Maintenance Instructions Sirona Dental Systems GmbH

ATTENTION !

Proper shielding of room and operator position is essential.


Since these requirements vary from state to state it is
the assembler's/installer's responsibility that all local radiation safety
requirements are met.

59 01 603 D3285
2 D3285.103.04.01.02
Sirona Dental Systems GmbH Maintenance Instructions

General
To stay in compliance with the DHHS requirements the ORTHOPHOS®3 / 3 Ceph must be maintained an-
nually following date of installation.
It is the responsibility of the user to insure that the equipment is maintained with the manufacturer's
recommended Maintenance Instructions to insure compliance with the Federal Performance Standard.
The manufacturer and the assembler/installer are relieved from responsibility in those cases where
noncompliance with the standard results from the user's failure to have the manufacturer's recommended
maintenance performed.
The actual maintenance inspection and consequent service must be accomplished by a trained serviceman.
Neither the inspection nor service is part of the equipment warranty.

Technical instructions required


Operating Instructions
Service Manual

Instruments and adjustment tools required


1. Digital multimeter Philips PM 2816 rms, Fluke 8000 A, or equivalent.
Accuracy: DC voltage ± 0.1 % of reading plus 0.02% of range
DC current ± 0.4 % of reading plus 0.1% of range
2. Electromechanical pulse counter,
model KESSLER ELLIS KT 203 ±1 pulse, or equivalent.
3. Needle phantom, adjusting film for tele-exposure (delivered with the unit, customer's property)
Alignment tool for X-ray beam, order no 89 31 800 (not delivered with the unit).

CAUTION RADIATION
Observe radiation protection guidelines as outlined in the Operating Instructions!
X-rays are generated, when the exposure button at the Multitimer is depressed.

Caution serviceman!
All PC-boards are fitted with electronic components sensitive to electrostatic discharge (ESD).
Electrostatic charges are unavoidable due to friction of clothing, carpeting etc.
To prevent damage of electronic components do not touch
same without putting on the unit mounted special wristlet.
Always handle circuit boards by the edge of same.
ESD

List of Contents Page


General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Visual Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Light Indicators and Audible Sound . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 – 6
Power Supply Adequacy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
kV-Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 – 9
Tube Current Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Exposure Time Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 – 13
Checking and Adjusting the X-Ray Beam for Panorama Exposure. . . . . . . 14 – 16
Phantom Radiograph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Checking the Ear Olives . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Checking the X-Ray Beam for Tele-Exposure . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Checking the Laser Light Beam . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Yearly Maintenance Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . appendix
59 01 603 D3285
D3285.103.04.01.02 3
Maintenance Instructions Sirona Dental Systems GmbH

This product complies with DHHS


regulations 21 CFR Subchapter J
applicable at date of manufacture.
Date of manufacture:

Sirona Dental Systems GmbH


Fabrikstr. 31
D-64625 Bensheim Germany Model- 18 90 123 D 3285
No.
Serial-
No.
Model- 30 03 246 D 3285 MADE IN GERMANY
No.
Serial- Model-No. 18 89 356 D 3285
No.
Serial-No
lll 80kV 10mA
↓ 2.5 Al / 80 IEC 522 1976
By 560 / 93 Rö - Typ D3285

Röhre / Tube
Siemens SR 80 / 10
Model-No. 28 09 163 V 7012
Serial-No.
For identification ‰ 0.5 IEC 336 / 82
turn wheel MADE IN GERMANY

Model-D clockwise.
No. 3285 oder Röhre / Tube
Serial- or CEI OPX / 105 / S
No. ou
o Model-No. 51 69 722 D3328
208 / 230 V~
max. 9,7 A
50 / 60 Hz Serial-No.
Software - Version : ‰ 0.5 IEC 336 / 82
Hardware - Version : MADE IN GERMANY
Sirona Dental Systems GmbH This product complies with DHHS
Fabrikstr. 31 regulations 21 CFR Subchapter J
D - 64625 Bensheim applicable at date of manufacture.
GERMANY Date of manufacture:
0123
Sirona Dental Systems GmbH
Fabrikstr. 31
D-64625 Bensheim Germany
Model- 18 90 149 D 3285
No.
Serial- Model-No. 18 88 408 D3200
No. Diaphragm 1
MADE IN GERMANY Serial-No.

Model- 59 38 436 D 3200


Ceph
No. X-RAY Model-No. 18 88 424 D3200
Serial- Multitimer Diaphragm 3
Serial-No.
No. 10 x 8 in. A
MADE IN GERMANY X-ray control.
This product complies with DHHS May be installed as
regulations 21 CFR Subchapter J
applicable at date of manufacture. remote control.*
Date of manufacture: Model-No. 18 88 366 D3200
Diaphragm 4
Sirona Dental Systems GmbH Serial-No.
Fabrikstr. 31 8 x 10 in. S
D-64625 Bensheim Germany

Visual Check
• Look for mechanical damage possibly affecting radiation safety.
• Verify that all labels are affixed and legible.
Defaced labels must be replaced.
Order same from Sirona (address, see rear)
in writing stating: Customer Name
Customer Address
All Model Numbers with Serial Numbers
still legible on the unit for identification purposes.
For serial numbers see also Warranty
Unit Passport.
Passport.
* REMARK for the following checks
If the Multitimer is installed as remote control outside the X-ray room:
Attach the Multitimer temporarily to the unit.
For this purpose remove plastic front cover of the unit's carriage and
the lower shielding, covering the plug.

59 01 603 D3285
4 D3285.103.04.01.02
Sirona Dental Systems GmbH Maintenance Instructions

I ON
O

C Ceph-mode only

X-RAY
Unit ON LED

A P I CS
kV 68
B
mA I0

Ready LED

Multitimer

Light Indicators at the Multitimer, Audible Sound at the Unit.


• Unit ON LED:
Depress the main switch into the ”I” position to turn unit ON.
The unit adjusts itself automatically, wait about 1 minute.
The ”Unit ON LED” in the upper left corner of the Multitimer will then indicate that the unit is ON.
• Digital displays at the Multitimer:
The program and exposure parameters employed with the last patient appear.
A shows you sequentially the P1, P6 and P11 exposure programs (with the diaphragm 1 set) and
the respective maximum exposure time. Press – + program keys for correct display check.
With diaphragms 3 and 4 the exposure times for tele-exposure light up.
B gives you the kV/mA matched value pair. The LED over the respective patient symbol must light up.
LED over the return key R must blink. (When LED blinks, unit is not ready for the exposure).
C lights up, when Ceph–mode is selected (see ORTHOPHOS 3C Operating Instructions for
Tele-Exposures).
See also Operating Instructions under ”Preparing the Exposure” subchapter ”Switching ON the Unit”.

continues

59 01 603 D3285
D3285.103.04.01.02 5
Maintenance Instructions Sirona Dental Systems GmbH

Radiation CAUTION RADIATION !


indication Observe
X-ray Radiation Protection Guide Lines
see Operating Instructions

Exposure
key

Multitimer X-RAY

P I
kV 68
mA I0

Ready LED

Return key R

• Make a panorama exposure: (diaphragm 1 is set at the diaphragm wheel)


- X-ray head must be in the initial position (If not, press return key R).
- Insert a panorama film cassette into the carriage and swing in the cassette holder. The LED over R
key must go out. For more details and possible error messages see Operating Instructions.
- Set the P1 exposure program using the upper – + keys.
- Select 68kV/10mA using the lower – + keys.
- CAUTION RADIATION. Depress the exposure key and hold until the exposure terminates auto-
matically.
The exposure ends when the LED over R key blinks and rotation and radiation automatically switch off.
The radiation indication X-ray must light up during the exposure period.
Simultaneously an audible beep must sound at the unit.
• Interrupt an exposure – deadman feature:
- Observe a cool-off time of 5 mins. between exposures (automatic exposure blockage).
- Setting same as above. Remove and reinsert the film cassette. The LED over R key must go out.
- CAUTION RADIATION. Depress the exposure key until X-ray lights up and subsequently release
– the exposure must terminate immediately. The LED over R key blinks.
• Defective light indicators constitute a safety hazard to the patient as well as to the operator.
The user is not permitted to use the unit, until repairs are made!

59 01 603 D3285
6 D3285.103.04.01.02
Sirona Dental Systems GmbH Maintenance Instructions

2. metal cover
6 screws

I ON
5. O

K1
6.– 9.
N
N
L
3. For details see
L
300VAC
1. Installation Instructions,
same chapter or
Operating Instructions.

4. 5.

Power Supply Adequacy


• To determine power supply adequacy, the line voltage drop during exposure must be measured.
1. Be sure power is disconnected at the central distribution panel!
2. Remove front plastic cover (for details see Service Manual). Remove lateral metal cover (6 srews).
3. Select 300VAC line voltage range on multimeter.
4. Connect measuring leads to terminal K1, L and N.
5. Connect power and switch unit ON. Wait 1 min. for self-adjustment of the unit.
6. Press key R at the Multitimer to return X-ray tube head into the initial position.
7. Remove and reinsert the film cassette.The LED over R key at the Multitimer must now go out.
8. Select P1 program and 80kV/10mA at the Multitimer.
9. CAUTION RADIATION! - Depress the exposure key at the Multitimer until meter reading is obtained.
Line voltage, no load: 187 – 208V Max. permissible line voltage drop: 9V
208 – 230V 8V
230 – 240V 7.5V
240 – 264V 7V
• - - Record reading. Turn unit OFF. Remove meter leads and refit lateral metal cover.
• If the voltage drop is not within the specified range advise the customer, that an adequate power
supply must be installed. Refer to Pre-Installation Instructions.

59 01 603 D3285
D3285.103.04.01.02 7
Maintenance Instructions Sirona Dental Systems GmbH

1.

I
O 2.
3. 4.– 7. 20VDC
com.
For details
see Operating Instructions.

KV– KV+

KV+

KV –

kV – Verification, kV–ramp during panoramic exposure


• During exposure the kV is encreased in the central region depending on kV/mA selected up to 12%.
This increase can be measured in VDC.
1. Remove metal cover (2 screws).
2. Connect digital voltmeter to KV+ and KV– and select range 20 VDC.
3. Switch unit ON. Wait 1 min. for self-adjustment of the unit.
4. Press return R key to return the X-ray head into the initial position.
5. Remove and reinsert the film cassette.
6. Select P1 program and 72kV/10mA at the Multitimer. LED above R key on Multitimer must be off.
7. CAUTION RADIATION! - Depress the exposure button until the exposure terminates automatically.
The following values must be obtained – see also diagram on next page.
up to 4.1 seconds: 7.2 V ±0.5V,
from 4.9 to 6.5 s: 7.9 V ±0.5V,
after 7.3 seconds: 7.2 V ±0.5V.
• Turn unit OFF and remove meter leads.
• If specified values cannot be obtained, see Service Manual, ”Radiograph Density in Central Region
Incorrect”

59 01 603 D3285
8 D3285.103.04.01.02
Sirona Dental Systems GmbH Maintenance Instructions

VDC ramp
1.6s

8 nominal 7.9 V
nominal 7.2 V
7

5 4.1s±0.4s
4 7.3s±0.6s
3 5.65s±0.5s 11.3s
2 complete P1
exposure time
1
s
0 1 2 3 4 5 6 7 8 9 10 11 exposure time

kV – ramp diagram
with program P1 and 72kV/10mA set on the Multitimer.

ATTENTION
The kV - ramp can be changed at the customer’s request.

59 01 603 D3285
D3285.103.04.01.02 9
Maintenance Instructions Sirona Dental Systems GmbH

2. 20mADC
I
3. O 5. com.

7.
4./6.
MA – MA+ MA+

MA –

jumper

1.

Tube Current Verification


1. Remove jumper from MA+/MA – test points on PC-board DX1.
2. Connect digital ammeter to MA+ and MA– and select range 20 mADC.
3. Switch unit ON and wait 1 min. for self-adjustment of the unit.
4. Press return key R to return the X-ray head into the initial position.
5. Remove and reinsert the film cassette.
6. Select P1 program and 72kV/10mA at the Multitimer. The LED above R key on Multitimer must be off.
• Measurement:
7. CAUTION RADIATION! Depress the exposure key and hold depressed until meter reading is
obtained.
The ammeter shall indicate 10mA ±0.5mA. - - Record reading.
• If specified value is obtained switch unit OFF. - Remove meter leads and replace jumper!
• If specified values cannot be obtained, see Service Manual, chapter ”Tube Current Verification”.

59 01 603 D3285
10 D3285.103.04.01.02
Sirona Dental Systems GmbH Maintenance Instructions

8.

I
O 5. No 1

4. Pulse counter
6./ 7.
Multitimer

T1 T0
3.
com.

2.
1. To N
To L
of 110VAC
power receptacle

Exposure Time Verification for Panorama Exposure


• The diaphragm 1 for panorama exposure is still adjusted at the diaphragm wheel.
1. Connect test point T1 of terminal on PCB DX1 to L of 110V power receptacle.
2. Connect common lead of pulse counter to N of 110V power receptacle.
3. Connect second measuring lead of pulse counter to test point T0.
4. Switch unit ON. Wait 1 min. for self-adjustment of the unit.
The X-ray head must be in the initial position (if not press return key R on Multitimer).
5. Remove and reinsert the film cassette.The LED above the R key on the Mutitimer must go out.
6. At the Multitimer select 72kV/10mA with P1 program.
• Measurement:
7. CAUTION RADIATION ! - Depress the exposure key and hold depressed until the exposure terminates
automatically.
The pulse counter shall indicate at 60Hz: nominal 11.3 sec.= 678 pulses±36
at 50Hz: nominal 11.3 sec.= 565 pulses±30
- - Record average pulse count.
8. Without Tele-Exposure: If specified values are obtained, switch unit OFF.
Remove meter leads and refit small metal cover (2 screws).
• If specified value cannot be obtained, see Service Manual, chapter ”Exposure Time Verification”.

59 01 603 D3285
D3285.103.04.01.02 11
Maintenance Instructions Sirona Dental Systems GmbH

I
3. O

1.
2. Pulse counter

Multitimer

T1 T0
com.

To N

To L
of 110VAC
power receptacle

59 01 603 D3285
12 D3285.103.04.01.02
Sirona Dental Systems GmbH Maintenance Instructions

Exposure Time Verification for Tele – Exposure


• The pulse counter is still connected as described on previous page.
1. Prepare unit for tele-exposure:
Adjust diaphragm 3 for tele-exposure on the diaphragm wheel.
Swing panorama cassette holder out of x-rax path.
Press R key to move x-ray head into position for tele-exposure.
For details see Operating Instructions for Tele-Exposures, chapter ”Preparing the Tele-Exposure”.
2.1 At the Multitimer select 72kV/10mA and 1.0 second.
• Make four exposures CAUTION RADIATION !
Depress the exposure key and hold depressed until the exposure terminates automatically.
- Record pulse count and reset pulse counter after each exposure.
1. exposure ......................... pulses Allow a 5 mins. cool-off time between exposures!
2. exposure ......................... pulses (automatic exposure blocage)
3. exposure ......................... pulses
4. exposure ......................... pulses
total ................................. pulses, divided by 4= average pulse count.
- - Record average pulse count.
At 60Hz: nominal 1.0 sec. = 60 pulses±2 at 50Hz: nominal 1.0 sec. = 50 pulses±2

2.2 At the Multitimer select 72kV/10mA and 2.0 seconds.


• Make four more exposures CAUTION RADIATION !
Depress the exposure key and hold depressed until the exposure terminates automatically.
- Record pulse count and reset pulse counter after each exposure.
1. exposure .........................pulses Allow a 5 mins. cool-off time between exposures!
2. exposure .........................pulses (automatic exposure blocage)
3. exposure .........................pulses
4. exposure .........................pulses
total .................................pulses, divided by 4= average pulse count.
- - Record average pulse count.
At 60Hz: nominal 2.0 sec.=120 pulses±2 at 50Hz: nominal 2 sec.=100 pulses±2
3. If specified values are obtained, switch unit OFF.
Remove meter leads and refit small metal cover (2 screws).
• If specified values cannot be obtained, see Service Manual, chapter ”Exposure Time Verification”.

59 01 603 D3285
D3285.103.04.01.02 13
Maintenance Instructions Sirona Dental Systems GmbH

I
O

3. 1.
2.

X-RAY

P I
kV 80
mA I0
Beam
alignment
tool

Checking and Adjusting the X-Ray Beam


1. Place beam alignment tool vertically in the cassette carriage and move behind the slit cover plate.

2. Set the diaphragm 1 on the wheel.


To move the wheel, press button D. X-radiation will only be activated when button D is engaged.

3. Turn unit ON and wait 1 min. for self-adjustment of the unit.

• Thereafter select Service - Routine S.02 at the Multitimer. See next page.

59 01 603 D3285
14 D3285.103.04.01.02
Sirona Dental Systems GmbH Maintenance Instructions

Multitimer

X-RAY X-RAY X-RAY

007
kVE2
S. 0 I
kV
S. 0 2
kV00
8.
mA04 mA mA

1 2 4
6. 2– 4 –1

4. Memory

5. S23 7. S23 9. S23


X-RAY X-RAY

S.0 2
kV02
10. 3.2 0
kV 80
c
s 3.

1. 4.
mA mA I0

2. S23 twice

• Select service routine S.02 at the Multitimer (X-ray head and cassette carriage remain stationary):
4. Press Memory key.
5. Press service key S23 until (approx. 4s) the digital displays disappear.
6. Within 3s press patient symbol keys in the sequence 2 – 4 – 1.
- Service mode is now selected.
• If the sequence is not observed when selecting the service routine, if a wrong key is pressed or the
time is exceeded, the system switches automatically to normal mode.
7. Press service key S23.
- Service routine S.0 I appears on the digital display.
8. Press + key to select service routine S.02.
9. Press service key S23.
- On the kV display appears 00.
10. 1. Press + key to select service code 02.
2. Press service key S23 twice (kV/mA blinking)
3. Select exposure time 3.20s.
4. Select 80kV.

59 01 603 D3285
D3285.103.04.01.02 15
Maintenance Instructions Sirona Dental Systems GmbH

Wheel

ε
1 PAN
Correct
position X Y Z
of X-ray
beam 10.
X-RAY FilmPAN 1
15x30cm

c
6x12 in. C
3.2 0 s
E PAN E
kV80 SID=49,7

mAI0 B

• Darken the room.


10. Activate X-radiation to check correct position of X-ray beam. CAUTION RADIATION!
• If X-ray beam is not in correct position adjust the X-ray beam position to the corresponding mark 1
of the alignment tool.
To do this, remove wheel cover of X-ray head (see Service Manual).
Activate X-radiation. CAUTION RADIATION!
Activate X-radiation only for as long as you need to recognize the X-ray beam position.
X Beam correction 'HIGH – LOW': Loosen 2 screws E by 1 turn. Set beam via screw A.
Retighten screws E.
Y Beam correction 'VERTICAL': Loosen 2 screws E. Set beam via screws B and C.
Retighten screws E.
Z Beam correction 'RIGHT – LEFT': Loosen 2 screws E. Set beam via screws B and C.
Retighten screws E.
• Remove beam alignment tool and switch unit OFF.
NOTE: Final check of X-ray beam position under ”Phantom Radiograph”.

59 01 603 D3285
16 D3285.103.04.01.02
Sirona Dental Systems GmbH Maintenance Instructions

I
O 3.
2.
Cassette 1.

Paper
Film
Paper

5. RADIATION 6. Process film

X-RAY

P I
kV 60 4.1 Un-
exposed
m I0 margin

4.2 7. a1 a2 a1=a2 ±0.5mm


±1/50”
84 - 85.5mm
Line distances 3 5/16” - 3 6/16”

Phantom Radiograph
• Diaphragm 1 is still adjusted at the diaphragm wheel.
1. Fit the needle phantom up to stop.
2. Cut out two paper strips as large as the film size.
Place the film between the paper strips in the dark room – making sure film and paper strips are
level with the bottom of the cassette! (The paper strips are needed to neutralize the intensifying screens).
Attach film cassette to the unit's cassette holder (For details see Operating Instructions).
3. Switch unit ON and wait 1 min. for self-adjustment of the unit.
4. Select P1 and lowest kV. Press return key R , the X-ray head moves into start position!
5. CAUTION RADIATION! Initiate the exposure for a complete rotating cycle. 84 - 85.5mm
0 3 5/16” - 3 6/16”
6. Process the film.
7. Check the film: Measure the
l___ .___ . ___ .___ . ___ .___ . ___. ___. _ l _ .
line distances on the film as shown.
• If the distances exceed the above tolerances, actuator M2 must be adjusted –
see Service Manual.

59 01 603 D3285
D3285.103.04.01.02 17
Maintenance Instructions Sirona Dental Systems GmbH

1. B
2.
E 3.
Transparent cap
small ball
5.2
E X-RAY

Black cap
2.0 0 cs
k 62 5.1
4. m I0

Checking the Ear Olives


• Metal balls in the adjustment caps show up as dots on an exposed film. Both dots must coincide.
1. Attach an intraoral film with adhesive tape to the cassette holder.
2. Turn button B counter clockwise and position ear olives in X-ray path.
3. Pull out ear olive holders to maximum distance and place adjusting caps E on ear olives.
4. Prepare unit for tele-exposure (diaphragm 4). For details see ORTHOPHOS 3C Operating Instructions
manual, chapter ”Preparing the Tele-Exposure”, steps 1. – 4.
5. Select 62kV/2.00s on the Multitimer and release the exposure. CAUTION RADIATION!
• Develop film and evaluate the position of the dots.
• If the dots do not coincide see Mounting Instructions, chapter 'Checking and Adjusting the Ear Ol-
ives' for correction.

59 01 603 D3285
18 D3285.103.04.01.02
Sirona Dental Systems GmbH Maintenance Instructions

1. D
120 3

3.1
A
3.1 3.2 S 3.2 4

X-RAY RADIATION

4.0 0 cs
k 80 2.
m I0
Left and lower On all sides an even
margin unexposed unexposed margin.
1/8”+1/16”

Checking the X-Ray Beam Positon for Tele – Exposure


1. Move the ear olive holder out of the X-ray beam path.
The cassette holder for panorama exposure is still out of the X-ray beam path .
2. Select 80kV/4.00s at the Multitimer.
3.1 Check A = ASYMMETRICAL with diaphragm 3.
Set diaphragm 3 on wheel (press button D and turn wheel). Set soft-tissue filter to position 120.
Insert opened cassette vertically up to the lefthand stop.
Activate X-RADIATION — only for as long as you need to recognize the beam position.
• If the X-ray beam is not in the correct position,
see Mounting Instructions, chapter 'Adjusting the X-Ray Beam for Tele – Exposure' for correction.
3.2 Check S = SYMMETRICAL with diaphragm 4.
Set diaphragm 4 on wheel.
Insert opened cassette vertically up to the righthand stop.
Activate X-RADIATION to recognize the beam position.
• If the X-ray beam is not in the correct position,
see Mounting Instructions, chapter 'Adjusting the X-Ray Beam for Tele – Exposure' for correction.
• Finally reattach covers.

59 01 603 D3285
D3285.103.04.01.02 19
Maintenance Instructions Sirona Dental Systems GmbH

1. 2.

60°

32° 5° 60°

Horizontal beam Vertical beam

Checking the Laser Light Beam


• Pressing the height adjustment button ↑ always automatically switches on the laser light localizers for one minute.
• Radiation of class 1 appears during tests.
Keep always a minimum distance between eyes and laser of 100mm.
• No stray radiation shall be emitted by defective optics or covers.
The laser must generate a line.
• No controls are available to adjust the level of laser radiation.
1. Horizontal laser light beam FH.
The light beam must meet the FH line of a sheet of paper which is fixed in the head support. It must be
moveable in the vertical direction by moving the laser module at the cover of ORTHOPHOS 3. Unwan-
ted movement of the FH-light localizer is not permitted.

2. Vertical laser light beam.


The light beam must meet the center of the head support in a vertical direction.

59 01 603 D3285
20 D3285.103.04.01.02
368,34,37'ITL
=IEVP]1EMRXIRERGI'LIGOPMWX
Customer: _________________________ Address: ________________________

Dealer: ____________________________ Address: ________________________

Date of original installation: ____________ Date of inspection: ________________

Report of Assembly FD 2579 # _________


___________________________________
_______
SCHEDULE Yes No Remarks

All manuals are present

Test instruments as required


Manufacturer Model Accuracy Last calibrated
Voltmeter
mAmeter
Pulse counter
Any mechanical damage noticed

All labels are present and legible

All indicator lights are OK

Radiation indicator X-ray lights up, audible buzzer OK

Deadman feature OK

Power supply adequate Line voltage: . . . . . . . . . .V


Voltage drop: . . . . . . . . . V
kV – Verification is OK

Tube current is within specified limits Measurement: . . . . . . . mA

Specified exposure time, panoramic (pulses) OK Pulse count: . . . . . . . . . . . . . .

Specified exposure times, ceph (average pulses) OK 1. Average pulse count: . . . . . .


2. Average pulse count: . . . . . .
X-ray beam position, panoramic OK

X-ray beam position, ceph OK

Phantom radiograph, line distances within tolerance

The unit is in compliance with MFG specified tests and safety

Technician: ________________ Dealer: ___________________


D3285.103.04.01.02 04.2001 59 01 603 D3285
368,34,37'ITL
=IEVP]1EMRXIRERGI'LIGOPMWX
Customer: _________________________ Address: ________________________

Dealer: ____________________________ Address: ________________________

Date of original installation: ____________ Date of inspection: ________________

Report of Assembly FD 2579 # _________


___________________________________
_______
SCHEDULE Yes No Remarks

All manuals are present

Test instruments as required


Manufacturer Model Accuracy Last calibrated
Voltmeter
mAmeter
Pulse counter
Any mechanical damage noticed

All labels are present and legible

All indicator lights are OK

Radiation indicator X-ray lights up, audible buzzer OK

Deadman feature OK

Power supply adequate Line voltage: . . . . . . . . . .V


Voltage drop: . . . . . . . . . V
kV – Verification is OK

Tube current is within specified limits Measurement: . . . . . . . mA

Specified exposure time, panoramic (pulses) OK Pulse count: . . . . . . . . . . . . . .

Specified exposure times, ceph (average pulses) OK 1. Average pulse count: . . . . . .


2. Average pulse count: . . . . . .
X-ray beam position, panoramic OK

X-ray beam position, ceph OK

Phantom radiograph, line distances within tolerance

The unit is in compliance with MFG specified tests and safety

Technician: ________________ Dealer: ___________________


D3285.103.04.01.02 04.2001 59 01 603 D3285
368,34,37'ITL
=IEVP]1EMRXIRERGI'LIGOPMWX
Customer: _________________________ Address: ________________________

Dealer: ____________________________ Address: ________________________

Date of original installation: ____________ Date of inspection: ________________

Report of Assembly FD 2579 # _________


___________________________________
_______
SCHEDULE Yes No Remarks

All manuals are present

Test instruments as required


Manufacturer Model Accuracy Last calibrated
Voltmeter
mAmeter
Pulse counter
Any mechanical damage noticed

All labels are present and legible

All indicator lights are OK

Radiation indicator X-ray lights up, audible buzzer OK

Deadman feature OK

Power supply adequate Line voltage: . . . . . . . . . .V


Voltage drop: . . . . . . . . . V
kV – Verification is OK

Tube current is within specified limits Measurement: . . . . . . . mA

Specified exposure time, panoramic (pulses) OK Pulse count: . . . . . . . . . . . . . .

Specified exposure times, ceph (average pulses) OK 1. Average pulse count: . . . . . .


2. Average pulse count: . . . . . .
X-ray beam position, panoramic OK

X-ray beam position, ceph OK

Phantom radiograph, line distances within tolerance

The unit is in compliance with MFG specified tests and safety

Technician: ________________ Dealer: ___________________


D3285.103.04.01.02 04.2001 59 01 603 D3285
368,34,37'ITL
=IEVP]1EMRXIRERGI'LIGOPMWX
Customer: _________________________ Address: ________________________

Dealer: ____________________________ Address: ________________________

Date of original installation: ____________ Date of inspection: ________________

Report of Assembly FD 2579 # _________


___________________________________
_______
SCHEDULE Yes No Remarks

All manuals are present

Test instruments as required


Manufacturer Model Accuracy Last calibrated
Voltmeter
mAmeter
Pulse counter
Any mechanical damage noticed

All labels are present and legible

All indicator lights are OK

Radiation indicator X-ray lights up, audible buzzer OK

Deadman feature OK

Power supply adequate Line voltage: . . . . . . . . . .V


Voltage drop: . . . . . . . . . V
kV – Verification is OK

Tube current is within specified limits Measurement: . . . . . . . mA

Specified exposure time, panoramic (pulses) OK Pulse count: . . . . . . . . . . . . . .

Specified exposure times, ceph (average pulses) OK 1. Average pulse count: . . . . . .


2. Average pulse count: . . . . . .
X-ray beam position, panoramic OK

X-ray beam position, ceph OK

Phantom radiograph, line distances within tolerance

The unit is in compliance with MFG specified tests and safety

Technician: ________________ Dealer: ___________________


D3285.103.04.01.02 04.2001 59 01 603 D3285
368,34,37'ITL
=IEVP]1EMRXIRERGI'LIGOPMWX
Customer: _________________________ Address: ________________________

Dealer: ____________________________ Address: ________________________

Date of original installation: ____________ Date of inspection: ________________

Report of Assembly FD 2579 # _________


___________________________________
_______
SCHEDULE Yes No Remarks

All manuals are present

Test instruments as required


Manufacturer Model Accuracy Last calibrated
Voltmeter
mAmeter
Pulse counter
Any mechanical damage noticed

All labels are present and legible

All indicator lights are OK

Radiation indicator X-ray lights up, audible buzzer OK

Deadman feature OK

Power supply adequate Line voltage: . . . . . . . . . .V


Voltage drop: . . . . . . . . . V
kV – Verification is OK

Tube current is within specified limits Measurement: . . . . . . . mA

Specified exposure time, panoramic (pulses) OK Pulse count: . . . . . . . . . . . . . .

Specified exposure times, ceph (average pulses) OK 1. Average pulse count: . . . . . .


2. Average pulse count: . . . . . .
X-ray beam position, panoramic OK

X-ray beam position, ceph OK

Phantom radiograph, line distances within tolerance

The unit is in compliance with MFG specified tests and safety

Technician: ________________ Dealer: ___________________


D3285.103.04.01.02 04.2001 59 01 603 D3285
368,34,37'ITL
=IEVP]1EMRXIRERGI'LIGOPMWX
Customer: _________________________ Address: ________________________

Dealer: ____________________________ Address: ________________________

Date of original installation: ____________ Date of inspection: ________________

Report of Assembly FD 2579 # _________


___________________________________
_______
SCHEDULE Yes No Remarks

All manuals are present

Test instruments as required


Manufacturer Model Accuracy Last calibrated
Voltmeter
mAmeter
Pulse counter
Any mechanical damage noticed

All labels are present and legible

All indicator lights are OK

Radiation indicator X-ray lights up, audible buzzer OK

Deadman feature OK

Power supply adequate Line voltage: . . . . . . . . . .V


Voltage drop: . . . . . . . . . V
kV – Verification is OK

Tube current is within specified limits Measurement: . . . . . . . mA

Specified exposure time, panoramic (pulses) OK Pulse count: . . . . . . . . . . . . . .

Specified exposure times, ceph (average pulses) OK 1. Average pulse count: . . . . . .


2. Average pulse count: . . . . . .
X-ray beam position, panoramic OK

X-ray beam position, ceph OK

Phantom radiograph, line distances within tolerance

The unit is in compliance with MFG specified tests and safety

Technician: ________________ Dealer: ___________________


D3285.103.04.01.02 04.2001 59 01 603 D3285
368,34,37'ITL
=IEVP]1EMRXIRERGI'LIGOPMWX
Customer: _________________________ Address: ________________________

Dealer: ____________________________ Address: ________________________

Date of original installation: ____________ Date of inspection: ________________

Report of Assembly FD 2579 # _________


___________________________________
_______
SCHEDULE Yes No Remarks

All manuals are present

Test instruments as required


Manufacturer Model Accuracy Last calibrated
Voltmeter
mAmeter
Pulse counter
Any mechanical damage noticed

All labels are present and legible

All indicator lights are OK

Radiation indicator X-ray lights up, audible buzzer OK

Deadman feature OK

Power supply adequate Line voltage: . . . . . . . . . .V


Voltage drop: . . . . . . . . . V
kV – Verification is OK

Tube current is within specified limits Measurement: . . . . . . . mA

Specified exposure time, panoramic (pulses) OK Pulse count: . . . . . . . . . . . . . .

Specified exposure times, ceph (average pulses) OK 1. Average pulse count: . . . . . .


2. Average pulse count: . . . . . .
X-ray beam position, panoramic OK

X-ray beam position, ceph OK

Phantom radiograph, line distances within tolerance

The unit is in compliance with MFG specified tests and safety

Technician: ________________ Dealer: ___________________


D3285.103.04.01.02 04.2001 59 01 603 D3285
368,34,37'ITL
=IEVP]1EMRXIRERGI'LIGOPMWX
Customer: _________________________ Address: ________________________

Dealer: ____________________________ Address: ________________________

Date of original installation: ____________ Date of inspection: ________________

Report of Assembly FD 2579 # _________


___________________________________
_______
SCHEDULE Yes No Remarks

All manuals are present

Test instruments as required


Manufacturer Model Accuracy Last calibrated
Voltmeter
mAmeter
Pulse counter
Any mechanical damage noticed

All labels are present and legible

All indicator lights are OK

Radiation indicator X-ray lights up, audible buzzer OK

Deadman feature OK

Power supply adequate Line voltage: . . . . . . . . . .V


Voltage drop: . . . . . . . . . V
kV – Verification is OK

Tube current is within specified limits Measurement: . . . . . . . mA

Specified exposure time, panoramic (pulses) OK Pulse count: . . . . . . . . . . . . . .

Specified exposure times, ceph (average pulses) OK 1. Average pulse count: . . . . . .


2. Average pulse count: . . . . . .
X-ray beam position, panoramic OK

X-ray beam position, ceph OK

Phantom radiograph, line distances within tolerance

The unit is in compliance with MFG specified tests and safety

Technician: ________________ Dealer: ___________________


D3285.103.04.01.02 04.2001 59 01 603 D3285
368,34,37'ITL
=IEVP]1EMRXIRERGI'LIGOPMWX
Customer: _________________________ Address: ________________________

Dealer: ____________________________ Address: ________________________

Date of original installation: ____________ Date of inspection: ________________

Report of Assembly FD 2579 # _________


___________________________________
_______
SCHEDULE Yes No Remarks

All manuals are present

Test instruments as required


Manufacturer Model Accuracy Last calibrated
Voltmeter
mAmeter
Pulse counter
Any mechanical damage noticed

All labels are present and legible

All indicator lights are OK

Radiation indicator X-ray lights up, audible buzzer OK

Deadman feature OK

Power supply adequate Line voltage: . . . . . . . . . .V


Voltage drop: . . . . . . . . . V
kV – Verification is OK

Tube current is within specified limits Measurement: . . . . . . . mA

Specified exposure time, panoramic (pulses) OK Pulse count: . . . . . . . . . . . . . .

Specified exposure times, ceph (average pulses) OK 1. Average pulse count: . . . . . .


2. Average pulse count: . . . . . .
X-ray beam position, panoramic OK

X-ray beam position, ceph OK

Phantom radiograph, line distances within tolerance

The unit is in compliance with MFG specified tests and safety

Technician: ________________ Dealer: ___________________


D3285.103.04.01.02 04.2001 59 01 603 D3285
368,34,37'ITL
=IEVP]1EMRXIRERGI'LIGOPMWX
Customer: _________________________ Address: ________________________

Dealer: ____________________________ Address: ________________________

Date of original installation: ____________ Date of inspection: ________________

Report of Assembly FD 2579 # _________


___________________________________
_______
SCHEDULE Yes No Remarks

All manuals are present

Test instruments as required


Manufacturer Model Accuracy Last calibrated
Voltmeter
mAmeter
Pulse counter
Any mechanical damage noticed

All labels are present and legible

All indicator lights are OK

Radiation indicator X-ray lights up, audible buzzer OK

Deadman feature OK

Power supply adequate Line voltage: . . . . . . . . . .V


Voltage drop: . . . . . . . . . V
kV – Verification is OK

Tube current is within specified limits Measurement: . . . . . . . mA

Specified exposure time, panoramic (pulses) OK Pulse count: . . . . . . . . . . . . . .

Specified exposure times, ceph (average pulses) OK 1. Average pulse count: . . . . . .


2. Average pulse count: . . . . . .
X-ray beam position, panoramic OK

X-ray beam position, ceph OK

Phantom radiograph, line distances within tolerance

The unit is in compliance with MFG specified tests and safety

Technician: ________________ Dealer: ___________________


D3285.103.04.01.02 04.2001 59 01 603 D3285