You are on page 1of 43

CHEMCHLOR® SCM

AUTO DOSING SYSTEM

OPERATION MANUAL

REVISION JUNE 2016

P age |1
TABLE OF CONTENTS

1.0 SAFETY INSTRUCTION 4


1.1 GENERAL 4
1.2 GENERAL SAFETY RULES 4
®
1.3 SYMBOLS USED IN THE CHEMCHLOR - SCM 4
2.0 CHEMCHLOR® STREAMING CURRENT MONITOR 5
2.1 FEATURES 5
2.2 PRODUCT SPECIFICATIONS 5
3.0 THEORY OF OPERATION 6
3.1 TREATMENT OF WATER FOR CLARIFICATION 7
3.2 CHARGE ANALYSIS 8
4.0 INSTALLATION 10
4.1 SAMPLING POINT 10
4.2 TYPICAL INSTALLATION 10
4.3 POWER SUPPLY REQUIREMENTS 11
5.0 OPERATION AND SETTINGS 12
5.1 SENSOR / SAMPLING CHAMBER 12
5.2 TOUCHSCREEN ANALYZER CONTROL PANEL 13
5.2.1 CHEMCHLOR® E-20 13
5.2.2 CHEMCHLOR® EX-10 14
5.3 TOUCHSCREEN ANALYZER SETTINGS 14
5.3.1 HOME SCREEN 15
5.3.2 SCU DISPLAY PAGE 16
5.3.3 ALARM DISPLAY PAGE 17
5.3.4 TRENDING PAGE 18
5.3.4.1 CHART DATA 19
5.3.4.2 TOTALIZER DATA 20

P age |2
5.3.5 SETTINGS 21
5.3.5.1 BASIC SETTING 22
5.3.5.2 ALARM SETTING 23
5.3.5.3 DATE/TIME SETTING 24
5.3.5.4 FLOW SETTING 25
5.3.5.5 METERING PUMP 27
5.3.5.6 INPUT SCALING SETTING 28
5.3.5.7 PID AUTO-TUNING 29
5.3.6 SCM SENSOR SETTING 31
6.0 CALIBRATION 32
6.1 CALIBRATION PROCEDURES 32
7.0 ROUTINE MAINTENANCE 33
7.1 CHEMICAL CLEANING 33
7.2 MANUAL CLEANING 33
8.0 TROUBLESHOOTING 35
8.1 DIAGNOSTIC TABLE 35
8.2 TECHNICAL AND CUSTOMER ASSISTANCE 36
9.0 ACCESSORIES AND REPLACEMENT PARTS LIST 37
APPENDIX A - CHEMCHLOR® E-20 ELECTRICAL DIAGRAM 38
APPENDIX B - CHEMCHLOR® EX-10 ELECTRICAL DIAGRAM 41

P age |3
1.0 SAFETY INSTRUCTION
1.1 GENERAL

The manufacturer is not responsible for any damages due to misapplication or misuse of this product
including, without limitation, direct, incidental and consequential damages, and disclaims such
damages to the full extent permitted under applicable law. The user is solely responsible to identify
critical application risks and install appropriate mechanisms to protect processes during a possible
equipment malfunction.
This manual contains basic instructions that must be followed during the commissioning, operation,
care and maintenance of the instrument. Please read this entire manual before unpacking, setting
up or operating this equipment. Pay attention to all danger and caution statements. Failure to do so
could result in serious injury to the operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do not use or install this
equipment in any manner other than that specified in this manual.

1.2 GENERAL SAFETY RULES

a) Read the operation manual thoroughly before starting up the system.

b) The CHEMCHLOR® - SCM must be used by trained and authorized personnel only.

c) The system may be used only in compliance with the technical data and specifications given in the
operating instructions.

d) It is forbidden to use the system for any other purpose, or to modify it in any way without the
consent from the manufacturer.

e) Keep equipment clean for best and safest performance. Follow instructions given in the manual for
proper maintenance.

f) Before servicing the system, disconnect equipment from power source.

1.3 SYMBOLS USED IN THE CHEMCHLOR® - SCM

Standard IEC symbols are used on the high voltage cover.

ISO 3864, No. B.3.6 Caution, risk of electric shock.


This symbol indicates that hazardous voltages may be present under this cover

ISO 3864, No.B3.1 Caution refers to accompanying documents.


This symbol is a reminder to read the sections in the manual referring to the electrical
connections, and potential hazards.

P age |4
2.0 SAFETY INSTRUCTION
2.1 FEATURES

CHEMCHLOR® SCM is CHEMKIMIA’s innovative engineering design which incorporate the SCM with
Programmable Logic Controller (PLC) and controlled with a touch screen monitor. This SCM system
with PLC is able to conduct auto dosing if connected with 4 - 20mA dosing pump. Another version
with pre-assembled inverter enables auto dosing using only basic dosing pump.

This fully Auto SCM dosing system will be able to handle the dosage of coagulant chemical (especially
ACH) without the aid of an operator to conduct jar test as the Auto Dosing System will regulate the
dosage of coagulant chemical continually and maintain the optimum dosage to the treatment plant
to work consistently and cost effectively.

CHEMCHLOR® SCM AUTOMATIC DOSING SYSTEM have features of :-

 Trending of SCU value


 Trending of raw and ACH (Aluminium Chlorohydrate - Coagulant) flow
 ACH calculated consumption per day, per month and per duration
 HMI control with auto dosing and manual (constant concentration & constant L/hr)
 Secure access to operation with password (CHEMCHLOR® SCM-E20 only)
 Data log function (CHEMCHLOR® SCM-E20 only)
 Web ready HMI for remote access (CHEMCHLOR® SCM-E20 only)

2.2 PRODUCT SPECIFICATIONS

Measurement Range ± 10.00 SCU


Accuracy ± 1 % of Full Scale
Repeatability 1%
Resolution 0.01 SCU
Response Time 1 second
Operating Temperature 0°C – 50°C
Wetted Materials HDPE, PTFE, Stainless Steel, Neoprene
Ingress Protection IP65 – according to EN 60529 / DIN VDE 0470-1
Power Source 100 - 240 VAC±10%, 50/ 60 Hz
Warranty 1 Year from date of shipment

P age |5
3.0 THEORY OF OPERATION
In a liquid form, water molecules move around each other at a fast rate. One effect of this fast
movement is the ability to suspend matter. This phenomenon is called “Brownian Motion”. It occurs
when microscopic particles are maintained dispersed in suspension due to their random
bombardment by the fast movement of water molecules. Typical particles found in raw water
entering WTP have finely divided clay particles and organic matter collectively called silt.

A second phenomenon which stabilizes the suspension is the surface charge of the suspended
matter. When a salt such as sodium chloride is place in water, complete dissolution occurs. This
system reaches a stable energy level when the individual sodium and chloride ions (Na+ and Cl-) are
separated in the water phase by being surrounded by water molecules.

In the case of large pseudo salts, e.g. Aluminosilicates (clay), only partial dissolution takes place due
to incomplete breakdown of the crystal to individual ions. The structure of these clays is similar to
silica or sand except that random silicon atoms in the crystal are replaced by aluminum atoms in the
cage structure, causing the clay to swell and crack between adjacent aluminum atoms in the crystal.
Thus a clay particle is formed with a size of less than 1 micron with a negative charge. This particle is
small enough to be maintained in suspension by Brownian Motion. The particles in the suspension
repel each other due to their surface charge, preventing them from coming together and
agglomerating, or flocking to form a larger particle, which would settle out. The result is an
energetically stable system and is the reason why the particles remain dispersed.

The counter ions (say sodium for the sake of argument) are separated from the large cage structure
because they are dissolved in the water. Clay particles have a negative charge associated with it,
while the counter ions, typically cat-ions (or positively charged ions) are dispersed in the water
phase.

In the case of most naturally occurring substances, the larger ion, when in suspension, has a negative
net charge (anionic). The smaller, counter ion is positive (cationic). The residual charge of the larger
particles is negative, which causes them to repel each other, preventing them from forming
agglomerates. The size of the particles never becomes large enough to settle out, so they remain
dispersed in suspension.

This phenomenon creates an energetically stable system. In order to cause the suspended particles
to agglomerate and settle out, the energy of the system must be upset. There are numerous
mechanical means to accomplish this, but the addition of chemical flocking agents to the suspension,
drastically reduces the time and is far more efficient.

Chemical additives perform two functions, charge neutralization & bridging. Both of these techniques
allow the small particles to floc and grow sufficiently that Brownian motion can no longer support
them. Due to the high density of the particle, flocs will form and settle as fine sludge.

P age |6
Additive Chemicals
+
+
-
- + +
+
+ +
- -
- + + +
+ -
- - - - +
-
- + + -
+ - -
+ - - -
+
-
+ - + - +
+ + +
+
- +
+ -
+ - +
+ +
+
+ -
- - +
-
+ +
+ + +
-
+
+
- +
-
- - - -
Contaminant Particles
+ + -
-
+
+ -
+
-
+
+

1. Charge Neutralization 2. Bridging

Figure 1: Effects of Chemicals

3.1 TREATMENT OF WATER FOR CLARIFICATION

Most water treatment chemicals consist of a cationic (positively charged) chemical e.g. aluminum
salts, ferric salts, polyamines or cationic polyacrylamides, some of which have the cationic part
tagged on to a long polymer chain. As stated earlier, raw water entering the WTP is an energetically
stable system of suspended particles with a net negative charge. Cationic chemicals are added to
bring the charge to neutral.

Before the development of Streaming Current technology, the best way to determine the optimum
dosage has been the jar test method. The jar test involves taking a representative sample of the
water being treated and placing it in several jars. Different amounts of clarifying chemicals are added
to each jar; stirred and comparing the clarity of the water in the different jars. Jar tests are time
consuming and it is difficult to reproduce the conditions of the WTP in a jar. The tests can take
several hours rendering them useless when plant personnel are really responding to rapid changes in
water quality. A typical curve of Turbidity vs. Chemical dosage is shown in Figure 3.
Some considerations when treating the water are the rate of floc formation, the size of the floc
formed, how fast the floc settles, and the clarity of the final settled and filtered water.

P age |7
Other techniques exist, such as a dosing curve, which indicates a recommended dosage for a given
water turbidity. This is generally built up over years of dosing experience with the water, but has the
disadvantage that turbidity caused by extremely small particles requires a higher dosage than that
caused by larger particles, and therefore can only be adapted for use with known type turbidity on
any given water.

TURBIDITY

CHEMICAL DOSAGE ppm

Figure 2: Turbidity vs Dosage

3.2 CHARGE ANALYSIS

Charge analysis is the measurement of the electro kinetic charge of a solution due to the presence of
charged particles. The electro kinetic charge can be measured by a number of different methods.

1. Applied electric field


Measurement: The relative mobility of the solid or liquid phase.

This is the first method developed for calculating the Zeta Potential. The motion of charged particles
under the influence of an electric field was observed and the potential required to achieve a certain
amount of particle mobility was measured.

A cell consisting of two flat plates separated by approximately 0.1 mm and having an electrode at
each end of the cell is filled with water containing suspended matter. When an electrical potential is
applied to the electrodes, the particles can be observed to drift toward one of the electrodes. The
Zeta Potential is calculated from the measured speed of particle drift.

P age |8
2. Induced Electrical Potential
Measurement: The potential developed as the result of forced movement of particles in the solution.
This is the method used by the CHEMCHLOR®- SCM. Continuous sample water is directed into an
annulus, inside which a displacement piston oscillates vertically at a fixed frequency. This action
causes the liquid to move between the two stainless steel electrodes. The suspended particles are
absorbed onto the walls under the action of Van der Waal’s and electrostatic forces. As the sample is
moved rapidly back and forth, mobile counter ions surrounding the colloids are sheared near the
surface of the walls and moved past the electrodes. The resultant A.C. signal or Streaming Current,
proportional to charge density, is electronically processed and displayed.

P age |9
4.0 INSTALLATION
4.1 SAMPLING POINT

Careful consideration must be given to where in the system the sample will be taken. The
CHEMCHLOR SCM requires sampling of the coagulated water (sample after introduction of coagulant)
and sometimes where pre-pH adjustment is completed. It is critical that sample point should be at
least 10 pipe diameters away from the dosing point to ensure ample mixing.

Equally important is the lag time from when a change in dosage occurs to when it shows up at the
sensor. If the lag time is too long the control of loop will not be possible. A lag time not greater than
10 minutes is recommended.

4.2 TYPICAL INSTALLATION

Raw Water Coagulation Flocculation Clarifier

Coagulant
Tank

P a g e | 10
The sampling water (coagulated water) shall be connected to the sample chamber at the flow rate of
600 to 800 LPH. Run the flow close to maximum as possible without overflowing the sample box and
then adjust the drain valve to drain off large particles while ensure sensor is always immersed in the
sample water. Low level of sample will activate the sampling pot low level alarm.

The sample flow must be free of large shells or other debris that might clog the orifices or damage to
the sensor.

Note: The sample flow must be free of large shells or other debris that might clog the orifices or
cause damage to the sensor. Supplying an adequate flow free of debris is the responsibility
of the installer.

Drains MUST be routed to a suitable drain. DO NOT reintroduce this water back into the
process stream.

4.3 POWER SUPPLY REQUIREMENTS

Insufficient voltage from factory power source may affect the power output of the motor and the
function of the controller. It is important to connect this machine to the correct voltage in the factory
power source. Use only an independent power source. Table for power supplies requirement:

Model Voltage Current Breaker capacity

CHEMCHLOR® E-20 100 - 240V 13 A 40 A

CHEMCHLOR® EX-10 100 - 240V 13 A 40 A

Note: Only qualified electricians should be allowed to perform the installation


of the instrument as it involves a line voltage that could endanger life.

P a g e | 11
5.0 OPERATION AND SETTINGS
The CHEMCHLOR SCM system consists of three major components, the Sensor, the Sampling
Chamber, and the Touch Screen Control Panel. Some system may also come with the Dosing Pump
skid depending on the customer requirements.

5.1 SENSOR / SAMPLING CHAMBER

The sensor module sits on top of the sampling chamber, with the probe end below the water level.
The sample chamber has two valves to adjust the flow :

 The inlet should be adjusted such that there is always a sample present at recommended flow
600-800 lph.
 The main drain needs to be adjusted open to allow the larger particles to drain while sample
water is measured
 The overflow pipe is usually connected to the main drain and is free flow, its purpose is to
keep the sample chamber from overflowing.

To prevent back flow and allow proper draining it is important that both drain are left open to the
atmosphere and kept as short as possible.

If heavy particulates can be present in the water it is important to install a 40 mesh strainer before
the inlet. The flow should also be kept low to allow large sand and larger debris to fall to the bottom
of the sample chamber and drain, without causing harm to the sensor.

P a g e | 12
5.2 TOUCHSCREEN ANALYZER CONTROL PANEL

Details are not provided on individual settings as most are self-explanatory. The control panel is
housing the touch screen for all the PLC settings and switch button as shown in diagram below. Most
of the parts are plug and play, connect the SCM sensor cable and main power cable to the control
panel port provided on the right hand side of the panel and system is ready to measure the charges
of the coagulated water once power is turned on. The power requirements for the system are
100 - 240 VAC (universal voltage).

5.2.1 CHEMCHLOR® E-20

ALARM INDICATOR
1) SCU HIGH ALARM
2) SCU LOW ALARM
3) SCM GENERAL ALARM

10" TOUCH SCREEN

POWER BUTTON

INDICATOR LAMPS M12 SIGNAL CONNECTORS


1) POWER ON
2) SCM SENSOR ON
3) MP 1 ON
4) MP 2 ON
5) BUZZER

OPERATION SWITCHES
1) AUTO / OFF / MAN
2) SCM SENSOR OFF / ON
3) LOCAL MP 1 OFF / ON
4) LOCAL MP 2 OFF / ON
5) GENERAL RESET

P a g e | 13
5.2.2 CHEMCHLOR® EX-10

7" TOUCH SCREEN

OPERATION SWITCHES /
POWER BUTTON
LAMPS
1) SYSTEM ON
2) SCM SENSOR ON M12 SIGNAL CONNECTORS
3) MP 1 ON
4) MP 2 ON

OPERATION SWITCHES /
LAMPS
1) AUTO / OFF / MAN
2) SCM SENSOR ALARM
3) ALARM RESET
4) BUZZER

P a g e | 14
5.3 TOUCHSCREEN ANALYZER SETTINGS

5.3.1 HOME SCREEN

7
1
8
2
9 3
10 4

11 5

12 6

13

14

15

HOME TREND SETTING ALARM LIST BRIGHTNESS LOGIN MUTE

1. Activate Auto Dosing Mode 9. Current Dosing Rate in L/hr

2. Activate Manual Dosing Mode 10. Raw Water Flow Rate

3. SCM Sensor Status 11. Coagulant Consumption

4. Link to SCU Display Page 12. Dosing Pump Status

5. Current SCU Value 13. Dosing Pump Dose Rate

6. Current SCU Setpoint 14. Coagulant Storage Tank Level

7. Alarm In Progress 15. Manual Dosing Rate

8. Current Dosing Rate in mg/L


P a g e | 15
5.3.2 SCU DISPLAY PAGE

HOME TREND SETTING ALARM LIST BRIGHTNESS LOGIN MUTE

1. Current SCU Reading

2. 24 Hour Live Trending Chart

3. Current SCU Setpoint

4. Access SCM Sensor Settings

P a g e | 16
5.3.3 ALARM DISPLAY PAGE

In the alarm display page, the user is able to view all the current ongoing alarms and also those that
have been triggered previously. The occurrence column shows the time when the alarm is first
triggered. The message column will display in detail the alarm while recover will show the time when
the alarm is recovered.

If the user pressed the Ack & Reset button, the time when the button is pressed will be stored into
the acknowledge column stating the time the alarm is attended by the operator. It is possible to
acknowledge an alarm after it is recovered.

If there are more than a page of recorded alarms, the user can press the Cursor button and then
access the previous records using the Up and Down buttons.

P a g e | 17
5.3.4 TRENDING DISPLAY PAGE (CHEMCHLOR® E-20 ONLY)

Inside the Chart/Data Menu, there are selection of trending charts and recorded data to view. The available
data to view are as divided according to the buttons as below :

1) SCM Value - Trending of SCM Present Value, Setpoint Value, High Alarm and Low Alarm Limits

2) Water Flow - Trending of Raw Water Flow Rate, Coagulant Dose Rate and Treated Water Flow Rate

3) Dosing Pump - Trending of MP 1 and MP 2 Dosing Rate

4) Totalizer - Totalizer Values for Coagulant, Raw Water and Treated Water

5) ACH Usage - Trending of Coagulant usage

P a g e | 18
5.3.4.1 CHART DATA (CHEMCHLOR® E-20 ONLY)

®
The diagram above shows the typical trending page for the recorded data in the CHEMCHLOR E-20 unit. The
chart is displayed in typical X-Y chart fashion. The view historical data, press the Real-time button to change it
to Historical. Once in historical mode, you can press the rewind forward and backward buttons to review past
and present records.

If there is a need to scrutinize a particular point in the chart, simply click on the point in the chart and a cursor
will appear indicating the current cursor location. The value for the cursor is shown at the bottom of the chart
inside the bounded box. To fine-tune the cursor position, press the forward or backward button flanking the
reading to adjust the cursor position in small increments. The value and recorded time will be updated
accordingly. To review data of different chart line color, simply press on the reading bounding box and the
reading of the next chart line will be shown. The color of the text will match the color of the chart line to
indicate which chart line the reading belongs to.

To view the chart in different time scale, press the 1 Hr, 8 Hr or 24 Hr scale and the chart will update
accordingly to the selection. The Legend button provides the legend for the chart lines shown on the screen.

P a g e | 19
5.3.4.2 TOTALIZER DATA (CHEMCHLOR® E-20 ONLY)

In the Totalizer Data page, the user is able to review the totalizer data for today, yesterday and also total
cumulative. Totalizer values are recorded for the Coagulant, Feed Water and also Treated Water. To reset the
totalizer data, please contact our representative for assistance.

P a g e | 20
5.3.5 SETTINGS PAGE

From the Settings Menu, there are 7 sub-menus to properly setup the SCM Analyzer to function as needed by
site requirements. The sub-menus are detailed as below :

1) Basic Setting - Settings covering basic stuffs from pump changeover time, password and factory reset

2) Alarm Setting - Allows the user to alter alarm delay time, alarm trigger setpoints

3) Date/Time - Adjustment of the system time to allow proper pump changeover and data recording

4) Flow Setting - Setting of the method used for the system to calculate raw water flow rate

5) Metering Pump - Settings detailing the metering pump used with the system

6) Input Scaling - Setting of associated external sensor measuring values

7) PID Auto-Tuning - Perform a PID Auto-Tuning sequence to tune the system to the site conditions

P a g e | 21
5.3.5.1 BASIC SETTING

Pump Changover Time - Key in the time in 24hr format for the metering pumps to changeover on daily basis

Modbus Address (1-127) - Set the system Modbus address if the optional output accessory is purchased

Settings Password - Key in the 4-digit password if the password function is enabled. Press the toggle button to
disable password function if it is enabled

Normal / Demo Mode - Switch between normal and demo mode of operation. During normal use, the system
is set to normal mode. When in demo mode, the system does not respond to actual
SCM sensor input. The user is able to key in the current PV value to observe the
changes of the system response in accordance to the SCU value. Please take note
that the pump still functions as usual in demo mode, only the SCM sensor input is
disabled.

Perform Factory Reset - Perform a factory reset of the settings in the system. Do not perform factory reset
unless it is necessary!

P a g e | 22
5.3.5.2 ALARM SETTING

SCU High Alarm - Setpoint of the SCU high alarm trigger

SCU Low Alarm - Setpoint of the SCU low alarm trigger

Alarm Delay Timer - Delay time in seconds before triggering an alarm to prevent false alarms

Buzzer On / Off Timer - Time in seconds to on and off the buzzer to create a unique buzzer call

P a g e | 23
5.3.5.3 DATE/TIME SETTING

Time - Set the current time in HH:MM:SS format HH is in 24 hours format

Date - Set the current date in DD/MM/YYYY format

Day of Week - Set the day of the week, 0-6 assuming 0 represents Sunday

When done setting, press the Save Date/Time button, otherwise the changes will not be saved.

P a g e | 24
5.3.5.4 FLOW SETTING

Flow Signal Option - Choose A if there is a 4-20mA flow meter input connected to the SCM Analyzer
Choose B if raw water pump start signal is connected to the SCM Analyzer
Choose C if there is no form of input for raw water flow rate available to system, user will
manually key in the flow rate to the system

Feed Pump 1 - 4 Capacity - Enter the pump capacity in % for up to 4 raw water pumps linked to the SCM
analyzer. If there is only one pump connected, just key in 100% for Feed Pump
1 capacity and 0% for all others. The total for the 4 pumps must add up to
100%. A simple arithmetic check is in place to facilitate data keyed in.

Feed Capacity When No Flow Signal - Enter the expected raw water flow rate if there is no sensors or pump
start signal available to the SCM Analyzer

Press the right arrow to proceed to the second page of the Flow Setting

P a g e | 25
Low Flow System Cut-off - Set the low cut-off value for raw water flow rate for the system to treat as no flow.
Any raw water flow rate below this value will be treated as zero flow by the system.

Low Flow Cut-off Delay - Delay time in seconds before system treats the system as zero raw water flow input.
When there is no flow to the system, the system will hold the last SCU value
(it will hold the last value till raw water resumes to facilitate PID control).
Also during zero raw water flow, the metering pump outputs are disabled, meaning
no coagulant will be dosed into the system.

P a g e | 26
5.3.5.5 METERING PUMP SETTING

MP1 & MP2 Min, Max Frequency - Enter the proper minimum and maximum speed of the connected
metering pumps in units of Strokes/min. The toggle switch at the side is
used to enable/disable the said metering pump.

MP Flow Capacity at 100% (Single Pump) - Enter the maximum dosing capacity of a single metering pump in
unit of L/hr.

Coagulant Specific Gravity - Enter the specific gravity of the coagulant used in the system. This value is used to
calculate the dose rate and also the consumed coagulant totalizer.

P a g e | 27
5.3.5.6 INPUT SCALING SETTING

Raw Water Flow - Enter the 4mA (Min Value) and 20mA (Max Value) of the connected raw water signal input.

Treated Water Flow - Enter the 4mA (Min Value) and 20mA (Max Value) of the connected treated water signal
input.

ACH Dose Flow - Enter the 4mA (Min Value) and 20mA (Max Value) of the connected coagulant flow input
signal.

SCM Instrument - Enter the minimum and maximum value of the measuring range for the SCM sensor (largest
allowable range, -10.00 to 10.00)

Please disregard the settings if there are no associated sensors connected to the SCM Analyzer

P a g e | 28
5.3.5.7 PID AUTO-TUNING

SCU Value, S1, S2 - Enter the first and second dosing point (in SCU) for the PID Auto-tuning sequence to be
carried out. Note that the entered points should be higher than current SCU reading
otherwise it will not work.

Dosing Rate, X1, X2 - Enter the dosing rate associated with the S1 and S2 reading points. The auto-tuning
process will run at dosing rate X1 until it achieve value of S1, then ramp up to X2 rate
till it achieve the point of S2.

Flow Rate, F - This value is non-editable and is updated according to current raw water flow rate.

Cal, Saved P Value - The value in red is the current calculated P value while the value in blue is the current
value used by the system. When the user click the save button, the value in red is
copied to value in blue.

Cal, Saved I Value - The value in red is the measured I value while the value in blue is the current value used
by the system. The value is obtained as time taken to achieve S2 from S1 during tuning
step started by pressing the Tune button. Save will copy the value in red to value in blue.

P a g e | 29
During the PID auto-tuning step triggered by pressing the Tune button, a pop-up window will appear showing
the progress of the tuning. During the step, you can set the Allowed Deviation to allow the system to proceed
to the next step if it is unable to achieve the target SCU setpoint S1 or S2. If the system unable to achieve
either S1 or S2 in 10 min, the system will generate and error and the auto-tuning process is halted. If the
system manage to achieve both values, the calculated I value will be updated in the PID Auto-tuning setting
page. Pop-up window will auto close after completion.

P a g e | 30
5.3.6 SCM SENSOR SETTING

Power Reset Trigger - Delay time to power reset the SCM sensor in case of SCM sensor alarm. After 3 times of
power reset within a span of 10 minutes, the system will no longer auto-reset the SCM
sensor and it would require manual intervention to remove the SCM sensor alarm.

Damping Samples (1-10000) - Sets the amount of samples used in the moving average algorithm to calculate
the SCU value. Increase the value if there is too much erratic movement on the
SCU value. Decrease to improve system response.

SCU Reading Offset - Set a offset to obtained SCU value.


Only use this function if necessary and always set it to 0 before performing a calibration!

Calibrate - Press this button to initate a calibration sequence. Only initiate this sequence after the sensor is
allowed to stabilize for a moment in the calibration solution. Calibration will take 60 seconds. After
successful calibration, the SCU reading should read close to 5.00SCU. If there is a problem during
the calibration, a calibration error message will be shown.

P a g e | 31
6.0 CALIBRATION
The CHEMCHLOR SCM sensor has been calibrated at the factory prior to shipment, however to
ensure accuracy it is recommended that the instrument be calibrated prior to being placed online.
Long term drift may occur in this instrument and it is recommended to calibrate every 3 months to
ensure the correct SCU measurement.

To facilitate the initial calibration, a calibration kit is supplied with this instrument. When prepared
according to the included instruction a +5.00 SCU cationic calibration solution is produced. Allow this
solution to stand for 30 minutes prior to use.

6.1 CALIBRATION PROCEDURES

Run the system using the semi or manual dosing mode to prevent interruption of the dosing while
performing the calibration. In preparation for the calibration ensure that previously operated sensors
are rinsed with and then operated in clean water for several minutes. Place the sensor in the Cationic
Standard and allow it to stand for about 15 minutes.

On the SCU display page, click the SCM setting menu. Once inside the SCM Sensor Setting page press
the CALIBRATE button to start the calibration process, which will take 60 seconds.

At the completion of the calibration, ensure the reading is +4.80 to +5.20 SCU. If the reading appears
unstable initiate another calibration sequence. If the +5.00 SCU calibration value is not achievable,
sensor may due for change.

Rinse the sensor with clean water and let it stand at the sampling pot for 15 minutes prior to
returning to auto dosing control mode.

P a g e | 32
7.0 ROUTINE MAINTENANCE
The most important maintenance procedure is to keep the sensor clean. The need for cleaning is
indicated when normal reading cannot be maintained. As a preventive measure, cleaning intervals of
14 days or less is recommended. There are two recommended cleaning methods.

7.1 CHEMICAL CLEANING

Pour the SCM-1 cleaning solution (Part No SCM224952) into a suitable container, large enough to
immerse the lower 1/2 of the probe body. Run the sensor in this solution for approximately 10
minutes. Then run the instrument for about 10 minutes in clean water.

For organic debris, replace the cleaning solution with a 5% chlorine solution.

Please refer to the Material Safety Data Sheet for proper handling of the SCM-1 cleaning solution.

7.2 MANUAL CLEANING

In extreme conditions, the cell will have to be removed and cleaned with an abrasive cleaning pad
such as Scotch-Brite® and a small brush. Always rinse the probe out with clean water prior to
starting.

1. To expose the cell and probe area, remove the bottom cap by turning it CCW (as viewed from
the bottom). Be careful to retain o-rings and seals.

2. Carefully pull the Cell out of the probe end, about 25-50 mm (1-2 inches)

3. Clean the inside of the cell with a stiff toothbrush and an abrasive pad such as (Scotch-Brite®).
The aim is to remove all debris and polish all stainless steel surfaces.

4. Rinse the cell with water.

5. Loosen the shaft-retaining nut with a 10mm wrench and from the bottom of the sensor
unscrew the probe.

6. Completely remove the nut then pull the probe out.

7. Polish the probe with the abrasive pad.

8. Reinstall the probe, adjusting the probe such that it misses the bottom of the cell by 1-2 mm,
then snug the 10mm retaining nut. DO NOT OVER-TIGHTEN.

P a g e | 33
9. Completely dry off any water that is in and around the cell cap, the bottom of the cell and the
O-rings and seals.

10. Reassemble the lower seals and install the cap very firmly by hand.

11. Rotate the motor slowly by hand to ensure the probe does not hit the cell bottom. If there
appears to be any contact adjust the probe up slightly.

12. If the probe does not rotate freely inside the cell, check for obstructions. This condition will
cause premature motor wear.

P a g e | 34
8.0 TROUBLESHOOTING

8.1 DIAGNOSTIC TABLE

Symptom Solutions

Analyzer display not lit. 1. Make sure the unit is plugged in and turned on.
2. Make certain that the power source is providing the
correct voltage.
3. Check if any breaker is in off position inside panel.

Sensor probe not moving 1. Check the interconnect cable connection at the sensor.
2. Check to ensure that the probe is not jammed (Sensor
alarm is triggered if it is powered).

Display response is slow 1. Decrease Damping Time.


2. Distance between dosing and sampling point too great.

Reading different than expected 1. Sensor requires cleaning.


2. Sensor requires calibration.
3. Check Probe and Cell for wear and tear. Replace parts
as required.

SCM alarm indication on 1. Check that sensor is operating correctly.


Analyzer 2. Check wiring connections at both the analyzer and the
sensor.

Reading fluctuating, unstable 1. Incomplete mixing of coagulant with sample water.


2. Sensor and or sample chamber require cleaning.
3. Check coagulate dosing operation.

Reading doesn’t change with a 1. Sensor may require cleaning.


change in dosing 2. Check sample flow to sample chamber.
3. Ensure complete coagulant mix with water.

P a g e | 35
8.2 TECHNICAL AND CUSTOMER ASSISTANCE

Please contact CHEMKIMA if you require any assistance regarding this instrument. Please state your
equipment serial number to facilitate us tracing down the equipment exact specifications.

CHEMKIMIA SDN. BHD.


No. 19, Jalan Kenanga 6, Seksyen BB 11,
Bandar Bukit Beruntung, 48300 Rawang,
Selangor, Malaysia

Phone : +6-03-60283888
Fax : +6-03-60283888
Email : chemkimia@chemkimia.com
Web : www.chemkimia.com

P a g e | 36
9.0 ACCESSORIES AND REPLACEMENTS PARTS LIST

The items shown below are recommended accessories and replacement parts.

Accessory Part Number

SCM Cleaning Solution SCM224952

SCM Sensor Controller SCM949254

SCM Sensor Complete SCM949253

SCM Wetted Sensor SCM441944

SCM Calibration Solution SCM202952

SCM Sensor Motor SCM949351

SCM Cable 4.0 meter SCM210534

To order any accessory or replacement part, please contact CHEMKIMIA. If for any reason technical
assistance is needed regarding this instrument, please do not hesitate to contact CHEMKIMIA.

CHEMKIMIA SDN. BHD.


No. 19, Jalan Kenanga 6, Seksyen BB 11,
Bandar Bukit Beruntung, 48300 Rawang,
Selangor, Malaysia

Phone : +6-03-60283888
Fax : +6-03-60283888
Email : chemkimia@chemkimia.com
Web : www.chemkimia.com

P a g e | 37
APPENDIX A
®
CHEMCHLOR E-20
ELECTRICAL DIAGRAM

P a g e | 38
INSIDE LEFT VIEW INSIDE VIEW FRONT VIEW RIGHT VIEW

EQUIPMENT LIST
NO DESCRIPTION BRAND MODEL

1 10" TOUCHSCREEN IDEC HG3G-AJT22TF-W

2 ALARM ANNUNCIATOR IDEC SLC30N-0103-DDF2B-R(1)A(1)G(1)

3 M12 CONNECTORS STA TECHNOLOGY M12P-5D-S

4 EARTH LEAKAGE CIRCUIT BREAKER ABB FH202 AC-40/0,1

5 SURGE PROTECTION DEVICE ABB OVR T2 1N 40 275-P

6, 7, 8 MINIATURE CIRCUIT BREAKER ABB SH201-C10

9 PROGRAMMABLE LOGIC CONTROLLER OMRON CP1H-XA40DT-D

10 SWITCHING DC POWER SUPPLY IDEC PS5R-SE24

11, 12 INVERTER DRIVE YASKAWA CIMR-JTB A 0002 BAA

P a g e | 39
CONNECTOR FOR E SERIES
(ARRANGEMENT AS SEEN FROM FRONT)
BUILT-IN VSD

P a g e | 40
APPENDIX B
®
CHEMCHLOR EX-10
ELECTRICAL DIAGRAM

P a g e | 41
INSIDE LEFT VIEW INSIDE VIEW FRONT VIEW RIGHT VIEW

EQUIPMENT LIST
NO DESCRIPTION BRAND MODEL

1 7" TOUCHSCREEN OMRON NB7W-TW00B

2 M12 CONNECTORS STA TECHNOLOGY M12P-5D-S

3 EARTH LEAKAGE CIRCUIT BREAKER ABB FH202 AC-40/0,1

4 SURGE PROTECTION DEVICE ABB OVR T2 1N 40 275-P

5, 6, 7 MINIATURE CIRCUIT BREAKER ABB SH201-C10

8 SWITCHING DC POWER SUPPLY OMRON S8VK-06024

9 PROGRAMMABLE LOGIC CONTROLLER OMRON CP1H-XA40DT-D

P a g e | 42
CONNECTOR FOR EX SERIES
(ARRANGEMENT AS SEEN FROM FRONT)

P a g e | 43

You might also like