Professional Documents
Culture Documents
discussions, stats, and author profiles for this publication at: https://www.researchgate.net/publication/257347180
CITATIONS READS
8 304
3 authors, including:
All content following this page was uploaded by Amit Agrawal on 30 December 2016.
The user has requested enhancement of the downloaded file. All in-text references underlined in blue are added to the original document
and are linked to publications on ResearchGate, letting you access and read them immediately.
Sensors and Actuators A 184 (2012) 78–85
a r t i c l e i n f o a b s t r a c t
Article history: Piezoelectric sensors are one of the most widely used sensors for vortex flowmeter application due to
Received 12 March 2012 their low cost. Various researchers have employed piezoelectric sensor for this application. However, the
Received in revised form 2 July 2012 location of the sensor and the performance of vortex flowmeter under disturbed conditions are seldom
Accepted 3 July 2012
reported. In the present study, experimental investigations are conducted with water as the working
Available online 14 July 2012
medium in a circular pipe of diameter 52.5 mm. The optimum position of the piezoelectric sensor behind
the trapezoidal bluff body is found out be 0.85 times the width of the bluff body. A new algorithm based
Keywords:
on empirical mode decomposition and autocorrelation decay rate is suggested to identify the vortex
Piezoelectric sensors
Vortex flowmeter
shedding frequency under low Reynolds numbers flow condition. The performance of the flowmeter is
Trapezoid also evaluated under different disturbed flow conditions to quantify the sensitivity of the flowmeter. The
EMD disturbances studied are single 90◦ bend, gate valve, globe valve, and two 90◦ out of plane bends. The
Autocorrelation overall uncertainty in the Strouhal number is within ±1.71%.
© 2012 Elsevier B.V. All rights reserved.
0924-4247/$ – see front matter © 2012 Elsevier B.V. All rights reserved.
http://dx.doi.org/10.1016/j.sna.2012.07.004
A. Venugopal et al. / Sensors and Actuators A 184 (2012) 78–85 79
Non-dimensional numbers
Reynolds number Um D 1. Optimization of the location of piezoelectric sensor placed
ReD
fd
behind a trapezoidal bluff body,
Staverage average Strouhal number St = Um 2. Explore signal processing methods for the detection of vortex
Stmax maximum Strouhal number
shedding frequency at different Reynolds numbers,
Stmin minimum Strouhal number
3. Evaluate the sensitivity of the performance of vortex flowmeter
Stmax +Stmin
Stmean 2 under various disturbed flow conditions.
Stfd Strouhal number under fully developed conditions
SNR signal to noise ratio
2. Experimental setup
Greek symbols
A closed loop water circuit is built to conduct the experi-
fluid density (kg/m3 )
ments in a circular pipe of diameter 52.5 mm. Appropriate test
dynamic viscosity (Pa s)
sections are designed and fabricated for conducting experimental
ı deviation (in percentage)
investigations with a piezoelectric sensor. A high accuracy water
deviation from fully developed conditions (in per-
flow dynamic calibration facility utilizing dynamic weighing is
centage)
employed in the present study. The system is capable of handling
flow rate in the range of 0–550 m3 /h. The accuracy of the present
facility depends on the diameter of the pipe and mass flow rate
flowmeter. However, the signal output from a piezoelectric sensor range. The collecting tank is equipped with an electronic plat-
consists of hydrodynamic noise, piping vibration and other inter- form scale load cell at the base. The collecting tank capacity is
ferences. Hence, power spectrum based signal processing methods 4000 kg. The system can respond to a time change of 0.1 s with 1 kg
sometimes fail to differentiate the vortex shedding signal from resolution. The uncertainty in the velocity estimation for the min-
other frequencies. In the recent past, Hilbert transform, which is imum Reynolds number covered in the present system is 0.34%.
a time-frequency domain analysis, is widely used for vortex shed- The details of the experimental setup and calibration facility are
ding frequency estimation. Empirical mode decomposition is the shown in Fig. 1. A mixed flow type of pump (capacity 78 m and
key step in the application of Hilbert–Huang transform proposed 114 m3 /h) is used for circulating water through the system from
in detail with theoretical basis by Huang et al. [10]. Sun et al. [11] an underground sump. The flow rate is controlled with the help of
proposed using Hilbert–Huang transform to estimate the vortex gate valves as shown in Fig. 1.
shedding frequency. The complex raw signal was first decomposed The load cell is interfaced to a personal computer through an
into intrinsic mode functions (IMF) with the help of empirical RS232 port. Two butterfly valves (150 mm) are provided down-
mode decomposition (EMD). The percentage error in the meter fac- stream of the collecting tank in order to drain or collect water in
tor obtained by this method was four times lower as compared the tank. Water from the tank is recirculated back to the sump. The
to the Fourier transform based method at low flow rates. Zheng experiments are conducted in a circular pipe of diameter 52.5 mm.
et al. [12] proposed an algorithm based on Hilbert transform and The test section is provided with an upstream pipe length of 35D
empirical mode decomposition for weak vortex signal. The residue to achieve fully developed inlet velocity profile. Flexible bellows
after every intrinsic mode decomposition was subjected to prob- are mounted at the inlet to the upstream pipe and outlet to the
ability density function estimate. A probability density of 5% was downstream pipe to dampen the pipe vibrations. A clamp on type
selected as the termination criteria for the shifting process. The ultrasonic flowmeter is installed 10D downstream of the test sec-
last residual component corresponds to the vortex shedding fre- tion for coarse adjustment of the mass flow rate. The output from
quency. The minimum Reynolds number covered with this method the sensor is logged to a personal computer with the help of a
was ReD = 6500. Sun and Zhang [13] proposed an energy ratio based Picoscope 2203. The bluff body employed in the present study is
method for diagnosing the vortex flowmeter performance based on trapezoidal in shape with a 72.5◦ included angle and a blockage
80 A. Venugopal et al. / Sensors and Actuators A 184 (2012) 78–85
Fig. 1. Test facility for vortex flowmeter. (1) Pump; (2) gate valve; (3) vortex flowmeter; (4) piezoelectric sensor; (5) ultrasonic flowmeter; (6) measuring tank; (7) platform
scale with shear beam load cell.
ratio of 0.27 as shown in Fig. 1 [7,8]. The blockage ratio is defined The Strouhal number is computed based on the average of these
as the ratio of bluff body width to pipe diameter. 44 sets of frequencies. The dynamic calibration facility is operated
for 10 min for a given Reynolds number. The linearity of the vortex
flowmeter is defined as:
2.1. Data reduction
Stmax − Stmin
For a given Reynolds number, the sensor signals are sampled Linearity (%) = × 100 (2)
2 × Stmean
at a rate of 1 kHz for 10 min. Subsequently, the sampled data are
divided into 44 blocks (each comprising 12,000 points) for Fourier
The deviation of the Strouhal number is defined as:
spectral analysis. The Strouhal number is calculated as
St − Staverage
fd %Deviation(ı) = × 100 (3)
St = (1) Staverage
Um
where f is the average frequency of 44 sets of readings for a given where Staverage is the average value of the Strouhal number for all
Reynolds number. Reynolds number covered.
A. Venugopal et al. / Sensors and Actuators A 184 (2012) 78–85 81
0.260 Table 1
x/d = 0.28 Linearity and deviation in St for various locations of piezoelectric sensor.
0.250 x/d = 0.42
Location (x/d) ReD Linearity (%) ı
x/d = 0.56
x/d = 0.71 0.42 26 × 103 –3 × 105 5.69 7
0.240 x/d = 0.85 0.56 26 × 103 –3.1 × 105 5.25 7.3
x/d = 1 27 × 103 –3.1 × 105
St
0.220
Sensor Output
Spectrum Analysis
(ΔP) α Um 2
ΔP Frequency ComputaƟon
Operating range
Y Output
No. of Peaks = 1
Vortex frequency
N
Noise
Velocity (Um) EMD
2
as the ratio for vortex shedding frequency (8.33 Hz) is 0.239. The
reason for such results is attributed to the broad spectrum of other
interferences, which make the autocorrelation correlation function
0 to decay faster. The method is modified and used in combination
0 100 200 300
Frequency (Hz) with EMD. The flow chart for the modified algorithm is shown in
0.0008 Fig. 5. In the present study, the decay rate of the autocorrelation
f = 9.31 Hz
ReD = 3.0 × 104 coefficient is chosen as the criterion for differentiating the vortex
PSD (W/Hz )
6
4 Raw Signal ReD= 2.8 × 104
Voltage (mV)
2
0
-2
-4
-6
0 0.2 0.4 0.6 0.8 1
Time (s)
2
IMF1 f = 84.33 Hz
1
Voltage (mV)
-1
-2
0 0.2 0.4 0.6 0.8 1
Time (s)
1.2
IMF2 f = 49.33 Hz
0.8
Voltage (mV)
0.4
0
-0.4
-0.8
-1.2
0 0.2 0.4 0.6 0.8 1
Time (s)
1.5
IMF3 f = 21.33 Hz
1
Voltage (mV)
0.5
0
-0.5
-1
0 0.2 0.4 0.6 0.8 1
Time (s)
0.4
0
-0.4
-0.8
0 0.2 0.4 0.6 0.8 1
Time (s)
0.3
distance is considered to be negligible as long as the performance
parameter is within the uncertainty intervals. The results of a
0.2
single 90◦ bend and 90◦ out of plane bends at various upstream dis-
turbances are shown in Fig. 9 with an uncertainty band of ±1.71%.
0.1 The performance parameter values are positive for 5D upstream
length and with the increase in upstream length (10D and 15D)
0 they tend to move toward negative values. This indicated that close
0 2 4 6 8 10
Peak Index to the disturbance the estimated vortex frequencies are higher than
under fully developed conditions. However, for all upstream length
Fig. 7. Autocorrelation decay of IMFs at ReD = 2.8 × 104 . the values variation with Reynolds number shows similar trend
84 A. Venugopal et al. / Sensors and Actuators A 184 (2012) 78–85
2
5D
Single 90° Bend 10D
1
15D
ζ
0 Out of plane 5D
90° Bend 10D
-1
-2
0 100000 200000 300000 400000
ReD
Fig. 9. Effect of single 90◦ bend and out of plane 90◦ bend at various upstream Fig. 12. Minimum upstream length requirements for various disturbances.
distances to the vortex flowmeter.
25% Closed
-1 10 D 50% Closed around the maximum frequency. The signal to noise ratio (SNR)
75% Closed value obtained under single 90◦ bend is positive and the spec-
-2 trum is distinct at 12 Hz. However, in the rest of the cases, multiple
peaks are observed. Hence, the SNR values are negative. This shows
-3
0 100000 200000 300000 400000 that three dimensional effects broaden the spectrum. The upstream
ReD length required for various disturbances are shorter compared to
most of the manufacturers recommended length. This is mainly
Fig. 10. Effect of gate valve for various gate openings at 5D and 10D to the vortex
because in the present study the upstream length requirement is
flowmeter.
quoted based on the uncertainty value of ±1.71%. Most of the man-
(i.e. decrease in value with increase in Reynolds number). The ufacturers quote the linearity of the flowmeter as the accuracy and
results at 10D for single 90◦ bend and 90◦ out of plane bends are the uncertainties are seldom reported. Hence, a direct comparison
with uncertainty band of ±1.71%. The results for gate valve at 5D is not feasible.
and 10D upstream to the flowmeter are presented in Fig. 10. The
results show that the most severe case is encountered with 25% f = 12 Hz
0.008 ReD = 3.9 × 10 4
PSD (W/Hz)
in Fig. 13. The signal quality parameter used is the signal to noise SNR = - 4.02
1.02f
Po 0
SNR (dB) = 10 log10 , Po = P(w) dw 0 100 200 300
Pr 0.98f Frequency (Hz)
0.0016
we
ReD = 2.8 × 104
PSD (W/Hz)
Pt = P(w) dw, Pr = Pt − Po
f = 9 Hz SNR = - 4.38
0 0.0008
0
2 5D 0 100 200 300
10D Frequency (Hz)
100% of Flow
15D 0.0016
PSD (W/Hz)
0 5D
50% of Maximum SNR = - 2.91
10D 0.0008
Flow
15D f = 9 Hz
0
-2 0 100 200 300
0 100000 200000 300000 Frequency (Hz)
ReD
Fig. 13. Power spectrum for lowest Reynolds number at 5D downstream various
Fig. 11. Effect of globe valve for 100% and 50% of maximum flow at various upstream flow disturbance. (1) Single 90◦ bend; (2) gate valve 75% closed; (3) two 90◦ out of
distances to the vortex flowmeter. plane bends; (4) globe valve 50% of maximum flow.
A. Venugopal et al. / Sensors and Actuators A 184 (2012) 78–85 85
6. Conclusions [10] N.E. Huang, Z. Shen, S.R. Long, The empirical mode decomposition and the
Hilbert spectrum for nonlinear and non-stationary time series analysis, Pro-
ceedings of the Royal Society of London (1998) 903–995.
Piezoelectric sensors are the cheapest available sensors for vor- [11] Z.Q. Sun, J.M. Zhou, P. Zhou, Application of Hilbert–Huang transformation to
tex flowmeter applications. The location of the sensor behind the denoising in vortex flowmeter, Journal of Central South University of Technol-
bluff body is crucial in obtaining good linearity and turndown ratio. ogy 13 (2006).
[12] D. Zheng, T. Zhang, J. Xing, J. Mei, Improvement of the HHT method and appli-
However, the optimal sensor location is not available in the open cation in weak vortex signal detection, Measurement Science and Technology
literature. At low Reynolds numbers, piping vibrations and electri- 18 (2007) 2709–2776.
cal interferences makes the detection of vortex shedding frequency [13] Z.Q. Sun, H.J. Zhang, Application of empirical mode decomposition based energy
ratio to vortex flowmeter state diagnosis, Journal of Central South University
difficult. The signal processing algorithm should be robust enough
of Technology 16 (2009) 154–159.
to identify the vortex shedding frequency and unambiguously dif- [14] J.J. Miau, C.C. Hu, J.H. Chou, Response of a vortex flowmeter to impulsive vibra-
ferentiate it from other undesirable frequencies. A new algorithm tions, Flow Measurement and Instrumentation 11 (2000) 41–49.
[15] J.J. Miau, C.W. Wu, C.C. Hu, J.H. Chou, A study of signal quality of vortex
to achieve this is proposed in the present work. The results from
flowmeter downstream of two elbow out of plane, Flow Measurement and
the present study are summarized as: Instrumentation 13 (2002) 75–85.
[16] A. Laneville, A. Strezelecki, P. Gajan, P. Hebrard, Signal quality of a vortex
1. The non-dimensional optimum location for the piezoelectric flowmeter exposed to swirling flows, Flow Measurement and Instrumentation
4 (1993) 151–154.
sensor behind the trapezoidal bluff body is 0.85 with a linearity [17] C.T. Yang, J.H. Shaw, M.C. Wu, Verification of a PZT vortex sensor in disturbed
of ±1.14% and a turndown ratio of 1:11. pipe flows, in: Proceedings of the 2nd International Symposium on Instrumen-
2. The upstream length required for various upstream disturbances tation Science and Technology, Jinan, China, August 2002.
[18] C.C. Hu, J.J. Miau, T.L. Chen, Determination of real-time vortex shedding fre-
with piezoelectric sensor with ±1.71% uncertainties are quency by a DSP, Journal of the Chinese Society of Mechanical Engineers 27
a. single 90◦ bend – 10D (2006) 335–342.
b. two 90◦ out of plane bends – 10D [19] J. Peng, F. Xin, C. Ying, Flow measurement of by a new type vortex flow meter
of dual triangulate bluff body, Sensors and Actuators A: Physical 115 (2004)
c. gate valve 75% closed – 10D 53–59.
d. globe valve 50% of maximum flow – 15D. [20] K.J. Xu, Q.L. Luo, G. Wang, S.S. Liu, Y.B. Kang, Frequency-feature based
3. Empirical mode decomposition combined with autocorrelation Antistrong-disturbance signal processing method and system for vortex
flowmeter with single sensor, Review of Scientific Instruments 81 (2010)
decay rate can be used for vortex frequency estimation. The 075104–075110.
method is shown to be robust and works reliably even at low [21] J.J. Miau, C.C. Yang, J.H. Hou, K.R. Lee, A T-shaped vortex shedder for vortex
Reynolds numbers (<30,000). flowmeter, Flow Measurement and Instrumentation 4 (1993) 259–267.
References Biographies