When analyzer gives a false alarm - Failed to leave initial position (LYSE syringe).
Mechanism: The motor initial position photocoupler always detects that the photocoupler is
blocked.
Declaration
Reference Standard
DH36X
2. Poor connection of lyse syringe B. Replace the motor wire if it’s open circuit
DH36X↑ DF50CRP↓
DF50CRP ↑ DH56,UN73/DH76,D5CRP/D7CRP↓
4. Lyse syringe moves with too large A. Apply fresh lubricating oil after cleaning
resistance. the sliding rail and screw rode.
If you have any problem with this file, please contact our engineer ASAP, thank you!