hz)
Temperature 35.00 °C
Column compartment
Initial valve position Port 1 & 6
Temperature 30.00 °C
Change valve position 7.48 minutes: Port 1 & 2
Sampler
Injection mode Standard injection
Injection volume 20.00 μL
Needle wash in vial with heptane
Binary pump
Flow 1.000 mL/min
Stop time 30.00 min
Solvent A Heptane, isocratic

Valve operation (Figure 1) A


In the OFF position, the sample is injected Pump NH2-column
into the column through the autosampler.
6
Once the DAH is eluted from the column,
the valve is switched to the ON position, Autosampler
1 5
which reverses the flow in the column,
backflushing the TRI+ aromatics to the
RID detector. 2 4
Detector
The time B (seconds) for switching to B 3
backflush is calculated based on an
Pump NH2-column
injection of the System Performance
Standard (SPS) without valve switching 6
using the following equation:
Autosampler
1 5
B = tA + 0.4 (tB – tA)

Where: tA is the retention time of 2 4


Detector
dibenzothiophene (DAH), in seconds. tB is
the retention time of 9-methylanthracene 3
(TRI+), in seconds.
Figure 1. Valve configuration for the IP 391 method. A) OFF Position 1 & 6 is the loading position. B) After
switching to ON position 1 & 2 the compounds are eluted and analyzed in backflush.

3
Sample preparation RID, Refractive index signal
×10 5 Mono-aromatics
Place between 0.9 g to 1.1 g of sample, 4.197
to the nearest 0.001 g, into a 10-mL 6 Di-aromatics
5.183
volumetric flask and fill to the mark with
n-heptane. Shake thoroughly to mix, and 5
Non-aromatics
filter the sample to remove the particulate 3.460
matter. 4

nRIU
Results and Discussion 3

With the final method, all four calibration 2


standards were measured (Figure 2). The
measurement showed a good separation 1
tri+-aromatics
15.158
of the inherent compounds, a clear
separation of the aromatic compounds 0
from the aliphatic cyclohexane, and a 0 2.5 5.0 7.5 10.0 12.5 15.0 17.5
separation of the poly nuclear aromatic Time (min)
hydrocarbons to a much higher retention Figure 2. Overlay of IP391 Calibration standard 1–4 (ASTM-P-0136-SET), cyclohexane (non-aromatics),
time. This is necessary for their o-xylene (MAH), dibenzothiophene (DAH), 9-methylanthracene (TRI+).
determination in real samples.

For the non-aromatic as well as aromatic A B


compounds (MAH, DAH, TRI+), individual ×10 5
Non-aromatics, RID1 A
×10 5 Mono-aromatics, RID1 A
calibration curves were created. 25 60 Area = 140.857805*Amt + 26032.465
Area = 47.8448872*Amt − 24597.273 4 4
50
They ranged from 1 to 50 mg/mL for 20
40
cyclohexane, 0.5 to 40 mg/mL for MAH 15

Area
Area

30
and DAH, and 0.1 to 4 mg/mL for TRI+ 10 3
20 3
aromatics (Figure 3). All calibration curves 5 10 2
for the aromatic hydrocarbons showed 0
1
2 Correlation: 0.99756
1
Correlation: 0.99987
0
linearity coefficients better than 0.9998. 0 20,000 40,000 0 10,000 20,000 30,000
To determine the limit-of-detection (LOD), Amount (µg/mL) Amount (µg/mL)
the signals and the noise were measured C D
for a 1/10 dilution of Standard 4. The ×10 5 Di-aromatics, RID1 A ×10 5 Tri+-aromatics, RID1 A
Area = 245.02653*Amt + 20896.64 Area = 333.880221*Amt + 16078.198
typical LOD for aromatic compounds 100 14
4 12 4
is 1 µg/mL (calculated for a signal-to- 80
10
noise ratio of 3). Statistical evaluation of 60 8 3
Area

Area

retention time and area for the aromatic 40 6


3
compounds by multiple injection of 20
4
2
2 2 1
calibration standard level 3 showed RSD 0 Correlation: 0.99997 0 Correlation: 0.99980
values typically below 0.06 % and 0.16 %, 0 10,000 20,000 30,000 0 1,000 2,000 3,000
respectively (Table 1). Amount (µg/mL) Amount (µg/mL)

Figure 3. Calibration curves for the four compounds tested.

Table 1. Statistical evaluation of retention time and area for the aromatic compounds by multiple injection
(n = 5) of calibration standard level 3.

Mono-aromatics Di-aromatics Tri+-aromatics


RT (min) 4.27 5.36 15.15
RT RSD (%) 0.06 0.06 0.02
Area RSD (%) 0.12 0.16 0.15
LOD (µg/mL) 1 0.7 1

4
Figure 4 displays a RI chromatogram RID, Refractive index signal
×10 5
of a real sample. This sample shows Non-aromatics
approximately 4.3 mg/mL of tri+aromatic
compounds. 6
Mono-aromatics
Conclusion 5

This Application Note demonstrates the 4


use of the Agilent 1260 Infinity Binary

nRIU
LC in combination with the Agilent 1260 3
Di-aromatics
Infinity RI Detector for the determination
2
mono-aromatic (MAH), di-aromatic (DAH),
and tri+-aromatic (TRI+) contaminants Tri-aromatics
1
in petroleum middle distillates and diesel
fuels comparable to IP 391(2000)/ASTM 0
D6591. The calibration of the RID showed
2 4 6 8 10 12 14 16
excellent linearity for MAH, DAH, and Time (min)
TRI+ aromatics. The retention time RSD is
typically below 0.06 % and the area RSD Figure 4. Real diesel oil sample showing aromatic compounds.
is typically below 0.16 %.

References
1. IP391(2000)/ASTM D6591–11,
Standard Test Method for
Determination of Aromatic
Hydrocarbon Types in Middle
Distillates—High Performance
Liquid Chromatography Method with
Refractive Index Detection.

2. http://www.astm.org/Standards/
D6591.htm

3. Agilent Application Note, Publication


number 5991-3172EN

4. Agilent Application note, Publication


number 5991-3171EN

5
www.agilent.com/chem
This information is subject to change without notice.

© Agilent Technologies, Inc., 2013-2017


Published in the USA, July 1, 2017
5991-3170EN

You might also like