You are on page 1of 86

MANUAL-2

[ FOR CALIBRATION ]

DONGMUN HIGH FREQUENCY GENERATOR

DIG SERIES (FOR 1-TUBE)

DONGMUN CO., LTD


DONGMUN

CONTENTS

INPUT POWER SECTION

1. Input Power Selection ------------------------------- 4

BOARD kV & mA CALIBRATION

1. Board Calibration ------------------------------- 6


1.1 HT Control Calibration ------------------------------- 7
1.2 HT Control Component Layout ------------------------------- 7
1.3 FILAMENT Calibration ------------------------------- 8
1.4 FILAMENT Control Component Layout ------------------------------- 8
1.5 Aging ------------------------------- 9
1.6 kV Calibration ------------------------------- 10
1.7 mA Calibration ------------------------------- 11
1.8 mA Oscilloscope Wave Form ------------------------------- 12
1.9 Bucky Type Selection ------------------------------- 14
1.10 Bucky OK Signal ------------------------------- 14

-1-
DONGMUN

CONTENTS

OPERATION PANEL CALIBRATION

1. Operation Panel Description ---------------------------- 16


2. DIP SW Function ---------------------------- 19
3. Radiography Selection in Normal Mode ---------------------------- 20
3-1. Setting Manual Radiography ---------------------------- 21
3-2. Setting APR Radiography ---------------------------- 23
3-3. Setting AEC Radiography ---------------------------- 24
4. Radiography ---------------------------- 25
5. Calibration Mode ---------------------------- 26
5-1. Display of Panel in Calibration Mode ---------------------------- 33
5-2. KVREF/MAREF Setting ---------------------------- 34
5-3. KVREF/MAREF Calibration Details ---------------------------- 36
5-4. Guide to Each Calibration Details ---------------------------- 44
5-5. Initialization of Calibration Data ---------------------------- 49
5-6. Initialization Details of Calibration Data ---------------------------- 50
5-7. APR Calibration Mode ---------------------------- 51
5-8. AEC Calibration Mode ---------------------------- 55
5-8. Counter Mode ---------------------------- 59
6. Error Code & Check Code ---------------------------- 62
7. Trouble Shooting Guide ---------------------------- 66
8. APR Initial Transmission Data ---------------------------- 69
9. HT-Controller A/D Reference ---------------------------- 70

WIRE CONNECTION

1. HV Tank Connection ---------------------------- 76


2. I/F Board Connection ---------------------------- 77
3. AEC Borad Connection ---------------------------- 79

-2-
DONGMUN

INPUT POWER SELECTION

-3-
DONGMUN

INPUT POWER SELECTION

[Fig. 1]

[Fig. 2]

Input Connection Remark


AC 200V ①(AC 0V) + ②(AC 200V)

-4-
DONGMUN
AC 220V ①(AC 0V) + ③(AC 220V)
AC 240V ①(AC 0V) + ④(AC 240V)

BOARD kV & mA SELECTION

-5-
DONGMUN

1. BOARD CALIBRATION

1.1 HT CONTROL CALIBRATION

VR1 Oscillator frequency calibration for filament drive


• Measuring Point:U18-3 PIN
• Adjust Frequency:36kHz ± 0.5kHz(55.6±1.6μsec).
• Caution:Measure as 2-PULSE since it is alternating current.

VR2 Filament LAMP DEAD TIME calibration


• Caution: Do not calibrate but just turn to the left (counter clock-wise).

VR3 Small Scale (10mA ~ 80mA) mA FEEDBACK to mAs adjustment with VF converter
• Measuring Point:TP5
• Adjust the VR3 until the TP5 set to 25kHz± 0.5kHz.
• Simple Adjustment: Connect mAs tester to the connector 8,9(mA TEST)of HV tank
and shoot x-ray and adjust mAs.
• Conditions: 40kV 50mA / 10mAs
• Measuring: 10 ± 0.2mAs

VR5 Large Scale(100mA ~ 999mA) mA FEEDBACK to mAs adjustment with VF converter


• Measuring Point:TP5
• Adjust the VR3 until the TP5 set to 2.5kHz ± 0.5kHz.
• Simple Adjustment : Connect mAs tester to the connector 8,9(mA TEST) of HV
tank and shoot x-ray and adjust mAs.
• Conditions: 40kV 50mA / 10mAs
• Measuring: 10 ± 0.2mAs

VR4 Oscillator Frequency Adjustment for kV DRIVE


• Measuring Point:U25-7 PIN.
• Adjust Frequency:40kHz ± 0.5kHz (50±1.2μsec)
• Caution:Measure as 2-PULSE since it is alternating current.

-6-
DONGMUN

1.2 HT CONTROL COMPONENT LAYOUT

TP15
GND

U25

VR6

TP8
kV FB

VR2

TP4
mA FB

VR5 VR3 VR1 U18

-7-
DONGMUN

1.3 FILAMENT DRIVE CALIBRATION

VR1 FILAMENT TRANS DIRVE Power Voltage Calibration


• Measuring Point : TP5, TP6
• Adjusting Point:DC240V±0.1%

1.4 FILAMENT DRIVE COMPONENT LAYOUT


VR1

TP5, TP6

-8-
DONGMUN

1.5 AGING

CONDITIONS SHOOTING TIME SHOOTING INTERVAL


40kV/50mA 0.01sec Two 20~30 Sec.
50kV/50mA 0.01sec Two 20~30 Sec.
60kV/50mA 0.01sec Two 20~30 Sec.
80kV/50mA 0.01sec Two 20~30 Sec.
40kV/50mA 0.1sec Two 50~60 Sec.
50kV/50mA 0.1sec Two 50~60 Sec.
60kV/50mA 0.1sec Two 50~60 Sec.
80kV/50mA 0.1sec Two 50~60 Sec.
90kV/50mA 0.1sec Two 20~30 Sec.
100kV/50mA 0.01sec Two 20~30 Sec.
110kV/50mA 0.01sec Two 20~30 Sec.
120kV/50mA 0.01sec Two 20~30 Sec.
150kV/50mA 0.01sec Two 20~30 Sec.

-9-
DONGMUN

1.6 kV CALIBRATION

Calibration Method: Act in the panel calibration Mode(From C1-00 to 20)


• Setting range:
mA Sec kV Remark
10mA 0.05sec 40kV, 80kV, 120kV, 150kV
12.5mA 0.05sec 40kV, 80kV, 120kV, 150kV
16mA 0.05sec 40kV, 80kV, 120kV, 150kV
20mA 0.05sec 40kV, 80kV, 120kV, 150kV
25mA 0.05sec 40kV, 80kV, 120kV, 150kV
32mA 0.05sec 40kV, 80kV, 120kV, 150kV
40mA 0.05sec 40kV, 80kV, 120kV, 150kV
50mA 0.05sec 40kV, 80kV, 120kV, 150kV
63mA 0.05sec 40kV, 80kV, 120kV, 150kV
80mA 0.05sec 40kV, 80kV, 120kV, 150kV
100mA 0.05sec 40kV, 80kV, 120kV, 150kV
125mA 0.05sec 40kV, 80kV, 120kV, 150kV
160mA 0.05sec 40kV, 80kV, 120kV, 150kV
200mA 0.05sec 40kV, 80kV, 120kV, 150kV 16kW Spec.
250mA 0.05sec 40kV, 80kV, 120kV, 150kV 20kW Spec.
320mA 0.05sec 40kV, 80kV, 120kV, 150kV
400mA 0.05sec 40kV, 80kV, 120kV, 150kV 32kW Spec.
500mA 0.05sec 40kV, 80kV, 120kV, 150kV 40kW Spec.
630mA 0.05sec 40kV, 80kV, 120kV 50kW Spec.
800mA 0.05sec 40kV, 80kV, 100kV
999mA 0.05sec 40kV, 80kV 80kW Spec.

• Measuring Point:HT-CONT TP8:kV FEEDBACK


• Converted Value:1V=33.3 kV
• Adjusting Value:±10% for the setting kV
• Caution: Execute the final test with correct mA value. kV could be changed by the mA.

- 10 -
DONGMUN

1.7 mA CALIBRATION

Calibration Method: Execute in panel calibration mode.(From C1-21to 41)


• Setting Range:
mA Sec kV Remark
10mA 0.05sec 40kV, 80kV, 120kV, 150kV
12.5mA 0.05sec 40kV, 80kV, 120kV, 150kV
16mA 0.05sec 40kV, 80kV, 120kV, 150kV
20mA 0.05sec 40kV, 80kV, 120kV, 150kV
25mA 0.05sec 40kV, 80kV, 120kV, 150kV
32mA 0.05sec 40kV, 80kV, 120kV, 150kV
40mA 0.05sec 40kV, 80kV, 120kV, 150kV
50mA 0.05sec 40kV, 80kV, 120kV, 150kV
63mA 0.05sec 40kV, 80kV, 120kV, 150kV
80mA 0.05sec 40kV, 80kV, 120kV, 150kV
100mA 0.05sec 40kV, 80kV, 120kV, 150kV
125mA 0.05sec 40kV, 80kV, 120kV, 150kV
160mA 0.05sec 40kV, 80kV, 120kV, 150kV
200mA 0.05sec 40kV, 80kV, 120kV, 150kV 16kW Spec.
250mA 0.05sec 40kV, 80kV, 120kV, 150kV 20kW Spec.
320mA 0.05sec 40kV, 80kV, 120kV, 150kV
400mA 0.05sec 40kV, 80kV, 120kV, 150kV 32kW Spec.
500mA 0.05sec 40kV, 80kV, 120kV, 150kV 40kW Spec.
630mA 0.05sec 40kV, 80kV, 120kV 50kW Spec.
800mA 0.05sec 40kV, 80kV, 100kV
999mA 0.05sec 40kV, 80kV 80kW Spec.

• Measuring point:HT-CONT TP4:mA FEEDBACK conversion


(10~80mA : 1V = 10mA, 100~999mA : 1V = 100mA)
• Adjusting value :±10% for the setting kV
• Caution : Execute the mA variation from C1-21 to 41.
In case of overshoot and undershoot、act from C2-21to 41 for the stable mA levels.

- 11 -
DONGMUN

1.8 mA OSCILLOSCOPE WAVE FORM

1) NORMAL CONDITIONS

2) UNDERSHOOT WAVE FORM


Operate in panel calibration mode (From C2-21 to 41). Decrease DATA(PANEL⑤).

- 12 -
DONGMUN

3) OVERSHOOT WAVE FORM


Act in Panel calibration mode (From C2-21 to 41). Increase DATA(PANEL⑤).

- 13 -
DONGMUN

1.9 BUCKY TYPE SELECTION

Select the bucky type by choosing no. 4 of DIP SW 4.


DIP SW 1-4 Bucky Type Remark
On Motor-Type Bucky
Off Solenoid-Type Bucky

1.10 BUCKY O. K. SIGNAL

Select the bucky O. K. signal by choosing no.1 of DIP SW 2.


DIP SW 2-1 Use or Non-use of signal Remark
On Use Bucky O. K. Signal
Off Not use Bucky O. K. Signal

- 14 -
DONGMUN

OPERATION PANEL CALIBATION

- 15 -
DONGMUN

1. OPERATION PANEL DESCRIPTION

22
10
12
16 5
9 3 4 11 1 2 6

23 24

13 14 15 20 17 21 7 8
18 19

1 Power ON switch 13 Setting switch for APR mode


2 Power OFF switch 14 Selection switch of Region and Display
3 KV Display/Setting switch 15 Setting switch of Direction and Display
4 mAs Display/Setting switch 16 Setting switch of a patient size and Display
5 ‘mAs/sec Display/Setting switch 17 Setting switch of AEC mode and Display
6 ‘mAs/sec selection switch 18 BACKUP Release switch
7 Preparation switch for Exposure 19 Setting switch of DENSITY and Display
8 X-ray Exposure switch 20 BUCKY1 & 2 selection switch
9 “OVER” LED (Red) 21 AEC Field Selection and Display
10 “ERROR” LED (Red) 22 Exposure hand switch
11 Ready LED (Green) 23 Small Focus Display
12 X-ray Exposure (Yellow) 24 Large Focus Display

- 16 -
DONGMUN

[1]Power on Switch
Switch on the power supply

[2]Power off Switch


Switch off the power supply

[3]kV Display / Setting Switch


Set the kV for the exposures. The calibrated kV will be displayed on the menu.

[4] mA Display / Setting Switch


Set the mA for the exposures. The calibrated mA will be displayed on the menu.

[5] mA/sec Display / Setting Switch


Set the mA/sec for the exposures. The calibrated mA/sec will be shown on the menu.

[6]mAs/sec selection Switch


Select mAs/sec exposure mode.
• mAs/sec selection「mAs」Mode
•「sec」Mode for the exposure time
Indicate the selected exposure mode on the menu.

[7]Exposure Ready Switch (READY Switch)


When this switch is pushed, the exposure ready action [READY] is executed.

[8]X-ray Exposure Switch(X-RAY Switch)


Press Exposure Ready Switch and when the [11] Exposure ready LED is lighted on, press
the x-ray exposure switch at the same time.

[9]OVER Indication(Red)
The LED is switched on when operation exceeds radiography conditions. The exposure
is prohibited while [the over LED] is lighted on. In case the LED is turned on due to
changed exposure conditions in the APR mode, the memory of condition should not be
saved. Under the AEC mode selection, if radiography time exceeds backup time, it
blinks. Backup release switch cancels over display at the time of AEC mode.

[10]ERROR Indicator(Red)
The LED is turned on when an error occurs.

[11]Ready Condition Indicator(Green)


The LED is turned on when X-ray exposure changes into a possible state.

- 17 -
DONGMUN

[12]X-ray Exposure Indicator


The light is switched on during the exposure.

[13]APR Mode setting switch


This switch sets APR mode as On or Off.
[14]Selection Switch of Parts and Display
It sets the body parts to be radiographed. LED displays the selected location.
[15]Setting switch of Direction and Display.
The direction of radiography is set up. LED displays the selected direction.
[16]Setting switch of a Patient Size and Display and Radiography memory switch
This switch is used for selecting patient size and LED displays the selected size.

If the target region, the radiography direction, and patient size are chosen in the APR mode,
radiography conditions (kV, mA, mAs/sec) shall be set up automatically. Each photographic
condition can be changed and memorized by the Patient Size setting switch (condition
memory switch).

[17]AEC Mode Set-up switch and display


This switch sets AEC mode as On or Off.

[18]BACKUP Release Switch


This switch resets BACKUP and CHECK CODE. Some parts of check code are excluded.

[19] Setting switch of Density and Display


The density level of radiography is set up.

[20] BUCKY1 & 2 Selection Switch and Display


BUCKY1&2 is selected and the selected BUCKY is displayed with LED.

[21] AEC Field Selection Switch and Display


AEC Field Combination Select and display

[22] Hand Switch for exposure


If the 1st step of switch is pressed, the rotor is moved. The filament of X-ray tube is
preheated, then the preparation for X-ray exposure shall performed. At the same
time the ready condition display is switched on. When the 2nd step of switch is
pushed, X-ray is exposed.

[23]Small-focus Display
The light is switched on when the small focus of the tube is selected.

- 18 -
DONGMUN

DIPSW-1 ON (DIPSW-2,3,4 OFF):Calibration Mode


DIPSW-2 ON (DIPSW-1,3,4 OFF):APR Calibration Mode
DIPSW-3 ON (DIPSW-1,2,4 OFF):AEC Calibration Mode
DIPSW-4 ON (DIPSW-1,2,3 OFF):RAM Monitor
DIPSW-5 ON :AEC selection admission
DIPSW-6 ON :Not in use
DIPSW-7 ON :Not in use
DIPSW-8 ON :Select Exposure End Buzzer
DIPSW-1,2,3,4 ON:Counter Mode

[24]Large- focus Display


The light is switched on when the large focus of the tube is selected.

2. DIP SW FUNCTION

Adjust DIP Switch SW32 (8P) located on the back front of control panel and operate the each
calibration.
DIP SW (SW 32: 8P) Function Display

DIP SW-1:ON (DIPSW-2,3,4 OFF) Calibration Mode


Execute Indicated kV levles(KVREF), mA levels (MAREF), set up of each calibration
levels and initialization of calibration data.

DIP SW-2:ON (DIPSW-1,3,4 OFF) APR Calibration Mode


Change APR Data.

DIP SW-3:ON (DIPSW-1,2,4 OFF) AEC Calibration Mode


Start AEC Data setting.

DIP SW-4:ON (DIPSW-1,2,3 OFF) RAM Monitor


Monitoring RAM data of operation Panel.

DIP SW-5:ON Admission of AEC selection


Enable to Select SW-ON:AEC Mode.

- 19 -
DONGMUN
DIP SW-8:ON Select Exposure End Buzzer
SW-OFF: Select Exposure On Buzzer.

DIP SW-1,2,3,4 ON:Counter Mode


Display Frequency of exposure, exposure time and frequency of Error & Check Code
occurrence.

3. RADIOGRAPHY SELECTION IN NORMAL MODE

This equipment is available with following exposure mode.

1) MANUAL EXPOSURE MODE


Execute x-ray exposure by manual settings

2) APR EXPOSURE MODE


Execute X-ray Exposure after setting up radiographic conditions with selection of target
regions, directions and patient size. You can save each exposure conditions and levels.

3.AEC Exposure Mode


Execute the X-ray exposure with Ion Chamber control.
You can calibrate the range with density setting switch.

- 20 -
DONGMUN

3-1.SETTING MANUAL RADIOGRAPHY

1) Select the manual radiography mode


Clear up the conditions by pushing [13] APR setting switch and [17] AEC setting switch when
the APR mode and AEC mode are selected.

2) Select “BUCKY1" or "BUCKY 2"


Push the BUCKY1 [20] (table BUCKY) or BUCKY2 [20] (Stand BUCKY) switch and select the
mode if a radiography using BUCKY should be used.

3) A setup of tube voltage "kV"


Set up the tube voltage with “kV” [3] setting switch.
Setting range: 40kV ~ 125kV or 150kV (1kV step)

4) A set up of tube current “mA”


Set up the tube current with “mA” [4] setting switch.
Setting range : 10、12.5、16、20、25、32、40、50、63、80、100、
125、160、 mA ....(Small Focus)
200、250、320、400、500、630, 800, 999* mA....(Large Focus)

*
Model mA Range Remark
DIG-161 50~200mA 50mA Step
DIG-201 10~250mA
DIG-321 10~400mA
DIG-401A 10~500mA
DIG-501(T) 10~630(800)mA (T) : 3-Phase
DIG-801T 10~999mA 3-Phase

5) Setup of radiography mode (mAs /sec radiography)


Select the radiography mode with mAs/sec [6] selection switch (display). The selected one
lights up. "mAs" radiography makes a radiograph by the current time product. "sec"
radiography makes radiography in set-up time.

6) A setup of mAs value at time of selection “mAs"


mAs value is chosen with the mAs/sec [5] setting switch.
- Setting range:0.1mAs ~ 999mAs

- 21 -
DONGMUN

- Step: 0.1、0.2、0.3、0.4、0.5、0.6、0.8、1.0、1.3、1.6、2.0、2.5、3.2、4.0、
5.0、6.3、8.0、10.0、12.5、16、20、25、32、40、50、63、80、100、125、
160、200、250、320、400、500、630、800、999* mAs

Model mAs Range Remark


DIG-161 0.4~320mAs
DIG-201 0.1~500mAs
DIG-321 0.1~630mAs
DIG-401A 0.1~630mAs
DIG-501(T) 0.1~800mAs
DIG-801T 0.1~999mAs

A setup of sec value at the time of selection “sec. radiography”


Sec Value is choosen with the mAs/sec [5] setting switch.
- Setting range:0.001sec ~ 10sec
- Step: 0.001、0.002、0.003、0.004、0.005、0.006、0.008、0.010、0.013、
0.016、0.020、0.025、0.032、0.040、0.050、0.064、0.080、0.100、
0.125、0.160、0.200、0.250、0.320、0.400、0.500、0.640、0.800、
1.00、1.25、1.60、2.00、2.50、3.20、4 .00、5.00、6.3、8.0 、10sec

Model sec Range Remark


DIG-161 0.003~6 sec
DIG-201 0.001~10 sec
DIG-321 0.001~10 sec
DIG-401A 0.001~10 sec
DIG-501(T) 0.001~10 sec
DIG-801T 0.001~10 sec

- 22 -
DONGMUN

3-2.SETTING APR RADIOGRAPHY

1) Select the APR mode


Set up the APR mode with APR [13] setting switch.
When it is selected, the LED of the switch upper-left part switches on.
2) Select Target region
Set up with Region [14] setting switch.
LED display on a panel moves up and down by pushing up and down switch.
Choose the target region to be taken radiography. The number of parts that can be
chosen out of nine.
3) Selection of the Directions
Set up with [15] Radiography Direction [15] setting switch.
The radiography direction is displayed on the right-hand side of the selected location.
Push the direction setting switches 1 2 3 4, and choose it out of four.
4) Select Patient Size
Patient size is chosen with Size [16] setting switch.
The patient size that can be chosen is out of LL L M S SS BABY.
5) Selection of DENSITY
The set-up value of mAs/sec step is changed by pushing [19] “DENSITY” switch.
.DENSITY:range –1 to -3: Step down
.DENSITY:±0 : No change
.DENSITY:range+1 to +3: Step up
6) Change of Radiography Conditions
The set-up radiography conditions can be changed by the same method as a setup of manual
radiography conditions. BUCKY setup at the time of APR radiography is decided beforehand.
A setting state is expressed as "BUCKY1, 2 selection switch" X-ray exposure will be forbidden
if the light is switched on and change radiography conditions.
7) Memory of the Changed Radiography Conditions
The radiography conditions changed by above 6) clause can be memorized. The memory
method pushes Patient Size setting switch chosen now for about 1 second. If a buzzer
sounds, it is an end of memorizing conditions. Patient Size switch chosen is displayed by
lighting of attached LED. If the patient size switch not chosen is pushed, it will be set as other
radiography conditions and memory will not be carried out. It does not memorize in the
following condition.
(1) OVER display lights up.
(2) DENSITY setting ≠0

- 23 -
DONGMUN

3-3.SETTING AEC RADIOGRAPHY

1) AEC Mode Enable


Select the mode with [17] AEC setting switch.

2) Selection of Field

Select the target region with setting switch. You can choose the field by pushing
switches.
The selected field of lon Chamber will be displated on the LED panel.
You cannot make radiography without selecting the field.

3) Time should be higher than 100msec under AEC mode.

4) You can select AEC 1 or AEC 2 by setting [20]. [21] bucky 1, 2.


It can not be the ready condition if Bucky and field are not selected.
The autual mAs value will be displayed on the [5] mAs / sec LED after AEC exposure
snded.

ERROR AND ERROR CHECK


-Error 29: Occur when there is no feedback from
lon-chamber
CAUTION -Check 10: Occur when there is no enough x-rays for
setting level. Make the radiography again after pushing
back-up switch and rearrange conditions in this case.

- 24 -
DONGMUN

4.RADIOGRAPHY

1) X-ray Radiography using Radiography switch of Control Panel


(1) Push [7] X-ray exposure preparation switch. [11] X-ray exposure ready LED is lighted
up after completing preparation.
(2) After READY LED lighting up, if [8] exposure switch and [7] exposure ready switch are
pressed together, X-ray shall exposed.
(3) During exposure, X-rays Exposure [12] displays (X-RAY) lights up, and the buzzer
turns on.
(4) A check code is displayed, when a finger is lifted from a switch on the way and
exposure is interrupted.

2) X-ray radiography that used Radiography hand-switch


(1) Press the [22] 1st step (preparation switch) of Hand Switch. [11] Ready LED is lighted up
after finishing preparation for exposure.
(2) After ready LED is lighting up, press the 2nd step of Hand Switch in order to expose
x-ray.
(3) [12] x-ray Exposure LED (X-RAY) lights up during the exposure, and the buzzer is
turning on.
(4) If a finger was lifted from a switch on the way, the check code will be displayed and
exposure is interrupted.

The selection of exposure on and off buzzer is operated by the DIP SW-8.
(SW-OFF/ select exposure on buzzer, SW-ON/ select exposure off buzzer)

- 25 -
DONGMUN

5.CALIBRATION MODE (DIPSW1: ON)

* CALIBRATION MODE OUTLINE

[C-1]KVREF&MAREF SETTING (1/2)

Code Description Remark


C-1 __0 [10mA] KVREF Shift Data
C-1 __1 [12.5mA] KVREF Shift Data
C-1 __2 [16mA] KVREF Shift Data
C-1 __3 [20mA] KVREF Shift Data
C-1 __4 [25mA] KVREF Shift Data
C-1 __5 [32mA] KVREF Shift Data
C-1 __6 [40mA] KVREF Shift Data
C-1 __7 [50mA] KVREF Shift Data
C-1 __8 [63mA] KVREF Shift Data
C-1 __9 [80mA] KVREF Shift Data
C-1 _10 [100mA] KVREF Shift Data
C-1 _11 [125mA] KVREF Shift Data

- 26 -
DONGMUN
C-1 _12 [160mA] KVREF Shift Data
C-1 _13 [200mA] KVREF Shift Data
C-1 _14 [250mA] KVREF Shift Data
C-1 _15 [320mA] KVREF Shift Data
C-1 _16 [400mA] KVREF Shift Data
C-1 _17 [500mA] KVREF Shift Data
C-1 _18 [630mA] KVREF Shift Data
C-1 _19 [800mA] KVREF Shift Data
C-1 _20 [999mA] KVREF Shift Data

[C-1]KVREF&MAREF SETTING (2/2)

Code Description Remark


C-1 _21 [10mA] MAREF Shift Data
C-1 _22 [12.5mA] MAREF Shift Data
C-1 _23 [16mA] MAREF Shift Data
C-1 _24 [20mA] MAREF Shift Data
C-1 _25 [25mA] MAREF Shift Data
C-1 _26 [32mA] MAREF Shift Data
C-1 _27 [40mA] MAREF Shift Data
C-1 _28 [50mA] MAREF Shift Data
C-1 _29 [63mA] MAREF Shift Data
C-1 _30 [80mA] MAREF Shift Data
C-1 _31 [100mA] MAREF Shift Data
C-1 _32 [125mA] MAREF Shift Data
C-1 _33 [160mA] MAREF Shift Data
C-1 _34 [200mA] MAREF Shift Data
C-1 _35 [250mA] MAREF Shift Data

- 27 -
DONGMUN
C-1 _36 [320mA] MAREF Shift Data
C-1 _37 [400mA] MAREF Shift Data
C-1 _38 [500mA] MAREF Shift Data
C-1 _39 [630mA] MAREF Shift Data
C-1 _40 [800mA] MAREF Shift Data
C-1 _41 [999mA] MAREF Shift Data

[C-2] SETTING EACH CALIBRATION LEVELS (1/4)

Code Description Remark


C-2 __0 Not in Use
C-2 __1 Not in Use
C-2 __2 Not in Use
C-2 __3 Not in Use
C-2 __4 Not in Use
C-2 __5 Not in Use
C-2 __6 Not in Use
C-2 __7 Not in Use
C-2 __8 Not in Use
C-2 __9 Not in Use
C-2 _10 Not in Use
C-2 _11 Not in Use
C-2 _12 Not in Use
C-2 _13 Not in Use
C-2 _14 Not in Use

- 28 -
DONGMUN
C-2 _15 Not in Use
C-2 _16 Not in Use
C-2 _17 Not in Use
C-2 _18 Not in Use
C-2 _19 Not in Use
C-2 _20 Not in Use

[C-2] SETTING EACH CALIBRATION LEVELS (2/4)

Code Description Remark


C-2 _21 [10mA] PREP_MAREF Calibration Data
C-2 _22 [12.5mA] PREP_MAREF Calibration Data
C-2 _23 [16mA] PREP_MAREF Calibration Data
C-2 _24 [20mA] PREP_MAREF Calibration Data
C-2 _25 [25mA] PREP_MAREF Calibration Data
C-2 _26 [32mA] PREP_MAREF Calibration Data
C-2 _27 [40mA] PREP_MAREF Calibration Data
C-2 _28 [50mA] PREP_MAREF Calibration Data
C-2 _29 [63mA] PREP_MAREF Calibration Data
C-2 _30 [80mA] PREP_MAREF Calibration Data
C-2 _31 [100mA] PREP_MAREF Calibration Data
C-2 _32 [125mA] PREP_MAREF Calibration Data
C-2 _33 [160mA] PREP_MAREF Calibration Data
C-2 _34 [200mA] PREP_MAREF Calibration Data
C-2 _35 [250mA] PREP_MAREF Calibration Data

- 29 -
DONGMUN
C-2 _36 [320mA] PREP_MAREF Calibration Data
C-2 _37 [400mA] PREP_MAREF Calibration Data
C-2 _38 [500mA] PREP_MAREF Calibration Data
C-2 _39 [630mA] PREP_MAREF Calibration Data
C-2 _40 [800mA] PREP_MAREF Calibration Data
C-2 _41 [999mA] PREP_MAREF Calibration Data

[C-2] SETTING EACH CALIBRATION LEVELS (3/4)

Code Description Remark


C-2 _42 Not in Use
C-2 _43 kV Feedback Monitor setting Data
C-2 _44 BUCKY1(Table)Selection Data
C-2 _45 BUCKY2(Stand)Selection Data
C-2 _46 BUCKY1 Exposure DELAY-TIME control Data
C-2 _47 BUCKY2 Exposure DELAY-TIME control Data
C-2 _48 Not in Use
C-2 _49 [AEC Mode] AEC1 Type Selection
C-2 _50 [AEC Mode] AEC2 Type Selection
C-2 _51 Not in Use
C-2 _52 Not in Use
C-2 _53 Not in Use
C-2 _54 Not in Use
C-2 _55 Not in Use
C-2 _56 Not in Use

- 30 -
DONGMUN
C-2 _57 Not in Use
C-2 _58 Not in Use
C-2 _59 Not in Use
C-2 _60 [AEC Mode] DENSITY=-3 Shift Data
C-2 _61 [AEC Mode] DENSITY=-2 Shift Data
C-2 _62 [AEC Mode] DENSITY=-1 Shift Data
C-2 _63 [AEC Mode] DENSITY=+1 Shift Data
C-2 _64 [AEC Mode] DENSITY=+2 Shift Data
C-2 _65 [AEC Mode] DENSITY=+3 Shift Data
C-2 _66 Not in Use
C-2 _67 Not in Use
C-2 _68 Not in Use
C-2 _69 Not in Use
C-2 _70 [AEC Mode] AEC1 Reference Level Data

[C-2] SETTING EACH CALIBRATION LEVELS (4/4)

Code Description Remark


C-2 _71 [AEC Mode] AEC2 Reference Level Data
C-2 _72 Not in Use
C-2 _73 Not in Use
C-2 _74 Not in Use
C-2 _75 Not in Use
C-2 _76 Not in Use
C-2 _77 Not in Use
C-2 _78 Not in Use
C-2 _79 Not in Use
C-2 _80 Tube & High-Voltage Cable Select Data
C-2 _81 kV Range Selection Data

- 31 -
DONGMUN

[C-3] INITIALIZATION OF CALIBRATION DATA

Code Description Remark


C-3 __ 0 Initialize of all Data

- 32 -
DONGMUN

5. DISPLAY OF PANEL

5-1. DISPLAY OF PANEL IN CALIBRATION MODE

[1] Power on Switch

[2] Power off Switch

[3] kV display/setting switch


Calibration Mode
kV

・ Calibration Mode Display [C-1 ⇔ C-2 ⇔ C-3]


・ Mode Selection Switch

[4]mA display/setting switch

- 33 -
DONGMUN
Calibration Mode
mA
・ Calibration numbers display
・ Select the numbers

[5] mAs/sec display/setting switch


mA s Calibration Mode C-1] [C-2]
sec
・ Calibration Data Display
・ Data Change Switch

[6] mAs/sec display/setting switch


Calibration Mode [C-3]

mAs
・ Data Format
sec

[18] BACKUP Cancellation Switch


Calibration Mode [C-1]
•Monitor Calibration ⇔ Monitor Conversion Switch

- 34 -
DONGMUN

5-2. KVREF/MAREF SETTING

• Push the [2]power off switch and turn the power off.
Press 1st switch of DIPSW(SW32 : 8P)and turn it on.

• Push the power on [1] switch and turn the power on.
Control panel is displayed as a calibration mode.

• Select the calibration mode with[3] kV setting switch.


The selected calibration mode is appeared on kV display on the panel.

C-1:KVREF&MAREF SETTING

• Select the numbers in the calibration mode with [4] mA setting switch.
The selected number in calibration mode is appeared on mA display on the panel.

*THE RANGE FOR SELECTION

__0~_20, _21~_41

The calibration data is shown on the [5] mAs/sec display after the calibration mode and
numbers are selected.

* CALIBRATION DATA SETTING RANGE

-20∼ 00 ∼+20

- 35 -
DONGMUN

* Execute the x-ray exposure with following methods after the data is selected in
KVREF/MAREF setting mode.

[18] Execute the conversion of monitor(the monitor after calibration)with backup release
switch. During the monitor recording, it is possible to make x-ray exposures. The setting
of mA during calibration is fixed for the time. APR•AEC Mode is not in option this time.

Calibtation Mode Calibtation Mode mA in the exposure


(KVREF setting) (MAREF setting) Monitor (Fixed)
C-1__0 C-1_21 10 mA
C-1__1 C-1_22 12.5 mA
C-1__2 C-1_23 16 mA
C-1__3 C-1_24 20 mA
C-1__4 C-1_25 25 mA
C-1__5 C-1_26 32 mA
C-1__6 C-1_27 40 mA
C-1__7 C-1_28 50 mA
C-1__8 C-1_29 63 mA
C-1__9 C-1_30 80 mA
C-1_10 C-1_31 100 mA
C-1_11 C-1_32 125 mA
C-1_12 C-1_33 160 mA
C-1_13 C-1_34 200 mA
C-1_14 C-1_35 250 mA
C-1_15 C-1_36 320 mA
C-1_16 C-1_37 400 mA
C-1_17 C-1_38 500 mA
C-1_18 C-1_39 630mA
C-1_19 C-1_40 800mA
C-1_20 C-1_41 999mA

As the mA setting is fixed, make exposure with selecting other conditions such as
kV•mAs/sec and check the movement of calibration data. When the exposure condition
exceeds the capacity, the ‘over’ LED is lighted on. While the light is turned on, exposure of
x-ray is forbidden. In case of rearrangement after x-ray exposure, operate the data setting
again with back up release switch and return the exposure monitor to calibration monitor.
• How to end the calibration mode
Turn it off with power off switch and the position will return to the DIP SW mode before the
start of Calibration Mode.

- 36 -
DONGMUN

5-3. KVREF/MAREF CALIBRATION DETAILS

C-1__0 : [10mA : Small Focus] KVREF Shift Data


Shift the KVREF value from the total base when the mA is 10.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
W It is possible to change on the screen with [18] backup switch(10 mA fixed).
•

C-1__1 : [12.5mA : small focus] KVREF Shift Data


Shift the KVREF value from the total base when the mA is 12.5.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
W
• It is possible to change on the screen with [18] backup switch(12.5 mA fixed).

C-1__2 : [16mA : small focus] KVREF Shift Data


Shift the KVREF value from the total base when the mA is 16.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
W It is possible to change on the screen with [18] backup switch(16 mA fixed).
•

C-1__3 : [20mA : small focus] KVREF Shift Data


Shift the KVREF value from the total base when the mA is 20.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
W It is possible to change on the screen with [18] backup switch(20 mA fixed).
•

C-1__4 : [25mA : small focus] KVREF Shift Data

Shift the KVREF value from the total base when the mA is 25.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-:
W KVREF goes down
• It is possible to change on the screen with [18] backup switch(25 mA fixed).

- 37 -
DONGMUN

C-1__5 : [32mA : small focus] KVREF Shift Data

Shift the KVREF value from the total base when the mA is 32.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-:
W KVREF goes down
• It is possible to change on the screen with [18] backup switch(32 mA fixed).

C-1__6 : [40mA : small focus] KVREF Shift Data

Shift the KVREF value from the total base when the mA is 40.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-:
W KVREF goes down
• It is possible to change on the screen with [18] backup switch(40 mA fixed).

C-1__7 : [50mA : small focus] KVREF Shift Data

Shift the KVREF value from the total base when the mA is 50.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-:
W KVREF goes down
• It is possible to change on the screen with [18] backup switch(50 mA fixed).
C-1__8 : [63mA : small focus] KVREF Shift Data
Shift the KVREF value from the total base when the mA is 63.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
• It is possible to change on the screen with [18] backup switch(63 mA fixed).

C-1__9 : [80mA : small focus] KVREF Shift Data

Shift the KVREF value from the total base when the mA is 80.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
• It is possible to change on the screen with [18] backup switch(80 mA fixed).

- 38 -
DONGMUN
C-1_10 : [100mA : small focus] KVREF Shift Data
Shift the KVREF value from the total base when the mA is 100.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
W It is possible to change on the screen with [18] backup switch(100 mA fixed).
•

C-1_11 : [125mA : small focus] KVREF Shift Data


Shift the KVREF value from the total base when the mA is 125.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
W It is possible to change on the screen with [18] backup switch(125 mA fixed).
•

C-1_12 : [160mA : small focus] KVREF Shift Data


Shift the KVREF value from the total base when the mA is 160.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
W It is possible to change on the screen with [18] backup switch(160 mA fixed).
•

C-1_13 : [200mA : large focus] KVREF Shift Data


Shift the KVREF value from the total base when the mA is 200.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
W It is possible to change on the screen with [18] backup switch(200 mA fixed).
•

C-1_14 : [250mA : large focus] KVREF Shift Data


Shift the KVREF value from the total base when the mA is 250.
Data range is from –20 to +20 and starting value is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
W
• It is possible to change on the screen with [18] backup switch(250 mA fixed).

C-1_15 : [320mA : large focus] KVREF Shift Data

- 39 -
DONGMUN
Shift the KVREF value from the total base when the mA is 320.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
W It is possible to change on the screen with [18] backup switch(320 mA fixed).
•

C-1_16 : [400mA : large focus] KVREF Shift Data


Shift the KVREF value from the total base when the mA is 400.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
W It is possible to change on the screen with [18] backup switch(400 mA fixed).
•

C-1_17 : [500mA : large focus] KVREF Shift Data


Shift the KVREF value from the total base when the mA is 500.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
W It is possible to change on the screen with [18] backup switch(500 mA fixed).
•

C-1_18 : [630mA : large focus] KVREF Shift Data


Shift the KVREF value from the total base when the mA is 630.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
W It is possible to change on the screen with [18] backup switch(630 mA fixed).
•

C-1_19 : [800mA : large focus] KVREF Shift Data


Shift the KVREF value from the total base when the mA is 800.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
W It is possible to change on the screen with [18] backup switch(800 mA fixed).
•

C-1_20 : [999mA : large focus] KVREF Shift Data


Shift the KVREF value from the total base when the mA is 999.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down - 40 -
• It is possible to change on the screen with [18] backup switch(999 mA fixed).
DONGMUN

C-1_21 : [10mA : small focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 10.
Data range is from –20 to +20 and starting point is 00.(Defalult)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(10 mA fixed).
•

C-1_22 : [12.5mA : small focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 12.5.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(12.5 mA fixed).
•

C-1_23 : [16mA : small focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 16.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
• It is possible to change on the screen with [18] backup switch(16 mA fixed).

C-1_24 : [20mA : small focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 20.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(20 mA fixed).
•

C-1_25 : [25mA : small focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 25.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(25 mA fixed).
•

- 41 -
DONGMUN

C-1_26 : [32mA : small focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 32.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(32 mA fixed).
•

C-1_27 : [40mA : small focus] MAREF Shift Data

Shift the MAREF value from the total base when the mA is 40.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W
• It is possible to change on the screen with [18] backup switch(40 mA fixed).

C-1_28 : [50mA : small focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 50.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(50 mA fixed).
•

C-1_29 : [63mA : small focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 63.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(63 mA fixed).
•

C-1_30 : [80mA : small focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 80.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(80 mA fixed).
•

C-1_31 : [100mA : small focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 100.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down - 42 -
• It is possible to change on the screen with [18] backup switch(100 mA fixed).
DONGMUN

C-1_32 : [125mA : small focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 125.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(125 mA fixed).
•

C-1_33 : [160mA : small focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 160.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(160 mA fixed).
•

C-1_34 : [200mA : large focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 200.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(200 mA fixed).
•

C-1_35 : [250mA : large focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 250.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(250 mA fixed).
•

C-1_36 : [320mA : large focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 320.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(320 mA fixed).
•

- 43 -
DONGMUN

C-1_37 : [400mA : large focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 400.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(400 mA fixed).
•

C-1_38 : [500mA : large focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 500.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(500 mA fixed).
•

C-1_39 : [630mA : large focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 630.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(630 mA fixed).
•

C-1_40 : [800mA : large focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 800.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(800 mA fixed).
•

C-1_41 : [999mA : large focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 999.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(999 mA fixed).
•

- 44 -
DONGMUN

5-4. GUIDE TO INDIVIDUAL CALIBRATION DETAILS

C-2_21 : [10mA : small focus] PREP/MAREF Calibration Data

Set the PREP/MAREF Calibration data in case the mA is 10.


Data range is from 000 to 010 and starting point is 002.
• In case of Data is 002, mAs –2 value shall be output when the exposure is ready.

C-2_22 : [12.5mA : small focus] PREP/MAREF Calibration Data

Set the PREP/MAREF Calibration data in case the mA is 12.5.


Data range is from 000 to 010 and starting point is 002.
• In case of Data is 002, mAs –2 value shall be output when the exposure is ready.

C-2_23 : [16mA : small focus] PREP/MAREF Calibration Data


Set the PREP/MAREF Calibration data in case the mA is 16.
Data range is from 000 to 010 and starting point is 002.
• In case of Data is 002, mAs –2 value shall be output when the exposure is ready.

C-2_24 : [20mA : small focus] PREP/MAREF Calibration Data

Set the PREP/MAREF Calibration data in case the mA is 20.


Data range is from 000 to 010 and starting point is 002.
• In case of Data is 002, mAs –2 value shall be output when the exposure is ready.

C-2_25 : [25mA : small focus] PREP/MAREF Calibration Data

Set the PREP/MAREF Calibration data in case the mA is 25.


Data range is from 000 to 010 and starting point is 002.
• In case of Data is 002, mAs –2 value shall be output when the exposure is ready.

C-2_26 : [32mA : small focus] PREP/MAREF Calibration Data

Set the PREP/MAREF Calibration data in case the mA is 32.


Data range is from 000 to 010 and starting point is 002.
• In case of Data is 002, mAs –2 value shall be output when the exposure is ready.

- 45 -
DONGMUN

C-2_27 : [40mA : small focus] PREP/MAREF Calibration Data

Set the PREP/MAREF Calibration data in case the mA is 40.


Data range is from 000 to 010 and starting point is 002.
• In case of Data is 002, mAs –2 value shall be output when the exposure is ready.

C-2_28 : [50mA : small focus] PREP/MAREF Calibration Data

Set the PREP/MAREF Calibration data in case the mA is 50.


Data range is from 000 to 010 and starting point is 002.
• In case of Data is 002, mAs –2 value shall be output when the exposure is ready.

C-2_29 : [63mA : small focus] PREP/MAREF Calibration Data

Set the PREP/MAREF Calibration data in case the mA is 63.


Data range is from 000 to 010 and starting point is 002.
• In case of Data is 002, mAs –2 value shall be output when the exposure is ready.

C-2_30 : [80mA : small focus] PREP/MAREF Calibration Data

Set the PREP/MAREF Calibration data in case the mA is 80.


Data range is from 000 to 010 and starting point is 002.
• In case of Data is 002, mAs –2 value shall be output when the exposure is ready.

C-2_31 : [100mA : small focus] PREP/MAREF Calibration Data

Set the PREP/MAREF Calibration data in case the mA is 100.


Data range is from 000 to 010 and starting point is 003.
• In case of Data is 003, mAs –3 value shall be output when the exposure is ready.

C-2_32 : [125mA : small focus] PREP/MAREF Calibration Data

Set the PREP/MAREF Calibration data in case the mA is 125.


Data range is from 000 to 010 and starting point is 003.
• In case of Data is 003, mAs –3 value shall be output when the exposure is ready.

C-2_33 : [160mA : small focus] PREP/MAREF Calibration Data

Shift the MAREF value from the total base when the mA is 160.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up - 46 -
Data=-: MAREF goes down
• It is possible to change on the screen with [18] backup switch(160 mA fixed).
DONGMUN

C-2_34 : [200mA : large focus] PREP/MAREF Calibration Data

Set the PREP/MAREF Calibration data in case the mA is 200.


Data range is from 000 to 010 and starting point is 005.
• In case of Data is 005, mAs –5 value shall be output when the exposure is ready.

C-2_35 : [250mA :large focus] PREP/MAREF Calibration Data

Set the PREP/MAREF Calibration data in case the mA is 250.


Data range is from 000 to 010 and starting point is 005.
• In case of Data is 005, mAs –5 value shall be output when the exposure is ready.

C-2_36 : [320mA : large focus] PREP/MAREF Calibration Data

Set the PREP/MAREF Calibration data in case the mA is 320.


Data range is from 000 to 010 and starting point is 005.
• In case of Data is 005, mAs –5 value shall be output when the exposure is ready.

C-2_37 : [400mA : large focus] PREP/MAREF Calibration Data

Set the PREP/MAREF Calibration data in case the mA is 400.


Data range is from 000 to 010 and starting point is 005.
• In case of Data is 005, mAs –5 value shall be output when the exposure is ready.

C-2_38 : [500mA :large focus] PREP/MAREF Calibration Data

Set the PREP/MAREF Calibration data in case the mA is 500.


Data range is from 000 to 010 and starting point is 005.
• In case of Data is 005, mAs –5 value shall be output when the exposure is ready.

- 47 -
DONGMUN
C-2_39 : [630mA :large focus] PREP/MAREF Calibration Data

Set the PREP/MAREF Calibration data in case the mA is 630.


Data range is from 000 to 010 and starting point is 005.
• In case of Data is 005, mAs –5 value shall be output when the exposure is ready.

C-2_40 : [800mA :large focus] PREP/MAREF Calibration Data

Set the PREP/MAREF Calibration data in case the mA is 800.


Data range is from 000 to 010 and starting point is 005.
• In case of Data is 005, mAs –5 value shall be output when the exposure is ready.

C-2_41 : [999mA :large focus] PREP/MAREF Calibration Data

Set the PREP/MAREF Calibration data in case the mA is 999.


Data range is from 000 to 010 and starting point is 005.
• In case of Data is 005, mAs –5 value shall be output when the exposure is ready.

C-2_43 : kV Feedback Monitoring Data

It sets the levels, which monitors kV Feedback in case of exposure.


Data range is from 001 to 070 and starting point is 070. [Monitoring 70%]

C-2_44 : Bucky 1(Table) Selection Data

Data range is from 001 to 002 and starting point is 000.


000=Prohibit Bucky 1(Table) Selection
001=Permit Bucky 1(Table) Selection-control delay time
002= Permit Bucky 1(Table) Selection- control input sign for bucky movement check

C-2_45 : Bucky 2(Stand) Selection Data


Data range is from 000 to 001 and starting point is 000.
000=Prohibit Bucky 2(stand) Selection
001=Permit Bucky 2(stand) Selection-control delay time
002= Permit Bucky 2(stand) Selection- control input sign for bucky movement check

C-2_46 : Bucky 1(Table) Exposure delay time control data

Use in case C-2 __44 Bucky 1(Table) selection Data delay time control.
Data range is from 000 to 200 and starting point is 020
When data is 020, delay time is 200msec.
- 48 -
• Operation in Bucky 1(Table) selection
Magnet for Bucky 1(Table) is called as ON under ready-up condition.
Turn the magnet off with x-ray signal input and the action after delay time is called as
HV-Enable on.
DONGMUN

C-2_47 : Bucky 2(stand) Exposure delay time control data

Use in case C-2 __45 Bucky 2(stand) selection Data delay time control.
Data range is from 000 to 200 and starting point is 020
When data is 020, delay time is 200msec.
• Operation in Bucky 1(Table) selection
Magnet for Bucky 2(table) is called as ON under ready-up condition.
Turn the magnet off with x-ray signal input and the action after delay time is called as
HV-Enable on.

C-2_49 : [AEC Mode] AEC 1 Type Selection Data


AEC 1 Type Selection Data, Data range is from 000 to 002 and starting point is 000.
000= AEC 1 –OFF : [AEC1 : Prohibited for selection]
001= AEC 2 –ON : [AEC1 : Photo control selection]
002= AEC 2 –ON : [AEC1 : Iron chamber control selection]

C-2_50 : [AEC Mode] AEC 2 Type Selection Data


AEC 2 Type Selection Data, Data range is from 000 to 002 and starting point is 000.
000= AEC 2 –OFF : [AEC2 : Prohibited for selection]
001= AEC 2 –ON : [AEC2 : Photo control selection]
002= AEC 2 –ON : [AEC2 : Iron chamber control selection]

C-2_60 : [AEC Mode] Density =-3, Shift Data


Set the level of density as –3.
Data range is from 050 to 150 and starting point is 070.
In case data is 070, VREF setting value is 70% of the original value.

C-2_61 : [AEC Mode] Density =-2, Shift Data


Set the level of density as –2.
Data range is from 050 to 150 and starting point is 080.
In case data is 080, VREF setting value is 80% of the original data.

C-2_62 : [AEC Mode] Density =-1, Shift Data


Set the level of density as –1.
Data range is from 050 to 150 and starting point is 090.
In case data is 090, VREF setting value is 90% of the original data.
- 49 -
DONGMUN

C-2_63 : [AEC Mode] Density =+1, Shift Data


Set the level of density as =+1.
Data range is from 050 to 150 and starting point is 110.
In case data is 110, VREF setting value is 110% of the original data.

C-2_64 : [AEC Mode] Density =+2, Shift Data


Set the level of density as =+2.
Data range is from 050 to 150 and starting point is 120.
In case data is 120, VREF setting value is 120% of the original data.

C-2_65 : [AEC Mode] Density =+3, Shift Data


Set the level of density as =+3.
Data range is from 050 to 150 and starting point is 130.
In case data is 130, VREF setting value is 130% of the original data.

C-2_70 : [AEC Mode] AEC1 Reference Level Data


Set the level AEC1 Reference Level.
Data range is from 000 to 255(hex) and starting point is 128(2.5V).
In case the data is 026, AEC Reference Level is 0.5V.
In case the data is 051, AEC Reference Level is 1.0V.
In case the data is 077, AEC Reference Level is 1.5V.
In case the data is 102, AEC Reference Level is 2.0V.
In case the data is 153, AEC Reference Level is 3.0V.
In case the data is 179, AEC Reference Level is 3.5V.
In case the data is 204, AEC Reference Level is 4.0V.
In case the data is 230, AEC Reference Level is 4.5V.
In case: the
C-2_71 data
[AEC is 255,
Mode] AEC1AEC Reference
Reference Level
Level is 5.0V.
Data
Set the level AEC2 Reference Level.
Data range is from 000 to 255(hex) and starting point is 128(2.5V).
In case the data is 026, AEC Reference Level is 0.5V.
In case the data is 051, AEC Reference Level is 1.0V.
In case the data is 077, AEC Reference Level is 1.5V.
In case the data is 102, AEC Reference Level is 2.0V.
In case the data is 153, AEC Reference Level is 3.0V.
In case the data is 179, AEC Reference Level is 3.5V.
In case the data is 204, AEC Reference Level is 4.0V.
In case the data is 230, AEC Reference Level is 4.5V.
In case the data is 255, AEC Reference Level is 5.0V.

C-2_80 : Tube & High Voltage Cable Select Data

- 50 -
DONGMUN

Select the Tube & High-Voltage Cable.


Data range is from 000 to 017 and starting point is 1(7239, 8m).
In case the data is 000, Tube is Toshiba E7239X, and 6m Cable.
In case the data is 001, Tube is Toshiba E7239X, and 8m Cable.
In case the data is 002, Tube is Toshiba E7239X, and 10m Cable.
In case the data is 003, Tube is Toshiba E7239X, and 12m Cable.
In case the data is 004, Tube is Toshiba E7239X, and 15m Cable.
In case the data is 005, Tube is Toshiba E7239X, and 18m Cable.
In case the data is 006, Tube is Toshiba E7252X, and 6m Cable.
In case the data is 007, Tube is Toshiba E7252X, and 8m Cable.
In case the data is 008, Tube is Toshiba E7252X, and 10m Cable.
In case the data is 009, Tube is Toshiba E7252X, and 12m Cable.
In case the data is 010, Tube is Toshiba E7252X, and 15m Cable.
In case the data is 011, Tube is Toshiba E7252X, and 18m Cable.
In case the data is 012, Tube is Varian RAD-14, and 6m Cable.
In case the data is 013, Tube is Varian RAD-14, and 8m Cable.
In case the data is 014, Tube is Varian RAD-14, and 10m Cable.
In case the data is 015, Tube is Varian RAD-14, and 12m Cable.
In case the data is 016, Tube is Varian RAD-14, and 15m Cable.
In case the data is 017, Tube is Varian RAD-14, and 18m Cable.

C-2_81 : kV Range Selection Data

Select the kV Range Data.


Data range is from 000 to 001 and starting point is 000(125kV).
In case the data is 000, kV Range is 40~125kV.
In case the data is 001, kV Range is 40~150kV.
- 51 -
DONGMUN

5-5. INITIALIZATION OF CALIBRATION DATA

• Push the [2]power off switch and turn the power off.
Press 1st switch of DIPSW(SW32 : 8P)and turn it on.

• Push the [1] power on switch and turn the power on.
Control panel is shown as a calibration mode.

• Select the calibration mode with[3] kV setting switch.


The selected calibration mode appears on the kV display on the panel.

C-3:Initiation of Calibration Data


• Execute the initialization of calibration data after selecting calibration mode and numbers.
Press the [6]mAs/sec switch for a second. When the check buzzer sounds, the data
initialization is completed.

• How to end the calibration mode


Turn it off with power off switch and the position will return to the DIP SW mode before the
start of calibration Mode.

5-6. DETAILS OF CALIBRATION DATA INITIALIZATION

C-3__0 : Initialization of the Whole Data


•(C-1__0~C-1_39)The initialization of KVREF & MAREF calibration data
•(C-2__0~C-2_811)The initialization of each data setting
•Initialization of AEC data
•Initialization of APR radiography data

Press the [6]mAs/sec switch for a second. When the check buzzer sounds, the data
initialization is completed.

- 52 -
DONGMUN

5-7. APR CALIBRATION MODE(DIP SW2:ON)

16 5
9 3 4 1 2

- 53 -
DONGMUN

17
21
13 14 18 19 20
15

1 Power ON switch 15 Setting switch of Direction and Display


2 Power OFF switch 16 Setting switch of a patient size and Display
3 kV setting switch display 17 Setting switch of AEC mode and Display
4 mA setting switch display 18 Release switch for BACKUP
5 ‘mAs/sec setting switch display 19 Setting switch of DENSITY and Display
9 ‘OVER’ LED (RED) 18 Release switch for BACKUP
13 APR mode setting switch 20 BUCKY1 & 2 selection and Display
14 Selection switch of region and Display 21 AEC Field Selection and Display

APR CALIBRATION MODE SETTING

(1)Notice about APR Calibration Mode

• The operation of x-ray Exposure is prohibited under the APR Calibration Mode.
Only the changes of APR radiography conditions are possible.
• The data is saved in memory area of APR Calibration Mode.
The radiography conditions could be called out from work area.
In this area, rearrangement of APR radiography condition is available.
When you complete the calibration work, save the changed APR data in the memory area
by pushing [17]APR Data memory switch.

- 54 -
DONGMUN

(2)Selection of APR Calibration Mode

• Push the [2]power off switch and turn the power off.
Press 1st switch of DIPSW(SW 32 : 8P)and turn it on.

• Push the power on [1] switch and turn the power on.
Control panel is displayed as a APR Calibration Mode.

- 55 -
DONGMUN

(3) Calibration for APR mode Radiography conditions 1

Operate the data calibration after the selection of target regions, exposure directions and
patient size. It saves the data memory as a whole unlike under the normal APR mode

• Loading radiography conditions (memory area → work area)


Push the[13]APR data uploading switch. (The check buzzer sounds)
• Select radiography regions
Select using [14] region switches. Nine options are available.
• Select radiography directions
Select using [15]radiography direction switch. You can choose out of four directions.
• Choose patient size
Select using [16]patient size switch. You can choose out of 6 sizes.
After the exposure regions, directions and size are chosen, the exposure conditions
should be set up. The mAs condition can not be selected.
• Changing radiography conditions
The calibrated radiography conditions(kV/mA/sec)are changed by setting switch.
In case of rearrangement, please be careful that the over LED is not lighted.
• Selection of BUCKY(Table/Stand)
When the bucky calibration is needed under APR mode, you can use [20]BUCKY1 or
BUCKY2 selection switch. The setting of bucky in APR mode is only possible under APR
calibration mode.(When the [C-2_44, C-2_45] mode is selective about bucky)
• When the changes about radiography condition is not possible.
When the density setting is not 0, rearrange it as 0.
• Memory of radiography conditions(work area→memory area)

[17]Push the APR Data memory switch. (The check buzzer sounds)
*The buzzer does not sound when the conditions can not be saved.
• When the OVER sign is included in the APR radiography conditions, the numbers 1-9
will be appeared on the kV, mA and sec display. The appeared numbers represents the
regions for exposure. Then, change the over sign on display and push the [17]APR
Data memory switch again.
• When the density is not 0, reset the density =0 and push the APR Data memory
switch.

- 56 -
DONGMUN

(4)Calibration for APR mode Radiography conditions 2

Change the sec setting data in the selected radiography area.


Add or subtract certain values from selected sec data about density setting.
Operate the memory of the whole data after the completion of calibration work.

• Loading radiography conditions (memory area → work area)


Push the[13]APR Data uploading switch. (The check buzzer sounds)
• Exposure location selections
[14]Select with target region switches. Nine options are available.
• Calibration of step value in sec setting
Select the [19]sec setting Shift switch. Operate the sec setting
(4directions×6Size=24sets)at the same time. The check buzzer sounds when the

[18]sec setting Shift switch is executed.


DENSITY=-3(Flash sign) Step value of sec setting should be -3.
DENSITY=-2( Flash sign ) Step value of sec setting should be -2.
DENSITY=-1( Flash sign ) Step value of sec setting should be -1.
DENSITY=0 *not in operation
DENSITY=+1( Flash sign ) Step value of sec setting should be +1.
DENSITY=+2( Flash sign ) Step value of sec setting should be +2.
DENSITY=+3( Flash sign ) Step value of sec setting should be +3.

• Memory of radiography conditions (work area → memory area)


Push the[17]APR Data memory switch. (The check buzzer sounds)

*In case the conditions can not be saved. (The check buzzer does not sound)
• When the OVER sign is included in the APR radiography conditions, the numbers from
1-9 will be appeared on the kV, mA and sec display. The appeared numbers represents
the regions for exposure. Then, change the over sign on display and push the [17]APR
Data memory switch again.
• When the density is not 0, reset the density =0 and push the APR Data memory switch.

• How to end the APR calibration mode


Turn it off with power off switch and the position will return to the DIP SW mode before the
start of Calibration Mode.

- 57 -
DONGMUN

5-8. AEC CALIBRATION MODE

16 5
3 4 1 2

17
14 15

1 Power ON switch 15 Setting switch of Direction and Display


2 Power OFF switch 16 Setting switch of a patient size and
Display
3 Tube kV setting / display switch 17 Setting switch of AEC mode and Display
4 Tube mA setting / display swich
5 ‘mAs/sec setting switch
14 ‘Selection switch of Region and
Display

- 58 -
DONGMUN

The Method of Setting AEC calibration Mode

1) The cautions under AEC calibration mode


-The x-ray exposure under AEC calibration mode is prohibited.
Only change the AEC data.
-Set the AEC calibration mode (DIPSW-1).
-Only the lon-chamber control is available with AEC.
Refer to the method of connection with attached AEC connection data.

2)Selection of AEC calibration mode


-Push the [2] power off switch and make the power off.
Only push the 1st switch of DIPSW (SW32 : 8P) and select the calibration mode.

-Push the [1] power on switch and make the power on.
The control panel will be changed into calibration mode.

- 59 -
DONGMUN

3)METHOD FOR AEC DATA SETTING

• AEC reference data setting


1) Arrange the setting value of C-2_70 : AEC1 Reference Level Data with reference
to
HT-Controller A/D Reference Table and calibrate the AEC level.
(The initial default value is set as 128(2.5V))
2) If you arrange the density with AEC level data, the AEC level will be calibrated in
accordance with % of density for the changed level data.
EX)
AEC Level Data : 128 (2.5V) AEC Level Data : 153 (3V)
Density % DEC Level(V) Density % DEC Level(V)
-3 70 1.75V -3 70 2.1V

-2 80 2V -2 80 2.4V
LEVEL
-1 90 2.25V -1 90 2.7V
Data
0 100 128 2.5V 0 100 153 3V
change
1 110 2.75V 1 110 3.3v
2 120 3V 2 120 3.6V
3 130 3.25V 3 130 3.9V

• AEC Density data setting

1) You can arrange the AEC level with % shift for Density setting by arranging the
C-2_60 ~ 65 : AEC Level Shift Data.
2) The data of AEC level shift will change the related AEC Reference Level Data (%)
under density 0.

EX)
AEC Level Data : 128 (2.5V) → AEC Level Data : 153 (3V)
Density % DEC Level(V) Shift Density Data DEC Level(V)
-3 70 70 1.75V Data -3 25 25 0.62V
-2 80 80 2V change -2 50 50 1.25V
-1 90 90 2.25V -1 75 75 1.87V
0 100 2.5V 0 100 2.5V

- 60 -
DONGMUN
1 110 110 2.75V 1 125 125 3.12v
2 120 120 3V 2 150 150 3.75V
3 130 130 3.25V 3 175 175 4.37V

• CAUTION FOR AEC LEVEL SHIFT


Do not exceed the data of density-2 than the data of density-1.

The size of data should be set higher or lower consecutively.

EX)

Default “Wrong” “Correct”


Density Data Densit Data Density Data
y
-3 70 -3 70 -3 85
-2 80 -2 90 -2 90
-1 90 -1 80 -1 95
0 0 0
1 110 1 130 1 105
2 120 2 120 2 110
3 130 3 110 3 115

4) How to end the AEC calibration mode


• Turn off the device with Power off switch
Return to the condition of DIP SW before the start of calibration Mode.

- 61 -
DONGMUN

5-9. COUNTER MORE (DIPSW1, 2, 3, 4 : ON)

5 1 2
4

17

- 62 -
DONGMUN

COUNTER MODE CALIBRATION

(1)Counter Mode Display


Display following counter values.
• Accumulated Frequency of X-ray exposures
• Accumulated time of X-ray exposures
• Accumulated Frequency of Error code
• Accumulated Frequency of Check code

(2)Counter Mode Selection


• Push the [2]power off switch and turn the power off.
Press 1•2•3•4 switch of DIPSW(SW32 : 8P)and turn it on.

• Push the power on [1] switch and turn the power on.
Control panel appears as a Counter Mode.

(3)Selection of Counter Mode

1 Power ON switch 5 ‘mAs/sec Setting switch


2 Power OFF switch 17 Setting switch of AEC mode and Display
3 Counter Mode
Error Code & Check Code
Display switch
4 mA setting switch
• Push the[15]APR direction setting switch(1•2•3•4)and select counter mode menu.

A. Switch: Accumulated Frequency of X-ray exposures


Each sign of[3][4][5]are as following under the exposure frequency display
000 kV 000 mA 00 sec

- 63 -
DONGMUN
[3]kV Display:Freqency
[4]mA Display:Frequency
[5]sec Display:Blank
Range:000000∼999999
(In case exceeding the maximum value : 000000)

- 64 -
DONGMUN

B. Switch: Accumulated Period of X-ray exposures


Each sign of[3][4][5]are as following under the exposure frequency time display
0 kV 00 mA 00 sec
[3]kV Display:Time [Hour]
[4]mA Display:Time [Minute]
[5]sec Display: Time [seconds]
Range:0 Hr. 00 Min. 00 Sec. ≈999 Hr. 59 Min. 59 Sec.
(In case exceeding the maximum value :0 Hr. 00 Min. 00 Sec. )

C. Switch: Accumulated Frequency of Error code


Each sign of[3][4][5]are as following under the Error code frequency display
Err kV 01 mA 0 sec
[3]kV Display, [4]mA Display :Error code set up
[5]sec Display:Frequency of error
Selection of error code is set by kV setting switch.
Range:Err01 ∼ Err30
Frequency :0 ∼9999, In case exceeding the maximum value :0

D. Switch: Accumulated Frequency of Check code


Each sign of[3][4][5]are as following under the check code frequency display
Err kV 01 mA 0 sec
[3]kV Display, [4]mA Display :Error code set up
[5]sec Display:Frequency of error
Selection of error code is set by kV setting switch.
Range:Ch-01 ∼ Ch-o9
Frequency :0 ∼9999, In case exceeding the maximum value :0

• How to end the Counter calibration mode


Turn it off with power off switch and the position will return to the DIP SW mode before the
start of Counter Mode.

- 65 -
DONGMUN

6. ERROR CODE AND CHECK CODE

When problems occurred during the use of equipment, they should be marked on the [3]kV,
[4]mA setting section. As well, the [10] error LED will be lighted on with the occurrence of
error.

• Error Code
The error code is displayed when problems, such as abnormalities, occur to equipment.
While the error code is displayed, radiography operation is forbidden for safety.
A power supply is re-switched off and on, then an error code is cleared. If the problem of
equipment is not solved at the time of a power-supply re-switched on, the display is lighted
on again.

• Check code
Check codes are warning displays when an error is on operation of equipment unlike an error
code. While the check code is displayed, radiography operation is forbidden for safety. Check
code is cleared when the backup release switch is switched on. Please refer to the check code
list for the case the clearance method is different.

- 66 -
DONGMUN

ERROR CODE LIST

Error Code Description


Err 01 CHARGE MONITOR ERROR
• When the power voltage drops below AC160V and DC150V.
Err 02 TIME OVER ERROR AT STANDBY(Serial Communication Fail Error)
• Standby position is not cleared after the 10 sec. at the operation panel
power on
Err 03 READY OUT SIGNAL ERROR AT STANDBY
• Defects of Operation panel or HT control board
Err 04 X-RAY OUT SIGNAL ERROR AT STANDBY
• Defects of Operation panel or HT control board
Err 05 PANEL READY SWITCH INPUT ERROR AT STANDBY
• When the operation panel ready switch input is ‘on’ position.
Err 06 PANEL X-RAY SWITCH INPUT ERROR AT STANDBY
• When the operation panel ready switch input is ‘on’ position.
Err 07 HAND S/W READY INPUT ERROR AT STANDBY
• When the hand s/w ready input is ‘on’ position.
Err 08 HAND S/W X-RAY INPUT ERROR AT STANDBY
• When the hand s/w x-ray input is ‘on’ position.
Err 09 ROTOR CURRENT ERROR AT NON-EXPOSURE POSITION
• TP18 (RTR_I_FB) of HT Control Board is higher than 1V at non-exposure
action
Err 10 FILLAMENT CURRENT ERROR AT NON-EXPOSURE POSITION
• TP2 (FIL_I_FB) of HT Control Board is higher than 1.5V at non-exposure
action
Err 11 kV FEEDBACK CURRENT ERROR AT NON-EXPOSURE POSITION
• TP8 (kV_FB) of HT Control Board is higher than 1V TP8 (33.3kV) at
non-exposure action
Err 12 mA FEEDBACK CURRENT ERROR AT NON-EXPOSURE POSITION
• TP4 (mA_FB) 10-80mA setting of HT Control Board is higher than o.5V
(5mA), In case of 100-500mA setting of HT Control Board is higher than
o.1V (11mA) TP8 (33.3kV) at non-exposure action
Err 13 ROTOR CURRENT ERROR AT EXPOSURE POSITION
• TP18 (RTR_I_FB) of HT Control Board is lower than 3V at exposure ready
action after the 0.8 sec. With ready switch input.
Err 14 FILLAMENT CURRENT ERROR AT EXPOSURE POSITION
• TP2 (FIL_I_FB) of HT Control Board is lower than 1.5V at exposure ready
action after the 1.5 sec. With ready switch input.
Err 15 kV FEEDBACK CURRENT ERROR AT NON-EXPOSURE POSITION
• TP8 (kV_FB) of HT Control Board is higher than 1V TP8 (33.3kV) at
non-exposure action
Err 16 mA FEEDBACK CURRENT ERROR AT EXPOSURE POSITION
• Exposure Ready Condition
TP4 (mA_FB) 10-80mA setting of HT Control Board is lower than o.5V
(5mA), In case of 100-500mA setting of HT Control Board is lower than
o.1V (11mA) after the 3 msec. with kV Feedback signal.
• During Exposure
TP4 (mA_FB) 10-80mA setting of HT Control Board is lower than o.5V
(5mA), In case of 100-500mA setting of HT Control Board is lower than
o.1V (11mA) .

- 67 -
DONGMUN
Err 17 Not Defined

Err 18 X-RAY TIME OVER ERROR AT SEC MODE EXPOSURE


• X-ray end signal is not reached to the panel from HT exceeding the
setting time at the sec mode exposure
Err 19 X-RAY TIME OVER ERROR AT mAs MODE EXPOSURE
• X-ray end signal is not reached to the panel from HT exceeding the
setting time calculated by the mA×(1/2) formula at the mAs mode
exposure.
Err 20 X-RAY TIME OVER ERROR AT AEC MODE EXPOSURE
• X-ray end signal is not reached to the panel from HT exceeding the
setting time calculated by backup-timer at the AEC mode exposure.
Err 21 HV INTERLOCK ERROR
• Unlink the HV trans connection
Err 22 IGBT FAULT ERROR
• HV Tans is under overloading condition about mA (350A).
Err 23 kV FEEDBACK OVER CURRENT ERROR AT EXPOSURE POSITION
• TP8 (kV_FB) of HT Control Board is higher than 20kV during the
exposure.
Err 24 mA FEEDBACK OVER CURRENT ERROR AT EXPOSURE POSITION
• TP8 (mA_FB) of HT Control Board is higher than [mA setting value plus
50%] during the exposure.
Err 25 NO ZERO CROSS ERROR
• Impossible to detect the frequency signal.
Err 26 ZERO CROSS OVER ERROR
• Frequency signal detection over.
Err 27 BUCKY 1 MOTION ERROR
• No signal input exceeding 3 sec. during the bucky 1 motion.
• Execute error check for the calibration mode [C-2_44]=002 setting.
Err 28 BUCKY 2 MOTION ERROR
• No signal input exceeding 3 sec. during the bucky 2 motions.
• Execute error check for the calibration mode [C-2_45]=002 setting.
Err 29 AEC A/D Data-In ERROR
• When Voltage Level is below 0.5V(AEC Board TP1) from Ion Chamber
within 100ms.
Err 30 PHOTO TIMER SIGNAL RESET ERROR
• Photo timer input signal does not reset at the AEC exposure ready
position (ready switch input).
• The input value of HT control board is higher than 1.0V.
Err 31 I/F_BOARD-IN ERROR (HT-CONT)
Err 32 AEC_BOARD-IN ERROR (HT-CONT)
Err 33 SMALL-FOCUS SELECT ERROR (HT-CONT)
Err 34 LARGE-FOCUS SELECT ERROR (HT-CONT)

- 68 -
DONGMUN

CHECK CODE LIST

CODE Description
Ch_01 TUBE THERMOSTAT OPERATION
• The temperature of the slim and long shaped light bulb is higher than 80C.
• Thermo switch is in open position.
• Release the check code at Thermo switch off condition.
Ch_02 When another switch is pushed during exposure
• Release the check code with back up reset switch.
Ch_03 When PREP. Switch on OP PANEL is detached during exposure
• Release the check code with back up reset switch.
Ch_04 When X-RAY switch on OP PANEL is detached during exposure
• Release the check code with back up reset switch.
Ch_05 When a hand switch (PREP) is detached during exposure
• Release the check code with back up reset switch.
Ch_06 When a hand switch (X-RAY) is detached during exposure
• Release the check code with back up reset switch.
Ch_07 Shortage of power-supply capacity for exposure
• The power voltage drop is below AC160V and DC 150V.
• Release the check code with input of power.
Ch_08 Simultaneous exposure prohibition
• Other x-ray system is executed simultaneously when the exposure ready switch
was pushed.
• Release the check code by off the power of other x-ray system.
Ch_09 Warning to X-rays room door opening
• The door of x-ray room was opened when the exposure ready switch was pushed.
• Stop the exposure of x-ray in this case and the check code is released when the
door is closed.
Ch_10 AEC Voltage Level Check
• After setting AEC Voltage Level, when voltage level which is input from Ion
Chamber when do Exposure does not reach and is ended by time and mAs that is
established to AEC End Signal .

- 69 -
DONGMUN

7. TROUBLE SHOOTING GUIDE

1. System & PCB Check Point (1/2)


Error No. Parts Check Point
Error 01 IGBT PC1(PC817)
LED LD1, LD2
CN404 Connection
U2(MIC4584)
POWER Q2, Q3(MTD3055EL)
CN208 Connection
MAIN R101(60W20Ω)
BR1(VUO150-16NO7)
SW1(Toggle Switch Position)
Error 02 POWER +12V
CN210 Connection
HT-Cont. CN103 Connection
CN104 Connection
MAIN J1 Connection
Error 03 OP Ready S/W
Error 04 OP X-Ray S/W
Error 05 MAIN J1 Connection
HT-Cont. U14(TLP521-4)
U10(74HC14)
CN103 Connection
Error 06 MAIN J1 Connection
HT-Cont. U14(TLP521-4)
U10(74HC14)
CN103 Connection
Error 07 OP Hand S/W Connection
MAIN J1 Connection
HT-Cont. U14(TLP521-4)
U10(74HC14)
CN103 Connection
Error 08 OP Hand S/W Connection
MAIN J1 Connection
HT-Cont. U14(TLP521-4)
U10(74HC14)
CN103 Connection

- 70 -
DONGMUN

1. System & PCB Check Point (2/2)


Error No. Parts Check Point
Error 09 - -
Error 10 - -
Error 11 - -
Error 12 - -
Error 13 HT-Cont. U29(VM48)
I/F Rotor Signal Connection
POWER F2(5A) Fuse
Error 14 FILAMENT RY1(CH2-MD12)
R11(5W2Ω)
Q2, Q5(IRFP40B)
Error 15 - -
Error 16 - -
Error 17 Not In Use
Error 18 HT-Cont. HT-Cont. PCB
Error 19 HT-Cont. U22(VFC32)
Error 20 HT-Cont. U22(VFC32)
Error 21 HT-Cont. CN107 Connection
Error 22 - -
Error 23 - -
Error 24 - -
Error 25 HT-Cont. HT-Cont. PCB
Error 26 HT-Cont. HT-Cont. PCB
Error 27 I/F CN501 Connection
HT-Cont. CN101 Connection
Error 28 I/F CN501 Connection
HT-Cont. CN101 Connection
Error 29 AEC/ABC CN603, 604 Connection
Error 30 HT-Cont. U12(74HC14)
Error 31 HT-Cont. U12(74HC14)
Error 32 HT-Cont. U12(74HC14)
Error 33 FILAMENT RY1(CH2-MD12)
Error 34 FILAMENT RY1(CH2-MD12)

- 71 -
DONGMUN

2. Check Point
Check Code No. Parts Check Point
Check 01 I/F Connection
Tube Tube Temperature
Check 02 OP Panel S/W Input
Check 03 OP Panel S/W Input
Check 04 OP Panel S/W Input
Check 05 Hand S/W Hand S/W Input
Check 06 Hand S/W Hand S/W Input
Check 07 MAIN AC Input
High-Voltage Capacitor Voltage Level
(DC 300~310V)
Check 08 I/F I/F Connection
Check 09 I/F I/F Connection
Check 10 Ion Chamber Ion Chamber Input Level

- 72 -
DONGMUN

8. APR INITIAL TRANSMISSION DATA

[C-3]It means the APR Initial Data transmitted by the initialization of calibration data.

*If possible, initial data is recommended to be 1 type, and in case you need several kinds of
initial data, different APR conditions about ROM should be prepared and correspond.

• APR Radiography Conditions


This data is transmitted in case of calibration data initialization[C-3__0][C-3__2].
Please refer to the APR radiography conditions Table for the contents of Data.

• Composition of APR Radiography Condition Data (9×4×6=216 sets)

LEVEL-1 LEVEL-2 LEVEL-3


Selections of Target Selections of Directions Selections of
region(9sets) (4sets) Patient size
(6sets)
Head 1 LL
Cervical Vertebral 2 L
Chest 3 M
Thoracic Vertebrae 4 S
Rib SS
Abdomen BABY
Vertebra

- 73 -
DONGMUN
Upper Limbs
Lower Limbs

9. HT-CONTROLLER A/S REFERENCE

* HT-Controller CPU Analog-to-Digital Input Value (1/5)


DEC A/D Input (A/D) X 2 DEC A/D Input (A/D) X 2
1 0.02 0.039 31 0.608 1.216
2 0.039 0.078 32 0.627 1.255
3 0.059 0.118 33 0.647 1.294
4 0.078 0.157 34 0.667 1.333
5 0.098 0.196 35 0.686 1.373
6 0.118 0.235 36 0.706 1.412
7 0.137 0.275 37 0.725 1.451
8 0.157 0.314 38 0.745 1.49
9 0.176 0.353 39 0.765 1.529
10 0.196 0.392 40 0.784 1.569
11 0.216 0.431 41 0.804 1.608
12 0.235 0.471 42 0.824 1.647
13 0.255 0.51 43 0.843 1.686
14 0.275 0.549 44 0.863 1.725
15 0.294 0.588 45 0.882 1.765
16 0.314 0.627 46 0.902 1.804
17 0.333 0.667 47 0.922 1.843
18 0.353 0.706 48 0.941 1.882

- 74 -
DONGMUN
19 0.373 0.745 49 0.961 1.922
20 0.392 0.784 50 0.98 1.961
21 0.412 0.824 51 1 2
22 0.431 0.863 52 1.02 2.039
23 0.451 0.902 53 1.039 2.078
24 0.471 0.941 54 1.059 2.118
25 0.49 0.98 55 1.078 2.157
26 0.51 1.02 56 1.098 2.196
27 0.529 1.059 57 1.118 2.235
28 0.549 1.098 58 1.137 2.275
29 0.569 1.137 59 1.157 2.314
30 0.588 1.176 60 1.176 2.353

* HT-Controller CPU Analog-to-Digital Input Value (2/5)


DEC A/D Input (A/D) X 2 DEC A/D Input (A/D) X 2
61 1.196 2.392 91 1.784 3.569
62 1.216 2.431 92 1.804 3.608
63 1.235 2.471 93 1.824 3.647
64 1.255 2.51 94 1.843 3.686
65 1.275 2.549 95 1.863 3.725
66 1.294 2.588 96 1.882 3.765
67 1.314 2.627 97 1.902 3.804
68 1.333 2.667 98 1.922 3.843
69 1.353 2.706 99 1.941 3.882
70 1.373 2.745 100 1.961 3.922
71 1.392 2.784 101 1.98 3.961
72 1.412 2.824 102 2 4
73 1.431 2.863 103 2.02 4.039
74 1.451 2.902 104 2.039 0.078
75 1.471 2.941 105 2.059 4.118
76 1.49 2.98 106 2.078 4.157
77 1.51 3.02 107 2.098 4.196
78 1.529 3.059 108 2.118 4.235

- 75 -
DONGMUN
79 1.549 3.098 109 2.137 4.275
80 1.569 3.137 110 2.157 4.314
81 1.588 3.176 111 2.176 4.353
82 1.608 3.216 112 2.196 4.392
83 1.627 3.255 113 2.216 4.431
84 1.647 3.294 114 2.235 4.471
85 1.667 3.333 115 2.255 4.51
86 1.686 3.373 116 2.275 4.549
87 1.706 3.412 117 2.294 4.588
88 1.725 3.451 118 2.314 4.627
89 1.745 3.49 119 2.333 4.667
90 1.765 3.529 120 2.353 4.706

* HT-Controller CPU Analog-to-Digital Input Value (3/5)


DEC A/D Input (A/D) X 2 DEC A/D Input (A/D) X 2
121 2.373 4.745 151 2.961 5.922
122 2.392 4.784 152 2.98 5.961
123 2.412 4.824 153 3 6
124 2.431 4.863 154 3.02 6.039
125 2.451 4.902 155 3.039 6.078
126 2.471 4.941 156 3.059 6.118
127 2.49 4.98 157 3.078 6.157
128 2.51 5.02 158 3.098 6.196
129 2.529 5.059 159 3.118 6.235
130 2.549 5.098 160 3.137 6.275
131 2.569 5.137 161 3.157 6.314
132 2.588 5.176 162 3.176 6.353
133 2.608 5.216 163 3.196 6.392
134 2.627 5.255 164 3.216 6.431
135 2.647 5.294 165 3.235 6.471
136 2.667 5.333 166 3.255 6.51
137 2.686 5.373 167 3.275 6.549

- 76 -
DONGMUN
138 2.706 5.412 168 3.294 6.588
139 2.725 5.451 169 3.314 6.627
140 2.745 5.49 170 3.333 6.667
141 2.765 5.529 171 3.353 6.706
142 2.784 5.569 172 3.373 6.745
143 2.804 5.608 173 3.392 6.784
144 2.824 5.647 174 3.412 6.824
145 2.843 5.686 175 3.431 6.863
146 2.863 5.725 176 3.451 6.902
147 2.882 5.765 177 3.471 6.941
148 2.902 5.804 178 3.49 6.98
149 2.922 5.843 179 3.51 7.02
150 2.941 5.882 180 3.529 7.059

* HT-Controller CPU Analog-to-Digital Input Value (4/5)


DEC A/D Input (A/D) X 2 DEC A/D Input (A/D) X 2
181 3.549 7.098 211 4.137 8.275
182 3.569 7.137 212 4.157 8.314
183 3.588 7.176 213 4.176 8.353
184 3.608 7.216 214 4.196 8.392
185 3.627 7.255 215 4.216 8.431
186 3.647 7.294 216 4.235 8.471
187 3.667 7.333 217 4.255 8.51
188 3.686 7.373 218 4.275 8.549
189 3.706 7.412 219 4.294 8.588
190 3.725 7.451 220 4.314 8.627
191 3.745 7.49 221 4.333 8.667
192 3.765 7.529 222 4.353 8.706
193 3.784 7.569 223 4.373 8.745
194 3.804 7.608 224 4.392 8.784
195 3.824 7.647 225 4.412 8.824
196 3.843 7.686 226 4.431 8.863

- 77 -
DONGMUN
197 3.863 7.725 227 4.451 8.902
198 3.882 7.765 228 4.471 8.941
199 3.902 7.804 229 4.49 8.98
200 3.922 7.843 230 4.51 9.02
201 3.941 7.882 231 4.529 9.059
202 3.961 7.922 232 4.549 9.098
203 3.98 7.961 233 4.569 9.137
204 4 8 234 4.588 9.176
205 4.02 8.039 235 4.608 9.216
206 4.039 8.078 236 4.627 9.255
207 4.059 8.118 237 4.647 9.294
208 4.078 8.157 238 4.667 9.333
209 4.098 8.196 239 4.686 9.373
210 4.118 8.235 240 4.706 9.412

* HT-Controller CPU Analog-to-Digital Input Value (5/5)


DEC A/D Input (A/D) X 2 DEC A/D Input (A/D) X 2
241 4.725 9.451 249 4.882 9.765
241 4.745 9.49 250 4.902 9.804
243 4.765 9.529 251 4.922 9.843
244 4.784 9.569 252 4.941 9.882
245 4.804 9.608 253 4.961 9.922
246 4.824 9.647 254 4.98 9.961
247 4.843 9.686 255 5 10
248 4.863 9.725

- 78 -
DONGMUN

WIRE CONNECTION

- 79 -
DONGMUN

- 80 -
DONGMUN

1. HV TANK CONNECTION

T2 T1 OIL INJECTION

CATHOD ANODE

GND

CORE

* CORE
No. Symbol Description Remark
1 mA TP+ mA Test Point (+) mA & mAs Meter
2 mA TP- mA Test Point (-) Connection
3 F-L Large Focus of Filament
4 F-COM Common
5 F-S Small Focus of Filament
6 mA +F/B mA Feedback (+)
7 mA –F/B mA Feedback (-)
8 kV +F/B kV Feedback (+)
9 INT Tank Interlock
10 GND Ground(Earth)

- 81 -
DONGMUN

2. I/F BODARD CONNECTION


TB502

CN503

CN502

TB501

TB503
TB504

CN501 CN505 CN504

- 82 -
DONGMUN

TB 504 (BR-944C) TB 503 (BR-944C)


PIN I/O Signal PIN I/O Signal
1 O BUCKY1 SPLY 1 I DOOR
2 O BUCKY1 DR 2 I GND
3 O BUCKY1 EXP1 3 I OTHER DEVICE 1
4 O BUCKY1 EXP2 4 I GND
5 O SPARE OUT1-1 5 I OTHER DEVICE 2
6 I SPARE OUT1-2 6 I GND
7 I SPARE OUT2-1 7 I SPARE IN 1
8 O SPARE OUT2-2 8 I GND
9 O BUCKY2 SPLY 9 I SPARE IN 2
10 O BUCKY2 DR 10 I GND
11 O BUCKY2 EXP1 11 O LIGHT IN USE-1
12 O BUCKY2 EXP2 12 O LIGHT IN USE-2
13 I BUCYK2 OK 13 O WARNING 1-1
14 I BUCYK2 OK (GND) 14 O WARNING 1-2
15 O BUCYK1 OK 15 O WARNING 1-3
16 O BUCYK1 OK (GND) 16 O WARNING 2-1
17 O +24V 17 O WARNING 2-2
18 O 24V GND 18 O WARNING 2-3
TB 502 (BR-944C) TB 501 (BR-944C)
PIN I/O Signal PIN I/O Signal
1 O ROTOR 1 1 25 VAC
2 O ROTOR 2 2 16 VAC
3 O ROTOR 3 3 0 VAC
4 O GND 4 220 VAC
5 I TUBE THERMAL 5 110 VAC
6 I TUBE THERMAL 6 0 VAC
7 O 25 VAC(10A)
8 O 16 VAC(10A)
9 O 0 VAC
10 O 220 VAC
11 O 110 VAC
12 O 0 VAC

- 83 -
DONGMUN

- 84 -
DONGMUN

- 85 -