UD150B-40/L-40
Service Manual
Shimadzu Corporation
This service manual is made by Shimadzu Medical Systems Nishinihon co. on request
from Medical Systems Division of Shimadzu Corporation. And Quality Control Section
of Medical Systems Division gives approval for publication.
This manual’s contents are for service engineer. Don’t hand in customers.
Revison History
Revision Date Contents
First Revision 2005.Jun.20
Prolouge
This service manual explain about “Xray High Voltage Generator
UD150B-40/V-40/L-40”.Please refer and use this in time of your installation, inspection
and repair.
There are relevant documents to this manual, as follow:
(4)Turn off power when take off and connect a connector and a Board.
Device will be broken when take off and connect a Board and connector while turning
on the power.
supplementary explanation
In case of B -40/V-40/L-40, it is able to measure VIN signal(CP14) with actual exposure
by oscilloscope and D.V.M.
However, likely other Generator, it is unable to measure F signal(CP12).
Contents
Chapter 4. Troubleshooting
4.1 Measured kV Over
4.2 Radio Over Current and filament checking
4.3 Measured kV Under
4.4 Cautions on installation, Adjustment, troubleshooting.
4.5 Preparation for test operation
Line Voltage Over Power supply voltage is over the S/W detection
tolerance level. UD-CONT2002 (5/15)
LV(CP29)
Charge Volt ERR Primary smoothing capacitor’s S/W detection
charging voltage is an error. • UD-CONT2002 (4,5/15)
VP1,VP2(CP36,CP31)
‚ TERMINAL2002(1/3)
VC+, VC0, VC-
(CP5,CP3,CP9)
Measured kV Over Measured tube voltage is over UD-CONT2002 (9/15)
the tolerance level. TKV(CP32)
Starter ERR Starter has a problem.
I.F. Over There is a problem with the S/W detection
filament heating current. UD CONT2002(10/15)
IFV_F,S,L (CP4,8,23)
Power Down Control circuit’s power supply S/W detection
voltage is an error. UD CONT2002(11/15)
-15VOK (No CP)
H.V.T. Not Connected There is a problem with the S/W detection
connection to the high-voltage UD CONT2002(1/15)
transformer. HVCK (No CP)
Arcing Trouble Electric discharge occurred H/W detection
repeatedly. UD CONT2002(9/15)
TGD1,TGD2(No CP)
Inverter thermal Error IGBT inverter is over 95°C. INVERTER2002(1/1)
J4 connector pin 1, pin 2
(2) Radiography condition messages
Message Meaning
mAs Over The mAs value is over 800 mAs.
mAs too small The mAs value is smaller than 0.5 mAs. Or, tube
current is smaller than the minimum value at the mAs
setting method.
Emission Over Over the emission property limit.
Generator Load Over Over the instrument’s rating.
mAs/time Over Radiography time is over 10 seconds due to a high mAs
value. Or, radiography time is over the tolerance level.
2 control range Over Over the settable range at the mAs setting method.
(3) Messages concerning X-ray tube’s stored heat quantity
1-1
Message Meaning
HU OVER Prodicted If the heat quantity predicted at the set X-ray
condition is added to the current stored heat quantity,
the total heat quantity will exceed the tolerance level.
HU Full Stored Stored heat quantity has reached to the maximum
tolerance level.
Thermal Over X-ray tube container’s temperature is over the
tolerance level.
***Notice!!***
Take actions for messages described in (2) and (3) following the operation manual.
(4) Messages displayed when applying current (messages displayed right after turning
On the power)
Message Meaning
Panel Battery WARNING Operation panel’s data storage battery needs to be
changed soon. (An error is displayed at a figure smaller
than about 3.75V.)
Cabinet Battery WARNING Operation panel’s data storage battery needs to be
changed soon. (An error is displayed at a figure smaller
than about 3.75V.)
Panel Battery EMPTY Operation panel’s data storage battery is empty. (This
message is displayed when a memory error occurs, not
by the lowering of the voltage.)
Cabinet Battery EMPTY Control cabinet’s data storage battery is empty. (This
message is displayed when a memory error occurs, not
by the lowering of the voltage.)
***Notice!!***
Messages above may be displayed even when battery voltage is normal after
initialization is carried out on the control cabinet or operation panel. In this case, turn
OFF/ON the power for confirmation.
1-2
2.Outline of UD150B-40/V-40/L-40
2.1 Comparison between B-30 series and B-40 series
All P.C.B. are redesigned at New UD series (UD150B-40/V-40/L-40). So, it isn’t
compatible with old X-ray high voltage generator(B-30 series etc.).
In case of B-30 series, Boards are parted into by the function. However, almost of
functions in B-40 series are processed by CPU on UD CONT2002.
2-1
2.2 kV control
Case of B-30 series, NEXSC output “setting tube voltage signal(RKV)”.
I/O CONT received the signal(RKV) and convert digital to analog signal(PKV).
O n the other hand, B-40 series output PKV directly from CPU, and convert digital
to
analog on UD CONT2002.
2.3 mA control
In case of B-30/L-30 series, NEXSC PCB output signals, RMA , FVR and LMT.
These signals are exchanged from digital signals to analog ,and use to dispose of
mA control.
B-40/L-40 series have these analog signals too, but does not use these signals for
mA control. CPU have FVR data matrix. CPU output FVR data from this matrix to
FPGA(M29) directly ,on conditions of input digital data. And FPGA hand over
Gate-pulse to mA POWER2002.
Concerning feedback motion, CPU read TMA and IFV value ,and output pulse for
heating filament directly that include feedback in case of B-40/L-40 series.
B-30/L-30 series feedback TMA and IFV on analog circuit.
2-2
Figure B. is “Block diagram of S/W control”.
Figure C. is “TmA measure circuit”.
Q14 Q23
SM
mA POWER-2002
A/D
CONVERTER
PCB
FPGA
(gate array
IFV
TMA
CPU
GAIN
MAM adjustment
UD CONT−2002 PCB
MA MA0
mA DETECTOR-2002
PCB
2-3
Figure A. is “Block diagram of H/W control”.
CPU
START
CPU compe nsate IFST
TmA by CPU.
value.
signal.(Q14,Q23)
A/D CONVERTER
value. filament.
2-4
Figure B. is “Block diagram of S/W control”.
X 線管に流れる X 線管管電流を測定する回路です。
図 D に示すように高圧回路の中性点カソード電流を管電流として計測します。
検出抵抗部は高圧発生器の MA DETECTOR-2002 基板に取り付けられています。
撮影時には、検出抵抗はリレーが働いて 2.5Ωと 500Ωの並列接続となり、透
K2
視時には、リレーが切れて 500Ωとなります。 K3
K2 K3 検出抵抗 TAM(CP27) 備 考
L OPEN CLOSE 2.5ohm 5V/1000mA 使用
M CLOSE OPEN 50 ohm 5V/50mA 現在未使用
H OPEN OPEN 500 ohm 5V/5mA 現在未使用
検出電流はオペアンプに入力され UD CONT−2002 の TMA に現れます。
透視時、約 1V/1mA、撮影時、約 1V/200mA の関係をもちます。
56.2Ω
2-5
H o w e v e r the number of retry is till three times. If arc happen 4 times,
exposure will be stopped with FAULT message.
LINE VOLTAGE
………Power source voltage is over more than 30% or is lower less than
20% ,this error message appear and stop Xray.
RmA OVER
………CPU measured MAM and if the value is over more than “RmA +
200mA”,
this message will appear.
2-6
UD CONT2002
LCD
FPGA KVRF
Set kV data and CPU Set sec is handed
set sec over to the timer.
PKV
KVT
D/A PKV Comparison
Convertor circuit.(94%)
TKV
KV_ADJ PCB
construction function
2-7
Figure E. is “Block diagram of Console Panel control”.
2-8
3.各基板等の機能説明
3.explanations of each PCB
3.1
3.1Main circuit
主回路…基本的には B-30 と同じだが、電源部に NoiseFilter が標準設置されてお
Fundamentally this circuit is almost same with B-30. Main different parts is
asり、IGBT
follows, 素子の種類・個数が変更されている。
3.2Noise-Filter
基板説明 is include as standard.
And the number and kind of IGBT are changed.
(1) Explanation
3.2 UD-CONT2002……CPU
of PCB 基板(メイン制御基板)
(1) UD CONT-2002
撮影条件設定,OLP,管電圧制御,管電流制御(透視/撮影),フィラメント加熱制御(S/W
This board is CPU board , and control as follows,
制御で加熱インバータの駆動パルス(PWM)を発生),撮影タイマ,メモリ回路,EMC
Setting X-ray condition , Over load protect(OLP) , kV control ,
mA制御,アドレス制御,I/O,D/A,A/D
control(Fluoroscopy and Radiography), Filament-heating control , sec control ,
EMC circuit , Address control , Memory circuit , I/O circuit , D/A and A/D circuit.
UD CONT2002 のisCheckPinMAP
Check-Pin-Map drawing at theは据説 2 section
second 章にも記載があります。
of “B-40 Installation Manual” too.
CP21 FB+
CP24
CP23 IFV_ L KVFB+
CP22 FB
VR1 PULSE ADJ CP26 KVFB
CP35 CHGREF
CP29 LV
CP33 IFV
CP31 VP2
VR9 FBRES L
CP36 VP1
VR10 mAS
CP41 PKV
CP38 MASCLK
CP37 VS
CP40 VN
CP39 mA−IN
VR7 MAM ADJ
3-1
(2) INVERTER
(2) INVERTER-CONT2002
CONT-2002
IGBT 回路(インバータ+スナバ回路+ゲート駆動回路),(大電流対応)
This board is IGBT circuit.(Inverter + snubber circuit + gate-drive circuit )
従来の
New IGBT
IGBT is とは違うタイプの IGBT 素子が使われており、
using that one is include 2 個 1 型になっています。
two IGBT. But the frequency is same
ただし、周波数は
a s B -30.(50kHz) B-30 と同じ 50kHz です。
CP1,3,5,7 [G]Gate
CP2,4,6,8 [E]Emitter
3-2
(3) kV ADJ-2002
(3) KV
This ADJ2002
board have volume resistance to agree with actual kV and measured kV.
kVADJ
And then 基板の時定数を調整することで高圧トランス内にある
kVFB signal is made by this board and CPU board. kV 検出回路の時定
数と比例した kVFB 信号を取出すための基板。(高圧トランス内検出回路のコンデ
ンサが個体差をもつため高圧トランスも個々で kVFB 信号にバラツキがあり、こ
Attention.
H . V. の個体差を吸収するための回路)
Trans and kV ADJ board are a pair. So at case of exchanging UD CONT-2002,
*注記*
use kV ADJ board that had already installed.
UD CONT2002 基板交換時
CP2(CO)
CP4 T1
CP1(CL)
GND
T2
J4 コネクタ T3
T4
J1 コネクタ
CP5 ‐ CP3 +
*注記*
Notes
一般撮影系の
Filament B-40/L-40
current では CL/CS(大焦点/小焦点)のフィラメント電流切替をこ
of CL/CS(Large Focus/Small Focus) is changed on this board.
の基板上で行っている。
3-3
(5)TERMINAL-2002
(5) TERMINAL2002
撮影台等とのインターフェイス
Interface with Bucky system.
充電コンデンサの電圧検出(VC+/VC0/VC−)の検出回路
Detection of electrolysis-condenser-voltage.(VC+ / VC- / VC0)
電源電圧(LV)の検出
Detection of Line voltage.(LV)
システム電源
Control ON/OFF
system power の制御(MG-C,MG-L,MG-RS,MG-D)
source.(MG-C, MG-L, MG-RS, MG-D)
SW1
SW2
CP3 VC0
CP5 VC+
3-4
Attention
補足 as follows, when you connect with Bucky.
Ex. In (例)24VDC
case of interface line is装置と信号をやり取りする場合(端子板コード“02”)
系で Bucky DC24V at Bucky system. ( terminal code is “02” )
UD UD
side(TERMINAL-2002)
側(TERMINAL2002) Bucky side (BF-10H/11H
メカ側(BF-10H/11H 等),etc.)
X1
1B2 Bucky start
request signal
1B2
X2
0V
X-ray start
request signal
1B1
(例)
Ex.AC100V
In case系で Bucky 装置と信号をやり取りする場合(端子板コード“02”)
of interface line is AC100V at Bucky system. ( terminal code is “02” )
Need to short.
UD UD
side(TERMINAL-2002)
側(TERMINAL2002) Bucky side (BF-10 ,etc.)
X2
L100 In case of Bucky
X1 work by AC100V
1B2
1B2
Relay
(K15A etc.) X3
The fifth terminal on
X3 terminal board.
Bucky start and X-ray
L0
start request signal
X2
0V
1B1
Need to short.
*注記*
Note
本基板上にスタータ起動信号・サーマル確認用の端子盤(X4)が存在しますが、こ
Don’t connect to X4 terminal board.
board have starter andType)ですので、B-40/L-40
Thisれは汎用装置用(1Tube ではここに接続しないで下さ
thermal circuit on X4 terminal ,but this is only for
い。
other X-ray system(1 tube type). So don’t use this terminal.
3-5
(6) TABLE i/F-2002(OP)
CP2
JUSD-2
JUSD-1
CP1 JGC4
CN8
X3
JP1
CP7 +5V X1
CP3 +24V
J132-1
CP5 −
CP6 + JH3
CP4 GND
X5
(7) FLUO-2002(OP)
透視関連の制御全般(IRIS
Not supported yet. 制御,透視用共振コンデンサ切替,II 電源,視野切り替え)
現在はまだサポートされていません。
3-6
(8) mA DETECTOR 2002
Detection of actual tube mA.
管電流検出,検出感度の切替
(9) TUBE SELECTOR-96(OP)
Tube selection of No1 and No2.
第1管球,第2管球の切替
(10) D150KH-40
このトランスを B-30 で使用することも可能。ただし追加配線が必要。
High Voltage trans.
CP7
CP6 GND
SW2
CP3 +3.3V
CP2 VCC
CP1 +15V
SW1
JP4,3,2,1 CP4 +5V
CP5 GND
3-7
(12) DAR i/F-2002
DAR-6000/循環器系メカとのパラレル
Not supported yet. I/F(撮影トリガ),
COM i/F2002 の CPU とバス接続による S/W 制御
現在はまだサポートされていません。
(14) PH i/F-2002(OP)
ホトタイマ制御を行ない、UD CONT 基板と通信します。 CP1 PH-DET
This board control PHOTO-TIMER of
ホトセルの制御もここで行ないます。
photo-multiplier type and photo-diode type.
J5 connector JM3
J6 connector J60
VR2 STC
CP2 +
15V
CP3 −
15V
CP4 GND
J4 CN8
CP12:CP13:CP14
J3
REF:GAIN_CONT1:GAIN_CONT2
CP15:CP16
GAIN_CONT3:GAIN_CONT4
J2 PH-POWER
CP6 FD1SIG CP9 FD2SIG
CP11 PHSIG2
CP7 FD3SIG
3-8
(15) PH POWER-2002(OP)
ホトマル型の受光部に高電圧を印可します。
This board supply power to the photo-multiplier.
VR1 CP1 DET1 -3
VR2 CP2 DET1 -2
VR3 CP3 DET1 -1
VR4 CP4 DET2 -3
VR5 CP5 DET2 -2
VR6 CP6 DET2 -1
VR7 CP7 DET3 -3
VR8 CP8 DET3 -2
SW1 VR9 CP9 DET3 -1
J1 DT1B
CP11 J2 DT1A
HV.SET J3 DT2B
J4 DT2A
J5 DT3B
J6 DT3A
CP14
J7 DT4
CP10 −HV
CP13 +
CP18 II2_IN
CP15 −15V
VR10
CP17 +15V
VR12 OFF_SET2
CP19 GND
VR11
CP19 II1_IN
CP16
VR13
3-9
(16) PH CONT-2002(OP)
ホトタイマの制御を行ないます。
This board control photo-timer.
CP29 PHM_1
CP27 −15V
CP26 +15V
CN1 DET4
CP14 +15V
CP7 GND
CP5 CHECK
CP15 IP OFF
CP6 REF IN
VR20 VR10
CN1 DT2A VR21 VR11 CP1 REF V+
VR3
VR5
VR22 VR12
CN1 DT3A
VR23 VR13 VR6 J1 ION DET1
CN1 DT3B VR4
VR19 VR9 J2 ION DET2
3-10
4. Troubleshooting
4.1 When Measured kV Over (TOV) occurs
TOV occurs
Problem
No problem
No problem
Check the waveform at kV+ and kV-.
Change the UD
From B (kV+(CP 52), kV-(CP 51), TKV (CP 32) on the
CONT2002 board.
UD CONT board )
Problem
No change
Situations change
No change
4-1
4.2 Checking the filament heating system control when ROC occurs
No change
Enter into the adjustment mode (FVR ) and Is K3 relay on mA Detector
turn OFF feedback to check the situation. 2002 normal ?
AC power supply
problem
Normal Problem
Check T1 power supply
Change mAPower2002 .
To B
4-2
4.3 When Measured kV Over (TUN) occurs
TUN occurs
Problem
Check the power fuse of Check IGBT gate
the inverter unit. waveform.
Normal
Problem
MGRS is not
Does MGRS operate and is charge operating
Check the J-MGRS
control resistance short-circuited
(Terminal2002) signal .
properly ?
MGRS is operating
Problem
Does MG-D operate properly ? Check the J-MGD
Are RC+ and RC- normal? (Terminal2002) signal .
Normal
Problem
Are A /D (CP7) and B/C (CP6) on UD Change UD
CONT board normal? CONT2002 .
Normal
To A
4-3
4.4 Notes for installation, adjustment, check, and troubleshooting
With UD150B-40/L-40, there are some notes for installation, adjustment, check, and
troubleshooting. Please make sure to follow them. Failing to do so may lead to
breakage of the instrument, secondary problems, or fire accidents.
2) Make sure to turn OFF tube current feedback before adjusting the tube current.
Make sure to turn OFF mAFC in the adjustment mode (FVR mode) and tube
current feedback before adjusting the radiography tube current. Proper
adjustment is impossible when it is ON.
4-4
Matters to be avoided at all costs
(1) Do not carry out X-ray operation when the high-voltage cable is disconnected.
Doing so with an inverter-type high-voltage generator will damage diode of the
IGBT unit or high-voltage transformer. Bad contact of a high-voltage cable may also
cause damage.
(2) Do not operate the filament heating circuit output side unloaded.
Turning ON the FET block of MA-POWER2002 unloaded may cause damage.
This is the same when J3 of MA-POWER2002 is disconnected.
(3) Do not connect the ground wire of the oscilloscope power plug.
Use the instrument without connecting the GND terminal of the oscilloscope’s
power outlet in order to prevent breakage. Do not perform measurements with
different reference potentials between CH1 and CH2.
Addition
Using this instrument (B-40/L-40), the inverter can be oscillated by connecting an
oscilloscope/DVM to the VIN (CP14) signal. (Connection to the F (CP12) signal is
impossible same as other models.)
4-5
4.5 Preparation for test operation
Prepare to test the Ready and X-ray operations without charging the electrolysis
condenser of the inverter unit. These connections must be restored when they
become unnecessary for other checking or adjustment. Restore them to their original
states after the preparation process is completed.
Addition
Steps 4 and 5 are unnecessary when a tube has been already attached.
4-6
5. Waveform collections
5.1 about filament heating control (Normal waveforms)
***Notice!!***
The waveform as follow isn’t appear in situation of “test operation”.
5-1
2) CL-CO Filament heating waveform (normal waveform)
This is a filament heating waveform at READY.
5-2
5.2 FET Gate-drive-circuit
***Notice!!***
The waveform as follow isn’t appear in situation of “test operation”. 1) FET Gate
pulse waveform( at Pre-heating)
下記条件時
This width depend on FVR value.
4μS
In case of under condition, width is 4μS.
5-3
2) FET Gate pulse waveform( at Ready Up)
下記条件時
This width depend on FVR value.
6.60μS
In case of under condition, width is 6.60μ
S.
5-4
5.3 Flash waveform (IFVL)
***Notice!!***
To check this waveform , it is need to prepare “test operation” setting.
(1)Select large focus.
(2)Connect probe-side of CH1 to IFVL(CP23) and connect clip-side to GND on UD
CONT2002.
(3)Set FVR Value to 4.0A .
C B A
B = 3/5 A
C = 4/3 A
Pre-heating
(Flash)
5-5
5.4 IGBT Gate-drive-pulse waveform(F)
“F”:”VIN”=”90.0kHz”:”8.25V”
***Notice!!***
To check this waveform , it is need to prepare “test operation” setting and need
to short circuit between CP20(VSIN) and CP3(+15V) on UD CONT2002.
11.1μs
5-6
6. Changing a PC board
Introduction
This section describes adjustment procedures to be taken at plants. Refer to this section
when there is a problem at a customer's site or readjustment is necessary after
replacement of a PC board or other parts. Adjustment work described in this section is
unnecessary at installation.
Use an about 18 meter-long high-voltage cable (low-capacity type) for adjustment work
at a plant. (Using a high-voltage cable other than 18 meter-long ones may hinder
adjustment work because normal waveforms cannot be obtained.)
6-1
6.1.1 Preparation for test operation
Prepare to test the Ready and X-ray operations without charging the electrolytic
capacitor of the inverter unit. Adjust the UD CONT2002 board with the X-ray tube and
starter connected. Make sure to restore connections when they become unnecessary for
other adjustments.
6-2
6.1.3 Adjustment of the oscillation pulse width
Carry out rough adjustment in this stage. Fine adjustment should be carried out later.
(In subsection 1.5)
(1) Connect the prove of an oscilloscope to CP6 B/C and CP7 A/D respectively.
(2) Perform “Ready Up” and “X-ray” operations and adjust VR1 PULSE ADJ so that the
pulse width of CP6 and CP7 becomes approximately 10 µs.
(Before X-ray operation, ma ke sure that the gate driver cable J8 and J10 to
INVERTER-2002 are disconnected.)
CP6 B/C
10 µs
CP7 A/D
10 µs
(3) Make sure that each signal is alternately outputted as shown in the figure above.
6-3
6.1.5 Adjustment of gate waveforms
Make sure that the instrument is ready for test operation j.
After turning OFF the power, connect J8, J10 To INVERTER UNIT on the UD
CONT-2002 board.
(1) Connect CH1 of an oscilloscope to CP2 E and CP1 G on the INVERTER-2002 board.
(Connect GND to CP2 E.) Connect CH2 to CP4 E and CP3 G on the
INVERTER-2002 Board.
(2) Short-circuit CP20 VSIN and CP3 +15V on the UD CONT board.
(Or, type “check 7FF” on the maintenance PC’s terminal screen to forcibly maximize
VS/VN 10V OUT.)
(3) Turn ON the power.
(4) Perform “Ready Up” and “X -ray” operations and measure a gate waveform at its
stationary state. (The stationary state is achieved in 0.1 sec after the X-ray
operation starts.)
(5) Adjust VR1 PULSE ADJ on the UD CONT 2002 Board so that the pulse width is
10.0 µs±0.05 µs for 5V. Also, make sure that the phase lag is 11.1 µs±0.5 µs for 0V.
(6) Make sure that each waveform has a peak voltage of 20V or lower, positive flat
portion of 13V or higher, and negative flat portion of -6V or lower.
(7) Turn OFF the power and connect CH1 of the oscilloscope to CP6 E and CP5 G on the
INVERTER-2002 board. Connect CH2 to CP8 E and CP7 G on the INVERTER-2002
board.
(8) Perform procedures (3) through (6). Adjust VR1 PULSE ADJ on UDCONT 2002
Board described in (5) if necessary.
(9) When using B-40 type, perform procedures (3) through (8) for the other
INVERTER-2002 board as well.
CAUTION
Adjust VR1 PULSE ADJ on the UDCONT 2002 board so that all IGBTs
satisfy the condition described in procedure (5).
6-4
10µs 10µs
11.1µs
6-5
6.2 Changing the mA POWER2002 Board
Waveforms cannot be measured for the adjustment and checking described below if
“preparation for test operation” is taken.
6-6
7. Changing the battery in LCD and PLC
If the message “Panel Battery WARNING” or “Cabinet Battery WARNING” is
displayed when the power is turned ON, the respective battery is flat. Change the
battery.
***Notice!!***
A warning message for flat battery is always displayed after initializing LCD/UD
CONT. This does not indicate that the battery is actually empty. Turn OFF/ON the
power to cancel the message.
Replacement battery
Lithium battery, CR2450 P/N 074-73307-02
All registered character strings and history information are initialized if the battery
on CPU board of the operation panel runs out. The calendar function is initialized
if the battery on UD CONT-2002 in the control cabinet runs out.
***Notice!!***
Changing a battery clears information included in “radiography condition currently
set,” “registered character strings information,” and “error history.”
Addition: ? With S/W Ver1.00 or Ver1.01, data cannot be downloaded to a PC. Make
sure to backup the data by writing it down.
? Domains for writing data are different for S/W “V1.00 and V1.01” and
“V1.10 or later.” Even when upgrading “V1.00 or V1.01” to “V1.10 or
later,” upload/download using a maintenance PC is impossible. Make
sure to backup data by writing it down.
Usually, change a battery on the control panel CPU board after turning OFF the
power.
When a battery actually runs down, the message “Panel Battery EMPTY” or
“Cabinet Battery EMPTY” is displayed.
7-1
(1) Disconnect the connector when the power is OFF and unfasten eight screws on
the back of the LCD.
Unfasten 8 screws.
(2) There are about ten tabs on the LCD cover. Carefully remove the cover without
breaking them. (Do not pull the upper cover (LED board) too strong because it is
still connected to the lower cover (LCD) with a flat cable. * Carry out this process
carefully to avoid any damage to the flat cable or other components.)
7-2
(3) Disconnect the flat cable and remove the upper cover (LED board) and LCD
main unit.
(4) Disconnect the brown flat cable connected to the LCD panel.
***Notice!!***
Do not put strong tension or distortion on the cable because it is very thin. If the
cable is damaged, the entire LCD unit needs to be replaced because it is
impossible to replace the cable only.
(5) Unfasten the screws to hold the LCD main unit (four places).
7-3
(6) Disconnect the power and signal cable connectors and detach the LCD panel and
board.
***Notice!!***
Do not pull the LCD panel and board with connectors too strong because they are
connected with the power and signal cables.
***Notice!!***
An inverter board for the power supply is attached on the back of the LCD panel.
When wiring path is changed, inverter noise may hinder normal operation of the
LCD.
Surely remember the current wiring path on the board.
(7) On the board, the TU-003U1-PC01C board and LCD-CONT board are attached
two-tiered. Battery can be changed in this state.
LCD-CONT board
Battery
TU-003U1-PC01C board
7-4
(8) To change the PLD, unfasten four screws on the corners of the LCD-CONT board
to detach the board.
Check the PLD in this state and replace it with a new one using a special tool.
TU-003U1-PC01C board
(LCD-CONT has been removed)
Sound control VR
PLD (Remove it
using a special tool.)
(9) Assemble the LCD with the reverse procedures of steps (1) through (8).
Turn ON the power and confirm operation.
7-5