Application Note
Energy & Chemicals - Petrochemicals
Authors Abstract
Edgar Naegele This Application Note shows that the Agilent 1260 Infinity Binary LC System is
Agilent Technologies Inc. able to determine aromatic hydrocarbons in jet fuel under normal phase conditions
Waldbronn, Germany with RI detection, and complies with the IP436/ASTM D6379 method.
LC method
Agilent 1260 Infinity Refractive Index Detector
Peak width > 0.2 minutes (4-second response time) (2.28 Hz)
Temperature 35.00 °C
Agilent 1260 Infinity Column Compartment
Valve position Port 1 & 6
Temperature 35.00 °C
Agilent 1260 Infinity Standard Autosampler
Injection volume 10.00 μL
Temperature 10 °C
Agilent 1260 Infinity Binary Pump
Flow 1.000 mL/min
Stop time 15.00 minutes
Solvent A Heptane
2
Valve operation Thermostatted Column Compartment
In the 1 & 6 position, the sample is
Pump NH2 column
injected onto the column through the
autosampler. The compounds o-xylene
(mono-aromatic hydrocarbons, MAH) Autosampler 1
and 1-methyl naphthalene (di-aromatic 6
hydrocarbons, DAH) are separated and
detected. The valve can be used for the
2
determination of MAH, DAB, and TRI+
(tri‑aromatic hydrocarbons) according to 5
IP391/ASTM D65913.
3
Sample preparation 4
0 2 4 6 8 10 12 14 min
Figure 3. Standard 4, IP436/ASTM D6379, o-xylene (MAH, 0.1 mg/mL), 1-mehtyl naphtalene
(DAH, 0.05 mg/mL).
3
The separation of the lowest level A
calibration standard containing Area MAH, RID1 A
0.1 mg/mL o-xylene (MAH) and Area = 724,878.245*Amt + 4,226.7994
4
1,000,000
0.05 mg/mL 1-mehtyl naphtalene (DAH)
showed their clear detection by RID at 800,000
the lowest concentration level (Figure 3).
600,000
The calibration curves of o-xylene (MAH)
3
and 1-methyl naphtalene (DAH) by RID 400,000
detection demonstrated good linearity 200,000 Correlation: 0.99990
for quantification (Figure 4). For o-xylene, 2
1
the linearity coefficient was 0.99990, 0
and for 1-methyl naphtalene the linearity 0 0.5 1 Amount (%)
coefficient was 0.99997. B
Area DAH, RID1 A
The precision performance of the system 700,000
Area = 1,335,789.36*Amt + 1,302.8679
was shown for a comparable application 4
600,000
including a valve, switching the column 500,000
to backflush, to separate the analyte
400,000
from the matrix. Typically, the retention
300,000
time RSD was below 0.1 % and the
area RSD was below 0.2 %. The typical 200,000 3
Correlation: 0.99997
limit-of-detection (LOD) for aromatic 100,000
12
hydrocarbons under the used method 0
is below 1 µg/mL (calculated for a 0 0.2 0.4 Amount (%)
signal‑to‑noise ratio of 3)2.
Figure 4. Calibration curves for A) MAH, o-xylene, and B) DAH, 1-methyl naphthalene by RID detection.
Conclusion
This Application Note demonstrates the
use of the Agilent 1260 Infinity Binary LC References 3. Determination of Total Aromatic
Hydrocarbons and Total Saturates with
System in combination with the Agilent 1. American Society for Testing
the Agilent 1260 Infinity Binary LC
1260 Infinity Refractive Index Detector and Materials (ASTM): ASTM
System with RI Detection According
for the determination of MAH and DAH D6379 – 11, Standard Test Method
to ASTM D7419, Agilent Technologies
contaminants in aviation kerosene and for Determination of Aromatic
Application Note, publication number
petroleum distillates boiling in the range Hydrocarbon Types in Aviation Fuels
5991-3171EN, 2014.
of 50 to 300 °C, according to IP436/ASTM and Petroleum Distillates—High
D6379. The calibration of the RID showed Performance Liquid Chromatography
excellent linearity for the range of 0.01 % Method with Refractive Index
to 1.5 % MAH and 0.005 % to 0.5 % DAH. Detection, http://www.astm.org/
Standards/D6379.htm
2. Determination of Aromatic
www.agilent.com/chem
Hydrocarbons in Petroleum Middle
Distillates with the Agilent 1260 This information is subject to change without notice.
Infinity Binary HPLC System with RID
Detection According to IP391(2000)/ © Agilent Technologies, Inc., 2014-2016
Published in the USA, September 1, 2016
ASTM D659, Agilent Technologies 5991-3172EN
Application Note, publication number
5991‑3170EN, 2013.