Image Manager- Function Panel
Vessel Analysis
Index
1 JiveX Information............................................................................................................ 3
1.1 General Information ................................................................................................. 3
1.2 System requirements ................................................................................................ 3
1.3 Software version CSI................................................................................................. 4
1.4 Requirements on image data .................................................................................... 4
1.5 License .................................................................................................................... 5
1.6 Interaction between Vessel Analysis-Panel and JiveX-Panel ....................................... 5
1.7 Terms and Abbreviations .......................................................................................... 7
2 Practical Application ..................................................................................................... 8
2.1 Overview ................................................................................................................. 8
2.2 Identify a vessel.......................................................................................................12
2.3 Adjust Vessel Centerline ..........................................................................................17
2.4 Vessel Analysis........................................................................................................18
2.4.1 General Functions ............................................................................................19
2.4.2 Adjust Contour .................................................................................................22
2.4.3 Aneurysm measurement ...................................................................................23
2.4.4 Stenosis measurement ......................................................................................26
2.4.5 Report..............................................................................................................28
In diagnostic radiology, the vessel analysis is very important for diagnosis. By MR-Angiography
(MR-A) and CT-Angiography (CT-A) sectional images (2D) of blood vessels can be acquired in a
very short time from head to toe. The challenge is to examine these blood vessels on possible
pathologies as quickly, reliably and efficiently as possible. In case of pathologies (aneurysm,
stenosis) these have to be measured and the result has to be presented with all necessary
information to the referring physician.
JiveX Vessel Analysis is a module which is available in the Diagnostic Advanced Client and in
the Demonstration Client. JiveX Vessel Analysis is a fully integrated third party vendor module
for detailed analysis, reporting and report documentation of blood vessels outside the heart. For
this, the external product ResMD with the clinical application Vessel Analysis of the company
Calgary Scientific (Canada) is integrated in the GUI of the RV Client.
The goal should be that the handling of the third party vendor module does not differ from the
use of the other functions of the RV Client.
System requirements of the workstation are identical to the Diagnostic Advanced Client.
Server-sided the Vessel Analysis needs an additional vessel analysis server (see document
“Hardware specification Vessel Analysis Server”).
The released software of the OEM product supplied by Calgary Scientific carries the version
number 4.5.2 and the “build number” 38177.
The “about“-dialog contains additional product information concerning the Calgary Scientific
Server (OEM).
In principle there are the same requirements for the image data as for the 3D VRT, MPR and
image registration.
Unlike the processing of image data in our implementations, the implementation of Calgary
Scientific requires completely equidistant spaces, i.e. the layer spacing may not vary. Our
programmed property „com.visustt.jiveX.viewer3d.maxSliceDeviation“ has no impact on the
behavior of the Calgary Server.
The image data of the Vessel Analysis must always have been first received by JiveX Server via
C-store. The Vessel Analysis with image data loaded from the local hard disc (via the file-open
dialog) is not possible.
The Vessel Analysis-license of the CS fixes the number of the simultaneously available
(concurrent) Vessel Analysis-licenses. They will be automatically released after a defined period
of time without the JiveX Client must be stopped.
If no more free Vessel Analysis-licenses are available on the server, the notice “No free Vessel
Analysis-licenses available” is indicated and the initialization of the Vessel Analysis is stopped.
Terminating a Vessel Analysis Session in the Client, the Vessel Analysis License will be
released. A Session is considered to be finished, if a Viewer Bean “Reset” is performed. The
Demonstration Client behaves differently: The Viewer Bean “Reset” doesn’t release the license,
but “timeouts“ become effective.
If you don’t move the mouse nor use the keyboard in an active Vessel Analysis Session, the
“Idle-Timer“ starts, that monitors the period of inactivity. It is defined in the “globalValues.xml“
in seconds and runs 20 minutes (default value). A change to the function tab does not affect this
timer.
After these 20 minutes an additional 5 minutes timer notice appear, which counts down in
second steps. The note is that the session will be closed in five minutes because of inactivity.
This value is defined in the “globalValues.xml”.
Once the Vessel Analysis is active, the activation of a function on all the other tabs of the
function panel or of a button in the toolbar, deactivates the active Vessel Analysis functions.
A Vessel Analysis session does not affect the functionality of the Captures. It can be created
neither Volatile Captures nor permanent Captures on the viewports of a Vessel Analysis-session.
The function buttons of the GUI creating Captures are disabled, if in the viewer a view is
shown, which is rendered by the Vessel Analysis server.
During a Vessel Analysis session the functions “Print“ and “Export“ are disabled. Only with
ending the Vessel Analysis session these functions can be used again.
VISUS “Reset“
Button and direct entry to change the series- and image matrix
(cMPR)
TOF This sort of sequence costs a lot of time and is not suitable
for organs which change their position during breathing.
2.1 Overview
Please load a CT-study, which contains a series of images with a layer thickness less than 3mm.
Now you have to select the series that you want to use for Vessel Analysis. You can scroll to the
interesting part (pathology) of the dataset.
In the function panel select the tab “Vessel Analysis“ and start the function with the button
“Vessel Analysis”. Alternatively you can start the Vessel Analysis via the context menu (right
mouse click).
! Notice !
If all buttons remain inactive after loading the study, the chosen series can’t be reconstructed.
Please make sure that you have not selected a series with less than 10 images or a series with a
scout image. The layer spacing should also be equidistant!
Once the reconstructions have been calculated, the various functions in the Vessel Analysis bar
are selectable.
In the middle section, the following functions are represented via four tabs:
“vessel detection”, “correction of the center line”, “correction of the contour line
measurement reporting” and the “creation of the diagnostic finding”.
In the section “Display” various “Default Color Presets”, the display of the “vessel tree”, the
“angiography view” and the “reset of the rotation of MPRs” can be activated.
The [B]-key on the keyboard for toggeling though the presets is active.
! IMPORTANT!
The functions of the buttons in the upper row remain active until they are deactivated or
the [Esc]-key is used.
Shortcuts as the [D]-key for distance measurement are not active in the Vessel Analysis.
Annotations can be selected by clicking on the line or using the handles at the end of
the lines. Generally a multi-selection is not possible. By clicking on an annotation or on
a measurement it will automatically change to the appropriate mode.
All annotations show at the end of a line so-called “handles“ onto which the cursor
changes. By clicking and dragging, the length and angle of the annotation can be
changed. The label of the annotations can be similarly touched and moved. A
movement of the entire annotation/measurement at angle measurement and text
annotations is not possible.
To start a vessel detection, please select the button “Identify a vessel“ in the section “Finder”.
Advice:
The start- and endpoints don’t have to be set in the same orientation (coronal, sagittal,
transversal, VR), different orientations can be chosen.
Pressing the [Shift]-key while doing the second mouse click,a straight line is drawn
between both markers without vessel identification. In this way a vascular obliteration
can be skipped. This function can also be used in the vessel extension.
All segmented vessels get a default name. Please select the previously segmented vessel and
rename it. When double clicking the name, you can type in the desired vessel name. Typing in,
e.g. the letter “a“, a popup menu is opened, where you can select the desired vessel name.
By using the cross behind the vessel name, you can activate or deactivate the display of the
vessels in the dataset. The asterisk behind the vessel name indicates that this vessel already has
been segmented.
Please now extend the segmented part of the vessel. For this, select the button “Extend a vessel“
and set an extension point into the vessel in a view of your choice. The centerline is updated in
all views.
To delete vessel segmentations completely, select the button “Remove the selected vessel from
the anatomical region“. The segmentation of the vessel and the entrance in the list are deleted.
The buttons “3D scalpel“ and “3D Clipping plane“ in the section “Tools” have the same
functionality as in the VRT in the Advanced Client. The shortcuts [S] and [C] on the keyboard
are active.
The grabbing dots of the clipping box are only displayed when the mouse cursor
touches the lines of the box.
To show only the vessel tree in the VRT orientation, select the button “Show the vessel tree“.
With the scrollbar you can fade in the surrounding tissue in the density you want.
In the second step you can correct the calculated center line of the selected vessel. For that the
blue bullets respectively the pink cross can be removed to the right position so that the system
recalculates the vessel. In the curved view (bottom right) additional blue bullets can be set for a
more distinct calculation. To replace the blue bullets or the crosses you only have to click with
the left mouse button on the desired place of the vessel. Only in the curved MPR you can
shorten the calculated vessel. For this purpose, one of the two outermost bullets is selected and
then positioned over the next bullet in the direction of the vessel center. The next stop needs to
be set beforehand; it defines up to which position the line should be shortened.
In the first viewport a stretched MPR (sMPR) and in the second viewport a curved MPR (cMPR)
of the vessel is displayed. With the [cursor up]-key and [cursor down]-key of the keyboard the
views can be flipped in 90° steps. For analyzing the vessel, different tools are available.
In the left viewport the diameter of the selected vessel is visualized. For that on the left and
right sides of the vessel a curve is shown.
In the left curve the minimum diameter of the vessel course (calculated in the cross
sections of the two dimensional measurement) is displayed.
The right curve displays the vessel diameter in mm² in the course of the vessel axis.
After activating the function, select the desired part of the vessel. On the right side a new cross
section of the vessel is displayed, which shows the sectional plane orthogonal to the vessel
center line. The area is calculated from the coverage of the vessel contour.
The length measurement along the center line of the vessel is drawn by clicking and dragging
with the left mouse button. The amount is displayed in a text label in [mm].
The measurement mode is deactivated after each measurement automatically. The button keeps
active. But several measurements can be done successively. Therefore activate and deactivate
the button.
The angle measurement along the centerline of the vessel is done by clicking and dragging with
the left mouse button. The two legs of the angle and the value in [deg] of the measurement are
displayed.
Along the center line of the vessel three markers are displayed. The length of the legs (outer
marker) and the angular point (inner marker) can be moved with the left mouse button by
grabbing these markers. The measurement mode is deactivated after each measurement. The
button keeps active. However, several measurements can be generated sequentially. Therefore
activate and deactivate the button.
In the display of the cross section the vessel contour can be circumscribed with a pressed left
mouse button to correct the automatically detected contour. The start and endpoint of the
freehand line have to cross each other. Then the manually correction will be adopted.
The puck minimizes or enlarges existing vessel contours. With a pressed left mouse button you
get a dot (puck). The size of this dot depends on the position of the first click. The more the first
click is set to the contour line, the smaller the dot will be. The more the first click is set into the
center of the detected vessel contour the bigger the dot will be. With a pressed left mouse
button you can move the dot (puck) and change the contour line.
After clicking on the button the centerline of the vessel automatically is displayed.
The design and the display of the aneurysm measurement is divided into three parts:
The section of the aneurysm, shown as a lilac line along the centerline of the vessel.
Proximal this section is defined by the khaki line and distal by the purple line,
which marks the passage to the normal diameter of the vessel in the distal part.
The distal (lower) aneurysmal neck. This part is marked in blue and is proximal
defined by the purple line and distal by the blue line, which marks the first relevant
distal vascular bifurcation.
The section of the aneurysm has three additional horizontal markers (green), which display the
maximum diameter of the contrast enhanced lumen, the minimum diameter of the contrast
enhanced lumen and the entire diameter of the vessel. The maximum entire diameter has to be
marked in the middle cross-section as a 2D measurement. Then the aneurysm measurement
has finished and can be saved in a report. Per report only one aneurysm measurement is
possible.
The transversal views of the horizontal lines are displayed in cross-sections of the
corresponding color at the left side of the viewport. The middle cross-section does not
Please select the button “Identification of the aneurysm diameter”. You can measure the
maximum diameter in the middle cross section in two dimensions. After the first measurement
you can automatically start with the second measurement. Changes are not possible. To make
changes, the whole measurement has to be repeated.
Once you have identified the maximum aneurysm diameter, the entire aneurysm measurement
is finished and can be stored in the report. Button “Save maximum aneurysm diameter to
report“
If an aneurysm measurement has to be removed, you draw the mouse over the middle part of
the horizontal marker, so that the line is highlighted. With the [Del]-key on the keyboard you
can now remove the measurement.
Please activate the button “Add vessel stenosis measurement“. To start this tool you have to
make a left mouse click on a MPR view. The start point has to be set above a stenosis. Two
blue lines appear. The second line has to be placed underneath the stenosis. Automatically the
stenosis will be calculated.
Additionally two cross-sections are displayed at the bottom left side of the viewport, which
display the normal and the constricted lumen. Above the upper cross-section the degree of the
stenosis is shown.
The measurement results are also displayed in the function panel. The measurement can be
changed at any time dragging the marker with the left mouse button to another position.
When saving the measurement the result is transferred to the report. Button “Save measured
stenosis to the result list“
2.4.5 Report
In the fourth part the report can be viewed, edited and saved.
To add a finding, select the button “Add diagnostic findings and assessments“. It opens a
dialog, where you can type in your finding. This finding is saved in the report.
Most of the functions can be activated also via the context menu.
JiveX Server and Viewer Version obligatory 4.5.2 (Released approximately end of
september 2012)
o Vessel Analysis
o Report
o Summary
o "...\Programme\csi-server\conf\clinicalapps\vesselanalysis.xml"
<?xml version="1.0"?>
<Configuration
xmlns:typeless="http://calgaryscientific.com/typeless/200
8">
<!-- LocalDicomCache>full_path_to_cache</LocalDicomCache
-->
<CrossSectionSize>70</CrossSectionSize>
<Report>
<Title>Gefäßanalyse</Title>
<DescriptionTitle>Befund</DescriptionTitle>
<NotesTitle>Zusammenfassende Beurteilung</NotesTitle>
</Report>
<Vessels>
<VesselList>
</VesselList>
</Vessels>
</Configuration>
Curved MPR in the second step of the vessel analysis can not be saved in the report
o %jcs_home%\resource_provider\vesselanalysis