Professional Documents
Culture Documents
(1) Update the firmware with Update-Kit2: V1.00R12 referring in 11.5 Updating procedure
III. Procedure of REGIUS 190/170 of <CS-3 Installation/Repair manual 1st edition>.
1. Upgrading JM/console *Do NOT use Update-Kit2:V1.00R07, 08, 09, 10 and 11 for updating, but make sure to use
V1.00R12.
If the console of the machine is CS-1/CS-3 and the software version is earlier than V2.00, *Do not upgrade using Update-Kit until overwriting the CF card is completed for (B) case of
upgrade the JM/Console using [CS-1CS-3 program V2.00R03 upgrading procedure] prior to the I. Target object. (If failed to do so, replace the MCB board and try the process (2)
upgrading Regius. Overwriting the CF cards again.)
-If the console of the machine is CS-2, upgrade the JM/console with CS-2 program V1.00R01 or *After overwriting the CF card according to the “”2) Overwriting CF cards,” never restore
later prior to upgrading Regius. the backup data that was saved before overwriting the CF card.
If you restore the backup data as the above case, you have to replace the CF card.
2. Upgrading Regius170/190
[Procedure required for all targeted machine] [Procedure required for all targeted machines]
(1) Confirm that the reader activates normally when connecting with JM/Console and also (1)Confirm that reader work normally when connected to updated JN/Console and that
Standby/Wakeup operates in keeping with Start/Stop mode. Standby/Wakeup operates in correlation with Start/Stop mode.
(2) In case version of firmware is unknown, check the “cf_version,mcb_version column” in (2) Open “r_version” table of JM and confirm the settings as follows:
table of “r_version” of JM.(When for case (B) of I . Target object I. the corresponding “cf_version = V100R11UPK”, mcb_version = V100R26
column should be blank). *3)
*.If speaker sound is customized prior to updating, re-set the setting referring the [Reader
(3) Pressing the “Operation” button will release the reader.
Firmware Setting Procedure-Supplement] attached.
*3).Refer to <CS-3Installation/Repair the 1st edition> 14.33 [Adjustment]-Setting to
confirm. IV. Attachment
*If (1) does not operate normally, change the network setting of JM/console of each table, and
then check it again. <CF card support tool instructions>
<Reader Firmware Setting Procedures-Supplement>