Professional Documents
Culture Documents
Technical Manual
Copyright application © 2006, 2007, 2008 by emd technologies Inc. All rights reserved.
emd technologies and the emd technologies logo are trademarks of emd technologies Inc.,
registered in the U.S.A. and other countries.
Contents of this publication may not be reproduced in any form or by any means, electronic or
mechanical, including photocopying and recording, or by any information storage or retrieval system
without the written permission of emd technologies Inc., 400 du Parc, Saint-Eustache (Quebec),
Canada, J7R 0A1.
IMPROVEMENTS
emd technologies reserves the right to make changes and improvements to its products and assumes
no obligation to incorporate any changes into units previously sold or shipped.
July 31/2008
A. GENERAL
The generator and operational instructions are to be used by qualified persons having been trained
in radiation protection. The generator must be operated in conformance with the manufacturer's
instructions and within local government regulations.
A.2 Documentation
The present manual describes the installation, start-up & setup instructions and maintenance
procedures for the generator. It also gives the performance and technical specifications of the
described product and provides the installer with all of the safety procedures necessary for a safe
installation and maintenance.
Operator’s ° Operator’s Manual P/N 9895.00 : RAD & Fluoro Configuration (English)
Manuals ° Operator’s Manual P/N 9895.01 : RAD & Fluoro Configuration (OEM)
° Operator’s Manual P/N 9895.02 : Biplane Configuration
No
tes ° Operator’s Manual P/N 9895.03 : RAD & Fluoro Configuration (Chinese)
° Operator’s Manual P/N 9895.04 : RAD & Fluoro Configuration (French)
° Operator’s Manual P/N 9895.05 : Cardiac Configuration (English)
° Operator’s Manual P/N 9895.11 : Cardiac Configuration (English, CV Console)
A.3 Contraindications
No contraindications are known for the products described in this manual.
A.4 Definitions
ABS : Automatic Brightness Stabilization, in Fluoro mode. Mode of operation in which the
factors of the Fluoroscopic beam is automatically adjusted to obtain an optimal image
brightness.
kV : Peak voltage applied between the anode and cathode of an x-ray tube, in thousands of
volts.
kW : Describes the product of both kV and mA output at the generator or loaded at the
terminals of the X-ray tube.
mA : Average x-ray tube current, in milliamperes, during the irradiation time. The operator
console can also display the peak mA rather than the average mA, during the optional
Pulsed Fluoro, if the user does want so.
Time : Irradiation time, in milliseconds (ms) or seconds (s). Irradiation time represents the time
interval between the instant that the tube potential has risen for the first time to a value
of 80% and the instant at which it finally drops below the same value.
User : Authority responsible for the use and maintenance of the generator.
WKS : Workstation (or image receptor). Transducer device that converts the X-ray beam energy
across the patient into visible images.
(1) Medical radiology – Terminology, IEC Standard Publication 788, First Edition, 1984.
B. SAFETY PROCEDURES
Only properly trained and qualified personnel should be permitted access to any internal parts. Live
electrical terminals are deadly; be sure line disconnect switches are opened and other appropriate
precautions are taken before attaching accessories.
Do not remove the flexible H. T. cables from the X-ray tube housing or Anode or Cathode
assemblies H.T. connector until the electrical mains have been disconnected for at least 1 minute.
The useful and scattered beams can produce serious or fatal bodily injuries to any persons in the
surrounding area if used by an unskilled operator. Adequate precautions must always be taken to
avoid exposure to the useful beam, as well as to leakage radiation from within the source housing or
to scattered radiation resulting from the passage of radiation through matter.
Those authorized to operate, participate in or supervise the operation of the equipment must be
thoroughly familiar and comply completely with the currently established safe exposure factors and
procedures described in publications. Failure to observe these warnings may cause serious or fatal
bodily injuries to the operator or those in the area.
According to the class of equipment, the Epsilon High-Voltage generator must be referenced to
an earth ground by a protective earth conductor. Failure to provide a separate earth ground can
result in electrical shock hazard causing injury or death.
C. ELECTROMAGNETIC COMPATIBILITY
The Epsilon EPS 45-80 High-Voltage X-ray generator complies with the IEC 60601-1-2 standard
(2nd version: 2001).
Compliance to any EMC standard does not guarantee that a device will not interfere with or will not
be affected by the operation of another electrical medical equipment placed in the nearby
environment. Therefore, the EPS 45-80 High-Voltage X-ray generator is not designed to work in
close vicinity with other electrical equipment or in the presence of portable and mobile RF
communication equipment. Stacking and adjacent positioning of equipment should be avoided.
EMC Instructions
1. When stacking or adjacent positioning of equipment or accessories is unavoidable,
make sure that the performance of the EPS 45-80 High-Voltage X-ray generator
and of the stacked equipment is not adversely affected.
2. The performance of the EPS 45-80 High-Voltage X-ray generator can be affected
by the presence of portable and mobile RF communication equipment.
3. Do not operate the EPS 45-80 High-Voltage X-ray generator without all covers
and enclosure perfectly fitted onto the High-Voltage Power Supply.
d = [1,2] P
Conducted RF 3 Vrms 3 Vrms
IEC 61000-4-6 150 kHz to 80 MHz d = [1,2] P , 80 MHz to 800 MHz
IEC 61000-4-3 80 MHz to 2.5 GHz where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Note 1 At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment, due to fixed RF transmitters, an electromagnetic site survey should
be considered. If the measured field strength in the location in which the EPS 45-80 High-Voltage X-ray generator is used
exceeds the applicable RF compliance level above, the EPS 45-80 High-Voltage X-ray generator should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting
or relocating the EPS 45-80 High-Voltage X-ray generator.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
D. IMPORTANT SYMBOLS
E. TABLE OF CONTENTS
IN THIS CHAPTER :
The EPS 45-80 High-Voltage generator is an Advanced High Frequency X-ray Generator designed
with a unique power inverter technology and modular conception.
As more power or custom features are needed, they are simply added using plug-in modules quickly
inserted to the generator's main core. The latter consists of an electronic controller driving an
encapsulated (oil-free) high-voltage unit. The generator comes in different configurations, best
suited to conventional RAD & Fluoro applications and to the latest digital procedures.
2. A modular concept allowing large OEMs to efficiently manage the system configuration and
costs in order to compete successfully in today’s competitive world.
4. An exceptionally high efficiency which allows for less severe mains line impedance requirements.
7. Epsilon PC Tools program used for installing, setting up and servicing the Epsilon High-
Voltage generator. The program features a system monitor mode, a history log and a statistics
register to record and track back important events occurring during the course of the generator
maintenance and normal operation.
8. RS-232 (or CAN) software providing full flexibility for customer control.
X-ray tube : One dual focal spot X-ray tube. Two-Tube option available.
Anode drive : Integrated Low or Dual Speed Rotor Controller. See options in § 1.1.3.
An elegant X-ray control panel that affords simple control of the EPS High-Voltage generator.
Depending on configuration, the console controls between 4 and 6 workstations with more than
2000 APR exams divided in 3 or 4 patient sizes.
Exposure buttons, optical density correction, and many feedbacks from the X-ray room and/or
X-ray tube are provided. The technique factors are quickly available for both pre-exposure and post-
exposure visualization.
The console also includes a built-in service mode and a diagnostic mode, displayed in a graphic
screen. Connection to the generator is carried out by a 23.6-m (50-foot) console cable (other cable
lengths are optional).
Microprocessor-controlled electronic board that powers the anode rotation at low (60 Hz) or dual
speed (60/180 Hz), for one or two X-ray tubes. Anode rotation is independent of mains line
frequency. The Epsilon Rotor Controller supports most common stator types.
The 2-Tube option provides a standalone X-ray tube switch and the necessary hardware & software
to drive two different X-ray tubes from a single generator. Optionally, the 2-tube assembly can be
enclosed in a decorative cabinet.
High Performance Radiography option is available on EPS 80 configurations only. It extends the
irradiation time from 10 seconds to 63 seconds, and the mAs factor from 1000 up to 12,500 mAs on
some very high capacity X-ray tubes.
High Performance Pulsed Fluoro is performed on standard X-ray tubes (no grid), using a
programmable surface dose rate and a variable pulse width. It is done in Dose Reduction or Image
Enhancement mode.
• Dose reduction lowers average patient dose at slower pulse rates while keeping the same
dose per pulse (the pulse count is reduced).
• Image enhancement provides clearer, less noisy images at slower pulse rates while
keeping the patient dose about constant.
• Source to Patient Distance Dose Compensation Option:
The generator adjusts the X-ray beam to compensate for the
source-to-patient distance (SPD) when the gantry position
varies during an exam. This provides constant surface dose
along the full SPD span.
mAs/pulse ∝ (1/SPD)2
• High Frame Rate option: Typically dedicated for high grade special procedures, this
option allows image acquisition at up to 90 frames per second, RAD & Fluoro.
Continuous Fluoro can be enabled using the service mode, for example to troubleshoot the Fluoro
imaging chain.
Provides a medical grade power supply rated 24 VAC, 150 VA (6.25A) and 24 VDC (2.7A), 65W.
Consists of a plug-in board and the necessary software to perform AEC exposures on up to 3 solid
state AEC chambers. This board includes also an additional interface for a standard ion chamber.
(B)
An X-ray Hand-Switch (C) provides the Epsilon operator console with a hand held push-button
switch that can be extended 3.8 m (8 feet) from the operator console. Does not permit Fluoro
activation.
Can be installed on the left or right side of the console.
Irradiation time From 1.0 ms (1) to 10 seconds (2), with 25% increment steps (41 stations):
1.0 (1); 1.25; 1.6; 2.0; 2.5; 3.2; 4.0; 5.0; 6.3; 8.0; ms
10; 12.5; 16; 20; 25; 32; 40; 50; 63; 80; ms
100; 125; 160; 200; 250; 320; 400; 500; 630; 800; ms
1.0; 1.25; 1.6; 2.0; 2.5; 3.2; 4.0; 5.0; 6.3; 8.0; 10.0 seconds.
(1) Minimum irradiation time in AEC = 5.0 ms
(2) Maximum irradiation time with optional High Performance Radiography = 63 seconds.
mAs From 0.4 to 1000 mAs (1), with 25% increment steps (up to 35 stations)
0.4; 0.5; 0.63; 0.8; 1.0; 1.25; 1.6; 2.0; 2.5; 3.2;
4.0; 5.0; 6.3; 8.0; 10; 12.5; 16; 20; 25; 32;
40; 50; 63; 80; 100; 125; 160; 200; 250; 320;
400; 500; 630 (2, 3); 800; 1000.
(1) Maximum mAs in the optional High Performance Radiography: 12 500 mAs.
(2) 600 mAs max. in AEC mode.
(3) 500 mAs maximum for 30 kW configuration.
Tube kW Both operative frame rate and run length impacts on the allowed tube kW load. The
generator ensures the X-ray tube operating parameters are within safe limits.
Notes (1) Available on optional configurations only. (2) Faster frame rates are available as an option.
(3) When the serial tube loading is synchronized using an external trigger, the rate of the tube's loading
will be slowed down if the triggers fail to come at the expected speed.
(4) The run length corresponds to the maximum total duration of repeated tube loadings allowed by the
generator for a distinct exposure switch actuation.
kW Small/Low
Small/High
Large/Low
Large/High
kW
-3 -2 -1 Normal +1 +2 +3
Max. kW > 30 kW – – – –
Max. kW > 30 37 45 kW – – –
Max. kW > 30 37 48 50 kW – –
Max. kW > 30 37 48 55 kW – –
Max. kW > 30 37 48 60 65 kW –
Max. kW > 30 37 48 60 70 kW –
Max. kW > 30 37 48 60 75 80 kW
Max. kV > 150 150 150 150 150 127 113 100
EPS 100
(*) Nominal electric power values (kW) are given for a loading time no less than 100 ms.
Tube loading:
100 kV, 400 mA, 10 ms
DUAL FLUORO: Permits to switch from pulsed to continuous Fluoro at the Epsilon operator
console, and vice versa.
FOCAL SPOT : Continuous Fluoro: Small filament (large filament if small filament opens);
Pulsed Fluoro: Small or large filament, as a function of the tube mA.
ALARM : Continuous alarm every 5-minute Fluoro interval. The alarm is shut down by
pressing a key on the Epsilon console or by using a digital input to the generator.
TIMER : Fluoro timer cumulated during the Fluoro exam. The timer is displayed and reset at
the Epsilon operator console.
ABS CURVES:
Four (4) service programmable kV/mA tracking curves are provided to improve the contrast during
ABS stabilization. The current ABS curve can be changed from the operator console.
mAs range From 0.025 up to 5.0 mAs @ any pulse rate and any kV.
Pulse rate 0.25 up to 30 p/s (7 pulse rates available at the operator console).
Up to 90 p/s with High Frame Rate option.
Continuous Fluoroscopy
• Photomultiplier tube; Requires the Fluoro PMT board P/N 9822.00. Gain is adjusted via a
high-voltage bias supply which can be varied from about -200V to about -950V.
• Photodiode; Requires the Photodiode Ion Chamber Fluoro board P/N 11030.000.
Usable input signal range is from about 5 mV to 1 V or from about -5 mV to -1 V.
Pulsed Fluoroscopy
• Photomultiplier tube: Requires the Fluoro PMT board P/N 9822.00. Gain is adjusted via a
high voltage bias supply which can be varied from about -200V to about -950V.
• Photodiode: Requires the Photodiode Ion Chamber Fluoro board P/N 11030.000.
Usable input signal range is from about 12.5 mV to 9 V or from about -12.5 mV to -9 V.
“Anode H.U. over 80%” alarm Pulsed tone (2 beeps per second)
“80 % of maximum run length” alarm Pulsed tone (2 beeps per second)
“Anode H.U. over 80%” alarm Pulsed tone (2 beeps per second)
3. The tone lasts for the duration of the X-ray exposure (1/4 second minimum).
1.2.5 Miscellaneous
Number of Tubes One X-ray tube standard; 2-tube configuration available as an option.
Models Most popular X-ray tube models are supported through a built-in X-ray tube
chart. Other X-ray tube models may be added by the installer.
Tube Filaments
Current range : Up to 8.1 RMS amperes
Filament power : Maximum 100 watts
Focal spot number : 2
Tube Derating
The tube can be derated to extend its usable lifetime, to troubleshoot specific conditions or other
reasons.
TEMPERATURE HUMIDITY
• Storage : - 25 to + 70°C • Storage : 30 to 95 %, non-condensing
• Operating : + 10 to + 40°C • Operating : 30 to 75 %, non-condensing
If the EPS generator is fit with the optional cable discharge unit (high-end Cardiac configurations),
this latter contains approximately 30 liters of dielectric oil. When comes the time to dispose of the
cable discharge unit, it must done in a way that respects the local environmental regulations in force.
141
5 1/2"
750 812
29 1/2"
32"
656
25 7/8"
174
6 7/8"
48
37
H.V. Cables
1 7/8" Access
81 303 1 1/2"
23 680
3 1/4" 12"
7/8" 26 3/4"
H.V. Cables
Access 781
30 3/4"
Front face of EPS Cabinet Side of EPS Cabinet
1 2
2
Upper left side cable access
(also possible on right side)
Upper front cable access
EPS generator inside decorative cabinet
(also possible on rear face)
Multiple High-Voltage Cables Access
Figure 1.7 – Decorative Cabinet dimensions and different possible cable access.
Epsilon High-Voltage Generators comply with the following standards and regulations (Table 11).
Note The conformity is based on the equipment configured and interconnected as described
in the present Technical Manual. Changes or modifications to this equipment not expressly
approved by emd technologies could void its conformity to applicable standards.
Notes:
1. Accuracy specifications were measured at 100 kV at half the nominal power with 35 feet
(10.7 m) H.V. cable between H.V. Power Supply and X-ray tube.
2. 0.5 kV ripple at 100 kV, 250 mA with 50' (15.25 m) High-Voltage cables
3. Irradiation time represents the time interval between the instant that the tube potential has risen
for the first time to a value of 80% and the instant at which it finally drops below the same value.
This section describes the physical installation and electrical supply of the generator. Interfacing, system
settings and calibration are described in subsequent chapters.
IN THIS CHAPTER :
The packing list enumerates all parts shipped with the generator and their serial numbers. A copy of
the list should be left on the site and a second copy should be filed at the installer's office. Inspect all
parts for any mechanical damage that could directly or indirectly impair the operation of the system.
1. X’s are placeholders for version numbers; item versions might differ according to generator
configuration;
2. Pre-installed on the Controller board; 3. Pre-installed on the Motherboard;
4. Pre-installed on the Console board; 5. Other cable lengths available.
1. X-ray Tube One 2-focal spot X-ray tube, Class 1 Type B (IEC Classification)
Two X-ray tubes when the “2-tube option, P/N 10020.000” is fitted
on the generator.
Use the same cable type and cable length for both anode and cathode sides
3. ABS/AEC Sensors
i. Ion chamber: AID ICX type compatible
Positive or negative integrating ramp
Active low or active high field selection, from 4 to 24 VDC
Active low or active high integrator reset signal, from 4 to 24 VDC
v. Varian flat panel detector: Recommended part number VIP-9 (Varian Image Processor)
With RG-59/U coaxial signal cable
1. Should the unit be permanently fixed on a surface or on the floor, you ought to leave the
following minimum clearance space (Figure 2.2):
8.75" on top of the unit, for servicing;
1" minimum elbowroom on each side, for proper ventilation;
No elbowroom necessary on back and front panels, after proper installation is
completed.
3. The console can be installed up to 100 feet away of the High-Voltage Generator (Section 2.6).
For longer installation setup, contact emd technologies service department.
Figure 2.3 – High-Voltage connection section of EPS 45-80 Generator (Line Access Panel removed).
Refer also to Figure 2.4 for details on Line Filter Board.
The generator requires a three-phase electric line with a protective earth ground (4 or 5 wires,
maximum wire gauge 0 AWG (50mm)).
2. Remove the Line Access Panel at the back of the Epsilon High-Voltage Generator to feed the
line wires through the cable clamp (Figure 2.3).
3. Connect the mains line of the disconnect switch to the Line Filter Board:
J2 (φ1), J3 (φ2), J4 (φ3) and J5 (Neutral) (i) lugs (ii) (Figure 2.4).
(i) For Delta line configuration, leave the neutral lug unconnected.
(ii) Lugs maximum wire gauge: 0 AWG (50 mm2).
5. Firmly attach the wires on the Line Filter Board Holder using cable-ties.
Figure 2.4
Line Filter Board
i) Power Line connection
ii) Tube Stator connection
iii) Tube Housing connection
iv) K1/K2 Relay connections
v) F3 & F4 fuses are installed with
Auxiliary Power Supplies option
only.
2. Label one of the two High-Voltage cables as "Anode" and the other one as "Cathode".
3. Take off the protective covers of the High-Voltage output on each receptacle. Do not discard them
as they can be used later for shipping.
4. Ensure proper isolation of the High-Voltage receptacles by coating them with insulating grease
or by using a lubricated silicone washer. Isolation is necessary to prevent electric arcs between
the receptacles and the Anode/Cathode modules.
5. Insert the high-voltage cable connector (anode) in the high-voltage receptacle (anode) and make
sure the contact pins are engaged in the holes of the receptacle. Tighten the cable nut while
holding the cable firmly in place. Use an appropriate Allen ranch to tighten the cable ring nut. If
necessary, consult the installation instructions provided by the H.V. cable manufacturer.
7. Attach both high tension cables to the H.V. cable retainers (Figure 2.3) using Tie-wraps.
8. To connect the high tension cables to the X-ray tube, follow the installation procedures
provided by the X-ray tube manufacturer.
Table 13-A Tube stator connection to Line Filter board P/N 9916.x
Usual cable color Signal Connector
Black Stator Main J6 pin 1 – Main T1
White Stator Common J6 pin 2 – Common T1
Green (Red) Stator Phase J6 pin 3 – Phase T1
Bare lug Stator Cable Shield Shield (on Line Filter Board Holder, Figure 2.4)
Table 13-B Tube housing connection to Line Filter board P/N 9916.00
Signal Connector
Tube Housing J8 pin 1 – Tube 1
Tube Housing Return J8 pin 2 – Return 1
Tube Housing Shield Shield (on Line Filter Board Holder, Figure 2.4)
2. Figure 2.5 below illustrates the back panel of the operator console. Use cable P/N 10015.xxx to
interconnect the Epsilon console (J7) with the generator controller (J4).
3. Use the two swivel leveling mounts on the console base to adjust the visualization
angle for the operator.
This adjustment would be refined later when the operator console will be turned on
for the first time.
No
tes At a later stage of the installation process, the operator console will be used to
setup the generator through a PC compatible computer.
1. Install the hand-switch holder on the hand-switch bracket using the two pairs of lock nuts
and HMS screws n°4 .
2. Unscrew the left or the right swivel leveling mount according to which side the hand-switch will
be installed.
3. Install the hand-switch bracket on the operator console using the swivel leveling mount ,
Sems HMS screw and the two stand-offs .
3. Hand-Switch Configuration: The operator console is usually shipped with its PREP/RAD
push-buttons disabled when the Hand-switch is connected.
To enable the console PREP/RAD push-button along with the hand-switch, move the jumper
on the console board P/N 9780.01 from position 2-3 to position 1-2 (Figure 2.7).
Figure 2.7 – PREP/RAD configuration on Console Board P/N 9780.01 (Hand-Switch option).
The Epsilon generator is normally shipped with the 0 VDC of all I/Os tied to the chassis ground of
the generator (Figure 2.9-A) using three (3) ground strap brackets. It is possible that this configuration
can be detrimental to certain peripheral equipment in case of tube arcs.
To isolate the field interface from the generator's chassis ground, undo the ground strap brackets
from the generator (Figure 2.9-B) and connect instead the client I/O system chassis ground to the
generator chassis ground provided on the Field Interfacing section of the Controller.
If the three (3) ground strap brackets are removed from the Controller, it is the client’s
responsibility to wire his main I/O ground to the client's ground studs ( ).
(Generator Section)
(Field Interfacing Section)
X-ray Room
F
Central Earth Ground
(Generator Section)
(Field Interfacing Section)
X-ray Room
F Central Earth Ground
Field Equipment
NO Ground Strap
(To prevent ground loop)
!
Field Equipment connected to
X-ray Room Earth Ground
It is recommended to provide the installed generator with mechanical enclosure to afford better
electrical security to the personnel. This step can be done using the Decorative Cabinet
P/N 10017.001 option.
Installation Note:
Final enclosure of the generator should be done at the end of the installation procedure only. It is
therefore better to postpone the final enclosure of the EPS generator after the Field-Calibration
procedure (Chapter 5) or the User Acceptance tests (Chapter 6) are performed. Refer also to Figure
1.7 for the decorative cabinet dimensions and different cable access configurations.
The interface of the peripheral equipment with the Epsilon generator is done on the Controller
board and optional expansion boards (Figure 3.1).
No
tes
Most signals are site-dependent and require software configuration using the
Epsilon PC Tools program (Chapter 4).
Progr. output 3 9
Analog signal for interfacing an ABS control device, typically
Remote AEC /
10 a flat panel, or other delayed feedback device. § 3.7.4
SPD Tracking
Can be used also for SPD tracking.
Progr. output 8 B6
Ground B4 0 VDC
GND 3, 4 Ground
N.U. 5, 6, 7 N.C.
Power on/off
8 The line is isolated from other console signal lines
Common
DTR 4
Ground 5, 6
RTS 7
CTS 8
N.U. 4 Needs generator power on/off signals (see J6 connector, next § 3.8.2
Table).
Ground 5, 6
Baud rate: Defaulted to 19, 200 (Configurable in PC Tools)
RTS 7
330Ω to 12 V 8
Ground 8 0 VDC
Connector J2:
Future expansion
Connector J3:
To Expansion Motherboard P/N 9824.00
The tube loading is requested by the activation of an exposure switch. After the exposure switch
gets activated, the generator will load the X-ray tube only if specific safety conditions are met (no
fault, interlock is ok, tube is ok, etc).
It is not possible to perform successive loadings of the tube without releasing first the exposure
switch between each tube single loading (single RAD) or sequential loading (serial RAD).
Fluoro Switches: Two Fluoro switch inputs are provided on the Fluoro boards (PMT or PhD),
Low & High Fluoro levels (Figure 3.2). When either Fluoro input comes low, Fluoro imaging starts
if all prerequisite conditions are acknowledged (optional interlock, Filament current and trigger in
Pulsed Fluoro). Both levels have their own Fluoro entrance exposure rate (Section 5.4) and dose
stabilization control (Section 5.5).
Note: Prep and RAD switches have priority over Fluoro footswitch.
Exposure Switch Ground: Exposure is allowed only when the exposure footswitch common signal
(J12 pin 3 or J2 pin 5, etc) is grounded via “Console Common Enabled” set from CPU (bit activated
thru Epsilon PC Tools configuration).
To assign the proper trigger input and configure its activation level for a given workstation, use the
Epsilon PC Tools Program (Section 4.5 for RAD & Section 4.6 for Fluoro). Refer to Figure 3.3b
for different types of trigger activation.
3.4.1 General
3.4.1.1 Programmable Input Types
The connection points shown in Table 41 are available inputs to interface the X-ray room with the
generator. Every digital signal can be set up as one of the following input types using the Epsilon
PC Tools program (=> Chapter 4):
1. Interlock (§ 3.4.2); 5. Fluoro Alarm Reset (§ 3.4.6);
2. Frame Rate input (§ 3.4.3); 6. Workstation Select (*);
3. AEC/Fluoro Lock Mode (§ 3.4.4); 7. Exam Select (*);
4. AEC Reset (§ 3.4.5); 8. I. I. Magnification inputs (**).
(*) Workstation & Exam Select inputs are dedicated to CBCT mode and shall
not be used otherwise. Refer to Technical Manual Supplement
P/N 11050.011 in the electronic documentation CD.
(**) I.I. Magnification inputs are used only on generators with High
Performance Pulsed Fluoro. Refer to Technical Manual Supplement
P/N 11050.012 in the electronic documentation CD.
The input threshold value will then be half the value of the reference (VTH = V REF/2). The selected
voltage is present as a reference on J14 between pins 1 & 14 and J15 between pins A1 and B4.
Refer to Table 41 for diagram. Do not use this voltage as an external voltage supply.
The location, number and names of interlocks can differ from the one shown
above using the Epsilon PC Tools program (=> Chapter 4).
For a given workstation, the interlock can be of 3 different types, each one corresponding to a
specific generator state described in Table 42. The generator will validate each interlock at a different
time during the exposure process, as depicted in Figure 3.6. Refer to Chapter 4 to program the
interlock inputs.
Figure 3.6 (1/2) — Interlock Ready & Start shown in a serial RAD exposure.
In AEC serial exposure (RAD), the first frame is done using the AEC feedback. The Lock input is
sampled at the end of each exposure:
• If false, the next exposure is performed in AEC mode;
• If true, the next exposure is performed at the latest measured mAs parameters.
In automatic Fluoro (ABS), kV/mA corrections are considered after each exposure pulse (Pulsed
Fluoro) or after every time interval (Continuous Fluoro). At each correction time, the following
conditions might produce one of the following actions.
4. Idle Mode
When the same Fluoro level is selected from Idle Mode that was just previously locked and the
Lock Input is still true, the generator continues with the same locked parameters. Selecting the
other Fluoro level starts the Fluoro sequence as in condition 2 above.
A7
J15 A8
Controller board
B1
B5, B6
A7
J3 Fluoro PMT A8
Board (Option) B1
Table 45- Selectable Polarity & Voltage on Programmable Outputs - P/N 11030.000
Connector Pin Typical Interfacing Description
4 Outputs can be active low or high (selectable polarity), and the voltage
J3 value of the output at high level is programmable from 4 to 24 V. This is
5 done using the Epsilon PC Tools program, as described in Chapter 4.
(*) I.I. Magnification outputs are used only on generators with High
Performance Pulsed Fluoro. Refer to Technical Manual Supplement
P/N 11050.012 in the electronic documentation CD.
Figure 3.10 – Typical X-ray On and Main Gate signals Interfacing Diagram.
Figure 3.11 – Momentary output activation associated to a F1 key on the Epsilon operator console.
These contacts are not available with the optional Auxiliary Power
Supplies P/N 10023.000. If this option is fitted on the generator,
refer to Technical Manual Supplement P/N 11050.000 or
P/N 11050.013 in the electronic documentation CD.
+ 24 V K1A
4
K1B 3
2
K2B 1
K2A
J7
K2 Progr. Out
The relays can be used as general purpose control lines. They are activated independently by the
programmable outputs of the generator; the status of each relay is thus dependent upon the various
operating conditions of the generator (such as IDLE, PREP, READY or FLUORO; see "Output
Configuration", Section 4.5.2).
+ 24 V
K1B
K1A
K1 Progr. Out
1 K1 N.O.
2 K1 Common
3 K1 N.C.
K2B K2A 4
K2 N.O.
K2 Progr. Out 5
K2 Common
6
K2 N.C.
7 K3 N.O.
8 K3 Common
K3B 9 K3 N.C.
10
K3 Progr. Out K3A K4 N.O.
11
K4 Common
12 K4 N.C.
J1
K4B
K4A
K4 Progr. Out
K5B
1 K5 N.O.
K6B 2 K5 Common
K6A
K6 Progr. Out 3 K5 N.C.
4 K6 N.O.
5 K6 Common
6 K6 N.C.
7 K7 N.O.
8 K7 Common
K7B 9 K7 N.C.
K7A 10
K7 Progr. Out K8 N.O.
11 K8 Common
12 K8 N.C.
J3
K8B
K8A
K8 Progr. Out
Notes:
1. The ion chamber feedback signal (pin 1) can ramp from 0 VDC up to (±) 10 VDC.
2. It is highly recommended that the ion chamber be electrically isolated from the chassis.
3. V High is set in Epsilon PC Tools: 4 VDC < VHIGH < 24 VDC.
4. Available ionization chambers are:
i. Controller board P/N 9798.01: Ion chambers # 1 and 2, J13 & J8 connectors;
ii. Universal Chambers Interface board P/N 11002.000
• Ion chamber # 3, J1 connector;
• Solid State A.E.C. chambers: Solid State Chambers # 4, 5 and 6; see Section 3.6.1.2.
iii. CPX Adaptor board P/N 9990.00: Ion chamber # 7, J101 connector;
iv. Photodiode Ion Ch. Fluoro board P/N 11030.000: Ion chambers # 8 and 9, J5 & J4 connectors
Figure 3.14 – Ion Chamber Interfacing Diagram.
1. Refer to Figure 3.14 to connect the ion chamber on the Controller (or optional board).
2. Using the Epsilon PC tools, select the activation polarity of the field select signal and integrator
reset signal (independent settings). Refer to Section 5.7.1 to setup the fields and integrator
reset signals as per the manufacturer specifications.
3. Refer to Section 5.7.2 to fine-tune the AEC using the film-screen correction tables and the gain
of the AEC feedback signal.
The board stores the resulting voltage for accurate feedback measurement. Using integrating
capacitors, up to three field inputs can be summed up (Field A, B and C). The input signal current
can be positive or negative.
2. For a phototube device, connect the high voltage supply of the phototube (RG-59/U coaxial) to
J1 BNC-connector on the Fluoro PMT board.
3. Refer to :
− Section 3.6.4 to interface SPD option, if fitted on the generator.
− Section 5.4 to adjust the Fluoro maximum surface dose.
− Section 5.5 to adjust the Fluoro stabilization dose.
Figure 3.19 – Photomultiplier Tube Interfacing Diagram on Fluoro PMT board P/N 9822.00.
To remotely control the Fluoro stabilization, the line at J3 pin B3 (0 to 12 VDC, Figure 3.20) can be
used; this stabilization method requires that the ABS source in the Fluoro setup be set to Rmt AEC
(Remote AEC, Section 4.5).
For an external console to be used with the generator, it must also provide the power on/off
switching of the generator. This is achieved on the Controller J6 connector (Figure 3.21).
Switching off the generator: Always turn off the generator by using the power off button on
the operator console or equivalent. When the tube is rotating, switching off the generator by
using the mains line contactor or any means other than the power off button will let the anode
coasting and might lead to hazardous consequences for the X-ray tube and/or anode starter.
In this section, the power is applied for the first time to the generator. Then, using the Epsilon PC
Tools, the complete setup of the generator is made to customize it as per the installation site
characteristics and specific peripherals.
IN THIS CHAPTER :
3. Do not press PREP, RAD or Fluoro during the System Setup Procedure.
After the generator is setup, it will be ready for the final calibrations (Chapter 5):
7. Initial Turn-On & Stator Type 10. Fluoro Dose Stabilization (Section 5.5)
Configuration (Section 5.1) 11. Fluoro kV/mA Curves (Section 5.6)
8. X-ray tube Calibration (Section 5.2)
9. Fluoro Entrance Exposure Rate (Section 5.4) 12. AEC Fine-Tuning (Section 5.7)
If the top cover is removed from the of EPS 45-80 generator, the utmost care must be taken as dangerous and
lethal potentials are present in offline circuits in several places! Every one of the heat sinks shown in the
picture below is at live AC line voltage!
Even when the generator is switched off at the operator console, mains voltages are still present in High-
Voltage Power Supply unit. These voltages are extremely dangerous, use the utmost caution.
1. Make sure the generator is installed as per the instructions of Chapters 2 & 3.
4. After power up, the operator console should display the initial power on screen shown in Figure
4.2 (1.), with the default workstation and initial tube loading factors.
• If the message displayed on the console is not “Small Fil. Calib. Required”, go to Section
7.4 to take an action as a function of the displayed message.
• If the message “Small Fil. Calib. Required” displays, press [Next] selection key twice
until the message “Anode H.U. = 0%” displays (Figure 4.2, 3.).
1. 2. 3.
Figure 4.2 – Typical sequence of power up screens on the Epsilon Operator Console.
5. Depress the Fault Reset key on the console until the Monitor Mode displays on the graphic
screen. Choose "Inverters" to invoke the Inverters Status screen (Figure 4.3). Make sure that no
BAD or FUSE status message displays on any Inverter board. Then, press "Exit" key and
proceed with the system setup of Sections 4.3 through 4.8.
Select
Inverters
Hold key to
Monitor
Switching off the generator: Always turn off the generator by using the power off button
on the operator console. When the tube is rotating, switching off the generator by using the
mains line contactor or any means other than the power off button will let the anode coasting
and might lead to hazardous consequences for the X-ray tube and/or anode starter.
1. Close all applications and resident programs currently running on your computer.
2. Explore the contents of the CD and double click on the Epsilon PC Tools.exe program file.
3. The Installation program will prompt you for specific information. At each prompt, select the
appropriate response and/or click “Next”.
NOTE For computer not equipped with a COM (RS-232) serial port, use an
USB adaptor. Recommended devices are:
• Edgeport/1 USB Converter, by Digi ConnectwareTM
• GUC232A USB to Serial Adapter, by IOGEAR
• RS-232 Optical Isolator (B & B electronics, P/N US09ML2)
1. Have installed on your service computer the Epsilon PC Tools program and the USB-Driver
for emd Operator Console (USB-Driver_EMD Operator_Console.zip). These two files can be
found on the electronic documentation CD (Figure a) or on your personal customer page on
emd technologies web site.
3. Connect the USB Printer cable between the laptop and the operator console.
4. Turn on the generator. The message “Find New Hardware” will display on the Notification
bar, followed by the “Found New Hardware Wizard”.
5. On the new Hardware wizard, choose “No, not this time” and then click Next button. Select
“Install from a list or specific location (Advanced)” and then click Next button (Figure b).
6. Select “Include this location in the search” and click the Browse button (Figure c). Locate the
unzipped driver folder on your local disk (USB-Driver for emd Operator Console).
7. If the compatibility window displays (Figure d, right), click the “Continue Anyway” button.
Note: On some computers, it is possible that the above installation steps should be repeated
once before final installation could be completed successfully.
8. After the installation is completed, click the Finish button (Figure e). Then, verify on which
port the driver has been installed on your laptop. This is done using the Windows control
panel (Figure f):
Control Panel/System/System Properties Hardware/Device Manager/Ports (COM &
LPT1)/USB Serial Port (EMD technologies settings) (COM“n”).
9. Open the PC Tools Program. Choose “Transfer/Settings” and select the correct COM port.
Click the OK button (figure g).
10. The communication with the operator console can now be established.
2. Select either “Open an existing configuration file” or “Create a new configuration file”
and click the OK button (Figure 4.4).
3. Locate the following elements on the screen (Figure 4.6): the title bar, the menu bar, the
different parameter configuration zones where the generator database can be edited, the status
line and connection status at the bottom of the screen.
Menu Bar
Database
Configuration
Parameters
Messages
1. 2. 3. 4.
4. Before beginning the setup of a generator, it is wise to save the current blank file by using the
menu "File" and "Save As". Use a file name that will clearly identify the generator. For
example, the file could be named EPS65RF.emd or Saint-Eustache Hospital.emd. The file will also
serve as a backup of the generator database (Section 4.2.3.2).
5. At any moment, a given configuration file can be retrieved by using the menu "File" and
choosing "Open".
Figure 4.7 – Connection Status & Possible Data Exchange using Epsilon PC Tools.
7. To transmit the whole set of parameters from the PC’s configuration file to a connected
generator, select "Transfer" and "Send to Generator".
To replace the data of the current configuration file (on the PC) with the complete generator
database data of the connected generator, select "Transfer" and "Get from generator".
Note that in Monitor/Transfer connection mode, the operator console is set in Service mode
(Figure 4.8). This mode is discussed later in the manual.
Important
The default parameters on a new machine will have to be customized to interface the generator
with the specific peripheral equipment and user’s working preferences.
Use the “File Save As” PC Tools command to save the configuration file on disk under a
meaningful name (for example Saint-Eustache.emd). PC Tools uses emd extension only.
While the service representative gradually configures the generator by editing the generator database,
it is recommended to backup the installation data by making a copy of the configuration file on the
hard disk. This will permit an eventual reload of the generator data from the computer in cases data
corruption has occurred.
The complete backup of the generator data is held onto 4 different files.
• Refer to Section 8.2.3.5 to make a complete backup.
• Refer to Section 7.7.3 for a description of the data files.
The EPS generator comes in several possible configurations, for example RAD-only, RAD/Fluoro,
2-tube configuration, and with a certain number of optional expansion boards.
1. Before beginning the setup of the generator, make sure you start with an existing configuration
file or, else, create a new one.
To create a new configuration file from a blank file:
• Connect the generator to the PC, start Epsilon PC Tools
and select “Create a new configuration file”.
• Save the current “untitled” blank file by using the menu
"File" and "Save As". Use a file name that will clearly
identify the EPS generator currently installed. Save as…
2. Double-click twice on the "Connection status" icon (bottom right-hand corner) of the Epsilon
PC Tool status bar until the status toggles to “Monitor/Transfer”.
Select Transfer/Get from generator to replace the current file data on the PC with the
complete configuration data of the connected generator. This is an important preliminary step
before continuing with the complete configuration of the generator described next. It makes
sure that the current working data file (*.emd file) has software options and other basic features
compatible with the ones shipped with the generator.
3. In the menu bar, select View/Options to invoke the Options configuration screen. This screen
shows up software and hardware options available for the EPS generator (Figure 4.9).
The appearance of this screen could vary as a function of the generator configuration.
Figure 4.9 – Options Configuration Screen.
4. Use the configuration sheet of your generator to set up the appropriate options (Figure 4.9).
Software Options
a) Power Options Configuration
i) Power Type : Set the Power type as per the following configuration table:
Max kW 80 kW 85 kW 90 kW 100 kW
The power type configuration is fixed and specified at the factory. If this figure is changed,
whenever the compatibility check is done or data is exchanged with the PC, the default value
will be restored.
iii) High Performance RAD: Checking the “High Performance RAD” option box will
enable this functionality on the generator.
iv) Hidden AEC Back-Up: Checking the “Hidden AEC Backup” option box will enable
this functionality on the generator. Refer to Section 7.7.2 for information.
v) Falling Load: This option allows the generator to initiate an AEC-mAs exposure at a
higher tube current value, after which the mA drops as the exposure proceeds.
5. Use the Compatibility Check button to verify that the software options are
well configured on your generator.
Note: If the selected kW and mA combination or other software option is improperly set,
Epsilon PC Tools won’t allow you to continue the rest of the generator configuration
Hardware Options
a) Expansion Boards Configuration (Figure 4.10):
i. First setup the Expansion Motherboard configuration
(None, Expansion Motherboard or CPX Adapter).
ii. When an Expansion Motherboard is present on the generator, set up the attached
expansion boards by using the drop down lists of the four slot positions. The slot number
corresponds to the position in the generator basket, from left to right. Boards shall
inserted from the first free position available beginning from J4 (right-most) connector.
b) Other Options:
i) Biplane Configuration: For biplane configuration, check the
Biplane Enabled option box to enable the functionality.
ii) Disable Auto Gain: This option box should be left unchecked (Auto Gain enabled) for
most ABS types. It should be disabled (option box checked) only on certain types of
ABS systems using the “Rem AEC signal” as ABS gain source (Controller J14, pin 10)
combined with a very slow feedback ABS device (or with delayed ABS response).
6. Whenever the options configuration is modified, the Apply button (Figure 4.9) must be pressed
to acknowledge the new setup.
NOTE: If the options configuration is set up using the “Get from generator” method
(Transfer/Connect/ Get from generator), it is possible that the obtained Hardware setup does
not fit the generator setup accurately, or that the compatibility check does not take into account
certain expansion boards. Therefore, it is better to make it manually by choosing which board is present in
which slot.
7. Use the File menu to save the file on the PC. If necessary, send the newly configured data on the
generator (Section 4.2.3).
The Tube Setup module is used to transfer the X-ray tube chart of the connected tube(s) to the
generator before the X-ray tube could be calibrated (Section 5.2) and the generator be fully
functional. This step is of the utmost importance as it ensures the X-ray tube operating
parameters are within safe limits.
2. With Epsilon PC Tools in communication mode (Monitor/Transfer), select View and Tube to
invoke the X-ray Tube module (Figure 4.11). A default Tube setup loads at the first installation.
3. Click Import… button to choose an X-ray tube from the tube library of the
Epsilon PC Tools program. If your tube is not listed in the tube library, contact
emd technologies service organization for a technical advice.
The tube library (*.tub) uses file names very similar to the manufacturer tube models.
4. Verify that the tube data displayed by the program does correspond to the tube chart of the
manufacturer.
Important For any discrepancy in the tube chart characteristics, contact our service organization
before initiating the tube calibration. Emd technologies does not guarantee the accuracy or adequacy of
the tube charts and is not responsible for any tube damage resulting from misuse of tube charts provided
with the generator.
5. Save the file to make this tube chart parts of the current configuration file (*.emd) for your
system and send the data to the generator (Section 4.2.3).
A tube can be derated to extend its usable lifetime, to troubleshoot specific conditions or other
reasons. Derating any of the X-ray tube factors necessarily limits the generator’s X-ray factors and
therefore might reduce the exposure techniques kW. For example, if the filament current value of
the small filament is derated to 3.5 A and the requested tube mA requires a filament heating higher
than this value, the generator might either switch to the large filament (autofocal spot mode) or, else,
it will refuse the current technique by displaying the following operator message: “Tube Fil Limit”.
To derate the X-ray tube operative parameters, perform the following procedure:
1. Connect the generator to the PC and open the configuration file for your generator.
2. With Epsilon PC Tools in communication mode (Monitor/Transfer), select View and Tube to
invoke the X-ray Tube module (Figure 4.11).
b. kW percent: Use this option to reduce the kW allowed on one or both tube filaments.
c. Maximum Filament Current: Use this option to reduce the maximum tube current (A)
on one or both filaments.
Tube Data Usable range
Maximum kV 50-100%
Small/ Large Filament kW Output 50-100%
Small /Large Filament Maximum Current 50-100%
Boost & Brake Times 500 -15000 ms
4. Save the configuration file and send the data to the generator (Section 4.2.3). If necessary, repeat
the process for the X-ray tube number 2, if any.
5. Whenever changes are brought about on an existing tube setup, it is possible to save the
modified tube characteristics on the PC. This is done by using the Export feature in the X-ray
tube Setup screen (Figure 4.11). The new data (tube data and derating values) is saved on disk
with a .tub filename extension and can be used later on by the import feature of the Tube Setup
module for another generator configuration.
Up to six (6) workstations will have to be configured as a function of the connected equipment and
dedicated application. This is detailed in Sections 4.5.1 & 4.5.2.
3. Click on the W.S. 1 tab to configure the first workstation. Set the "Enable W.S." checkbox to
enable the workstation. Clear the checkbox to disable the workstation; then choose another
workstation to continue.
For Operator Console P/N 9777.31 only: Note that the following
associations are done between the workstation numbers (MODE in Figure
below) and their physical locations on the operator console. Workstation 5 is
not available with this operator console and should be disabled in PC Tools.
Workstations (MODE) on
Operator Console
P/N 9777.31
b. Default W.S.: Set the "Default Workstation" checkbox to load the current workstation on
the console at power on. Only one workstation can be defaulted.
c. Fluoro: To enable Fluoro on the current workstation, set "Enable Fluoro" checkbox
Note that Fluoro is available on Tube 1 only.
Note: When the "Default Fluoro" checkbox is set ( ), the console will be defaulted, on
that workstation, with the display of the Fluoro parameters (instead of RAD).
d. Tube: If a tube switch is fitted on the generator, assign a tube number to the current
workstation. Tube 1 (or Tube 2)
e. Anode speed: Specify the anode speed defaulted at the workstation selection. For cardiac
workstations configuration, it is highly recommended to select automatic anode speed since
cardiac exams usually imply high kW, or high mA.
High Speed (or Low, or Auto)
f. Continuance: Set the desired continuance timer. At the expiration of the timer, the anode will
be braked. Note that the timer is reset to zero each time irradiation is done
(up to 60 minutes)
g. Exposure switch:
Enable the exposure switch(es) to be used with current workstation (=> § 3.2).
Note: The console must be selected if the Hand-Switch option is fitted on the generator.
5. Assign the trigger input attached to the workstation (Section 3.3) and specify its activation mode:
Trigger mode: Start rising/falling, stop timer (i) Trigger source:
Start rising/falling, stop rising/falling (ii) Choose the input pin desired for each
Internal (S/W) (iii) workstation or select “Line” (iv).
i. Stopping the exposure using a timer is typical to most exposure modes. The exposure stops by
a counter (mAs or AEC) or timer.
ii. Stopping the exposure using a trigger’s edge is typically used in tomography applications.
iii. When Internal (S/W) trigger mode is selected, the Epsilon generator will perform exposures at
the operator selected frame rate and exposure parameters and no external trigger is required.
iv. When Line is used as a trigger source, the zero crossings of the mains supply are used to
synchronize the exposures and no external trigger is required.
v. The generator has an operator selected frame rate and will not permit exposures to be performed
any faster than that selected frame rate. If external triggers come at a slower rate than the
selected frame rate, exposures will follow the trigger rate but if triggers come at a faster rate that
selected, then some triggers will be missed.
AEC Devices
7. In the AEC Control zone, assign the AEC device working with the current workstation, if any.
According to the type of AEC sensor and the AEC feedback input location, make the following
selections:
ION # 1 (1) (Controller board, J13); Solid State Chamber # 4 (1, 2) (J3 connector)
ION # 2 (1) (Controller board, J8); Solid State Chamber # 5 (1, 2) (J4 connector)
Remote AEC (Do not select this option); Solid State Chamber # 6 (1, 2) (J5 connector)
ION # 3 (1, 2) (J1 connector); Phototube (PMT)(3) (J5 connector)
ION # 8 (1, 4) (J5 connector); Photodiode (PhD)(4) (J6 connector)
ION # 9 (1, 4) (J4 connector);
(1) For an AEC chamber, select the film/screen and input fields that will be allowed with the ion
chamber or solid state chamber:
Selection of film-screens; Slow ; Medium ; Fast ;
Selection of active fields; Left ; Center ; Right .
(2) On Universal chambers Interface board P/N 11002.000.
(3) On Fluoro PMT board P/N 9822.00:
Select the source for the PMT gain adjustment:
i. Potentiometers located on Fluoro PMT board
ii. Software (uses an internal D/A to set the gain value)
iii. External (J3B, pin 3 on Fluoro PMT board)
(4) On Photodiode Ion Chamber Fluoro board P/N 11030.000.
MULTI-ZOOM FUNCTION (For generators with High Performance
Pulsed Fluoro): If an ion chamber is to be used in conjunction with
selectable I.I. magnification sizes, the “Flat” film screen selection
must be done.
9. Click on the Frame Rate tab to set the RAD frame rate operation available for serial exposures.
i. Setup the frame rate using the horizontal selection bars (Figure 4.13, left). Up to seven (7)
frame rates are available. For optimum friendliness, enable only those frame rates that will
actually be used.
Note: Only one set of frame rate selections is provided for all serial workstations. This
configuration should be done once only (do not repeat it for all workstations).
ii. For remote selection or remote transmission of frame rate, assign the proper bit configuration
(up to 3 bits) for the communication (Figure 4.13, right).
Note: The generator uses the console to select the frame rate when the three bits are set
open. Therefore, do not set any frame rate value to all 3 bits unchecked (not set).
11. Use the File menu to save the workstation configuration on your hard disk and send the data to
the generator.
1. Make sure the Expansion board configuration is done in the Options setup (Section 4.3.1).
2. In the menu bar, select View and Inputs to invoke the Programmable Inputs configuration screen
(Figure 4.15).
5. For each output line interfaced with your equipment, specify the type of output and with which
workstation(s) it will be used (see Table 49). Select RAD and/or Fluoro applications, as
necessary.
i. Beware that the software configuration matches the hardware interfaced done in Chapter 3.
ii. For programmable outputs (Table 49, item 4), setup the output as a function of the different
generator states.
iii. For the frame rate transmission lines coding (Table 49, item 3), refer to Figure 4.13.
iv. If the Fluoro Photodiode board (PhD tab) is present on the generator,
set up the signal activation polarity and choose a voltage (4 to 24 V) for
the High Level output. Active low polarity will put a 0 VDC at the
output pin when the signal is active.
6. Save the data on your hard disk and send it to the generator (§ 4.2.3).
Both “Run” and “RUN-X” output will go true in a RAD sequence when all of the following conditions
are true:
1. Both Prep and RAD switches are pressed. 4. All Ready Interlock Inputs have allowed the
2. The anode is turning at the correct speed. Ready State.
3. The selected filament is at proper temperature. 5. All Start and X-Ray Interlock inputs are true.
6. No fault is detected.
Once true, a “Run” output will go false when any of Once true, a “Run-X” output will go false when any
the following condition occurs: of the following condition occurs:
1. EITHER the Prep or the RAD switch is
1. BOTH Prep and RAD switches are released.
released.
2. Any X-Ray Interlock goes false. 2. Any X-Ray Interlock goes false;
A false Ready or Start Interlock cannot stop a A false Ready or Start Interlock cannot stop a
RUN sequence in progress. RUN-X sequence in progress.
3. The number of exposures allowed in the Note: The RUN-X output stays TRUE after the
sequence has been performed. In the case of a number of exposures allowed in the sequence has
single exposure, the Run output goes false after been performed. Only the release of both exposure
the exposure. switches will disable the RUN-X output.
The sequence is halted (and the outputs toggled to false) by any generator fault.
For EPS RAD/Fluoro generators, perform the Fluoro Setup configuration as per the following
instructions.
PC TOOLS … VIEW
Done => Options
Done => Workstations
See Section 4.7 APR
Make sure the options, the tube setup and workstations
Done => Inputs
are configured before Fluoro setup is done. Done => Outputs
Here… Fluoro
See Section 5.7 AEC Chambers
Done => Tube
• ALARM TIMER: Set the Timer for the Fluoro alarm (up to 5 minutes).
• PULSE RATE: For optimum friendliness in Pulsed Fluoro, enable only those pulse rates that
will be used.
• MIN./MAX./STARTING KV: Use these settings to configure the kV range and starting kV in
Fluoro. The starting kV is used at the beginning of a new patient exam only.
• NOMINAL INPUT: Do not setup the nominal voltage at this moment. Refer to Section 5.5.
• TRIGGER: Adjust the trigger type (mode) and source (if any).
I. Trigger in Continuous Fluoro:
1. If no trigger is used, set the trigger mode to “Internal (Software)” (see below). The
Fluoro starts whenever the exposure ready conditions are met.
2. Continuous Fluoro can also be synchronized with a trigger provided by the camera or
other acquisition device. In this mode, Fluoro is usually started and terminated on a
signal front (Falling or Rising, see below).
3. In continuous Fluoro, it is not recommend to use a “Start rising/or falling, stop timer”
trigger type.
2. The generator has an operator selected frame rate and will not permit exposures to be
performed any faster than that selected frame rate. If external triggers come at a slower
rate than the selected frame rate, exposures will follow the trigger rate but if triggers
come at a faster rate that selected, then some triggers will be missed.
• ADDITIONAL PULSED FLUORO SETUP: If Pulsed Fluoro is fitted on your generator, perform the
following additional setups.
I. Select the Fluoro mA display mode (Figure 4.18): 1. Average mA; 2. Peak mA.
For the moment, do not adjust the ABS curves. Wait until Section 5.6 to program
the Fluoro kV/mA tracking curves.
For the moment, do not adjust the ABS stabilization. Wait until Section 5.5 to
program the Fluoro kV/mA tracking curves.
The Epsilon APR program uses a node-lattice network to create a hierarchic multi-level exams
database.
You can customize it the way you want, creating and editing nodes at any level. It is worth
remembering that every newly created node has its own set of parameters. Every node corresponds
to an exam. This results in a quick change of the set of parameters during the setup of an APR
exposure in normal utilization mode.
On a first installation, the number of exams corresponds to the number of workstations. The default
parameters are displayed and edited in the right-hand pane for every patient sizes (Figure 4.19).
The Exam setup of the right pane corresponds to the selected node of the APR arborescence
(here, Digital /Spine/LW Ribs, 3rd level node). Each node has its own set of exam parameters.
Figure 4.19 –
APR Programming screen showing a typical APR.
1. Connect the generator to the computer and open the PC Tools configuration file for the
generator.
4. Select New Child. The new node (titled “Exam N”) is placed under the workstation level, in
level position two. The new exam possesses the same APR parameters than its parent. Doing
"New Child" again will result in a new sibling, in level position three. If, instead, you want to add
a second level to the workstation, return to the first level, then use the “New Child” feature
again.
Naming an exam
5. To change the name of a node, select the exam, right-click and choose Rename (or simply use
the F2 key). The focus is brought inside the name zone, and you can now type over a new name
(do not use more than 10 characters).
Exam Deletion
6. The delete an exam, select it and press the delete key or use the right-click menu and choose
"delete".
Any deletion of an exam will erase all exams defined below at inferior levels!
Copy/Paste Exam
7. Use the “Copy and Paste” function to duplicate a full branch of the APR to another branch.
Exam Setup
8. For every exam (node), specify the different parameters that will be defaulted whenever this
exam is selected on the console. This is done by editing the Radiographic Parameters
Configuration pane shown in Figure 4.19.
i. Choose the default exposure mode. If “Disabled” option is selected, no exposure will be
allowed for that APR node. The disabled option can be used to disable the exposure switches
during the APR navigation by the operator.
ii. Select both the frame rate and run length values. If these parameters are not used in the
currently chosen exposure mode, they might come of use if the operator switches the
exposure mode during the exam setup.
iii. Setup the ion chamber (defaulted parameters upon an AEC exam with ion chamber).
iv. According to the selected exposure mode, choose the following tube loading factors by
default: kV, mA, mAs (and mAs ideal time (*)), irradiation time and optical density (**).
(*) mAs Ideal Time = Time value used in mAs exposure (the smallest the time counterpart of the mAs
factor is, the greatest the mA will be).
(**) Optical density = Seven clinically significant optical density corrections are provided with
the EC mode, varying from the standardized optical density N= 1. See Figure 4.20.
Note that all of the above setups are default parameters only. They can always
be overridden in normal utilization mode by the user.
9. Repeat the exam configuration process for every available node on all workstations.
Use the Copy/Paste feature of the right-mouse button to quickly reproduce an exam set at another place on the
APR tree.
10. Use the File menu to save the APR configuration on the PC. Send the data to the console.
The Console Setup module is used to adjust the contrast and brightness of the operator console.
1. Put the operator console in service mode and select Console/Setup (Figure 4.21).
2. Adjust the contrast and brightness of the console as desired (Figure 4.21-C). Adjust also the
sound level of the console's alarm and beeping. Press Save selection key to apply new settings.
A B C
1. Connect the generator to the computer and open the PC Tools configuration file for the
generator.
3. Set Date and Time: Use this button to synchronize the generator clock with the PC.
1. Invoke the service mode on the operator console and select Console/Language (Fig. 4.23).
A B C
2. Use the selection key shown in Figure 4.23-C to toggle the active language on the Epsilon
operator console. Press Save selection key to apply the new language selection.
1. Connect the generator to the computer and open the PC Tools configuration file for the
generator.
ii. Import a second language file from Epsilon PC Tools files. As a second language, the file
“FRENCH.SL” can be loaded for a ready to use French language.
4. Once a second language has been created or imported (and/or exported), it can be transmitted
to the operator console by saving the emd file (File=>Save or Save as) and transferring the data
to the console (Transfer => Connect=> Send to Generator).
• As long as the alphabet chosen is compatible with the local keyboard setting, the proper
characters should be displayed by both PC Tools and the Epsilon console.
• In Epsilon PC Tools, it is also possible to translate:
o Workstation names in the Workstation configuration screen;
o Interlock names in the Input configuration screen;
o Anatomical region names in the APR configuration screen.
This chapter describes the final steps before the generator can be operated normally and the user's acceptance
tests (Chapter 6) be performed.
IN THIS CHAPTER :
Most sections of this chapter require the production of X-rays. Make sure to apply
all safety precautions to protect personnel from X-ray radiation.
Make sure the EPS 45-80 High Voltage generator is installed and configured as per the
instructions of Chapters 2, 3 and 4.
REQUIRED MATERIAL
• One copy of the Installation Report (Section 9)
• The Epsilon PC Tools Program with :
o One PC compatible computer running Windows TM
o One RS-232 interlink cable or a standard USB printer cable,
depending on the Epsilon Console version.
• Standard field service tool kit
• One 3½ digit digital multimeter
• Optionally, one 20 MHz dual trace storage oscilloscope with 2 probes
• One mAs meter such as Keithley/Victoreen meters (or equivalent)
• One X-ray dose meter (reading up to 20 Roentgens/min.)
Perform the following verification and test before calibrating the X-ray tube for the first time.
Make sure the Anode Starter Board and stator configuration matches the rotor type
of the X-ray tube.
• For Epsilon Anode Starter board P/N 9962.xx, perform Section 5.1.1.1.
• For EPS Starter board P/N 11104.xxx, perform Section 5.1.1.2.
NOTE: For R and S standard rotor types only, verify the following dip-switch
configuration (SW1) on the Anode Stator board before doing any exposure. Refer
to the Configuration Table 50 below.
1. Make sure that the EPS Starter Board model fitted on your generator matches your tube rotor
type (Figure 5.1(b)).
EPS Starter board P/N 11104.001 : Low Speed Starter (most common stator types)
EPS Starter board P/N 11104.002 : High Speed Starter (most common stator types)
EPS Starter board P/N 11104.003 : Low Speed Starter, Low Impedance (Q & E Stators)
EPS Starter board P/N 11104.004 : High Speed Starter, Low Impedance (Q & E Stators)
2. Make sure that the configuration of the EPS Starter Board installed with your generator is
appropriate for your tube rotor type. Refer to the Configuration Tables 51 & 52 to adjust, if
necessary, the configuration of the stator type using SW1 dipswitch and the jumper cables (10)
provided with the board. This step requires that the board is first taken out of the generator
upper basket.
“XS-R”
F to P G to M A to R H to J B to S C to N
“TS1”
“DT1” B to P C to M Q to L I to N F to R G to J A to S H to K D to T
“IAE” F to P G to M I to N Q to L B to J C to R E to O D to S
“Q” B to P C to M D to Q E to N F to R G to J A to S H to K
“E” B to P C to M D to R E to N F to S G to J
Figure 5.2 –
Quick Fit Tabs (A to T) on
Anode Starter board P/N 11104.xxx
50 kV @ 10 mA x 200 ms
3. Trigger on the rising edge of channel 2 with a time base of about 40 ms/division.
4. Press the Prep/RAD switch. Verify the exposure parameters on the scope.
The X-ray tube calibration consists for the generator in finding the exact values of filament current
that will produce the different tube mA stations, for the series of kVs. At the same time, the
calibration process ensures that the X-ray tube operating parameters are kept within safe limits.
1. On the operator console, invoke the service mode by pressing and holding down WS1 and
power on keys. Release WS1 key when the Service mode displays on the console (Figure 5.3-a).
2. Select the Tube Calibration Menu. Verify the veracity of the current tube chart data using
Epsilon PC Tools (Tube Setup menu).
A B C D
3. Make sure the tube anode rotation works well (Figure 5.3-b). Select the focus (Figure 5.3-c).
4. Press Prep & RAD footswitch to start the calibration for the selected filament (Figure 5.3-d).
The calibration for each filament is done in two steps, with no user intervention, and takes about
5 minutes. The Prep & RAD switches can be released at any time and the calibration be
concluded later by actuating the exposure switches anew.
5. At the end of a filament calibration, the “Calib. Done” message displays together with the total
exposure time for the calibration. Repeat the process for the other focus.
6. When the calibration is completed for both foci, perform the Exposure test of Section 5.3.
The following exposure tests must be done after an X-ray tube calibration. If a scope is available,
verify the kV and main gate waveforms.
4. Press the Prep/RAD switch. Verify the exposure parameters on the scope.
2. Make an exposure and compare the exposure mAs post-display on the operator console with the
reading of the mAs meter. The mAs value should be 200 ± 10.2 mAs.
A B C D
Figure 5.5 – Fluoro Maximum Entrance Exposure Rate Adjustment Module in Service Mode.
ABS Feedback Gain Adjustment: The end of a frame in Pulsed Fluoro is detected when the
voltage ramping at the ABS feedback input reaches the nominal input value. In continuous Fluoro, a
voltage level is taken into account instead of a voltage ramp.
To adjust the Fluoro image stabilization, perform either procedure 5.5.1 or procedure 5.5.2
according to the equipment interfaced to the generator.
− Use § 5.5.1 to adjust the Fluoro Dose stabilization in Continuous or Standard Pulsed Fluoro;
− Use § 5.5.2 to adjust the Fluoro Dose stabilization in High Performance Pulsed Fluoro.
2. Select Fluoro calibration, Dose Stabilization and Low Level (Figure 5.6, B and C).
3. Press the Low Level Fluoro footswitch and adjust the Fluoro image using the appropriate
Fluoro stabilizer device (Table 53):
• Using a D/A device: With the Epsilon Console in service mode, manually set the gain
of the PMT from 1 thru 100%. Press [Save] selection key when done (Figure 5.6-D).
• Using a potentiometer: Turn the proper potentiometer on the Fluoro PMT board.
Increasing (clockwise) the phototube sensitivity will decrease the image brightness.
Decreasing (counter-clockwise) the phototube sensitivity will increase the image
brightness.
4. If High Fluoro level is fitted on your generator, return to screen of Figure 5.6-C, select High
Fluoro Level and repeat step 3.
3. Make sure photodiode is selected for the Fluoro feedback device. Set the gain level(s) for the
feedback device by adjusting the nominal input with the corresponding cursor (Low & High
levels, if applicable).
FLUORO
1.
4.
3.
2.
Figure 5.7 – Fluoro Dose Stabilization Adjustment with a Photodiode ABS Device.
1. Connect the generator to the PC and launch the Epsilon PC Tools program.
2. Open the configuration file for your generator. Use View menu to invoke the Fluoro
configuration screen. Click on General Fluoro Setup tab.
3. Use a field procedure to adjust the nominal input voltage for Low Level Fluoro that will
provide optimum Fluoro image brightness. Save the data and send it to the generator.
4. Do Fluoro and verify the image brightness. Repeat the procedure for High Level Fluoro, if fitted
on the generator.
1.
FLUORO
4.
2. 3.
Nominal Input
From 0.1 to 10 V
Figure 5.8 – Fluoro Dose Stabilization Adjustment with an Ion Chamber in Pulsed Fluoro.
In Fluoro ABS, the Epsilon RF generator uses a kV to mA tracking curve to set the Fluoro mA as a
function of the kV variation. Four (4) different curves are provided, ABS1 thru ABS4, to be
customized as per your Fluoro application (Figure 5.9). The quality of Fluoro contrast can be judged
by the amount of blacks and whites on the picture and is determined by the proportion of mA
forming the fluoroscopic X-ray beam.
Each ABS curve is built from five points from 40 to 125 kV on the X-axis, and percent of maximum
mA on the Y-axis. Of the five kV points, only the middle 3 kV stations can be edited (40 and 125
kV stations are fixed). The maximum mA allowed starts at 100 % at the maximum 125 kV, and can
be raised or lowered at the inferior kV stations. The program will not let you programming a
kV/mA curve too steep that would compromise the Fluoro stabilization.
1. Connect the generator to the PC and launch the Epsilon PC Tools program.
2. Open the configuration file for your generator. Select View/Fluoro/ABS Curves Setup window
(Figure 5.9).
3. For ABS1, select the three middle kV stations to set the points of inflection of the curve.
4. For every kV station, set the percent of the maximum mA. As you drag the mA % horizontal
cursor (or click on the right/left arrows for smaller increments), the curve is redrawn
automatically in the Fluoro kV-mA tracking curves zone.
5. Repeat the drawing process for every ABS curve that you will use in Fluoro.
ABS2 through ABS4 curves can be disabled, if not used. A disabled ABS curve will not be
available on the operator console.
The following sections describe how to setup the ion chamber and calibrate the AEC mode using
the Epsilon PC Tools program.
3. Refer to the manufacturer specifications to setup the ion chamber control voltage and signal
polarity (Epsilon PC tools: View => AEC Chambers/Hardware tab, Figure 5.10, left):
• Control Voltage: Operating voltage used for the field selection and integrator reset of the
ion chamber. The control voltage (V-Out High) can be adjusted from 4 to 24 VDC.
• Signal polarity: Depending on the logic used for the activation of the ion chamber reset
and field select signals, choose the proper polarity using the Reset Polarity and Field Enable
Polarity radio buttons. Refer to Figure 5.11 for two examples of field selection configuration.
Active-Low Field Select Polarity; 14.5 VDC 14.5 VDC 14.5 VDC Not used
Control Voltage set to 14.5 VDC
0 VDC 14.5 VDC 14.5 VDC
14.5 VDC 0 VDC 14.5 VDC
Figure 5.11 – Field Selection Table for two (2) different Ion Chamber Setups. Note that the above values are
given as examples only and do not correspond to recommended figures for a default ion chamber.
The AEC calibration requires that the voltage level should be Examples of AEC Gain
adjusted for a neutral density and no film correction (see Nominal Input AEC Gain
procedure in Section 5.7.2.2). The nominal input figure applies 0.1 V 1/10
1.0 V 1
independently for each of the three film-screen combinations.
10 V 10
For an ion chamber, the effect of the film-screen correction is additive to the effect of the AEC
feedback gain. See Table 54 and description in the following paragraph (B. Film-Screen Correction).
The compensation effect of the correction factor applies in shortening or lengthening the end of the
exposure in the following manner:
Examples are given below in Table 54. Refer to Section 5.7.2.2 for the adjustment procedure of the
film-screen correction.
(*) The combined action of AEC gain and film-screen correction is effective up to 10 VDC. For AEC
calibration superior to that level, the exposure will still end at 10 VDC.
Equipment Required
- A clean flat-bottom plastic bucket, with a minimum diameter of 6”, which can be filled with
up to 14" of water. The bucket must be leveled.
- An X-ray densitometer capable of measuring optical film density from 0.5 to 2.5 with accuracy
of ± 0.05 O.D.
Procedure
1. Connect the generator to the PC and launch the Epsilon PC Tools program.
2. Open the configuration file for your generator. Use View menu to invoke the AEC Setup
module (Figure 5.10).
i. Select the proper ion chamber (# 1 or 2) and choose one of the three film-screen speeds.
ii. Set the nominal input voltage for that film-screen to 5.0 V.
iii. Set the correction factor for 81 kV at 0 (no correction).
iv. Set the optical density to “N” (optimum).
6. Place the water phantom (filled with 3" of water) in the center field of the ion chamber, with a
SID of about 36".
7. With no X-ray film and a strong collimation, perform an A.E.C. exposure and adjust the water
level inside the bucket so that the exposure time falls between 100 and 300 ms.
8. Insert the film-screen cassette and repeat the same A.E.C. exposure with the same water level
used in the previous step. Process the film and verify that the optical density of the exposed film
corresponds to the desired density (ex. 1.0 ± 0.1).
9. If the proper optical density is not achieved, adjust the Nominal input voltage for the current
film-screen speed. Save the data and send it to the generator. Repeat the same 81 kV-exposure.
Note: Once the desired optical density is obtained for the 81-kV exposure with the current film-
screen speed, the remaining eleven kV stations will be adjusted using the film-screen factor only
(from -80 to + 130), and Nominal input voltage should not be moved.
Ion Chamber Gain Adjustment: If the density correction is below 120 % or above 149 %, the gain of the
ion chamber must be adjusted. The ion chamber gain adjustment must be performed once only.
To adjust the gain of the ion chamber, repeat the same A.E.C. exposure at 81 kV and the factor of 0 % and
change the gain value of the ion chamber until the desired optical density is achieved.
10. With the same water level used in step 7, perform an A.E.C. exposure at 42 kV and find the
appropriate correction factor (from -80 to + 130) until the desired optical density of the exposed
film is achieved.
Remember to save the data and send it to the generator before doing an exposure whenever the
film-screen correction factor is adjusted!
Hint: Use the following formula as indicative means to correct the density % for a given kV station.
12. Repeat steps 2 thru 10 for a second film-screen speed, if used with the current ion chamber.
Adjust first the Nominal input voltage for the 81-kV exposure, and then, proceed with all kV
stations by adjusting only the film-screen factor. In step 4, don't forget to change the film-screen
speed to the new one.
13. Repeat the procedure for the last film-screen speed, if necessary. Repeat the procedure for the
second ion chamber, if any.
This section applies only to generators not using an Ion chamber as AEC sensor.
On certain workstations, the generator might be interfaced with a photomultiplier tube to control
the image brightness in AEC exposures. The image brightness is in direct relation with the gain of
the photomultiplier tube (PMT) controlled by the high voltage supply adjusted by a dedicated
potentiometer or a D/A device on the Fluoro PMT board.
3. Press RAD and adjust the image brightness using either i) a potentiometer on the Fluoro PMT
board (Figure 5.12) or ii) a gain set by a D/A set in service mode (console).
i) Turn the proper potentiometer on the Fluoro PMT board.
i. Increasing (clockwise) the phototube sensitivity will decrease the image brightness.
ii. Decreasing (counter-clockwise) the phototube sensitivity will increase the image
brightness.
ii) Use the Epsilon Console in Service mode to set the proper gain of the D/A device.
The testing procedure applies to the final generator, that is, to the generator that has been assembled,
installed, programmed and calibrated as per the instructions given in this manual, Sections 2 thru 5.
PRE-REQUISITE CHECKUP
• Make sure the Epsilon generator is installed as per Chapters 2 & 3 of this manual.
• Make sure the Epsilon generator is setup and calibrated as per Chapters 4 & 5 of this
manual.
• Make sure the Installation Report is completed as per Chapter 9 of this manual.
3. Verify that :
The X-ray tube selection is functional; OK ; NO
The focal spot selection is functional; OK ; NO
The APR selection is functional. OK ; NO
4. Choose a serial exposure mode workstation:
a) Verify that the frame rate selection is functional; OK ; NO
b) Verify that the run length selection is functional. OK ; NO
5. Set the console in 2 or 3-pt mode:
a) Verify that mAs key toggles between 2-point & 3-point mode; OK ; NO
a1. Verify Parameter keys up & down action:
kV, mA and time in 3-pt mode and mAs in 2-pt mode. OK ; NO
a2. Verify the mA, mAs, ms and s indicators. OK ; NO
b) Verify that AEC key toggles between AEC and 2/3-point mode. OK ; NO
6. Set the console in AEC mode:
a) Verify that the AEC indicator lights green; OK ; NO
b) Verify that the ion chamber setup keys are functional; OK ; NO
c) Verify the density up & down keys action. OK ; NO
7. Press the power off key (the generator must turn off). OK ; NO
Initials: _____
Test Equipment
Use the following calibrated measurement devices:
• Non-invasive or invasive kV meters
• mA/mAs meter
• X-ray Dose Meter
6.2.1 kV Accuracy
Acceptance Test Logbook (2/11)
1. Perform the following 3-point exposures (3) using the loading factors of Table 55
Table 55 and verify the accuracy of the tube kV (intermittent mode). The
error of the measured X-ray Tube Voltage of the different loading factors Passed
combination shall not be greater than ± (5% + 2 kV). Failed
2. Do Fluoro using the loading factors of Table 56 and verify the accuracy of Table 56
the tube kV in continuous mode. The error of the measured X-ray Tube
Voltage of the different loading factors combination shall not be greater Passed
than ± (5% + 2 kV). Failed
Initials: _____
6.2.2 mA Accuracy
Initials: _____
Initials: _____
Initials: _____
This test verifies the reproducibility of the radiation output of the generator.
Test Equipment
The following material is required to
X-Ray Tube
perform Section 6.3.
• X-ray Dosimeter
• Aluminum Attenuator
• Calculator Collimator
SID
Test Setup (Figure 6.1) and Protocol 135 cm Al Attenuator
1. Set the SID to about 135 cm.
2. Make 10 successive irradiations in less
than an hour, at the loading factors of
Table 61, column A. Do the exposure
shots at the same focal spot.
Ion Chamber (used for AEC
3. To simulate the presence of a patient shots only)
during the test, use an aluminum Radiation Probe
attenuator of sufficient size to intercept
the whole of the X-ray beam. Figure 6.1 – Setup for Sections 6.3 thru 6.6.
See Table 61 for attenuation thickness.
4. Repeat the test three more times, using the loading factors and attenuation thicknesses of
columns B, C & D.
NOTE: If the linearity tests of section 6.4 are intended, it is better to proceed with the
following tests sequence: A, B, C, E, D and F for more straightforward manipulations. These
tests correspond to Result Tables 6A, 6B, 6C, 7A, 6D and 7B.
Initials: _____
∑ (X − X )
2
Exp. shot 2 i =1
i
n-1
Exp. shot 3
Exp. shot 4
s=
Exp. shot 5
Exp. shot 6
Reproducibility
Exp. shot 7
n
∑ (X − X )
2
Exp. shot 8 i =1
i
n-1
C=
Exp. shot 9 X
Exp. shot 10
∑ (X − X ) c=
2
X = i
=
i =1
If c < 0.05 => Test Passed If c > 0.05 => Test Failed
Initials: _____
∑ (X − X )
2
Exp. shot 2 i =1
i
n-1
Exp. shot 3
Exp. shot 4
s=
Exp. shot 5
Exp. shot 6
Reproducibility
Exp. shot 7
n
∑ (X − X )
2
Exp. shot 8 i =1
i
n-1
C=
Exp. shot 9 X
Exp. shot 10
∑ (X − X ) c=
2
X = i
=
i =1
If c < 0.05 => Test Passed If c > 0.05 => Test Failed
Initials: _____
∑ (X − X )
2
Exp. shot 2 i =1
i
n-1
Exp. shot 3
Exp. shot 4
s=
Exp. shot 5
Exp. shot 6
Reproducibility
Exp. shot 7
n
∑ (X − X )
2
Exp. shot 8 i =1
i
n-1
C=
Exp. shot 9 X
Exp. shot 10
∑ (X − X ) c=
2
X = i
=
i =1
If c < 0.05 => Test Passed If c > 0.05 => Test Failed
Initials: _____
∑ (X − X )
2
Exp. shot 2 i =1
i
n-1
Exp. shot 3
Exp. shot 4
s=
Exp. shot 5
Exp. shot 6
Reproducibility
Exp. shot 7
n
∑ (X − X )
2
Exp. shot 8 i =1
i
n-1
C=
Exp. shot 9 X
Exp. shot 10
∑ (X − X ) c=
2
X = i
=
i =1
If c < 0.05 => Test Passed If c > 0.05 => Test Failed
Initials: _____
This test verifies the linearity of the radiation output of the generator.
Test Equipment
The following material is required to perform Section 6.4. Refer also to Figure 6.1.
• X-ray Dosimeter
• Aluminum Attenuator
• Calculator
Protocol
1. Use the same setup than Section 6.3.
2. Make 10 successive irradiations in less than an hour, at the loading factors of Table 66, column
E. To simulate the presence of a patient during the test, use an aluminum attenuator of sufficient
size to intercept the whole of the X-ray beam.
3. Repeat the test using the loading factors and attenuation thicknesses of column F.
Table 66- Tests for verifying linearity (see also Table 61)
Test setting E F
X-ray Tube Voltage 50% of highest 80% of highest
X-ray Tube Current or Two mA stations above Two mA stations above
Current Time product test C test D
Irradiation Time Between 0.01 s and 0.32 s for the two settings
Aluminum Attenuation 24 mm Al 40 mm Al
Result Calculations
X1 X2
The Absorbed Doses for the 10 exposure shots shall not differ by mAs1 mAs2
more than 0.1 for the combination of loading factors of Table 66 < 0.1
X1 X2 (
(Linearity tests E & F). See opposite formula. (mAs1+ mAs2
Initials: _____
Exp. shot 1
Exp. shot 2
Exp. shot 3
Dose/mAs (Test C): Dose/mAs (Test E):
Exp. shot 4
XC XE
mAs1 = mAs2 =
Exp. shot 5
Exp. shot 6
Linearity "L" :
Exp. shot 7
Exp. shot 8 XC − XE
mAs1 mAs2
< 0.1
XC + XE
Exp. shot 9 mAs1 mAs2
Exp. shot 10
Xe = mAs2 = L=
If "L" < 0.1 => Test Passed If "L" > 0.1 => Test Failed
Initials: _____
Exp. shot 1
Exp. shot 2
Exp. shot 3
Dose/mAs (Test D): Dose/mAs (Test F):
Exp. shot 4
Xd Xf
mAs1 = mAs2 =
Exp. shot 5
Exp. shot 6
Linearity "L" :
Exp. shot 7
Exp. shot 8 Xd − X f
mAs1 mAs2
< 0.1
Xd + X f
Exp. shot 9 mAs1 mAs 2
Exp. shot
10
Xf = mAs2 = L=
If "L" < 0.1 => Test Passed If "L" > 0.1 => Test Failed
Initials: _____
Use the following test to demonstrate the constancy of the AEC mode during direct radiography.
Test Equipment
The following material is required to perform Section 6.5.
• 12 identical films and cassette; or an X-ray dosimeter(*)
• Water phantom (or equivalent)
• Ion Chamber compatible with Epsilon generator
Protocol
Acceptance Test Logbook (8/11)
Test performed by: ______________________________
Date: _________ Beginning of Test Time: _________ End of Test Time: _________
Initials: _____
Table 70- Dose Constancy at same Phantom Thickness and varying Tube kVs
Phantom Exp. O.D. Diff. O.D.
Tube kV O.D.
Thickness No. versus O.D. Avrg differs by:
80 15 1 ______ ______
60 15 6 ______ ______
100 15 7 ______ ______
120 15 8 ______ ______
Compliance ± 0,15
Optical Density Average = _____
Test Passed Test Failed
Table 71- Dose Constancy at same Phantom Thickness versus Adjacent Tube kVs
Phantom Exp. Adjacent Exp. O.D.
KV No.
O.D. No.
O.D.
Thickness kV differs by:
60 15 cm 6 _____ 80 1 _____ _____
Compliance ± 0.15
Test Passed Test Failed
Compliance ± 0.20
Test Passed Test Failed
Initials: _____
Table 73- Dose Constancy as a Function of Different kVs and Phantom Thicknesses
Phantom Exp. O.D. Diff. O.D.
Tube kV O.D.
Thickness No. versus O.D. Avrg differs by:
60 10 cm 9 ______ ______ ______
60 15 cm 6 ______ ______ ______
80 15 cm 1 ______ ______ ______
80 20 cm 11 ______ ______ ______
100 15 cm 7 ______ ______ ______
100 20 cm 12 ______ ______ ______
120 10 cm 10 ______ ______ ______
120 15 cm 8 ______ ______ ______
Compliance ± 0.20
Optical Density Average = _______
Test Passed Test Failed
Initials: _____
2. Repeat the procedure to measure the Fluoro surface dose at 125 kV, ____ R/min ;
maximum pulse rate in High level.
3. If Image Enhancement mode is used, perform Fluoro at every pulse rate
inferior to the maximum and verify that the dose measured does not
exceed the dose measured at maximum pulse rate
• For Low level Fluoro OK ; NO
OK ; NO
• For High level Fluoro
Initials: _____
IN THIS CHAPTER :
7.1 WARNING ..................................................................................................................................................................................................... 7-3
7.2 EXPOSURE INHIBITION ........................................................................................................................................................................ 7-3
7.2.1 Exposure Limits ..................................................................................................................................................................................... 7-3
7.2.2 Interlock Conditions.............................................................................................................................................................................. 7-3
7.3 FAULT CONDITIONS ............................................................................................................................................................................... 7-4
7.3.1 Abnormal End of Exposure................................................................................................................................................................. 7-4
7.3.2 Operator End of Exposure .................................................................................................................................................................. 7-4
7.3.3 Alarms...................................................................................................................................................................................................... 7-5
7.4 OPERATOR MESSAGES........................................................................................................................................................................... 7-5
7.4.1 Informative Messages ............................................................................................................................................................................ 7-5
7.4.2 Inverter Problem Message / Power Reduction Mode...................................................................................................................... 7-6
7.4.3 Messages and Meanings ........................................................................................................................................................................ 7-7
7.5 DIAGNOSTIC FEATURES ..................................................................................................................................................................... 7-13
7.5.1 Power On Self-Test ............................................................................................................................................................................. 7-13
7.5.2 Operator Console’s Service Mode..................................................................................................................................................... 7-13
7.5.2.1 Invoking the System Monitor in Console’s Service Mode..........................................................................................................7-14
7.5.2.2 Anode Monitoring ............................................................................................................................................................................7-15
7.5.2.3 IGBT Code ........................................................................................................................................................................................7-15
Resetting the Generator/Basket Status ..............................................................................................................................................7-15
7.5.2.4 Software Revision .............................................................................................................................................................................7-18
7.5.2.5 History Log ........................................................................................................................................................................................7-18
7.5.2.6 Three-Phase Inverter Board Register.............................................................................................................................................7-19
I- Inverter Board Status Register Updating........................................................................................................................................7-19
II- Inverter Board Power On Self-Test ..............................................................................................................................................7-20
III- Inverter Board Test on PREP ......................................................................................................................................................7-20
7.5.2.7 Statistics ..............................................................................................................................................................................................7-21
7.5.2.8 User’s Monitor Mode .......................................................................................................................................................................7-22
7.5.3 Epsilon PC Tools’ Diagnostic Mode ................................................................................................................................................ 7-22
7.5.3.1 Monitor Module ................................................................................................................................................................................7-22
7.5.3.2 History Log ........................................................................................................................................................................................7-25
7.5.3.3 Tube Calibration Viewer ..................................................................................................................................................................7-27
7.5.3.4 Generator Calibration Module ........................................................................................................................................................7-28
7.5.4 Diagnostic Points ................................................................................................................................................................................. 7-30
7.5.4.1 Controller Board P/N 9798.01 .......................................................................................................................................................7-30
Test Points: .............................................................................................................................................................................................7-31
Typical Waveforms: ...............................................................................................................................................................................7-32
LEDs on Controller board: ..................................................................................................................................................................7-33
Input Voltage Reference & Threshold – J9 Connector: ..................................................................................................................7-33
7.5.4.2 3-Phase Inverter Boards P/N 9736.01 ..........................................................................................................................................7-33
Generator Power On LED:..................................................................................................................................................................7-33
7.5.4.3 Optional Fluoro Boards ...................................................................................................................................................................7-34
7.5.4.4 Anode Starter Board.........................................................................................................................................................................7-35
7.5.4.5 Optional Cable Discharge Interface Board P/N 11206.000 ......................................................................................................7-36
7.6 REPLACEMENT OF HARDWARE AND SOFTWARE.................................................................................................................. 7-37
7.6.1 Three-Phase Inverter Board & EPS Starter Board Replacement ................................................................................................. 7-38
7.6.2 Generator Basket Replacement.......................................................................................................................................................... 7-42
7.6.2.1 Preliminary Step (Generator Configuration Data Backup).........................................................................................................7-42
7.6.2.2 Replacement Procedure ...................................................................................................................................................................7-42
7.6.2.3 Firmware Upload and Final Verification.......................................................................................................................................7-48
7.6.3 Optional Boards Replacement ........................................................................................................................................................... 7-49
7.6.3.1 Fluoro Board Replacement..............................................................................................................................................................7-49
7.6.3.2 Anode Starter Board : Refer to Section 7.6.1................................................................................................................................7-49
7.6.3.3 Relay Board and Miscellaneous boards..........................................................................................................................................7-49
7.6.4 Fuse and Battery Replacement ........................................................................................................................................................... 7-50
7.6.4.1 Battery Function and Replacement ................................................................................................................................................7-50
7.6.4.2 Fuse Replacement .............................................................................................................................................................................7-51
Continued =>
7.1 WARNING
Workstations have their own set of programmable interlocks to prevent unsafe exposure. An
interlock condition always results in the invalidation of the exposure process.
Interlock conditions are apparent by the lighting of
the interlock lamp and by the display of an interlock
message on the operator console (Figure 7.1).
Figure 7.1 –
Fault (left) and Interlock (right) conditions in RAD
To reset a fault, the operator either presses the [Fault Reset] key or re-activates the exposure
switch.
If backup mAs or backup time happens before normal phototiming detection, the generator
aborts the exposure immediately and displays the "END of EXP. : B.U.T." message in the graphic
display. In this condition, the generator can only be reset by pressing the [Fault Reset] key.
NO FAULT
Blinking of parameter’s decimal point.
Actual parameter value after exposure
EXPOSURE FAULT
1. [Fault Reset] key’s LED turns on.
Abnormal termination of exposure
Fault LED and post-display of parameters
7.3.3 Alarms
“Anode H.U. over 80%” alarm Pulsed tone (2 beeps per second)
(1) This alarm is heard in the X-ray room only and requires the presence of an external buzzer.
(2) The alarm is shut off by pressing the Alarm Reset key at the Epsilon Operator Console.
Alarms in RAD
“80 % of maximum run length” alarm Pulsed tone (2 beeps per second)
“Anode H.U. over 80%” alarm Pulsed tone (2 beeps per second)
(3) The tone lasts for the duration of the X-ray exposure (1/4 second minimum).
To help understanding the nature and severity of a fault, refer first to Table 77 (Operator Messages).
Refer also to the diagnostic mode of the Epsilon PC Tools, Section 7.5.3.
When this condition occurs, usually at power on, the generator is completely disabled and the only
possible action available to the operator is the acknowledgement of the inverter fault by doing a
power off/on cycle. The acknowledgment of an inverter problem by cycling the power is a
palliative means only: the generator is available for x-rays but it will probably operate at reduced
power output capacity. In that mode, the message “Power Reduced” displays (Figure 7.2, right) and
will continue to be displayed each time the generator is cycled off/on again. To get rid of this
problem and restore the generator full power, a service intervention is necessary (see § 7.5.2.6).
Inverter Problem
Figure 7.2 – Inverter N°6 Problem (left) followed by a “Power Reduced” message (right).
Refer to Section 7.5.2.6 for more information.
UUUU
UUU
1. Cathode H.V. Cable; Small Fil.
3. X-ray Tube;
UUUU
UUU
Large Fil.
2. At the end of the self-test, the console ends up into normal display mode with the default
workstation and tube loading factors. Verify that no message displays on the console.
− If a message is reported, make sure it does not refer to a generator fault. The generator
will be able to expose if the Fault and Limit lamps are off.
− If a fault is latched and the generator cannot be reset to normal operation, refer to the
Operator Message Table 77 or enter the Monitor Mode (Sections 7.5.2 & 7.5.3).
In service mode, invoke the monitor mode by pressing System and Monitor keys (Figure 7.4). The
different options are discussed in the following sections.
B B
C
A
E
2. Select "System" and "Anode” to call upon the Anode monitoring screen
Press the LOW key to boost the Press the HIGH key to boost
anode to low speed. the anode to high speed.
Press the Brake key to apply a DC
brake on the anode.
The IGBT code displayed on top of many service mode screens can be used to have a IGBT
picture of the status of different parts of the generator and X-ray tube path. 6 0 0 0
• I: 3-Phase Inverter board code (Inverters & Dummy Inverters). This code gives a quick
indication of the Inverter board status (Table 78). The Inverter board’s status shall be
manually updated, for example after replacement of an Inverter board.
NOTE: A more complete picture of the 3-Phase Inverter board functionality is
available in a different register, as described in Section 7.5.2.6. This hardware register
will give you the exact status of every board installed in the six Inverter board slots.
• G: The generator status includes the rotor controller, tube switch, console and internal
cables; see Table 79-a. This status can be reset in the Hardware register module,
described below.
• B: The electronic controls status (referred to as the basket) includes the Power Supply,
Controller board and Expansion boards; see Table 79-b. This status can be reset in the
Hardware register module, described below.
• T: The X-ray tube status (open filament(s)); see Table 80.
Table 78- 3-Phase Inverter Board Possible Conditions as per [IGBT] code
Code Description Status
0 No good Inverter boards in the generator FAULT
1 One good Inverter board in the generator
2 2 good Inverter boards in the generator
3 3 good Inverter boards plus 3 Dummy boards in the generator NICE
4 4 good Inverter boards plus 2 Dummy boards in the generator
5 5 good Inverter boards plus one Dummy board in the generator
6 6 good Inverter boards in the generator
7 3 good Inverter boards in the generator SAD
One or more Inverter board(s) with a blown fuse
Number of Dummy board(s) in the generator between 0 and 2
8 4 good Inverter boards in the generator
One or more Inverter board(s) with a blown fuse
Number of Dummy board(s) in the generator between 0 and 2
9 Five good Inverter boards plus one Inverter board with a blown fuse
A 3 good Inverter boards plus between 1 and 3 Inverter(s) in bad condition CHECK
Between 0 and 2 Inverter board(s) with a blown fuse
Between 0 and 2 Dummy board(s)
B 4 good Inverter boards ; One or more Inverter board(s) in bad condition
0 or one Inverter board(s) with a blown fuse
0 or one Dummy board(s)
C 5 good Inverter boards plus one Inverter board in bad condition
D 3 good Inverter boards plus at least one Inverter board in uncertain condition FAULT
E 4 good Inverter boards plus at least one Inverter board in uncertain condition
F 5 good Inverter boards with one Inverter board in uncertain condition
Generator Possible Status as per [IGBT] code Basket Possible Status as per [IGBT] code
Code Description Code Description
0 All functional 0 All functional
1 Stator drive connection error 1 Battery weak
2 Internal cable connection error 2 Power Supply filament trouble
3 Conditions 1 + 2 3 Conditions 1 + 2
4 Tube switch connection problem 4 Expansion board trouble(s)
5 Conditions 1 + 4 5 Conditions 1 + 4
6 Conditions 2 + 4 6 Conditions 2 + 4
7 Conditions 1 + 2 + 4 7 Conditions 1 + 2 + 4
8 Console H/W problem 8 User Power Supply problem
9 Rotor Controller H/W Fault 9 Controller Board unusable
A Options incompatibility A Power Supply Board unusable
B Conditions 9 + A B Conditions 9 + A
C Tube Switch or Cable Discharge problem C Expansion Board(s) unusable
D Conditions 9 + C D Conditions 9 + C
E Conditions A + C E Conditions A + C
F Conditions 9 + A + C F Conditions 9 + A + C
Figure 7.9 – In Service Mode, select System Setup, Monitor (Figure 7.4), Hardware and Inverters
to invoke the Inverters register screen. The status of the 3-Phase inverter boards can be changed
using the selection keys 1 thru 6 corresponding to the Inverter board positions in the generator.
To replace a 3-Phase Inverter board or change a blown fuse, refer to the board
or fuse replacement procedure given in Section 7.6.1 or 7.6.4.2 respectively. Do
not try to replace a board or a fuse without reading and understanding that procedure.
Hint: The Inverter Board Status Register can also be reset using Epsilon PC Tools (Section 7.5.3.1).
Table 81- Possible actions done by the generator after every Inverter Power-On Test
Power On Compared against EEPROM Status
Test Result
STOP DUMMY FUSE BAD OK
STOP 1. No change 1. Write STOP 1. Write STOP 1. Write STOP 1. Write STOP
2. Stop 2. Stop 2. Stop 2. Stop 2. Stop
Dummy 1. No change 1. No change 1. Write STOP 1. Write STOP 1. Write STOP
2. Stop 2. Continue 2. Stop 2. Stop 2. Stop
IfR 1. No change 1. Write STOP 1. No change 1. No change 1. Write FUSE
2. Stop 2. Stop 2. Continue 2. Continue 2. Stop
IFr 1. No change 1. Write STOP 1. Write BAD 1. No change 1. Write BAD
2. Stop 2. Stop 2. Stop 2. Continue 2. Stop
IFR 1. No change 1. Write STOP 1. Write BAD 1. No change 1. No change
2. Stop 2. Stop 2. Stop 2. Continue 2. Continue
Legend:
Dummy : A Dummy board is installed
I fuse rail : 3-Phase Inverter board installed, bad fuse (f), bad rail (r) supply
I fuse Rail : 3-Phase Inverter board installed, bad fuse (f), good rail (R) supply
I Fuse rail : 3-Phase Inverter board installed, good fuse (F), bad rail (r) supply
I Fuse Rail : 3-Phase Inverter board installed, good fuse (F), good rail (R) supply
Continue : Do not display Operator message and go to normal mode
No change : Do not change the EEPROM current status
Stop : Disable all exposure switches and display "Inverter Problem # _" message.
Write BAD /FUSE / STOP : Change the EEPROM current status for BAD / FUSE or STOP
7.5.2.7 Statistics
The generator keeps statistics of the number of exposures, tube arcs occurrence, Inverters
temperature and certain operational limits. The data is stored in a FLASH memory that can be
visualized in service mode (Figure 7.10) and reset in PC tools. Use the Page Up ( ) & Page Down
( ) keys to display the different pages of this statistics module.
Statistics Legend:
• SML : The number of exposure done on the small focal spot.
• LRG : The number of exposure done on the large focal spot.
• AG : Number of Anode-to-Ground arcs.
• AK : Number of Anode-to-Cathode arcs.
• KG : Number of Cathode-to-Ground arcs.
• > 100% : Number of occurrences of Tube Heat Units greater than a hundred percent.
• MAXHU : Number of times an exposure was not allowed because it would otherwise have
exceeded the Maximum Tube Heat Units.
• HSING : Number of occurrences of Tube Housing Inhibit conditions.
• OV KV : Number of occurrences of Over KV fault.
• 1: 24 : Maximum Temperature on 3-Phase Inverter in slot position 1 was 24°C. Idem
for positions 2 thru 6.
Statistics Reset: Statistics can only be reset by using Epsilon PC Tools. See Section 7.5.3.2.
1. Connect the PC to the generator and launch the Epsilon PC Tools program.
2. Start the communication between the console and the generator using
Transfer and Connect.
3. Use Tools and Service menu to enable the Diagnostic module.
4. Choose the Monitor tab (Figure 7.12). The following information is displayed as a function of
the generator configuration.
Data Monitored
o Devices zone: The cursor indicates the activity of the following µprocessors
or software module:
CIO = Client µprocessor, on the Controller board
X-RAY = X-ray µprocessor, on the Controller board
CON = Console µprocessor, on the Console board
CIO-P2 and XRAY-P2 are used for biplane system only
CON2 is not used
o State zone: The selected radio button indicates in which state the generator is in now:
Init = Initialization of the generator;
Idle = The generator is in wait mode;
Fault = The generator is in fault mode;
Prep = The prep switch is active;
Ready = The prep and RAD switch are active and the generator is in ready state,
waiting for a trigger to make an exposure.
Run = An exposure is in progress (this could be Fluoro if Fluoro checkbox is set).
o In the Filament current zone, the main and standby filament currents are given in
Amperes. The currently selected main filament is indicated on the operator console (focal
spot lamp).
o In the optional Cable Discharge Temp zone, the temperature at the cathode & anode
cable discharge
o Inverters zone: Gives the status of the 3-Phase Inverter Boards. See also Section 7.5.2.6.
1. First column (Board)
3-Phase Inverter board number, from position 1 thru position 6, as installed on the
Motherboard.
2. Second column (Status)
Indicates the status of the Inverter board installed at the given position (Table 83).
3. Third column (Sel.)
Indicates which Inverter boards are presently selected for the given generator's state
shown in the State zone. Not all boards are used at all times.
4. Fourth column (Temp.)
Indicates the estimated temperature at the surface of every Inverter boards (34°C in Figure
7.12). The ambient temperature is also given (22°C at the sensor location).
Fix button: Use this button to reset the Inverter Status Register, for instance when an
Inverter board is replaced. The generator does a re-initialization process to validate the
reset.
o I/O zone: Gives the status of the different programmable inputs and outputs for the given
generator's state shown in the State zone.
History data must be fetched from the generator using the Query button when the PC is connected
to the generator. The log can be deleted (Delete) or it can be exported (Export) as a text file for
later reference. When the log is deleted, all events are deleted from the log and the following statistic
figures (displayed in Service mode, see Section 7.5.2.7) are zeroed:
1. Maximum Tube H.U.;
2. Tube H.U. greater than 100%;
3. Tube Housing;
4. Over KV;
5. Maximum Inverter Temperature.
Other information or actions can be obtained from the History Log screen:
Set Date and Time: Use this button to synchronize the generator clock with the PC.
Exposure Counters: Number of exposures done on both filaments.
Firmware: Names, part numbers and revision numbers of installed firmware.
Arc Counters: Number of tube arcs (for AG, AK and KG figures).
AG : Anode-to-Ground arcs
AK : Anode-to-Cathode arcs
KG : Cathode-to-Ground arcs.
Inv. Max. Temp.: The maximum temperature every 3-Phase Inverter board has reached
during the history log period.
HU Max.: Number of times an exposure was not allowed because it would otherwise
have exceeded the Maximum Tube Heat Units.
HU 100%: Number of occurrences of Tube Heat Units greater than a hundred
percent.
Housing: Number of occurrences of Tube Housing Inhibit conditions.
Over kV: Number of occurrences of Over KV fault.
Power Quality The Power Quality figure has an effect over the maximum kW admitted
for an exposure. The power quality percent is measured and refreshed
during idle mode. It varies from about 80% to about 125% and is based on
the following instantaneous four (4) measurements or values:
1. Number of good Inverters.
2. Maximum kW Figure set in the generator configuration thru
the Options setup in Epsilon PC Tools.
3. Instantaneous Line Voltage.
4. Measured Line Impedance.
• When the power quality figure reaches or exceeds 100%, the generator will be capable of
doing the most demanding exposure based on the maximum power output of the
generator configuration and inverter board status.
• When the power quality figure falls below 100%, the generator might not be capable of
doing certain demanding exposures. This could be the result of a bad Inverter, a bad line
condition, too much impedance on line or other. When the generator inhibits a kW-
demanding exposure, it displays the message “Generator kW Limit” at the operator
console.
The curves can be obtained from the generator using the Get from generator button or loaded
from a computer using the Load from file button. Once the tube calibration curves are obtained
from the generator, they can be stored to a connected PC using the Save to file button.
Figure 7.14 – Tube Calibration Viewer of the Epsilon PC Tools' Service Module.
It is also possible to transfer the tube calibration from a connected computer to the generator, using
the Send to generator button. This last action should be used very wisely, as it is not intended to replace
a tube calibration process. Doing a tube calibration is normal and unavoidable from time to time as the
characteristics of the tube vary with aging. Before sending a tube calibration file to the generator, make sure
that the tube chart of the sent tube matches the tube physically installed on the generator and configured in Tube Setup
of Epsilon PC Tools.
To Tube Calibration Viewer is best used by using the right mouse button to zoom in or out, display
the coordinates of a specific curve's point at the location of the cursor (Show Cursor), or print the
picture. The curve coordinates (x and y) are displayed using the scientific notation, for example:
x= 5.385E+00
y = 6.389E+02
Calibration data consists of different information such as the Fluoro Maximum dose allowed, certain
console parameters (brightness, volume& contrast), the software A/D device gain factors for RAD
and Fluoro ABS stabilization, and others. When you open this screen for the first time, make sure
to get data from the generator first before doing any change or saving to disk. Sending wrong
data to the generator can compromise the Fluoro entrance dose, the image stabilization or kV
and mA accuracies.
BAUD Rate
Leave this parameter at 57600 unless the communication fails. If necessary, change the baud rate.
Console Parameters
Use the Console parameters section to adjust any of the following setups.
a. The Operator Console Language : Defaulted to English as main language and French as
second language. Note that French second language can be changed to most any language by the
client using the Roman, Greek or Cyrillic alphabet. Refer to Section 4.8.2.
b. Contrast, Volume and Brightness parameters: Adjustable for better operator console visibility
and listening. These parameters can also be set using the console service menu (see Section 4.8).
c. mAs Copy Factor: Used in console software version 9811.03 only. The "Copy Fluoro Factors"
function is available in manual Pulsed Fluoro. In that mode, depressing F1 function key sets up
the RAD technique in two-point (mAs) mode with the kV equals to the actual Fluoro kV. The
RAD mAs parameter is a multiple of the actual Fluoro mAs; that factor can be modified by
depressing F1 key for a moment and by using the mAs Factor Setup menu or by using the “mAs
Copy factor” slider. The factor varies from 1.6 up to 10.0.
d. “Info Display” selection: Used to default one of the six possible data to be displayed to the
operator in the Information Display area (Figure 7.16). This information can be changed at any
moment by the operator, using the left-hand third selection key (I.D.A. key) on the console
Figure 7.16 – Graphic screen on the Epsilon Operator Console. Different information can be
displayed in the Information display area using the IDA key.
Test Points:
Table 84- Test Points on Controller Board
Test Point Description Range / Scaling
TP1(*) Voltage Controlled Oscillator 0–5V
TP2(**) kV Total 1 V = + 40 kV
TP3(**) kV Anode 1 V = + 20 kV
TP4(**) kV Cathode 1 V = - 20 kV
TP5 kV Select 1 V = + 40 kV
TP9 Tube current 1 V = 250 mA
TP10(**) Tube current ITUBE < 50 mA => 1 V = 15.6 mA
ITUBE > 50 mA => 1 V = 250 mA
TP6(*) Clock Inverter Phase Square pulses α Phase frequency ( Freq=> Power)
1 Hz @ TP6 = 2 Hz Ripple Frequency on Inverters
TP11(*) Permanent clock inverter Square pulses α Main frequency ( Freq=> Power)
TP12(*) Clock Inverter Main Square pulses α Main frequency ( Freq=> Power)
TP7 Main I FIL. feedback 1V=1A
TP13 Filament enabled 5 V = Filaments enabled (Normal state)
0 V = Filament Disabled (Filament Fault)
TP8 Standby IFIL. feedback 1V=1A
TP16 Main I FIL. select 1V=2A
TP14/15 mAs meter +/– Use an mAs meter to read the real tube mAs
TP17(**) Main gate 5 V = X-ray On ; 0 V = X-ray Off
TP18 Ground 0 VDC
TP19(*) I/F frequency ITUBE < 60 mA => 100 kHz = 1 A
ITUBE > 60 mA => 1.6 kHz = 1 mA
TP20(*) AEC EOX
= AEC End of exposure
TP21 Not used ---
TP22 Client I/O ground Client I/O ground
TP23(*) CPU EOX
= End of exposure (Fault)
TP24 Ready to expose (I/O’s) 5 V = OK to expose ; 0 V = Inhibit exposure
TP25 EIP Not used
TP26 Not used ---
TP27(**) AEC/ABS ramp: 0.1 to 4 VDC 1.0 V = 100% Brightness or normal density
When the film density is modified by a factor or when film/screen correction is applied, the
voltage read at TP27 at the end of exposure is modified by the same factor or correction.
(*) For emd technologies usage only. (**) See also typical waveforms in Figure 7.18.
Typical Waveforms:
Typical Exposure @ 100 kV, 500 mA Typical Exposure @ 100 kV (Anode + Cathode)
When LEDs on top of the 3-Phase Inverter boards are lit or blink, dangerous and lethal potentials are
present in offline circuits in several places of these boards! High Voltage, AC line-connected
potentials are present in these circuits. Do not manipulate these boards in this condition.
When the "AC Power" LED lights green on the High-Voltage Power Supply unit, on top of the Line
Filter board, dangerous and lethal potentials are present in offline circuits inside the generator! High
Voltage, AC line-connected potentials are present in these circuits. Make sure that the
generator enclosure is installed before operating the unit.
Figure 7.19 –
Warning LEDs on Inverter Boards P/N 9736.01 and
on X-Ray Control Panel.
In a fault condition, the red LED 2-bit code pattern is repeated continuously. The code can be
interpreted as follows:
1. X1 pulses of 250 ms on red LED where the number of pulses X1 indicates the decimal
number of the first digit.
2. 1/2 s delay between first and second digits.
3. X2 pulses of 250 ms on red LED where the number of pulses X2 indicates the decimal
number of the second digit.
4. 2 second-pause.
5. Steps 1 to 4 are repeated.
Before replacing a board inside the High-Voltage generator, make sure the generator power is turned
off at the mains entry panel for at least 5 minutes before removing the top panel of the generator
for board replacement.
Do not install or remove any component or board when the
power is on. Even when the generator is switched off at the
operator console, mains voltages are still present in High-
Voltage Power Supply unit. These voltages are extremely
dangerous, use utmost caution.-
• For the 45-80 kW configuration, the number of 3-Phase Inverter boards varies
from 3 to 6. In this configuration, no empty slots should be allowed in the 3-Phase
Inverter board locations. Any missing board should be replaced by a "Dummy
Inverter Jumper Board", P/N 9930.00.
• For the 30 kW configuration, only two 3-Phase Inverter boards are present on the
generator (no other boards in positions 3 thru 6).
1. Make sure the generator is turned off at the mains line disconnect box.
Warning ! When LEDs on top of the Inverter boards are lit or blink, dangerous and lethal
potentials are present in offline circuits in several places of these boards! High Voltage,
AC line-connected potentials are present in these circuits. Do not manipulate these
boards in this condition.
Figure 7.22 –
Access panels of Epsilon generator.
7. Reinstall the new board within the card slots and press it firmly downward to make sure the
ejectors clip in and the connection with the Motherboard is secured. Do not push 3-Phase Inverter
boards using their heat sinks.
Note: It is recommended to respect the board order identified on the Motherboard (Figure 7.24).
8. Reinstall the Inverter Board Retainer between the heat sinks as shown in Figure 7.23.
9. If 3-Phase Inverter boards were replaced, it is imperative to update the Inverter Board
Status register. Refer to Section 7.5.2.6 – Inverter Board Status Register Updating – to update
the register.
Technical Manual P/N 9894.00.G6 7-39
EPS 45-80 High-Voltage Generator
3
4
2
1
1. Card Ejector (x 12); 2. Inverter Boards Retainer (x 1); 3. Anode Starter board (x 1);
4. Three-Phase Inverter Board (3-6).
Note: Any missing Inverter board should be replaced by a Dummy Inverter Jumper board, (except
for EPS 30 kW generators).
10. Before replacing the covers, do a power on test to validate the installation.
11. Turn off the generator using the console power off button. Turn all powers off and wait five
minutes to completely discharge the generator. Replace the two panels on the top of the
generator.
Important
If the Controller board is in trouble and the firmware cannot be read, remember that a
backup of the installation should exist in virtue of the Installation procedure.
Refer to Section 8.2.3.5 or to Section 4.2.3.2.
1. Make sure the generator is turned off at the mains line disconnect box.
3. Take off any optional board from the Expansion Motherboard (Fluoro board, Relay board).
6. Disconnect all cables located inside the basket (Figure 7.29 and Figure 7.30).
1. Power supply 4. Filament (small and large)
2. Filament (small and large) 5. Anode connector (P18) from J18
3. Cathode connector (P17) from J17 6. Cathode connector (P19) from J19
Cathode
Anode
J19
Power Supply
Controller
J17 J18
Large Filament
Push to
release
Small Filament
7. On the new basket, reconnect the cables that were undone in step 6.
a. Reconnect first the four (4) filament cables (small and large).
b. Place the cables so that they will run at the bottom of the chassis assembly after the basket
is placed to its final position (see 1, Figure 7.31).
8. Reconnect the power supply cable. Place the cable so that it runs at the bottom of the chassis
assembly when the basket will be placed to its final position (see 2, Figure 7.31).
3 3
1
2
10. Flip the basket assembly back to its position in the chassis assembly (see Figures 7.32 thru 7.34).
Be careful of not pinching the cables when reassembling the basket in position
11. Place the basket assembly toward the main chassis and move it upward to insert the top
connector (Figure 7.33).
12. On the generator, reinstall all screws taken out in step 4 (Figure 7.25).
13. Reconnect all cables on the Controller board.
14. If any, reinstall the board(s) on the Expansion Motherboard (Fluoro board, Relay board, ..).
2. Verify that the Power-ON neon lamp on the High-Voltage Power Supply unit is turned off.
4. Replace the old Fluoro PMT board with the new one.
6. Perform the following field calibration before resuming to normal operation of the generator.
2. Verify that the Power-ON neon lamp on the High-Voltage Power Supply unit is turned off.
4. Replace the old board with the new board and reinstall the connections.
The battery has a useful life time of about 10 years. When replacement is necessary, a message is
displayed on the operator console. To replace the battery:
2. Take off the Controller Access Panel, as seen in Figure 7.35 below (4 screws to undo).
Locating pin
Figure 7.35 –
Replacement of Battery on the Controller.
4. To replace the battery, use the following part number: M4T32-BR12SH Lithium Battery
(120mAh, ST Manufacturer). The battery can be inserted in its socket in only one way as it is pin
oriented by a locating pin (see above Figure, right).
2. For replacement of an Inverter board fuse, refer to Section 7.6.1 to take out the 3-Phase
Inverter board. Pay great attention to all precautions given in that section.
Inverter Board Fuses: If one fuse is blown on the 3-Phase Inverter board, you should
consider replacing it. Replace a fuse on an Inverter board only when the Inverter Board Status
indicates “FUSE” (Section 7.5.2.6). Refer also to Section 7.5.2.6 to update the Inverter board
register in the Service Mode. Change the status from FUSE to OK and press “Save” key.
3. To replace a fuse, use the following part number and/or fuse rating:
Motherboard Fuses:
F1, F2, F3
MINIATURE FUSE 5AMPS 500VAC,
Dimensions: 1/4” diameter x 1 1/4” length (6.35 mm x 32 mm).
The four items form a group, and an element of one group or revision cannot be mixed with an
element of another group or revision. When updating a firmware, make sure that it is compatible
with the other two firmwares and with the current Epsilon PC Tools revision. If necessary, call emd
technologies for additional information, or verify the software compatibility on your customer web
page.
Epsilon Loader
No
tes
This module, when used inappropriately, can lead to serious or fatal impairment
of the generator. Do not use this module unless you are sure to have a backup or
an adequate copy of the generator firmware (Section 8.2.3.5).
Updating the CIO or Console firmware will result in the replacement of important
configuration data! Make sure to have a backup of the generator configuration data before
loading new software.
Firmware Devices Important Notes
9810.01.S19 CIO Will reset all the hardware setup, the tube calibration and the Fluoro dose
calibration.
9809.01.S19 X-RAY Configuration data not replaced.
9811.xx.S19 Console Will replace all APR data by the cold load parameters.
If you do not have the firmware and Epsilon PC Tools with you on a CD or on your laptop, you
will have to download them from emd technologies web site using the following procedure:
Epsilon Loader
1. Make sure the X-ray tube anode is not rotating. If necessary, brake it.
3. Open the Epsilon PC Tools program and start the Epsilon Loader
module by using Tools and Update Software (Figure 7.36).
Figure 7.36 –
6. Click on the appropriate Update device firmware button (CIO, XRAY or CONSOLE) to
launch the data transfer from the PC to the generator.
Note: CIO2, XRAY2 devices are used only on a biplane generator configuration.
CONSOLE2 device is not used on standard generator configurations.
7. Select the desired firmware version to be put into the µprocessor and click OK. Repeat, if
necessary, for the other firmware devices.
1. Make sure the generator is turned off at the mains line disconnect box.
3. Refer to Section 7.6 to remove the Inverter boards (or Dummy board, if any)
from positions # 2, 4 and 6.
Epsilon Options
Chip
pin1
7. Use Epsilon PC Tools program to change the software options of the generator. The generator
needs to know that new software features are now installed. Refer to Chapter 4.
7.7.1.1 kV Output
7.7.1.3 mA Control
Selection of other exposure parameters: The normal mAs calculator is used to determine mA and
time for parameters given mANOM & tBU
1. If tBU < tI, then mA = mANOM => The falling load function is off;
2. If tBU < τFIL then mA = mANOM => The falling load function is off;
Hence, when tBU larger than tI and τFIL => The falling load function is on.
In falling load mode, the normal mA/t calculator is used to determine mA max and mA min for
exposure:
Maximum mA for exposure of 0.7 * τFIL = mA(HIGH);
Minimum mA for exposure => mA(LOW) = mANOM2/mA(HIGH);
The falling load function is on.
Notes:
i. When falling load is active, if real mA <
mA(LOW), the filament loop will start
regulating to mA(LOW).;
ii. τFIL = 66% filament inertia time (τ = Tau)
Each time the generator receives an emd file from PC Tools, it verifies the manufacturer and part
number of the X-ray tube. If it is different from that saved during the last X-ray tube calibration, the
tube calibration table for both filaments are erased and the maximum Fluoro dose data is replaced
with the default values.
The emd file data stored in the console is limited to the APR tree and associated exam data. All of
the exam data can be modified by the user and saved in the console. Each exam sent from the
console to the generator is completely checked for suitability by the generator.
The emd file data in the Client I/O firmware is stored with an overall check sum fixed at the time
of data transfer from PC Tools. At power on and periodically during operation, the check sum is
verified and if an error is detected a message displays on the console. This error is extremely rare and
is similar to finding a check sum error in the code section of the flash memory. In the case of a code
check sum error, the processor enters the boot loader mode and the generator is unusable. With a
configuration check sum error, the generator is also unusable and Epsilon PC Tools must be used to
re-configure it. The following message displays:
PC
(Epsilon PC Tools)
A.
J16 Controller
PC Tools Communication
Between Laptop & Generator XRAY I/O µC
µC
RS-232
J4
CAN
USB
J7
J6
Console (CAN)
B. PC
J16 Controller (Epsilon PC Tools)
XRAY I/O µC
µC
C.
J16 Controller
J4
XRAY I/O µC
µC
J7 J4
PC
J6
J7
J6
Console
(RS232)
NOTE: USB connection is available on Epsilon Operator Consoles P/N 9777.21 and above
only. For an operator console without USB support and when standard RS-232 serial port is not
available on the computer, an USB to serial adapter can be used to connect the computer or
notebook to the generator. Recommended devices are:
• Edgeport/1 USB Converter, by Digi ConnectwareTM
• GUC232A USB to Serial Adapter, by IOGEAR
• RS-232 Optical Isolator, by B & B electronics
Before leaving a site after an intervention, the service representative can perform the following tests
to make sure the system works fine.
1. Perform the exposure tests of Section 5.3 (Output kV and mAs Verification) on the small and
large filaments.
2. Perform Fluoro in ABS mode and verify the image stabilizes using different phantom sizes (1
mm and 2 mm Cu are suggested).
4. Perform an A.E.C. exposure and verify the exposure does not terminate by BUT.
IN THIS CHAPTER :
A maintenance program is necessary to insure safe operation of the unit and to maintain the
Epsilon generator specifications. See table below to ensure proper maintenance of the generator at
regular time intervals.
Table 89-
Maintenance Schedule
Time Interval Description
Monthly 1. Cleaning the operator console (Section 8.2.1)
2. Verification of exposure Indicators (Section 8.2.2)
• Visual exposure indicator
• Audible exposure indicator
• Fluoro timer alarm
• Deadman exposure switch
Annual 1. Labels inspection (8.2.3.1)
2. Cable Tightness (8.2.3.2)
3. System Grounding Integrity (8.2.3.3)
4. Field Calibrations (8.2.3.4)
5. Generator Data Backup (Section 8.2.3.5)
6. Validation of AEC Mode (Section 8.2.3.6)
When necessary:
• System Backup 1. Perform a backup of the generator data (Section 8.2.3.5)
Warning
• Always switch off the mains power to the Epsilon generator and wait 5 minutes before
cleaning.
• Do not allow water or other liquids to enter the console or High-Voltage Power Supply unit
since this may cause short circuits or corrosion.
Epsilon Console: Spray cleaners or disinfectants are not recommended. Do not use abrasives,
solvents or corrosive cleaning or disinfecting agents. Clean painted parts with a clean cloth
dampened with a mild detergent solution. Clean the plastic faceplate with a clean cloth dampened
with a glass or plastic cleaner. Do not allow any liquid into switch recesses. Wipe the back panel and
connecting cable(s) with a moistened cloth.
High-Voltage Power Supply: It is recommended to remove the dust on the exposed panels of the
unit once a year using a vacuum cleaner.
1. Press the PREP switch. Verify that the Prep lamp lights green to
indicate the preparation is under way for the current workstation.
When Prep is ready, the lamp starts blinking.
2. Press the RAD switch. Verify that the X-ray On lamp lights up yellow
during the emission of X-rays and turns off at the end.
3. Verify that the buzzer sounds (continuous alarm) during the exposure (if enabled) and turns
off at the end of the exposure.
2. The Fluoro Cumulated Timer functions. Make sure the timer stops when the Fluoro footswitch
is released. Verify the function of the [Timer Reset] key.
3. Verify that the alarm pulses (pulsed alarm) for the first 5 minutes and then changes for
continuous tone after 5 minutes of Fluoroscopy. Verify the function of the [Alarm Reset] key.
2. Perform the Fluoroscopic Maximum Entrance Exposure Rate Calibration (Section 5.4).
4. The DAP Interface option, if present, shall be checked periodically. Refer to the DAP chamber
manufacturer & to the electronic document P/N 10050.014 to get the maintenance schedule &
program.
The complete back up of the generator set-up data is held in 4 distinct files. Use the following
procedure to obtain the complete backup of the generator.
1. Start the generator in Service mode by holding down WS1 key while switching on the power on
the console. After the power cycle self test, the generator will enter into the service mode.
3. Start PC Tools program and choose “create a new configuration file” and click the OK
button. This will load a default file (untitled.emd) and load the “Options screen”.
4. Connect to the console with the computer by choosing “Transfer” and “Connect”. When
connected, the PC Tools status at the right side of the screen shows “Monitor/Transfer”. Click
again on “Transfer” and select “Get from generator”. The generator configuration set up shall
be transferred to your computer. When finished, the console shall return to the normal operator
mode, usually displaying the “WS1” menu.
5. emd File (*.emd): Save the generator data of this emd file by doing “File/Save as ….”.
Use a filename recognizable in the future.
6. LOG File (log.txt): Get now the History Log of the generator by clicking in
the menu bar “Tools” / “Service”/ “History Log” and then clicking the Query
button. Export this file (a text file) using the Export button; save it in a
recognizable location for future reference.
7. TC File (*.tc): Get now the Tube Calibration file from the generator by clicking in the menu
bar “Tools” / “Service”/ “Tube Calib Viewer” and clicking the Get from Generator button.
Save the file in a recognizable location on your computer for future reference
using the Save to file button.
8. GC File (*.gc): Get now the Generator Calibration file from the generator by
clicking in the menu bar “Tools” / “Service”/ “Generator Calibration” and
clicking the Get from Generator button. Save the file in a recognizable location on your
computer for future reference using the Save to file button.
l- Quick method
1. Set-up the following 3-point AEC exposure: near 100 kV, 500 mA and 1.0 BUT.
2. By reducing the collimator aperture or by introducing aluminum phantom into the X-ray beam,
put enough filtration so that the exposure last for about 0.5 s.
3. Reduce the kV to about 60 kV.
4. Make an exposure and verify that the exposure is stopped by backup mAs detection (600 mAs).
5. Fully open the collimator (or take off the aluminum phantom).
6. Repeat the same AEC exposure and verify the exposure is stopped by phototiming feedback
(irradiation mA x time product less than 600 mAs).
If a unit must be shipped, clean it and install the protective cover on the H.V. receptacle before
packaging the assembly.
Use a robust cardboard box for packaging. The original shipment box (25” 20” 20”) contains
space for two protective Styrofoam pieces.
X-ray Software
P/N 9809.01
Console Cable
Universal Chamber
P/N 10015.xxx
Interface P/N 11002.000
Optional parts
Standard parts
Installer name
Site address
AEC Sensor(s)
Photomultiplier Tube ( ) Photodiode ( ) Flat Panel Detector ( )
Manufacturer Model:
Multi meter
Manufacturer Model Serial N°:
Last Calibration Date / /
Oscilloscope
Manufacturer Model Serial N°:
Last Calibration Date / /
X-ray Dosimeter
Manufacturer Model Serial N°:
Last Calibration Date / /
mAs meter
Manufacturer Model Serial N°:
Last Calibration Date / /
Fluoro F. R. bit # 0
Fluoro F. R. bit # 1
Fluoro F. R. bit # 2
RAD F. R. bit # 0
RAD F. R. bit # 1
RAD F. R. bit # 2
Interlock Fluoro
Interlock Ready
Input Assignation
Interlock X-ray
F. Alarm Reset
Interlock Start
Interlock Idle
AEC Reset
AEC Lock
Workstation
Fluoro
1 2 3 4 5 6
J14- 2
J14- 3
J14- 4
J14- 5
J15A-2
J15A-3
J15A-4
J15A-5
J15B-2
J15B-3
+ J3A-2
+ J3A-3
+ J3A-4
+ J3A-5
& J3-1
& J3-2
+ : On Fluoro PMT Board P/N 9822.00
& : On Photodiode Ion Chamber Fluoro Board P/N 11030.000
Fluoro F. R. bit #1
Fluoro F. R. bit #2
RAD F. R. bit #1
RAD F. R. bit #2
X-ray-On Alarm
Ready State
Fault State
Prep State
Start State
Run State
X-ray On
Idle State
Workstation
1 2 3 4 5 6 Fluoro
J14- 7
J14- 8
J14- 9
J15A-7
J15A- 8
J15B-1
J15B-5
J15B-6
+ J3A- 7
+ J3A-8
+ J3B-1
+ J3B-5
& J3-4
& J3-5
(LFB) J7- 1 & 2
(K1-K8)
+ : On Fluoro PMT Board P/N 9822.00
& : On Photodiode Ion Chamber Fluoro Board P/N 11030.000
Refer to Section 6.