[ FOR CALIBRATION ]
CONTENTS
-1-
DONGMUN
CONTENTS
WIRE CONNECTION
-2-
DONGMUN
-3-
DONGMUN
[Fig. 1]
[Fig. 2]
-4-
DONGMUN
AC 220V ①(AC 0V) + ③(AC 220V)
AC 240V ①(AC 0V) + ④(AC 240V)
-5-
DONGMUN
1. BOARD CALIBRATION
VR3 Small Scale (10mA ~ 80mA) mA FEEDBACK to mAs adjustment with VF converter
Measuring Point:TP5
Adjust the VR3 until the TP5 set to 25kHz± 0.5kHz.
Simple Adjustment: Connect mAs tester to the connector 8,9(mA TEST)of HV tank
and shoot x-ray and adjust mAs.
Conditions: 40kV 50mA / 10mAs
Measuring: 10 ± 0.2mAs
-6-
DONGMUN
TP15
GND
U25
VR6
TP8
kV FB
VR2
TP4
mA FB
-7-
DONGMUN
TP5, TP6
-8-
DONGMUN
1.5 AGING
-9-
DONGMUN
1.6 kV CALIBRATION
- 10 -
DONGMUN
1.7 mA CALIBRATION
- 11 -
DONGMUN
1) NORMAL CONDITIONS
- 12 -
DONGMUN
- 13 -
DONGMUN
- 14 -
DONGMUN
- 15 -
DONGMUN
22
10
12
16 5
9 3 4 11 1 2 6
23 24
13 14 15 20 17 21 7 8
18 19
- 16 -
DONGMUN
[1]Power on Switch
Switch on the power supply
[9]OVER Indication(Red)
The LED is switched on when operation exceeds radiography conditions. The exposure
is prohibited while [the over LED] is lighted on. In case the LED is turned on due to
changed exposure conditions in the APR mode, the memory of condition should not be
saved. Under the AEC mode selection, if radiography time exceeds backup time, it
blinks. Backup release switch cancels over display at the time of AEC mode.
[10]ERROR Indicator(Red)
The LED is turned on when an error occurs.
- 17 -
DONGMUN
If the target region, the radiography direction, and patient size are chosen in the APR mode,
radiography conditions (kV, mA, mAs/sec) shall be set up automatically. Each photographic
condition can be changed and memorized by the Patient Size setting switch (condition
memory switch).
[23]Small-focus Display
The light is switched on when the small focus of the tube is selected.
- 18 -
DONGMUN
2. DIP SW FUNCTION
Adjust DIP Switch SW32 (8P) located on the back front of control panel and operate the each
calibration.
DIP SW (SW 32: 8P) Function Display
- 19 -
DONGMUN
DIP SW-8:ON Select Exposure End Buzzer
SW-OFF: Select Exposure On Buzzer.
- 20 -
DONGMUN
*
Model mA Range Remark
DIG-161 50~200mA 50mA Step
DIG-201 10~250mA
DIG-321 10~400mA
DIG-401A 10~500mA
DIG-501(T) 10~630(800)mA (T) : 3-Phase
DIG-801T 10~999mA 3-Phase
- 21 -
DONGMUN
- Step: 0.1、0.2、0.3、0.4、0.5、0.6、0.8、1.0、1.3、1.6、2.0、2.5、3.2、4.0、
5.0、6.3、8.0、10.0、12.5、16、20、25、32、40、50、63、80、100、125、
160、200、250、320、400、500、630、800、999* mAs
- 22 -
DONGMUN
- 23 -
DONGMUN
2) Selection of Field
Select the target region with setting switch. You can choose the field by pushing
switches.
The selected field of lon Chamber will be displated on the LED panel.
You cannot make radiography without selecting the field.
- 24 -
DONGMUN
4.RADIOGRAPHY
The selection of exposure on and off buzzer is operated by the DIP SW-8.
(SW-OFF/ select exposure on buzzer, SW-ON/ select exposure off buzzer)
- 25 -
DONGMUN
.
- 26 -
DONGMUN
C-1 _12 [160mA] KVREF Shift Data
C-1 _13 [200mA] KVREF Shift Data
C-1 _14 [250mA] KVREF Shift Data
C-1 _15 [320mA] KVREF Shift Data
C-1 _16 [400mA] KVREF Shift Data
C-1 _17 [500mA] KVREF Shift Data
C-1 _18 [630mA] KVREF Shift Data
C-1 _19 [800mA] KVREF Shift Data
C-1 _20 [999mA] KVREF Shift Data
- 27 -
DONGMUN
C-1 _36 [320mA] MAREF Shift Data
C-1 _37 [400mA] MAREF Shift Data
C-1 _38 [500mA] MAREF Shift Data
C-1 _39 [630mA] MAREF Shift Data
C-1 _40 [800mA] MAREF Shift Data
C-1 _41 [999mA] MAREF Shift Data
- 28 -
DONGMUN
C-2 _15 Not in Use
C-2 _16 Not in Use
C-2 _17 Not in Use
C-2 _18 Not in Use
C-2 _19 Not in Use
C-2 _20 Not in Use
- 29 -
DONGMUN
C-2 _36 [320mA] PREP_MAREF Calibration Data
C-2 _37 [400mA] PREP_MAREF Calibration Data
C-2 _38 [500mA] PREP_MAREF Calibration Data
C-2 _39 [630mA] PREP_MAREF Calibration Data
C-2 _40 [800mA] PREP_MAREF Calibration Data
C-2 _41 [999mA] PREP_MAREF Calibration Data
- 30 -
DONGMUN
C-2 _57 Not in Use
C-2 _58 Not in Use
C-2 _59 Not in Use
C-2 _60 [AEC Mode] DENSITY=-3 Shift Data
C-2 _61 [AEC Mode] DENSITY=-2 Shift Data
C-2 _62 [AEC Mode] DENSITY=-1 Shift Data
C-2 _63 [AEC Mode] DENSITY=+1 Shift Data
C-2 _64 [AEC Mode] DENSITY=+2 Shift Data
C-2 _65 [AEC Mode] DENSITY=+3 Shift Data
C-2 _66 Not in Use
C-2 _67 Not in Use
C-2 _68 Not in Use
C-2 _69 Not in Use
C-2 _70 [AEC Mode] AEC1 Reference Level Data
- 31 -
DONGMUN
- 32 -
DONGMUN
5. DISPLAY OF PANEL
- 33 -
DONGMUN
Calibration Mode
mA
・ Calibration numbers display
・ Select the numbers
mAs
・ Data Format
sec
- 34 -
DONGMUN
Push the [2]power off switch and turn the power off.
Press 1st switch of DIPSW(SW32 : 8P)and turn it on.
Push the power on [1] switch and turn the power on.
Control panel is displayed as a calibration mode.
C-1:KVREF&MAREF SETTING
Select the numbers in the calibration mode with [4] mA setting switch.
The selected number in calibration mode is appeared on mA display on the panel.
__0~_20, _21~_41
The calibration data is shown on the [5] mAs/sec display after the calibration mode and
numbers are selected.
-20∼ 00 ∼+20
- 35 -
DONGMUN
* Execute the x-ray exposure with following methods after the data is selected in
KVREF/MAREF setting mode.
[18] Execute the conversion of monitor(the monitor after calibration)with backup release
switch. During the monitor recording, it is possible to make x-ray exposures. The setting
of mA during calibration is fixed for the time. APR•AEC Mode is not in option this time.
As the mA setting is fixed, make exposure with selecting other conditions such as
kV•mAs/sec and check the movement of calibration data. When the exposure condition
exceeds the capacity, the ‘over’ LED is lighted on. While the light is turned on, exposure of
x-ray is forbidden. In case of rearrangement after x-ray exposure, operate the data setting
again with back up release switch and return the exposure monitor to calibration monitor.
How to end the calibration mode
Turn it off with power off switch and the position will return to the DIP SW mode before the
start of Calibration Mode.
- 36 -
DONGMUN
Shift the KVREF value from the total base when the mA is 25.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-:
W KVREF goes down
It is possible to change on the screen with [18] backup switch(25 mA fixed).
- 37 -
DONGMUN
Shift the KVREF value from the total base when the mA is 32.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-:
W KVREF goes down
It is possible to change on the screen with [18] backup switch(32 mA fixed).
Shift the KVREF value from the total base when the mA is 40.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-:
W KVREF goes down
It is possible to change on the screen with [18] backup switch(40 mA fixed).
Shift the KVREF value from the total base when the mA is 50.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-:
W KVREF goes down
It is possible to change on the screen with [18] backup switch(50 mA fixed).
C-1__8 : [63mA : small focus] KVREF Shift Data
Shift the KVREF value from the total base when the mA is 63.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
It is possible to change on the screen with [18] backup switch(63 mA fixed).
Shift the KVREF value from the total base when the mA is 80.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
It is possible to change on the screen with [18] backup switch(80 mA fixed).
- 38 -
DONGMUN
C-1_10 : [100mA : small focus] KVREF Shift Data
Shift the KVREF value from the total base when the mA is 100.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
W It is possible to change on the screen with [18] backup switch(100 mA fixed).
- 39 -
DONGMUN
Shift the KVREF value from the total base when the mA is 320.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
W It is possible to change on the screen with [18] backup switch(320 mA fixed).
- 41 -
DONGMUN
Shift the MAREF value from the total base when the mA is 40.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W
It is possible to change on the screen with [18] backup switch(40 mA fixed).
- 43 -
DONGMUN
- 44 -
DONGMUN
- 45 -
DONGMUN
Shift the MAREF value from the total base when the mA is 160.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up - 46 -
Data=-: MAREF goes down
It is possible to change on the screen with [18] backup switch(160 mA fixed).
DONGMUN
- 47 -
DONGMUN
C-2_39 : [630mA :large focus] PREP/MAREF Calibration Data
Use in case C-2 __44 Bucky 1(Table) selection Data delay time control.
Data range is from 000 to 200 and starting point is 020
When data is 020, delay time is 200msec.
- 48 -
Operation in Bucky 1(Table) selection
Magnet for Bucky 1(Table) is called as ON under ready-up condition.
Turn the magnet off with x-ray signal input and the action after delay time is called as
HV-Enable on.
DONGMUN
Use in case C-2 __45 Bucky 2(stand) selection Data delay time control.
Data range is from 000 to 200 and starting point is 020
When data is 020, delay time is 200msec.
Operation in Bucky 1(Table) selection
Magnet for Bucky 2(table) is called as ON under ready-up condition.
Turn the magnet off with x-ray signal input and the action after delay time is called as
HV-Enable on.
- 50 -
DONGMUN
Push the [2]power off switch and turn the power off.
Press 1st switch of DIPSW(SW32 : 8P)and turn it on.
Push the [1] power on switch and turn the power on.
Control panel is shown as a calibration mode.
Press the [6]mAs/sec switch for a second. When the check buzzer sounds, the data
initialization is completed.
- 52 -
DONGMUN
16 5
9 3 4 1 2
- 53 -
DONGMUN
17
21
13 14 18 19 20
15
The operation of x-ray Exposure is prohibited under the APR Calibration Mode.
Only the changes of APR radiography conditions are possible.
The data is saved in memory area of APR Calibration Mode.
The radiography conditions could be called out from work area.
In this area, rearrangement of APR radiography condition is available.
When you complete the calibration work, save the changed APR data in the memory area
by pushing [17]APR Data memory switch.
- 54 -
DONGMUN
Push the [2]power off switch and turn the power off.
Press 1st switch of DIPSW(SW 32 : 8P)and turn it on.
Push the power on [1] switch and turn the power on.
Control panel is displayed as a APR Calibration Mode.
- 55 -
DONGMUN
Operate the data calibration after the selection of target regions, exposure directions and
patient size. It saves the data memory as a whole unlike under the normal APR mode
[17]Push the APR Data memory switch. (The check buzzer sounds)
*The buzzer does not sound when the conditions can not be saved.
• When the OVER sign is included in the APR radiography conditions, the numbers 1-9
will be appeared on the kV, mA and sec display. The appeared numbers represents the
regions for exposure. Then, change the over sign on display and push the [17]APR
Data memory switch again.
• When the density is not 0, reset the density =0 and push the APR Data memory
switch.
- 56 -
DONGMUN
*In case the conditions can not be saved. (The check buzzer does not sound)
• When the OVER sign is included in the APR radiography conditions, the numbers from
1-9 will be appeared on the kV, mA and sec display. The appeared numbers represents
the regions for exposure. Then, change the over sign on display and push the [17]APR
Data memory switch again.
• When the density is not 0, reset the density =0 and push the APR Data memory switch.
- 57 -
DONGMUN
16 5
3 4 1 2
17
14 15
- 58 -
DONGMUN
-Push the [1] power on switch and make the power on.
The control panel will be changed into calibration mode.
- 59 -
DONGMUN
1) You can arrange the AEC level with % shift for Density setting by arranging the
C-2_60 ~ 65 : AEC Level Shift Data.
2) The data of AEC level shift will change the related AEC Reference Level Data (%)
under density 0.
EX)
AEC Level Data : 128 (2.5V) → AEC Level Data : 153 (3V)
Density % DEC Level(V) Shift Density Data DEC Level(V)
-3 70 70 1.75V Data -3 25 25 0.62V
-2 80 80 2V change -2 50 50 1.25V
-1 90 90 2.25V -1 75 75 1.87V
0 100 2.5V 0 100 2.5V
- 60 -
DONGMUN
1 110 110 2.75V 1 125 125 3.12v
2 120 120 3V 2 150 150 3.75V
3 130 130 3.25V 3 175 175 4.37V
EX)
- 61 -
DONGMUN
5 1 2
4
17
- 62 -
DONGMUN
Push the power on [1] switch and turn the power on.
Control panel appears as a Counter Mode.
- 63 -
DONGMUN
[3]kV Display:Freqency
[4]mA Display:Frequency
[5]sec Display:Blank
Range:000000∼999999
(In case exceeding the maximum value : 000000)
- 64 -
DONGMUN
- 65 -
DONGMUN
When problems occurred during the use of equipment, they should be marked on the [3]kV,
[4]mA setting section. As well, the [10] error LED will be lighted on with the occurrence of
error.
Error Code
The error code is displayed when problems, such as abnormalities, occur to equipment.
While the error code is displayed, radiography operation is forbidden for safety.
A power supply is re-switched off and on, then an error code is cleared. If the problem of
equipment is not solved at the time of a power-supply re-switched on, the display is lighted
on again.
Check code
Check codes are warning displays when an error is on operation of equipment unlike an error
code. While the check code is displayed, radiography operation is forbidden for safety. Check
code is cleared when the backup release switch is switched on. Please refer to the check code
list for the case the clearance method is different.
- 66 -
DONGMUN
- 67 -
DONGMUN
Err 17 Not Defined
- 68 -
DONGMUN
CODE Description
Ch_01 TUBE THERMOSTAT OPERATION
The temperature of the slim and long shaped light bulb is higher than 80C.
Thermo switch is in open position.
Release the check code at Thermo switch off condition.
Ch_02 When another switch is pushed during exposure
Release the check code with back up reset switch.
Ch_03 When PREP. Switch on OP PANEL is detached during exposure
Release the check code with back up reset switch.
Ch_04 When X-RAY switch on OP PANEL is detached during exposure
Release the check code with back up reset switch.
Ch_05 When a hand switch (PREP) is detached during exposure
Release the check code with back up reset switch.
Ch_06 When a hand switch (X-RAY) is detached during exposure
Release the check code with back up reset switch.
Ch_07 Shortage of power-supply capacity for exposure
The power voltage drop is below AC160V and DC 150V.
Release the check code with input of power.
Ch_08 Simultaneous exposure prohibition
Other x-ray system is executed simultaneously when the exposure ready switch
was pushed.
Release the check code by off the power of other x-ray system.
Ch_09 Warning to X-rays room door opening
The door of x-ray room was opened when the exposure ready switch was pushed.
Stop the exposure of x-ray in this case and the check code is released when the
door is closed.
Ch_10 AEC Voltage Level Check
After setting AEC Voltage Level, when voltage level which is input from Ion
Chamber when do Exposure does not reach and is ended by time and mAs that is
established to AEC End Signal .
- 69 -
DONGMUN
- 70 -
DONGMUN
- 71 -
DONGMUN
2. Check Point
Check Code No. Parts Check Point
Check 01 I/F Connection
Tube Tube Temperature
Check 02 OP Panel S/W Input
Check 03 OP Panel S/W Input
Check 04 OP Panel S/W Input
Check 05 Hand S/W Hand S/W Input
Check 06 Hand S/W Hand S/W Input
Check 07 MAIN AC Input
High-Voltage Capacitor Voltage Level
(DC 300~310V)
Check 08 I/F I/F Connection
Check 09 I/F I/F Connection
Check 10 Ion Chamber Ion Chamber Input Level
- 72 -
DONGMUN
[C-3]It means the APR Initial Data transmitted by the initialization of calibration data.
*If possible, initial data is recommended to be 1 type, and in case you need several kinds of
initial data, different APR conditions about ROM should be prepared and correspond.
- 73 -
DONGMUN
Upper Limbs
Lower Limbs
- 74 -
DONGMUN
19 0.373 0.745 49 0.961 1.922
20 0.392 0.784 50 0.98 1.961
21 0.412 0.824 51 1 2
22 0.431 0.863 52 1.02 2.039
23 0.451 0.902 53 1.039 2.078
24 0.471 0.941 54 1.059 2.118
25 0.49 0.98 55 1.078 2.157
26 0.51 1.02 56 1.098 2.196
27 0.529 1.059 57 1.118 2.235
28 0.549 1.098 58 1.137 2.275
29 0.569 1.137 59 1.157 2.314
30 0.588 1.176 60 1.176 2.353
- 75 -
DONGMUN
79 1.549 3.098 109 2.137 4.275
80 1.569 3.137 110 2.157 4.314
81 1.588 3.176 111 2.176 4.353
82 1.608 3.216 112 2.196 4.392
83 1.627 3.255 113 2.216 4.431
84 1.647 3.294 114 2.235 4.471
85 1.667 3.333 115 2.255 4.51
86 1.686 3.373 116 2.275 4.549
87 1.706 3.412 117 2.294 4.588
88 1.725 3.451 118 2.314 4.627
89 1.745 3.49 119 2.333 4.667
90 1.765 3.529 120 2.353 4.706
- 76 -
DONGMUN
138 2.706 5.412 168 3.294 6.588
139 2.725 5.451 169 3.314 6.627
140 2.745 5.49 170 3.333 6.667
141 2.765 5.529 171 3.353 6.706
142 2.784 5.569 172 3.373 6.745
143 2.804 5.608 173 3.392 6.784
144 2.824 5.647 174 3.412 6.824
145 2.843 5.686 175 3.431 6.863
146 2.863 5.725 176 3.451 6.902
147 2.882 5.765 177 3.471 6.941
148 2.902 5.804 178 3.49 6.98
149 2.922 5.843 179 3.51 7.02
150 2.941 5.882 180 3.529 7.059
- 77 -
DONGMUN
197 3.863 7.725 227 4.451 8.902
198 3.882 7.765 228 4.471 8.941
199 3.902 7.804 229 4.49 8.98
200 3.922 7.843 230 4.51 9.02
201 3.941 7.882 231 4.529 9.059
202 3.961 7.922 232 4.549 9.098
203 3.98 7.961 233 4.569 9.137
204 4 8 234 4.588 9.176
205 4.02 8.039 235 4.608 9.216
206 4.039 8.078 236 4.627 9.255
207 4.059 8.118 237 4.647 9.294
208 4.078 8.157 238 4.667 9.333
209 4.098 8.196 239 4.686 9.373
210 4.118 8.235 240 4.706 9.412
- 78 -
DONGMUN
WIRE CONNECTION
- 79 -
DONGMUN
- 80 -
DONGMUN
1. HV TANK CONNECTION
T2 T1 OIL INJECTION
CATHOD ANODE
GND
CORE
* CORE
No. Symbol Description Remark
1 mA TP+ mA Test Point (+) mA & mAs Meter
2 mA TP- mA Test Point (-) Connection
3 F-L Large Focus of Filament
4 F-COM Common
5 F-S Small Focus of Filament
6 mA +F/B mA Feedback (+)
7 mA –F/B mA Feedback (-)
8 kV +F/B kV Feedback (+)
9 INT Tank Interlock
10 GND Ground(Earth)
- 81 -
DONGMUN
CN503
CN502
TB501
TB503
TB504
- 82 -
DONGMUN
- 83 -
DONGMUN
- 84 -
DONGMUN
- 85 -