You are on page 1of 2

Fuse Replacement

The Stellaris Vision Enhancement System has two (2) user-replaceable fuses. If
an over-current condition should occur which opens these fuses, they should be
replaced with fuses of the same value as the original fuses. A blown fuse may
be indicated by any of the following conditions:

¾ With the system off, using a known good outlet, no power is sent to the
system when you flip the main power switch to “on”. (i.e. the stand-by
power switch is not lit up, no faint fan noise from the lower rear of the
system, etc.).

¾ With the system on, it switches to low-power mode and the UPS turns on
with no discernible cause (i.e. the cord is still plugged in; the rest of the
operating suite still has power, etc.)

A blown fuse is usually noticeable after removal from the system by obvious
discoloration within the fuse and/or an obviously broken fuse-wire within the fuse.

Note: If damage is apparent to either fuse, both should be replaced to


ensure proper operation.

1. Remove the power cord from the Stellaris Vision Enhancement System.
The presence of the power cord will physically prevent the removal of the
fuse drawer.

2. Using a flat-blade screwdriver, carefully remove the fuse drawer. Replace


the fuses. Refer to Figure 1.

3. Snap the fuse drawer back into place.

4. Replace the cord and the system should be ready to run again.
Figure 1