Load/Unload Problems
Answer ID 27849 | Published 20-May-2011 | Updated 12-Nov-2012
Boards :
Problem Solution
DDB card failure - 1. Reload the DDB altera use send all button.
The DDB doesn’t respond
2. Check that the GUI in the machine is defined correctly for TSP/ VLSI.
You need to make sure that it is defined correctly.
3. Check all the wiring from the DDB to the boards in the L-Box.
MCF card start up error Restart the machine, and then check if the error occurs again. If it does, replace
the board.
Communication problems between the 1. Check the network cable and the TSP/VLSI to the machine.
machine to the Brisque
2. Check the PC's TCP/IP definition.
3. Perform a TSP/VLSI configuration:
Local board returns a put and get error See Get and Put local board related error.
CPU bios is corrupt. 1. Configure the CPU BIOS after replacing the board.
DDB returns illegal buffer error message. See DDB returns faulty error message.
Startup problems with the Lotem Controller 1. Replace the NUC board.
software.
2. Try to replug the boards and wiring .It might be a communication problem
between the boards.
Expose parameters are not valid Make sure that the IO OBJECTS\ DDB CARDS\ DDB Card\ Minimum DDB
Divider parameter value is 1.