Whiteness Meter
PCE-WNM 100
INSTRUCTION MANUAL
1. APPLICATIONS & FEATURES * Pass strict inspection and testing, the zeroing operation amplifier, inputted to
* Whiteness Meter is mainly used to instrument conforms to the JJG 512- the calibration circuit for energy
directly measure the whiteness value of 2002 Whiteness Meter Verification calibration. Then, the whiteness value of
object or powder with flat surface. Regulation. Also conforms to GB3978, the sample is directly shown on the digital
* Can be widely used in whiteness GB3979, GB7973, GB7974, ISO2470, display.
measurement of textile printing and GB8904.2, QB1840, GB2913, 3. NOTE
dyeing, paint, chemical building GB13025.2,GB1543, ISO2471, This instrument is a metrological
materials, paper board, plastic products, appliance, is strictly forbidden to
GB10339, GB5950, GB12911, GB2409
white cement, ceramics, enamel, talc disassemble, please read the instructions
WHITENESS METER powder, starch, flour, salt, detergent,
and other standards.
* Able to memory 254 groups of carefully before use.
cosmetics and other substances. * There should be no corrosive gas or
measurement data. vibration source in the operating
* It consists of Light Source, Optical * Use USB data output and RS-232 data
PCE-WNM 100 System, Detection System, Data Process output to connect with PC.
environment.
* There should be no light exposure and
and Display System. The whiteness * Provide Bluetooth data output choice. magnetic field interference.
Calculation Formula of the instrument is 2. OPERATING PRINCIPLE * The ambient air should be dry and no
blue light whiteness R457. The light source of the instrument is long- dust, etc.
* Use of high quality imported components, life LED with 457nm, conforming to the * If the instrument is not to be used for any
good reliability. light source requirement in the JJG 512- extended period, remove batteries to
* The optical path design is reasonable, 2002 National Metrological Whiteness prevent battery leakage corroding
simple. Adopt filter system with Meter Verification Regulation. internal circuit.
The measurement of the instrument adopts * To keep the measuring sensor clean, not
correction, the whiteness can be
the principle of photoelectric effect. The to touch the optical element with hand,
reflected quantitatively. diffuse reflectance Radiant Brightness
* No preheating after starting, so as not to affect the spectral
Energy Value of the sample surface is characteristics.
measurements can be taken after simple
receive by the optical cell. Then, the value
calibration, easy to operate. is amplified by high-precision auto
1 2 3

4. SPECIFICATIONS 5-13
Range: 0~120
Resolution: 0.1 5-14
Illumination condition: 45/0
Illuminant: LED 457 nm 5-11 M
5-15
Whiteness Formula: 5. PANEL DESCRIPTIONS 5-12 R

Blue Light Whiteness WB=R457


Measuring Aperture: 5-1 5-11 Measurement Symbol
18x11 mm Rectangle 5-12 Reading Symbol
5-6
Zero Drift: ≤0.1 5-7 5-13 Coupling Symbol
Stability of Display: ≤0.5 5-8 5-14 Measurement Value
Repeatability of Measurement: ≤0.1 5-2 5-15 Data Group
Operating Condition: 5-3 6. MEASUREMENT PROCEDURE
Temperature: 0~40 ℃ 5-4 5-9 6.1 Important Notice
Humidity: <85 %RH 5-10 6.1.1 Before using the instrument, it is
Power Supply: necessary to carefully read the manual,
Lithium Battery 5-5 and operate in strict accordance with
Size: 140x45x75 mm the stipulated working conditions and
Weight: 305 g (Including Batteries) 5-1 Display 5-6 Save Key
Standard Accessaries: usage methods.
5-2 Power key 5-7 Delete Key
Main Unit 5-3 Calibration Key 5-8 Read/Plus Key
6.1.2 The standard white board is used for
Standard White Board 5-4 Minus Key 5-9 RS232C Interface calibration of the instrument. It is
Optical Cleaning Cotton Cloth 5-5 Sensor of Button 5-10 AC Interface necessary to keep the surface clean with
AC Adaptor no exposure to dust. Avoid artificially
Carrying Case affecting its nominal value.
Operation Manual 6.1.3 Keep the optical components of the
4 5 instrument clean, do not rub or touch
6
them with hands. If there is dust on 6.2.1 If the uneven surface of the sample can powder forming apparatus. reading is stable. (for the details of Data
optical components, use wash ear ball not be improved, or there are difference 6.2.4 For fibrous materials, such as cotton, Storing, see 6.4.1)
to blow off dust on the surface or wipe between the horizontal textures and chemical fiber, wool, silk, etc. Firstly 6.3.2 After being used for a period of time,
with lens wiping paper. If there is oil or vertical textures, measure the sample at sort them out, comb into a vertical the instrument needs to be calibrated.
mildew on optical components, use surface, and then placed in a self-made Just place the Standard White Board on
different positions or angles, adopt the
absorbent cotton dipped in absolute test box for measurement. Sampling horizontal table, place the instrument
alcohol to wipe the surface. When the average whiteness value as the
method (including quantity) should be on the Standard White Board, attach the
instrument is not to be used for any whiteness value.
unified. It is suggested to prepare a few sensor tightly on the standard white
extended period, sheathe it in a dust 6.2.2 For paper, cloth and various fiber
more samples for measurement and Board, press and hold the CAL
cover, which is self-made. fabric, it is necessary to overlap them Calibration Key for about 1 second,
adopt the average value.
6.1.4 Before starting up, clean the measuring till opaque. If the whiteness value does 6.3 MEASUREMENT & CALIBRATION signal CAL. is display on the screen,
aperture with optical cleaning cotton not change when increasing the number 6.3.1 I n s t r u m e n t h a s b e e n d e b u g g e d , the calibration is completed.
cloth. In order to avoid contamination of samples, it is opaque. calibrated before leaving the factory, 6.4 Data Storing and Browsing
of the whiteboard and test samples. 6.2.3 For powder or small granular samples, can be used for a period of time (length 6.4.1 After starting, the Measurement
6.1.5 Test sample surface should be flat or shall be put in powder containers, Symbol is on the display, indicating the
of the time depends on the
even. The consistencyof the flatten with bright and clean glass instrument is in a state of measurement,
longitudinal and transverse directions environment). There is no need to
plate. Different preparing conditions preheat after starting, measurement can data storage can be operated.
of the tested samples should be
bring different test results, so, in order be directly taken. Press the power key to Meanwhile, the display also shows the
maintained in the repeatability test. For
samples without flat or even to establish the whiteness relationship start, the Measurement Symbol M is number of stored data. When
surface(such as powder, particles, between the same kind of samples, the displayed on the screen, indicating it is measuring, the sensor is completely
fiber), it is necessary to prepare a preparing method must be unified, in the measuring state. When measuring, attached to the sample surface, the
sample board. For details see 6.2 including weight, particle size and place the instrument on the sample, fully coupling sign appears on the screen,
Sample Preparation. pressing method. So that the samples attached the sensor to the surface of the after the reading is stable, press the
6.2 Sample Preparation have approximate density and surface sample, the coupling sign appears on the Save Key, the data is stored in the
roughness. If higher accuracy is screen. Read or store the data when the instrument, the stored data groups plus
7
required, adopt the constant pressure 10
8 9
one. The instrument can store up to 254 seconds in the measuring state. Then, all Operation Instructions .
groups of data. the data stored will be deleted. Err1, 8. MAINTENANCE
6.4.2 To browse the stored data, just press the which indicates all the data has been 8.1 Standard White Board
Read/Plus Key, the Measurement removed, is also displayed on the The configured Standard White Board
Symbol disappears on the display, screen. should be regularly sent to the higher
while the Reading Symbol appears, 7. TRANSFERRING READINGS TO A measurement units or industry test center
indicating it is in browsing state. Press COMPUTER station for approval. It is recommended to
the Read/Plus Key or Minus Key to 7.1 USB,RS232 Connection Operation approve once half a year. If contaminated,
browse the data in the instrument. 7.1.1 When RS-232C Data Cable with after cleaning it should be timely
6.4.3 To quit browsing state, just press the Software is purchased, data can be inspected and re-calibrated.
Power Key. transfered to computer. 8.2 Illuminant Lens
6.5 Data Deleting 7.1.2 Put the RS232 CD into CD-ROM of It is forbidden to touch it with hands, so as
There are two ways to delete the data, computer and install the software. not to leave a trace, effecting spectral
single delete and delete all. 7.1.3 Connect instrument with computer transmittance. After being used for a long
A. Single data can be deleted in the time, use cotton dipped in Absolute
through USB,RS232 Cable.
browsing state. Press Read/Plus Key or Alcohol to carefully wipe the lens surface
7.1.4 Switch on your gauge and ensure the
Minus Key to call out the data to be with tweezers, and then use dry cotton to
Reading Screen is displayed.
deleted, press the Delete Key, the wipe and clean it.
7.1.5 Start the software and follow the
current data is immediately deleted.
instructions included with the software
After deleting the last group of data,
Demo. EXE.
Err1 is displayed on the screen, which
7.2 Bluetooth Connection Operation
indicates that all the data has been
When Bluetooth connection module and
removed.
the adapter is purchased, data can be
B. To move all the data, just press and transferred to the computer. For detailed
hold the DEL delete key for about 3 operation please see Bluetooth Adapter 13
11 12