*These authors contribute equally to this work and act as co-first authors.
1
Directorial Board, Pasteur Institute in Ho Chi Minh City, Vietnam
2
Microbiology and Immunology Department, Pasteur Institute in Ho Chi Minh City, Vietnam
3
Medical Analysis Department, Pasteur Institute in Ho Chi Minh City, Vietnam
4
Medical Testing and Calibration Centers, Ho Chi Minh City Pasteur Institute, Ho Chi Minh,
Vietnam
5
Planning Division, Pasteur Institute in Ho Chi Minh City, Vietnam
6
Faculty of Pharmacy, University of Medicine and Pharmacy at Ho Chi Minh City, Vietnam
DISCLAIMER
This paper was submitted to the Bulletin of the World Health Organization and was posted to the
COVID-19 open site, according to the protocol for public health emergencies for international
concern as described in Vasee Moorthy et al. (http://dx.doi.org/10.2471/BLT.20.251561).
The information herein is available for unrestricted use, distribution and reproduction in any
medium, provided that the original work is properly cited as indicated by the Creative Commons
Attribution 3.0 Intergovernmental Organizations licence (CC BY IGO 3.0).
RECOMMENDED CITATION
Lan PT, Cuong HQ, Linh HT, Hieu NT, Anh NH, Ton T, et al. Development of standardized
specimens with known concentrations for severe acute respiratory syndrome coronavirus 2
Realtime-RT-PCR testing validation. [Preprint]. Bull World Health Organ. E-pub: 20 April
2020. doi: http://dx.doi.org/10.2471/BLT.20.259630
Author for Correspondence:
167 Pasteur Street, District 3, Ho Chi Minh City, Vietnam (email: cuonghqpasteur@gmail.com)
Word count:
Objective: To develop a procedure for creating the standard specimens of known concentrations
for validation of Realtime RT-PCR testing.
Methods: We synthesized transcribed RNA based on the GenBank in order to create the
standard curve for estimating concentration among RNA extracted from an inactivated virus in
biosafety level 3.
Findings: We found that intra-assay precision, accuracy, and linearity met the accepted criterion
with RSD less than 15%, Besides, liner regression meets the accepted R2 of 0.98. The
concentrations of transcribed RNA and RNA extracted from the inactivated virus were quite
equivalent.
Conclusion: Standardized specimens of known concentration will improve the testing capacity
and support early diagnosis of infection as well as limiting the spread of the disease. In the long-
term, the standardized specimens of known concentration are applicable for evaluation,
verification, validation, and external quality assessment.
Methods
In this study, the sequence of SARS-CoV-2 virus was used based on the database of the NCBI
gene bank (GeneBank NC_045512.2). These sequences must ensure the coverage of the primer
and probe sequences was first announced by Corman, et.al [9].
The synthesized, length of the gene fragments were connected to the pJET 1.2 plasmid through
gene and vector homologous sequences using the cloning eClone system (Phu Sa Corp).
Eventually, the recombinant plasmid was transformed into Escherichia coli.
E. coli bacteria transformed was spread on LB agar plates containing 100 µg/ml ampicillin.
Gene-carried Bacteria were also selected through the characteristics of the gene carrying
plasmids. Plasmid pJET1.2 contains an enzyme gene that kills host cells, which enhances the
ability to select plasmid carrying genes. When a gene segment is inserted into the plasmid, the
lethal enzyme generation system is broken down. As a result, only bacterial cells with plasmids
carrying the inserted gene segment can form colonies. Plasmids that do not carry the inserted
gene are expressed as restriction enzymes, killing host E. coli cells after transformation.
Gene colonies are usually opaque when cultured on LB agar supplemented with 100 ug/ml of
ampicillin antibiotic. Collect these 5 colonies for PCR test with specific primers and pJET1.2.Fw
and pJET1.2.Rv primers on a pJET 1.2 plasmid. The modified vector-carrying strains were
grown on Luria-Bertani broth overnight. And recover with FavorPrep Plasmid Extraction Mini
Kit (Favorgen) according to the manufacturer's recommended procedures.
The transformation results were confirmed by genetic sequencing with specific primers and
plasmid pJET1.2 primers for all products by the Sanger sequencing technique on ABI 3130
device. The sequence needs to be synthesized to select the correct sequences to use for RNA
synthesis.
Straight-line plasmid fragments after treatment with a restriction enzyme were transcribed into
RNA by the catalysis of the enzyme T7 RNA polymerase (Thermo Fisher). The procedure
complied with the manufacturer's instructions. RNA after transcription was purified and
preserved in diethyl-pyrocarbonate water. RNA was then tested on a 1.2% agarose gel
electrophoresis system using 0.5X TBE buffer.
Quantity of pure RNA product (ng/µl) was determined in triplicate on a Nanodrop (Denovix).
Based on the size, nature of the target RNA structure, the yield for each control was calculated
using the Avogadro conversion factor (6,022x1023): Number of copies = (weight (ng) x
6,022x1023)/ (length x 330 x 109).
Culture
SARS-CoV-2 virus was isolated from throat swab samples of patients with Covid-19 infection
[10]. In brief, these specimens were placed in viral transport medium and refrozen, then inoculated
into Vero E6 cell culture. Vero E6 cells were cultured in Dulbecco’s minimal essential medium
(DMEM) supplemented with 10% heat-inactivated fetal bovine serum (FBS) and anti/anti
antibiotic [11]. On the isolation day, 50 µl serum-free DMEM was added into columns 2-12 of a
96-well culture plate. One-hundred µl clinical specimens were pipetted into column 1, and then
serially diluted 2-fold across the plate. After trypsinization, Vero E6 cells was suspended in
DMEM supplemented with 10% FBS, 200 IU/ml penicillin-streptomycin, and 5 µg/ml
amphotericin B with 2.5 x 105 cells/ml. One hundred µl of cell suspension were directly added to
the clinical swab dilutions and gently mixed by pipetting. The inoculated cultures were grown in
a humidified atmosphere of 5% CO2 at 37 °C. The cytopathic effects (CPE) were daily observed
in cells after incubation for three days. Standardized plaque assays were applied for SARS-CoV-
2 according to SARS-CoV and MERS-CoV protocols [12, 13].
Real-Time Reverse Transcription Polymerase Chain Reaction Assay for SARS-CoV-2
RNA synthesis placed into a collection tube with 150 μL of virus preservation solution. In brief,
40 μL of cell lysates were transferred into a collection tube followed by vortex for 10 seconds.
After standing at room temperature for 10 minutes, the collection tube was centrifugated at 1000
rpm/min for 5 minutes. The suspension was used for real-time reverse transcription-polymerase
chain reaction (RT-PCR) assay of 2019-nCoV RNA. The E target gene was amplified and tested
during the real-time RT-PCR assay with forward primer 5′-ACAGGTACGTTAA
TAGTTAATA GCGT-3’; reverse primer 5′-ATATTGCAGCAGTACGCACACA-3’; and the
probe 5-′FAM ACACT AGCCATCCTTACTGCGCTTCGBBQ-3′. All oligonucleotides were
synthesized and provided by Tib-Molbiol (Berlin, Germany).
Real-time RT-PCR was performed in triplicate on LightCycler® 480 System using a 2019-nCoV
nucleic acid detection kit according to the manufacturer’s protocol (TIB-Molbiol reagents).
Reaction mixture contained 5 μL of RNA, 12.5 μL of 2 × reaction buffer provided with the
Superscript III one step RT-PCR system with Platinum Taq Polymerase (Invitrogen, Darmstadt,
Germany; 0.4 mM of each deoxyribont triphosphates (dNTP) and 3.2 mM magnesium sulfate),
1 μL of reverse transcriptase/Taq mixture from the kit, 0.4 μL of a 50 mM magnesium sulfate
solution (Invitrogen) [9]. RT-PCR assay was performed under the following conditions:
incubation at 55 °C for 03 minutes and 95 °C for 30 second, 45 cycles of denaturation at 95 °C
for 03 seconds, and extending and collecting fluorescence signal at 60 °C for 12 seconds. A cycle
threshold value (Ct-value) less than 36 was defined as a positive test result, and a Ct-value of 40
or more was defined as a negative test [9].
Create cDNA based on Synthesis artificial
the publicly available RNA Inactivated SARS‐
Culture SARS‐CoV‐2
SARS‐CoV‐2 (Optical density CoV‐2
sequences measurement)
Estimate copy/µL
Made up Liner based on liner
regression equation regression equation Produce standard
specimens
Analysis
Data were entered using Epi-Data version 3.1 (EpiData Association, Odense, Denmark, 2005),
and all statistical analyses were carried out using Stata version 13.0 (StataCorp, TX, 2013). The
obtained sequences were manually edited with Chromas Lite 2.1.1. (Technelysium Pty Ltd,
South Brisbane, AU).
The results were summarized using mean and standardized deviation (SD), and relative
standardized deviation (RSD) for continuous variables. Linear regression analysis was performed
to estimate the linear regression equation. In this study, the percentage of RSD for accuracy,
intermediate precision and repeatability was less than 15% [14, 15].
Results
DNA fragments are synthesized in vitro by assembling single stranded oligonucleotide fragments
of 60-70 nucleotides length by the catalysis of Phusion High-Fidelity DNA Polymerase (Thermo
Fisher) using PCA technique. After assembly, product inspection was 2% gel agarose
electrophoresis. After assembly, the size of the DNA fragments was determined by
electrophoresis on 2% agarose gel, then transformed into pJET1.2 plasmid and screened the
transformation. The successfully transformed plasmid was extracted and sequenced by E gene
synthesized in vitro. The sequencing results showed that the E gene was synthesized to have the
same sequence as the originally designed E gene (gene sequencing data was illustrated in
supplementary (S1)). The result of aligning the E gene sequence with the E gene sequence on
Gene Bank showed that the two sequences are similar to each other with Query cover of 100%
and E value of 0.0 successful in vitro synthesis (Figure 1). E gene sequencing data was illustrated
in supplementary (Figure S1).
Figure 1. Alignment results of synthetic E gene sequence compared with E gene sequence on
Gene Bank.
Estimated concentration of RNA
Based on A260/A280 ratio of 2.1 and the concentration was 21.83 ± 0.74 ng/ul. The number of
copies/µL was 7.94x 1011. We calculated the number of copy/µL of 157.48 x 109
Gen E/Sars CoV-2 398 bp 1.97 0.97 21.83 ± 0.74 157.48 x 109
In the present study, we tested a dilution series of nine replicates RNA synthesis per
concentration. The Ct-value ranged from 22.4 to 23.69, and the number of copies was 79358.30
copies/µL at the level of 103 dilutions. When dilution tends to small, the Ct-values were more
likely to increase.
There was quite consistent fluorescence at 22, 24, and 28 cycles per concentration (figure 1).
Repeatability
Intra-assay precision was determined from assay results in each concentration run. The SD and
RSD of each concentration were 0.49- 0.95 cycles and 2.12- 2.69%, which reach the accepted
criterion of RSD<15% (Table 2).
Intermediate precision
Similarly, intra-assay precision was performed from assay results in four different runs and
three different days. The SD and RSD of each concentration were 0.06- 0.72 cycles and 0.36-
2.11%, which meets the accepted criterion of RSD less than 15% (Table 3).
Mean SD RSD %
Numbers
No. Repeatability
(copy/µl) (Ct-value) (Ct-value) (CT)
Three different concentrations (80 µL, 100 µL, 120 µL) were obtained by adding different
amounts of copy/ µL into the RNA synthesized specimens. The Realtime-RT-PCR was
performed by 7 replicates on different days. The percentage of RSD was 0.39-0.92% with an
RSD of <15%. The results also exhibited that the synthesized RNA specimens had a good
extraction efficiency.
Table 4. Accuracy test per each concentration for synthesized RNA.
Ct-values
Dilution
Mean RSD
No. (copy/µ 1st SD
2nd 3rd 4th 5th 6th 7th Ct-value %
L)
time time time time time time
Time
1 80 29.44 29.17 29.34 28.94 28.97 29.70 29.37 29.28 0.27 0.92
2 100 29.00 29.42 29.23 29.16 29.36 29.71 29.79 29.38 0.29 0.98
3 120 29.08 28.84 28.98 29.07 28.79 28.89 28.89 28.93 0.11 0.39
Table 5. The values of Ct-value per each concentration for synthesized RNA.
In this study, the linearity was created based on the Ct-value versus log copy. The linear
regression was obtained from five independent assays performed on different days. We
estimated the Linear regression equation with Y =– 3.68xlog copy/µl + 40.81, the R of the
standard linear reached -0.9989 which meets the accepted criterion of R2>0.98.
40
35
Cycle threshold (CT)
25
20
1 2 3 4 5 6
Log_copy
Figure 2. The linear regression based on the Ct-value versus log Copy.
In this study, intra-assay precision, the accuracy, and linearity met the accepted criterion with
RSD less than 15% (Table.5).
Result Required
Accuracy 0.39-0.92%
In order to estimate the concentration of RNA extracted from SARS-COV-2 virus isolated from
Vero E6 cells, Realtime RT-PCR was used in a dilution series of nine replicates run and in three
different days to estimate the Ct-values based on standard curves above. The percentage of RSD
was 0.26-5.21% with an RSD of <15%.
Table 6. Concentration of RNA extracted from SARS-COV-2 virus.
Ct-values Mean SD
RSD % Log
No. (Ct- (Ct- copy/µL
Day 1 Day 2 Day 3 (Ct-values)
values) values)
Discussion
This is the first study to develop standardized specimens of known concentration for SARS-
COV-2 testing. We followed the guideline on bioanalytical method validation [14, 15, 17].
Although, the A260/280 ratio was less than 2.1, it may be contained residual extraction reagent
or presence of proteins, it partly estimates the number of copies per reaction based on the
standardized specimens, especially in Covid-19 has spread around the world.
Policy implications
While waiting for a vaccine is completed or medicine is proven to be approval, diagnostics are a
life-threatening weapon in the fight against this pandemic and, once ready to use, standardized
specimens of known concentration will improve the testing capacity and support screening,
earlier diagnosis of infection and isolation. In the long-term, the standardized specimens of
known concentration are capable for evaluation, verification, validation, and external quality
assessment.
Conclusions
In summary, synthesized RNA based on the genes published was used to build the standard
curve for estimating concentration among RNA extracted inactivated virus under guideline on
bioanalytical method validation. Together with our findings provided a potential opportunity for
reliable testing to diagnose Covid-19.
This is a simulation study, no patient data were included, only throat specimens for virus culture
published in the previous study [10]. The emergence of Covid-19 is a significant health threat
that requires a rapid public health response to protect the entire community from this severe and
halt virus spread. For this reason, our request of a waiver of ethical approval.
Competing interests
Acknowledgment
We would like to thank Dr Ho Huu Tho at Military Medical institution and colleagues at Pasteur
Institute Ho Chi Minh City for supporting the study.
Authors’ contribution
HQC, PTL, HTL, NDH, NTTT developed the idea of the producing standardized specimens of
known concentration. HQC, HTL, NDH, NTTT, NHA, NTH performed the analysis and VTT,
NTTT, DTHT, TCD, NDT check the results. HDN, HQC, HTL, NTTT, VTT, TCD drafted the
manuscript, revised the manuscript. All authors were responsible for the final content. Both
authors have read and approved the final manuscript.
Supplementary data
Table S1. The estimation of Copy/µl concentration by Ct-value based on the liner regress
equation.
1 20 5.655 451743
2 21 5.383 241631
3 22 5.111 129245
4 23 4.840 69131
5 24 4.568 36977
6 25 4.296 19779
7 26 4.024 10579
8 27 3.753 5659
9 28 3.481 3027
10 29 3.209 1619
11 30 2.938 866
12 31 2.666 463
13 32 2.394 248
14 33 2.122 133
15 34 1.851 71
16 35 1.579 38
17 36 1.307 20
No. CT Copy/µl Log copy/µl
18 37 1.035 11
19 38 0.764 6
20 39 0.492 3
21 40 0.220 2
Figure S1. E gene sequencing data