EQUIPMENT TYPE QUANTITY SPECIFICATION

SUCTION MACHINE Electrical characteristics

A 240VV, 60Hz mono-phase electrical source compliant with EN60601 or equivalent.


 Compatible with Nema socket standards.  Built-in protections against over-
voltage and over-current line conditions.

Operational characteristics

Small and lightweight.  Jar of approximately 1 liter, made of shatterproof material,


5 autoclavable, with spillover protection system cover  Rechargeable battery, with
built in battery charger and preferably level indicator  Battery capacity: 60 minutes
or more.  Indicative Sound Level: < 70 dBA.  Adjustable vacuum range
approximate 0-600 mmHg , with analogue pressure gauge  Expected suction
capacity indicatively 25 l/min.

Expected configuration – Accessories and Consumables-:

 Autoclavable bottles with overflow protection system, quantity 2  Complete


autoclavable tubing set, quantity 3 ,.  Disposable bacterial filter, hydrophobic, kit
of 10.  Disposable naso-gastric suction probes, complete with probetube
connectors, kit of 10
ELECTROSURGICAL Electrical characteristics
UNIT, BIPOLAR- A240V, 60Hz mono-phase electrical source compliant with EN60601 or equivalent.
MONOPOLAR  Compatible with Nema socket standards.  Built-in protections against over-
voltage and over-current line conditions.

Operational characteristics
2
Mounted on a 4 wheels cart, antistatic, with brakes and space for accessories lodging
(shelf / drawer).  Microprocessor controlled, with self test and automatic
monitoring of currents  Monopolar and bipolar functionalities  Hand and pedal
activation   Indicative expected power output 300 W  Intuitive power control of
different levels and modes selection from main panel  Pre-existing range of cut and
coagulation programs  Audio-visual indicators of operation, alarms and errors  The
equipment shall provide the following standard modes:  Pure cut,  Pure
coagulation,  Blended cut  Blended Coagulation  Spray  Bipolar cut  Bipolar
coagulation  Bipolar blended

Expected configuration accessories and Consumables-


EQUIPMENT TYPE
Cut-coag pedal  Bipolar pedal  2 reusable neutral plate adult, complete with cable
and standard connector pin  1 reusable neutral plate pediatric, complete with cable
and standard connector pin  3 sterilizable handles, with finger buttons cut-coag,
with standard 3 pins connector .
QUANTIT SPECIFICATION
Y
MONITOR, Electrical characteristics
PHYSIOLOGICAL,
PORTABLE (NIBP, SPO2, A 240V, 60Hz mono-phase electrical source, EN60601 compliant or equivalent 
RESP, ECG, Compatible with Nema socket standards.  Built-in protections against over-voltage
TEMPERATURE and over-current line conditions.  Integrated rechargeable battery back-up with at
least 1 hour of autonomy without an electrical source input.

Operational characteristics

 Monitoring parameters: Respiration, ECG, Oxygen Saturation, NIBP, and


Temperature:  Adult and pediatric programmed settings  Preferably no touch
screen but Knob and button controls.  Clear display of measured values and alarms
(visual and audible).  Indicative monitor size: 12’’ wide.  Simultaneous
visualization of at least 6 parameters: 2 ECG curves, respiration curve, SpO2 curve,
NIBP measurements and temperature measurements.  Possibility to chose the data
and curves to visualize  Trends of data for events recalling

Monitoring – Physiological Parameters:

 Oxygen Saturation (SpO2); technology to be specified (Philips, Nellcor…), with


alarm settings and adjustments.  Non-invasive blood pressure (NIBP) – Automatic
and Manual measurements, with alarm settings and adjustments.  ECG Module: 5
standard ECG leads with possibility of 12 leads, ST segment analysis for arrhythmia
detection and recognition, with alarm settings and adjustments.  Respiration:
through standard impedance of ECG electrodes, with alarm settings and
adjustments.  Temperature measurement with alarm settings and adjustments.

Connectors, cables and probes compatible with the anesthesia monitor (part of Item
25256361)  Possibility to expand to extra module with additional parameters (i.e.:
IBP, CO2)

Expected configuration – Accessories and Consumables-

Each Monitor should be supplied with the following:

1 ECG 5 leads cable  1 ECG 12 leads cable  100 disposable adhesive electrodes  2
reusable adult non-invasive pressure cuffs  1 reusable pediatric non-invasive
pressure cuff  2 NIBP tube and connectors  1 reusable adult oxygen saturation
(SpO2) sensor  1 reusable pediatric oxygen saturation sensor  2 oxygen saturation
(SpO2) extension cables  2 reusable adult temperature probe (Skin type).  Wall
bracket support arm complete with wall fixation set and basket for accessories. (Set
should permit easy detachment and monitor position adjustment
EQUIPMENT TYPPE QUANTITY SPECIFICATION
TABLE, SURGICAL, Operational characteristics
MAJOR
 Strong, durable, stainless steel frame structure operating table  Frame material:
stainless steel 316/316L or other stainless steel with greater corrosion resistance. 
4 articulated sections: head, back, pelvis and separate legs sections.  Wheels, with
stable stopping system for the basement  Approximate Dimensions: 2000 x 500
mm  Workload support of approximately 200 Kg in all operating positions. 
Expected basic movements mechanical / hydraulic (pedal, crank lever or rapid gas
actuator):  Height adjustment, indicative range 750-950 mm  Trendelemburg,
antitrendelemburg indicative +/- 30 deg  Lateral tilt indicative +/- 20 deg  Back
section adjustment, indicative +80/-30 deg  Head section adjustment and
detachable  Leg section adjustment in height and spread and detachable  Table
longitudinal sliding preferable (not restrictive)
2
 Stainless steel lateral rails for accessories connection by clamp  Removable
mattress covering and antistatic, impermeable, washable, material.  Mattress
resistant to corrosion, water, detergent soap, ethylic alcohol solution with or without
nitrite and to the hypochlorite of sodium.  Table compatible for x-ray radioscopy
studies.

Expected configuration – Accessories and Consumables-:

 Stainless steel anesthesia arch, complete with clamp  2 adjustable narcosis arm
lateral support, with antistatic cushion, complete with stainless steel clamp  2
adjustable foot support, with antistatic cushion, complete with stainless steel clamp
 Stainless steel removable basin  Fixation belts  2 lateral support, stainless steel,
with antistatic cuschion

 2 shouder support, stainless steel, with antistatic cushion  Stainless steel


extension support with antistatic cushion for arm/hand operation  Basic traction
system for orthopedic and trauma procedures

LIGHTS, SURGICAL, Electrical characteristics


DUAL, FLOOR
 A 240V, 60Hz mono-phase electrical source compliant with EN60601 or equivalent.
2  Compatible with Nema socket standards.  Built-in protections against over-
voltage and over-current line conditions.  Integrated rechargeable battery back-up
with autonomy of not less than 2 hours

Operational characteristics

 Mobile unit with at least 4 castor anti-static wheels and two brakes.  Sturdy and
stable in the various positions  LED bulbs, long lasting life  Expected light intensity
at 1 meter distance from the lamp at approximately 90k Lux.  Single head lamp
(approx diameter 50 cm), with single or multiple lighting elements, mounted on an
articulated arms (vertical approx 160 cm height, horizontal approx. 100 cm long)
Sliding contacts for continuous movements  Expected allowed movements and
ranges:  Titling of horizontal arm of approximately +/- 45deg ;  rotation of
horizontal arm around vertical axis of indicatively 260° and stable in any position. 
Pivoting of support arc arm  Rotation of the head  Exposed surface
characteristics:  Low profile, free of sharp edges;  Grabbing sides for movements
control  washable (high IP index);  Resistant to corrosion, water, detergent soap,
alcohol solution  Light regulation system with:  control of light brightness at least
between 50% and 100%;  Manual Focus adjustment.  Control panel at head lamp

Low profile free of sharp edges;  grabbing sides for movements and positioning
control  washable (high IP index);  Resistant to corrosion, water, detergent soap,
ethylic alcohol solution and to the hypochlorite of sodium.  Control panel at light
head for:  On off  Brightness control from low to maximum

 Focus manual control  Light field diameter: not less than 20 cm at 1 meter
distance from the light source.  Expected range of proper depth of illumination of
indicatively 4050 cm  Manual mobilization of the lamp through removable
autoclave handle.

Expected configuration

 4 removable autoclave sterilizable handles covers.

ANESTHESIA MACHINE Important:

This unit should be compatible with basic and advanced ICU physiological monitors
(Items #17117305 & #25256078).

Electrical characteristics

 A 120V, 60Hz mono-phase electrical source, compliant with EN60601 or


equivalent.  Built-in protections against over-voltage and over-current surges in
line conditions.  Integrated rechargeable battery back-up with at least 1 hour of
autonomy without an electrical source input. Automatic passage from line
alimentation to battery operating modes.

Operational characteristics  Microprocessor controlled system.  Trolley 


Stainless steel or ABS construction material.  Easy to clean, resistant to corrosion,
water, detergent soap, ethylic alcohol solution and to the hypochlorite of sodium. 
Mounted on 4 castor antistatic wheels, at least two of them with brakes  Drawers
for accessories lodging  Back cylinders support  Lateral rail for clamp connection

 Anesthesia Module  Closed circuit system with possibility to work in open circuit.
 Equipment with independent gases (Air, Oxygen and N2O) inputs.  Gas-specific
connectors for hoses gas supply and cylinders connections  Analogue pressure
gauges for input pressure indication for each gas  One or two independent system
integrated compatible vaporizers Detachable patient module with CO2 absorbent
canister  Bellow module  Oxygen 100% flush controlAnalogue rotameters,
preferably two for each gas

 Passive circuit with Waste Anesthetic Gasses Disposal (WAGD) device for exhaled
gasses.  At least the following safety systems:  oxygen-N2O gases minimum
interlocked mixture guaranteed  oxygen leakage or low pressure alarm with
simultaneous stop of N2O gas delivery;  adjustable Pressure Limiting (APL) valve 
compressed Air leakage or low pressure alarm with automatic passage of the units
using air to the oxygen alimentation;

 Ventilation Module  Digital display for ventilation parameters of approximately 8


inches   Flow generator with pneumatic actuator.  The equipment will be able
to work with compressed air and oxygen pressures of about 3.5 – 5 bar  Expected
standard ventilation modes:  manual / spontaneous  positive End Expiratory
Pressure  pressure Controlled Ventilation  volume Controlled Ventilation 
synchronized Mandatory Ventilation  At least the following programmable
parameters:  Type of patient (adult, pediatric)  Pressure values and limits  tidal
volume;  minute volume;  respiratory rate;  oxygen concentration;  peep;  I:
E ratio  Pause  Ventilation parameters monitoring:  Values and alarms of pressure
data and settings  oxygen values, alarms and settings  values and alarms of
volume and flow data  trends and numerical data charts and curves

Vital Parameters Monitoring  Monitoring parameters: Respiration, ECG, Oxygen


Saturation, NIBP, Temperature, CO2:  Adult and pediatric programmed settings

 Preferably no touch screen but Knob and button controls.  Clear display of
measured values and alarms (visual and audible).  Indicative monitor size: 13’’
wide.  Simultaneous visualization of at least 6 parameters: 2 ECG curves, respiration
curve, SpO2 curve, NIBP measurements temperature measurements, capnography 
Possibility to chose the data and curves to visualize  Trends of data for events
recalling

 Monitoring – Physiological Parameters:

 Oxygen Saturation (SpO2); technology to be specified (Philips, Nellcor…), with


alarm settings and adjustments.  Non-invasive blood pressure (NIBP) – Automatic
and Manual measurements, with alarm settings and adjustments.  ECG Module: 5
standard ECG leads, analysis for arrhythmia detection and recognition, with alarm
settings and adjustments.  Respiration: through standard impedance of ECG
electrodes, with alarm settings and adjustments.  Temperature measurement with
alarm settings and adjustments  CO2 measurement with alarm settings and
adjustments (alternatively in anesthesia unit)  Connectors, cables and probes
compatible with the ICU monitor (Item 25256078)

Expected configuration – Accessories and Consumables-:

 Anesthesia unit:  3 m long gas hose, with female DISS connectors to the wall
outlet (color coding green for oxygen, yellow for air, blue for nitrous protoxide) 
Passive scavenging long hose  2 sterilizable complete adult breathing circuits (with
sampling luer-lock Y piece connector)  1 sterilizable pediatric breathing circuit (with
sampling luer-lock Y piece connector)  Set of manual ventilation balloon 
Combination of Halothane plus Isoflurane vaporizers for one unit. Halothane plus
Sevoflurane for the other unit. Preferably with halogenate-specific fillers connectors

 5 Jars of of approximately 4 kg of CO2 absorbent  50 disposable bacterial filters 


Basic set of masks, different sizes  Oxygen cell  Set of 5 flow sensors  Test lung, 1
Liter

 Monitoring:  ECG 5 leads cable  100 disposable adhesive electrodes  2


reusable adult non-invasive pressure cuffs  1 reusable pediatric non-invasive
pressure cuff  2 NIBP tube and connectors  1 reusable adult oxygen saturation
(SpO2) sensor  1 reusable pediatric oxygen saturation sensor  2 oxygen saturation
(SpO2) extension cables  1 reusable skin temperature probe  1 reusable anal
temperature probe  Kit of 20 capnography sampling trap and tubes with luer-lock
connector

Digital 5 Description
Product description
Upper The Contec08A includes everything you need to connect to your computer
to review oximetry reports and blood pressure reports, including trend
Arm graphs and analytical diagrams. It supports up to 3 users and each user
can store up to 100 records each associated with the current date and
Blood time. when memory is full, another record will delete the first record, on a
first-in-first-out basis. Just plug your oximeter into your computer with the
Pressure included USB cable and you'll be able to create a report with the included
software.

Monitor, It has an auto-inflating cuff for easy measurement of blood pressure. The
blood pressure trend chart and graph can be displayed right on the colour
Pulse screen, without connecting to a computer. The monitor runs on 4 AA
batteries (not included) and shuts off after 5 minutes idle time.

Rate & The Contec 08A includes a full featured fingertip pulse oximeter that is
perfect for CPAP users who want to monitor their pulse and oxygen
SpO2 saturation periodically during the night, although this unit doesn't record
oxygen saturation. This type of oximeter can help you as a CPAP user

Meter - determine the effectiveness of your current CPAP pressure.


The Contec 08A features a colour, high-resolution 3" LCD panel. The
easy-to-read display shows time, date, blood pressure, pulse rate, blood
One oxygen saturation, pulse waveform, trend graph, trend table, battery level
indicator, and alarm indicator. It also can sound an alarm when the blood

Machine, pressure moves outside a pre-set range (both systolic/diastolic high and
low)

Multiple Specifications:
Measurement Method: Oscillometry

Functions Measurement Mode: Manual


Patient range adult and pediatric
Measurement unit: mmHg/kPa selectable
Measurement Pressure Range: 0mmHg?280mmHg, Pulse
Rate:40~240bpm,Spo2:35%~100% Measurement Accuracy: Blood
Pressure: ± 3 mmHg, Pulse Rate: large of 3bpm and ± 3%,Spo2:±2% in
70%~100% range Resolution: 1mmHg
Measurement type: SYS/DIA/MAP
Legal Disclaimer
Legal Disclaimer Description The price does not including the Customs
Tax Fee, if it has, the buyer need to pay it by yourself. Thanks for your
support!
Features & details
 Automatic cuff inflation and deflation
 With Pulse rate and SpO2 Check - Built-in flash memory up to
100 records for each user
 Error detection and Alarm set; Battery status indicator; auto
power off when idle
 Clinically proven accuracy - EC Directive and CE Approved
 Sync with PC-based Software via USB port; accurate analysis
systems