You are on page 1of 8

Fracture Resistance of Various Thickness e.

max CAD
Lithium Disilicate Crowns Cemented on Different
Supporting Substrates: An In Vitro Study
Sara Elizabeth Chen, DMD ,1 Albert Christopher Park, DDS,1 Jingxu Wang, BDS, PhD,2
Kent L. Knoernschild, DMD, MS, FACP,1 Stephen Campbell, DMD, MMSc, FACP,1 &
Bin Yang, DMD, MSc, PhD, FACP1
1
University of Illinois at Chicago – Restorative Dentistry, Chicago, IL, United States
2
Department of Stomatology, The First Affiliated Hospital of Guangzhou Medical University, Guangzhou, China

Keywords ABSTRACT
Monolithic; occlusal thickness; load-bearing
capacity; preparation design; all ceramic
Purpose: To investigate the influence of abutment material properties on the fracture
crown; ultra-thin crown. resistance and failure mode of lithium disilicate (IPS e.max) CAD/CAM (computer-
aided design/manufacturing) crowns on traditionally and minimally prepared simu-
Correspondence lated tooth substrates.
Bin Yang, DMD, MSc, PhD, FACP, University Materials and Methods: Thirty lithium disilicate (IPS e.max) CAD/CAM crowns
of Illinois at Chicago – Restorative Dentistry, were divided into three groups (n = 10): TD: traditional thickness crowns ce-
Chicago, IL, USA. mented on Paradigm MZ100 abutments; MD: minimal thickness crowns cemented
E-mail: yangbin@uic.edu on Paradigm MZ100 abutments; ME: minimal thickness crowns cemented on e.max
abutments. The 3Shape system was used to scan, design and mill all abutments and
There are no funders to report for this crowns with a die space set to 40 µm. Traditional thickness crowns were designed
submission. based on manufacturer guidelines with 1.5 mm occlusal thickness and 1.0 mm mar-
The authors deny any conflicts of interest in
gins. Minimal thickness crowns were designed with 0.7 mm occlusal thickness and
regards to the current study.
0.5 mm margins. MZ100 composite and e.max abutments were selected to simulate
dentin and enamel substrates, respectively, based on their elastic-modulus. Variolink
Accepted August 7, 2019 Esthetic was used to cement all samples following manufacturer’s instructions. A
universal testing machine was used to load all specimens to fracture with a 3 mm
doi: 10.1111/jopr.13108 radius stainless steel hemispherical tip at a crosshead speed 0.5 mm/minute along the
longitudinal axis of the abutment with a 1 mm thermoplastic film placed between
the loading tip and crown surface. Data was analyzed using ANOVA and Bonferroni
post hoc assessment. Fractographic analysis was performed with scanning electron
microscopy (SEM).
Results: The mean fracture load (standard deviation) was 1499 (241) N for TD; 1228
(287) N for MD; and 1377 (96) N for ME. Statistically significant difference between
groups did not exist (p = 0.157, F = 1.995). In groups TD and MD with low e-modulus
abutments, the dispersion of a probability distribution (coefficient of variation: CV)
was statistically higher than that of group ME with high e-modulus abutments. SEM
illustrated larger micro-fracture dimensions in Group MD than Group ME.
Conclusion: Minimal thickness e.max crowns did not demonstrate statistical differ-
ence in fracture resistance from traditional thickness crowns. Fracture mechanisms of
minimal thickness e.max crowns may be affected by the e-modulus of the substrate.
Minimal thickness e.max crowns may be a viable restorative option when supported
by high e-modulus materials.

esthetic rehabilitation for young patients that do not warrant


Introduction excessive removal of sound tooth structure. Additional tooth
reduction to create restorative space may not be appropriate
Minimal thickness restorations offer a treatment modality for
following a substantial amount of pathological tooth loss if
patients with limited restorative space, advanced wear, and
progressive tooth eruption has compensated for wear/erosion.1

Journal of Prosthodontics 0 (2019) 1–8 


C 2019 by the American College of Prosthodontists 1
Fracture Resistance of Various Thickness e.max Crowns Chen et al

Minimal reduction to avoid pulp exposures and the use of ultra-


thin restorations may be the best treatment alternative.2-4 Previ-
ous studies have shown minimum thickness all ceramic crowns
may have fracture strengths suitable for use as a single posterior
restoration.1,4-6
Therapy that includes minimally invasive crown preparations
and lithium disilicate crowns could lead to predictable care
and improved survival. Traditional crown preparation requires
removal of 1.5 to 2 mm of tooth structure, where all enamel
is typically removed leaving the restoration to be supported by
dentin. Minimally invasive crown preparations require much
less tooth reduction and provide the potential for the restorations
to be supported by enamel. Theoretical analysis7 suggests that
smaller mismatch in elastic modulus between the abutment
and the restoration, can significantly increase the load-bearing
capacity of monolithic lithium disilicate crowns. Therefore,
conservative preparation not only could preserve tooth structure
but also provide a desirable enamel surface for bonding and
lithium disilicate crown support.
Clinically relevant, controlled in vitro or in vivo studies that
assess the influence of preparation design and supporting tooth
structure on lithium disilicate crown fracture strength are not
available. Laboratory studies indicate the elastic modulus of
Paradigm MZ1008,9 composite (3M, St. Paul, MN) is similar
to dentin,10 whereas, the elastic modulus of e.max11 (Ivoclar
Vivadent Inc., Amherst, NY) is similar to enamel.12,13 Sim-
ulation of support from dentin or enamel, respectively, could
be accomplished with such materials in a controlled study to
determine the influence of preparation design and underlying
support on fracture resistance.
Figure 1 A, Study design. B, 3 Shape design of traditional thickness
The aim of this in vitro study is to investigate the influence
crown (left) and minimal thickness crown (right). C, Milled traditional
of abutment material properties on the fracture resistance and
thickness MZ100 abutment (left), minimal thickness MZ100 abutment
failure mode of e.max CAD/CAM crowns on traditionally and
(middle), minimal thickness e.max abutment (right). D, Cross section
minimally prepared simulated tooth substrates. The null hy- of Group TD before cementation (left), Group MD before cementation
potheses were: (1) restoration thickness has no influence on the (middle), Group ME after cementation (right). Bite-wing radiographs of
fracture strength of CAD lithium disilicate crowns, and (2) frac- specimens were taken before and after cementation.
ture mechanisms of minimal thickness lithium disilicate CAD
crowns are not affected by the elastic modulus of the substrate.
Abutment fabrication

Material and methods A mandibular first molar typodont tooth was prepared with a
traditional thickness design using a diamond bur to include
Thirty lithium disilicate (IPS e.max CAD MT, Ivoclar Vi- rounded, smooth and flat occlusal anatomy with 1 mm wide
vadent Inc.) CAD/CAM (computer-aided design/computer- rounded shoulder margins. The traditionally prepared typodont
aided manufacturing) crowns were divided into three groups tooth was used to create the minimal thickness abutment, to
(n = 10). Group TD: traditional thickness crowns cemented on maintain the same rounded internal line angles, taper, and oc-
Paradigm MZ100 (3M, St. Paul, MN) abutments. Group MD: clusal geometry; while only changing the width of the finish
minimal thickness crowns cemented on MZ100 abutments. line. The original typodont was modified by circumferentially
Group ME: minimal thickness crowns cemented on e.max CAD trimming 0.5 mm from the external edge of the finish line to
abutments (Fig 1a). reduce the finish line width from 1 to 0.5 mm. The axial walls,
The 3Shape CAD system (3Shape, Copenhagen, Denmark) axial-cervical line angle and coronal preparation design were
was used to scan, design and mill all abutments and crowns unaltered.
with the die space set to 40 and 30 µm extra occlusal die The pre-modified traditionally prepared typodont tooth and
space. This allowed all abutments and restorations to have the minimal thickness preparation were scanned with a labora-
standardized dimensions and geometries. Traditional thickness tory scanner (3 Shape D700, Copenhagen, Denmark) and STL
crowns were designed based on manufacturer guidelines with a (Standard Tessellation Language) files of the abutments were
1.5 mm uniform occlusal thickness, and 1.0 mm rounded shoul- created. Ten traditional thickness abutments were milled in de-
der margins. Minimal thickness crowns were designed with a tail mode with a PlanMill40 milling unit (Planmeca, Helsinki,
0.7 mm uniform occlusal thickness and 0.5 mm rounded shoul- Finland) using MZ100 blocks for E4D. Twenty minimal thick-
der margins (Fig 1b). ness abutments were milled: 10 with MZ100 blocks and 10

2 Journal of Prosthodontics 0 (2019) 1–8 


C 2019 by the American College of Prosthodontists
Chen et al Fracture Resistance of Various Thickness e.max Crowns

with IPS e.max CAD blocks (Fig 1c). All milled lithium dis- Measurement of fracture load
ilicate abutments were sintered in an oven (Programat CS2;
Specimen holders were created by cross sectioning a 1-inch
Ivoclar Vivadent, Inc.) following the manufacturer’s sintering
diameter PVC pipe to 10 mm tall pieces. The specimens were
guidelines.
mounted in the center of the PVC pipe with self-curing or-
thodontic resin. A putty matrix was used to secure all specimens
in a reproducible position and to ensure the top of the crowns
Crown fabrication were leveled. The abutments were embedded in orthodontic
3Shape software was used to scan a milled MZ100 traditional resin 1 mm below the margin of the crown.
thickness abutment and a traditional thickness crown was de- The fracture load of the specimens was measured using a
signed on the abutment. The resulting traditional thickness universal testing machine (Instron 5582; Instron, Co, Norwood,
crown STL file was used to mill 10 traditional thickness crowns MA). A 3 mm radius stainless steel hemispherical tip was ap-
with IPS e.max CAD MT blocks. A minimal thickness MZ100 plied to the center of each crown. A 1 mm thermoplastic film
abutment was scanned and a minimal thickness crown was was placed between the loading tip and crown surface to dis-
designed. The resulting STL file was used to mill 20 mini- tribute forces and ensure a broad even contact.16 The force
mal thickness crowns with IPS e.max CAD MT blocks. All was loaded along the longitudinal axis at a crosshead speed of
crowns were milled in detail mode with the sprue located on 0.5 mm/min until audible fracture. The loading values versus
the axial surface 2 mm away from the margins. Restoration crosshead position were plotted continuously during loading
thicknesses were designed and milled as specified, 0.7 mm and a slight drop in load on the graph in conjunction with the
for minimal thickness restorations and 1.5 mm for traditional audible fracture noise defined the fracture. All crowns were
crowns. The occlusal surfaces were polished with a porcelain then inspected for visible fracture lines on the restoration sur-
polishing kit (Dialite Extra-Oral Porcelain Polishing, Brasseler, face with and without the use of trans-illumination. Fractures
Savannah, GA). All milled blue stage lithium disilicate crowns were categorized as catastrophic failure, fracture lines visible
were sintered following the manufacturer’s sintering guidelines. only with transillumination, and no fracture lines visible.
Thicknesses were verified post milling using an Iwanson spring
caliper. Data analysis
Data was analyzed with statistical software (IBM SPSS version
Cementation 22, Armonk, NY) using a one-way analysis of variance and post
hoc Bonferroni correction. The coefficient of variation (CV) for
Before cementation, all MZ100 composite abutments were sub- each group was calculated. A CV of 15% or greater was defined
merged in water for 7 days.14 A Variolink Esthetic DC System as a critical measure of dispersion of a probability distribution.
Kit (Ivoclar Vivadent, Inc, Albany, NY) was used to cement
all specimens following manufacturer’s instructions. All e.max
Fractographic analysis
surfaces (crowns and e.max minimal abutments) were steam
cleaned and dried. All e.max surfaces were then etched with In order to prepare specimens for fractographic analysis, one
5% hydrofluoric acid (IPS Ceramic etching gel) for 20 seconds. specimen from group ME and group MD were cross-sectioned
The 5% hydrofluoric acid was rinsed with water for 15 seconds to a 1 mm thin slice along the long axis of the crown with an
and dried with oil-free air. Ceramic primer (Monobond Plus) IsoMet 1000 precision wafering diamond saw (Buehler, Lake
was applied for 60 seconds. Once the pre-treatment for all e.max Bluff, IL) at 250 RPM. The cross section was taken from the
abutments and crowns was finished, the MZ100 abutments were center of the specimen where the load was applied. The speci-
prepared for cementation. The MZ100 abutments were steam mens were mounted on aluminum stubs with carbon adhesive
cleaned, air dried, and then etched with 35% phosphoric acid for tabs and sputter coated with 5.0 nm of gold/palladium at a low-
30 seconds. The 35% phosphoric acid etch was rinsed with wa- pressure argon atmosphere. Morphology, origin and magnitude
ter for 15 seconds and dried with oil free air. The MZ100 abut- of fractures were analyzed using a variable pressure scanning
ments were scrubbed with a bonding agent (Adhese Universal electron microscope (Hitachi S-3000N, Krefeld, Germany).
Vivapen) for 20 seconds and oil/moisture free compressed air
was used to disperse the bond evenly on the MZ100 abutment. Results
A thin layer of resin luting cement (Variolink Esthetic) was
applied simultaneously to the occlusal surface of the abutment The mean fracture load (standard deviation) was 1499 (241) N
with a microbrush and the intaglio surface of the crown. The for Group TD; 1228 (287) N for Group MD; and 1377 (96)
crown was firmly seated onto the abutment and held with finger N for Group ME (Fig 2). No statistically significant difference
pressure for 1 min to allow chemical cure. Excess cement was between groups was observed (p = 0.157, F = 1.995). Abut-
removed with a microbrush and the margins were covered with ments with high elastic modulus (Group ME) showed slightly
an air-blocking barrier (Ivoclar Liquid Strip). A curing light higher fracture resistance with a narrower standard deviation
with 500 to 1000 mW/cm2 was applied to crown for 30 seconds compared to abutments with low elastic modulus (Group TD
on each surface. The air-blocking barrier was rinsed with water and MD). The coefficient of variation (CV) was 16.6% for
and the cemented specimens were left at room temperature group TD and 23.3% for group MD. The CV’s of these two
for 30 minutes, and then stored in distilled water at ambient groups were greater than 15% (critical measure of dispersion
temperature for 1 week before fracture testing.15 of a probability distribution). In contrast the CV of group ME

Journal of Prosthodontics 0 (2019) 1–8 


C 2019 by the American College of Prosthodontists 3
Fracture Resistance of Various Thickness e.max Crowns Chen et al

Figure 2 Mean fracture load. * indicates no statistical difference (p >


0.05) among the three groups. Coefficient of variation for group ME
(6.9%) is significantly lower than the coefficient of variation for groups
TD (16.6%) and MD (23.3%).

was 6.9%. The dispersion of a probability distribution in groups


TD and MD was statistically higher than that of group ME.
Catastrophic failures were defined in this study as specimens
with clearly visible fracture lines at the surface of the restora-
tion. Only four specimens were defined as catastrophic failures
after loading the specimens to audible fracture sounds with a
deviated loading graph. All catastrophic failures still remained
intact even when visible occlusal surface fracture lines were
observed (Fig 3). Catastrophic failures were from Group TD
(3 specimens) and Group MD (1 specimen). No specimens had
visible fracture lines on the restoration surface in Group ME.
When further trans-illumination inspection was conducted, ad-
ditional fracture lines were identified. Group TD had 5 speci-
mens with very small fractures (1-2 mm in length) within the
occlusal portion of the restoration and Group ME had 2 spec-
imens with tiny fracture lines (1 mm in length) within the
axial walls of the restorations. Overall group MD had the most
specimens with visible fractures but group TD had the most
number of catastrophic failures (Table 1).
Different fracture modes were demonstrated through SEM
observation. Figure 4 illustrates the fracture mode from Group
MD and ME. Fracture origin was visualized at the interface

Table 1 Fracture categories

Group TD Group MD Group ME


Fracture category (n = 10) (n = 10) (n = 10)

Catastrophic Failures 3 1 0
Fracture lines visible only 0 5 2
with transillumination
No visible fracture lines with 7 4 8
Figure 3 Fracture categories. A, Catastrophic fracture from Group TD.
audible fracture sound
B, Fracture lines visible only using transillumination from Group MD.
Fractures were categorized as catastrophic failure, fracture lines visible only C, Sample from Group ME that displayed no visible fracture lines with
with transillumination, or no fracture lines visible at all. Catastrophic failures and without transillumination. These fractures were identified by ini-
were defined as fracture lines clearly visible at the surface of the restoration tial audible fracture sound upon loading and verified by dip in force vs
without the need to use transillumination. crosshead plotting graphs.

4 Journal of Prosthodontics 0 (2019) 1–8 


C 2019 by the American College of Prosthodontists
Chen et al Fracture Resistance of Various Thickness e.max Crowns

A between the crown and luting resin cement in both groups.


The fracture defects in group ME were smaller in dimensions
than in group MD. One specimen in group TD was loaded
beyond initial audible crack to complete catastrophic failure
in order to examine with SEM analysis. Figure 5 illustrates
the remaining piece of traditional thickness crown bonded to
the abutment surface. The fracture lines (arrows labeled a, b,
c, d) were wider when close to the cementation interface than
B the fractures lines close to the restoration surface, indicating
fracture origin at or near the interface between the crown and
luting resin cement with propagation towards the restoration
occlusal surface. All groups showed similar fracture trends;
originating at the internal interface between the crown and
luting resin cement, and propagating towards the restoration
occlusal surface.

C
Discussion
This study demonstrated lithium disilicate crowns fabricated for
minimally reduced preparations had fracture resistance similar
to traditionally prepared lithium disilicate crowns when a sin-
gle load-to-failure is used. Such results justify further in vitro
investigation with fatigue loading in a simulated oral environ-
ment and clinical investigation with well-designed trials. The
D mean fracture load values in this study ranged from 1228 N to
1499 N. This suggests that all crowns even with minimal thick-
ness of <1 mm can have fracture loads high enough to withstand
the average maximum posterior biting force of 600 to 900 N
in healthy young adults.17-20 The in vitro findings cannot be
directly applied to clinical conditions since load-to-failure test-
ing provides information only at extreme conditions.6,16 Further
E in-depth investigation is indicated.
Reduced thickness restorations with sufficient fracture
strength may have a tremendous impact on clinical applications.
Minimal thickness all ceramic restorations may be warranted in
clinical situations of limited restorative space, excessive wear,
or desire to avoid pulp exposure. Monolithic lithium disilicate
restorations of various thicknesses can be bonded to dentin or
enamel substrates depending on the amount of tooth reduction.
F Failure due to tensile stresses is more sensitive to ratios of elas-
tic moduli between the restorative material and substrate, and
less attributed to the intrinsic material strength and thickness.15
In this study, 3M Paradigm MZ100 composite blocks were se-
lected as the first abutment material because it has an elastic
modulus of 18GPa,8,9 which is comparable to human dentin
with an elastic modulus of 16 to 18 GPa.10 Milling MZ100
composite abutments allowed for minimal defects and voids
in the composite compared to traditional methods of pack-
Figure 4 SEM Group MD and ME. Randomly selected samples from
able composites. Lithium disilicate was selected as the second
Group MD (minimal thickness crown on MZ100 abutment) and ME (min-
abutment material because it has an elastic modulus of 95
imal thickness crown on e.max abutment) are illustrated. Cross sections
Gpa11 which is similar to human enamel with an elastic mod-
from group MD (left column) and group ME (right column) were exam-
ined with SEM under low and high magnifications. The layer of Variolink
ulus of 91 GPa.12,13 In addition to using CAD/CAM tech-
Esthetic cement is labeled with “V” and the e.max crown is labeled
nology to control the material properties; the geometry, dimen-
with “Em.” No fracture lines were detected in the cement or abutment sions, internal fit, and die space of all abutments and crowns,
layers, but micro-fracture lines were found in e.max ceramic (labeled and loading to fracture in this study were standardized and
in dotted boxes in A,D). Fracture origin was visualized at the interface controlled.
between the crown and luting resin cement in both groups. Fracture The difference in fracture load and fracture mode between
defects in Group MD (B, C) were larger in dimension than defects in groups can be compared directly within the study. The mean
group ME (E, F). fracture load (standard deviation) was 1499 (241) N for Group

Journal of Prosthodontics 0 (2019) 1–8 


C 2019 by the American College of Prosthodontists 5
Fracture Resistance of Various Thickness e.max Crowns Chen et al

TD; 1228 (287) N for Group MD; and 1377 (96) N for Group
A
ME (Fig 2). There was no statistically significant difference
among the three groups. Therefore, the 1st null hypothesis,
that no statistically significant difference existed in the fracture
strength between the three groups was accepted.
Minimal thickness crowns luted to enamel may be more pre-
dictable and consistent than traditional thickness restorations or
minimal thickness restorations luted to dentin. The coefficient
of variation (CV) is a measure of dispersion of a probability dis-
tribution, and a CV less than 15% indicates potentially greater
predictability in results. The coefficients of variation of groups
TD (16.6%) and MD (23.3%) are greater than 15% while the CV
B of group ME (6.9%) is less than 15%. The CV of groups TD and
MD, and therefore the dispersion of a probability distribution
is statistically higher than that of group ME. The observations
from this study indicate a positive pattern suggesting an indica-
tion for lithium disilicate crowns to be supported by a material
with high elastic modulus (e.g., enamel) to achieve predictable
results.
The only group without any catastrophic fractures was group
ME - minimal thickness e.max crowns on e.max abutments
(enamel analog). On the other hand, group MD, minimal
thickness e.max crowns on MZ100 abutments (dentin analog),
demonstrated the most visible cracks at the restoration sur-
C face. Therefore, the 2nd null hypothesis, fracture mechanisms
of minimal thickness lithium disilicate CAD crowns is not af-
fected by the elastic modulus of the substrate is rejected.
All groups demonstrated similar fracture mechanisms from
SEM observation. In Figure 4, no fracture lines were detected
in the abutment materials, but microfracture lines were found
in the lithium disilicate ceramic, with origins at the cemen-
tation interface. The fracture lines in Figure 5 (labeled with
arrows a, b, c, and d) are wider when close to the cementation
interface than the fracture lines close to the surface, indicating
fracture origin at or near the interface between the crown and
luting resin cement with propagation towards the restoration
D
occlusal surface.21 All restorations regardless of the magnitude
of the fracture defects seem to have exhibited fracture origins
at the crown/cement interface. Interestingly, fracture defects
seen in Group MD were larger in magnitude than in Group
ME (Fig 4). Furthermore, group ME had a significantly lower
CV and was the only group without catastrophic fractures. This
further suggests that minimal thickness restorations may have
more predictable mechanical behavior and fewer catastrophic
failures when supported by a substrate with a higher elastic
modulus.
Previous studies2,4-6,22-24 have investigated fracture loads
of monolithic lithium disilicate CAD posterior restorations
Figure 5 SEM Group TD. A randomly selected traditional thickness with various abutment materials, loading techniques, ceramic
crown was purposely fractured beyond initial audible crack sound in thicknesses, and cementation methods. Although it is difficult
order to create a catastrophic failure to analyze under SEM at x25, to compare previous studies to the current study due to varying
x90, x300, x500 magnifications. Images illustrate the remaining piece
methodologies, most have reported fracture load ranges higher
of crown bonded to the abutment surface. The substrate is labeled with
than what was reported in the current study. This may be
“S,” the layer of Variolink Esthetic resin cement is labeled with “V,”
attributed to anatomical crown designs in some studies with
and the e.max crown is labeled with “Em”. The fracture lines (arrows
occlusal thicknesses of >1.5 mm in cusp areas. Also, some of
labeled a, b, c, d) are wider when close to the cementation interface than
the fracture lines close to the surface, indicating fracture origin (labeled
these studies were investigating fracture strength on titanium
with *) at or close to the interface between the crown and cement with
abutments, which have an elastic modulus much higher than
propagation towards the restoration occlusal surface. Dotted box in 5B MZ100 composite. Few studies25-27 have looked at ultra-thin
indicates area that is magnified in panels C and D. monolithic e.max crowns at such an extreme and uniform

6 Journal of Prosthodontics 0 (2019) 1–8 


C 2019 by the American College of Prosthodontists
Chen et al Fracture Resistance of Various Thickness e.max Crowns

occlusal thickness as this study, but the fracture loads reported American Dental Association for providing their Universal
were similar to what was found in this study. The ability to Testing machine.
carefully create standardized dimensions and geometries for
a variety of substrate materials provides reproducibility in
studying the fracture resistance for complex geometries.
Although previous studies have demonstrated that CAD sys-
References
tems are able to successfully mill ultra-thin restorations,1,3 we 1. Schlichting LH, Maia HP, Baratieri LN, et al: Novel-design
encountered minor chipping of margins in the minimal thick- ultra-thin CAD/CAM composite resin and ceramic occlusal
ness groups using a three-axis mill. This is consistent with veneers for the treatment of severe dental erosion. J Prosthet
Magne et al 2015, who also experienced minor chipping of Dent 2011;105:217-225
ceramic margins when milling ultra-thin feather-edge margins. 2. Magne P, Schlichting LH, Maia HP, et al: In vitro fatigue
This may be a limitation of three-axis mills and current material resistance of CAD/CAM composite resin and ceramic posterior
properties. It is anticipated that improvements will be made as occlusal veneers. J Prosthet Dent 2010;104:149-157
3. Edelhoff D, Sorensen JA: Tooth structure removal associated
the technology advances. Some have increased the bulk in the
with various preparation designs for posterior teeth. Int J
marginal area as part of the design and milling, with subsequent Periodontics Restor Dent 2002;22:241-249
hand finishing to thin these areas to avoid this chipping. An ad- 4. Magne P, Carvalho AO, Bruzi G, et al: Fatigue resistance of
ditional limitation in this study occurred due to the nature of ultrathin CAD/CAM complete crowns with a simplified
cross-sectioning specimens. Unfortunately, we were only able cementation process. J Prosthet Dent 2015;114:574-579
to view a single plane of the fracture. Ideally, we would be able 5. Kim JH, Lee SU, Park JS, et al: Fracture load of monolithic
to analyze the fractures in three dimensions to fully observe CAD/CAM lithium disilicate ceramic crowns and veneered
and understand the fracture mechanisms. zirconia crowns as a posterior implant restoration. Implant Dent
More investigation is needed to determine the true reliability 2013;22:66-70
of ultra-thin e.max posterior restorations intra-orally. Although 6. Choi S, Yoon HI, Park EJ: Load-bearing capacity of various
CAD/CAM monolithic molar crowns under recommended
the present study was able to digitally standardize the dimen-
occlusal thickness and reduced occlusal thickness conditions. J
sions and geometries for all crown and abutment materials, Adv Prosthodont 2017;9:423-431
limited information for the long-term performance can be pro- 7. Ma L, Guess PC, Zhang Y: Load-bearing properties of
vided. The current results can provide information on the ini- minimal-invasive monolithic lithium disilicate and zirconia
tial performance of ultra-thin restorations on differing substrate occlusal onlays: finite element and theoretical analyses. Dent
materials at extreme loading conditions. Additional clinical pa- Mater 2013;29:742-751
rameters such as masticatory patterns, occlusal anatomy and 8. Khawi H, Kaizer M, Chughtai A, et al: Polymer infiltrated
antagonist conditions should be considered. Additional studies ceramic network structures for resistance to fatigue fracture and
incorporating a larger sample size, thermal cycling, fatigue test- wear. Dent Mater 2017;32:1352-1361
ing, and ultimately controlled clinical trials to evaluate intraoral 9. Kim JW, Kim JH, Thompson VP, et al: Sliding contact fatigue
damage in layered ceramic structures. J Dent Res
survival are needed.
2007;86:1046-1050
10. Kinney JH, Marshall SJ, Marshall GW: The mechanical
properties of human dentin: a critical review and re-evaluation of
Conclusions the dental literature. Crit Rev Oral Bio Med 2003;14:13-29
11. Zhang Y, Kim JW, Bhowmick S, et. al: Competition of fracture
Minimal and traditional thickness e.max crowns demonstrated mechanisms in monolithic dental ceramics: flat model systems.
no significant difference in their fracture resistance. The only J Biomed Mater Res Part B: Appl Biomater 2009;88:402-411
group without catastrophic fractures was the minimal thickness 12. Cuy JL, Mann AB, Livi KJ, et al: Nanoindentation mapping of
e.max crowns on e.max abutments (enamel analog). Minimal the mechanical properties of human molar tooth enamel. Arch
thickness e.max crowns supported by a substrate with a higher Oral Biol 2002;47:281-291
13. Marshall GW Jr, Balooch M, Gallagher RR, et al: Mechanical
elastic modulus had slightly higher fracture strengths than mini-
properties of the dentinoenamel junction: AFM studies of
mal thickness e.max crowns supported by substrates with lower nanohardness, elastic modulus, and fracture. J Biomed Mater Res
elastic modulus. Fracture mechanisms of minimal thickness 2001;54:87-95
e.max crowns appear to be affected by the elastic modulus of the 14. Blackburn C, Rask H, Awada A: Mechanical properties of
substrate. Minimal thickness e.max crowns may behave more resin-ceramic CAD-CAM materials after accelerated aging.
predictability when compared to traditional thickness restora- J Prosthet Dent 2018;119;954-958
tions as noted by their smaller coefficient of variation. Mini- 15. Kassem A, Atta O, El-Mowafy O: Fatigue resistance and
mal thickness e.max crowns may be a viable restorative option microleakage of CAD/CAM ceramic and composite molar
when supported by high elastic modulus substrates such as crowns. J Prosthodont 2012;21;28-32
enamel. 16. Kelly JR: Clinically relevant approach to failure testing of
all-ceramic restorations. J Prosthet Dent 1999;81:652-661
17. Ferrario VF, Sforza C, Zanotti G, et al: Maximal bite forces in
healthy young adults as predicted by surface electromyography.
Acknowledgments J Dent 2004;32:451-457
18. Waltimo A, Kononen M: A novel bite force recorder and
The authors would like to thank Ivoclar Vivadent, Inc. for maximal isotnetric bite force values for healthy young adults.
supplying the dental materials used in this study and the Scctnd J Dent Res 1993;101:171-175

Journal of Prosthodontics 0 (2019) 1–8 


C 2019 by the American College of Prosthodontists 7
Fracture Resistance of Various Thickness e.max Crowns Chen et al

19. Braun S, Hnat WP, Marcotte MR: A study of bite force, part 1: 24. Bindl A, Luthy H, Mormann W: Strength and fracture pattern of
relationship to various physical characteristics. Angle Orthod monolithic CAD/CAM- generated posterior crowns. Dent Mater
1995;65:367-372 2006;22:29-36
20. Gibbs CH, Mahan PE, Lundeen HC, et al: Occlusal 25. Zimmerman M, Egli G, Zaruba M, et al: Influence of material
forces during chewing and swallowing as measured thickness on fractural strength of CAD/CAM fabricated ceramic
by sound transmission. J Prosthet Dent 1981;46:443- crowns. Dent Mater J 2017;36:778-783
449 26. Guess PC, Schultheis S, Zhang Y, et al: Influence of preparation
21. Kelly JR, Campbell S, Bowen HK: Fracture-surface analysis of design and ceramic thicknesses on fracture resistance and failure
dental ceramics. J Prosthet Dent 1989;62:536-541 modes of premolar partial coverage restorations. J Prosthet Dent
22. Silva NRFA, Bonfante EA, Martins LM, et al: Reliability of 2013;110:264-273
reduced-thickness and thinly veneered lithium disilicate crown. 27. Sasse M, Krummel A, Klosa K, et al: Influence of restoration
J Dent Res 2012;91:305-310 thickness and dental bonding surface on the fracture resistance of
23. Zhang Y, Mai Z, Barani A: Fracture-resistant monolithic dental full-coverage occlusal veneers made from lithium disilicate
crowns. Dent Mater 2016;32:442-449 ceramic. Dent Mater 2015;31:905-915

8 Journal of Prosthodontics 0 (2019) 1–8 


C 2019 by the American College of Prosthodontists

You might also like