You are on page 1of 6

Peer Reviewed

The Effects of a Short-Radius Elbow on


Electromagnetic Flowmeter Accuracy
KADE J. BECK,1 RYAN P. WELLER,1 STEVEN L. BARFUSS,1 AND MICHAEL C. JOHNSON1

1
Utah State University, Logan, Utah

This study was conducted to show the effects that a short- pipe installed upstream of the meter and downstream of a
radius elbow has on electromagnetic flowmeter accuracy. short-radius elbow. However, nearly all meters in this
Previous laboratory studies have shown that non-ideal pip- study were repeatable to within 0.55% when installed in
ing affects the accuracy of electromagnetic flowmeters. The straight pipe. The results illustrate that when a magnetic
results of this study indicated that many of the meters flowmeter is to be installed in either straight pipe or piping
tested did not meet their own manufacturer’s specified with upstream disturbances, a laboratory calibration will
accuracy either in straight pipe or with the required straight improve the installed accuracy of the meter.

Keywords: accuracy, calibration, disturbance, electromagnetic

This article analyzes the effects of a short-radius upstream and downstream pipe length between a distur-
elbow on electromagnetic (magnetic) flowmeter accu- bance and the magnetic meter vary between manufacturers
racy. Kelner (2003) showed that upstream piping con- and researchers. Lomas and Liptak (2003) suggest that a
figurations may cause flow measurement errors. The conservative estimate of 3–5 pipe diameters (3D–5D)
purpose of this research was to explore meter perfor- upstream and 2–3 pipe diameters (2D–3D) downstream of
mance for a common—but not optimal—magnetic the magnetic meter is needed. Lomas and Liptak do not
meter installation. The results of this study are not quantitatively define why this is a conservative installation
intended to be used for or against specific magnetic guideline. Thorn et al. (1999) state that the general recom-
flowmeter manufacturers. However, this study is mendation for meter installation is 5D upstream and down-
intended to provide a general sense of “off-the-shelf” stream of the meter to ensure good accuracy. However,
magnetic flowmeter accuracy. Thorn and colleagues did not quantify “good accuracy.”
Magnetic flowmeters are a type of fluid flowmeter Hanson and Schwankl (1998) claim that manufac-
that uses Faraday’s law of induction to measure volu- turers generally recommend between 8D and 10D
metric flow rates. Faraday’s law of induction states that upstream and 2D downstream and assert that this stan-
as a conductor of width D with velocity V passes dard has been accepted as a general guideline for mag-
through a magnetic field B, a flux e is created (Eq 1). netic meter installation without clearly identifying its
origin. Although there are speculations regarding the
e = DVB ð1Þ source of this guideline, researchers still debate its origin
(Hanson & Schwankl 1998).
In magnetic flowmeters, two magnets in the meter The British Standards Institution BS7526 (BSI 1991)
spool generate a magnetic field, which is held constant. requires 10D of straight pipe upstream and 5D down-
The fluid passing through the meter spool is the conduc- stream of the magnetic meter. Because of the variance in
tor, and the conductor width is the inner diameter of research recommendations, a standard installation con-
the meter spool. Consequently, the velocity of the fluid figuration may not be conclusive.
is directly proportional to the magnetic flux produced. Other researchers have also noted that the same
Magnetic flowmeters measure this flux as a voltage hydraulic disturbance caused different measurement
using two or more electrodes. effects in various magnetic flowmeters (Luntta &
To improve accuracy, meter manufacturers often provide Halttunen 1989). Furthermore, Halttunen and Luntta
installation guidelines for users. Recommendations for (1993) conclude that “the general behavior [of magnetic

E12 B E CK E T A L . | J UL Y 20 18 • 1 10 :7 | JO U RN A L A W WA
flowmeters] can be estimated, but if the accuracy of the first test established how the off-the-shelf meter per-
meter is important, the estimation must be based on formed in a straight pipe with more than 20D of straight
individual knowledge of the meter.” Not surprisingly, pipe upstream and 10D downstream of the meter. The
this article demonstrates that measurement errors vary second test introduced a disturbance by installing a 90
for the same hydraulic disturbance among various short-radius elbow at 3D upstream of the meter (30 in.
meters—even from the same manufacturer. from the downstream flange of the elbow to the
The study presented here was conducted in Logan, Utah, upstream flange of the meter). For this setup, each meter
at the Utah Water Research Laboratory. The results repre- was always installed so that the electrodes were in plane
sent data from seventeen 10 in. magnetic flowmeters from with the elbow (i.e., the elbow was in the horizontal
13 different manufacturers. The focus of this article is not plane, and the meter electrodes were horizontal).
on the 17 flowmeters specifically, but more on the applica- It should be noted that installing a meter downstream of
tion of the magnetic flowmetering technology. The authors a 90 short-radius elbow is only one of an infinite number
want to emphasize that magnetic flowmeters continue to be of non-ideal installation conditions in which a magnetic
an excellent means for managing and measuring water. flowmeter may be installed. Table 1 indicates the straight
Meter manufacturers often provide installation recom- upstream pipe requirements for 11 of the 17 meter specifi-
mendations to help users produce flow measurements cations, which recommend ≤3D of straight pipe upstream
within a given accuracy range. Even the manufacturers of the meter in order for the meter to meet its own
of magnetic flowmeters provide varying recommenda- accuracy requirement. Accordingly, for this research, the
tions on the required number of pipe diameters that authors selected an upstream length of 3D as a comparison
should be installed, both upstream and downstream of for all meters (see Figure 1). It is significant to note that if
the meter. Table 1 summarizes the installation and accu- any external disturbances (e.g., noise, interference) were
racy parameters of each of the meters that were tested. present, each of the meters was subject to them because the
By following the installation recommendations, manufac- tests were all conducted under the same laboratory condi-
turers assert that meter accuracy specifications will be tions. Therefore, all comparative tests provided relative
met within the manufacturer’s velocity range. As indi- differences regardless of external disturbances. All tests
cated in Table 1, the 17 meter manufacturer specifica- provided adequate pipe downstream of the meter to meet
tions varied from 0 diameters (0D) to 5D of straight pipe manufacturer specifications.
upstream of the meter and 0D to 3D of straight pipe
downstream of the meter. Personal communications with
meter manufacturers and the authors’ experience suggest RESULTS
that these variances are largely due to specific meter vari- Test flow rates were determined using gravimetric
ables, such as signal strength, number and location of the weight tanks as a reference. The weight tanks are traceable
meter electrodes, actual inside diameter of the meter to the National Institute of Standards and Technology and
spool, and axial rotation of the spool in relation to the have uncertainties of approximately 0.11%.
upstream disturbance when the meter is installed. Figures 2–5 present the data for the straight pipe and
3D tests. The horizontal red lines indicate the manufac-
turer’s published specified accuracy. The meters are repre-
TEST PROCEDURE sented anonymously by the number shown on the
Two specific tests were conducted for each of the horizontal axis. The vertical axis represents meter error as
17 meter calibrations. All test setups used a 10 in., stan- a percentage difference from the laboratory velocity. The
dard wall, carbon steel pipe with flanged couplings. The test velocities calculated during each test are represented

TABLE 1 Manufacturer meter installation specifications

Meter number 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17
Accuracy—% 1.0 1.0 1.0 1.0 0.5 0.5 0.5 0.5 0.4 0.4 0.3 0.25 0.25 0.2 0.2 0.2 0.2
MFR 2 2 2 2 3 1 3 5 5 5 5 5 3 3 3 5 0
recommended
upstream pipe—
diameters
MFR 1 1 1 1 2 0 — 2 3 3 2 2 2 2 1 2 0
recommended
downstream
pipe—diameters

MFR—manufacturer

Dash indicates that the value was not reported.

B EC K ET A L. | JU LY 2 01 8 • 11 0: 7 | J O UR N AL AW W A E13
off-the-shelf test in conditions that are significantly better
FIGURE 1 Test setup than each manufacturer requires. Although many meters
did not satisfy their own stated specifications, most of the
meters were repeatable. To illustrate this, the range from
each meter’s initial accuracy test (Figure 2) was
centered—or corrected—to zero. The bias was removed
so a direct comparison between meters could be made.
This method is a common approach to ensure in-field
accuracies after a laboratory calibration.
Figure 3 shows that after mathematically removing
the bias from original straight pipe test data, 14 of the
17 meters satisfied the accuracy specifications defined
by the manufacturer. The three meters that did not sat-
isfy their specified accuracies after correction failed
because the deviation of the lowest flow rate expanded
the range significantly. The horizontal black lines in
Figure 3 represent accuracy bands of 0.55%. All but
one of the meters in this study fall within this band after
the bias correction.
Figure 4 illustrates the test data with the meter
installed 3D downstream of the short-radius elbow
(3D test) and with the straight pipe bias removed. As
seen in Figure 4, even with the straight pipe bias
D—diameters
removed, only one of the 17 meters satisfied its own
accuracy specifications. This implies that using an
off-the-shelf magnetic meter will not always guarantee
by varying colors and shapes on the figures. All velocities accurate results.
shown are within the velocity range specified by the man- Figure 4 also shows that every meter in this study
ufacturer. Table 1 is inserted for convenience in each responded differently to the same non-ideal meter installa-
figure. tion, which confirms the research performed by Luntta
Figure 2 represents the data collected in straight pipe and Halttunen (1989). As noted previously, the authors’
conditions. Only three of the 17 meters in this study sat- experience suggests that this result is most likely due to
isfied the accuracy limits given by manufacturers in an the variances in each meter manufacturer’s design, such as

FIGURE 2 Straight data

4.00%

3.50%

3.00%
Velocity 0.96-1.04 fps
2.50%
Velocity 3.70-4.15 fps
2.00%
Deviation—%

1.50% Velocity 7.86-8.15 fps

1.00%
Velocity 11.5-12.3 fps
0.50%
Velocity 15.4-16.4 fps
0.00%

-0.50% Manufacturer accuracy


specification
-1.00%

-1.50%

-2.00%
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18
Meter Number

Meta number 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17
Accuracy – ±% 1.0 1.0 1.0 1.0 0.5 0.5 0.5 0.5 0.4 0.4 0.3 0.25 0.25 0.2 0.2 0.2 0.2
MFR. recommended US pipe – diameters 2 2 2 2 3 1 3 5 5 5 5 5 3 3 3 5 0
MFR. recommended DS pipe – diameters 1 1 1 1 2 0 - 2 3 3 2 2 2 2 1 2 0

DS—downstream, MFR—manufacturer, US—upstream

E14 B E CK E T A L . | J UL Y 20 18 • 1 10 :7 | JO U RN A L A W WA
FIGURE 3 Mathematically corrected straight data

1.50%

1.00%

Velocity 0 .96 -1.0 4 f ps


0.50%
Velocity 3 .70 -4.1 5 f ps
Deviation—%

0.00% Velocity 7 .86 -8.1 5 f ps

Velocity 1 1.5 -12.3 f ps


-0.50%
Velocity 1 5.4 -16.4 f ps

-1.00% Manufa cturer accuracy


specification

-1.50%
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18
Meter Number

Meta number 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17
Accuracy –±% 1.0 1.0 1.0 1.0 0.5 0.5 0.5 0.5 0.4 0.4 0.3 0.25 0.25 0.2 0.2 0.2 0.2
MFR. recommended US pipe–diameters 2 2 2 2 3 1 3 5 5 5 5 5 3 3 3 5 0
MFR. recommended DS pipe–diameters 1 1 1 1 2 0 - 2 3 3 2 2 2 2 1 2 0

DS—downstream, MFR—manufacturer, US—upstream

signal strength, number and location of the meter elec- the bias from the raw 3D test data by centering the
trodes, and the actual inside diameter of the meter spool. ranges on zero. After applying the new bias adjustment
Figure 5 presents the results without the straight pipe from the raw 3D test data, 10 of the 17 meters in this
bias correction and, instead, mathematically removes study satisfy accuracy specifications at 3D downstream

FIGURE 4 Mathematically corrected 3D test deviation using corrected straight pipe data

2.00% Velocity 0 .45 -0.5 2 f ps

1.50% Velocity 1 .60 -1.7 6 f ps

1.00% Velocity 2 .85 -2.9 8 f ps

Velocity 4 .05 -4.3 0 f ps


0.50%
Velocity 5 .20 -5.4 3 f ps
0.00%
Velocity 6 .41 -6.6 8 f ps
Deviation—%

-0.50% Velocity 7 .63 -8.0 6 f ps


-1.00% Velocity 8 .77 -9.0 1 f ps

-1.50% Velocity 9 .84 -10.1 f ps

Velocity 1 1.1 -11.3 f ps


-2.00%
Velocity 1 2.3 -12.7 f ps
-2.50%
Velocity 1 3.3 -13.8 f ps
-3.00%
Velocity 1 4.5 -15.2 f ps
-3.50% Velocity 1 5.6 -16.5 f ps

-4.00% Velocity 1 8.3 -20.4 f ps


0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18
Manufa cturer accuracy
Meter Number specification

Meta number 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17
Accuracy–±% 1.0 1.0 1.0 1.0 0.5 0.5 0.5 0.5 0.4 0.4 0.3 0.25 0.25 0.2 0.2 0.2 0.2
MFR. recommended US pipe–diameters 2 2 2 2 3 1 3 5 5 5 5 5 3 3 3 5 0
MFR. recommended DS pipe–diameters 1 1 1 1 2 0 - 2 3 3 2 2 2 2 1 2 0

DS—downstream, MFR—manufacturer, US—upstream

B EC K ET A L. | JU LY 2 01 8 • 11 0: 7 | J O UR N AL AW W A E15
FIGURE 5 Mathematically corrected 3D test deviation using corrected raw 3D data

1.50% Velocity 0 .45 -0.5 2 f ps

Velocity 1 .60 -1.7 6 f ps

Velocity 2 .82 -2.9 8 f ps


1.00%
Velocity 4 .05 -4.3 0 f ps

Velocity 5 .19 -5.4 3 f ps


0.50%
Velocity 6 .41 -6.6 8 f ps
Deviation—%

Velocity 7 .63 -8.0 6 f ps

0.00% Velocity 8 .77 -9.0 1 f ps

Velocity 9 .84 -10.2 f ps

Velocity 1 1.1 -11.3 f ps


-0.50%
Velocity 1 2.3 -12.7 f ps

Velocity 1 3.3 -13.8 f ps


-1.00% Velocity 1 4.4 -15.2 f ps

Velocity 1 5.6 -16.5 f ps

-1.50% Velocity 1 8.3 -20.4 f ps


0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 Manufa cturer accuracy
Meter Number specification

Meta number 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17
Accuracy - ±% 1.0 1.0 1.0 1.0 0.5 0.5 0.5 0.5 0.4 0.4 0.3 0.25 0.25 0.2 0.2 0.2 0.2
MFR. recommended US pipe - diameters 2 2 2 2 3 1 3 5 5 5 5 5 3 3 3 5 0
MFR. recommended DS pipe - diameters 1 1 1 1 2 0 – 2 3 3 2 2 2 2 1 2 0

DS—downstream, MFR—manufacturer, US—upstream

of a 90 short-radius elbow. Meters 5 and 13 are out- user to appropriately correct the data and account for
side of specification by 0.01%. Meters 15 and 17 appear that unique installation condition. Figure 5 shows an
to fall outside accuracy specifications due to the lack of example of this calibration and demonstrates that a sim-
repeatability of the meter for this installation condition. ulated laboratory calibration will significantly improve
From Table 1, the manufacturers of these two meters meter measurement performance.
(15 and 17) claim an accuracy of 0.2% with 3 or Figures 3 and 5 demonstrate that magnetic flowme-
more diameters of straight pipe. It should be noted that ters remain an excellent choice for flow measurement.
these two meters (15 and 17) would pass an accuracy Although the technology’s application affects the ability
specification of 0.4% for this specific installation con- to accurately measure flow to the resolution claimed,
dition. Meters 1, 6, and 12 are outside of specification the magnetic flowmeters in this study demonstrate an
because the lowest flow rate expands the range signifi- approximate repeatability of 0.55% and 0.75% in
cantly. The black lines in Figure 5 represent 0.75%. straight pipe and 3D downstream of an elbow,
All but three of the meters (1, 6, and 12) fall within respectively.
this band. The results of this study are limited to 10 in. mag-
netic flowmeters from 13 manufacturers installed in
straight pipe as well as 3D downstream of a short-
CONCLUSIONS radius elbow. Further research regarding additional
The authors of this study recommend laboratory cali- disturbances would add to the understanding of mag-
brations to ensure the accuracy of magnetic flowmeters. netic flowmeter accuracy. Expanding the data to differ-
A comparison of Figures 3 and 4 clearly shows that ent sizes of magnetic flowmeters as well as different
introducing a 90 elbow upstream of a 10 in. magnetic distances downstream of the elbow may provide addi-
flowmeter consistently causes the meter to be less accu- tional insights.
rate even though the results are repeatable. In this study, The results of this study add supporting data that
Figure 4 shows that applying a correction from a magnetic flowmeters do not always meet their own
straight pipe calibration is not sufficient to ensure accu- accuracy specifications and that installation conditions
racy in non-ideal flow conditions. affect meter accuracy even when the appropriate length
If a magnetic flowmeter is to be installed in piping of piping is installed upstream and downstream of the
that is less than ideal, the meter can be laboratory cali- meter. This study also demonstrates that laboratory cal-
brated in the same piping configuration, allowing the ibrations and the corresponding meter corrections are

E16 B E CK E T A L . | J UL Y 20 18 • 1 10 :7 | JO U RN A L A W WA
one viable solution for producing more accurate mea- REFERENCES
surements when the meter is installed for use. BSI (British Standards Institution), 1991. Methods of Evaluating the
Performance of Electromagnetic Flowmeters, BSI 31. BS7526
(BS 7526/ISO 9104). BSI, London.
ABOUT THE AUTHORS
Kade J. Beck is pursuing a master’s Halttunen, J.E. & Luntta, E.A., 1993. Modeling in the Analysis of
degree in civil engineering at Utah Installation Effects on Flowmeters. Proc. 6th International
Conference on Flow Measurement, Taejon, Korea.
State University in Logan, Utah. He
began working at the Utah Water Hanson, B.R. & Schwankl, L.J., 1998. Error Analysis of Flowmeter
Research Laboratory in October 2015 Measurements. Journal of Irrigation and Drainage Engineering, 124:
5:248. https://doi.org/10.1061/(ASCE)0733-9437(1998)124:5(248).
as a research assistant. Ryan P. Weller
is a graduate research assistant and Kelner, E., 2003. Flow Meter Installation Effects. http://asgmt.com/
Steven L. Barfuss (to whom correspondence may be wp-content/uploads/pdf-docs/2007/1/033.pdf (accessed Apr.
28, 2017).
addressed) is a research professor at Utah State
University, 8200 Old Main Hill, Logan, UT 84322 Lomas, D.J. & Liptak, B.G., 2003 (4th ed.). Flow Measurement.
USA; steve.barfuss@usu.edu. Michael C. Johnson is a Instrument Engineers Handbook. CRC Press, Boca Raton, Fla.
research professor at Utah State University. Luntta, E.A. & Halttunen, J.E., 1989. The Effect of Velocity Profile on
Electromagnetic Flow Measurement. Sensors and Actuators, 16:4:
335. https://doi.org/10.1016/0250-6874(89)85004-8.
https://doi.org/10.1002/awwa.1055
Thorn, R.; Melling, A.; Kochner, H.; Haak, R.; Husain, Z.D.;
Wass, D.J.; Wadlow, D. et al., 1999. Flow Measurement. The
PEER REVIEW Measurement, Instrumentation and Sensors Handbook. CRC
Date of submission: 09/19/17 Press, Boca Raton, Fla. https://doi.org/10.1201/9780415876179.
Date of acceptance: 03/13/18 ch28.

B EC K ET A L. | JU LY 2 01 8 • 11 0: 7 | J O UR N AL AW W A E17

You might also like