Dymind Global Customer Service Center QA-F00XX

0x230341C0-Failed to leave initial position (LYSE syringe)

Use Background

When analyzer gives a false alarm - Failed to leave initial position (LYSE syringe).

Mechanism: The motor initial position photocoupler always detects that the photocoupler is
blocked.

Declaration

14/4/2021, V2.0, For reference only.

Reference Standard

Model Possibility Solution

If we find the syringe block the photocoupler,


then we can check the status of it in
software, Status > Sensor, if it displays Not
Blocked.

A. Clean the photocoupler.

B. Replace the photocoupler.


1. Photocoupler behind the lyse
C. Reconnect or replace the photocoupler
syringe assembly has failure.
wire.

DH36X

A. Reconnect the motor wire.

2. Poor connection of lyse syringe B. Replace the motor wire if it’s open circuit

motor wire. tested by multi-meter.

Method: Put the range to Diode, then test


the two ends of the wire from the same

Dymind Global Customer Service Center Roy


Dymind Global Customer Service Center QA-F00XX

color, the wire is ok if there’s a beep from


multi-meter, or broken.

Do a self-test for the syringe, replace the


motor if it doesn’t move. Self-test > Syringe >
Lyse Syringe.
3. The motor controls lyse syringe has
failure.

A. Apply fresh lubricating oil after cleaning


the sliding rail and screw rode.

4. Lyse syringe moves with too large


resistance. B. Check the O-ring of syringe, replace it if
damaged.

C. Replace the syringe assembly without


motor.

DH36X↑ DF50CRP↓

Dymind Global Customer Service Center Roy


Dymind Global Customer Service Center QA-F00XX

If we find the syringe block the photocoupler,


then we can check the status of it in
software, Status > Sensor, if it displays Not
Blocked.

A. Clean the photocoupler.

B. Replace the photocoupler.


1. Photocoupler behind the lyse
syringe assembly has failure. C. Reconnect or replace the photocoupler
wire.

A. Reconnect the motor wire.

DF50CRP B. Replace the motor wire if it’s open circuit


tested by multi-meter.

Method: Put the range to Diode, then test


the two ends of the wire from the same
color, the wire is ok if there’s a beep from
2. Poor connection of lyse syringe multi-meter, or broken.
motor wire.

Do a self-test for the syringe, replace the


3. The motor controls lyse syringe has
motor if it doesn’t move. Self-test > Syringe >
failure.
Lyse Syringe.

Dymind Global Customer Service Center Roy


Dymind Global Customer Service Center QA-F00XX

A. Apply fresh lubricating oil after cleaning


the sliding rail and screw rode.

4. Lyse syringe moves with too large


resistance. B. Check the O-ring of syringe, replace it if
damaged.

C. Replace the syringe assembly without


motor.

DF50CRP ↑ DH56,UN73/DH76,D5CRP/D7CRP↓

If we find the syringe block the photocoupler,


then we can check the status of it in
software, Status > Sensor, if it displays Not
DH56,
Blocked.
1. Photocoupler behind the lyse
UN73/DH76,
syringe assembly has failure. A. Clean the photocoupler.
D5CRP/D7CRP
B. Replace the photocoupler.

C. Reconnect or replace the photocoupler


wire.

Dymind Global Customer Service Center Roy


Dymind Global Customer Service Center QA-F00XX

A. Reconnect the motor wire.

B. Replace the motor wire if it’s open circuit


tested by multi-meter.

Method: Put the range to Diode, then test


the two ends of the wire from the same
color, the wire is ok if there’s a beep from
2. Poor connection of lyse syringe multi-meter, or broken.
motor wire.

Do a self-test for the syringe, replace the


motor if it doesn’t move. Self-test > Syringe >
Lyse Syringe.
3. The motor controls lyse syringe has
failure.

4. Lyse syringe moves with too large A. Apply fresh lubricating oil after cleaning
resistance. the sliding rail and screw rode.

Dymind Global Customer Service Center Roy


Dymind Global Customer Service Center QA-F00XX

B. Check the O-ring of syringe, replace it if


damaged.

C. Replace the syringe assembly without


motor.

If you have any problem with this file, please contact our engineer ASAP, thank you!

Relative spare part code

Model Part Name P/N Picture

Sensor(Vertical) 23.99.0007C

All models above Syringe motor 24.02.0003A

O-ring for Lyse Syringe /