Professional Documents
Culture Documents
Phosphorescence
LUMINESCENCE and
SPECTROSCOPY Chemiluminescence
Pertemuan 6
PAK 325-KIMIA-KELAS A – Dr. Matlal F. Alif 16 Maret 2015
Photoluminescence
What happens after a molecule has absorbed light ?
Excitation More
Energy
Absorbance
molecule returns to the ground or lower energy state via a non-radiative
transition such as vibration, collision with other molecules, etc. These give
off the energy absorbed rather than the emission of light.
Fluorescence
Some energy is lost through various processes (e.g. non-radiative
transitions) and then light is given off.
Phosphorescence
The molecule transitions from an excited triplet state to an lower energy
singlet state and gives off light. Non-radiative transitions intervene.
What is luminescence ?
Luminescence is the emission of photons from electronically excited state.
Luminescence is divided into two types, depending upon the nature of the ground and
the excited states.
In a singlet excited state, the electron in the higher energy orbital has the opposite
spin orientation as the second electron in the lower orbital. These two electrons are
said to be paired. Return to the ground state from an excited singlet state does not
require an electron to change its spin orientation.
In a triplet state these electrons are unpaired, that is, their spins have the same orient
ation. A change in spin orientation is needed for a triplet state to return to the singlet
ground state.
So diamagnetic S1 paramagnetic T1
Types of luminescence
(classification according to the means by which energy is supplied to excite the luminescent molecule)
1) Photoluminescence : Molecules are excited by interaction with photons of radiation.
Fluorescence :
Prompt fluorescence : S1→ S0 + hν
The release of electromagnetic energy is immediate or from the singlet state.
Delayed fluorescence : S1→ T1→ S1→ S0 + hν
This results from two intersystem crossings, first from the singlet to the triplet,
then from the triplet to the singlet.
Phospholuminescence : T1→ S0 + hν
A delayed release of electromagnetic energy from the triplet state.
2) Chemiluminescence : The excitation energy is obtained from the chemical energy of
reaction.
3) Bioluminescence : Chemiluminescence from a biological system: firefly, sea pansy, jellyfish
, bacteria, protozoa, crustacea.
4) Triboluminescence : A release of energy when certain crystals, such as sugar, are broken.
5) Cathodoluminescence : A release of energy produced by exposure to cathode rays
6) Thermoluminescence : When a material existing in high vibrational energy levels emits energy at a t
emperature below red heat, after being exposed to small amounts of thermal energy.
A) Introduction
Fluorescence Phosphorescence
Example of
Phosphorescence
0 sec 1 sec
Fluorescence process
A: So + hν → S1 or S2 Radiation process
Molecular fluorescence spectrometry is bas
ed on the emission of light by molecules that h
ave become electronically excited subsequent
to the absorption of visible(400~700nm), U
V(200~400nm), or NIR (700 ~ 1100nm) radi
ation. Excitation process to the excited state f
rom the ground state is very fast, on the orde
r of 10–15 s.
VR: vibrational relaxation,
non-radiational process, 10–11 s ~10–10 s
.
IC : internal conversion, S2→ S1 S1→ S0
non-radiative process, 10–12 s.
ST : intersystem crossing, S1→ T1
Jablonski diagram.
F : fluorescence, S1→ S0 + hν 10–10~10–6 s.
P : phosphorescence, T1→ S0 + hν
10–4 s ~104 s.
Deactivation Processes:
a) vibrational relaxation: solvent collisions
- λ emission > λ excitation (Stokes shift)
- vibrational relaxation is efficient and goes to
lowest vibrational level of electronic state
within 10-12s or less.
- significantly shorter life-time then
electronically excited state
- fluorescence occurs from lowest vibrational
level of electronic excited state, but can go to
higher vibrational state of ground level.
b) internal conversion:
- crossing of e- to lower electronic state.
- S1 to S0 would also happen .
- efficient, therefore many compounds don’t
fluoresce (aliphatic)
- especially probable if vibrational levels of
two electronic states overlap, can lead to
predissociation or dissociation.
- dissociation: direct excitation
(absorption) to vibrational state with
enough energy to break a bond
- predissociation: relaxation to
vibrational state of a lower electronic
state with enough energy to break a bond
c) external conversion:
d) intersystem crossing:
• spin of electron is reversed
- change in multiplicity in molecule occurs (singlet to triplet)
- enhanced if vibrational levels overlap
- more common if molecule contains heavy atoms (I, Br)
e) Phosphorescence:
Deactivation from an ‘triplet” electronic
state to the ground state producing a
photon
PHOTOLUMINESCENCE
AND STRUCTURE
The presence of the benzene ring and the nature of substituents on it
seem to favour the fluorescent behaviour of the molecule.
The halogen substituents tend to decrease the fluorescence and shift
the fluorescence bands to longer wavelengths; the effects increase
with increase in the atomic mass of the substituted halogen.
PHOTOLUMINESCENCE
AND STRUCTURE
Compounds with fused ring are found to be especially
fluorescent, and the extent of fluorescence is found to be
directly proportional to the number of rings in the
molecule
The structural rigidity in a molecule favours fluorescence
Structural factors affecting fluorescence
1. Fluorescence is expected in molecules that are aromatic or multiple conjugated dou
ble bonds with a high degree of resonance stability.
2. Fluorescence is also expected in polycyclic aromatic systems.
3. Substituents such as –NH3, –OH, –F, – OCH3, – NHCH3, and – N(CH3)2 groups, ofte
n enhance fluorescence.
4. On the other hand, these groups decrease or quench fluorescence completely :
–Cl, –Br, –I, –NHCOCH3, – NO2, – COOH.
5. Molecular rigidity enhances fluorescence. Substances fluoresce more brightly in a g
lassy state or viscous solution. Formation of chelates with metal ions also promotes flu
orescence. However, the introduction of paramagnetic metal ions gives rise to phosph
orescence but not fluorescence in metal complexes.
6. Changes in the system pH, if it affects the charge status of chromophore, may influ
ence fluorescence.
T he fluorescence observed with rigid cyclic molecules
with pi-bonds is found to be enhanced by electron
donating groups e.g., −NH2, OR, – OH and OCH3,
T he electron withdrawing groups such as COOH, NO2,
N=N and Br, I and CH2COOH tend to reduce it.
2-(0-Hydroxyphenyl)- NH3
Cd2+ 365 Blue 2
benzoxazole
Li+ 8-Hydroxyquinoline 370 580 0.2 Mg
Sn4+ Flavanol 400 470 0.1 F-, PO43-, Zr
B, Be, Sb, colored
Zn2+ Benzoin - green 10
ions
OH
OH HO
N O O OH
C C
HO N N
OH H
O SO3Na
Sample cell
Power Source Excitation monoc
hromator Slit
supply
Emission monochromator
Detector
Data processor
4) Detectors
Photomultiplier
Photoconductive target vidicon
Return beam vidicon
Intensified target vidicon
Stephen G. Schulman , (Alan Townshend Edt.), Encyclopedia of analytical science, Vol. 3, Academic Press, L
ondon, pp. 1358-1365.
Schematic of a fibre optic based multichannel fluorometer.
IDA=512 element intensified linear photodiode array detector, L=lens, OF1 and OF2
= the excitation and emission fibres.
Stephen G. Schulman , (Alan Townshend Edt.), Encyclopedia of analytical science, Vol. 3, Academic Press, London,
p 1396.
Generation of fingerprint excitation-emission matrix. a) EEM of pure component, compound A, b) E
EM of pure component, compound B, c)fingerprint EEM of a mixture of compound A and B, d) isom
etric projection of fingerprint in c). Shelly et al., Clinical Chemistry 26, 1127-1132, 1980.
Applications
1) Direct measurement --- metal cations as fluorescent chelates
2) Indirect measurement where the fluorescence of the substance
being determined is measured prior to and after quenching
3) Indirect measurement where the fluorescence of the
determined substance is enhanced by the addition of a reacting
material.
4) Tracer techniques --- bioengineered anlysis.
FISH(fluorescence in situ hybridization)
5) SFS( spectral fluorescent signatures)
Fluorometers
- simple, rugged, low cost, compact
- source beam split into reference and sample beam
- reference beam attenuated ~ fluorescence intensity
Perkin-Elmer 204
TOTAL FLUORESCENCE
INSTRUMENT
Use array detector (CCD) to collect total fluorescence spectrum
TOTAL FLUORESCENCE
SPECTRUM
INSTRUMENTATION FOR
PHOSPHORESCENCE
MEASUREMENT
T he basic difference between fluorescence and
phosphorescence is that the phosphorescence
emission occurs at a different time frame and can
be measured only if the sample is solid or is at
liquid nitrogen temperatures.
The basic instrumentation for phosphorescence is
similar to that of fluorescence; however, two
aspects of the measurement need to be modified.
The first is the sampling technique and second
being the recording procedure.
SAMPLING
S ince most of the measurements in
phosphorescence are carried out in rigid media at
cryogenic temperatures of liquid nitrogen we
need to use solvents that have certain special
characteristics.
It is the most important requirement are
good solubility of the analyte.
The solvent must form a clear rigid glass at 77 K i.e., the
temperature of measurement.
In addition, it should be highly pure so that there is practically nil
background phosphorescence
Ethanol is an excellent solvent for polar molecules though it
may require addition of small quantities of acid or base
to produce a clear solid.
On the other hand a mixture of diethyl ether, isopentane
and ethanol in the ratio of 5: 5: 2 respectively, commonly
called EPA is an excellent choice for non-polar
compounds.
PHOSPHORIMETRY
Spectrophophorimeter is similar to a Spectrofluorimeter except that
the former instrument must be fitted with
1) A Rotating-shutter device commonly called a phosphoroscope and
2) a sample system which is maintained at liquid nitrogen
temperature.
PHOSPHOROSCOPE
PHOSPHOROSCOPE
1) The Rotating-Can Phosphoroscope:
• it consists of hollow cylinder having one or more slit which are equally
spaced in the circumference.
• This is rotated by a variable-speed motor.
• when the rotating-can is rotated by a motor the sample is first illuminated
and then darkened.
• Whenever, there is a dark, phosphorescence radiation passes to the
monochromator and be measured
2)The Becquerel or rotating disc phosphoroscope:
It has two discs which are mounted on a common axis turn by a
variable speed motor.
Both the discs are having openings equally spaced in their
circumference.
On moving becqueral disc the sample is first illuminated and
then darkened.
Common problems of fluorescence measurements
D.A. Harris, C.L. Bashford , Spectrophotometry & spectroflurimetry- a practical approach, IRL Press, Oxford, U
K, 1987, p.18-20.
Problems with photoluminescence
1) Self-quenching
Self-quenching results when luminescing molecule collide and lo
se their excitation energy by radiationless transfer. Serious offen
ders are impurities, dissolved oxygen, and heavy atoms or param
agnetic species (aromatic substances are prime offenders).
2) Absorption of radiant energy
Absorption either of the exciting or of the luminescent radiation
reduces the luminescent signal. Remedies involve (a) dilution the s
ample, (b) viewing the luminescence near the front surface of the
cell, and (c) using the method of standard additions for evaluatin
g samples.
John A. Dean, Analytical Chemistry Handbook, McGraw-Hill, 1995, New York, p.5.55
3) Self-absorption
Attenuation of the exciting radiation a sit passes through the c
ell can be caused by too concentrated an analyte. The remedy i
s to dilute the sample and note whether the luminescence increas
es or decreases. If the luminescence increases upon sample diluti
on, one is working on the high-concentration side of the luminesc
ence maximum. This region should be avoided.
4) Excimer formation
Formation of a complex between the excited-state molecule a
nd another molecule in th ground state, called an excimer, causes
a problem when it dissociates with the emission of luminescent r
adiation at longer wavelengths than the normal luminescence. Dil
ution helps lesson this effect.
John A. Dean, Analytical Chemistry Handbook, McGraw-Hill, 1995, New York, p.5.55
CHEMILUMINESCENCE
- chemical reaction yields an electronically excited species that emits
light as it returns to ground state.
- relatively new, few examples
A + B à C* à C + hν
Examples of Chemical Systems giving off light:
Luminol (used to detect blood)
NH2 O NH2
C COO-
NH O2/OH-
+ hν + N2 + H2O
NH
C
COO-
O
S
HO
N
O
Luciferin (firefly)
HO
“Glowing” Plants
Luciferase gene cloned into plants
OTHER APPLICATIONS
Determination of nitrogen monoxide
NO + O3 → NO2* + O2
Determination of sulfur