Professional Documents
Culture Documents
Note: 40899
Petrochemical Series - Analysis of trace elements
in naphtha using the iCAP 6000 Series ICP
Matthew Cassap, ICP Applications Specialist, Thermo Fisher Scientific, Scientific Instruments, Cambridge, UK
Instrumentation
The Thermo Scientific iCAP 6500 Radial ICP (Figure 1) was
used for this analysis. The radial was chosen because the
interferences from carbon based emissions can be reduced
by optimizing the radial viewing height. The IsoMist (below
in Figure 1) is a Peltier cooled spray chamber which was
used in conjunction with a glass concentric nebulizer for this
analysis.
Figure 2: The plasma aspirating naphtha after the auxiliary and nebulizer gas
flows have been optimized.
The instrument was then calibrated and the spiked
sample analyzed. A detection limit study was carried out
by analyzing ten replicates of the calibration blank and
multiplying the standard deviation of this analysis by 3. The
subarray plots for each of the wavelengths were examined
to ensure freedom from interference and to confirm that
the background correction points were set appropriately.
Figure 1: The iCAP 6500 Radial ICP with the GE IsoMist peltier cooled spray.
Parameter Setting Results In addition to these
Tubing Orange/white solvent flex offices, Thermo Fisher
Naphtha RSD on Instrument
RF Power 1150 W spike at three replicates Detection Scientific maintains
Nebulizer Gas Flow 0.7 L/min 2.4mg/kg of the spike Limit a network of represen-
Auxiliary Gas Flow 1.5 L/min mg/kg % µg/kg tative organizations
Coolant Gas Flow 12 L/min Ag 338.389 nm 2.39 0.558 0.8 throughout the world.
Radial Viewing Height 8 mm Al 308.215 nm 2.37 0.188 2.6
Nebulizer Glass concentric B 208.595 nm 2.17 1.207 14
Spray chamber (IsoMist Kit) Ba 223.527 nm 2.35 0.637 1.3
Ca 184.006 nm 2.35 1.426 8.1
Table 1: Instrument conditions
Cd 214.438 nm 2.38 0.970 1
Africa
Cr 267.716 nm 2.37 0.246 0.8 +43 1 333 5034 127
Cu 324.754 nm 2.41 0.207 0.4 Australia
Fe 238.204 nm 2.39 0.546 1.8 +61 2 8844 9500
Mg 279.553 nm 2.40 0.087 0.08 Austria
+43 1 333 50340
Mn 293.930 nm 2.38 0.227 0.8
Belgium
Mo 281.615 nm 2.38 0.233 2.2 +32 2 482 30 30
Na 589.592 nm 2.41 0.546 13 Canada
+1 800 530 8447
Ni 231.604 nm 2.37 0.751 5.0
China
P 178.284 nm 2.40 0.442 15 +86 10 5850 3588
Pb 220.353 nm 2.35 0.391 16 Denmark
Si 212.412 nm 2.40 0.467 10.5 +45 70 23 62 60
Sn 283.999 nm 2.37 0.456 20 Europe-Other
+43 1 333 5034 127
Ti 334.941 nm 2.37 0.153 0.4 France
V 309.311 nm 2.37 0.387 0.9 +33 1 60 92 48 00
Zn 213.856 nm 2.39 0.235 0.8 Germany
+49 6103 408 1014
India
Figure 3: The calibration plot of P 178.284 nm The results for the spiked naphtha are all within acceptable +91 22 6742 9434
limits of the spiked value (± 5 %) with the exception of Italy
+39 02 950 591
boron which is within 10 % of the spiked value. A stabilizer Japan
is often added to organic solvents when analyzing for boron. +81 45 453 9100
However, in this case a stabilizer was not used and this Latin America
+1 608 276 5659
could explain the slightly low recovery and the instability
Middle East
of this element, which is also reflected in the high %RSD. +43 1 333 5034 127
All of the %RSDs are below 2 % with the vast majority Netherlands
below 0.5 %. Figure 3 gives a calibration plot for +31 76 579 55 55
South Africa
phosphorus, indicating excellent precision and linearity at +27 11 570 1840
relatively low concentrations. Figure 4 shows the subarray Spain
window for vanadium at 309.311 nm, indicating the peak +34 914 845 965
center and background points. This window makes Sweden / Norway /
Finland
method development extremely simple. +46 8 556 468 00
Switzerland
Conclusion +41 61 48784 00
Figure 4: The sub-array plot of V 309.311 nm showing the calibration blank UK
The analysis of naphtha on the iCAP 6500 Radial ICP is +44 1442 233555
and standards.
made simple by the addition of a Peltier cooled sprayed USA
chamber set to -5 ºC. This reduces the volatility of the +1 800 532 4752
solvent, which in turn lowers the plasma loading when the www.thermo.com
solvent is aspirated. The iCAP 6000 Series is capable of
detecting single figure sub ppb concentrations of various
elements within this complex and challenging matrix.
• Duo IsoMist – 8423 120 51641
• Radial IsoMist – 8423 120 51651 FM 09032
Thermo Electron Manufacturing
©2008 Thermo Fisher Scientific Inc. All rights reserved. IsoMist is a trade mark of Glass Expansion, Batman Street, Melbourne, Australia. All other trademarks are the property of Thermo Fisher Ltd (Cambridge) is ISO Certified.
Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details.
AN40899_E 08/08C