You are on page 1of 4

Tips & Tricks GPC/SEC:

How to Treat Your RI


Detector
Daniela Held, PSS Polymer Standards Service GmbH, Mainz, Germany

The refractive index (RI) detector is the most common detector in gel
permeation chromatography/size-exclusion chromatography (GPC/SEC).
The advantage of this universal detector is that it detects everything; the
disadvantage is that it detects everything. This instalment of “Tips & Tricks”
offers some advice when working with RI detectors.
Gel permeation chromatography/size-exclusion differential refractometers. They measure the
chromatography (GPC/SEC) is the method deflection of a light beam from the difference in
of choice for determining the molar mass the refractive index between pure solvent and
distribution of synthetic and natural polymers. solvent with sample. This difference is referred
Every GPC/SEC system requires at least one to as ∆n and is expressed in refractive index
concentration detector in the setup to allow for units (RIU).
the detection of the fractions eluting from the Figure 1 shows how that can be achieved
column. Typical detectors used are refractive by using a detector flow cell divided diagonally
index detectors (RIs), ultraviolet (UV), diode array into two parts. The cell therefore comprises a
detectors (DAD), photodiode array detectors sample-containing side filled with the column
(PDA), or evaporative light scattering detectors effluent and a non-sample-containing reference
(ELSDs). Each of these detectors has its side. If both sides contain the same mobile
advantages and disadvantages. This instalment phase the detector is in optical balance. When
of Tips & Tricks compares the RI to other sample elutes from the column, the refractive
concentration detectors and provides practical index in the sample-containing side will change
Photo Credit: Benbou/Shutterstock.com

tips when using it. and the light beam will be deflected.
RI detectors are therefore very universal
How Does an RI Detector Work? detectors. Compared to UV detectors they
In general, RI detectors respond to a change do not require chromophores in samples.
of the refractive index, n. Most detectors are They can detect all types of samples including

2 IInstrumental
nstrumental Innovations
Innovations 2 NNews
21 ews 24 Tips
Tips and
and Tricks
Tricks 28 Mallet
Mallet 34 QQ&A:
&A: SSteinhaus
teinhaus

24
24
3
37 Moreau
Moreau a
and
nd LLoftus
oftus 41 HHTC
TC Event
Event Preview
Preview 43 Training
Training a
and
nd E
Events
vents 45 SStaff
taff
The Column www.chromatographyonline.com Tips & Tricks

Figure 1: RI detector cell diagonally split into the reference and the sample part. The RI Figure 2: RI chromatogram of a mixture of four PMMA reference materials with different
detects difference in the refractive index between pure solvent and solvent with sample. molar masses. The system peaks at the end of the chromatogram indicating the end of size
exclusion mode are typical. Drift, noise, wander, and S/N are compared to the requirements of
ISO13885-1, when applicable.

polysaccharides, starches, and polyesters. pressure. Therefore, the refractive index will
Compared to ELSDs, which do not require change with solvent composition, traces of
chromophores either, they are much more contamination, degassing level, temperature
linear in response and evaporating the solvent fluctuations, and more. One consequence of detectors or ELSDs are. They need relatively leading to band broadening, and the signal
is not required. Therefore, RIs can be used with this is that an RI detector can only be applied high concentrations compared to other itself is pressure sensitive.
all types of solvents with or without salt. In in isocratic mode—solvent gradients are not detectors. The vast majority of an RI needs flushing or
addition, RIs are able to detect oligomers and possible. Another consequence is that the RI purging of the reference cell. To achieve
contaminants as residual monomer, which could will respond to any change in the experimental How to Handle an RI Detector a stable baseline, the solvent in the reference
be evaporated when using ELSDs. conditions with a change in its signal and a The cell design and the principle of the RI cell and the mobile phase must be the same
variation of its baseline. It sometimes appears detector have some consequences for practical quality. The reference cell must be flushed or
Disadvantages of RI Detectors as if RIs are unstable or generate unstable work. In a multidetector setup, which is very purged regularly, and at the very least when
At a fixed wavelength of incident light the baselines, when in fact, RIs are simply displaying typical in GPC/SEC, the RI often has to be the the mobile phase in the reservoir is changed.
changes in refractive index are generally linear the instabilities of the environment and the rest last detector in the daisy-chain (1,2). The main The best time to purge or flush the RI is
in relation to the changes in medium density. of the system. reasons for this are that the cell can withstand directly prior to the analysis as the last action
However, the density of a medium is also It should be also kept in mind that RIs only low backpressures, the cell and the tubing after the system and the columns have been
affected by composition, temperature, and are in general not as sensitive as UV in the detector often have large dimensions thoroughly flushed and had time to stabilize. If

2 IInstrumental
nstrumental Innovations
Innovations 2 NNews
21 ews 24 Tips
Tips and
and Tricks
Tricks 28 Mallet
Mallet 34 QQ&A:
&A: SSteinhaus
teinhaus

25
25
3
37 Moreau
Moreau a
and
nd LLoftus
oftus 41 HHTC
TC Event
Event Preview
Preview 43 Training
Training a
and
nd E
Events
vents 45 SStaff
taff
The Column www.chromatographyonline.com Tips & Tricks

It is good practice in GPC/SEC to prepare the same dn/dc, so the same concentration
Figure 3: RI detector wander example. A faulty degasser was responsible for the fluctuations.
As RIs also respond to the density of a medium the degassing level is important. Similar the samples in solvent taken from the mobile should result in the same area. This is true for
signals have been reported for malfunctioning ovens (heating cycle) or laboratories with bad phase reservoir to reduce the system peaks to this example because the highest molar mass
air conditioning. a minimum. The larger the difference between (eluting around 6 mL) has only half of the
the solvent used for sample preparation and the concentration compared to the three other
mobile phase, the larger the system peaks will lower molar mass samples resulting in only half
be. System peaks are identified by measuring of the peak area (4).
a blank sample within the sequence and 2) The larger the dn/dc, the larger the signal
overlaying this with the samples. It is important area. For example, if a polystyrene (PS) mixture
that the blank is treated the same way as the with the same molar masses and concentrations
samples themselves including all filtration steps. is measured, the PS signals will be higher than
Problems that can occur with RI detectors the PMMA signals because of the higher dn/
are mainly the same as for other detectors in dc of PS compared to PMMA (conditions: THF,
GPC/SEC. Typical are baseline drift (the steady 35 °C). If the dn/dc is 0 (isorefractive samples),
movement of the baseline, either up or down such as for polydimethylsiloxane (PDMS) in THF,
the scale), baseline wander (variation of the no signals or only small signals resulting from
detector with frequencies between 6 and end groups will be obtained when using an RI.
60 cycles/h), and detector noise (variation of the Here another solvent, such as toluene, needs
detector signal with frequencies above 1 cycle/s) to be used to obtain representative signals with
(3). The inset in Figure 2 shows the values good S/N. As the dn/dc is negative for PDMS
for baseline drift, signal noise, signal wander, in toluene, negative signals will be obtained.
and signal-to-noise ratio (S/N), if applicable It is possible to either inverse the signal at the
compared to the requirements of ISO13885-1. RI directly or in the corresponding software to
purge or flush is performed while the system (or for a mixture of four different polymethyl S/N Ratio: When discussing the S/N, it is evaluate them properly.
the columns) are still equilibrating, the solvent methacrylate (PMMA) reference materials important to understand the RI detector Baseline Drift: As mentioned earlier, RIs
in the reference cell will be of a different quality with different molar masses. The system response to a sample. The RI signal intensity show instabilities of the environment and
to the sample cell. No optical balance can be peaks at the end of the chromatogram are depends on the concentration and the the rest of the system. There are three
achieved before the run and low signal quality very typical. These peaks can be only negative refractive index increment, dn/dc. major sources for baseline drift: low mobile
may be a problem during the run. (as in the example), but a combination of For a sufficient S/N two parts need to be phase quality, temperature differences
negative and positive peaks is also common. considered: between the columns and the RI cell, or a
Troubleshooting for RI Detectors The system peaks indicate the end of size 1) The larger the concentration, the larger not yet equilibrated system. Using freshly
Figure 2 shows a common RI trace obtained exclusion mode. the signal area. In Figure 2 all PMMAs have prepared, high-quality solvent, thoroughly

2 IInstrumental
nstrumental Innovations
Innovations 2 NNews
21 ews 24 Tips
Tips and
and Tricks
Tricks 28 Mallet
Mallet 34 QQ&A:
&A: SSteinhaus
teinhaus

26
26
3
37 Moreau
Moreau a
and
nd LLoftus
oftus 41 HHTC
TC Event
Event Preview
Preview 43 Training
Training a
and
nd E
Events
vents 45 SStaff
taff
The Column www.chromatographyonline.com Tips & Tricks

Figure 4: RI trace obtained for an RI with a defective purge valve. The observation of a The air conditioning installed in the laboratory GPC/SEC light scattering setups they are
negative signal with a steep increase to a positive signal is typical. Replacement of the purge could also evoke detector wander. often used to estimate the refractive index
valve was required to solve the problem. Other Problems: A very interesting increment, dn/dc.
troubleshooting example is shown in Figure 4. • The disadvantages of RIs are that they
Here the signal of the internal standards (like not very sensitive, they are affected by
that of the other samples in the following environmental conditions, and they have a
injections) starts negative and then becomes reference cell.
directly positive with a steep increase in • RI detectors require high solvent quality for
between. The reason for this behaviour was stable baselines.
located in a defective purge valve. The reference • Purging or flushing of the RI reference cell
part of the cell was therefore not completely should be done as the last action prior to
shut during analysis and the sample was flowing the analysis and only for fully equilibrated
through both cell parts, resulting in one positive systems.
and one negative signal. The same behaviour
would be observed if a sample was run in References
activated RI purge or flush mode. A replacement 1. D. Held and P. Kilz, The Column 8(18), 9–12 (2012).
of the purge valve took care of this problem. 2. D. Held and W. Radke, The Column 13(11), 9–14
The signal split does not have to be as (2017).
pronounced as in the example. Sometimes the 3. D. Held, The Column 8(6), 13–18 (2012).
process starts with a small negative dip before 4. D. Held, The Column 10(10), 12–15 (2014).
the signal starts. However, once the valve is 5. D. Held, The Column 9(2), 2–5 (2013).
fully damaged the signal will be as shown in
thermostatted columns and RI detector cells, After removing the degasser completely from the example. Daniela Held studied polymer chemistry
and providing enough equilibration time for the setup, the signal quality was stable again. in Mainz, Germany, and works in the PSS
the system should take care of this issue. Similar traces are also observed when the Summary software and instrument department.
Baseline Wander: Worse than drift—which degasser is not working properly (because • RI detectors are universal detectors with a She is also responsible for education and
can be easily corrected (especially when of low solvent quality or malfunctioning of linear response suitable for many different customer training.
using the recommended two-step evaluation the vacuum pump) or if the column oven types of macromolecules. They do not
procedure [5])—is detector wander. Figure 3 temperature is not stable enough (heating require chromophores because they respond
shows a nice example of detector wander. cycles). In both cases working without a to a change in refractive index. E-mail: DHeld@pss-polymer.com
The reason for the signal fluctuations was a degasser or column oven yields better results • RIs are typically used as concentration Website: www.pss-polymer.com
degasser not suited for this GPC/SEC solvent. than working with low quality components. detectors in GPC/SEC. When applied in

2 IInstrumental
nstrumental Innovations
Innovations 2 NNews
21 ews 24 Tips
Tips and
and Tricks
Tricks 28 Mallet
Mallet 34 QQ&A:
&A: SSteinhaus
teinhaus

27
27
3
37 Moreau
Moreau a
and
nd LLoftus
oftus 41 HHTC
TC Event
Event Preview
Preview 43 Training
Training a
and
nd E
Events
vents 45 SStaff
taff

You might also like