You are on page 1of 4

Gastrointest Radiol 16:49-52 (1991)

Gastrointestinal
Radiology
9 Springer-VerlagNewYorkInc.1991

Oral Cholecystography: Comparison of Conventional Screen-Film


with Photostimulable Imaging Plate Radiographs
Deep N. Srivastava, Akhilesh Kulshrestha, Ratni B. Gujral, and Rajendra V. Phadke
Department of Radiodiagnosis, Sanjay Gandhi Postgraduate Institute of MedicalSciences,Lucknow,India

Abstract. Oral cholecystography was performed on tains a dominant position because a high-quality
ultrasonically proven cases of choMithiasis. The radiographic image contains at least 4-6 mega-
conventional screen/film system and computed ra- bytes, whereas CT images contain 0.5 megabytes,
diography (CR) using the imaging plate were used magnetic resonance imaging 0.3 megabytes, and
under practically identical conditions. The diag- ultrasound only 0.07 megabytes [3]. A technologi-
nostic accuracy of the two techniques was assessed cal difficulty to overcome is to eliminate such big
objectively, and the image quality was assessed differences in the amount of recorded information
subjectively. The CR image proved to be better through the development of a practical digital ra-
than the conventional screen/film image, even with diography to replace the analog x-ray film/screen
the low-dose exposure. This paper describes our system [i.e., a computed radiography (CR)].
experience in the area of the biliary tract system A number of studies have reported the applica-
with CR - a digital radiographic system using pho- tion of digital techniques to plain radiography
tostimulable phosphor. [4-7]. Because of the wide-exposure latitude (sys-
tem dynamic range) of the photostimulable imag-
Key words: Gallbladder, radiography - Computed
ing plate technique, radiation dose to the patient
radiography, technique - Abdomen, imaging sys-
can be reduced by decreasing mAs levels compared
tems.
with those required for a conventional screen/film
combination. Due to the automatic sensitivity con-
trol, quality images of the same standard are pro-
Diagnosis of biliary tract diseases has made a re- duced which are comparable every time and re-
markable progress following the recent develop- takes are avoided. In addition, digital radiographs
ment of medical imaging techniques, such as ultra- can be manipulated and postprocessing can be per-
sound and computed tomography (CT). Sonogra- formed to alter the enhancement of edges, contrast,
phy, a noninvasive and simple procedure without and brightness of the image.
the risk of radiation, has become the method of A prospective evaluation of diagnostic accura-
choice in biliary tract imaging. Despite the increas- cy and image quality of screen/film and the CR
ing availability and popularity of ultrasound, oral system in oral cholecystography was undertaken
cholecystography still remains the most common to compare the results.
investigation for suspected diseases of the gall
bladder because of its reliability in detection of
Materials and Methods
gall stones and assessment of gallbladder function,
particularly when oral dissolution therapy is to be Oral cholecystographywas performed on ultrasonicallyproven
started [1, 2]. Technological innovations have led cases of cholelithiasis.The conventionalscreen/filmsystemand
the CR systemwereused under practicallyidenticalconditions.
to development of new digital imaging systems in The study was conducted on a remote-controlledtable of Shi-
medicine, but still conventional radiography main- madzu unit (ZS-40)having provision of an imagingplate maga-
zine. This machine is specially designed for gastroenterology
Address offprint requests to: Deep N. Srivastava, M.D., Depart- work with conventional screen/film, as well as imaging plate
ment of Radiodiagnosis, Sanjay Gandhi Postgraduate Institute magazine for CR. The CR system used (model 901; Fuji) has
of Medical Sciences,Lucknow-226001,India been describedin detail elsewhere[8]. Conventionalscreen/film
50 D.N. Srivastava et al. : Computed Radiographic Imaging

Fig. 1. Normal gallbladder opacified by oral cholecystography. screen/film image was graded (+), an equivalent image (+),
A Enhanced CR image obtained with a one-half exposure dose and an inferior image ( - ) . A few examples derived from our
is shown on the top, and the normally processed conventional clinical experience with the CR system are herein illustrated
radiograph on the bottom. B Enhanced image with one-quarter (Figs. 1 and 2).
dose (top) as compared with conventional view.

Fig. 2. Gallbladder containing numerous radiolucent stones. A Results


Enhanced CR images at full exposure (top) as compared to
standard film (bottom). B Enhanced image at a one-half dose The details of the observations are given in Ta-
exposure and its comparative radiograph (bottom row). C En- bles 1 and 2. The quality of image was improved
hanced image at one-quarter exposure and its comparative ra-
diograph (bottom row). in more than three quarter of the cases in CR as
compared to the conventional screen/film combi-
nation (Table 1). There was no loss in diagnostic
information by reducing the exposure time up to
radiography was performed by using a high-speed K Y O K K O
screen and double-coated polyester-based x-ray film, the expo- one fourth; however, the noise (granularity) was
sure conditions were a focal film distance (FFD) of 115 cm, increased by reducing the exposure dose by more
70-85 kV, and 140-160 mA. For CR the imaging plates were than one half (Table 2). This increase in granular-
used (instead of conventional screen/film) with the same, one ity can be explained by the fact that in such low-
half, and one fourth the exposure dose. The radiation exposure
dose images, the granularity is mainly determined
was reduced by decreasing the exposure time. Other parameters,
such as F F D and kV, were kept the same. Two images were by the quantum noise of the x-ray exposure.
obtained which were set as appropriate in advance (left image In many of the cases the size and number of
close to the conventional screen/film image and right image stones were better appreciated on CR when com-
with intensive image processing). Four parameters were taken pared to the conventional screen/film system,
for evaluation of the gall bladder (e.g., wall, fundus, body,
neck, and gall stone). The image of each region was compared
which may prove to be a helpful feature in evalua-
with the conventional screen/film image and was graded on tion of patients who are on oral dissolution thera-
a three-point scale. A CR image superior to the conventional PY.

Table 1. Comparative evaluation of CR vs. conventional radiographs

Image quality grades Area of interest

Margins Fundus Body Neck Gall stone (number, size, outline)

Superior ( + ) 80% 77% 77% 83.3 % 80%


Equivalent ( _ ) 6.6% 6.6% 6.6% 6.6% 16.6%
Inferior ( - ) 13.3% 16.6% 16.6% 10% 3.3%
D.N. Srivastava et al. : Computed Radiographic Imaging 51

Table 2. Evaluation of CR images at different exposure doses I X-RAY TUBE I


9
Exposure Same dose 1/2 dose 1/4 dose I PATIENT I
t
IMAGING PLATE
Diagnostic Optimum Optimum Optimum (re~ords an X-ray Image)
information t
Image quality Optimum Optimum Reduced IMAGE READER
Granularity Optimum Optimum Increased (Converts X-ray Image to electric signals)
t
IMAGE PROCESSOR
(does gradation and frequency processing)
t
IMAGE RECORDER ~ AUTOMATIC
Discussion (Converts electric signals to
light for film recording) 11 FILM PROCESSOR

The basic concept of the CR system is to reduce


Fig. 3. Generalized block diagram of CR system.
the exposure dose to a minimum; to provide rich
diagnostic information by producing consistently
sharp images that offer a wide latitude and allow
image processing; and to convert x-ray image in- cessing may also be normally controlled by using
formation into digital signals to be stored for fu- an image-processing console.
ture retrieval and distant transmission. The major clinical merit of the CR system is
In Fig. 3 a generalized block diagram of the that it presents constantly clear images of high
CR system is shown. The basic x-ray image detec- quality; this was confirmed with respect to the gall
tor is the imaging plate, made by densely applying bladder in this study (Tables 1 and 2). Calcifica-
crystals of phosphors on a support layer of a po- tions were shown clearly due to the wide dynamic
lymer material. When exposed to x-rays, the energy range and the image processing of the CR system.
is temporarily stored within the crystals and emit- However, the gall stones particularly at the base
ted as fluorescent blue light when an He-Ne laser of the gall bladder and in cases where the density
beam scanning light is irradiated on it. The optical of the image of the contrast medium was low could
signals are then converted to time-serial electric be more easily demonstrated. The images at one
signals, and after A/D (analog to digital) conver- half and one fourth of the normal exposure dose
sion, image processing, D/A (digital to analog) were practically equivalent to conventional screen/
conversion, and conversion to optical signals, an film images as a result of the high sensitivity.
x-ray image is recorded by a laser beam on a film. Costs involved in digital imaging have not been
The residual image can be erased from the imaging fully analyzed [9, 10]. The costs in CR system are
plate allowing it to be used again and again. dependent upon the number of images that one
The major image-processing techniques of CR wishes to save as hard copy. The cost will be fur-
consist of gray-scale conversion and spatial-fre- ther reduced because the number of repeated ex-
quency enhancement for the specified spatial-fre- aminations can be dramatically reduced with the
quency range. A CR processed by the image pro- use of the imaging plate technique. In a cost analy-
cessor is printed as a hard copy on a single-coated sis by Nudelman et al. of a photoelectronic digital
transparent film by a high-resolution laser printer. radiology department, it was predicated that, al-
Two differently processed images obtained from though start-up costs are high, over time the
one x-ray exposure are printed on a single film. photoelectronic-digital systems promise to be sig-
We find it appropriate to process one image to nificantly less expensive than a film-based system
resemble the characteristic curve of a conventional for long-term hospital use [11].
screen/film radiograph (normally processed im-
age), mainly by gradation conversion, and to pro-
cess the complementary image by gray-scale con- References
version to expand the image contrast and its spa-
1. Sutton D : A Text Book of Radiology and Imaging. London:
tial-frequency enhancement (enhanced image). Churchill Livingstone, 1987
Normally processed images reflect the degree of 2. Berk RM, Clemett AH : Radiology of the Gall Bladder and
x-ray attenuation of the various tissues, but the Bile Duets. Philadelphia: W.B. Saunders Company, 1977
enhanced image allows better visualization of the 3. Takeshi I : Jpn Radiat Technol Soe J 40:1004~1015, 1984
4. MacMahon H, Vyborny CJ, Metz CE, et al. : Digital ra-
contour of the object. The image processor au- diography of subtle pulmonary abnormalities: An ROC
tomatically selects the type of processing required study of the effect of pixel size on observer performance.
for a given examination. The type of image pro- Radiology 158: 21-26, 1986
52 D.N. Srivastava et al. : Computed Radiographic Imaging

5. Goodman LR, Foley WD, Wilson CR, Rimm AA, Lawson 9. Dwyer SJ, Templeton AW, Martin NL, et al.: The cost
TL: Digital and conventional chest images: Observer per- of managing digital diagnostic images. Radiology
formance with film digital radiography system. Radiology 144:313-318, 1982
158:27-34, 1986 10. Merritt CRB: Computed radiography: A new approach to
6. Lares PM, Cocklin ML: Spatial resolution requirements plain film imaging. Diagn Imaging 7:58-65, 1985
for digital chest radiographs: ROC study of observer 11. Nudelman S, Healy J, Capp MP: A study of photoelec-
performance in selected cases. Radiology 158:11 19, tronic digital radiology. II. Cost analysis of a photoelec-
1986 tronic digital vs. film-based system for radiology. Proc IEEE
7. Curtis D J, Ayella R J, Whitley HG, et al. : Digital radiogra- 70:708-714, 1982
phy in trauma using small dose exposure. Radiology
132: 587-591, 1979
8. Tateno Y: Clinical application of FCR. J Med Imag 4
(suppl 1):7-15,/984 Received: April 2, 1990; accepted: May 4, 1990

You might also like