Model Q-DRT Series

Rotating Trays for Portable


Digital Detectors

Installation and Operation


Manual

Manual Part No. DC30-092 Revision E


0

This manual is copyrighted and all rights are reserved. No portion of this document
may be copied, photocopied, reproduced, translated, or reduced to any electronic
medium or machine readable form without prior consent in writing from Quantum
Medical Imaging, LLC. (QMI)

Copyright© 2013 QMI

Quantum Medical Imaging, LLC Made in U.S.A.


2002-B Orville Drive North
Ronkonkoma, New York 11779 USA
Phone: (631) 567-5800
Fax: (631) 567-5074
E-mail: qmiinfo@carestream.com
www.quantummedical.net
Revision History

REVISION DATE TYPE OF MODIFICATION

A 2009-06-17 Initial Release.

B 2009-09-22 Incorporated Q-DRT-CL-R / Q-DRT-CR-R

C 2010-04-29 Incorporated ECO 2123, 2173, 2285, 2286

D 2010-12-27 Incorporated ECO 2432

E 2013-08-05 Incorporated ECO 2476, 2482, 2486, 2647,


2721, 2845, obsoleted non rotating models

Page Page Page


Number Rev Number Rev Number Rev
i - xvi E
1-1 thru 1-6 E
2-1 thru 2-4 E
3-1 thru 3-4 E
4-1 thru 4-16 E
5-1 thru 5-14 E

Model Q-DRT Series Digital Detector Trays Revision E


Quantum Medical Imaging, LLC i
Revision History

THIS PAGE INTENTIONALLY LEFT BLANK

Revision E Model Q-DRT Series Digital Detector Trays


ii Quantum Medical Imaging, LLC
Table of Contents

SAFETY NOTICES
GENERAL SAFETY INFORMATION ............................................................ v
WARNINGS, CAUTIONS, NOTES............................................................... v
REGULATORY COMPLIANCE ...................................................................vii
COMPATIBILITY ..............................................................................vii
INTENDED OPERATOR .....................................................................vii
TRAINING ...................................................................................... viii
ACCOMPANYING DOCUMENTATION ................................................. viii
ENVIRONMENTAL PROTECTION ............................................................ viii
ABBREVIATION DEFINITION .................................................................. ix

WARRANTY INFORMATION
WARRANTY STATEMENT ........................................................................ xi
WARRANTY STATEMENT (Continued) ..................................................... xiii
WARRANTY EXCLUSIONS ...................................................................... xiii
BUYER’S REMEDIES ..............................................................................xiv
WARRANTY RETURN PROCEDURE .........................................................xiv
EQUIPMENT IN TRANSIT .......................................................................xiv
VOIDING WARRANTY..............................................................................xv

CHAPTER 1 - INTRODUCTION
OVERVIEW .......................................................................................... 1-3
KEY FEATURES .............................................................................. 1-3
Intended Use ................................................................................. 1-3
MAIN COMPONENTS ............................................................................ 1-4
Q-DRT DETECTOR TRAY CHARACTERISTICS ......................................... 1-5

CHAPTER 2 - SPECIFICATIONS
PRODUCT SPECIFICATIONS ................................................................. 2-3
General Specifications .................................................................... 2-3
Design .................................................................................... 2-3
Physical Characteristics .................................................................. 2-3
ENVIRONMENTAL SPECIFICATIONS ................................................ 2-4
System Operating Environment ................................................. 2-4
Non-Operating Environment ..................................................... 2-4
COMPATIBILITY STATEMENT ............................................................... 2-4

Model Q-DRT Series Digital Detector Trays Revision E


Quantum Medical Imaging, LLC iii
Table of Contents

CHAPTER 3 - ASSEMBLY & INSTALLATION


OVERVIEW .......................................................................................... 3-3

CHAPTER 4 - OPERATION
OPERATION OF Q-DRT ROTATING TRAYS ............................................ 4-3
OPERATION OF Q-DRT ROTATING TRAYS MOUNTED TO
VERTICAL WALL STANDS ............................................................... 4-3
Q-DRT ROTATING PORTABLE DETECTOR
TRAY OPERATION ................................................................... 4-3
OPERATION OF Q-DRT ROTATING TRAYS MOUNTED TO
RADIOGRAPHIC TABLES ................................................................. 4-6
IMAGING IN PORTRAIT ORIENTATION ..................................... 4-6
CHANGING DETECTOR ORIENTATION TO LANDSCAPE ............... 4-8
OPTIONAL REMOVABLE GRID OPERATION ...................................... 4-9

CHAPTER 5 - MAINTENANCE
OVERVIEW .......................................................................................... 5-3
USER MAINTENANCE ..................................................................... 5-3
PLANNED MAINTENANCE ............................................................... 5-3
REPLACEMENT PARTS AND ORDERING INFORMATION .................... 5-3
ORDERING INFORMATION ............................................................. 5-4
Removal and Replacement Procedure for Phenolic Panel
(Wall Stands Only) ................................................................... 5-7
Removal and Replacement of Detent Bumpers
(Rotary Models Only) ............................................................... 5-8
Removal and Replacement of the Sliding Tray Latch
(Rotary Models Only) ............................................................... 5-9
Removal and Replacement of the Chin Rest
(Wall Stands Only .................................................................. 5-10
Removal and Replacement of Push/Pull Catch .......................... 5-12
Removal and Replacement of Receptor Lock Micro Switch
(Table Models Only) ............................................................... 5-13

Revision E Model Q-DRT Series Digital Detector Trays


iv Quantum Medical Imaging, LLC
Safety Notices

GENERAL SAFETY INFORMATION


Quantum products are designed to meet stringent safety standards. All medical elec-
trical equipment requires proper installation, operation, and maintenance (particularly
with regard to safety).
It is vital that the user read, understand, note, and where applicable, strictly observe
all Warnings, Cautions, Notes and Safety markings within this document and on the
equipment, and that the user strictly follow all safety directions in this manual to help
ensure the safety of users and patients.
Every reasonable precaution has been taken during manufacture to safeguard the
health and safety of persons who will operate this equipment. The following precau-
tions must be observed at all times.
WARNINGS, CAUTIONS, NOTES
The following samples show how warnings, cautions, and notes appear in this docu-
ment. The text explains their intended use.

WARNING Indicates injury or death is possible if the instruc-


tions are not obeyed. Instructs users to refer to
documentation if displayed without warning text.

CAUTION Indicates that damage to equipment is possible if


the instructions are not obeyed.

NOTE Notes provide advice and highlight unusual


points. A note is not intended as an instruction.

The purpose of safety icons, such as those shown below, is to indicate at a glance the
type of caution, warning or danger.

WARNING Ionizing radiation: indicates the possibility of


increased levels of radiation.

WARNING Dangerous voltage: indicates the presence of high


voltage.

WARNING Warning, hot surface.

Model Q-DRT Series Digital Detector Trays Revision E


Quantum Medical Imaging, LLC v
Safety Notices

WARNING
Quantum Medical Imaging, LLC disclaims all responsibility from any injury
resulting from improper application of this equipment.
This equipment is sold to be used exclusively under the prescribed direction
of a person who is licensed by law to operate equipment of this nature.
This equipment must be used in accordance with all safety procedures
described in this manual and must not be used for purposes other than
those described herein. In the United States, Federal law restricts this
device to sale, distribution, and use by or on order of a licensed physician.
Quantum Medical Imaging, LLC cannot assume responsibility for any mal-
functioning of this equipment resulting from improper operation, mainte-
nance, or repair, or from damage or modification of its components.
Failure to observe these warnings may cause serious injuries.

WARNING
X-rays are hazardous to both patient and operator unless established safe
exposure factors and operating instructions are observed.
Only qualified and authorized personnel shall operate this system. In this
context, qualified means those legally permitted to operate this equipment
in the jurisdiction in which the equipment is being used, and authorized
means those authorized by the authority controlling the use of the equip-
ment. Full use must be made of all radiation protection features, devices,
systems, procedures and accessories.
It is important that everyone having anything to do with x-radiation be
properly trained and fully acquainted with the recommendations of the
National Council on Radiation Protection and Measurements as published in
NCRP Reports available from NCRP Publications, 7910 Woodmont Avenue,
Suite 800, Bethesda, Maryland 20814-3095 (www.ncrp.com), and of the
International Commission on Radiological Protection (www.icrp.org), and
take adequate steps to protect against injury.

WARNING
X-ray equipment may cause injury if used improperly. The instructions con-
tained in this manual must be read and followed when operating this unit.
Personal radiation monitoring and protective devices are available. You are
urged to use them to protect against unnecessary x-ray exposure.

Revision E Model Q-DRT Series Digital Detector Trays


vi Quantum Medical Imaging, LLC
Safety Notices

REGULATORY COMPLIANCE
This certified Quantum Medical Imaging, LLC medical device has been designed, man-
ufactured, and calibrated to comply with governing Federal Regulations 21 CFR Sub-
chapter J and the performance standards attendant thereto. Upon installation, all
certified products require the filing of Form FD-2579 "Report of Assembly of a Diag-
nostic X-ray System" by the assembler (i.e., the installer) with the appropriate agen-
cies; the "Installation Quality Assurance Checklist" must also be completed and
properly distributed upon installation. A copy of each form (pink copy) is provided to
the user. The installation report is completed by the installer and returned to Quantum
Medical Imaging, LLC.
Those responsible for the planning of x-ray equipment installations must be thor-
oughly familiar and comply completely with NCRP Report No. 49, "Structural Shielding
Design and Evaluation for Medical Use of X-Rays and Gamma Rays of Energies up to
10 MeV", as revised or replaced in the future. Those authorized to operate, test, par-
ticipate in or supervise the operation of the equipment must be thoroughly familiar
and comply completely with the currently established safe exposure factors and pro-
cedures described in publications such as Subchapter J of Title 21 of the Code of Fed-
eral Regulations, "Diagnostic X-Ray Systems and Their Major Components," and NCRP
Report No. 102, “Medical X-Ray, Electron Beam and Gamma Ray Protection for Ener-
gies Up to 50 MeV—Equipment Design and Use” as revised or replaced in the future.
This equipment must only be used in rooms that comply with all applicable laws or
regulations that have the force of law, concerning electrical safety for this type of
equipment.
Scheduled maintenance is essential to the assurance of continued integrity of this
equipment with respect to regulatory compliance. The continuance of certified perfor-
mance to the regulatory standard is incumbent upon the user's diligent conformance
to recommended maintenance instructions.
The following symbols may be used for marking on this equipment and/or equipment
documentation:

Attention, consult accompanying documents

COMPATIBILITY
The equipment described in this manual must only be used in combination with
other equipment or components if these are expressly recognized by Quantum
Medical Imaging, LLC as compatible.

INTENDED OPERATOR
This equipment is intended to be installed, used and operated only in accordance
with the safety procedures given within this manual for the purpose for which it

Model Q-DRT Series Digital Detector Trays Revision E


Quantum Medical Imaging, LLC vii
Safety Notices

was designed. Before attempting to work with this equipment, read, understand,
note and strictly observe all warnings, cautions and safety markings on the equip-
ment.
Users include those persons who actually handle the equipment and those who
have authority over the equipment.

TRAINING
Users of this equipment shall have received adequate training on its safe and
effective use before attempting to work with the equipment. Training require-
ments may vary from country to country. The User shall make sure that training is
received in accordance with local laws or regulations that have the force of law.

ACCOMPANYING DOCUMENTATION
The documentation consists of an Installation & Operation manual (this docu-
ment) and related documentation:
• Service Manual P/N DC30-007: Contains technical and service documentation
for radiographic tables, into which this product may be installed
• Service Manual P/N DC30-040: Contains technical and service documentation
for casette holders, into which this product may be installed
The documentation shall be kept with the system for easy reference.

ENVIRONMENTAL PROTECTION
This equipment may contain certain materials and chemical compounds incidental to
the manufacture of electrical and electronic equipment, and improper "end-of-life"
disposal of such equipment can result in environmental contamination. Therefore, this
equipment should not be disposed of as ordinary household waste, but should instead
be delivered to a designated electrical and electronic waste disposal or recycling cen-
ter. For further information on disposing of electrical and electronic waste, contact the
cognizant authority within the jurisdiction.

Revision E Model Q-DRT Series Digital Detector Trays


viii Quantum Medical Imaging, LLC
Safety Notices

ABBREVIATION DEFINITION
The following abbreviations and acronyms may be found in this document. Their def-
inition is explained below.
DR Digital Receptor

hPa Hectopascal

kg Kilograms

mm Millimeters

in. Inch(es)

lb Pound(s)

VAC Volts, Alternating-Current

Model Q-DRT Series Digital Detector Trays Revision E


Quantum Medical Imaging, LLC ix
Safety Notices

THIS PAGE INTENTIONALLY LEFT BLANK

Revision E Model Q-DRT Series Digital Detector Trays


x Quantum Medical Imaging, LLC
Warranty Information

WARRANTY STATEMENT
Quantum Medical Imaging, LLC (herein after known as “QMI”) warrants to the buyer
that any new product manufactured by QMI will be free from defects in material and
workmanship, and will substantially conform to the applicable specifications in effect
on the date of shipment when subjected to normal, proper and its intended use by
properly trained personnel. QMI shall be the sole judge in determining whether said
equipment or component is defective by reason of manufacture.
All QMI products shall be so warranted for a period of 12 months from the date of
original installation, such date to be evidenced by means of a completed Warranty
Card returned to QMI within 30 days of installation. In no case shall the warranty
extend beyond 15 months from the date of shipment. If the attached warranty card is
not so returned to QMI, then the warranty period will be deemed to have commenced
on the date of shipment (the invoice date) and extend for a period of twelve months.
The buyer should submit only one such card per system or major component pur-
chased.
WARRANTY CARD

Model Q-DRT Series Digital Detector Trays Revision E


Quantum Medical Imaging, LLC xi
Warranty Information

Clip or Copy Warranty Card on reverse side and submit to QMI:

Quantum Medical Imaging, LLC


2002-B Orville Drive North
Ronkonkoma, N.Y. 11779 USA
631 567-5800 voice
631 567-5074 fax
qmiservice@carestream.com

Revision E Model Q-DRT Series Digital Detector Trays


xii Quantum Medical Imaging, LLC
Warranty Information

WARRANTY STATEMENT (Continued)


Replacement components furnished by QMI to the Buyer/Dealer during the warranty
period shall be warranted for the remainder of the original product warranty or 90
days, whichever is longer. This warranty extends only to the original purchaser and is
not transferable unless expressly authorized in writing by Quantum Medical Imaging,
LLC.
Products manufactured by parties other than QMI, whereby QMI acts solely as distrib-
utor or reseller, are warranted exclusively by their manufacturers according to each of
their independent warranty terms and conditions.
Warranty consideration can only be given for defective QMI products properly
returned to the factory in accordance with the QMI Returned Materials Procedure
(refer to Dealer Price Book or contact QMI customer service).
WARRANTY EXCLUSIONS
The foregoing warranties are exclusive and in lieu of all other warranties, whether
written, oral, express, implied or statutory. NO IMPLIED WARRANTY OF MERCHANT-
ABILITY OR FITNESS FOR A PARTICULAR PURPOSE SHALL APPLY. Quantum Medical
Imaging, LLC Warranty is exclusive of:
1) Failure of the Buyer/Dealer to prepare the site and operating environment in
accordance with applicable instructions and recommendations of QMI.
2) Failure of Buyer/Dealer to provide the proper incoming power required to support
the equipment in accordance with the requirements of QMI.
3) Modification of QMI products performed by a party other than QMI.
4) Combining products deemed by QMI to be incompatible.
5) Improper or extraordinary use of a product, improper maintenance of the prod-
uct, or failure to comply with any applicable instructions and recommendations of
Quantum Medical Imaging, LLC.
6) Misuse, abuse, tampering, or negligent storage or handling of a product by the
Buyer, its employees, agents, or contractors.
7) Fuses, glassware, high voltage cables and other items deemed by QMI to be
expendable.
Acts of God, fires, floods, power failure or electrical power surges. Strikes, sabotage,
labor disturbances, war, riots, acts of civil or military authority, or other causes beyond
the reasonable control of QMI.
Installation, routine troubleshooting and repair are also excluded from warranty.
Technical service and maintenance is the responsibility of the Dealership selling the
equipment.
The Manufacturer is hereby relieved of all responsibility for damage during shipment
of the product following the freight carrier’s pick-up for transportation to the delivery
point.

Model Q-DRT Series Digital Detector Trays Revision E


Quantum Medical Imaging, LLC xiii
Warranty Information

BUYER’S REMEDIES
If QMI determines that a product fails to meet any specification during the applicable
warranty period, QMI shall correct any such failure as follows:
A) By repairing, adjusting, or replacing any defective or non-conforming component
or product.
B) By making available any necessary repair or replacement parts or assemblies for
exchange.
Quantum Medical Imaging, LLC shall have the option to furnish either new or rebuilt
replacement parts or assemblies for exchange. All returned parts shall become the
property of Quantum Medical Imaging, LLC upon exchange.
The preceding Paragraphs set forth the Buyer’s sole remedies and QMI’s sole liability
for claims based upon failure of the product to meet any warranty, whether the claim
is on contract, warranty, tort (including negligence and strict liability) or otherwise,
and however instituted.
Upon the expiration of the applicable warranty period, all such liability shall terminate.
In no event shall QMI be liable for special or consequential damages arising out of the
use of or inability to use its equipment, whatsoever.
The warranties and remedies available to the buyer are conditioned upon claims
under this warranty being made in accordance with the aforementioned warranty
statement.
WARRANTY RETURN PROCEDURE
A fully completed Field Returned Material Evaluation Form must be returned with any
defective product or any returned item. All returns must include the Serial Number of
the Equipment and the Specific Part Number written on the Field Returned Material
Evaluation Form. All freight charges resulting from Warranty Returns are the respon-
sibility of the Buyer/Dealer.
EQUIPMENT IN TRANSIT
QMI assumes no responsibility for equipment damaged in transit to or from QMI. To
protect the Buyer/Dealer, the receiver of any equipment should examine all cartons
and crates carefully at the time of delivery. If damage is apparent, make a notation
on the delivery receipt, request an inspection by the freight carrier, and if applicable,
file an appropriate carrier claim. Should concealed damage be detected, immediately
notify the carrier and request an inspection. The purchaser (Buyer/Dealer/Customer)
is fully responsible for the filing of freight damage claims to the freight carrier.
QMI assumes no responsibility for any loss or damage to products once they have
been shipped from our factory. As such, the Buyer/Dealer and Customer remain fully
responsible for payment to QMI for all invoices, according to our standard payment
terms, regardless of freight damage or processing of an insurance claim, by the dealer
or customer.

Revision E Model Q-DRT Series Digital Detector Trays


xiv Quantum Medical Imaging, LLC
Warranty Information

VOIDING WARRANTY
Tampering with, or any attempt at installation, maintenance, repair, service, reloca-
tion, or alteration of or to a QMI product, when performed by any person or entity
other than Quantum Medical Imaging, LLC or its Certified Dealer without the written
approval of an Authorized Person at Quantum Medical Imaging, LLC, shall immedi-
ately Void and Cancel all warranties with respect to the affected product.

Model Q-DRT Series Digital Detector Trays Revision E


Quantum Medical Imaging, LLC xv
Warranty Information

THIS PAGE INTENTIONALLY LEFT BLANK

Revision E Model Q-DRT Series Digital Detector Trays


xvi Quantum Medical Imaging, LLC
Chapter
1

INTRODUCTION

1-1
1-2
Chapter 1 Introduction

OVERVIEW

NOTE: The user should read this manual in its


entirety prior to using this equipment. It should
be kept in a location near the equipment and be
readily accessible to those who operate it.

This document is intended to assist users in the safe and effective operation of the
equipment described herein. Pay special attention to all the information described in
the Safety section (refer to SAFETY NOTICES).
This manual is written for trained users of the Q-DRT Series rotating X-ray image
detector trays for portable digital detector panels, hereinafter referred to as the Q-
DRT Trays, and for authorized field service personnel. Quantum Medical Imaging, LLC
assumes no liability for use of this document if any unauthorized changes to the con-
tent or format have been made.
KEY FEATURES
It is imperative that all safety procedures described in this manual be strictly
adhered to, in order to ensure the safety of both patient and user. The significant
physical and performance characteristics of the Q-DRT Trays are as follows:
• Q-DRT models to accommodate popular portable 14" x 17" Digital
Receptor (DR) panels, including, Agfa/Varian 4336, Carestream DRX-1
portable wireless, Vatech (Samsung) Xmaru 1417, and Thales 3543.
• All Q-DRT Trays feature a convenient drawer slide for easy installation
and removal of portable detector panel.
• Q-DRT models are compatible with Quantum’s Model QT-740 & QT-750
Radiographic Tables, and QW-420 Wall Stands where 14" x 17" portable
image detectors are used.
• Various options are available for left or right side loading trays on
Quantum QW-420 vertical wall stands, and fixed or elevating, float-top
QT-740 and QT-750 radiographic tables, including rotation of the
cassette holder to accommodate either portrait or landscape exposures.
• Removable grid and ion chamber options are available.
This product is intended to be used and operated only in accordance with the
safety procedures given within this manual for the purpose for which it was
designed. The intended use is given below. Nothing stated in this manual reduces
user's professional responsibilities for sound judgment and best practice.
INTENDED USE
Model Q-DRT Trays are intended for integration with new or existing Quantum
radiographic tables and wallstands, and are to be used during the performance of
radiographic examinations.
Use of the equipment for purposes other than those intended and expressly

Model Q-DRT Series Digital Detector Trays Revision E


Quantum Medical Imaging, LLC 1-3
Chapter 1 Introduction

stated by the manufacturer, as well as incorrect use or operation, may relieve the
manufacturer or his agent from all or some of the responsibility for resultant non-
compliance, damage or injury.
MAIN COMPONENTS
See Figure 1-1. The Q-DRT Trays contain:
1. Detector tray cover (Wall Stand models only)
2. Phenolic Panel (Wall Stand models only)
3. Sliding drawer (wall stand with left load shown)
4. Drawer slide handle
5. Portable detector locking retainer and spring block
6. Heavy duty drawer slide system
7. Detector tray turntable (rotating models only)
8. Removable grid and ion chamber (optional).

Revision E Model Q-DRT Series Digital Detector Trays


1-4 Quantum Medical Imaging, LLC
Chapter 1 Introduction

3 1

2
5

6 7
Note:
1. A cable guide for tethered receptors is provided on certain models
(location varies with tray model).

Figure 1-1. Q-DRT Tray - Rotating Unit Shown

Q-DRT DETECTOR TRAY CHARACTERISTICS


The Q-DRT Image Detector Tray Series is designed to provide a fixture for portable image
detectors available in the Radiographic Equipment market. The following table describes
which style is available for each portable detector manufacturer shown, including Quan-
tum’s Type Number for each:

Model Q-DRT Series Digital Detector Trays Revision E


Quantum Medical Imaging, LLC 1-5
Chapter 1 Introduction

Table 1-1: Q-DRT Trays for Portable Image Detectors

Detector Detector Tray Rotating or Loading Quantum Tray


Manufacturer Number Mounting Non-rotating Characteristics Type Number

Varian 4336 Wall Rotating Right Q-DRT-VR-R

Varian 4336 Wall Rotating Left Q-DRT-VL-R

Varian 4336 Table Rotating Table Q-DRT-VT-R

Carestream DRX-1 Wall Rotating Right Q-DRT-DRXR-R

Carestream DRX-1 Wall Rotating Left Q-DRT-DRXL-R

Carestream DRX-1 Table Rotating Table Q-DRT-DRXT-R

Carestream DRXT Table Rotating Table Q-DRT-WOD-R

Carestream DRXT Right Rotating Right Q-DRT-WOD-L

Carestream DRXT Wall Rotating Left Q-DRT-WOD-T

Vatech Xmaru 1417 Wall Rotating Right Q-DRT-SAMR-R


(Samsung)

Vatech Xmaru 1417 Wall Rotating Left Q-DRT-SAML-R


(Samsung)

Vatech Xmaru 1417 Table Rotating Table Q-DRT-SAMT-R


(Samsung)

FXRD Rotating Right Q-DRT-VWR-R


Vieworks Portable Wall

FXRD Rotating Left Q-DRT-VWL-R


Vieworks Portable Wall

FXRD Rotating Table Q-DRT-VWT-R


Vieworks Portable Table

Revision E Model Q-DRT Series Digital Detector Trays


1-6 Quantum Medical Imaging, LLC
Chapter
2

SPECIFICATIONS

2-1
2-2
Chapter 2 Specifications

PRODUCT SPECIFICATIONS
The following are specifications for Q-DRT Trays, which are compatible with
Quantum QW-420 Wall Stands and QT-740 and QT-750 Radiographic Tables.
GENERAL SPECIFICATIONS
Design
ENCLOSURE: Four detector tray models (see Table 1-1)
Material and Finish: Aluminum base with powder coat steel cover.

DRAWER SLIDES: heavy duty, 3-member slides with caged in-line


ball bearings.
Dimensions: 3.0 in. (W) x 0.75 in. (H), (76 mm [W] x 19 mm
[H]) with 18.5 in. (457 mm) slide extension (two
mounted normal to base plate).

PHYSICAL CHARACTERISTICS
The following table describes the physical characteristics of four Q-DRT Tray
designs:

Weight
Detector Style Overall Dimensions
(without detector)

Rotating
Table: 584 mm (W) x 559 mm (H) x 73 mm (DP)
Varian 23.0 in. (W) x 22.0 (H) x 2.87 (DP) 18 kg (40 lb)
4336
Wall Stand: 561 mm (W) x 612 mm (H) x 79 mm (DP)
22.1 (W) x 24.1 (H) x 3.1 (DP) 21 kg (48 lb)

Rotating
Table: 584 mm (W) x 559 mm (H) x 73 mm (DP)
Carestream 23.0 in. (W) x 22.0 (H) x 2.87 (DP) 18 kg (40 lb)
DRX-1
Wall Stand: 561 mm (W) x 612 mm (H) x 79 mm (DP)
22.1 (W) x 24.1 (H) x 3.1 (DP) 21 kg (48 lb)

Rotating
Table: 584 mm (W) x 559 mm (H) x 73 mm (DP)
Vatech 23.0 in. (W) x 22.0 (H) x 2.87 (DP) 18 kg (40 lb)
(Samsung)
Wall Stand: 561 mm (W) x 612 mm (H) x 79 mm (DP)
22.1 (W) x 24.1 (H) x 3.1 (DP) 21 kg (48 lb)

Model Q-DRT Series Digital Detector Trays Revision E


Quantum Medical Imaging, LLC 2-3
Chapter 2 Specifications

Weight
Detector Style Overall Dimensions
(without detector)

Rotating
Table: 584 mm (W) x 559 mm (H) x 73 mm (DP)
Vieworks 23.0 in. (W) x 22.0 (H) x 2.87 (DP) 18 kg (40 lb)

Wall Stand: 561 mm (W) x 612 mm (H) x 79 mm (DP)


22.1 (W) x 24.1 (H) x 3.1 (DP) 21 kg (48 lb)

ENVIRONMENTAL SPECIFICATIONS
The following are environmental specifications for Model Q-DRT Image Detec-
tor Tray:
System Operating Environment
Ambient
Temperature: +10 ° C to +40 ° C (+50 ° F to 104 ° F)
Relative
Humidity: 30 to 75%, Non-condensing
Atmospheric
Pressure: 700 hPa to 1060 hPa
Non-Operating Environment
Ambient
Temperature: -18 ° C to +70 ° C (-0.4 ° F to 158 ° F)
Relative Humidity: 20 to 95%, non-condensing
Atmospheric
Pressure: 500 hPa to 1060 hPa
COMPATIBILITY STATEMENT
The Quantum Medical Imaging, LLC Digital Image Detector Trays, marketed under
the Q-DRT designation, are compatible with all Quantum Medical Imaging, LLC Model
QW-420 Series Wall Stands, and Models QT-740 and QT-750 Fixed and Elevating
Radiography Tables.

Revision E Model Q-DRT Series Digital Detector Trays


2-4 Quantum Medical Imaging, LLC
Chapter
3

ASSEMBLY &
INSTALLATION

3-1
3-2
Chapter 3 Assembly & Installation

OVERVIEW

NOTE: Examine all cartons and crates carefully at


time of delivery. If damage is apparent, have
delivery driver write a "Demand Shipment Note"
on copies of the freight bill, sign, and file appro-
priate carrier claim. Should you discover con-
cealed damage, immediately notify the
transporting agent and ask for an "Inspection of
Damage." Carrier will not accept concealed dam-
age claim if filed after 15 days from date of
receipt of merchandise.

The Model Q-DRT Series Trays for Model QW-420 Wall Stands, and QT-740 and QT-
750 Radiographic Tables are completely installed and tested at the factory prior to
shipment. Therefore, all relevant unpacking, and assembly and installation instruc-
tions for Q-DRT Trays are found in the respective Table or Wall Stand Service Manuals
DC30-007 or DC30-002. Refer to Chapter 4, OPERATION, of this manual for details as
to proper use of the Q-DRT Trays installed on the specific table or wall stand ordered.

Model Q-DRT Series Digital Detector Trays Revision E


Quantum Medical Imaging, LLC 3-3
Chapter 3 Assembly & Installation

THIS PAGE INTENTIONALLY LEFT BLANK

Revision E Model Q-DRT Series Digital Detector Trays


3-4 Quantum Medical Imaging, LLC
Chapter
4

OPERATION

4-1
4-2
Chapter 4 Operation

OPERATION OF Q-DRT ROTATING TRAYS


This chapter describes the operation of Q-DRT Rotating Trays, including models that
support detectors designed by several different manufacturers. All Q-DRT models
(hereinafter referred to as "detector tray") utilize a manually operated drawer slide to
provide access to the tray for detector installation. They have a mechanical compo-
nent that provides the capability to position the detector in either portrait or land-
scapes orientation, without removing it from the tray. Q-DRT Trays for wall stand use
and radiographic table use are described herein.

WARNING! All movable assemblies and parts of


this equipment must be operated with reasonable
care. Manufacturer’s equipment recommenda-
tions described in this manual must be observed.

OPERATION OF Q-DRT ROTATING TRAYS MOUNTED TO


VERTICAL WALL STANDS

Figure 4-1. Q-DRT Rotating Tray - Wall Stand Type


Q-DRT ROTATING PORTABLE DETECTOR TRAY OPERATION
The rotating detector tray option provides the capability of changing the
Detector’s orientation (portrait to landscape), while the detector is
securely mounted in the tray. To load a Detector into the tray, proceed as
follows:

Model Q-DRT Series Digital Detector Trays Revision E


Quantum Medical Imaging, LLC 4-3
Chapter 4 Operation

TO RELEASE TRAY
LOCK, PUSH IN ON
DETECTOR TRAY
HANDLE, UNTIL A
"CLICK" IS HEARD.
THEN PULL TRAY OUT
UNTIL ANOTHER
"CLICK" IS HEARD
(AT FULL EXTENSION)

Figure 4-2. Q-DRT Rotating Tray Motion Characteristics


1. Pull back on the handle until another "click" is heard. This indi-
cates that the tray is locked in its fully extended (OPEN) position
(see Figure 4-2).
2. Grasp the Detector by its handle and place it in the tray, with the
handle facing out from the side of the wall stand. Ensure that the
corners of the Detector are aligned with two (2) fixed, white plas-
tic retainers mounted in the tray closest to the tray cover.
3. Carefully apply pressure by pushing the Detector’s handle toward
the fixed, white plastic retainers to seat the Detector.

Revision E Model Q-DRT Series Digital Detector Trays


4-4 Quantum Medical Imaging, LLC
Chapter 4 Operation

SPRING-LOADED
RETAINER

FIXED
RETAINERS

CLAMPING
RETAINER

Figure 4-3. Q-DRT Rotating Tray with Detector Installed - Portrait Orientation
4. While maintaining pressure against the fixed retainers, move the
handle of the Detector toward the floor of the tray, to seat it in
front of the spring-loaded retainer and the clamping retainer clos-
est to the Detector handle (see Figure 4-3).
5. Set the clamping retainer to lock the Detector in place. The
Detector is prepared for a portrait exposure.
6. If necessary (for landscape exposure), and while the tray is fully
extended, rotate the detector handle downward for right-loading
configurations, or upward for left-loading configurations, until the
tray enters the detent. The foregoing procedures should result in
secure mounting of the Detector, allowing for the tray to slide
back into its tray, set for the desired orientation.

Model Q-DRT Series Digital Detector Trays Revision E


Quantum Medical Imaging, LLC 4-5
Chapter 4 Operation

Figure 4-4. Left-Loading Q-DRT Tray


Rotate Clockwise for Landscape Orientation
7. Pull out on the tray handle until a click is heard. This releases the
slide locks, and allows the tray to slide back into the tray.

NOTE: The Detector should be inserted or


extracted from the rotating tray only when the
tray is in portrait orientation.

OPERATION OF Q-DRT ROTATING TRAYS MOUNTED TO


RADIOGRAPHIC TABLES
The Q-DRT rotating tray is designed specifically to support near effortless posi-
tioning of the Detector in portrait or landscape orientation. The rotating tray elim-
inates the need for the X-ray technician to remove the detector to change its
orientation. To insert the Detector into the Q-DRT rotating tray, proceed as fol-
lows:
IMAGING IN PORTRAIT ORIENTATION
1. Pull out the Detector tray handle, extracting the drawer from the
enclosure, until a "click" is heard. This locks the tray in the fully
extended (OPEN) position (see Figure 4-15).
2. Grasp the Detector by the handle and place it on the tray, so the
corners furthest away from the Detector handle aligns with the
two white plastic, fixed retainers mounted to the floor of the tray
closest to the table.
3. Apply slight pressure to the detector handle, as to seat the Detec-
tor in the fixed retainers.

Revision E Model Q-DRT Series Digital Detector Trays


4-6 Quantum Medical Imaging, LLC
Chapter 4 Operation

Figure 4-5. Q-DRT Rotating Tray Mounted to Radiographic Table


4. Lay the opposite corners (closest to the Detector handle) in front
of the spring-loaded retainer (left), and the clamping retainer
(right). Twist the clamping retainer to lock the Detector in the
tray.

DETECTOR
CABLE
GUIDE IN
RETAINER
(Not
applicable if
wireless
detector)

POSITIONING
RETAINERS

Figure 4-6. Detector Installed in Tray - Portrait Orientation


5. Insert the Detector’s cable into slot as shown.
6. The Detector is prepared for an exposure in portrait orientation,
as shown in Figure 4-7
7. With the Detector securely in place, pull the tray handle out until
a "click" is heard. This releases the tray slide lock, allowing the
tray to close.

Model Q-DRT Series Digital Detector Trays Revision E


Quantum Medical Imaging, LLC 4-7
Chapter 4 Operation

Figure 4-7. Q-DRT Rotating Tray with Detector Installed - Portrait Orientation

CHANGING DETECTOR ORIENTATION TO LANDSCAPE


For landscape orientation exposures, it is assumed that the steps
described above (Imaging in Portrait Orientation) have been completed,
and the tray is resting in its fully extended (open) position, with the
Detector mounted, clamping retainer set, and cable properly secured.
The following instructions describe how to change the Detector’s orienta-
tion (to "landscape") without removing the detector from the tray. Pro-
ceed as follows:
1. Grasp the Detector by the handle and apply sufficient pressure
(clockwise direction) to cause the tray to rotate out of its portrait
orientation detent (see Figure 4-8), and reposition it approxi-
mately 90° into the detent for landscape orientation (Detector to
left and perpendicular to the table). Once the tray is properly
seated in its landscape detent position, the Detector is prepared
for a landscape exposure.

Revision E Model Q-DRT Series Digital Detector Trays


4-8 Quantum Medical Imaging, LLC
Chapter 4 Operation

Figure 4-8. Setting Q-DRT Rotating Tray to Landscape Orientation


2. Pull the Detector tray handle out until a "click" is heard. This
releases the tray slide lock, allowing the tray to close.
3. To return the Detector back to the portrait orientation, pull the
tray (by its handle) back out to its full extended position.
4. Grasp the Detector handle and rotate it counter-clockwise from
the landscape detent to the portrait detent position (approxi-
mately -90°).
5. Release the tray slide lock by pulling out on the tray handle, then
slide the tray closed.

NOTE: The Detector should be inserted or


extracted from the rotating tray only when the
tray is in portrait orientation.

OPTIONAL REMOVABLE GRID OPERATION


Q-DRT Trays can be configured with a removable grid panel which is designed for
installation on rails inside the cover. The grid panel is mounted in a frame and fit-
ted with a locking mechanism.
Installation of the grid panel is accomplished as follows:
1. Align and insert the top and bottom flanges of the grid frame in the tracks,
using care to place the flange between the grid guide clips and the rear
surface of the phenolic panel (see Figure 4-9).
2. Push the grid panel assembly into the cover. Note that the lock actuator rod
will move the handle as it is seated (locking the grid).

Model Q-DRT Series Digital Detector Trays Revision E


Quantum Medical Imaging, LLC 4-9
Chapter 4 Operation

3. To remove the grid panel, move the handle in the direction indicated by the
arrow, until it stops, then, withdraw the grid assembly from the cover.
4. While the grid is not in use, store it carefully to protect the grid surface, and
the symmetry of the frame.

Phenolic Panel
(rear surface)

Grid Guide Clip

Slide handle to disengage


lock and remove grid

T:\Storage\Manuals\Q-DRT Series Cassette Trays\Installation - Operation Manual\Photos\Grid Panel Operation.cdr

Figure 4-9. Removable Grid Panel Operation

Revision E Model Q-DRT Series Digital Detector Trays


4-10 Quantum Medical Imaging, LLC
Chapter
5

MAINTENANCE

5-1
5-2
Chapter 5 Maintenance

OVERVIEW
The Q-DRT Digital Detector Trays will function reliably with virtually no mechanical
maintenance, however a list of wear parts is provided in this section.

WARNING! Only properly trained service


personnel should service or maintain this
equipment. Failure to follow manufacturer's
or service personnel's recommendations
may result in serious injury.

USER MAINTENANCE
The equipment user is responsible for the basic cleanliness of the equipment. On
a regular basis, the Q-DRT Tray, along with Radiographic Table or Wall Stand to
which it is mounted, should be wiped clean. Painted metal surfaces should be
cleaned using a clean cloth slightly moistened in warm soapy water (use mild
soap). Wipe with a clean wet cloth. Use only alcohol-based liquid disinfectants.

CAUTION! Always test cleaning and/or dis-


infecting agents, such as wipes, sprays and
hand sterilizers, on a small inconspicuous
area of the equipment prior to use. Proceed
with cleaning only if no damage occurred to
the test area.

PLANNED MAINTENANCE
Routine inspection of the Q-DRT tray should be performed by qualified service
personnel on an annual basis.
A complete series of inspections and functional checks was conducted at the time
of installation to insure proper operation of the system. The following inspection
and adjustment procedures are recommended to maintain the system in its origi-
nal operating condition.
• Check for evidence of loose hardware
• Check all bearings and bearing surfaces for cleanliness and corrosion
• Conduct a general inspection for worn or damaged parts
• Visually inspect for wear and cleanliness
REPLACEMENT PARTS AND ORDERING INFORMATION
Table 5-1 provides a list of replaceable parts for Q-DRT Rotating Trays. Figures
5-1 and 5-1a show the locations of the replaceable parts described therein.
Removal and replacement instructions are provided as required.

Model Q-DRT Series Digital Detector Trays Revision E


Quantum Medical Imaging, LLC 5-3
Chapter 5 Maintenance

CAUTION! When servicing wall stand-


mounted Q-DRT models, it is advisable to
remove the portable detector from the
rotating tray, and then remove the entire
assembly from the vertical trolley of the
wallstand, before performing replacement
procedures. Ensure that the tray is resting
on a horizontal surface that supports it
when the Sliding Tray is extended.

ORDERING INFORMATION
To order replacement parts for the Q-DRT Tray, contact the Service
Department at:
Quantum Medical Imaging, LLC
2002-B Orville Drive North
Ronkonkoma, New York 11779 USA
Phone: (631) 567-5800 (x2)
Fax: (631) 567-5074
e-mail: qmiservice@carestream.com
When ordering replacement parts, supply the following information:
• Model, reference, and serial number of equipment
• Part number
• Part description
• Quantity required
• P.O. Number
• Shipping Instructions
When ordering components or parts not listed in Table 5-1, a complete description
of the part, including its function and location should be provided with the model
number and serial number of the unit.

Revision E Model Q-DRT Series Digital Detector Trays


5-4 Quantum Medical Imaging, LLC
Chapter 5 Maintenance

Table 5-1. Table Replaceable Parts

ITEM DESCRIPTION PART NUMBER QTY

1 Phenolic Panel (Wall Stands only): 1


Rotating Trays:
Right Load Models ME31-024-1
Left Load Models ME31-024-2

2 Detent Bumper ME30-251 2

3 Tray Latch FA30-025 1

4 Chin Rest (Wall Stands only) ME12-107 1

5 Double-sided Tape (for Chin Rest) CS30-008 AR*

6 Tray Handle 1
Rotating Models ME10-681

7 Push / Pull Catch FA30-026 2

8 Sub-miniature Switch EL50-041 1

9 Push/Pull Catch Housing (shroud) ME10-727 2

10 Removable Grid (Optional)


Carestream Evolution System: AY20-336 1
Quantum Systems: AY20-325 1

* As required

Model Q-DRT Series Digital Detector Trays Revision E


Quantum Medical Imaging, LLC 5-5
Chapter 5 Maintenance

5
2
3
4

2 7
8

6 9

T:\Storage\Manuals\Q-DRT Series Casette Trays\Installation - Operation Manual\Replaceable Parts - Q-DRT Rotating Tray.cdr

Figure 5-1. Q-DRT Rotating Tray Trays - Replaceable Parts Diagram

Revision E Model Q-DRT Series Digital Detector Trays


5-6 Quantum Medical Imaging, LLC
Chapter 5 Maintenance

10

T:\Storage\Manuals\Q-DRT Series Cassette Trays\DRAWINGS\Q-DRT Rotating Housing\Q-DRT Rotating Housing - Grid Assy - Spare Parts.cdr

Figure 5-1a. Q-DRT Rotating Tray Trays - Replaceable Parts Diagram


Removal and Replacement Procedure for Phenolic Panel (Wall
Stands Only)
1. Remove four (4) 10/32 x 1/4" Hex Button Head screws from the
corners of the Cover, and two (2) 10/32 x 1-1/4" Hex Button
Head screws with spacers, to release the Cover from the Tray,
and lift the Cover to expose the Phenolic Panel (Panel).
2. Remove four (4) 10/32 x 1/4", Flat Head (100°) screws securing
the Phenolic Panel to the Rotating DR Tray, and lift the Panel
away from its position.
3. Position the replacement Panel over the appropriate tapped holes
and re-install the four (4) 10/32 x 1/4" Flat Head (100°) screws.
4. Position the Cover over the Phenolic Panel and re-install four (4)
10/32 x 1/4" Hex Button Head screws and two (2) 10/32 x 1-1/4"
Hex Button Head screws and spacer to secure the Cover.

Model Q-DRT Series Digital Detector Trays Revision E


Quantum Medical Imaging, LLC 5-7
Chapter 5 Maintenance

NOTE: The 10/32 x 1-1/4" Hex Button Head


screws and spacer are installed on the side oppo-
site the chin rest.

Removal and Replacement of Detent Bumpers (Rotary Models


Only)
This procedure requires that the portable DR detector be removed from
the Rotating Tray. Proceed as follows:
1. Grasp the drawer slide handle and pull the drawer out of the tray
to its fullest extension (a "click" will be heard).
2. Remove from the floor of the Rotating Tray, three (3)
5/16-18 x 1/4" Undercut Flat Head screws spaced at 120°, and
securing the Rotating Tray to the underlying Turntable mecha-
nism on the Sliding Tray.
3. Lift the Rotating Tray away from the Sliding Tray to expose the
Detent Bumpers, which are located adjacent to the Detent
Plunger, attached to the floor of the Sliding Tray.

DETENT
PLUNGER

DETENT
BUMPER

Figure 5-2. Detent Bumper and Plunger Assembly (Two Per Tray)
4. Remove the worn bumper, by removing one Phillips Flat Head
screw (see Figure 5-2).
5. Install the new Detent Bumper and secure it with the same Phil-
lips Flat Head screw removed in previous step.

Revision E Model Q-DRT Series Digital Detector Trays


5-8 Quantum Medical Imaging, LLC
Chapter 5 Maintenance

6. Position the Rotating Tray on the Sliding Tray, making sure that
the Detent Wheel mounted to the bottom of the Rotating Tray is
positioned between the two Detent Bumpers, then align the thru-
holes in the Rotating Tray with the tapped holes in the Turntable.
Secure the Rotating Tray to the Turntable, using three (3)
5/16-18 x 1/4" undercut Flat Head screws.
7. Inspect the installation by rotating the Tray through it’s entire
range of motion (detent to detent).
8. Ensure that the Rotating Tray is in its Portrait detent position, and
release the Sliding Tray catch by gently pulling out on the handle,
then pushing the handle in to close the drawer slide into the Tray.
9. Reinstall the assembled Q-DRT Tray on the vertical trolley of the
wall stand.
Removal and Replacement of the Sliding Tray Latch (Rotary
Models Only)
This procedure requires that the portable DR detector be removed from
the Rotating Tray. Proceed as follows:
1. Refer to Removal and Replacement of Detent Bumpers procedure
(steps 1 through 3) for instructions for gaining access to the Slid-
ing Tray Latch.
2. Once the disassembly of the Sliding Tray has been completed,
remove four (4) 4-40 x 1/4" Pan Head screws securing the Latch
to the floor of the Sliding Tray, and lift the Latch assembly out or
its position, noting its orientation before removal (see Figure
5-3 below).
3. Position the replacement Latch (in the same orientation as the
original Latch) over the through holes, and re-install four (4) 4-40
x 1/4" Pan Head screws.

NOTE: Orientation of the Latch must correspond


to the orientation of the keeper mounted to the
back wall of the Tray. On right-loading units, the
hook of the keeper is oriented up, and down on
left loading units.

4. Assemble the unit, following the instructions provided in Removal


and Replacement of Detent Bumpers in this section (steps 6
through 9).

Model Q-DRT Series Digital Detector Trays Revision E


Quantum Medical Imaging, LLC 5-9
Chapter 5 Maintenance

REAR WALL OF
HOUSING

LATCH
KEEPER

LATCH
MECHANISM

FLOOR OF SLIDING TRAY

Figure 5-3. Q-DRT Tray - Slide Tray Latch Mechanism


Removal and Replacement of the Chin Rest (Wall Stands Only
The Chin Rest on Q-DRT Trays are affixed to the Cover with double-sided
tape. The Chin Rest is removed and replaced as follows:
1. Grasp the Chin Rest, using the thumb and forefinger of both
hands, and apply gentle and constant upward pressure to over-
come the adhesive of the double-sided tape between the Cover
and the underside of the Chin Rest.
2. Once the Chin Rest has been removed, inspect the double sided
tape to determine if the tape should be replaced. If tape replace-
ment is required, remove the existing tape and any residual adhe-
sive remaining on the Cover.
3. Apply strips of double sided tape (P/N CS30-008) to the under-
side of the Chin Rest as shown in Figure 5-4.

Revision E Model Q-DRT Series Digital Detector Trays


5-10 Quantum Medical Imaging, LLC
Chapter 5 Maintenance

DOUBLE -SIDED
TAPE
(1" x 3-1/2")
P/N: CS30-008

BOTTOM
VIEW

Figure 5-4. Chin Rest - Location of Double-sided Tape


4. When Reinstalling, position the Chin Rest on the Tray Cover as
shown in Figure 5-5, and seat the against the surface of the
Cover.

Figure 5-5. Q-DRT Wall Stand Tray Chin Rest - Final Installation

Model Q-DRT Series Digital Detector Trays Revision E


Quantum Medical Imaging, LLC 5-11
Chapter 5 Maintenance

Removal and Replacement of Push/Pull Catch


The Push/Pull Catch is located on the underside of the Sliding Tray on all
Q-DRT models. The Slide Tray must be fully extended with the portable
detector removed from Tray, when replacing the Catch. In order to gain
access to the Push/Pull Catches, follow instructions 1 through 3, provided
in the Removal and Replacement of the Detent Bumpers previously
described in this chapter, and then proceed as follows.
1. On the top of the Slide Tray, locate and remove four (4)
4-40 x 3/4" Flat Head screws (82°) for each of two Push/Pull
Catches and Shroud, mounted to the bottom of the Slide Tray
(see Figure 5-6).

LOCATION OF PUSH/
PULL CATCH
HARDWARE

PUSH/PULL
CATCH AND
SHROUD (2X)

Figure 5-6. Q-DRT Rotating Model - Slide Tray Shown without Rotating Tray

Revision E Model Q-DRT Series Digital Detector Trays


5-12 Quantum Medical Imaging, LLC
Chapter 5 Maintenance

2. Once the screws holding the Catch to the Sliding Tray have been
removed, slide the Catch out from under the Tray.
3. The replacement part should be slid into position via the same
access point, and aligned with the through holes in the floor of
the Tray.
4. Install the Catch and Shroud with four (4) 4-40 x 3/4" (82°) Flat
Head screws.
5. Inspect the functionality of the Sliding Tray with the new Catch,
then reassemble the unit, using the instructions provided in
Removal and Replacement of Detent Bumpers in this section
(steps 6 through 9).
Removal and Replacement of Receptor Lock Micro Switch (Table
Models Only)

CAUTION: This procedure involves dis-


connection and re-connection of electri-
cal wiring. Before attempting the
following procedure, secure all power to
the Radiographic Table.

Q-DRT Trays designed for use on Radiographic Tables utilize a electro-


magnetic receptor lock to immobilize the tray, once it is in the desired
position. The lock is released via a lever-actuated micro switch. This
switch, and associated actuator lever, are replaceable parts in the Recep-
tor Lock assembly. Refer to Figure 5-1, for location information, and pro-
ceed as follows:
1. Remove the Phenolic Table Top, by gently lifting it from one cor-
ner and working the hook and loop tape loose around the Table.
2. Release the Receptor Lock Mechanism from its angle mounting
bracket by removing four (4) 10-32 x 1/2" screws from the rear of
the bracket.
3. Remove the cover of the lock mechanism by removing four
(4) 6-32 x1/4" Filister screws (two on each side).
4. Locate the micro switch, which is mounted to right side of the
assembly, and note electrical connections to the switch, then
remove the wiring.
5. Remove two (2) 2-56 x 3/8" Pan Head screws and flat washers
from the Switch Plate and the Switch.
6. Install the replacement Switch and reconnect the wiring.
7. Restore electrical power to the Switch, and Inspect the operation
of the Receptor Lock.

Model Q-DRT Series Digital Detector Trays Revision E


Quantum Medical Imaging, LLC 5-13
Chapter 5 Maintenance

NOTE: It should not be necessary to adjust the


Switch Plate if it is not disturbed during replace-
ment of the Switch. However, if the stroke neces-
sary (of the Receptor Lock Handle) is too short or
too long, adjust the slotted switch plate for the
desired Lock Handle stroke for switch activation.

8. Make any necessary adjustments.


9. Re-install the cover, and reattach the Phenolic Table Top.

Revision E Model Q-DRT Series Digital Detector Trays


5-14 Quantum Medical Imaging, LLC