SPECIFICATIONS:
The square field X-Ray Beam Limiting Device provides for the stepless adjustment of X-Ray
field dimension to the size of the image receptor or to that of the anatomical area of interest.
The collimator body contains 6 pairs of lead faced shutters controlled by means of two
knobs. Four pairs of shutters are positioned near the entrance window to control the X-ray
beam and eliminate the off-focus radiation.
The other two pairs of shutters are positioned near the exit window to provide a sharply
defined X-ray field.
The direct visualization of the X-Ray field is given by a light-beam which corresponds to the
X-Ray beam, within a tolerance of two percent of the selected distance. The light field center
is provided by the intersection of two perpendicular silk-screened lines in the Lexan window
and projected on the light field by the light beam.
To activate the light field, press the area marked with the light symbol. The light will switch on
for 30 seconds before automatically turning off. Average illumination is not less than 160 lux
(16 foot-candles); the edge contrast ratio is four to one.
The field size at the different Focus-Film Distances (FFD) is indicated by the dial indicators.
Collimator and Potter Bucky alignment is performed by means of a laser light fed for 30
seconds by the electronic timer.
Variable filtration, choice of 4 different values: 0.3 mm. aluminum equivalent
1 mm. al. plus 0.1 mm. Copper
1 mm. al. plus 0.2 mm. Copper
2 mm. aluminum
Maximum protection against radiation leakage 150 kVp(leakage less than 40mr/hr)
Maximum X-Ray field coverage at 90 cm.: 43 x 43 cm
Indicator dial accuracy corresponds to x-ray field size within 2% of
SID used
Lamp power requirement 20 volts AC or DC 6.5 A
Lamp on time 30 seconds
Projecting lamp type: OSRAM HLX 64642 - 150 watts 24 volts 2
prong plug-in..
Lamp holder Bender 902
Laser light Melles Griot model 56HVL502/P1
Accessories: Two guides
Retractable steel measuring tape
Included with the collimator Manual collimator body.
Instructions manual.
Tube port mounting flange
X-Ray collimator model 989001002093 is for assembly on stationary equipment.
This product complies with DHHS Performance Standard "21 CFR Subchapter J" when
assembled and maintained in accordance with the instructions contained in this Instructions
Manual.