Shenzhen emperor electronic technology Co., Ltd.
Foreword
Dear customer,
Thanks for your purchase of Emperor Ultrasound Scanner. This operation manual describes you a
whole view of this model by introducing its design, dimensions, operation instruction, maintenance,
packaging, transportation and storage. You are strongly advised to read it carefully before using the
scanner.
The mistakes are unavoidable when it is printed. We beg your pardon if such careless omission were
made due to some reasons out of control. We reserve the right to make modification without prior
notification, and you may find out some contents that are different from the actual operation of the
scanner.
Please contact us when you find out any mistake in this manual.
NOTE: The manufacturer will not be responsible for any damage or hurt to the scanner if
the operator or the distributor does not follow our instructions.
-1-
Shenzhen emperor electronic technology Co., Ltd.
Symbol Description
Type B Equipment.
Ground.
Equal Potentiometer.
! Important.
Fuse.
-2-
Shenzhen emperor electronic technology Co., Ltd.
-3-
Shenzhen emperor electronic technology Co., Ltd.
A Information
1.1 Appearance
EMP-
860
1.2 Features
z The following technology are applied in this unit to further widen image focusing area and
increase the information volume, accordingly it obtains a finer image display at a higher
resolution.
* 4-sector dynamic focusing, four focus points can be operated independently or combined freely
to change the focus position.
* Image up/down conversion, left/right conversion, black/white conversion, image depth scroll,
etc.
z High-frequency beam-former to ensure best resolution of the image.
z Regulated power supply is adopted, SVGA monitor, flicker-free,Protection screen filter to make
operation more comfortable. And the screen could be switched to larger or smaller size.
z Easy and flexible operation with newly designed keyboard to better meet the demand of our
customers.
-4-
Shenzhen emperor electronic technology Co., Ltd.
1.3 Functions
z The image display is categorized into single-B, B/B, B/M and M modes. The image may be
frozen. Parallel images may be displayed on the same screen and either one image or the two can
be frozen as desired.
z In addition to the measurement of such parameters as distance, perimeter, area, volume (ellipse
method), angle, heart rate, EF slope, etc., the measurement of fetal parameter is especially added,
6 methods (GS, CRL, BPD, HC, AC, FL, HL) are provided for the measure of fetal age and
weight and EDCB (Expected Date of Child Birth).
z Software packages including obstetrics, gynecology, urology, cardiology, the reports generated
automatically.
z Optional wide-frequency electronic convex array probe, electronic linear array probe and
transvaginal probe.
z 4-sector emission focusing and dynamically reception focusing. The position of the focus point
is changeable freely.
z In zoom ×1.0, single B-Mode, the zoomed area could be auto-optimized and picture in picture
real time.
z The machine software can be upgraded by U-disk, and store or load image on the U-disk.
z Connecting with laser printer to print the needed image and report.
● The unit is designed and manufactured strictly in accordance with the National Standard
GB10152—1997 <B Mode Ultrasound Diagnostic Instrument>and Manufacturers’
Standard<Electrical Linear/convex B Mode Ultrasound Diagnostic Equipment>Q/19SZ
EP01-2001.
● Only the standard three-pin socket should be used and have its protection terminal connected with
grounding wire. If not, you are required to have the scanner in well grounded by the piece of
grounding wire provided by the manufacturer.
Note: Follow the experience technician’s instruction when connecting the earth wire.
-5-
Shenzhen emperor electronic technology Co., Ltd.
-6-
Shenzhen emperor electronic technology Co., Ltd.
(3)
: Ambient Conditions
Electrical Diagram
This instruments is applicable to the diagnosis of the organs such as liver, gall-bladder, kidney, spleen,
pancreas, heart, thyroid gland, breast, uterus, bladder, ovary, etc. It has been widely used in examining
abdomen and obstetrics, urology, cardiology, gynecology, small parts etc. In various hospitals at all level.
1.8 Inapplicability
This device is not applicable to the diagnosis of the pneumatic organs that contain gas, such as lung,
stomach, intestines, etc.
This transvaginal probe is not applicable to the diagnosis of the patient of vaginal inflammation, such as
trichomoniasis elytritis, fungus elytritis, venereal diseases; bachelorette; malformed vaginal;
intermenstruum; postmenopausal vaginal atrophy, colporrhagia; front placental, etc.
-7-
Shenzhen emperor electronic technology Co., Ltd.
B Installation
2) Check the unit, accessories and material attached as per Packing List (See the table below)
S/NO Item Name QTY Remark
1 Main Unit 1
2 Convex Probe(C080-60E) 1 STDU
3 Probe Holder 2
4 Power Cord 1
5 Grounding Wire 1
6 Ultrasound Gel 1
7 Screw(ф3×8) 4 Spare Parts
11 Quality Certification 1
12 Operation Manual 1
4) Laser printer
5) Foot-switch
6) Biopsy guide
7) Trolley
8) U-disk
-8-
Shenzhen emperor electronic technology Co., Ltd.
Frequency
Elements
Degree
Main
Probe Scanning 3-Step
Model Method Multi-frequency
2.3 Installation
USB Host
USB Device
Foot-switch
L
P
T
Don’t connect or
Sa
Cl
R
E
E
N
S
I
ZS
Em
Laser Printer
N
o
r
m
a
l
l
V
I
D
E
O
Probe Socket
Video Output
Power Socket Grounding Wire On/Off
1) Keep this device in dry conditions, free of any erosive gas or strong magnetic field interference.
2) Use corresponding AC power, connect one end of the 3-pin power cord with power socket on the rear
of the unit while the other in the wall socket, and the grounding wire is properly connected with the
3-pin power line.
Warning:1. When the unit is used together with other medical equipment, use supplied
grounding wire to connect the grounding of the unit with the counterpart of other
-9-
Shenzhen emperor electronic technology Co., Ltd.
2. Connect one end of supplied grounding wire to the grounding of the unit while
the other to earth for better protection. Do not take water hose or pipes as
grounding wire, otherwise, preliminary protection measures will be disabled and
the patient is likely to be electrically shocked.
3) To connect external large-scale monitor for better observation of medical display, insert the video
plug of the monitor into Video Output of the unit.
4) How to connect the video graphic printer to the unit. Connect the video input of the printer to
video output on the rear of the unit.
5) How to connect the Laser printer to the unit. Connect the unit and the laser printer with the laser
printer’s data string.
Note: The printer should be operated according its operation manual when it is power on.
The Laser printers are Hp LaserJet 1150 and Hp LaserJet 1160, and compatible with
the two models.
Importance: Please use the appointed printer otherwise the user has to bear the
responsibility of damaging the inner power.
6) Probe holder installation: Fasten the U shape faucet of container to the probe holder which is on the
right side of the main unit, and then fasten the bottom slip.
Probe Holder
7) Probe installation: Insert the probe into the probe connector properly in horizontal direction, and
drawing the probe out of the socket in horizontal direction.
8) Biopsy attachments installation:
Put biopsy needle through biopsy guide from top to down, adjust the screw2 to make it slid able.
- 10 -
Shenzhen emperor electronic technology Co., Ltd.
Warning:Diagnosis with biopsy attachments is doing under high risk, and all the
operators should be trained and qualified professionally. Manufacturer is
All the biopsy attachments should be sterilized before using, and operator
should work with sterilized glove. Biopsy line is only provided for reference.
9) USB Device: Connect the unit and the computer with the USB cable, The frozen image will be
Note:Before you transfer images to your PC, please install DRIVE PROGRAM for IMAGE
TRANAFER.THRU USB prepared by our company, please visit our website to download
and install it. Detailed operation instructions available as well. For our website,
Http://www.china-emperor.com.
USB Cable
- 11 -
Shenzhen emperor electronic technology Co., Ltd.
2.4 Important
Before connecting and operating the device this section must be read carefully.
Warnings
1) Keep the unit in dry and clean conditions, free of corrosive gas or strong magnetic interference
conditions. Any operation or storage in the environment, where the atmospheric pressure is too
high, temperature and humility exceed stipulated values or badly ventilated or there are
2) Operation room should be relatively dark, and avoid any direct strong shining on the monitor so as
3) The instrument should be evenly placed on the horizontal platform, any shock or concussion when
4) The frequency and voltage of AC is as required, and the current is strong enough. Do not use the
5) Power supply system in the operation room should be kept in good conditions and the unit should be
grounded to the earth well (The grounded pole is on the rear board).
6) Be careful not to place the patient into contact with the ultrasound scanner or other devices. If the
ultrasound scanner or other devices are defective, there is a risk of elective shock.
7) Do not pull out or insert power cord when the unit is still on. Do not take off the probe by the cable
- 12 -
Shenzhen emperor electronic technology Co., Ltd.
8) The device of grounding out of the room will be much helpful to weaken the electric and magnetic
interference.
9) Constant switching on should be gapped for 2~3 minutes. Otherwise the device will be locked and
10) Switch off the unit when trouble occurs and inform the qualified engineers from local distributor or
our factory. Do not open the unit at discretion, or our company has the right not to repair it.
Note: Do not plug or unplug the probes when the power is on to avoid any possible damage
to the probe or the unit. Do not plug or unplug at discretion after being connected.
12) Disconnect the probe while removing the equipment so as to avoid damage.
13) Avoid any shock or strike on the valuable and fragile probes. When diagnosis is interrupted, put the
probe on the hook and do remember to press down the freeze key to keep the probe in frozen state.
14) Inspect the probe carefully after a fall. Contact the manufacturer if it is necessary.
15) To protect against fire hazards, replace fuses only with the same type and rating.
16) The use of non-manufacturer’s components with this device may result in damage to the
components.
17) Handle it carefully while removing the equipment. If the equipment is repeatedly subject to
18) Installation, extensions, re-adjustments modifications or repairs must be carried our only by
authorized persons.
20) You are advised to unplug the power cord if the unit will not be used for a rather long period of
time.
Warning:
1) Do not plug or unplug the power cord when the unit is switched on. If you need
switch on the unit just after you switch it off, wait for 2~3 minutes before you
switch it on again.
- 13 -
Shenzhen emperor electronic technology Co., Ltd.
2) Do not operate the unit under the circumstance that there are combustible substances
Precautions
1) Cleaning the probe: first remove the ultrasound gel with a soft tissue, and then gently wipe the
2) Do not put the probe into the ultrasound gel when stop using it.
3) To avoid possible damage the probe must not be coiled to a diameter of less than 9cm(3.5 inch).
4) Do not use the equipment in places subject to intense electric or magnetic field. If the equipment is
5) Do not use the equipment near devices generating high frequency (such as medical telemeters and
cordless telephones). If so, the equipment may malfunction or adversely affect such devices.
7) For correct image geometry, only monitor properly adjusted by the manufacturer may be used on
the device.
8) Check the probe carefully after an accidental fall. A dangerous situation may arise due to damaged
You are kindly advised to do the following-mentioned checks before switching on the unit for
right diagnosis.
1) Ensure that grounding wire is properly connected and no broken; the grounding nut tightly screwed;
2) Ensure that operation condition is required and no electrical devices such as X ray machine, ECG
around to avoid possible interference;
3) Ensure that power cord are properly connected and not tangled with other wires.
- 14 -
Shenzhen emperor electronic technology Co., Ltd.
C Introduction
Plug power cord in the power socket and probe pin in the probe socket. Open the keyboard by push two
clips in left/right direction, and then put down the foldaway keyboards on the level table-board.
3.1 Interface
Turn on the scanner, the indicator lights on (on the additional monitor first if it is connected) . The system
starts self-checking, the sentence “System-testing, please wait…” will be displayed on the screen.
“Checking completed!
Memory OK!
Keyboard OK!
Mouse OK!”
The intelligent system will identify the probe information automatically. If there is no probe connected or
the probe is connected in mistake way, “No probe found” will be shown on the screen in “Probe info”
window. The unit will automatically recognize the probe I when there are two probes connected. Probes
- 15 -
Shenzhen emperor electronic technology Co., Ltd.
Time, Date
Probe info.
Gray Bar
Zoom
Frame.
Right/Left
Patient ID.
Depth
EMP-
860
Contrast
Bright
Cover Keyboard
Diagnosis
Patient lying on his back in bed, tightly contact probe with the to-be-detected part after being pasted with
ultrasound gel. NOTE: Press the probe gently to avoid damage. And you will have a tangent image
of that part displayed on the screen. By changing direction and slope of the probe, different tangent
images can be obtained. Freeze the required tangent image as diagnosis basis by pressing down the key
FREEZE. During this course if you find out the brightness, contrast or gain is not as desired, repeatedly
- 16 -
Shenzhen emperor electronic technology Co., Ltd.
adjust the knobs: Brightness, Contrast of the monitor or Gain (6 TGC, Gain) on the right top of keyboard
Note: Move left the TGC knobs will decrease the gain, otherwise increase the gain.
! @ # $ % * ( ) ? / - + TGC
1 2 3 4 5 6 7 8 9 0 - =
Back
Q W E R T Y U I O P Space
Caps
Lock Enter
A S D F G H J K L
;
GS CRL BPD HC AC FL ” ’
.
Function
Z X C V B N M
:
< >
Space Esc
GAIN
4
3
2
M
1 B M
B B
B
ABC
Q K
∽ Alphanumeric keys & shortcut keys for body marks
__ When the indicator of key Function is off, press down this keys for inputting
English letters.
__ Press one of them whenever the indicator of the key FUNCTION is on, the
corresponding body mark will be shown on the screen, and select among in
- 17 -
Shenzhen emperor electronic technology Co., Ltd.
Note: Press down the same shortcut keys for body marks could change
the direction of the probe mark, press the direction keys could
move the arrow position. Hide the body mark you have to enter
the MENU(Setting → body mark ).
GS FL
Z ∽ N Alphanumeric keys & shortcut keys for Obstetrics measurement
__ When the indicator of key Function is off, press down this keys for inputting
English letters.
__ Press one of them whenever the indicator of the key FUNCTION is on, the
Dual-function
__ When the indicator of key Function is off, press down this key for inputting
__ Press down the alphanumeric key M as the indicator of the key FUNCTION is
on, the system will enter the function to measure angle directly.
- 18 -
Shenzhen emperor electronic technology Co., Ltd.
Dual-function
__ When the indicator of key Function is off, press down this key for inputting
__ Press down the alphanumeric key L as the indicator of the key FUNCTION is
on, the image angle in B-mode will be changed. There are five levels
0 Alphanumeric Keys
All the above keys are dual-function. The characters on the upper of the keys
will be input when CAPS LOCK is press down and its indicator is on.
Example: Press down the key CAPS LOCK to change content to display. If the
indicator of the key CAPS LOCK is on, press down this key to show“/”, otherwise,
Direction Keys
__ It is for moving the caliper at difference direction when system is in
measurement.
__ It is for changing the body mark at difference direction when body mark is
making.
__ When indicator of the key CAPS LOCK is on, press down the left direction key,
’
.
”
,
+
= - Character keys
- 19 -
Shenzhen emperor electronic technology Co., Ltd.
Function
Function
When indicator of the key is on, press down the corresponding key to enable into
Caps
Lock
Caps Lock
__ Switch between capital and lowercase. If the indicator of this key is on, all the
Back
Space
Letters Delete
__ Move the cursor behind the wrong letter first when the annotation is doing,
press down it to delete letter. Press down this key to delete a wrong letter
once a time.
__ The key immediately reselects the menu item that was active before.
E.g., Select Obstetrics from the main menu, then Twins, then Fetus 1. Now you
can perform the measurement and press Backspace to jump immediately to the
menu where Fetus 1 can be selected.
Enter
Enter /Confirmation
__ Press it will move the cursor at the first on the next line.
__ Press it for confirmation in Menu.
Patient ID
Activating this key will display a cursor on the screen. Patient ID can be edited
Note: Press down the key, the all data of report will be cleared away.
- 20 -
Shenzhen emperor electronic technology Co., Ltd.
Probe selection
To select the probe and proper working frequency by pressing it. A menu
appeared on the screen, and all information of probes will be shown in the menu.
Space
Space
Menu
Press down this key, the menu will be appeared on the screen. Select the wanted
item by pressing the corresponding key.
Esc
Escape
Press down this key you can exit operation. Such as exit from a menu, or from a
measurement.
1 st
1 Focus Point
Press down this key when the image is unfrozen, the first focus point “A” is
selected for changing position along the scale by moving the trackball, finally
press down the else keys to settle it.
Note: The focus “A” will move between the top of the image and the second
focus. The scale is on the left on the screen.
2 th
2 Focus Point
Press down this key when the image is unfrozen, the second focus point “B” is
selected for changing position along the scale by moving the trackball, finally
press down the else keys to settle it.
Note: The focus “B” will move between the first focus and the third
focus.
3 th
3 Focus Point
- 21 -
Shenzhen emperor electronic technology Co., Ltd.
Press down this key when the image is unfrozen, the third focus point “C” is
selected for changing position along the scale by moving the trackball, finally
press down the else keys to settle it.
Note: The focus “C” will move between the second focus and the fourth
focus.
4 th
4 Focus Point
Press down this key when the image is unfrozen,, the fourth focus point “D” is
selected for changing position along the scale by moving the trackball, finally
press down the else keys to settle it.
Note: The focus “B” will move between the third focus and the bottom
of the image.
B
B-mode
Single-B display mode. Press down this key, then the screen goes back to
B B
B+B-mode
The double B-mode key selects dual images on the screen, if the screen is
already in dual image mode, press down the same key is used to select the live
B M
B+M-mode
__ Both B and M images will be displayed on the screen when this key is
pressed. Both images are real-time.
__ In B+M-mode, the same key is used to change the scanning speed of M-mode.
Continuously pressing down the same key to switch the scanning speed in
seven levels.
- 22 -
Shenzhen emperor electronic technology Co., Ltd.
M
M-mode
__ In M-mode, the same key is used to change the scanning speed of M-mode.
Continuously pressing down the same key to switch the scanning speed in
seven levels.
Distance
Volume
Heart Rate
EF slope
Start/Stop
- 23 -
Shenzhen emperor electronic technology Co., Ltd.
Image Magnification
Image Resume
Depth Scroll up
Left/Right
Up/Down
Black/White
Press down this key you can reverse the image black/white.
Local Zoom
Refer the detail in the instruction of Local Zoom and Picture In Picture.
- 24 -
Shenzhen emperor electronic technology Co., Ltd.
Save
Refer the detail in the instruction of Image store, load and delete.
ABC
Clear
Press down this key you can clear text annotation on the screen. Some fixed
characters could not be cleared by pressing this key upon they were input, such as
Freeze
Cine-memory
- 25 -
Shenzhen emperor electronic technology Co., Ltd.
D Operation Instruction
Probe 1:
1 ---- C – 2.5M
2 ---- C – 3.5M
3 ---- C – 5.0M
Probe 2:
4 ……
5 ……
6 ……
Esc…Exit
1) Press alphanumeric key 1,2……or 6 to choose the probe or the frequency of the probe(s) you have
connected.
2) If there is no probe connected, it will display “No Probe Found” when the device is switched on.
In single-B mode and ×1.0, press down the key local zoom,a menu will display on the screen:
1——Local Zoom
2——Picture In Picture
— If the screen is already under local zoom, the key LOCAL ZOOM is used to select change the
2)— press down the key Alphanumeric “2”,enter into the picture in picture.
- 26 -
Shenzhen emperor electronic technology Co., Ltd.
— If the screen is already under picture in picture, the key LOCAL ZOOM is used to select
change the zoom level, continuously pressing down it to switch in three levels(×1.2,×1.5,
×2.0).
Press down the key FREEZE, then press down the key STORE, the screen will displayed as following:
S——Store
L——Load
D——Del
— Press down the key Alphanumeric “S” to have a substrate menu appeared on the screen.
— Select alphanumeric key “1” to save the image into the host, select alphanumeric key
NOTE:If the user have not input the file name ,but press down the key ENTER
directly, the system will produce a file name automatically: patient name +
ID.
— Press down the key Alphanumeric “L” to have a substrate menu appeared on the screen.
S——Store
L——Load
1 – Load from host
D——Del
2 – Load from U Disk
— Select alphanumeric key “1” to load the image from the host, select alphanumeric key
- 27 -
Shenzhen emperor electronic technology Co., Ltd.
NOTE:The user could take basic operations on the loaded image, such as measurement,
etc. The stored image could be transferred between host and U-disk.
— Press down the key alphanumeric “D” to have a substrate menu appeared on the screen.
S——Store
L——Load 1 – Delete from host
D——Del 2 – Delete from U Disk
— Select alphanumeric key “1” to delete the image from the host, select alphanumeric key
NOTE: There are three ways of selecting the various functions or items in each menu and/or
submenu list.
1. Select the item that you prefer by scrolling the trackball up and down or press the
direction key up and down, then press down the key ENTER;
Or 2. Select the item that you prefer by scrolling the trackball up and down or press
the direction key up and down, then press down the direction key Right to enter, or
Or 3. Input the first character(number or letter) in front of each item that you need.
__ Press it when the image is frozen, the system will realize cine recall.
__ Use the trackball to scroll forward or backward through cine-memory. Quickly move the
__ During auto-replay, press trackball could pause replay to get the image wanted and do some
measurements.
- 28 -
Shenzhen emperor electronic technology Co., Ltd.
1)Press down the key FUNCTION and then activate alphanumeric key “M”, the image will be frozen
“ Dist:
Dist:
Angle:”
2)Move the cursor with trackball to the origination position where the angle should be measured.
3)Press down the key START/STOP, then the cursor will disappear, but it will appear again in case if you
move trackball or press down the DIRECTION KEYS to start measurement. Move the cursor to the final
position and press down the key START/STOP, do it again to finish the measurement. The screen will
show:
“ Dist: ***.*mm
Dist: ***.*mm
Angle: ***.*°”
cursor, and the position could be changed by moving the trackball and pressing
3)The result of last four measurements will be separately displayed on the right
1. Press down the key DISTANCE to conduct distance measurement. The screen display “Dist”;
- 29 -
Shenzhen emperor electronic technology Co., Ltd.
2. Press down the corresponding directional keys or scroll trackball to move the cursor “+” to the
3. Press down the key START/STOP and the cursor “+” disappears that the system start the measurement.
Press down the corresponding direction key or scroll Trackball, and mark “+” reappears and move the
cursor to the desired stopping point. Press down the key START/STOP to end the measurement. The
“Dist=***.*mm”
As the above display shows, move the cursor from point A to B, the unit will display the result after
systematic calculation.
of first cursor, and the position could be changed by moving the trackball and
3)The result of last four measurements will be separately displayed on the right
1) Press down the key perimeter and area measurement and indicator lights to start area measurement,
“Area:
Length: ”;
2) Press down the corresponding directional keys or scroll trackball to move the cursor “+” to the
- 30 -
Shenzhen emperor electronic technology Co., Ltd.
3) Press down the key START/STOP, the cursor disappears and the system starts the measurement. Then
press down the corresponding direction keys or scroll Trackball, the cursor “+” reappears, move the
cursor, the screen displays measured track in desired direction. When the track closes, press down the
key START/STOP to end the measurement. It displays perimeter and area results covered by the closed
curve.
Length: ***.* mm
+
“Area” refers to the area (mm2);
Refer to the above display, follow the steps previously mentioned track the cursor along the edge of
fetal’s transvaginal, you will have the perimeter and the area encircled by the curve.
2)The result of last four measurements will be separately displayed on the right
NOTE:Please choice the method of the volume measurement before you beginning to volume
measure. There have two methods three axis and single plane.
Single plane
1. Press down the key VOLUME and indicator lights to start measures. The screen prompts:
LA: SA:
Volume:
Area:
Per(perimeter):
2. Press down the corresponding directional keys or scroll trackball to move the cursor “+” to the
- 31 -
Shenzhen emperor electronic technology Co., Ltd.
3. Press down the key START/STOP, the cursor disappears and the system starts measurement. a ellipse
will display on the screen, and then scroll trackball to Change the size, position, shape of ellipse , and
then Press the key START/STOP again, the values of area/perimeter, volume will be displayed
Per(Perimeter): ***.* mm D
“LA” Axes is long axes (mm), “SA” Axes is short axes (mm), Volume: ***.* cm3 is volume value,
2)The result of last four measurements will be separately displayed on the right
Three axis
NOTE:Choice the three axis, the volume data will display after the X axis、Y axis and
Z axis be measured.
Press down the key VOLUME, the following menu will display on the screen:
1——X axis
2——Y axis
3——Z axis
Choice the X axis, the measure method is allowing to the distance method, and then measure Y axis
Following measurements can only used to get the probable growth state of fetus in a certain period, so
error will be generated if fetus is malformation or abnormality. All the parameters of following
- 32 -
Shenzhen emperor electronic technology Co., Ltd.
measurements should be calculated together to get a correct growth state of fetus as a whole. It will be
The parameters of crown-rump, head circumference, abdomen circumference and femur length are usually
used in mid and late pregnancy. Sum up all the parameters and then divide the result by the number of
measurement, then you will get the fetus week. Refer to《Family Planning Ultrasound Diagnostics, Second
edition》
GS (Predict fetus week and EDCB according to the size of womb, applicable within 7 weeks)
CRL (Predict fetus week and EDCB according to the crown-rump, applicable within 8-12 weeks. Error
will be 3~5 days. Refer to《Family Planning Ultrasound Diagnostics, Second edition》)
BPD (Predict fetus week and EDCB according to the bi-parietal diameter, applicable in 12 fetus weeks
to one month old. Fetus week<20 weeks,Error will be 1~1.5week;Fetus week is 20~30week,Error
will be1.5~2 weeks;Fetus week>30 weeks,Error will be 3~3.5weeks. Refer to《Family Planning
HC (Predict fetus week and EDCB according to head circumference, applicable in 12 fetus weeks to one
month old, and this parameter can be used to measure fetus weight. Fetus week<20 weeks,error will be
will be 3~3.5 weeks. In the last 6 weeks, the result of HC will be more reliable than the result of BPD.
AC (Predict fetus week and EDCB according to abdomen circumference, applicable within36~42
weeks, and this parameter can be used to measure fetus weight. AC is made up the shortage of PD or HC.
Error will be 3~3.5 weeks. Refer to《Family Planning Ultrasound Diagnostics, Second edition》)
FL (Predict fetus week and EDCB according to femur length, applicable within 14~22 weeks, and this
parameter can be used to measure fetus weight. Error will be ±6.7days. Refer to《Family Planning
- 33 -
Shenzhen emperor electronic technology Co., Ltd.
Remark 1: In following measurements, the unit will automatically display the fetal age and EDCB and
weight after systematic calculation. The parameters are displayed in the following forms:
Expected Date of Child Birth =EDCB xxxxxx(I.e. EDCB: 010902 means Expected Date of
1. Press down the key FUNCTION and then alphanumeric key “Z”, the lower part of the screen displays:
GS
GS=000.0mm
Following to the distance measurement to measure “GS” , the system will automatically display fetal
GS
BL
GS
YS
UT. UT.
Gs Measurement Illustration
Left: Longitudinal Section of UT for measurement of GS’s Longitudinal Radius
Right: Transverse Section of UT for measurement of lateral Radius
Note:The Bladder is required full. Measure the compete fetal bladder with maximum lateral
2. Press down the key FUNCTION and then alphanumeric key “X”, the lower part of the screen displays:
CRL
CRL=000.0mm
Rotate or tilt the probe to have better image of the fetal in the womb and ensure that the max diameter
of crown-rump is displayed and then freeze the picture. Following to the distance measurement to
measure “CRL”, the screen will automatically display fetal age and EDCB after systematic calculation.
- 34 -
Shenzhen emperor electronic technology Co., Ltd.
AMN
FH FT
3. Press down the key FUNCTION and then alphanumeric key “C”, the lower part of the screen displays:
BPD
BPD=000.0mm
From top to bottom locate the standard section in which fetal head is an ellipse, centerline of the brain
centers and the two sides symmetrical. Take the maximum diameter vertical to centerline of the brain
from echo external edge of nearer skull to echo interior edge of farer one for BPD measurement with the
method as same as distance measurement. The unit will automatically display fetal age and EDCB after
systematic calculation.
Note: Choose a standard section for measurement, if the section selected is too high, the
value is too low; if the section selected is too low, the value is rather low.
4. Press down the key FUNCTION and then alphanumeric key “V” , the lower part of the screen displays:
HC
SA=000.0mm LA=000.0mm
SA refers to short axis of fetal head; LA refers to long axis of fetal head (Unit: mm).
- 35 -
Shenzhen emperor electronic technology Co., Ltd.
Measure the maximum front-rear (LA) and horizontal diameter (SA) with standard section of frozen
images. The system will automatically display fetal age, fetal weight and EDCB after systematic
calculation as per Head circumference=(SA+LA)×1.57, The unit will automatically display fetal age
Front-rear length
Lateral Radius
5. Press down the key FUNCTION and alphanumeric key “B”, lower part of the screen displays:
AC
D1=000.0mm D2=000.0mm
D1 and D2 respectively refer to front-rear diameter and horizontal diameter of the cross-sectioned
Measure the maximum front-rear (LA) and horizontal diameter (SA) with standard section of frozen
images. Take the distance from interior edge of the abdomen to the opposite. The system will
automatically display fetal age, fetal weight and EDCB after systematic calculation as per abdomen
circular=(D1+D2)×1.57, The unit will automatically display fetal age and EDCB after systematic
calculation.
Liver
Spine
Stomach
AC Measurement Illustration
6. Press down the key FUNCTION and then alphanumeric key “N”, lower part of the screen:
FL
FL=000.0mm
- 36 -
Shenzhen emperor electronic technology Co., Ltd.
FL refers to the length between the two ends of femur (Unit: mm).
Take the length between the two ends of femur and follow the method described as same as distance
measurement. The system will automatically display fetal age and EDCB after systematic calculation.
FL
FL Measurement Illustration
Note: 1) The femur is displayed very clearly and avoid slope measurement of the femur.
2) If the section selected is too high, the value is too low; if the section selected
7. Press down the key FUNCTION and then alphanumeric key “H”, lower part of the screen:
HL
HL=000.0mm
HL refers to the length between the two ends of humeral (Unit: mm).
Take the length between the two ends of humeral and follow the method described as same as distance
measurement. The system will automatically display fetal age and EDCB after systematic calculation.
2. Measuring methods could be switched freely in your measurement. Select the method you
wanted to start measurement following steps.
3. Each measurement has value range, it will display”!!! INVALID VALUE!!!” if the result
exceeds the limitation, it will return to the previous step. The user has the choice to select
measurement mode or exit.
4. Fetal Weight is applicable to the over-20-week fetal, and the unit will display the result after
systematic judgment.
- 37 -
Shenzhen emperor electronic technology Co., Ltd.
1) Press down the key HEART RATE, and indicator on to start area measurement, the screen prompts:
displays:
“Heart Rate
Time” ;
2) Move the cursor “+” to the starting point of measurement by pressing directional keys or scrolling
3) Press down the key START/STOP, the cursor “+” will disappear. Press down the corresponding
direction keys or scroll Trackball, “+” reappears. Select another measurement point as desired. The
measurement result will be dynamically displayed on the screen. In the diagram, horizontal distance
refers to time and vertical one to the amplitude. Measured HR and EF will be automatically displayed
Display in M-mode.
+
EF can be obtained by measuring the height
& width of peak & bottom values of the wave
2)The result of last four measurements will be separately displayed on the right
- 38 -
Shenzhen emperor electronic technology Co.,Ltd
T——Time Setting On
H——Hospital Name Off
1——Settings Ö D——Doctor Name
F——Focus Number Ö On
2——Function Ö
B——Biopsy Line Ö Off
3——Obstetrics Ö
4——Gynecology Ö C——Center Line Ö
On
5——Urology Ö N——Measure Line Ö
Off
6——Cardiology Ö M——Body Mark Ö
V——Vol. Method Ö
On
(Main menu) R——Factory Setting
Off
Select Body Mark
Three axis
P——Print Single Plane
T——Text
R——Ratio Measure Suspend Mode
H——Histogram 1——10 minutes later
G——Gamma Correction 2——20 minutes later
S——Suspend Mode 3——30 minutes later
V——Version 4——60 minutes later
U——Software Update 5——No Suspend Mode
Press <ESC> to exit
- 39 -
Shenzhen emperor electronic technology Co.,Ltd
5.1 Settings
Note: There are three ways of selecting the various functions or items in each menu and/or
submenu list.
1. Select the item that you prefer by scrolling the trackball up and down or press the
direction key up and down, then press down the key ENTER;
Or 2. Select the item that you prefer by scrolling the trackball up and down or press the
direction key up and down, then press down the direction key Right to enter, or press
Or 3. Input the first character(number or letter) in front of each item that you need.
NOTE: Press down the key Esc to exit all the menus that you enter.
Return to the main menu and select Settings, upon pressing the key ENTER, following menu will display
on the screen.
1——Settings Ö
2——Function Ö
T——Time Setting
3——Obstetrics Ö
H——Hospital Name
4——Gynecology Ö
D——Doctor Name
5——Urology Ö
F——Focus Number Ö
6——Cardiology Ö
B——Biopsy Line Ö
C——Center Line Ö
N——Measure Line Ö
M——Body Mark Ö
V——Vol. Method Ö
R——Factory Setting
Time Setting
The date, week and time display in the right top corner of the screen.
The cursor appears on the screen, you can input the new date, week and time.
E.g. In this case, if you input 010105 6 181818, and then press the key ENTER, the screen will be changed
- 40 -
Shenzhen emperor electronic technology Co.,Ltd.
01/01/05 Saturday
18:18:18
The key ENTER is used to finish your setting, but the key ESC is used to escape in midway.
Note: Numerical keys 1 stands for Monday, 2 for Tuesday ……7 for Sunday;
Hospital Name
Select Hospital Name, and then press the key ENTER.
Edit the hospital name and press the key ENTER, the result that you have edited will be automatically
stored and appear on the right bottom corner of the screen.
Doctor Name
Select Doctor Name, and then press the key ENTER.
Edit the doctor name and press the key ENTER, the result that you have edited will be automatically stored
and appear on the right bottom corner of the screen.
Focus Number
Select Focus Number, and then press down the key ENTER.
1——Settings Ö
2——Function Ö
T——Time Setting
3——Obstetrics Ö
H——Hospital Name
4——Gynecology Ö
D——Doctor Name
5——Urology Ö
F——Focus Number Ö
6——Cardiology Ö
B——Biopsy Line Ö
1 focus
C——Center Line Ö
2 focus
N——Measure Line Ö
3 focus
M——Body Mark Ö
4 focus
V——Vol. Method Ö
R——Factory Setting
Steps: Select focus number firstly, and then press down corresponding focus key, scroll trackball to
- 41 -
Shenzhen emperor electronic technology Co.,Ltd.
“B” will turn to be “b”, and then move the trackball to change the position of the “b” focus. Pressing the
else keys to fix the focus position, meanwhile “b” turn to be “B”.
NOTE:In M mode and B/M mode, not more than two focuses could be selected.
Biopsy Line
Select Biopsy Line, and then press down the key ENTER.
1——Settings Ö
2——Function Ö
T——Time Setting
3——Obstetrics Ö
H——Hospital Name
4——Gynecology Ö
D——Doctor Name
5——Urology Ö
F——Focus Number Ö
6——Cardiology Ö
B——Biopsy Line Ö
C——Center Line Ö
On
N——Measure Line Ö
Off
M——Body Mark Ö
V——Vol. Method Ö
R——Factory Setting
NOTE:1) Biopsy line will not display except when convex probe in connected in B mode.
2) Biopsy line should be calibrated carefully before using. Please follow the
instruction on the screen to adjust the position and direction of the biopsy line.
Center Line
Select Center Line, and then press down the key ENTER.
1——Settings Ö
2——Function Ö
T——Time Setting
3——Obstetrics Ö
H——Hospital Name
4——Gynecology Ö
D——Doctor Name
5——Urology Ö
F——Focus Number Ö
6——Cardiology Ö
B——Biopsy Line Ö
C——Center Line Ö
N——Measure Line Ö
On
M——Body Mark Ö
Off
V——Vol. Method Ö
R——Factory Setting
- 42 -
Shenzhen emperor electronic technology Co.,Ltd.
NOTE: Center line will not display when linear probe and/or transvaginal probe connected.
Measure Line
Select Measure Line, and then press down the key ENTER.
1——Settings Ö
2——Function Ö
T——Time Setting
3——Obstetrics Ö
H——Hospital Name
4——Gynecology Ö
D——Doctor Name
5——Urology Ö
F——Focus Number Ö
6——Cardiology Ö
B——Biopsy Line Ö
C——Center Line Ö
N——Measure Line Ö
M——Body Mark Ö
On
V——Vol. Method Ö
Off
R——Factory Setting
Body Mark
Select Body Mark, and then press down the key ENTER.
1——Settings Ö
2——Function Ö
T——Time Setting
3——Obstetrics Ö
H——Hospital Name
4——Gynecology Ö
D——Doctor Name
5——Urology Ö
F——Focus Number Ö
6——Cardiology Ö
B——Biopsy Line Ö
C——Center Line Ö
N——Measure Line Ö
M——Body Mark Ö
V——Vol. Method Ö
On
R——Factory Setting
Off
Select Body Mark
Select Off to hide Body Mark (including 18 body marks on the keyboard).
Select Select body Mark, and then press the key ENTER, 20 body marks available will display on the
- 43 -
Shenzhen emperor electronic technology Co.,Ltd.
screen.
Body pregnant (right) Body pregnant (Left) Fetus (back) Fetus (back) Parasternal long axis
Head (right) Head (face) Arm (right) Arm (Left) Leg (back)
- 44 -
Shenzhen emperor electronic technology Co.,Ltd.
Note:
1)Select the needed body mark by scrolling the trackball. Pressing the key ENTER
2)In unfrozen and B/M mode, Press down or to change the direction of
arrow which is showing the scanning direction;In frozen and B/M mode,and the
Vol. Method
Select Vol. Method,then press down the key ENTER,the user can be choose the volume measurement
Factory Setting
Select Factory Setting,then press down the key ENTER, the setting which the user could be resetting will
5.2 Function
Return to the main menu and select Function. Upon pressing the key ENTER, following menu will display
on the screen.
- 45 -
Shenzhen emperor electronic technology Co.,Ltd.
1——Settings Ö
2——Function Ö
3——Obstetrics Ö
P——Print
4——Gynecology Ö
T——Text
5——Urology Ö
R——Ratio Measure
6——Cardiology Ö
H——Histogram
G——Gamma Correction
S——Suspend Mode
V——Version
U——Software Update
Freeze the image, enter into the menu to select Print and confirm with the key ENTER.
Note: 1) Connected Laser printer after switched on the scanner for 10 minutes.
Text
Select Text, and then press down the key ENTER, a cursor which is moveable with trackball will appear
on the screen. Text could be input or edited in different places within image area on the screen and then
Note: During input process, the key ENTER is used to changed to a new line, the key ESC
to finish the input and escape the operation, the key CLEAR to clear the test. And
Ratio Measure
Select ratio measure, and then press down the key ENTER, the screen will show two- distance
measurement. According to the methods for distance measurements, you will get two distance values. The
ratio between the second distance and the first distance values will automatically appear on the screen.
Histogram
Select Histogram, the screen will display a square, press the alphanumeric key “H” to enlarge or reduce
the square, scroll trackball to change the position of the square, then press the key ENTER, the screen will
display a histogram on the image, and then press the key CLEAR to clear it.
- 46 -
Shenzhen emperor electronic technology Co.,Ltd.
Gamma Correction
Eight correction curves are available, and pressing corresponding number to select different ones which
will be stored into the system automatically even switched off the system.
Suspend Mode
Select Suspend Mode, then press down the key ENTER. The following menu will display on the screen:
Suspend Mode
1——10 minutes later
2——20 minutes later
3——30 minutes later
4——60 minutes later
5——No Suspend Mode
Press <ESC> to exit
Example: If you select 1, in case the machine standbys for more than 10 minutes, it will be enter into the
suspend state. In this case, the probe and the machine will not work, and press any key to exit.
If you select 2, in case the machine standbys for more than 20 minutes, it will be enter into the suspend
state. In this case, the probe and the machine will not work, and press any key to exit; etc.
Version
Software Update
Connected U-disk, enter into menu to select Software Update, press key ENTER to confirm. Follow the
instruction on the screen. Software update need a few minutes, any operation is forbidden to avoid
damage or malfunction.
Note:Please format U-disk before using it. In this course, please ensure the power supply
on level off.
- 47 -
Shenzhen emperor electronic technology Co.,Ltd
F Obstetrics
Note: There are three ways of selecting the various functions or items in each menu and/or
submenu list.
1. Select the item that you prefer by scrolling the trackball up and down or press the
direction key up and down, then press down the key ENTER;
Or 2. Select the item that you prefer by scrolling the trackball up and down or press the
direction key up and down, then press down the direction key Right to enter, or press
or 3. Input the first character(number or letter) in front of each item that you need.
Note: Press down the key Esc to exit all the menus that you enter.
Return to the main menu and select Obstetrics. Upon pressing the key ENTER, following menu will
1——Settings Ö
2——Function Ö
3——Obstetrics Ö
4——Gynecology Ö
Z——GS
5——Urology Ö
X——CRL
6——Cardiology Ö
C——BPD
V——HC
B——AC
N——FL
H——HL
I——AFI Ö
T——Twins Ö
P——LMP Measure
O——Obstetrics Report
S——Select Methods Ö
U——User Methods Ö
Gestational Sac (GS)
Select GS, then press down the key ENTER, the screen will display as following:
GS
GS=000.0mm
Press down the corresponding directional keys or scroll trackball as per your requirement to move the
cursor to the starting point of each measurement. When measure “GS”, The system will automatically
display fetal age and EDCB after systematic calculation, at the same time the measurement data will be
stored in the report. For the same patient, measure “GS” again, the newest measurement data display on
- 49 -
Shenzhen emperor electronic technology Co.,Ltd.
the report.
Select CRL, then press down the key ENTER, the screen will display as following:
CRL
CRL=000.0mm
Rotate or tilt the probe to have better image of the fetal in the womb and ensure that the max diameter of
crown-rump is displayed and then freeze the picture. Measure “CRL” in proper order, the screen will
automatically display fetal age and EDCB after systematic calculation, at the same time the measurement
data will be stored in the report. For the same patient, measure “CRL” again, the newest measurement
Select BPD, then press down the key ENTER, the screen will display as following:
BPD
BPD=000.0mm
From top to bottom locate the standard section in which fetal head is an ellipse, centerline of the brain
centers and the two sides symmetrical. Take the maximum diameter vertical to centerline of the brain from
echo external edge of nearer skull to echo interior edge of farer one for “BPD” measurement with the
method described in Remark 2. The unit will automatically display fetal age and EDCB after systematic
calculation, at the same time the measurement data will be stored in the report. For the same patient,
measure “BPD” again, the newest measurement data display on the report.
Select HC, then press down the key ENTER, the screen will display as following:
HC
SA=000.0mm LA=000.0mm
SA refers to short axis of fetal head; LA refers to long axis of fetal head (Unit: mm).
Measure the maximum front-rear (LA) and horizontal diameter (SA) with standard section of frozen
images. The system will automatically display fetal age, fetal weight and EDCB after systematic
- 50 -
Shenzhen emperor electronic technology Co.,Ltd.
calculation, at the same time the measurement data will be stored in the report. For the same patient,
measure “HC” again, the newest measurement data display on the report.
Select AC, then press down the key ENTER, the screen will display as following:
AC
D1=000.0mm D2=000.0mm
D1 and D2 respectively refer to front-rear diameter and horizontal diameter of the cross-sectioned
Measure the maximum front-rear (LA) and horizontal diameter (SA) with standard section of frozen
images. Take the distance from interior edge of the abdomen to the opposite. The system will
automatically display fetal age, fetal weight and EDCB after systematic calculation, at the same time the
measurement data will be stored in the report. For the same patient, measure “AC” again, the newest
Select FL, then press down the key ENTER, the screen will display as following:
FL
FL=000.0mm
FL refers to the length between the two ends of femur (Unit: mm).
Take the length between the two ends of femur and follow the method described in Remark 2. The system
will automatically display fetal age, fetal weight and EDCB after systematic calculation, at the same time
the measurement data will be stored in the report. For the same patient, measure “FL” again, the newest
Select HL, then press down the key ENTER, the screen will display as following:
HL
HL=000.0mm
HL refers to the length between the two ends of humeral (Unit: mm).
- 51 -
Shenzhen emperor electronic technology Co.,Ltd.
Take the length between the two ends of humeral and follow the method described in Remark 2. The
system will automatically display fetal age, fetal weight and EDCB after systematic calculation, at the
same time the measurement data will be stored in the report. For the same patient, measure “HL” again,
Select AFI, then press down the key ENTER, the screen will display as following:
1——Settings Ö
2——Function Ö
3——Obstetrics Ö
4——Gynecology Ö
Z——GS
5——Urology Ö
X——CRL
6——Cardiology Ö
C——BPD
V——HC
B——AC
N——FL
H——HL
I——AFI Ö
T——Twins Ö
Distance 1
P——LMP Measure
Distance 2
O——Obstetrics Report
Distance 3
S——Select Methods Ö
Distance 4
U——User Methods Ö
Note: The measurement method of AFI refers to the distance measure. Select Distance 1 for
the first quadrant measurement. Repeat these measurements for the second to the
fourth quadrant.
Twins
Select Twins, then press down the key ENTER, the screen will display as following:
- 52 -
Shenzhen emperor electronic technology Co.,Ltd.
1——Settings Ö
2——Function Ö
3——Obstetrics Ö
4——Gynecology Ö
Z——GS
5——Urology Ö
X——CRL
6——Cardiology Ö
C——BPD
V——HC
B——AC
N——FL
H——HL
I——AFI Ö
T——Twins Ö
P——LMP Measure
Fetus 1
O——Obstetrics Report
Fetus 2
S——Select Methods Ö
U——User Methods Ö
Select fetus1, the data will be saved as one record in the report of the first fetus.
Select fetus2, the data will be saved as one record in the report of the second fetus.
Note: If you select fetus2 in the twins’ option, the report will display two fetus’ data;
LMP Measurement
Select LMP Measurement, and then press down the key ENTER, in the right top corner of screen will
display as following:
After inputting the last menstrual period (LMP), the Expected Date of Child Birth (EDCB) will
automatically display on the screen. The data will not be saved in the report.
- 53 -
Shenzhen emperor electronic technology Co.,Ltd.
Obstetrics Report
Hospital Name
GS:
CRL:
BPD:
HC:
AC:
FL:
HL:
LMP:
AFI:
Fetal weight:
Doctor:
Select Obstetrics Report, and then press the key ENTER, the report will be display on the screen:
Note: The content of the report couldn’t be changed or erased by user. If you selected
fetus2 in the twins’ option, the report will display as report2; otherwise, the report
Note:
1. Once you choose each software packages, all the measurement data will be saved in the
report.
2. For the same patient, the data in the report will be updated if you carry out the
measurement upon the same item again, and the system will show the newest data as well.
3. When you press down the key New Patient, the data that saved in the report for the
4. When connect the laser printer, Press down the alphanumeric key P to print the report.
- 54 -
Shenzhen emperor electronic technology Co.,Ltd.
Select Methods
Select Select Methods, upon pressing the key ENTER, following menu will display on the screen.
1——Settings Ö
2——Function Ö
3——Obstetrics Ö
4——Gynecology Ö
Z——GS
5——Urology Ö
X——CRL
6——Cardiology Ö
C——BPD
V——HC
B——AC
N——FL
H——HL
I——AFI Ö
T——Twins Ö
P——LMP Measure
O——Obstetrics Report
S——Select Methods Ö
U——User Methods Ö
GS Ö
CRL Ö
BPD Ö
HC Ö
AC Ö
FL Ö
HL Ö
EFW Ö
For example, for GS measurement, there are four formulas: Hellman, Rempen, Tokyo University and
User.
GS Ö
CRL Ö
Hellman
BPD Ö
Rempen
HC Ö
Tokyo University
AC Ö
User
FL Ö
HL Ö
EFW Ö
Note: 1. CRL, BPD, HC, AC, FL and HLhave similar formulas for users’ option. The default
- 55 -
Shenzhen emperor electronic technology Co.,Ltd.
User Methods
Select User Methods, upon pressing the key ENTER, following menu will display on the screen.
Z——GS
X——CRL
C——BPD
V——HC
B——AC
N——FL GS
H——HL CRL
I——AFI BPD
Ö
T——Twins HC
Ö
P——LMP Measurement AC
O——Obstetrics Report FL
S——Select Methods HL
Ö
U——User Methods Ö
Users could define their own formula for the seven OB measurements. The setting will be stored into the
Input the fetus age and date. Press ENTER to confirm and press ESC to exit.
User’s setting will be saved into the User formula which is selectable in Select Method.
- 56 -
Shenzhen emperor electronic technology Co.,Ltd
G Gynecology
Length
AP Diameter
Width
- 57 -
Shenzhen emperor electronic technology Co.,Ltd
Note: There are three ways of selecting the various functions or items in each menu and/or
submenu list.
1. Select the item that you prefer by scrolling the trackball up and down or press the
direction key up and down, then press down the key ENTER;
Or 2. Select the item that you prefer by scrolling the trackball up and down or press the
direction key up and down, then press down the direction key Right to enter, or press
Or 3. Input the first character(number or letter) in front of each item that you need.
Note: Press down the key Esc to exit all the menus that you enter.
Return to the main menu and select Gynecology. Upon pressing the key ENTER, following menu will
1——Settings Ö
2——Function Ö
3——Obstetrics Ö
4——Gynecology Ö
5——Urology Ö
U——Uterus Ö
6——Cardiology Ö
L——Left Ovary Ö
R——Right Ovary Ö
E——Endometrial Thickness
G——Gynecology Report
Uterus
Selecting Uterus will generate a submenu, from which three measurements for the uterus can be
performed. Press ENTER and the following menu will appear on the screen.
Length
AP Diameter
Width
a) Length
The uterus is displayed in the sagittal plane, The length is measured from the fundus to the cervix of the
Uterus.
b) AP Diameter
- 58 -
Shenzhen emperor electronic technology Co.,Ltd.
The uterus is displayed in the sagittal plane. With this measurement the AP diameter of the corpus of the
Uterus is measured.
c) Width
The fundus of the uterus is displayed in the transverse plane, With this measurement the width of the
Note: Select one of the three measurements and the menu will disappear from the screen,
A caliper cross will be displayed in the image and you can now perform the
measurement.
Left Ovary
Selecting Left Ovary will generate a submenu, from which three measurements for the uterus can be
performed. Press ENTER and the following menu will appear on the screen.
Length
AP Diameter
Width
a) Length
The length of the left ovary is measured with the ovary displayed in the sagittal plane.
b) AP Diameter
The AP diameter of the left ovary is measured with the ovary displayed in the sagittal plane of the
transverse plane.
c) Width
The width of the left ovary is measured with the ovary displayed in the transverse plane.
Note: Select one of the three measurements and the menu will disappear from the screen,
a caliper cross will be displayed in the image and you can now perform the
measurement.
Right Ovary
Selecting Right Ovary will generate a submenu, from which three measurements for the
uterus can be performed. Press ENTER and the following menu will appear on the screen.
- 59 -
Shenzhen emperor electronic technology Co.,Ltd.
Length
AP Diameter
Width
a) Length
The length of the right ovary is measured with the ovary displayed in the sagittal plane.
b) AP Diameter
The AP diameter of the right ovary is measured with the ovary displayed in the sagittal plane of the
transverse plane.
c) Width
The width of the right ovary is measured with the ovary displayed in the transverse plane.
Note: Select one of the three measurements and the menu will disappear from the screen,
a caliper cross will be displayed in the image and you can now perform the
measurement.
Endometrial Thickness
Select Endometrial Thickness and the menu will disappear in the screen, a caliper cross will be
displayed in the image and you can now perform the measurement.
The Endometrial Thickness is measured at the height of the fundus of the uterus with the uterus
Gynecology Report
Select Gynecology Report, then press down the key ENTER, t the report will displayed on the screen.
- 60 -
Shenzhen emperor electronic technology Co.,Ltd.
Hospital Name:
Patient: Section: Date:
Uterus Length:
Uterus AP Diameter:
Uterus Width:
left ovary Length:
left ovary AP Diameter:
left ovary Width:
Right ovary Length:
Right ovary AP Diameter:
Right ovary Width:
Endometrial Thickness:
Doctor:
Note:
1. Once you choose each software packages, all the measurement data will be saved in the
report.
2. For the same patient, the data in the report will be updated if you carry out the
measurement upon the same item again, and the system will show the newest data as well.
3. When you press down the key New Patient, the data that saved in the report for the
4. When connect the laser printer, Press down the alphanumeric key P to print the report.
- 61 -
Shenzhen emperor electronic technology Co.,Ltd
H Urology
- 62 -
Shenzhen emperor electronic technology Co.,Ltd
Note: There are three ways of selecting the various functions or items in each menu and/or
submenu list.
1. Select the item that you prefer by scrolling the trackball up and down or press the
direction key up and down, then press down the key ENTER;
Or 2. Select the item that you prefer by scrolling the trackball up and down or press the
direction key up and down, then press down the direction key Right to enter, or press
Or 3. Input the first character(number or letter) in front of each item that you need.
Note: Press down the key Esc to exit all the menus that you enter.
Return to the main menu and select urology. Upon pressing the key ENTER, following menu will display
on the screen.
1——Settings Ö
2——Function Ö
3——Obstetrics Ö
4——Gynecology Ö
5——Urology Ö
6——Cardiology Ö
B——Bladder Ö
L——Left Kidney
R——Right Kidney
P——Prostate Ö
U——Urology Report
S——Set Bladder Coefficients
Bladder
Selecting Bladder will generate a submenu,from which four measurements for the Bladder can be
performed,press ENTER and the following menu will appear on the screen.
Transversal
Sagittal
Oblique 1
Oblique 2
a) Transversal
Select Transversal and the menu will disappear from the screen, the screen will display as following:
Bladder Transversal
- 63 -
Shenzhen emperor electronic technology Co.,Ltd.
H= B=
Press down the corresponding directional keys or scroll trackball as per your requirement. The
measurement method of the height(H) and breadth(B) refer to the distance measurement. When measure
“Bladder Transversal”, the data will automatically display on the screen, at the same time the
measurement data will be saved in the report. For the same patient, measure “Bladder Transversal” again,
b) Sagittal
Select sagittal and the menu will disappear in the screen, the screen will display as following:
Bladder sagittal
H= B=
Press down the corresponding directional keys or scroll trackball as per your requirement. The
measurement method of the height(H) and breadth(B) are order to the distance measurement. When
measure “Bladder sagittal”, the data will automatically display on the screen, at the same time the
measurement data will be stored in the report. For the same patient, measure “Bladder sagittal” again, the
c) Oblique 1
Select Oblique 1 and the menu will disappear from the screen, the screen will display as following:
Bladder Oblique 1
H= B=
Press down the corresponding directional keys or scroll trackball as per your requirement. The
measurement method of the height(H) and breadth(B) are order to the distance measurement. When
measure “Bladder Oblique 1”, the data will automatically display on the screen, at the same time the
measurement data will be stored in the report. For the same patient, measure “Bladder Oblique 1” again,
d) Oblique 2
Select Oblique 2 and the menu will disappear from the screen, the screen will display as following:
Bladder Oblique 2
H= B=
Press down the corresponding directional keys or scroll trackball as per your requirement. The
measurement method of the height(H) and breadth(B) are order to the distance measurement. When
- 64 -
Shenzhen emperor electronic technology Co.,Ltd.
measure “Bladder Oblique 2”, the data will automatically display on the screen, at the same time the
measurement data will be stored in the report. For the same patient, measure “Bladder Oblique 2” again,
Left Kidney
Select Left Kidney and the menu will disappear from the screen, the screen will display as following:
Left Kidney
L= H=
Press down the corresponding directional keys or scroll trackball as per your requirement. The
measurement method of the length(L) and height(H) are order to the distance measurement. When
measure “Left Kidney”, the data will automatically display on the screen, at the same time the
measurement data will be stored in the report. For the same patient, measure “Left Kidney” again, the
Right Kidney
Select Right Kidney and the menu will disappear from the screen, the screen will display as following:
Right Kidney
L= H=
Press down the corresponding directional keys or scroll trackball as per your requirement. The
measurement method of the length(L) and height(H) are order to the distance measurement. When
measure “right Kidney”, the data will automatically display on the screen, at the same time the
measurement data will be stored in the report. For the same patient, measure “right Kidney” again, the
Prostate
Selecting Prostate will generate a submenu,from which four measurements for the Bladder can be
performed,press ENTER and the following menu will appear on the screen.
Transversal
Sagittal
Oblique 1
Oblique 2
- 65 -
Shenzhen emperor electronic technology Co.,Ltd.
Note: The prostate’ masurement method is order to the bladder’ measurement method.
Urology Report
Hospital Name:
Patient: Section: Date:
Bladder Transverse Sagittal Oblique 1 Oblique 2
Height:
Breadth:
Vb1= γ=1.23
Vb2= α=1.00 β=0.00
Vbmean2=
Vbmean3=
Vbmean4=
Left kidney Right kidney
Length:
Height:
Prostate Transverse Sagittal Oblique 1 Oblique 2
Height:
Breadth:
Vp1=
Vp2=
Vpmean2=
Vpmean3=
Vpmean4=
Doctor:
Note:
1. Once you choose each software packages, all the measurement data will be saved in the
report.
2. For the same patient, the data in the report will be updated if you carry out the
measurement upon the same item again, and the system will show the newest data as well.
3. When you press down the key New Patient, the data that saved in the report for the
4. When connect the laser printer, Press down the alphanumeric key P to print the report.
- 66 -
Shenzhen emperor electronic technology Co.,Ltd.
V=1/6×п×H×B2
With this method the bladder is scanned in, the transverse plane only. The estimated bladder volume is:
Vb1=V×γ
Where Vb1 is the measured volume in the transverse plane, γ is the coefficient.
a) According to this method one first measures the H/B in the transverse plane and then the length in the
Vb2=αV +β
b) Volume calculation according to the average for the measured transverse and sagittal volume:
Where Vob1 is the measured volume in the Oblique 1 plane. Where Vob2 is the measured volume in the
Oblique 2 plane.
When the volume of the bladder is calculated according to measurements made in two planes, a correction
factor is introduced into the formula for calculation of the bladder or bladder geometry yielding.
Select Set Bladder Coefficients, in the right top corner of screen will display as following:
The correction factors for α、β、γcan now be entered. Pressing the key ENTER to finish
the set.
α 1.00
β 0.00
γ 1.23
Note: Please preset the coefficient before you start your measurement if you need. The
user-set coefficient will be lost and it returns to the default. The users are usually
- 67 -
Shenzhen emperor electronic technology Co.,Ltd.
I Cardiology
Diastole
Systole
Aortic Cusp Separation
AV Ejection Time
D-E Excursion
E-F0 Slope
E-F Slope
A-C Interval
E-IVS Distance
E-F Slope
B-C Slope
A-Wave Amplitude
PV Ejection Time
Note: There are three ways of selecting the various functions or items in each menu and/or
submenu list.
1. Select the item that you prefer by scrolling the trackball up and down or press the
direction key up and down, then press down the key ENTER;
Or 2. Select the item that you prefer by scrolling the trackball up and down or press the
direction key up and down, then press down the direction key Right to enter, or press
Or 3. Input the first character(number or letter) in front of each item that you need.
Note: Press down the key Esc to exit all the menus that you enter.
Return to the main menu and select Cardiology. Upon pressing the key ENTER, following menu will
1——Settings Ö
2——Function Ö
3——Obstetrics Ö
4——Gynecology Ö
5——Urology Ö
6——Cardiology Ö
B——Left Ventricle B Ö
M——Left Ventricle M Ö
A——Left Atrium Ö
T——Mitral Valve Ö
P——pulmonary Valve Ö
C——Cardiology Report
S——Bmode Method Ö
Before you start your measurements for the Left Ventricle in B-mode, the measurement method must be
chosen.
- 69 -
Shenzhen Emperor Electronic Technology Co., Ltd.
Bmode Method
Display the Cardiology’s submenu, and then select Bmode Method, the following menu will be
displayed.
These three items represent three different sets of formulas from which the following diagnostic can be
calculated:
a) End-diastolic volume;
b) End-systolic volume;
c) Stroke volume;
d) Ejection fraction;
e) Cardiac output.
After select Dodge Single Plane, then return to the main menu and Select Left Ventricle B, then press
B——Left Ventricle B Ö
1——Settings Ö
M——Left Ventricle M Ö
2——Function Ö Diastole 2 Chamber
A——Left Atrium Ö
3——Obstetrics Ö Systole 2 Chamber
T——Mitral Valve Ö
4——Gynecology Ö Diastole 4 Chamber
P——pulmonary Valve Ö
5——Urology Ö Systole 4 Chamber
C——Cardiology Report
6——Cardiology Ö Outflow Tract Diameter
S——Bmode Method Ö
This menu displays the different measurements for the left ventricle during End-diastole and
End-systole, only the Diastole 4 Chamber, Systole 4 Chamber and Outflow Tract Diameter can be
selected.
For these calculations of the left ventricle an apical four chamber view must be displayed.
- 70 -
Shenzhen Emperor Electronic Technology Co., Ltd.
In four chamber views the length is derived from the midpoint of the mitral annulus to the most distant
point, presumed to be the apex of the posterior position in the short view.
When all the necessary measurements are stored the scanner will perform the necessary calculations
according to dodge single plane formula, the results will be displayed in the cardiology report.
After selecting Dodge Biplane, then return to the main menu and Select Left Ventricle B, then press
1——Settings Ö
2——Function Ö B——Left Ventricle B Ö
3——Obstetrics Ö M——Left Ventricle M Ö
4——Gynecology Ö Diastole 2 Chamber
A——Left Atrium Ö
5——Urology Ö Systole 2 Chamber
T——Mitral Valve Ö
6——Cardiology Ö Diastole 4 Chamber
P——pulmonary Valve Ö
Systole 4 Chamber
C——Cardiology Report
Outflow Tract Diameter
S——Bmode Method Ö
This menu displays the different measurements in the 4 chamber vie and the 2chamber view for the
For these calculations of the left ventricle an apical two chamber view must be displayed.
- 71 -
Shenzhen Emperor Electronic Technology Co., Ltd.
In 4 chamber and 2 chamber views, the length is derived from the midpoint of the mitral annulus to the
most distant point the apex of the posterior position in the short axis view.
When all the necessary measurements are stored the scanner will perform the necessary calculation
according the dodge biplane formula. The results will be displayed in the cardiology report.
After select Bullet, then return to the main menu and Select Left Ventricle B, then press down the key
1——Settings Ö
2——Function Ö B——Left Ventricle B Ö
3——Obstetrics Ö M——Left Ventricle M Ö
4——Gynecology Ö Diastole 2 Chamber
A——Left Atrium Ö
5——Urology Ö Systole 2 Chamber
T——Mitral Valve Ö
6——Cardiology Ö Diastole 4 Chamber
P——pulmonary Valve Ö
Systole 4 Chamber
C——Cardiology Report
Outflow Tract Diameter
S——Bmode Method Ö
This menu displays the different measurements for the left ventricle during End-diastole and
End-systole, only the Diastole 4 Chamber, Systole 4 Chamber and Outflow Tract Diameter can be
selected.
For all calculation of the left ventricle you must display an apical four chamber view.
- 72 -
Shenzhen Emperor Electronic Technology Co., Ltd.
How to different parameters for the calculation of the volume of the left ventricle according the bullet
In four chamber views the length is derived from the midpoint of the mitral annulus to the most distant
point, presumed to be the apex of the posterior position in the short axis view:
Select Left Ventricle, then press down the key ENTER, the screen will display as following:
1——Settings Ö
2——Function Ö
3——Obstetrics Ö
4——Gynecology Ö
B——Left Ventricle B Ö
5——Urology Ö
M——Left Ventricle M Ö
6——Cardiology Ö
A——Left Atrium Ö
Diastole
T——Mitral Valve Ö
Systole
P——pulmonary Valve Ö
C——Cardiology Report
S——Bmode Method Ö
This menu displays the different measurements for the left ventricle during End-diastole and End-systole.
For these calculations of the left ventricle a parasternal view of the left ventricle must be displayed.
- 73 -
Shenzhen Emperor Electronic Technology Co., Ltd.
Select Left Atrium, then press down the key ENTER, the screen will display as following:
1——Settings Ö
2——Function Ö
3——Obstetrics Ö
4——Gynecology Ö
5——Urology Ö
6——Cardiology Ö
B——Left Ventricle B Ö
M——Left Ventricle M Ö
A——Left Atrium Ö
T——Mitral Valve Ö
Diastole
P——pulmonary valve Ö
Systole
C——Cardiology Report
Aortic Cusp Separation
S——Bmode Method Ö
AV Ejection Time
The menu displays the different measurements for the left atrium and aorta during End-diastole and
End-systole.
- 74 -
Shenzhen Emperor Electronic Technology Co., Ltd.
Select Mitral Valve, then press down the key ENTER, the screen will display as following:
1——Settings Ö
2——Function Ö
3——Obstetrics Ö
4——Gynecology Ö
5——Urology Ö
6——Cardiology Ö
B——Left Ventricle B Ö
M——Left Ventricle M Ö
A——Left Atrium Ö
T——Mitral Valve Ö
P——pulmonary valve Ö
D-E Excursion
C——Cardiology Report
E-F0 Slope
S——Bmode Method Ö
E-F Slope
A-C Interval
E-IVS Distance
This menu displays the different measurements for the Mitral Valve.
For these calculations of the Mitral Valve a parasternal long axis view of the left ventricle, left atrium and
- 75 -
Shenzhen Emperor Electronic Technology Co., Ltd.
E-F Slope = Early diastolic posterior motion(measured with faster heart rates)
A-C Interval = Time interval between the peak of the A wave and the C point
E-IVS Distance = Distance between the IV septum and the anterior mitral valve
Select Pulmonary Valve, then press down the key ENTER, the screen will display as following:
1——Settings Ö
2——Function Ö
3——Obstetrics Ö
4——Gynecology Ö
5——Urology Ö
6——Cardiology Ö
B——Left Ventricle B Ö
M——Left Ventricle M Ö
A——Left Atrium Ö
T——Mitral Valve Ö
P——pulmonary valve Ö
C——Cardiology Report
E-F Slope
S——Bmode Method Ö
B-C Slope
A-Wave Amplitude
PV Ejection Time
This menu displayed the different measurements for the pulmonary valve.
- 76 -
Shenzhen Emperor Electronic Technology Co., Ltd.
Cardiology Report
Select Cardiology Report, then press down the key ENTER, the screen will display as following:
- 77 -
Shenzhen Emperor Electronic Technology Co., Ltd.
Hospital Name:
Patient: Section: Date:
Left Ventricle B left Ventricle M
End Diastolic V:
End Systolic V:
Stroke Volume:
Ejection Fraction:
Left Atrium
Diastole Systole
LA diameter:
LA posterior wall:
Aortic Diameter:
AC Separation
AV Ejection Time:
Mitral Valve
D-E Excursion:
E-F0 Slope:
E-F Slope:
A-C Interval:
E-IVS Distance:
Pulmonary Valve
E-F Slope:
B-C Slope:
A-Wave Amplitude:
PV Ejection Time:
Doctor:
Note:
1. Once you choose each software packages, all the measurement data will be saved in the
report.
2. For the same patient, the data in the report will be updated if you carry out the
measurement upon the same item again, and the system will show the newest data as well.
3. When you press down the key New Patient, the data that saved in the report for the
4. When connect the laser printer, Press down the alphanumeric key P to print the report.
- 78 -
Shenzhen Emperor Electronic Technology Co., Ltd.
J Maintenance
8.1) Check power supply regularly. Do not switch on this unit when the power exceeds the stipulated
value;
8.2) Check the power cord and probe wire regularly, change them immediately if they are found out
broken or damaged;
8.3) Unplug the power cord before cleaning the unit; remember to disconnect the probe from the unit
before cleaning the probe with soft and dry cloth. Do not use wet cloth, or the cloth with detergent
8.4) Clean and disinfect the probe by tampon with medical alcohol after each examination and place it in
the probe holder or in its original storage case. Never use any organic liquid for cleaning;
8.5) Do not presumptuously discover the inner for cleaning, or shake or disassemble the parts of the unit;
8.6) Please check the ventilation fans for proper operation. Contact our customer service center
immediately if it dose not work. Do not use the equipment with defective fan so as to avoid
unpredictable result;
8.7) In order to keep good ventilation, certain distance must be kept between the back of equipment and
8.8) The unit should be placed evenly and stably. Handle it with care and do not put any heavy object on
the unit. While removing the equipment, one hand holds the top handle and the other hand holds the
bottom;
8.9) The contact agent for the probe is medical ultrasound coupling which is used to make sure a complete
contact between the probe and body parts and to obtain good clinic image. Some ultrasound gels may
be toxic to human reproductive cells. It is recommended that you check with the product’s
8.10) Do not dip the probe in any other conductive solutions to avoid any corrosion of the probe and its
protective sleeve;
- 79 -
Shenzhen Emperor Electronic Technology Co., Ltd.
8.11) Do not disconnect the probe from the unit at discretion when they are properly connected;
8.12) Weaken the brightness and contrast if the unit is not used after switching on to ensure its life;
8.13) Ask /Call qualified engineer(s) to repair the unit when it breakdowns;
8.14) Power on the unit at least twice, each lasting for not less than one hour every week if the unit will
8.15) Probe cleaning: Remove the ultrasound gel with a soft tissue and then gently wipe the probe dry by
using a new tissue or dry cloth. When further cleaning is required only a mild detergent or hand
soap may be used together with some water and a soft tissue cloth;
8.16) Probe maintenance: Inspect the probe carefully after a fall. A dangerous situation may arise due to
Warning: Do not drop or shock the probe slightly since the damaged chip is not reparable;
8.17) How to clean the trackball: If the trackball can be not moved effectively but image displayed is ok,
check if there is any dust or dirt inside it since they can easily get in. Please follow the following
steps:
1.Remove the trackball cover by rotating it unclockwisely with a tiny screw driver in the recessed
spot;
5.Put the trackball in place and direct the recess at the protrusion, and turn the cover clockwisely
with a tiny screwdriver in the recessed spot until the cover in its original potions.
- 80 -
Shenzhen Emperor Electronic Technology Co., Ltd.
Rotation Point
8.18) Service life will be six years calculated from delivery date which is stated in the quality certificate.
8.19)The device should be disused complying to relevant legal regulations and principles when it is out of
using lifetime.
- 81 -
Shenzhen Emperor Electronic Technology Co., Ltd.
K Trouble-shooting
9.1) How to change the fuse
Break out the fuse from the power socket, replace the old fuse with a new one, and then fix the
9.2) Please refer to the following table for other troubles that may happen:
3 Image isn’t clear enough 1. Adjust the 6 TGC and the gain;
4 Image isn’t clear in near, Adjust the 6 TGC and the gain;
medium or far distance
5 Good image but the fan 1. Brush off all dirt or dust inside the fan;
doesn’t work when the unit
is on 2. Change a new one if the fan is broken down;
- 82 -
Shenzhen Emperor Electronic Technology Co., Ltd.
10.2) Take out the units and power them on for four hours if the units have been put in stock for more
than 6 months, and put them back according to the instructions marked outside the carton.
10.3) Do not stack the units or let the units contact the ground directly. Remember to leave some space
10.4) Shock absorption has been deliberately considered when the parking is designed and they are
suitable for air, railway, road and sea freight. Do not shock or place the products upside down and
- 83 -
Shenzhen Emperor Electronic Technology Co., Ltd.
Appendix I
Table 1-1 Relations between Fetal Age and GS’s Average Inner Radius
Fetal Age Average Inner Radius Fetal Age Average Inner Radius
4.2 0.4 7.3 2.6
4.3 0.5 7.4 2.7
4.5 0.6 7.6 2.8
4.6 0.7 7.8 2.9
4.8 0.8 7.9 3.0
4.9 0.9 8.1 3.1
5.1 1.0 8.2 3.2
5.2 1.1 8.3 3.3
5.3 1.2 8.5 3.4
5.5 1.3 8.6 3.5
5.6 1.4 8.8 3.6
5.8 1.5 8.9 3.7
5.9 1.6 9.1 3.8
6.1 1.7 9.2 3.9
6.2 1.8 9.3 4.0
6.3 1.9 9.5 4.1
6.4 2.0 9.6 4.2
6.6 2.1 9.8 4.3
6.8 2.2 9.9 4.4
6.9 2.3 10.1 4.5
7.1 2.4 10.2 4.6
7.2 2.5 10.3 4.7
- 84 -
Shenzhen Emperor Electronic Technology Co., Ltd.
- 85 -
Shenzhen Emperor Electronic Technology Co., Ltd.
Fetal Age No.of samples Average Fetal Age No.of samples Average
13 35 2.52 27 51 6.98
14 36 2.83 28 50 7.24
15 34 3.23 29 50 7.50
16 33 3.62 30 47 7.83
17 34 3.27 31 48 8.06
18 44 4.25 32 48 8.17
19 38 4.52 33 39 8.50
20 36 4.88 34 55 8.61
21 41 5.22 35 57 8.70
22 50 5.45 36 55 8.81
23 44 5.80 37 53 9.00
24 52 6.05 38 53 9.08
25 41 6.39 39 51 9.21
26 50 6.68 40 33 9.28
- 86 -
Shenzhen Emperor Electronic Technology Co., Ltd.
- 87 -
Shenzhen Emperor Electronic Technology Co., Ltd.
- 88 -
Shenzhen Emperor Electronic Technology Co., Ltd.
- 89 -
Shenzhen Emperor Electronic Technology Co., Ltd.
Appendix II
Mode B M
Parameter
P-(MPa) 2.06 2.08
Ispta(mW/cm )
2
6.40 7.01
1 SEG FOC FOC:8cm 1 SEG FOC FOC:8cm
System setting ×1.0 ×1.0
Work Freq:3.5MHz Work Freq:3.5MHz
Lp(mm) 60 60
Wpb6(‖)(mm) 3.36 3.36
Wpb6(⊥)(mm) 7.27 7.36
Prr(kHz) —— 3.13
Srr(Hz) 25.00 ——
Output beam dimensions
12.03×13.98 12.05×13.98
(mm×mm)
ƒ awf(MHz) 2.78 2.84
APF(%) 100 100
AIF(%) 100 100
Pmax(mW) 25.21 27.48
Iob(mW/cm )
2
15.70 17.08
Power-up Mode B B
Initialization Mode B B
Acoustic output freeze Yes Yes
Ltt(mm) —— ——
Lts(mm) —— ——
- 90 -
Shenzhen Emperor Electronic Technology Co., Ltd.
Mode B M
Parameter
P-(MPa) 3.41 3.41
Ispta(mW/cm )
2
17.28 17.24
1 SEG FOC FOC:4cm 1 SEG FOC FOC:4cm
System setting ×1.0 ×1.0
Work Freq:6.5MHz Work Freq:6.5MHz
Lp(mm) 30 30
Wpb6(‖)(mm) 1.82 1.74
Wpb6(⊥)(mm) 2.07 2.16
Prr(kHz) —— 5.44
Srr(Hz) 42.00 ——
Output beam dimensions
7.32×8.54 9.88×8.13
(mm×mm)
ƒ awf(MHz) 3.73 3.70
APF(%) 100 100
AIF(%) 100 100
Pmax(mW) 31.80 22.04
Iob(mW/cm )
2
53.27 28.74
Power-up Mode B B
Initialization Mode B B
Acoustic output freeze Yes Yes
Ltt(mm) —— ——
Lts(mm) —— ——
- 91 -
Shenzhen Emperor Electronic Technology Co., Ltd.
Mode B M
Parameter
P-(MPa) 1.99 1.99
Ispta(mW/cm )
2
3.24 4.10
1 SEG FOC FOC:4cm 1 SEG FOC FOC:4cm
System setting ×1.0 ×1.0
Work Freq:6.5MHz Work Freq:6.5MHz
Lp(mm) 32 32
Wpb6(‖)(mm) 3.55 2.99
Wpb6(⊥)(mm) 2.14 2.06
Prr(kHz) —— 5.41
Srr(Hz) 41.70 ——
Output beam dimensions
5.01×8.43 5.48×8.87
(mm×mm)
ƒ awf(MHz) 3.68 3.66
APF(%) 100 100
AIF(%) 100 100
Pmax(mW) 7.15 7.76
Iob(mW/cm )
2
16.92 15.98
Power-up Mode B B
Initialization Mode B B
Acoustic output freeze Yes Yes
Ltt(mm) —— ——
Lts(mm) —— ——
- 92 -