Shenzhen emperor electronic technology Co., Ltd.

Foreword

Dear customer,
Thanks for your purchase of Emperor Ultrasound Scanner. This operation manual describes you a
whole view of this model by introducing its design, dimensions, operation instruction, maintenance,
packaging, transportation and storage. You are strongly advised to read it carefully before using the
scanner.

The mistakes are unavoidable when it is printed. We beg your pardon if such careless omission were
made due to some reasons out of control. We reserve the right to make modification without prior
notification, and you may find out some contents that are different from the actual operation of the
scanner.

Please contact us when you find out any mistake in this manual.

Before using the scanner, you are advised to:


Make sure that all the parts and documentations are packed before installation. Please contact the
distributor or manufacturer if there is any thing left, and keep all the documentations well for
future reference.

NOTE: The manufacturer will not be responsible for any damage or hurt to the scanner if
the operator or the distributor does not follow our instructions.

-1-
Shenzhen emperor electronic technology Co., Ltd.

Symbol Description

Type B Equipment.

○ Cut off Power Supply.

Connect Power Supply.

Ground.

Equal Potentiometer.

! Important.

Fuse.

-2-
Shenzhen emperor electronic technology Co., Ltd.

S/No. Chapter Page


A Information 3
1.1 Appearance 3
1.2 Features 4
1.3 Functions 5
1.4 Safety Standard 5
1.5 Technical Specifications 6
1.6 Block Diagram 7
1.7 Application Scope 7
1.8 Inapplicability 7
B Installation 8
2.1 Check when Unpack 8
2.2 Optional Parts 8
2.3 Installation 9
2.4 Important 12
2.5 Check before Switching on the Unit 14
C Introduction 15
3.1 Interface 15
3.2 The Keyboard 17
3.2.1 The alphanumeric keys 17
3.2.2 Alphanumeric Keys/Character Keys 19
3.2.3 Setting Keys 20
3.2.4 Focus Points 21
3.2.5 Display Mode 22
3.2.6 Function Keys 23
D Operation Instruction 26
4.1 General Settings 26
4.1.1 Select the probe and the probe frequency 26
4.1.2 Local Zoom and Picture in picture 26
4.1.3 Image Store, Load & Delete 27
4.1.4 Image Cine-memory 28
4.2 Various Measurements 29
4.2.1 Angle Measurement 29
4.2.2 Distance Measurement 29
4.2.3 Area and Perimeter Measurement 30
4.2.4 Volume Measurement 31
4.3 Fetal age, weight and EDCB measurement 32
4.4 M-mode Measurement 38
E Settings and Function 39
5.1 Settings 40
5.2 Function 45
F Obstetrics 48
G Gynecology 57
H Urology 62
I Cardiology 68
J Maintenance 79
K Trouble Shooting 82
L Packaging, Storage and Transportation 83
Appendix I 84
Appendix II 90

-3-
Shenzhen emperor electronic technology Co., Ltd.

A Information

1.1 Appearance

EMP-
860

1.2 Features

z The following technology are applied in this unit to further widen image focusing area and
increase the information volume, accordingly it obtains a finer image display at a higher
resolution.

* 4-sector dynamic focusing, four focus points can be operated independently or combined freely
to change the focus position.

* Dynamic logarithm compress, TGC control and Wave filtering.

* 8 Gamma correction curves, pre-processing, transitional frame, histogram, interpolation, digital


scanning (DSC), post-processing, etc.

* Image up/down conversion, left/right conversion, black/white conversion, image depth scroll,
etc.
z High-frequency beam-former to ensure best resolution of the image.

z High-density probe and wider frequency band optimize image quality.

z Regulated power supply is adopted, SVGA monitor, flicker-free,Protection screen filter to make
operation more comfortable. And the screen could be switched to larger or smaller size.

z Easy and flexible operation with newly designed keyboard to better meet the demand of our
customers.

z Light-touch keyboard and trackball.

z This unit is attractively designed, plastic injection, small-sized, lightweight.

-4-
Shenzhen emperor electronic technology Co., Ltd.

1.3 Functions

z The image display is categorized into single-B, B/B, B/M and M modes. The image may be
frozen. Parallel images may be displayed on the same screen and either one image or the two can
be frozen as desired.

z The direct operation keys for easy and quick operation.

z In addition to the measurement of such parameters as distance, perimeter, area, volume (ellipse
method), angle, heart rate, EF slope, etc., the measurement of fetal parameter is especially added,
6 methods (GS, CRL, BPD, HC, AC, FL, HL) are provided for the measure of fetal age and
weight and EDCB (Expected Date of Child Birth).

z Software packages including obstetrics, gynecology, urology, cardiology, the reports generated
automatically.

z Optional wide-frequency electronic convex array probe, electronic linear array probe and
transvaginal probe.

z 4-sector emission focusing and dynamically reception focusing. The position of the focus point
is changeable freely.

z In zoom ×1.0, single B-Mode, the zoomed area could be auto-optimized and picture in picture
real time.

z Dual-direction 256-frame cine-loop, 16-image built-in memory.

z Optional biopsy line, the line can be adjusted manually.

z The image can be uploaded to computer through the USB port.

z The machine software can be upgraded by U-disk, and store or load image on the U-disk.

z Connecting with laser printer to print the needed image and report.

z The unit could be connected with video graphic printer, foot-switch.

1.4 Safety Standard

● The unit is designed and manufactured strictly in accordance with the National Standard
GB10152—1997 <B Mode Ultrasound Diagnostic Instrument>and Manufacturers’
Standard<Electrical Linear/convex B Mode Ultrasound Diagnostic Equipment>Q/19SZ
EP01-2001.

● Protection against electric shock:Class I, Type B


Protection against harmful ingress of water:Ordinary

● Only the standard three-pin socket should be used and have its protection terminal connected with
grounding wire. If not, you are required to have the scanner in well grounded by the piece of
grounding wire provided by the manufacturer.

Note: Follow the experience technician’s instruction when connecting the earth wire.

-5-
Shenzhen emperor electronic technology Co., Ltd.

1.5 Technical Specifications


(1)
: Technical Specification
Display Mode B, B/B, B/M, M
Gray Scales 256
Frame/Rate 36 frame/second
Monitor 10″SVGA (B/W)
Video Output PAL
Digital Scanning Converter 800×600
Zoom ×1.0,×1.2,×1.5, ×2.0
Body Marks 38 types indicating probe positions
General Measurement Distance, area/perimeter, volume(ellipse method), angle, heart
rate, Time, EF slope, fetal age , weight and EDCB
Software Measurement Obstetrics :GS、CRL、BPD、HC、AC、FL、HL,and LMP measure
EDD etc;
Gynecology :Uterus、Left/Right Ovary、Endometrial Thickness;
Urology :Bladder、Left/ Right Kidney、Prostate;
Cardiology: Left Ventricle 、 Left Atrium、 Mitral Valve 、
Pulmonary Valve;
Character Date, week, time, probe frequency, frame rate, ID No., doctor
name, hospital name, annotation, value, body mark, probe
position, full-screen edit, character note etc.
Obstetrics, gynecology, urology, cardiology’s report generated
automatically
Power Supply AC 220V/50Hz 110V/60Hz
Input 115VA
Continuous Operation Period ≥8 hours
Dimensions 380mm(width) *335mm (height) *465mm (depth)
Weight Approx. 12kg
(2)
: Specifications of the Standard Configuration Probe

Probe Model C080-60E


Probe Frequency Standard=2.5/3.5/5.0 MHz
Probe Type R60/60°electronic convex Array probe
Number of Element 80
Radius 60 mm
Angle 60°
Scanning Scope 190 mm Scanning Angle: 60°
Lateral Resolution ≤ 2 mm
Axial Resolution ≤ 1 mm
Blind Zone ≤ 5 mm
Max. Measure Depth ≥ 190 mm
Geometric Position Precision ≤ 4%( Horizontal), ≤4% (Vertical)
Area Tolerance ≤ 14 %
perimeter Tolerance ≤ 10 %

-6-
Shenzhen emperor electronic technology Co., Ltd.

(3)
: Ambient Conditions

Ambient Temperature Relative Humidity Atmospheric Pressure


Operation +5 ℃~+40℃ < 80% 700hPa-1060hPa
Transportation -40℃~+55℃ < 95% 700hPa-1060hPa
Storage -20℃~+40℃ < 93% 700hPa-1060hPa

1.6 Block Diagram

Probe Receiving, A/D Processing MAddress


Focus &
Enlarge

Video Video Frame


Remitting Control B Address
Synchronous Synthesis RAM

Crystal CPU Symbol, D/A Grey Read/Write


Character Scale

Keyboard (trackball) Monitor Power Supply

Electrical Diagram

1.7 Application Scope

This instruments is applicable to the diagnosis of the organs such as liver, gall-bladder, kidney, spleen,
pancreas, heart, thyroid gland, breast, uterus, bladder, ovary, etc. It has been widely used in examining
abdomen and obstetrics, urology, cardiology, gynecology, small parts etc. In various hospitals at all level.

1.8 Inapplicability

This device is not applicable to the diagnosis of the pneumatic organs that contain gas, such as lung,
stomach, intestines, etc.

This transvaginal probe is not applicable to the diagnosis of the patient of vaginal inflammation, such as
trichomoniasis elytritis, fungus elytritis, venereal diseases; bachelorette; malformed vaginal;
intermenstruum; postmenopausal vaginal atrophy, colporrhagia; front placental, etc.

-7-
Shenzhen emperor electronic technology Co., Ltd.

B Installation

2.1 Check when unpack


1) Follow the indications marked outside the carton box when unload;

2) Check the unit, accessories and material attached as per Packing List (See the table below)
S/NO Item Name QTY Remark
1 Main Unit 1
2 Convex Probe(C080-60E) 1 STDU
3 Probe Holder 2
4 Power Cord 1
5 Grounding Wire 1
6 Ultrasound Gel 1
7 Screw(ф3×8) 4 Spare Parts

8 Screw(ф3×14, with Gaskets) 2 Spare Parts

9 Fuse(Φ5×20 F2A/250V) 2 Spare Parts

10 Quality Inspection Report 1

11 Quality Certification 1
12 Operation Manual 1

3) Start the installation if no problem found out in your check.

2.2 Optional Parts

1) 5.0/6.5/7.5MHZ,80-element multifrequency electronic transvaginal probe(C080-13D)

2) 5.0/6.5/7.5MHz, 80-element multifrequency electronic linear array probe(L080-57A)

3) Video Graphic Printer

4) Laser printer

5) Foot-switch

6) Biopsy guide

7) Trolley

8) U-disk

-8-
Shenzhen emperor electronic technology Co., Ltd.

Frequency (MHz) Others Applications

Frequency

Elements

Degree
Main
Probe Scanning 3-Step
Model Method Multi-frequency

C080-60E electronic 3.5


2.5/3.5//5.0MHz 80-element
R60
Convex MHz 60°

L080-57A electronic 6.5


5.0/6.5/7.5MHz 80-element
57m
Linear MHz m

C080-13D electronic 6.5


5.0/6.5/7.5MHz 80-element
R13
0
Convex MHz 110

Main Application Alternative Application

2.3 Installation

The real panel of the device, as follows:

USB Host
USB Device
Foot-switch
L
P
T

Don’t connect or
Sa
Cl
R
E
E
N
S
I
ZS
Em

! disconnect the probe

Laser Printer
N
o
r
m

a
l
l
V
I
D
E
O

when the scanner is on!


n O
O f
f

Switch the Screen


AA
CC
21
21
0
V0
//
56
00
HH
zz
V

Probe Socket
Video Output
Power Socket Grounding Wire On/Off

1) Keep this device in dry conditions, free of any erosive gas or strong magnetic field interference.

2) Use corresponding AC power, connect one end of the 3-pin power cord with power socket on the rear
of the unit while the other in the wall socket, and the grounding wire is properly connected with the
3-pin power line.

Warning:1. When the unit is used together with other medical equipment, use supplied
grounding wire to connect the grounding of the unit with the counterpart of other

-9-
Shenzhen emperor electronic technology Co., Ltd.

equipment so as to prevent the patients from electrical shock. If current leakage


occurs to other equipment, it helps lead it the ground;

2. Connect one end of supplied grounding wire to the grounding of the unit while
the other to earth for better protection. Do not take water hose or pipes as
grounding wire, otherwise, preliminary protection measures will be disabled and
the patient is likely to be electrically shocked.

3) To connect external large-scale monitor for better observation of medical display, insert the video
plug of the monitor into Video Output of the unit.

4) How to connect the video graphic printer to the unit. Connect the video input of the printer to
video output on the rear of the unit.

5) How to connect the Laser printer to the unit. Connect the unit and the laser printer with the laser
printer’s data string.

Note: The printer should be operated according its operation manual when it is power on.
The Laser printers are Hp LaserJet 1150 and Hp LaserJet 1160, and compatible with
the two models.

Importance: Please use the appointed printer otherwise the user has to bear the
responsibility of damaging the inner power.

6) Probe holder installation: Fasten the U shape faucet of container to the probe holder which is on the
right side of the main unit, and then fasten the bottom slip.

Probe Holder

7) Probe installation: Insert the probe into the probe connector properly in horizontal direction, and
drawing the probe out of the socket in horizontal direction.
8) Biopsy attachments installation:

Fasten biopsy body to the probe closely with screw1;

Fasten biopsy guide to biopsy body with screw2;

Put biopsy needle through biopsy guide from top to down, adjust the screw2 to make it slid able.

- 10 -
Shenzhen emperor electronic technology Co., Ltd.

Warning:Diagnosis with biopsy attachments is doing under high risk, and all the
operators should be trained and qualified professionally. Manufacturer is

not responsible for the risk during this operation.

All the biopsy attachments should be sterilized before using, and operator

should work with sterilized glove. Biopsy line is only provided for reference.

Biopsy line and biopsy needle have be checked and calibrated to be

superposition before using, otherwise, it is forbidden to use. Please contact

with manufacturer or local distributors.

9) USB Device: Connect the unit and the computer with the USB cable, The frozen image will be

uploaded to the computer through the USB Device.

Note:Before you transfer images to your PC, please install DRIVE PROGRAM for IMAGE

TRANAFER.THRU USB prepared by our company, please visit our website to download

and install it. Detailed operation instructions available as well. For our website,

Http://www.china-emperor.com.

USB Cable

- 11 -
Shenzhen emperor electronic technology Co., Ltd.

Plug “A” is to connect Plug “B” is to connect


the USB of the PC USB device on the back
panel of the unit

10) USB Host:The connecter for U-disk.

11)Foot-switch:Freeze or unfreeze the image through step the foot-switch.

2.4 Important

Before connecting and operating the device this section must be read carefully.

Warnings

1) Keep the unit in dry and clean conditions, free of corrosive gas or strong magnetic interference

conditions. Any operation or storage in the environment, where the atmospheric pressure is too

high, temperature and humility exceed stipulated values or badly ventilated or there are

combustibles, explosive substances and highly corrosive articles, are prohibited.

2) Operation room should be relatively dark, and avoid any direct strong shining on the monitor so as

to obtain good clinical image.

3) The instrument should be evenly placed on the horizontal platform, any shock or concussion when

being moved should be avoided.

4) The frequency and voltage of AC is as required, and the current is strong enough. Do not use the

equipment while AC is higher than 250 V.

5) Power supply system in the operation room should be kept in good conditions and the unit should be

grounded to the earth well (The grounded pole is on the rear board).

6) Be careful not to place the patient into contact with the ultrasound scanner or other devices. If the

ultrasound scanner or other devices are defective, there is a risk of elective shock.

7) Do not pull out or insert power cord when the unit is still on. Do not take off the probe by the cable

but the plug.

- 12 -
Shenzhen emperor electronic technology Co., Ltd.

8) The device of grounding out of the room will be much helpful to weaken the electric and magnetic

interference.

9) Constant switching on should be gapped for 2~3 minutes. Otherwise the device will be locked and

you could open it again after 2~3 minutes.

10) Switch off the unit when trouble occurs and inform the qualified engineers from local distributor or

our factory. Do not open the unit at discretion, or our company has the right not to repair it.

11) Connect the probe when power supply is off.

Note: Do not plug or unplug the probes when the power is on to avoid any possible damage

to the probe or the unit. Do not plug or unplug at discretion after being connected.

12) Disconnect the probe while removing the equipment so as to avoid damage.

13) Avoid any shock or strike on the valuable and fragile probes. When diagnosis is interrupted, put the

probe on the hook and do remember to press down the freeze key to keep the probe in frozen state.

14) Inspect the probe carefully after a fall. Contact the manufacturer if it is necessary.

15) To protect against fire hazards, replace fuses only with the same type and rating.

16) The use of non-manufacturer’s components with this device may result in damage to the

components.

17) Handle it carefully while removing the equipment. If the equipment is repeatedly subject to

excessive shock, mechanical parts may be damaged.

18) Installation, extensions, re-adjustments modifications or repairs must be carried our only by

authorized persons.

19) The device must be used as per Operation Manual.

20) You are advised to unplug the power cord if the unit will not be used for a rather long period of

time.

Warning:

1) Do not plug or unplug the power cord when the unit is switched on. If you need

switch on the unit just after you switch it off, wait for 2~3 minutes before you

switch it on again.

- 13 -
Shenzhen emperor electronic technology Co., Ltd.

2) Do not operate the unit under the circumstance that there are combustible substances

as anesthetic, oxygen or hydrogen, etc to avoid possible explosion or fire.

Precautions

1) Cleaning the probe: first remove the ultrasound gel with a soft tissue, and then gently wipe the

probe dry using a new tissue or dry cloth.

2) Do not put the probe into the ultrasound gel when stop using it.

3) To avoid possible damage the probe must not be coiled to a diameter of less than 9cm(3.5 inch).

4) Do not use the equipment in places subject to intense electric or magnetic field. If the equipment is

used in such location, the monitor will be adversely affected.

5) Do not use the equipment near devices generating high frequency (such as medical telemeters and

cordless telephones). If so, the equipment may malfunction or adversely affect such devices.

6) Do not install it near a heater or in direct sunlight.

7) For correct image geometry, only monitor properly adjusted by the manufacturer may be used on

the device.

8) Check the probe carefully after an accidental fall. A dangerous situation may arise due to damaged

insulation of the probe surface. There is a risk of electrical shock.

2.5 Check before switching on the unit

You are kindly advised to do the following-mentioned checks before switching on the unit for
right diagnosis.

1) Ensure that grounding wire is properly connected and no broken; the grounding nut tightly screwed;

2) Ensure that operation condition is required and no electrical devices such as X ray machine, ECG
around to avoid possible interference;

3) Ensure that power cord are properly connected and not tangled with other wires.

- 14 -
Shenzhen emperor electronic technology Co., Ltd.

C Introduction

Plug power cord in the power socket and probe pin in the probe socket. Open the keyboard by push two

clips in left/right direction, and then put down the foldaway keyboards on the level table-board.

3.1 Interface

Turn on the scanner, the indicator lights on (on the additional monitor first if it is connected) . The system

starts self-checking, the sentence “System-testing, please wait…” will be displayed on the screen.

When the system is test over, the screen will be displayed:

“Checking completed!

Memory OK!

Keyboard OK!

Mouse OK!”

The intelligent system will identify the probe information automatically. If there is no probe connected or

the probe is connected in mistake way, “No probe found” will be shown on the screen in “Probe info”

window. The unit will automatically recognize the probe I when there are two probes connected. Probes

could be plugged in either of the probe socket.

- 15 -
Shenzhen emperor electronic technology Co., Ltd.

Time, Date

Probe info.
Gray Bar
Zoom

Frame.

Right/Left

Patient ID.

Depth

Brightness and Contrast Controllers


Adjust the knobs ” Bright” and “Contrast” for better image on screen.

EMP-
860

Contrast
Bright
Cover Keyboard

Diagnosis

Patient lying on his back in bed, tightly contact probe with the to-be-detected part after being pasted with

ultrasound gel. NOTE: Press the probe gently to avoid damage. And you will have a tangent image

of that part displayed on the screen. By changing direction and slope of the probe, different tangent

images can be obtained. Freeze the required tangent image as diagnosis basis by pressing down the key

FREEZE. During this course if you find out the brightness, contrast or gain is not as desired, repeatedly

- 16 -
Shenzhen emperor electronic technology Co., Ltd.

adjust the knobs: Brightness, Contrast of the monitor or Gain (6 TGC, Gain) on the right top of keyboard

till clear images are displayed.

Note: Move left the TGC knobs will decrease the gain, otherwise increase the gain.

3.2 The Keyboard

! @ # $ % * ( ) ? / - + TGC
1 2 3 4 5 6 7 8 9 0 - =

Back
Q W E R T Y U I O P Space

Caps
Lock Enter
A S D F G H J K L

;
GS CRL BPD HC AC FL ” ’

.
Function
Z X C V B N M

:
< >
Space Esc

GAIN

4
3
2
M
1 B M
B B
B

ABC

3.2.1 The Alphanumeric keys

Q K
∽ Alphanumeric keys & shortcut keys for body marks

All are dual-function

__ When the indicator of key Function is off, press down this keys for inputting

English letters.

__ Press one of them whenever the indicator of the key FUNCTION is on, the

corresponding body mark will be shown on the screen, and select among in

this body marks freely.

- 17 -
Shenzhen emperor electronic technology Co., Ltd.

Foetus Foetus Foetus Foetus

Foetus Foetus Body female Body male

Body female Back Body(right) Body(left)

Breast Neck(with chin up) Uterus(sliver) ovary

Prostate(sliver) thyroid gland

Note: Press down the same shortcut keys for body marks could change
the direction of the probe mark, press the direction keys could
move the arrow position. Hide the body mark you have to enter
the MENU(Setting → body mark ).

GS FL
Z ∽ N Alphanumeric keys & shortcut keys for Obstetrics measurement

All are dual-function

__ When the indicator of key Function is off, press down this keys for inputting

English letters.

__ Press one of them whenever the indicator of the key FUNCTION is on, the

system is switched into the corresponding item of obstetrics measurement.

M Alphanumeric M & shortcut key for Angle measurement

Dual-function

__ When the indicator of key Function is off, press down this key for inputting

English letter “M”.

__ Press down the alphanumeric key M as the indicator of the key FUNCTION is

on, the system will enter the function to measure angle directly.

- 18 -
Shenzhen emperor electronic technology Co., Ltd.

L Alphanumeric L & Shortcut Key for Angle Changing

Dual-function
__ When the indicator of key Function is off, press down this key for inputting

English letter “L”.

__ Press down the alphanumeric key L as the indicator of the key FUNCTION is

on, the image angle in B-mode will be changed. There are five levels

circularly by pressing it in series.

3.2.2 Alphanumeric Keys/Character Keys


/

0 Alphanumeric Keys

All the above keys are dual-function. The characters on the upper of the keys
will be input when CAPS LOCK is press down and its indicator is on.

Example: Press down the key CAPS LOCK to change content to display. If the

indicator of the key CAPS LOCK is on, press down this key to show“/”, otherwise,

it will display “0”, so are the following numerical /Character keys.


!

1 ∽ 9 Alphanumeric Keys/Character Keys


:

Direction Keys
__ It is for moving the caliper at difference direction when system is in
measurement.
__ It is for changing the body mark at difference direction when body mark is
making.

__ It is for inputting the corresponding character when system is in annotation.

__ When indicator of the key CAPS LOCK is on, press down the left direction key,

input the letter“‹ ”, otherwise, it is used as direction key.


.


,
+

= - Character keys

Note: The operation is same as the alphanumeric keys.

- 19 -
Shenzhen emperor electronic technology Co., Ltd.

3.2.3 Setting Keys

Function
Function

When indicator of the key is on, press down the corresponding key to enable into

the corresponding function in directly.

Caps
Lock
Caps Lock

__ Switch between capital and lowercase. If the indicator of this key is on, all the

letters input will be in capital, otherwise in lowercase;

__ It can be used as shift key.

Back
Space
Letters Delete

__ Move the cursor behind the wrong letter first when the annotation is doing,
press down it to delete letter. Press down this key to delete a wrong letter
once a time.
__ The key immediately reselects the menu item that was active before.
E.g., Select Obstetrics from the main menu, then Twins, then Fetus 1. Now you
can perform the measurement and press Backspace to jump immediately to the
menu where Fetus 1 can be selected.

Enter
Enter /Confirmation

__ Press it will move the cursor at the first on the next line.
__ Press it for confirmation in Menu.

Patient ID

Activating this key will display a cursor on the screen. Patient ID can be edited

or modification in this state.

Note: Press down the key, the all data of report will be cleared away.

- 20 -
Shenzhen emperor electronic technology Co., Ltd.

Probe selection

To select the probe and proper working frequency by pressing it. A menu
appeared on the screen, and all information of probes will be shown in the menu.

Space
Space

Menu

Press down this key, the menu will be appeared on the screen. Select the wanted
item by pressing the corresponding key.

Esc
Escape

Press down this key you can exit operation. Such as exit from a menu, or from a

measurement.

3.2.4 Focus points

1 st
1 Focus Point

Press down this key when the image is unfrozen, the first focus point “A” is
selected for changing position along the scale by moving the trackball, finally
press down the else keys to settle it.
Note: The focus “A” will move between the top of the image and the second
focus. The scale is on the left on the screen.

2 th
2 Focus Point

Press down this key when the image is unfrozen, the second focus point “B” is
selected for changing position along the scale by moving the trackball, finally
press down the else keys to settle it.
Note: The focus “B” will move between the first focus and the third
focus.

3 th
3 Focus Point

- 21 -
Shenzhen emperor electronic technology Co., Ltd.

Press down this key when the image is unfrozen, the third focus point “C” is
selected for changing position along the scale by moving the trackball, finally
press down the else keys to settle it.
Note: The focus “C” will move between the second focus and the fourth
focus.

4 th
4 Focus Point

Press down this key when the image is unfrozen,, the fourth focus point “D” is
selected for changing position along the scale by moving the trackball, finally
press down the else keys to settle it.
Note: The focus “B” will move between the third focus and the bottom
of the image.

3.2.5 Display Mode

B
B-mode

Single-B display mode. Press down this key, then the screen goes back to

single-B display mode.

B B
B+B-mode

The double B-mode key selects dual images on the screen, if the screen is

already in dual image mode, press down the same key is used to select the live

image at the left side or at the right side on the screen.

B M
B+M-mode

__ Both B and M images will be displayed on the screen when this key is
pressed. Both images are real-time.
__ In B+M-mode, the same key is used to change the scanning speed of M-mode.
Continuously pressing down the same key to switch the scanning speed in
seven levels.

- 22 -
Shenzhen emperor electronic technology Co., Ltd.

M
M-mode

__ Only M image displayed on the screen when you press it.

__ In M-mode, the same key is used to change the scanning speed of M-mode.

Continuously pressing down the same key to switch the scanning speed in

seven levels.

3.2.6 Function keys

Distance

Used for distance measurement.

Circ & Area

Used for area, perimeter measurement.

Volume

Used for volume measurement.

Heart Rate

Used for heart rate, interval measurement in M-mode.

EF slope

Used for EF slope measurement in M-mode.

Start/Stop

To start or stop one measurement.

- 23 -
Shenzhen emperor electronic technology Co., Ltd.

Image Magnification

__ In real display, image magnifications could be four (×1.0,×1.2,×1.5, ×


2.0) levels when system is in B, B+B and B+M mode.
__ Only one image is enlarged and shown on the screen by pressing the key
when the system is in B+B mode. You have to press down the key B+B-mode
to another image enlarged and shown on.

Image Resume

__ Resume the amplificatory images by pressing the key from ×2.0,×1.5, ×


1.2,×1.0 time.
__ Same operation as the key IMAGE MAGNIFICATION described above.

Depth Scroll down

Press it to have the amplificatory image scroll down.

Depth Scroll up

Press it to have the amplificatory image scroll up.

Left/Right

Press down this key to conversion the image left/right.

Up/Down

Press down this key to conversion the image up/down.

Black/White

Press down this key you can reverse the image black/white.

Local Zoom

Refer the detail in the instruction of Local Zoom and Picture In Picture.

- 24 -
Shenzhen emperor electronic technology Co., Ltd.

Save

Refer the detail in the instruction of Image store, load and delete.

ABC
Clear

Press down this key you can clear text annotation on the screen. Some fixed

characters could not be cleared by pressing this key upon they were input, such as

Patient ID, Doctor name, Hospital name etc.

Freeze

Press down this key enables freeze or unfreeze the image.

Cine-memory

__ System will start to replay images after pressing the key.

__ Use the trackball to scroll forward or backward through cine-memory.

Quickly move the trackball either to left or right to start auto-replay.

- 25 -
Shenzhen emperor electronic technology Co., Ltd.

D Operation Instruction

4.1 General settings

4.1.1 Select the probe and the probe frequency


When the device is connected with multiple probes, press down the key PROBE, a menu will be displayed

at right side of the screen.

Probe 1:
1 ---- C – 2.5M
2 ---- C – 3.5M
3 ---- C – 5.0M
Probe 2:
4 ……
5 ……
6 ……
Esc…Exit

1) Press alphanumeric key 1,2……or 6 to choose the probe or the frequency of the probe(s) you have

connected.

2) If there is no probe connected, it will display “No Probe Found” when the device is switched on.

3) Press down the key ESC to exit this program.

4.1.2 Local Zoom and Picture In picture

In single-B mode and ×1.0, press down the key local zoom,a menu will display on the screen:

1——Local Zoom
2——Picture In Picture

1)— press down Alphanumeric “1”,enter into the local zoom.

— If the screen is already under local zoom, the key LOCAL ZOOM is used to select change the

zoom level, continuously pressing down it to switch in three levels(×1.2,×1.5,×2.0).

— The operation should follow the instructions on the screen.

2)— press down the key Alphanumeric “2”,enter into the picture in picture.

- 26 -
Shenzhen emperor electronic technology Co., Ltd.

— If the screen is already under picture in picture, the key LOCAL ZOOM is used to select

change the zoom level, continuously pressing down it to switch in three levels(×1.2,×1.5,

×2.0).

— The operation should follow the instructions on the screen.

4.1.3 Image Store, Load & Delete

Press down the key FREEZE, then press down the key STORE, the screen will displayed as following:

S——Store
L——Load
D——Del

1) How to store the image?

— Press down the key Alphanumeric “S” to have a substrate menu appeared on the screen.

S——Store 1 – Save to host


L——Load 2 – Save to U Disk
D——Del

— Select alphanumeric key “1” to save the image into the host, select alphanumeric key

“2” to save the image into U-disk.

— The operation should follow the instructions on the screen.

NOTE:If the user have not input the file name ,but press down the key ENTER

directly, the system will produce a file name automatically: patient name +

ID.

2) How to load the image?

— Press down the key Alphanumeric “L” to have a substrate menu appeared on the screen.

S——Store
L——Load
1 – Load from host
D——Del
2 – Load from U Disk

— Select alphanumeric key “1” to load the image from the host, select alphanumeric key

“2” to load the image from U-disk.

- 27 -
Shenzhen emperor electronic technology Co., Ltd.

— The operation should follow the instructions on the screen.

NOTE:The user could take basic operations on the loaded image, such as measurement,

etc. The stored image could be transferred between host and U-disk.

3) How to delete the image?

— Press down the key alphanumeric “D” to have a substrate menu appeared on the screen.

S——Store
L——Load 1 – Delete from host
D——Del 2 – Delete from U Disk

— Select alphanumeric key “1” to delete the image from the host, select alphanumeric key

“2” to delete the image from U disk.

— The operation should follow the instructions on the screen.

NOTE: There are three ways of selecting the various functions or items in each menu and/or

submenu list.

1. Select the item that you prefer by scrolling the trackball up and down or press the

direction key up and down, then press down the key ENTER;

Or 2. Select the item that you prefer by scrolling the trackball up and down or press

the direction key up and down, then press down the direction key Right to enter, or

press down the direction key LEFT to return to previous menu;

Or 3. Input the first character(number or letter) in front of each item that you need.

4.1.4 Image Cine-memory

__ Press down the key FREEZE to frozen the image.

__ Press it when the image is frozen, the system will realize cine recall.

__ Use the trackball to scroll forward or backward through cine-memory. Quickly move the

trackball either to left or right to start auto-replay.

__ During auto-replay, press trackball could pause replay to get the image wanted and do some

measurements.

- 28 -
Shenzhen emperor electronic technology Co., Ltd.

4.2 Various Measurements

4.2.1 Angle measurement

1)Press down the key FUNCTION and then activate alphanumeric key “M”, the image will be frozen

automatically and you will start angle measurement.

The screen will show:

“ Dist:

Dist:

Angle:”

2)Move the cursor with trackball to the origination position where the angle should be measured.

3)Press down the key START/STOP, then the cursor will disappear, but it will appear again in case if you

move trackball or press down the DIRECTION KEYS to start measurement. Move the cursor to the final

position and press down the key START/STOP, do it again to finish the measurement. The screen will

show:

“ Dist: ***.*mm

Dist: ***.*mm

Angle: ***.*°”

4)Exit the operation.

NOTE:1)The type of the cursors are different during each measurement.

2)During measurements,pressing down key START/STOP to confirm the position of first

cursor, and the position could be changed by moving the trackball and pressing

the key DISTANCE.

3)The result of last four measurements will be separately displayed on the right

side of the screen.

4.2.2 Distance Measurement

1. Press down the key DISTANCE to conduct distance measurement. The screen display “Dist”;

- 29 -
Shenzhen emperor electronic technology Co., Ltd.

2. Press down the corresponding directional keys or scroll trackball to move the cursor “+” to the

origination position of the measurement;

3. Press down the key START/STOP and the cursor “+” disappears that the system start the measurement.

Press down the corresponding direction key or scroll Trackball, and mark “+” reappears and move the

cursor to the desired stopping point. Press down the key START/STOP to end the measurement. The

result will real-time show on the screen:

“Dist=***.*mm”

As the above display shows, move the cursor from point A to B, the unit will display the result after

systematic calculation.

4. Exit the operation.

NOTE:1)The type of the cursors are different during each measurement.

2)During measurements,pressing down the key START/STOP to confirm the position

of first cursor, and the position could be changed by moving the trackball and

pressing the key DISTANCE.

3)The result of last four measurements will be separately displayed on the right

side of the screen.

4.2.3 Area and Perimeter Measurement

1) Press down the key perimeter and area measurement and indicator lights to start area measurement,

The screen prompts:

“Area:

Length: ”;

2) Press down the corresponding directional keys or scroll trackball to move the cursor “+” to the

- 30 -
Shenzhen emperor electronic technology Co., Ltd.

starting point of the measurement;

3) Press down the key START/STOP, the cursor disappears and the system starts the measurement. Then

press down the corresponding direction keys or scroll Trackball, the cursor “+” reappears, move the

cursor, the screen displays measured track in desired direction. When the track closes, press down the

key START/STOP to end the measurement. It displays perimeter and area results covered by the closed

curve.

Area: ***.* mm2

Length: ***.* mm
+
“Area” refers to the area (mm2);

“Length” refers to the perimeter (mm);

Refer to the above display, follow the steps previously mentioned track the cursor along the edge of

fetal’s transvaginal, you will have the perimeter and the area encircled by the curve.

4) Exit the operation.

NOTE:1)The type of the cursors are different during each measurement.

2)The result of last four measurements will be separately displayed on the right

side of the screen.

4.2.4 Volume measurement

NOTE:Please choice the method of the volume measurement before you beginning to volume

measure. There have two methods three axis and single plane.

Single plane
1. Press down the key VOLUME and indicator lights to start measures. The screen prompts:

LA: SA:
Volume:
Area:

Per(perimeter):

2. Press down the corresponding directional keys or scroll trackball to move the cursor “+” to the

starting point of the measurement;

- 31 -
Shenzhen emperor electronic technology Co., Ltd.

3. Press down the key START/STOP, the cursor disappears and the system starts measurement. a ellipse

will display on the screen, and then scroll trackball to Change the size, position, shape of ellipse , and

then Press the key START/STOP again, the values of area/perimeter, volume will be displayed

automatically on the screen.


LA: ***.* mm SA: ***.* mm

Volume: ***.* cm3 C

Area: ***.* mm2 + A B

Per(Perimeter): ***.* mm D

“LA” Axes is long axes (mm), “SA” Axes is short axes (mm), Volume: ***.* cm3 is volume value,

Area ***.*mm2 is area value, Per: ***.*mm is the perimeter.

4. Exit the operation.

NOTE:1)The type of the cursors are different during each measurement.

2)The result of last four measurements will be separately displayed on the right

side of the screen.

Three axis

NOTE:Choice the three axis, the volume data will display after the X axis、Y axis and

Z axis be measured.

Press down the key VOLUME, the following menu will display on the screen:

1——X axis
2——Y axis
3——Z axis

Choice the X axis, the measure method is allowing to the distance method, and then measure Y axis

and the Z axis.

4.3 Fetal age, Weight and EDCB Measurement

Following measurements can only used to get the probable growth state of fetus in a certain period, so

error will be generated if fetus is malformation or abnormality. All the parameters of following

- 32 -
Shenzhen emperor electronic technology Co., Ltd.

measurements should be calculated together to get a correct growth state of fetus as a whole. It will be

helpful to reduce error to make EDCB more correct.

The parameters of crown-rump, head circumference, abdomen circumference and femur length are usually

used in mid and late pregnancy. Sum up all the parameters and then divide the result by the number of

measurement, then you will get the fetus week. Refer to《Family Planning Ultrasound Diagnostics, Second

edition》

GS (Predict fetus week and EDCB according to the size of womb, applicable within 7 weeks)

CRL (Predict fetus week and EDCB according to the crown-rump, applicable within 8-12 weeks. Error

will be 3~5 days. Refer to《Family Planning Ultrasound Diagnostics, Second edition》)

BPD (Predict fetus week and EDCB according to the bi-parietal diameter, applicable in 12 fetus weeks

to one month old. Fetus week<20 weeks,Error will be 1~1.5week;Fetus week is 20~30week,Error

will be1.5~2 weeks;Fetus week>30 weeks,Error will be 3~3.5weeks. Refer to《Family Planning

Ultrasound Diagnostics, Second edition》)

HC (Predict fetus week and EDCB according to head circumference, applicable in 12 fetus weeks to one

month old, and this parameter can be used to measure fetus weight. Fetus week<20 weeks,error will be

1~1.5 week;fetus week is 20~30 weeks,error will be1.5~2 weeks;Fetus week>30weeks,error

will be 3~3.5 weeks. In the last 6 weeks, the result of HC will be more reliable than the result of BPD.

Refer to《Family Planning Ultrasound Diagnostics, Second edition》)

AC (Predict fetus week and EDCB according to abdomen circumference, applicable within36~42

weeks, and this parameter can be used to measure fetus weight. AC is made up the shortage of PD or HC.

Error will be 3~3.5 weeks. Refer to《Family Planning Ultrasound Diagnostics, Second edition》)

FL (Predict fetus week and EDCB according to femur length, applicable within 14~22 weeks, and this

parameter can be used to measure fetus weight. Error will be ±6.7days. Refer to《Family Planning

Ultrasound Diagnostics, Second edition》)

HL (Predict fetus week and EDCB according to humeral length.)

- 33 -
Shenzhen emperor electronic technology Co., Ltd.

Remark 1: In following measurements, the unit will automatically display the fetal age and EDCB and

weight after systematic calculation. The parameters are displayed in the following forms:

Fetal age GW=xW(weeks)+xD(days); Fetal Weight WT=xkg;

Expected Date of Child Birth =EDCB xxxxxx(I.e. EDCB: 010902 means Expected Date of

Child Birth is September 1,2002)

2: In following measurements are as same as the distance measurement.

1. Press down the key FUNCTION and then alphanumeric key “Z”, the lower part of the screen displays:

GS
GS=000.0mm

Following to the distance measurement to measure “GS” , the system will automatically display fetal

age and EDCB after systematic calculation.

GS
BL
GS
YS

UT. UT.
Gs Measurement Illustration
Left: Longitudinal Section of UT for measurement of GS’s Longitudinal Radius
Right: Transverse Section of UT for measurement of lateral Radius

Note:The Bladder is required full. Measure the compete fetal bladder with maximum lateral

and longitudinal radiuses.

2. Press down the key FUNCTION and then alphanumeric key “X”, the lower part of the screen displays:

CRL
CRL=000.0mm

Rotate or tilt the probe to have better image of the fetal in the womb and ensure that the max diameter

of crown-rump is displayed and then freeze the picture. Following to the distance measurement to

measure “CRL”, the screen will automatically display fetal age and EDCB after systematic calculation.

- 34 -
Shenzhen emperor electronic technology Co., Ltd.

AMN

FH FT

CRL Measurement Illustration

3. Press down the key FUNCTION and then alphanumeric key “C”, the lower part of the screen displays:

BPD

BPD=000.0mm

From top to bottom locate the standard section in which fetal head is an ellipse, centerline of the brain

centers and the two sides symmetrical. Take the maximum diameter vertical to centerline of the brain

from echo external edge of nearer skull to echo interior edge of farer one for BPD measurement with the

method as same as distance measurement. The unit will automatically display fetal age and EDCB after

systematic calculation.

BPD Measurement Illustration

Note: Choose a standard section for measurement, if the section selected is too high, the

value is too low; if the section selected is too low, the value is rather low.

4. Press down the key FUNCTION and then alphanumeric key “V” , the lower part of the screen displays:

HC

SA=000.0mm LA=000.0mm

SA refers to short axis of fetal head; LA refers to long axis of fetal head (Unit: mm).

- 35 -
Shenzhen emperor electronic technology Co., Ltd.

Measure the maximum front-rear (LA) and horizontal diameter (SA) with standard section of frozen

images. The system will automatically display fetal age, fetal weight and EDCB after systematic

calculation as per Head circumference=(SA+LA)×1.57, The unit will automatically display fetal age

and EDCB after systematic calculation.

Front-rear length
Lateral Radius

Schematic Diagram for Head Circumference (HC ) Measurement

5. Press down the key FUNCTION and alphanumeric key “B”, lower part of the screen displays:

AC

D1=000.0mm D2=000.0mm

D1 and D2 respectively refer to front-rear diameter and horizontal diameter of the cross-sectioned

abdomen, (Unit: mm).

Measure the maximum front-rear (LA) and horizontal diameter (SA) with standard section of frozen

images. Take the distance from interior edge of the abdomen to the opposite. The system will

automatically display fetal age, fetal weight and EDCB after systematic calculation as per abdomen

circular=(D1+D2)×1.57, The unit will automatically display fetal age and EDCB after systematic

calculation.

Liver
Spine

Stomach

AC Measurement Illustration

6. Press down the key FUNCTION and then alphanumeric key “N”, lower part of the screen:

FL

FL=000.0mm

- 36 -
Shenzhen emperor electronic technology Co., Ltd.

FL refers to the length between the two ends of femur (Unit: mm).

Take the length between the two ends of femur and follow the method described as same as distance

measurement. The system will automatically display fetal age and EDCB after systematic calculation.

FL

FL Measurement Illustration

Note: 1) The femur is displayed very clearly and avoid slope measurement of the femur.

2) If the section selected is too high, the value is too low; if the section selected

is too low, the value is rather low.

7. Press down the key FUNCTION and then alphanumeric key “H”, lower part of the screen:

HL

HL=000.0mm

HL refers to the length between the two ends of humeral (Unit: mm).

Take the length between the two ends of humeral and follow the method described as same as distance

measurement. The system will automatically display fetal age and EDCB after systematic calculation.

Remark: 1. Press down the key CLEAR/FREEZE/ESC to exit the measurement;

2. Measuring methods could be switched freely in your measurement. Select the method you
wanted to start measurement following steps.

3. Each measurement has value range, it will display”!!! INVALID VALUE!!!” if the result
exceeds the limitation, it will return to the previous step. The user has the choice to select
measurement mode or exit.

4. Fetal Weight is applicable to the over-20-week fetal, and the unit will display the result after
systematic judgment.

5. The data will be stored in the Obstetrics’ report.

6. Please refer to the appendix I.

- 37 -
Shenzhen emperor electronic technology Co., Ltd.

4.4 M-Mode Measurement

Heart rate, Time and EF slope measurement

1) Press down the key HEART RATE, and indicator on to start area measurement, the screen prompts:

displays:

“Heart Rate

Time” ;

2) Move the cursor “+” to the starting point of measurement by pressing directional keys or scrolling

trackball with frozen image;

3) Press down the key START/STOP, the cursor “+” will disappear. Press down the corresponding

direction keys or scroll Trackball, “+” reappears. Select another measurement point as desired. The

measurement result will be dynamically displayed on the screen. In the diagram, horizontal distance

refers to time and vertical one to the amplitude. Measured HR and EF will be automatically displayed

after systematic calculation (HR: b/m, EF: m/s);

Display in M-mode.

Heart -rate can be obtained by


++
measuring peak- peak distance
+ +

+
EF can be obtained by measuring the height
& width of peak & bottom values of the wave

4) Press down the key START/STOP again to end the measurement;

NOTE:1)The type of the cursors are different during each measurement.

2)The result of last four measurements will be separately displayed on the right

side of the screen.

- 38 -
Shenzhen emperor electronic technology Co.,Ltd

E Settings and Function


1 Focus
2 Focus
3 Focus
4 Focus

T——Time Setting On
H——Hospital Name Off
1——Settings Ö D——Doctor Name
F——Focus Number Ö On
2——Function Ö
B——Biopsy Line Ö Off
3——Obstetrics Ö
4——Gynecology Ö C——Center Line Ö
On
5——Urology Ö N——Measure Line Ö
Off
6——Cardiology Ö M——Body Mark Ö
V——Vol. Method Ö
On
(Main menu) R——Factory Setting
Off
Select Body Mark

Three axis
P——Print Single Plane
T——Text
R——Ratio Measure Suspend Mode
H——Histogram 1——10 minutes later
G——Gamma Correction 2——20 minutes later
S——Suspend Mode 3——30 minutes later
V——Version 4——60 minutes later
U——Software Update 5——No Suspend Mode
Press <ESC> to exit

- 39 -
Shenzhen emperor electronic technology Co.,Ltd

5.1 Settings
Note: There are three ways of selecting the various functions or items in each menu and/or

submenu list.

1. Select the item that you prefer by scrolling the trackball up and down or press the

direction key up and down, then press down the key ENTER;

Or 2. Select the item that you prefer by scrolling the trackball up and down or press the

direction key up and down, then press down the direction key Right to enter, or press

down the direction key LEFT to return to previous menu;

Or 3. Input the first character(number or letter) in front of each item that you need.

NOTE: Press down the key Esc to exit all the menus that you enter.

Return to the main menu and select Settings, upon pressing the key ENTER, following menu will display

on the screen.

1——Settings Ö
2——Function Ö
T——Time Setting
3——Obstetrics Ö
H——Hospital Name
4——Gynecology Ö
D——Doctor Name
5——Urology Ö
F——Focus Number Ö
6——Cardiology Ö
B——Biopsy Line Ö
C——Center Line Ö
N——Measure Line Ö
M——Body Mark Ö
V——Vol. Method Ö
R——Factory Setting

Time Setting

Select Time Setting, and press the key ENTER.

The date, week and time display in the right top corner of the screen.

The cursor appears on the screen, you can input the new date, week and time.

04/08/08 Thursday day/month/year week


08:08:08 hour : minute : second

E.g. In this case, if you input 010105 6 181818, and then press the key ENTER, the screen will be changed

as follows, the updated data are stored in memory:

- 40 -
Shenzhen emperor electronic technology Co.,Ltd.

01/01/05 Saturday
18:18:18
The key ENTER is used to finish your setting, but the key ESC is used to escape in midway.

Note: Numerical keys 1 stands for Monday, 2 for Tuesday ……7 for Sunday;

Time is preset in 24-hour format.

Hospital Name
Select Hospital Name, and then press the key ENTER.
Edit the hospital name and press the key ENTER, the result that you have edited will be automatically
stored and appear on the right bottom corner of the screen.

Doctor Name
Select Doctor Name, and then press the key ENTER.
Edit the doctor name and press the key ENTER, the result that you have edited will be automatically stored
and appear on the right bottom corner of the screen.

Focus Number
Select Focus Number, and then press down the key ENTER.

1——Settings Ö
2——Function Ö
T——Time Setting
3——Obstetrics Ö
H——Hospital Name
4——Gynecology Ö
D——Doctor Name
5——Urology Ö
F——Focus Number Ö
6——Cardiology Ö
B——Biopsy Line Ö
1 focus
C——Center Line Ö
2 focus
N——Measure Line Ö
3 focus
M——Body Mark Ö
4 focus
V——Vol. Method Ö
R——Factory Setting
Steps: Select focus number firstly, and then press down corresponding focus key, scroll trackball to

change the position of focus and press down key ENTER.

Example: How to select two focus


st
In the above-mentioned menu, Select 2 focus will generate A, B focus. Pressing 1 Focus Point, the “A”
will turn to be “a”, and then move the trackball to change the position of the “a” focus. Pressing the else
st
keys to fix the focus position, meanwhile “a” turn to be “A”. Secondly, press down 2 Focus Point, the

- 41 -
Shenzhen emperor electronic technology Co.,Ltd.

“B” will turn to be “b”, and then move the trackball to change the position of the “b” focus. Pressing the
else keys to fix the focus position, meanwhile “b” turn to be “B”.

NOTE:In M mode and B/M mode, not more than two focuses could be selected.

Biopsy Line

Select Biopsy Line, and then press down the key ENTER.

1——Settings Ö
2——Function Ö
T——Time Setting
3——Obstetrics Ö
H——Hospital Name
4——Gynecology Ö
D——Doctor Name
5——Urology Ö
F——Focus Number Ö
6——Cardiology Ö
B——Biopsy Line Ö
C——Center Line Ö
On
N——Measure Line Ö
Off
M——Body Mark Ö
V——Vol. Method Ö
R——Factory Setting

Select On to display biopsy line.

Select Off to hide biopsy line.

NOTE:1) Biopsy line will not display except when convex probe in connected in B mode.

2) Biopsy line should be calibrated carefully before using. Please follow the

instruction on the screen to adjust the position and direction of the biopsy line.

Center Line

Select Center Line, and then press down the key ENTER.

1——Settings Ö
2——Function Ö
T——Time Setting
3——Obstetrics Ö
H——Hospital Name
4——Gynecology Ö
D——Doctor Name
5——Urology Ö
F——Focus Number Ö
6——Cardiology Ö
B——Biopsy Line Ö
C——Center Line Ö
N——Measure Line Ö
On
M——Body Mark Ö
Off
V——Vol. Method Ö
R——Factory Setting

Select On to display Center line.

- 42 -
Shenzhen emperor electronic technology Co.,Ltd.

Select Off to hide Center line.

NOTE: Center line will not display when linear probe and/or transvaginal probe connected.

Measure Line

Select Measure Line, and then press down the key ENTER.

1——Settings Ö
2——Function Ö
T——Time Setting
3——Obstetrics Ö
H——Hospital Name
4——Gynecology Ö
D——Doctor Name
5——Urology Ö
F——Focus Number Ö
6——Cardiology Ö
B——Biopsy Line Ö
C——Center Line Ö
N——Measure Line Ö
M——Body Mark Ö
On
V——Vol. Method Ö
Off
R——Factory Setting

Select On to display Measure line.

Select Off to hide Measure line.

Body Mark

Select Body Mark, and then press down the key ENTER.

1——Settings Ö
2——Function Ö
T——Time Setting
3——Obstetrics Ö
H——Hospital Name
4——Gynecology Ö
D——Doctor Name
5——Urology Ö
F——Focus Number Ö
6——Cardiology Ö
B——Biopsy Line Ö
C——Center Line Ö
N——Measure Line Ö
M——Body Mark Ö
V——Vol. Method Ö
On
R——Factory Setting
Off
Select Body Mark

Select On to display Body Mark.

Select Off to hide Body Mark (including 18 body marks on the keyboard).

Select Select body Mark, and then press the key ENTER, 20 body marks available will display on the

- 43 -
Shenzhen emperor electronic technology Co.,Ltd.

screen.

Body pregnant (right) Body pregnant (Left) Fetus (back) Fetus (back) Parasternal long axis

Papillary High short


Mitral valve level Apical four Apical two
muscle level axis chamber view chamber view

Subcostal four Parasternal Subparasternal Subparasternal


Head (left)
chamber view application parallel application parallel application transverse

Head (right) Head (face) Arm (right) Arm (Left) Leg (back)

Foetus Foetus Foetus Foetus Foetus

Foetus Body female Breast Neck(with chin up) Body male

- 44 -
Shenzhen emperor electronic technology Co.,Ltd.

Body female Body(right) Body(left) Back thyroid gland

Prostate(sliver) Uterus(sliver) ovary

Note:

1)Select the needed body mark by scrolling the trackball. Pressing the key ENTER

to confirm or the key ESC to exit.

2)In unfrozen and B/M mode, Press down or to change the direction of
arrow which is showing the scanning direction;In frozen and B/M mode,and the

bodymark is display on the screen,press direction keys to move the position of

arrow which is showing the scanning position.

Vol. Method

Select Vol. Method,then press down the key ENTER,the user can be choose the volume measurement

method, the three axis and single plane.

Factory Setting

Select Factory Setting,then press down the key ENTER, the setting which the user could be resetting will

resume the first setting.

5.2 Function

Return to the main menu and select Function. Upon pressing the key ENTER, following menu will display

on the screen.

- 45 -
Shenzhen emperor electronic technology Co.,Ltd.

1——Settings Ö
2——Function Ö
3——Obstetrics Ö
P——Print
4——Gynecology Ö
T——Text
5——Urology Ö
R——Ratio Measure
6——Cardiology Ö
H——Histogram
G——Gamma Correction
S——Suspend Mode
V——Version
U——Software Update

Print

Freeze the image, enter into the menu to select Print and confirm with the key ENTER.

Note: 1) Connected Laser printer after switched on the scanner for 10 minutes.

2) Please freeze the image before printing.

3) Please use the manufacturer defined model of laser printer.

Text

Select Text, and then press down the key ENTER, a cursor which is moveable with trackball will appear

on the screen. Text could be input or edited in different places within image area on the screen and then

press the key ENTER in the end.

Note: During input process, the key ENTER is used to changed to a new line, the key ESC

to finish the input and escape the operation, the key CLEAR to clear the test. And

the LEFT and RIGHT direction keys to move the cursor.

Ratio Measure

Select ratio measure, and then press down the key ENTER, the screen will show two- distance

measurement. According to the methods for distance measurements, you will get two distance values. The

ratio between the second distance and the first distance values will automatically appear on the screen.

Histogram

Select Histogram, the screen will display a square, press the alphanumeric key “H” to enlarge or reduce

the square, scroll trackball to change the position of the square, then press the key ENTER, the screen will

display a histogram on the image, and then press the key CLEAR to clear it.

Note: Press the key ESC if you want to exit midway.

- 46 -
Shenzhen emperor electronic technology Co.,Ltd.

Gamma Correction

Select Gamma Correction, then press down the key ENTER.

Eight correction curves are available, and pressing corresponding number to select different ones which

will be stored into the system automatically even switched off the system.

Suspend Mode

Select Suspend Mode, then press down the key ENTER. The following menu will display on the screen:

Suspend Mode
1——10 minutes later
2——20 minutes later
3——30 minutes later
4——60 minutes later
5——No Suspend Mode
Press <ESC> to exit

Example: If you select 1, in case the machine standbys for more than 10 minutes, it will be enter into the

suspend state. In this case, the probe and the machine will not work, and press any key to exit.

If you select 2, in case the machine standbys for more than 20 minutes, it will be enter into the suspend

state. In this case, the probe and the machine will not work, and press any key to exit; etc.

If you select 5 , it will enter into No Suspend mode.

Version

Select Version to display software version.

Software Update

Connected U-disk, enter into menu to select Software Update, press key ENTER to confirm. Follow the

instruction on the screen. Software update need a few minutes, any operation is forbidden to avoid

damage or malfunction.

Note:Please format U-disk before using it. In this course, please ensure the power supply

on level off.

- 47 -
Shenzhen emperor electronic technology Co.,Ltd

F Obstetrics

1——Settings Ö Z——GS Distance 1 Hellman Hadlock


2——Function Ö X——CRL Distance 2 Rempen Robinson
3——Obstetrics Ö C——BPD Distance 3 Tokyo University Hansmann
4——Gynecology Ö V——HC Distance 4 User Lasser
5——Urology Ö B——AC Tokyo University
6——Cardiology Ö N——FL Fetus 1 Hadlock User
H——HL Fetus 2 Merz
I——AFI Ö Lasser Hadlock
T——Twins Ö GS Ö Merz
Tokyo University
P——LMP Measure CRL Ö Lasser
User
O——Obstetrics Report BPD Ö Tokyo University
S——Select Methods Ö HC Ö User
Hadlock
U——User Methods Ö AC Ö
Merz Hadlock
FL Ö
Lasser Merz
HL Ö
Tokyo University Jeanty
EFW Ö
User Tokyo University
GS User
CRL
BPD Jeanty
1 Hadlock User
HC
2 Hadlock
AC
3 Hadlock
FL
4 Hadlock
HL
Shepard
Hansmann
Tokyo University
- 48 -
Shenzhen emperor electronic technology Co.,Ltd

Note: There are three ways of selecting the various functions or items in each menu and/or

submenu list.

1. Select the item that you prefer by scrolling the trackball up and down or press the

direction key up and down, then press down the key ENTER;

Or 2. Select the item that you prefer by scrolling the trackball up and down or press the

direction key up and down, then press down the direction key Right to enter, or press

down the direction key LEFT to return to previous menu;

or 3. Input the first character(number or letter) in front of each item that you need.

Note: Press down the key Esc to exit all the menus that you enter.

Return to the main menu and select Obstetrics. Upon pressing the key ENTER, following menu will

display on the screen.

1——Settings Ö
2——Function Ö
3——Obstetrics Ö
4——Gynecology Ö
Z——GS
5——Urology Ö
X——CRL
6——Cardiology Ö
C——BPD
V——HC
B——AC
N——FL
H——HL
I——AFI Ö
T——Twins Ö
P——LMP Measure
O——Obstetrics Report
S——Select Methods Ö
U——User Methods Ö
Gestational Sac (GS)
Select GS, then press down the key ENTER, the screen will display as following:

GS

GS=000.0mm

Press down the corresponding directional keys or scroll trackball as per your requirement to move the

cursor to the starting point of each measurement. When measure “GS”, The system will automatically

display fetal age and EDCB after systematic calculation, at the same time the measurement data will be

stored in the report. For the same patient, measure “GS” again, the newest measurement data display on

- 49 -
Shenzhen emperor electronic technology Co.,Ltd.

the report.

Crown-Rump Length (CRL)

Select CRL, then press down the key ENTER, the screen will display as following:

CRL

CRL=000.0mm

Rotate or tilt the probe to have better image of the fetal in the womb and ensure that the max diameter of

crown-rump is displayed and then freeze the picture. Measure “CRL” in proper order, the screen will

automatically display fetal age and EDCB after systematic calculation, at the same time the measurement

data will be stored in the report. For the same patient, measure “CRL” again, the newest measurement

data display on the report.

Biparietal Diameter (BPD)

Select BPD, then press down the key ENTER, the screen will display as following:

BPD

BPD=000.0mm

From top to bottom locate the standard section in which fetal head is an ellipse, centerline of the brain

centers and the two sides symmetrical. Take the maximum diameter vertical to centerline of the brain from

echo external edge of nearer skull to echo interior edge of farer one for “BPD” measurement with the

method described in Remark 2. The unit will automatically display fetal age and EDCB after systematic

calculation, at the same time the measurement data will be stored in the report. For the same patient,

measure “BPD” again, the newest measurement data display on the report.

Head Circumference (HC)

Select HC, then press down the key ENTER, the screen will display as following:

HC

SA=000.0mm LA=000.0mm

SA refers to short axis of fetal head; LA refers to long axis of fetal head (Unit: mm).

Measure the maximum front-rear (LA) and horizontal diameter (SA) with standard section of frozen

images. The system will automatically display fetal age, fetal weight and EDCB after systematic

- 50 -
Shenzhen emperor electronic technology Co.,Ltd.

calculation, at the same time the measurement data will be stored in the report. For the same patient,

measure “HC” again, the newest measurement data display on the report.

Abdominal Circumference (AC)

Select AC, then press down the key ENTER, the screen will display as following:

AC

D1=000.0mm D2=000.0mm

D1 and D2 respectively refer to front-rear diameter and horizontal diameter of the cross-sectioned

abdomen, (Unit: mm).

Measure the maximum front-rear (LA) and horizontal diameter (SA) with standard section of frozen

images. Take the distance from interior edge of the abdomen to the opposite. The system will

automatically display fetal age, fetal weight and EDCB after systematic calculation, at the same time the

measurement data will be stored in the report. For the same patient, measure “AC” again, the newest

measurement data display on the report.

Femur Length (FL)

Select FL, then press down the key ENTER, the screen will display as following:

FL

FL=000.0mm

FL refers to the length between the two ends of femur (Unit: mm).

Take the length between the two ends of femur and follow the method described in Remark 2. The system

will automatically display fetal age, fetal weight and EDCB after systematic calculation, at the same time

the measurement data will be stored in the report. For the same patient, measure “FL” again, the newest

measurement data display on the report.

Humeral Length (HL)

Select HL, then press down the key ENTER, the screen will display as following:

HL

HL=000.0mm

HL refers to the length between the two ends of humeral (Unit: mm).

- 51 -
Shenzhen emperor electronic technology Co.,Ltd.

Take the length between the two ends of humeral and follow the method described in Remark 2. The

system will automatically display fetal age, fetal weight and EDCB after systematic calculation, at the

same time the measurement data will be stored in the report. For the same patient, measure “HL” again,

the newest measurement data display on the report.

Amniotic Fluid Index (AFI)

Select AFI, then press down the key ENTER, the screen will display as following:

1——Settings Ö
2——Function Ö
3——Obstetrics Ö
4——Gynecology Ö
Z——GS
5——Urology Ö
X——CRL
6——Cardiology Ö
C——BPD
V——HC
B——AC
N——FL
H——HL
I——AFI Ö
T——Twins Ö
Distance 1
P——LMP Measure
Distance 2
O——Obstetrics Report
Distance 3
S——Select Methods Ö
Distance 4
U——User Methods Ö

Note: The measurement method of AFI refers to the distance measure. Select Distance 1 for

the first quadrant measurement. A distance measurement is performed in order to get

the first quadrant measurement. Repeat these measurements for the second to the

fourth quadrant.

Twins

Select Twins, then press down the key ENTER, the screen will display as following:

- 52 -
Shenzhen emperor electronic technology Co.,Ltd.

1——Settings Ö
2——Function Ö
3——Obstetrics Ö
4——Gynecology Ö
Z——GS
5——Urology Ö
X——CRL
6——Cardiology Ö
C——BPD
V——HC
B——AC
N——FL
H——HL
I——AFI Ö
T——Twins Ö
P——LMP Measure
Fetus 1
O——Obstetrics Report
Fetus 2
S——Select Methods Ö
U——User Methods Ö

Select fetus1, the data will be saved as one record in the report of the first fetus.

Select fetus2, the data will be saved as one record in the report of the second fetus.

Note: If you select fetus2 in the twins’ option, the report will display two fetus’ data;

otherwise, the report will display only one fetus’ data.

LMP Measurement

Select LMP Measurement, and then press down the key ENTER, in the right top corner of screen will

display as following:

LMP: 01D 01M 01Y


EDCB: 01D 01M 01Y

After inputting the last menstrual period (LMP), the Expected Date of Child Birth (EDCB) will

automatically display on the screen. The data will not be saved in the report.

- 53 -
Shenzhen emperor electronic technology Co.,Ltd.

Obstetrics Report

Hospital Name

Patient: Section: Date:

GS:

CRL:

BPD:

HC:

AC:

FL:

HL:

Average Fetal Age:

LMP:

AFI:

HC/AC Ratio: 89%-131%for12-40weeks

FL/AC Ratio: 20%-24%for21-40weeks

FL/BPD Ratio: 71%-87%for23-40weeks

Fetal weight:

Doctor:

Select Obstetrics Report, and then press the key ENTER, the report will be display on the screen:

Note: The content of the report couldn’t be changed or erased by user. If you selected

fetus2 in the twins’ option, the report will display as report2; otherwise, the report

will display as report1.

Note:

1. Once you choose each software packages, all the measurement data will be saved in the

report.

2. For the same patient, the data in the report will be updated if you carry out the

measurement upon the same item again, and the system will show the newest data as well.

3. When you press down the key New Patient, the data that saved in the report for the

previous patient will be removed.

4. When connect the laser printer, Press down the alphanumeric key P to print the report.

- 54 -
Shenzhen emperor electronic technology Co.,Ltd.

Select Methods

Select Select Methods, upon pressing the key ENTER, following menu will display on the screen.

1——Settings Ö
2——Function Ö
3——Obstetrics Ö
4——Gynecology Ö
Z——GS
5——Urology Ö
X——CRL
6——Cardiology Ö
C——BPD
V——HC
B——AC
N——FL
H——HL
I——AFI Ö
T——Twins Ö
P——LMP Measure
O——Obstetrics Report
S——Select Methods Ö
U——User Methods Ö
GS Ö
CRL Ö
BPD Ö
HC Ö
AC Ö
FL Ö
HL Ö
EFW Ö

There are various formulas available for different measurements.

For example, for GS measurement, there are four formulas: Hellman, Rempen, Tokyo University and

User.

GS Ö
CRL Ö
Hellman
BPD Ö
Rempen
HC Ö
Tokyo University
AC Ö
User
FL Ö
HL Ö
EFW Ö

Note: 1. CRL, BPD, HC, AC, FL and HLhave similar formulas for users’ option. The default

setting for all the measurements is Tokyo University formula.

- 55 -
Shenzhen emperor electronic technology Co.,Ltd.

2.The last formula User has to be defined in User Methods in advance.

User Methods

Select User Methods, upon pressing the key ENTER, following menu will display on the screen.

Z——GS
X——CRL
C——BPD
V——HC
B——AC
N——FL GS
H——HL CRL
I——AFI BPD
Ö
T——Twins HC
Ö
P——LMP Measurement AC
O——Obstetrics Report FL
S——Select Methods HL
Ö
U——User Methods Ö

Users could define their own formula for the seven OB measurements. The setting will be stored into the

User formula in Select Method.

For example: GS measurement

Pressing down the key ENTER to get following menu:

User defined table GS User defined table GS


Please input fetal age: 009 days continue?
Please input data: 026.0 mm
Press<enter>to confirm Press<Esc> to exit Press<Y>to continue Press<enter> to exit
009/026.0

Input the fetus age and date. Press ENTER to confirm and press ESC to exit.

Press Y to continue edit, but press ENTER to finish you setting.

User’s setting will be saved into the User formula which is selectable in Select Method.

- 56 -
Shenzhen emperor electronic technology Co.,Ltd

G Gynecology

1——Settings Ö U——Uterus Ö Length


2——Function Ö L——Left Ovary Ö AP Diameter
3——Obstetrics Ö R——Right Ovary Ö Width
4——Gynecology Ö E——Endometrial Thickness
5——Urology Ö G——Gynecology Report
6——Cardiology Ö Length
AP Diameter
Width

Length
AP Diameter
Width

- 57 -
Shenzhen emperor electronic technology Co.,Ltd

Note: There are three ways of selecting the various functions or items in each menu and/or
submenu list.

1. Select the item that you prefer by scrolling the trackball up and down or press the

direction key up and down, then press down the key ENTER;

Or 2. Select the item that you prefer by scrolling the trackball up and down or press the

direction key up and down, then press down the direction key Right to enter, or press

down the direction key LEFT to return to previous menu;

Or 3. Input the first character(number or letter) in front of each item that you need.

Note: Press down the key Esc to exit all the menus that you enter.

Return to the main menu and select Gynecology. Upon pressing the key ENTER, following menu will

display on the screen.

1——Settings Ö
2——Function Ö
3——Obstetrics Ö
4——Gynecology Ö
5——Urology Ö
U——Uterus Ö
6——Cardiology Ö
L——Left Ovary Ö
R——Right Ovary Ö
E——Endometrial Thickness
G——Gynecology Report

Uterus

Selecting Uterus will generate a submenu, from which three measurements for the uterus can be

performed. Press ENTER and the following menu will appear on the screen.

Length
AP Diameter
Width

The three measurements are:

a) Length

The uterus is displayed in the sagittal plane, The length is measured from the fundus to the cervix of the

Uterus.

b) AP Diameter

- 58 -
Shenzhen emperor electronic technology Co.,Ltd.

The uterus is displayed in the sagittal plane. With this measurement the AP diameter of the corpus of the

Uterus is measured.

c) Width

The fundus of the uterus is displayed in the transverse plane, With this measurement the width of the

fundus of the uterus is measured.

Note: Select one of the three measurements and the menu will disappear from the screen,

A caliper cross will be displayed in the image and you can now perform the

measurement.

Left Ovary

Selecting Left Ovary will generate a submenu, from which three measurements for the uterus can be

performed. Press ENTER and the following menu will appear on the screen.

Length
AP Diameter
Width

The three measurements are:

a) Length

The length of the left ovary is measured with the ovary displayed in the sagittal plane.

b) AP Diameter

The AP diameter of the left ovary is measured with the ovary displayed in the sagittal plane of the

transverse plane.

c) Width

The width of the left ovary is measured with the ovary displayed in the transverse plane.

Note: Select one of the three measurements and the menu will disappear from the screen,

a caliper cross will be displayed in the image and you can now perform the

measurement.

Right Ovary

Selecting Right Ovary will generate a submenu, from which three measurements for the

uterus can be performed. Press ENTER and the following menu will appear on the screen.

The three measurements are:

- 59 -
Shenzhen emperor electronic technology Co.,Ltd.

Length
AP Diameter
Width

a) Length

The length of the right ovary is measured with the ovary displayed in the sagittal plane.

b) AP Diameter

The AP diameter of the right ovary is measured with the ovary displayed in the sagittal plane of the

transverse plane.

c) Width

The width of the right ovary is measured with the ovary displayed in the transverse plane.

Note: Select one of the three measurements and the menu will disappear from the screen,

a caliper cross will be displayed in the image and you can now perform the

measurement.

Endometrial Thickness

Select Endometrial Thickness and the menu will disappear in the screen, a caliper cross will be

displayed in the image and you can now perform the measurement.

The Endometrial Thickness is measured at the height of the fundus of the uterus with the uterus

displayed in the sagittal or transverse plane.

Gynecology Report

Select Gynecology Report, then press down the key ENTER, t the report will displayed on the screen.

- 60 -
Shenzhen emperor electronic technology Co.,Ltd.

Hospital Name:
Patient: Section: Date:
Uterus Length:
Uterus AP Diameter:
Uterus Width:
left ovary Length:
left ovary AP Diameter:
left ovary Width:
Right ovary Length:
Right ovary AP Diameter:
Right ovary Width:
Endometrial Thickness:
Doctor:

Note:

1. Once you choose each software packages, all the measurement data will be saved in the

report.

2. For the same patient, the data in the report will be updated if you carry out the

measurement upon the same item again, and the system will show the newest data as well.

3. When you press down the key New Patient, the data that saved in the report for the

previous patient will be removed.

4. When connect the laser printer, Press down the alphanumeric key P to print the report.

- 61 -
Shenzhen emperor electronic technology Co.,Ltd

H Urology

1——Settings Ö B——Bladder Ö Transversal


2——Function Ö L——Left Kidney Sagittal
3——Obstetrics Ö R——Right Kidney Oblique 1
4——Gynecology Ö P——Prostate Ö Oblique 2
5——Urology Ö U——Urology Report
6——Cardiology Ö S——Set Bladder Coefficients
Transversal
Sagittal
Oblique 1
Oblique 2

- 62 -
Shenzhen emperor electronic technology Co.,Ltd

Note: There are three ways of selecting the various functions or items in each menu and/or

submenu list.

1. Select the item that you prefer by scrolling the trackball up and down or press the

direction key up and down, then press down the key ENTER;

Or 2. Select the item that you prefer by scrolling the trackball up and down or press the

direction key up and down, then press down the direction key Right to enter, or press

down the direction key LEFT to return to previous menu;

Or 3. Input the first character(number or letter) in front of each item that you need.

Note: Press down the key Esc to exit all the menus that you enter.

Return to the main menu and select urology. Upon pressing the key ENTER, following menu will display

on the screen.

1——Settings Ö
2——Function Ö
3——Obstetrics Ö
4——Gynecology Ö
5——Urology Ö
6——Cardiology Ö
B——Bladder Ö
L——Left Kidney
R——Right Kidney
P——Prostate Ö
U——Urology Report
S——Set Bladder Coefficients

Bladder

Selecting Bladder will generate a submenu,from which four measurements for the Bladder can be

performed,press ENTER and the following menu will appear on the screen.

Transversal
Sagittal
Oblique 1
Oblique 2

The four measurements are:

a) Transversal

Select Transversal and the menu will disappear from the screen, the screen will display as following:

Bladder Transversal

- 63 -
Shenzhen emperor electronic technology Co.,Ltd.

H= B=

Press down the corresponding directional keys or scroll trackball as per your requirement. The

measurement method of the height(H) and breadth(B) refer to the distance measurement. When measure

“Bladder Transversal”, the data will automatically display on the screen, at the same time the

measurement data will be saved in the report. For the same patient, measure “Bladder Transversal” again,

the newest measurement data display on the report.

b) Sagittal

Select sagittal and the menu will disappear in the screen, the screen will display as following:

Bladder sagittal

H= B=

Press down the corresponding directional keys or scroll trackball as per your requirement. The

measurement method of the height(H) and breadth(B) are order to the distance measurement. When

measure “Bladder sagittal”, the data will automatically display on the screen, at the same time the

measurement data will be stored in the report. For the same patient, measure “Bladder sagittal” again, the

newest measurement data display on the report.

c) Oblique 1

Select Oblique 1 and the menu will disappear from the screen, the screen will display as following:

Bladder Oblique 1

H= B=

Press down the corresponding directional keys or scroll trackball as per your requirement. The

measurement method of the height(H) and breadth(B) are order to the distance measurement. When

measure “Bladder Oblique 1”, the data will automatically display on the screen, at the same time the

measurement data will be stored in the report. For the same patient, measure “Bladder Oblique 1” again,

the newest measurement data display on the report.

d) Oblique 2

Select Oblique 2 and the menu will disappear from the screen, the screen will display as following:

Bladder Oblique 2

H= B=

Press down the corresponding directional keys or scroll trackball as per your requirement. The

measurement method of the height(H) and breadth(B) are order to the distance measurement. When

- 64 -
Shenzhen emperor electronic technology Co.,Ltd.

measure “Bladder Oblique 2”, the data will automatically display on the screen, at the same time the

measurement data will be stored in the report. For the same patient, measure “Bladder Oblique 2” again,

the newest measurement data display on the report.

Left Kidney

Select Left Kidney and the menu will disappear from the screen, the screen will display as following:

Left Kidney

L= H=

Press down the corresponding directional keys or scroll trackball as per your requirement. The

measurement method of the length(L) and height(H) are order to the distance measurement. When

measure “Left Kidney”, the data will automatically display on the screen, at the same time the

measurement data will be stored in the report. For the same patient, measure “Left Kidney” again, the

newest measurement data display on the report.

Right Kidney

Select Right Kidney and the menu will disappear from the screen, the screen will display as following:

Right Kidney

L= H=

Press down the corresponding directional keys or scroll trackball as per your requirement. The

measurement method of the length(L) and height(H) are order to the distance measurement. When

measure “right Kidney”, the data will automatically display on the screen, at the same time the

measurement data will be stored in the report. For the same patient, measure “right Kidney” again, the

newest measurement data display on the report.

Prostate

Selecting Prostate will generate a submenu,from which four measurements for the Bladder can be

performed,press ENTER and the following menu will appear on the screen.

Transversal
Sagittal
Oblique 1
Oblique 2

- 65 -
Shenzhen emperor electronic technology Co.,Ltd.

Note: The prostate’ masurement method is order to the bladder’ measurement method.

Urology Report

Select Urology Report, the report will displayed on the screen:

Hospital Name:
Patient: Section: Date:
Bladder Transverse Sagittal Oblique 1 Oblique 2
Height:
Breadth:
Vb1= γ=1.23
Vb2= α=1.00 β=0.00
Vbmean2=
Vbmean3=
Vbmean4=
Left kidney Right kidney
Length:
Height:
Prostate Transverse Sagittal Oblique 1 Oblique 2
Height:
Breadth:
Vp1=
Vp2=
Vpmean2=
Vpmean3=
Vpmean4=
Doctor:

Note:

1. Once you choose each software packages, all the measurement data will be saved in the

report.

2. For the same patient, the data in the report will be updated if you carry out the

measurement upon the same item again, and the system will show the newest data as well.

3. When you press down the key New Patient, the data that saved in the report for the

previous patient will be removed.

4. When connect the laser printer, Press down the alphanumeric key P to print the report.

Explanation of the report

The bladder and prostate’s measurement method is Height (H)/Breadth (B).

The corresponding volume is calculated according to the formula:

- 66 -
Shenzhen emperor electronic technology Co.,Ltd.

V=1/6×п×H×B2

Volume calculation according to measurements made in one plane:

With this method the bladder is scanned in, the transverse plane only. The estimated bladder volume is:

Vb1=V×γ

Where Vb1 is the measured volume in the transverse plane, γ is the coefficient.

a) According to this method one first measures the H/B in the transverse plane and then the length in the

sagittal plane. The formula to calculate to volume now is:

Vb2=αV +β

The values for αand βcan be set.

b) Volume calculation according to the average for the measured transverse and sagittal volume:

Vbmean2= (Vb1 + Vb2)/2

c) Volume calculation according to measurements made in three or four planes:

Vbmean3= (Vb1 + Vb2 + Vob1)/3

Vbmean4= (Vb1 + Vb2 + Vob1 + Vob2)/4

Where Vob1 is the measured volume in the Oblique 1 plane. Where Vob2 is the measured volume in the

Oblique 2 plane.

Set Bladder Coefficients

When the volume of the bladder is calculated according to measurements made in two planes, a correction

factor is introduced into the formula for calculation of the bladder or bladder geometry yielding.

Select Set Bladder Coefficients, in the right top corner of screen will display as following:

The correction factors for α、β、γcan now be entered. Pressing the key ENTER to finish

the set.

α 1.00
β 0.00
γ 1.23

Note: Please preset the coefficient before you start your measurement if you need. The

user-set coefficient will be lost and it returns to the default. The users are usually

recommended to make measurement with the default coefficient.

- 67 -
Shenzhen emperor electronic technology Co.,Ltd.

I Cardiology

1——Settings Ö B——Left Ventricle Ö Diastole 2 Chamber


2——Function Ö M——Left Ventricle Ö Systole 2 Chamber
3——Obstetrics Ö A——Left Atrium Ö Diastole 4 Chamber
4——Gynecology Ö T——Mitral Valve Ö Systole 4 Chamber
5——Urology Ö P——Pulmonary Valve Ö Outflow Tract Diameter
6——Cardiology Ö C——Cardiology Report
S——Bmode Method Ö Diastole
Systole

Diastole
Systole
Aortic Cusp Separation
AV Ejection Time

D-E Excursion
E-F0 Slope
E-F Slope
A-C Interval
E-IVS Distance

E-F Slope
B-C Slope
A-Wave Amplitude
PV Ejection Time

Dodge Single Plane


Dodge Biplane
Bullet
- 68 -
Shenzhen emperor electronic technology Co.,Ltd.

Note: There are three ways of selecting the various functions or items in each menu and/or

submenu list.

1. Select the item that you prefer by scrolling the trackball up and down or press the

direction key up and down, then press down the key ENTER;

Or 2. Select the item that you prefer by scrolling the trackball up and down or press the

direction key up and down, then press down the direction key Right to enter, or press

down the direction key LEFT to return to previous menu;

Or 3. Input the first character(number or letter) in front of each item that you need.

Note: Press down the key Esc to exit all the menus that you enter.

The cardiology measurement package of the scanner contains measurements for:

a) The left ventricle in end-diastolic and in end-systolic phase;

b) The left atrium and aorta;

c) The mitral valve;

d) The pulmonary valve.

Return to the main menu and select Cardiology. Upon pressing the key ENTER, following menu will

display on the screen.

1——Settings Ö
2——Function Ö
3——Obstetrics Ö
4——Gynecology Ö
5——Urology Ö
6——Cardiology Ö
B——Left Ventricle B Ö
M——Left Ventricle M Ö
A——Left Atrium Ö
T——Mitral Valve Ö
P——pulmonary Valve Ö
C——Cardiology Report
S——Bmode Method Ö

Left Ventricle B (B-mode)

Before you start your measurements for the Left Ventricle in B-mode, the measurement method must be

chosen.

- 69 -
Shenzhen Emperor Electronic Technology Co., Ltd.

Bmode Method

Display the Cardiology’s submenu, and then select Bmode Method, the following menu will be

displayed.

Dodge Single Plane


Dodge Biplane
Bullet

These three items represent three different sets of formulas from which the following diagnostic can be

calculated:

a) End-diastolic volume;

b) End-systolic volume;

c) Stroke volume;

d) Ejection fraction;

e) Cardiac output.

Left Ventricle B (Dodge Single Plane)

After select Dodge Single Plane, then return to the main menu and Select Left Ventricle B, then press

down the key ENTER, the screen will display as following:

B——Left Ventricle B Ö
1——Settings Ö
M——Left Ventricle M Ö
2——Function Ö Diastole 2 Chamber
A——Left Atrium Ö
3——Obstetrics Ö Systole 2 Chamber
T——Mitral Valve Ö
4——Gynecology Ö Diastole 4 Chamber
P——pulmonary Valve Ö
5——Urology Ö Systole 4 Chamber
C——Cardiology Report
6——Cardiology Ö Outflow Tract Diameter
S——Bmode Method Ö

This menu displays the different measurements for the left ventricle during End-diastole and

End-systole, only the Diastole 4 Chamber, Systole 4 Chamber and Outflow Tract Diameter can be

selected.

For these calculations of the left ventricle an apical four chamber view must be displayed.

- 70 -
Shenzhen Emperor Electronic Technology Co., Ltd.

In four chamber views the length is derived from the midpoint of the mitral annulus to the most distant

point, presumed to be the apex of the posterior position in the short view.

When all the necessary measurements are stored the scanner will perform the necessary calculations

according to dodge single plane formula, the results will be displayed in the cardiology report.

The dodge single plane formula is described in the calculation formulas.

Left Ventricle B (Dodge Biplane)

After selecting Dodge Biplane, then return to the main menu and Select Left Ventricle B, then press

down the key ENTER, the screen will display as following:

1——Settings Ö
2——Function Ö B——Left Ventricle B Ö
3——Obstetrics Ö M——Left Ventricle M Ö
4——Gynecology Ö Diastole 2 Chamber
A——Left Atrium Ö
5——Urology Ö Systole 2 Chamber
T——Mitral Valve Ö
6——Cardiology Ö Diastole 4 Chamber
P——pulmonary Valve Ö
Systole 4 Chamber
C——Cardiology Report
Outflow Tract Diameter
S——Bmode Method Ö

This menu displays the different measurements in the 4 chamber vie and the 2chamber view for the

left ventricle during End-diastole and End-systole.

For these calculations of the left ventricle an apical two chamber view must be displayed.

- 71 -
Shenzhen Emperor Electronic Technology Co., Ltd.

In 4 chamber and 2 chamber views, the length is derived from the midpoint of the mitral annulus to the

most distant point the apex of the posterior position in the short axis view.

When all the necessary measurements are stored the scanner will perform the necessary calculation

according the dodge biplane formula. The results will be displayed in the cardiology report.

The dodge biplane formula is described in the section calculation formulas.

Left Ventricle B (Bullet)

After select Bullet, then return to the main menu and Select Left Ventricle B, then press down the key

ENTER, the screen will display as following:

1——Settings Ö
2——Function Ö B——Left Ventricle B Ö
3——Obstetrics Ö M——Left Ventricle M Ö
4——Gynecology Ö Diastole 2 Chamber
A——Left Atrium Ö
5——Urology Ö Systole 2 Chamber
T——Mitral Valve Ö
6——Cardiology Ö Diastole 4 Chamber
P——pulmonary Valve Ö
Systole 4 Chamber
C——Cardiology Report
Outflow Tract Diameter
S——Bmode Method Ö

This menu displays the different measurements for the left ventricle during End-diastole and

End-systole, only the Diastole 4 Chamber, Systole 4 Chamber and Outflow Tract Diameter can be

selected.

For all calculation of the left ventricle you must display an apical four chamber view.

- 72 -
Shenzhen Emperor Electronic Technology Co., Ltd.

How to different parameters for the calculation of the volume of the left ventricle according the bullet

formula must be measured is shown below:

In four chamber views the length is derived from the midpoint of the mitral annulus to the most distant

point, presumed to be the apex of the posterior position in the short axis view:

The bullet formula is described in the section calculation formulas.

Left Ventricle M (M-mode)

Select Left Ventricle, then press down the key ENTER, the screen will display as following:

1——Settings Ö
2——Function Ö
3——Obstetrics Ö
4——Gynecology Ö
B——Left Ventricle B Ö
5——Urology Ö
M——Left Ventricle M Ö
6——Cardiology Ö
A——Left Atrium Ö
Diastole
T——Mitral Valve Ö
Systole
P——pulmonary Valve Ö
C——Cardiology Report
S——Bmode Method Ö

This menu displays the different measurements for the left ventricle during End-diastole and End-systole.

For these calculations of the left ventricle a parasternal view of the left ventricle must be displayed.

- 73 -
Shenzhen Emperor Electronic Technology Co., Ltd.

Left Atrium (M-mode)

Select Left Atrium, then press down the key ENTER, the screen will display as following:

1——Settings Ö
2——Function Ö
3——Obstetrics Ö
4——Gynecology Ö
5——Urology Ö
6——Cardiology Ö
B——Left Ventricle B Ö
M——Left Ventricle M Ö
A——Left Atrium Ö
T——Mitral Valve Ö
Diastole
P——pulmonary valve Ö
Systole
C——Cardiology Report
Aortic Cusp Separation
S——Bmode Method Ö
AV Ejection Time

The menu displays the different measurements for the left atrium and aorta during End-diastole and

End-systole.

- 74 -
Shenzhen Emperor Electronic Technology Co., Ltd.

Mitral Valve (M-mode)

Select Mitral Valve, then press down the key ENTER, the screen will display as following:

1——Settings Ö
2——Function Ö
3——Obstetrics Ö
4——Gynecology Ö
5——Urology Ö
6——Cardiology Ö
B——Left Ventricle B Ö
M——Left Ventricle M Ö
A——Left Atrium Ö
T——Mitral Valve Ö
P——pulmonary valve Ö
D-E Excursion
C——Cardiology Report
E-F0 Slope
S——Bmode Method Ö
E-F Slope
A-C Interval
E-IVS Distance

This menu displays the different measurements for the Mitral Valve.

For these calculations of the Mitral Valve a parasternal long axis view of the left ventricle, left atrium and

mitral valve must be displayed.

The different parameters to be measured in this menu are shown below:

- 75 -
Shenzhen Emperor Electronic Technology Co., Ltd.

D-E Excursion = Mitral valve excursion or opening amplitude

E-F0 Slope = Early diastolic posterior motion

E-F Slope = Early diastolic posterior motion(measured with faster heart rates)

A-C Interval = Time interval between the peak of the A wave and the C point

E-IVS Distance = Distance between the IV septum and the anterior mitral valve

leaflet at maximum opening

Pulmonary Valve (M-mode)

Select Pulmonary Valve, then press down the key ENTER, the screen will display as following:

1——Settings Ö
2——Function Ö
3——Obstetrics Ö
4——Gynecology Ö
5——Urology Ö
6——Cardiology Ö
B——Left Ventricle B Ö
M——Left Ventricle M Ö
A——Left Atrium Ö
T——Mitral Valve Ö
P——pulmonary valve Ö
C——Cardiology Report
E-F Slope
S——Bmode Method Ö
B-C Slope
A-Wave Amplitude
PV Ejection Time
This menu displayed the different measurements for the pulmonary valve.

- 76 -
Shenzhen Emperor Electronic Technology Co., Ltd.

The different parameters to be measured in this menu are shown below:

E-F Slope = Time interval of the diastole phase of the atrium

B-C Slope = Opening slope of the pulmonary valve

A-Wave Amplitude = Diastolic dip resulting from atrial contraction

PV Ejection Time = Time between point B and E

Cardiology Report

Select Cardiology Report, then press down the key ENTER, the screen will display as following:

- 77 -
Shenzhen Emperor Electronic Technology Co., Ltd.

Hospital Name:
Patient: Section: Date:
Left Ventricle B left Ventricle M
End Diastolic V:
End Systolic V:
Stroke Volume:
Ejection Fraction:
Left Atrium
Diastole Systole
LA diameter:
LA posterior wall:
Aortic Diameter:
AC Separation
AV Ejection Time:
Mitral Valve
D-E Excursion:
E-F0 Slope:
E-F Slope:
A-C Interval:
E-IVS Distance:
Pulmonary Valve
E-F Slope:
B-C Slope:
A-Wave Amplitude:
PV Ejection Time:
Doctor:

Note:

1. Once you choose each software packages, all the measurement data will be saved in the

report.

2. For the same patient, the data in the report will be updated if you carry out the

measurement upon the same item again, and the system will show the newest data as well.

3. When you press down the key New Patient, the data that saved in the report for the

previous patient will be removed.

4. When connect the laser printer, Press down the alphanumeric key P to print the report.

- 78 -
Shenzhen Emperor Electronic Technology Co., Ltd.

J Maintenance

8.1) Check power supply regularly. Do not switch on this unit when the power exceeds the stipulated

value;

8.2) Check the power cord and probe wire regularly, change them immediately if they are found out

broken or damaged;

8.3) Unplug the power cord before cleaning the unit; remember to disconnect the probe from the unit

before cleaning the probe with soft and dry cloth. Do not use wet cloth, or the cloth with detergent

and germicide containing alcohol or corrosive substances;

8.4) Clean and disinfect the probe by tampon with medical alcohol after each examination and place it in

the probe holder or in its original storage case. Never use any organic liquid for cleaning;

8.5) Do not presumptuously discover the inner for cleaning, or shake or disassemble the parts of the unit;

8.6) Please check the ventilation fans for proper operation. Contact our customer service center

immediately if it dose not work. Do not use the equipment with defective fan so as to avoid

unpredictable result;

8.7) In order to keep good ventilation, certain distance must be kept between the back of equipment and

the wall, and the air filter must be cleaned regularly;

8.8) The unit should be placed evenly and stably. Handle it with care and do not put any heavy object on

the unit. While removing the equipment, one hand holds the top handle and the other hand holds the

bottom;

8.9) The contact agent for the probe is medical ultrasound coupling which is used to make sure a complete

contact between the probe and body parts and to obtain good clinic image. Some ultrasound gels may

be toxic to human reproductive cells. It is recommended that you check with the product’s

manufacturer to determine appropriate use;

8.10) Do not dip the probe in any other conductive solutions to avoid any corrosion of the probe and its

protective sleeve;
- 79 -
Shenzhen Emperor Electronic Technology Co., Ltd.

8.11) Do not disconnect the probe from the unit at discretion when they are properly connected;

8.12) Weaken the brightness and contrast if the unit is not used after switching on to ensure its life;

8.13) Ask /Call qualified engineer(s) to repair the unit when it breakdowns;

8.14) Power on the unit at least twice, each lasting for not less than one hour every week if the unit will

not be used for a long period of time;

8.15) Probe cleaning: Remove the ultrasound gel with a soft tissue and then gently wipe the probe dry by

using a new tissue or dry cloth. When further cleaning is required only a mild detergent or hand

soap may be used together with some water and a soft tissue cloth;

8.16) Probe maintenance: Inspect the probe carefully after a fall. A dangerous situation may arise due to

damaged insulation of probe and probe lens;

Warning: Do not drop or shock the probe slightly since the damaged chip is not reparable;

8.17) How to clean the trackball: If the trackball can be not moved effectively but image displayed is ok,

check if there is any dust or dirt inside it since they can easily get in. Please follow the following

steps:

1.Remove the trackball cover by rotating it unclockwisely with a tiny screw driver in the recessed
spot;

2.Take out the trackball;

3.Rinse the trackball in warm water and let it dry;

4.Clean the shafts with alcohol cotton;

5.Put the trackball in place and direct the recess at the protrusion, and turn the cover clockwisely
with a tiny screwdriver in the recessed spot until the cover in its original potions.

- 80 -
Shenzhen Emperor Electronic Technology Co., Ltd.

Cover Rotation Point Trackball

Rotation Point

8.18) Service life will be six years calculated from delivery date which is stated in the quality certificate.

8.19)The device should be disused complying to relevant legal regulations and principles when it is out of

using lifetime.

- 81 -
Shenzhen Emperor Electronic Technology Co., Ltd.

K Trouble-shooting
9.1) How to change the fuse

Break out the fuse from the power socket, replace the old fuse with a new one, and then fix the

cover by turning clockwise. (onlyφ5×20 F2A/250V allowed).

9.2) Please refer to the following table for other troubles that may happen:

S/No. Breakdown Description Solution

1 The indictor is still off and 1 Check power supply;


no display on the screen
after the unit is switched on 2 Check the power cord and plug;

3 Check if the fuse is broken;

4Turn the knobs “BRIGHT” and “CONTRAST”;

2 Interrupted Stripe and 1Check if it power supply is interfered by other electrical


snow-shaped interference equipment;
on the screen
2 Check if operation condition is free from any electrical or
magnetic interference;

3.Check if power supply, probe and plugs are well connected or


not;

4.Far field interference is strong when you hold the probe in


hand and is ok when you put it down; Please investigate the
interference source;

5.Check if the grounding wire is connected or not;

3 Image isn’t clear enough 1. Adjust the 6 TGC and the gain;

2. Adjust the brightness and contrast;

3. Clean the organic glass of the display;

4 Image isn’t clear in near, Adjust the 6 TGC and the gain;
medium or far distance

5 Good image but the fan 1. Brush off all dirt or dust inside the fan;
doesn’t work when the unit
is on 2. Change a new one if the fan is broken down;

6 Good image but the Refer to How to Clean the trackball.


trackball doesn’t work well

- 82 -
Shenzhen Emperor Electronic Technology Co., Ltd.

L Packaging, Storage and Transportation


10.1) Put the unit in protective poly foam and then put in the carton box.

10.2) Take out the units and power them on for four hours if the units have been put in stock for more

than 6 months, and put them back according to the instructions marked outside the carton.

10.3) Do not stack the units or let the units contact the ground directly. Remember to leave some space

between the products and walls and roofs.

10.4) Shock absorption has been deliberately considered when the parking is designed and they are

suitable for air, railway, road and sea freight. Do not shock or place the products upside down and

be free of rains or snows.

- 83 -
Shenzhen Emperor Electronic Technology Co., Ltd.

Appendix I

Table 1-1 Relations between Fetal Age and GS’s Average Inner Radius

Fetal Age Average Inner Radius Fetal Age Average Inner Radius
4.2 0.4 7.3 2.6
4.3 0.5 7.4 2.7
4.5 0.6 7.6 2.8
4.6 0.7 7.8 2.9
4.8 0.8 7.9 3.0
4.9 0.9 8.1 3.1
5.1 1.0 8.2 3.2
5.2 1.1 8.3 3.3
5.3 1.2 8.5 3.4
5.5 1.3 8.6 3.5
5.6 1.4 8.8 3.6
5.8 1.5 8.9 3.7
5.9 1.6 9.1 3.8
6.1 1.7 9.2 3.9
6.2 1.8 9.3 4.0
6.3 1.9 9.5 4.1
6.4 2.0 9.6 4.2
6.6 2.1 9.8 4.3
6.8 2.2 9.9 4.4
6.9 2.3 10.1 4.5
7.1 2.4 10.2 4.6
7.2 2.5 10.3 4.7

- 84 -
Shenzhen Emperor Electronic Technology Co., Ltd.

Table 1-2 Relations between CRL(cm)and Fetal Age

Fetal Age+Days Head-rump Radius Fetal Age+Days CRL


6+2 0.55 10+2 3.32
6+3 0.61 10+3 3.46
6+4 0.68 10+4 3.60
6+5 0.75 10+5 3.74
6+6 0.81 10+6 3.89
7+0 0.89 11+0 4.04
7+1 0.96 11+1 4.19
7+2 1.04 11+2 4.35
7+3 1.12 11+3 4.51
7+4 1.20 11+4 4.67
7+5 1.29 11+5 4.83
7+6 1.38 11+6 5.00
8+0 1.47 12+0 5.17
8+1 1.57 12+1 5.34
8+2 1.66 12+2 5.52
8+3 1.76 12+3 5.70
8+4 1.87 12+4 5.88
8+5 1.97 12+5 6.06
8+6 2.08 12+6 6.25
9+0 2.19 13+0 6.43
9+1 2.31 13+1 6.63
9+2 2.42 13+2 6.82
9+3 2.54 13+3 7.02
9+4 2.67 13+4 7.22
9+5 2.79 13+5 7.42
9+6 2.92 13+6 7.63
10+0 3.05 14+0 7.83
10+1 3.18

- 85 -
Shenzhen Emperor Electronic Technology Co., Ltd.

Table1-3 Relations between BPD(cm)and Fetal Age

Fetal Age No.of samples Average Fetal Age No.of samples Average
13 35 2.52 27 51 6.98
14 36 2.83 28 50 7.24
15 34 3.23 29 50 7.50
16 33 3.62 30 47 7.83
17 34 3.27 31 48 8.06
18 44 4.25 32 48 8.17
19 38 4.52 33 39 8.50
20 36 4.88 34 55 8.61
21 41 5.22 35 57 8.70
22 50 5.45 36 55 8.81
23 44 5.80 37 53 9.00
24 52 6.05 38 53 9.08
25 41 6.39 39 51 9.21
26 50 6.68 40 33 9.28

- 86 -
Shenzhen Emperor Electronic Technology Co., Ltd.

Table1-4 Relations between HC(cm) and Fetal Age

Head Circumference Fetal Age Head Circumference Fetal Age


8.0 13.4 22.0 23.9
8.5 13.7 22.5 24.4
9.0 14.0 23.0 24.9
9.5 14.3 23.5 25.4
10.0 14.6 24.0 25.9
10.5 15.0 24.5 26.4
11.0 15.3 25.0 26.9
11.5 15.6 25.5 27.5
12.0 15.9 26.0 28.0
12.5 16.3 27.0 29.2
13.0 16.6 27.5 29.8
13.5 17.0 28.0 30.3
14.0 17.3 28.5 31.0
14.5 17.7 29.0 31.6
15.0 18.1 29.5 32.2
15.5 18.4 30.0 32.8
16.0 18.8 30.5 33.5
16.5 19.2 31.0 34.2
17.0 19.6 31.5 34.9
17.5 20.0 32.0 35.5
18.0 20.4 32.5 36.3
18.5 20.8 33.0 37.0
19.0 21.2 33.5 37.7
19.5 21.6 34.0 38.5
20.0 22.1 34.5 39.2
20.5 22.5 35.0 40.0
21.0 23.0 35.5 40.8
21.5 23.4 36.0 41.6

- 87 -
Shenzhen Emperor Electronic Technology Co., Ltd.

Table1-5 Relations between AC(cm) and Fetal Age

Abdomen Circumference Fetal Age Abdomen Circumference Fetal Age


10.0 15.6 23.5 27.7
10.5 16.1 24.0 28.2
11.0 16.5 24.5 28.7
11.5 16.9 25.0 29.2
12.0 17.3 25.5 29.7
12.5 17.8 26.0 30.1
13.0 18.2 26.5 30.6
13.5 18.6 27.0 31.1
14.0 19.1 27.5 31.6
14.5 19.5 28.0 32.1
15.0 20.0 28.5 32.6
15.5 20.4 29.0 33.1
16.0 20.8 29.5 33.6
16.5 21.3 30.0 34.1
17.0 21.7 30.5 34.6
17.5 22.2 31.0 35.1
18.0 22.6 31.5 35.6
18.5 23.1 32.0 36.1
19.0 23.6 32.5 36.6
19.5 24.0 33.0 37.1
20.0 24.5 33.5 37.6
20.5 24.9 34.0 38.1
21.0 25.4 34.5 38.7
21.5 25.9 35.0 39.2
22.0 26.3 35.5 39.7
22.5 26.8 36.0 40.2
23.0 27.3 36.5 40.8

- 88 -
Shenzhen Emperor Electronic Technology Co., Ltd.

Table 1-6 Relations between FL(cm)and Fetal Age

Fetal Age No.of Average Fetal Age No.of Average


Samples Samples
13 35 1.17 27 53 5.10
14 31 1.38 28 49 5.35
15 34 1.74 29 50 5.61
16 33 2.10 30 47 5.77
17 34 2.52 31 48 6.03
18 44 2.71 32 46 6.20
19 38 3.03 33 41 6.43
20 38 3.35 34 54 6.62
21 39 3.64 35 56 6.71
22 50 3.82 36 56 6.95
23 43 4.21 37 55 7.10
24 53 4.36 38 56 7.20
25 40 4.65 39 53 7.34
26 50 4.87 40 34 7.40

- 89 -
Shenzhen Emperor Electronic Technology Co., Ltd.

Appendix II

Acoustic Output Information For 2.5/3.5/5.0MHz Convex probe(C080-60E)

Mode B M
Parameter
P-(MPa) 2.06 2.08
Ispta(mW/cm )
2
6.40 7.01
1 SEG FOC FOC:8cm 1 SEG FOC FOC:8cm
System setting ×1.0 ×1.0
Work Freq:3.5MHz Work Freq:3.5MHz
Lp(mm) 60 60
Wpb6(‖)(mm) 3.36 3.36
Wpb6(⊥)(mm) 7.27 7.36
Prr(kHz) —— 3.13
Srr(Hz) 25.00 ——
Output beam dimensions
12.03×13.98 12.05×13.98
(mm×mm)
ƒ awf(MHz) 2.78 2.84
APF(%) 100 100
AIF(%) 100 100
Pmax(mW) 25.21 27.48
Iob(mW/cm )
2
15.70 17.08
Power-up Mode B B
Initialization Mode B B
Acoustic output freeze Yes Yes
Ltt(mm) —— ——
Lts(mm) —— ——

- 90 -
Shenzhen Emperor Electronic Technology Co., Ltd.

Acoustic Output Information For 5.0/6.5/7.5MHz Linear probe(L080-57A)

Mode B M
Parameter
P-(MPa) 3.41 3.41
Ispta(mW/cm )
2
17.28 17.24
1 SEG FOC FOC:4cm 1 SEG FOC FOC:4cm
System setting ×1.0 ×1.0
Work Freq:6.5MHz Work Freq:6.5MHz
Lp(mm) 30 30
Wpb6(‖)(mm) 1.82 1.74
Wpb6(⊥)(mm) 2.07 2.16
Prr(kHz) —— 5.44
Srr(Hz) 42.00 ——
Output beam dimensions
7.32×8.54 9.88×8.13
(mm×mm)
ƒ awf(MHz) 3.73 3.70
APF(%) 100 100
AIF(%) 100 100
Pmax(mW) 31.80 22.04
Iob(mW/cm )
2
53.27 28.74
Power-up Mode B B
Initialization Mode B B
Acoustic output freeze Yes Yes
Ltt(mm) —— ——
Lts(mm) —— ——

- 91 -
Shenzhen Emperor Electronic Technology Co., Ltd.

Acoustic Output Information For 5.0/6.5/7.5MHz Transvaginal probe(C080-13D)

Mode B M
Parameter
P-(MPa) 1.99 1.99
Ispta(mW/cm )
2
3.24 4.10
1 SEG FOC FOC:4cm 1 SEG FOC FOC:4cm
System setting ×1.0 ×1.0
Work Freq:6.5MHz Work Freq:6.5MHz
Lp(mm) 32 32
Wpb6(‖)(mm) 3.55 2.99
Wpb6(⊥)(mm) 2.14 2.06
Prr(kHz) —— 5.41
Srr(Hz) 41.70 ——
Output beam dimensions
5.01×8.43 5.48×8.87
(mm×mm)
ƒ awf(MHz) 3.68 3.66
APF(%) 100 100
AIF(%) 100 100
Pmax(mW) 7.15 7.76
Iob(mW/cm )
2
16.92 15.98
Power-up Mode B B
Initialization Mode B B
Acoustic output freeze Yes Yes
Ltt(mm) —— ——
Lts(mm) —— ——

- 92 -