Safety Update – US004/16/S
S Family VC31B Patch SW Field Update
Update within deadline 3 months
Mandatory ☒ Yes ☐ No
Affected ☒ Warranty ☒ Contract ☒ Others
Update by ☐ Customer ☐Apps ☐ RSC ☒ CSE
Remote update ☒ Yes ☐ No
Intranet download available ☒ Yes ☐ No
Update material required ☐ Yes ☒ No
Material free of charge ☒ Yes ☐ No ☐ No, credit if returned
Return of parts ☐ Yes ☒ No
Application training needed ☐ Yes ☒ No ☐ Recommended
Application training time N/A
Estimated completion time 1.25 h Number of CSEs: 1
Remarks: NA
Scope
Material number ACUSON S1000 PN 10441701,
ACUSON S2000 PN 10041461 and 10440017,
ACUSON S3000 PN 10441730
Software version VC30C, VC31A
Component material number N/A
Related to Update Instructions US013/15/S, US014/15/S, US016/15/S
Document Version
Siemens reserves the right to change its products and services at any time.
In addition, manuals are subject to change without notice. The hardcopy documents correspond
to the version at the time of system delivery and/or printout. Versions to hard‐ copy
documentation are not automatically distributed. Please contact your local Siemens office to
order a current version or refer to our website http://www.health‐ care.siemens.com.
Disclaimer
Siemens provides this documentation “as is” without the assumption of any liability under any
theory of law.
The content described herein shall be used by qualified personnel who are employed by Siemens
or one of its affiliates or who are otherwise authorized by Siemens or its affiliates to use such
documents.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or
authorized by Siemens or one of its affiliates are not entitled to use this documentation without
prior written authority.
Copyright
“© Siemens, 2016” refers to the copyright of a Siemens entity such as: Siemens Healthcare
GmbH - Germany
Siemens Shenzhen Magnetic Resonance Ltd. - China Siemens Shanghai Medical Equipment Ltd. -
China Siemens Medical Solutions USA Inc. - USA
1 Preparation
1.1 General Information
This Update will provide ACUSON S Family software version VC31B utilizing Remote Update
Handling and conventional on site visits with internet downloadable software and user
documentation.
This SW patch provides additional new service features to continue to support and improve
utilization of SRS.
• New version of the Manage Node Package (MNP) this is required to continue to support
SRS RUH going forward.
• Re-introduce syngo Remote Assist powered by eSieLink, Siemens Ultrasound
implementation of TeamViewer.
To take advantage of these new SRS features we recommend and encourage the countries to take
the opportunity and get their systems connected and configured for SRS at the same time this
update is performed.
Further information related to SRS and syngo Remote assist can be found on our service marketing
homepage.
https://intranet.healthcare.siemens.com/cms/USD/en/Departments/CS/Pages/ProgramsandProductS
erviceStrategy.aspx
1.1.4 Prerequisites
Minimum System Software version VC30C is required. Systems with VC30A, VC30B need to follow
up and perform Update Instruction US029/14/S.
It is recommended to contact the customer and ask them to archive all patient data prior to
arrival.
No update material is available from CSML for this update. Software patch and user
documentation must be downloaded from the Siemens intranet prior to arrival on-site.
The latest version of this document can be found in the Services Knowledge Base (SKB).
https://skb.healthcare.siemens.com/default.aspx
· You will need to sign on with PKI or Windows Login
· Check Document Types “UI”
· Type in the update number US004/16/S in the search field, click Search
https://bps-healthcare.siemens.com/productinformation/
https://productinformation.healthcare.siemens.com/
a. Use PKI card or Windows Login to login (If required)
b. Select US
c. Select Common US
d. Select SW Patches
e. Select S Family
f. Select VC31B (click) on the VC31B_450.0.118_IntranetUpdate_Rev02.zip file
11. Select a destination to save the file (in this example your desktop) click Save.
3. Select Windows Explorer, if necessary use the Browse button to navigate to Windows
Explorer then select Windows Explorer from the list and click OK.
Note: If this screen is not displayed, reference Appendix B to set up Windows Disc Image
Burner as default for iso-files.
6. Verify the status “The disk image has been successfully burned to disc”. The disk should
automatically eject. Click Close.
i Every CD-R/DVD has to be copied from the original file. Copying from CD-R/DVD
to CD-R/DVD is prohibited!
2 Update
2.1 Work Steps
3.2 Protocol
This update has been completed successfully. The customer has been informed of the benefits or
improvements resulting from this update.
Customer: ................................................................................................................
Notification: .....................................
Remark: ................................................................................................................
By pushing the description button, the details pertaining to the SW being installed will be
displayed.
Defer: Selecting the Defer option will postpone the SW installation. Customer can select Defer
option for up to 30 days at which time installation will be reported as failed.
Note: If the Defer option is selected, the GUI will appear on the next system power cycle or
reboot. The GUI will not appear after resuming from Standby.
Yes: Selecting the Yes button will immediately install the SW. The progress bar appears on the
screen.
During the installation the system will reboot and this screen will appear
Highlight Windows Disc Image Burner and Choose defaults for this program.
healthcare.siemens.com/services
siemens.com/healthcare