FC-1400
OPERATION
MANUAL
Rev. 1.2
FC-1400 Operation Manual
Terms of Warranty
- This product is manufactured and passed through strict quality control and
inspection.
- We provide a 1-year warranty period for main body, but Accessory provides a
6 months warranty period.(2 years in Europe)
- We will repair or replace any part of the FC-700 found to be defective in usual
operating circumstance for free to you.
- This warranty does not apply to any defect caused by improper abuse, misuse or
exposure to poor management.
Caution
How to reach us …
The following are telephone numbers and addresses for contacting various service, product
supplies and sales personnel
※ In the event of a malfunction or failure, contact Service Dept. of GIMA SPA along with the
model name, serial number, date of purchase and explanation of failure.
For a special emphasis on agreement, terms are defined as listed below in operation
manual. Users should operate the equipment according to all the Warning and Caution
instructions.
Warning
Be informed that it may cause serious injury or death to the patient, property damage,
material losses against the “Warning” sign.
Caution
Be informed that it may cause no harm in life but lead to injury against the “Caution” sign.
Note
Be informed that “Note” sign is used for notifying some important contents relating to
installations, use, maintenance for users, but not dangerous.
Note
The equipment should not be placed in the vicinity of electric generator, X-ray,
broadcasting apparatus to eliminate the electric noise during operation. Otherwise, it may
cause incorrect result.
Self-power line is important for FC-1400. To use same power source with other electric
instruments may cause incorrect result.
Note
Note
Accessory equipment connected to the analog and digital interfaces must be certified
according to the respective IEC standards ( e.g. IEC 950 for data processing equipment
and IEC 601-1 for medical equipment ). Furthermore all configurations shall comply with
the system standard EN 60601-1-1:1993.
If in doubt, consult the technical service department or your local representative.
Safety Symbols
- The International Electrotechnical Commission (IEC) has established a set of symbols
for medical electronic equipment which classify a connection or warn of any potential
hazards. The classifications and symbols are shown below.
Symbols contents
Stand by
Table of Contents
Terms of Warranty 2
How to reach us 3
Definition of Warning, Caution, Note 4
General Precaution on Environment 5
General Precaution on Electric Safety 6
Safety Symbols 7
Chapter 1. Introduction 10
1) Overview 10
2) Features 10
3) Configuration 11
4) Description 12
5) Display Screen 16
6) Description of Operation panel 18
7) Power 20
8) Description of Print form 21
9) CTG Analysis Results Report 22
10) Installation 23
1) Basic Instruction 25
2) Tracing the saved data 26
3) FHR(Fetal Heart Rate) Measurement 27
4) UA Measurement 30
5) Fetal Movement Measurement 30
6) Print 30
7) Alarm Sound 30
8) Volume Adjusting 30
9) Warning Message 31
10) Battery replacement (option) 31
Chapter 3. Setup 32
1) Alarm/FHR setup 32
2) TOCO/FM setup 33
3) Print/Screen setup 34
4) Patient I.D setup 35
5) Time and Date setup 36
6) Screen Brightness setup 37
Chapter 6. Specifications 43
Product Warranty 45
Without prior notice, the specifications and functions are subject to change to enhance
the product in this manual.
Chapter 1. Introduction
1) Overview
FC-1400 is a fetal monitors, providing continuous monitoring, display, and recording of fetal
fetal heart rate(FHR), fetal movement(FM), and uterine activity(UA) for antepartum testing
and monitoring. FC-1400 irradiates the ultrasound wave to maternal abdomen, and detects
the Doppler effect signal reflected from the heart of the fetus. FC-1400 extracts FHR and FM
from this signal and provides the fetal heart beat sound with internal speaker. FC-1400
measures the uterine activity of a pregnant woman using toco sensor. FC-1400 displays
FHR, UA and FM with waves and numbers on the color LCD screen and saves them in
internal flash memory.
2) Features
1. Measures Fetal Heart Rate and Fetal Movement simultaneously.
2. Using the US probe for twins, measuring FHR and FM of twins is possible. (Optional)
3. Traces the saved data on the color LCD screen without wasting the record paper.
4. Prints the data you select during tracing the saved data at high speed.
5. Provides the clear heart beat sound and the accurate FHR because of the robust
ultrasound probe which endures noise.
6. Since it adopted 9-crystal and 1Mhz extending irradiation area, the discontinuity of FHR
minimized even during patient’s moving.
7. Detects and print fetal movement automatically analyzing Doppler effect signal.
8. Prints Grid and signals simultaneously when using Fax paper.
9. Provides RS-232 to communicate with the central delivery monitoring system.
10. Using the rechargeable battery, constant delivery monitoring is possible in lights-out
case. (Optional)
3) Configurations
FC-1400 consists of the following. Unpack the package and check the followings are
included. Also, be sure to check any damage to the main body and accessories.
▣ Default
① FC-1400 main body
② US Probe (1EA)
③ TOCO Probe (1EA)
④ Patient Marker (1EA)
⑤ Record Paper (2EA)
⑥ Power Adapter (1EA)
⑦ Power Cord (1EA)
⑧ Ultrasound Gel (1EA)
⑨ Probe Belt (2EA)
⑩ Manual (1EA)
▣ Optional
① US Probe for twins (1EA)
② Probe Belt (1EA)
③ Rechargeable Battery (1EA)
4) Description
▣ General
① ⑥
⑤ ④ ③
① Main Body
② TOCO Probe
③ US Probe
④ US Probe for twins ( Optional )
⑤ Patient Marker
⑥ Power Adaptor
▣ Top view
① ②
③ ④
▣ Front view
▣ Rear view
① ② ③
Warning
① ② ③
Note
To avoid an expected electric shock, do not open the equipment cover or disassemble the
equipment. . Please do not use the device when the probe is damaged. Refer servicing
to qualified personnel of GIMA SPA
5) Display Screen
⑫
④
⑪
⑩
⑤
⑨
⑧ ⑦ ⑥
① Setting the location of base line of FHR wave
② Screen Contrast setting menu
③ Power Status Area
④ FHR numeric area
- Display FHR numeric
- Display Heart Rhythm and signal level
- Display speaker volume status
- Display alarm On/Off
- Display offset value when detecting heart beat of twins-fetus
⑤ TOCO numeric area
- Display relative value of UA
- Display On/Off of automatic FM detecting
- Display the base line of UA
⑥ Printer setting menu
⑦ Patient ID setting menu
⑧ Time and date setting menu
⑨ Marking area of UA and the degree of FM in wave
- FM1, FM2: Fetal Movement, Wave that shows the degree of FM
- FMD1, FMD2: Fetal Movement Dot, Mark in dots when the intensity of FM exceeds
the set value.
- +10: Show the UA reference value
⑩ Clinician Mark marking area
⑪ Patient Mark marking area
⑫ FHR Wave marking area
- +20: 20 added in FHR waves
▣ Monitoring Mode
US1 Connected US1 Connected US1 not Connected US1 not Connected
US2 Connected US2 not Connected US2 Connected US2 not Connected
TOCO Connected TOCO Connected TOCO Connected TOCO not Connected
⑨ ⑩ ③ ④
② ①
⑤ ⑥ ⑦
7) Power
When DC is connected to the device, DC power indication light turns light green and when
battery is installed inside, it automatically starts to charge the battery.
When you turn on the device without DC power, battery provides power and the remains of
battery is showed in “Power Status Area” of screen.
When battery has a power lack, it alarms consecutive 3 times and shows battery capacity
shortage warning message on the screen. Connect DC cable as soon as possible and
charge the battery.
If you don’t connect DC cable within 1 minute after battery capacity shortage warning, the
device automatically turns off.
Battery charging and discharging time are as follows
●Charging time when the device is running: 14 hrs.
●Charging time when the device turns off: 6 hrs 30mins.
●Consecutive Battery Using time: 2 hrs 30mins.
The device has a built-in battery so that date and time can be saved even after turning off
the power. It uses Type CR2032 3V Lithium Battery.
Note
To protect the environment, please do not dispose batteries but request at medical
engineering department and follow the appropriate process to dump at designated spots.
10) Installation
▣ Connect Power
The product uses power adapter. The input of the adapter is DC 100 to 250V, 50-60hz, 1.2A
of free voltage and output is 18V, 2.5A. Plug the input cord of adapter into power plug and
connect the output cord into the device.
▣ Connect Cable
There are three types of cable, US probe cable, TOCO probe cable, Patient Marker cable.
Connect US probe cable to US1, US2 on the right side of the device, connect TOCO probe
cable to TOCO connector on the right side of the device. To avoid any confusion, US probe
cable and TOCO probe cable have grooves in different direction. Connect Patient marker
cable to MARK connector on the back side.
Note
When US or TOCO cable is connected the power of device must be turned off
1) Basic Instruction
① Connect power cable to the device and press the power button.
② Make sure that all the settings are as you want.
③ Change the settings as you would like.
④ Put some ultrasound gel on US probe, find heart beat of fetus and fix it around the
abdomen of pregnant woman using a fixing belt.
⑤ Fix TOCO probe on Fundus which is located 10 cm above the belly button of the
maternal abdomen using a fixing belt.
⑥ Give Patient Marker to the woman and let her press it when she feels the movement of
fetus.
⑦ Press “zero point adjusting button” and set up a base line of UA.
⑧ Adjust speaker volume so that you can hear fetal heart beat clearly.
⑨ Start to record on paper by pressing “Print” button when accurate heart rate shows up on
the screen.
If power switched on, FHR, UA and FM data is saved in the memory automatically and you
can trace the saved data on the screen. Follow the steps.
① Press the tracing mode button in monitoring mode and icon pops up on the top left
3) FHR Measurement
It gets fetal heart sound by using ultrasound Doppler effect and calculates heart rate by
minutes and saves it. Please get rid of air layer by putting enough of ultrasound gel on US
probe to avoid diminution of ultrasound in the air.
▣ FHR Measurement
① Put a fixing belt under the waist of the woman in the state of maternity.
② Put enough of ultrasound gel on US probe to get rid of bubble between the maternal
abdomen and the surface of US probe.
③ Find fetal back by fumbling the maternal abdomen and place US probe on it. Place US
probe as the picture below when fetus is looking right and left.
Note
When the Doppler Probe is put not on the back but on the breast part of a fetus, accurate
ultrasonic waves can’t be caught from the fetal heart and the fetal heat beat can be
frequently missed.
④ Place US probe where you can hear fetal heart beat quite loudly and clearly, the heart
turns on green LED in FHR marking area and blinks to fetal heart beat and adjust
speaker volume suitably.
⑤ Stick the button on the top of US probe to the fixing belt hole and fix it to avoid slipping of
US probe.
Note
Fix the Probe Cable toward the head part of a pregnant woman in order to prevent it from
being damaged, and for moving to some degree comparatively.
4) UA Measurement
You can detect UA by using pressure sensor externally sticked. Put TOCO probe on the
maternal abdomen and it detects the relative pressure which changes as the maternal
abdominal contraction and record UA.
Connect TOCO probe to TOCO connector on the right side of the device.
If TOCO probe is not connected to the device, you will see “TOCO off” in TOCO marking
area. If you connect TOCO probe to the device, you will see “TOCO” in the area and it is
ready to detect UA. When you see FM(Fetal Movement), it shows “TOCO+FM”.
▣ UA detecting
Note
If the UC Probe is connected to the equipment but not used, unreliable value may be
indicated on the UC indication section.
▣ Automatic Detaction
6) Print
Real time printing function has AUTO NST (NON-Stress Test)printing function and
Monitorting printing function. AUTO NST printing function is effective for un UA test and prints
FHR, UC and FM on paper and automatically stops. User can start and stop to record in
Monitoring Printing Function. Please see Printing set up in 3, 3,
7) Alarm Sound
Alarm sound occurs when FHR goes beyond the previously set normal FHR upper limit and
lowest limit and exceeds the alarm delay time. Please use the alarm button on operation
panel for alarm On/Off. Please see 3.1 for Alarm setting.
8) Volume Adjusting
You can hear FHR sound detected by US probe through built-in speaker and adjust volume
by choosing US1, US2 by pressing volume control button and turning the knob. Volume has
16 sound levels from 0 to 15. Press volume control button again.
9) Warning Message
It notifies you with sound in situations listed below when using the device.
① When US cable is came off the device in use,
② When running out of paper in printing the message” ” pops up.
Rechargeable battery of this device uses NiMH type of 12V, 2600mA. Charging time is 6
hours and usable for 2 hours. This battery is consumable thing so that it must be replaced
every 6 months or use over 500 times. When battery replace, contact agency or service
center of GIMA SPA.
U/S CONFIGURATION
US1
ALARM
FHR LOW 120 BPM
FHR HIGH
DELAY TIME
160 BPM 140
20 SEC
LINE COLOR
US1
US 1
US 2 US2 ( + 20 )
GUIDE
US2 OFFSET
OFFSET 20 140
TOCO + FM
[ OK ] [ Cancle ]
30 (10)
01/JAN/00 00:02:21 123456789 3cm/min
① Turn the knob and select FHR marking area in monitoring mode and press the knob
and US Configuration box pops up.
② Turn the knob and select FHR Low. Turn the knob, set lowest limit of alarm and press
the knob.
③ Turn the knob and select FHR high. Turn the knob, set upper limit of alarm and press
the knob.
④ Delay Time is the duration until alarm rings after detected FHR goes beyond the limit.
⑤ Line Color is US1 and US2 channel. It sets the color of FHR wave which is extracted
from input Doppler sound.
⑥ US2 Offset is US2 channel. You can use it when you add offset on FHR wave
extracted from input Doppler sound. You can choose offset among 0, 10 and 20.
2) TOCO/FM SetUp
TOCO/FM CONFIGURATION
US1
TOCO
BASELINE 5
FM
GRATH OFF
140
THRESHOLD 30%
LINE COLOR US1
TOCO US2 ( + 20 )
FM1
FM2
DEMO
MONITOR NUM.
OFF 140
1
TOCO + FM
[ OK ] [ Cancel ]
30 (10)
① Turning the knob and select TOCO marking area in monitoring mode. TOCO
Configuration pops up when pressing the knob.
② Select “FM+FMD” in order to display both FM(fetal movement) and FMD(fetal
movement dot);selecting ”FMD” would show FMD parameter only.
③ UA Color sets the color of UA wave.
④ FM Color sets the color of FM wave.
⑤ TOCO Baseline Set sets UA baseline when pressing zero point adjusting button.
Baseline takes one among 0, 5, 10.
⑥ Select “ON” to display demo data, and “OFF” to display real measurements.
⑦ Select the monitor’s own identification number. This number will be used to identify
the monitor within the central monitoring system.
⑧ To save the settings, select OK.
⑨ Not to save the settings, select Cancel.
PRINT CONFIGURATION
SPEED US1
SPEED 1 CM/MIN
PAPER
GRID OFF
140
SIZE B5
US1
CONTRAST LOW
AUTO PERIOD PRINT US2 ( + 20 )
REAL TIME 20 MIN
TRACE 00 MIN
CTG ON 140
TOCO + FM
[ OK ] [ Cancle ]
30 (10)
01/JAN/00 00:02:21 123456789 3cm/min
① Turn the knob select print setting menu in monitoring menu. Press the knob and Print
Configuration pops up.
② By “Speed”, you can control the printing speed in real time printing. It also controls
the speed of FHR wave shown on the screen. It takes one among 1 cm/min, 2
cm/min, 3cm/min which means shows 1 minute data in 1cm, 2cm, 3cm.
③ By “Paper grid”, you can select paper grid printing. If you choose ON, grid is printed
on the paper together with waves.
④ By “Paper size”, you can select paper size. There are two types of paper, A4 and B5.
⑤ By “Paper contrast ”, you can adjust a contrast of waves printed. If you choose 1, it
prints normally and if you select 2, twice as thick as that.
⑥ By “Auto Period Print Real time”, you can stop printing automatically when printing
time is passed. It keeps printing until pressing print button when it is 0.
⑦ By “Auto Period Print Trace”, you can select the amount of data to print when
pressing print button in tracing mode.
⑧ An item to set whether you want to use Cardio-Toco Graph (CTG) analysis function
or not. If you choose ON, Base FHR values will be printed out 10 minutes after you
press print button. If you press print button after 10 minutes, CTG analysis results will
be printed out
⑨ To save the settings, select OK.
⑩ Not to save the settings, select Cancel.
4) Patient ID Setup
① Turn the knob and select patient ID setting menu in monitoring mode. You can set ID by
pressing the knob. Here, icon changes into icon.
② Turn the knob and move to where you want to set figures.
③ Press the knob, turn the knob, select figures and press the knob again.
④ Repeat number 2 and 3, complete ID settings and press the knob when the cursor
locates on icon to return to monitoring mode changing into icon.
US1
140
US1
US2 ( + 20 )
140
TOCO + FM
30 (10)
① Turn the knob and select Time and Date setting menu in monitoring mode. You can set
time by pressing the knob. Here, icon changes into icon.
③ Press the knob, select figures by turning the knob and press the knob again.
④ Repeat number 2 and 3, complete Time and Date setting and if you press the the knob
when the cursor locates on icon, it returns to monitoring mode changing into
icon again.
US1
140
US1
US2 ( + 20 )
140
TOCO + FM
30 (10)
① Turn the knob and select icon in monitoring mode. You can adjust the brightness of
the screen by pressing the knob.
② You can change the brightness of the screen by turning the knob.
2) ACCELERATION
Visually apparent abrupt increase(onset to peak is < 30sec) of FHR above baseline. Peak
is ≥ 15 bpm. Duration is ≥ 15 bpm and <2 min. Peak of 10 bpm and duration 10sec is
acceleration
3) LATE DECELERATION
Visually apparent gradual decrease( onset to nadir is ≥ 30 sec.) of FHR below baseline.
Return to baseline associated with a uterine contraction. Nadir of deceleration occurs after
the peak of the contraction. Generally, the onset, nadir and recovery of the deceleration
occur after same time as the onset, peak, and recovery of the contraction.
4) EARLY DECELERATION
Visually apparent gradual decrease(onset to nadir is ≥ 30 sec.) of FHR below baseline.
Return to baseline associated with a uterine contraction. Nadir of deceleration occurs at
the same time as the peak of the contraction. Generally, the onset, nadir, and recovery of
the deceleration occur at the same time as the onset, peak and recovery of the contraction.
5) Variable DECELERATION
Visually apparent abrupt decrease ( onset to nadir is <30sec) in FHR below baseline.
Decrease is ≥15 bpm. Duration is ≥ 2 min and < 10 min.
Notice
Maintain the device and the electrode clean by rubbing the device with an alcohol cloth at
least once a month. Please do not use lacquer, thinner, ethylene, oxide.
Keep the cable away from dirt or filth, cleanse it with a cloth soaked with warm water
(40˚C/104 F) and cleanse it with clinical alcohol once a week.
Please do not soak the probe cable in liquid or cleanser and keep the device and cables
from any types of liquid.
3) Regular Inspection
Like every medical equipment, please take a regular test for FC-1400 for safety. For test
categories, please see the service manual provided by us.
Chapter 6. Specifications
■ Environmental Specifications
■ Power Specifications
■ Performance Specifications
Ultrasound
Input signal : Ultrasound Pulsed Doppler
Uterine Activity
Input Source : External Transducer with strain gauge
Physical Characteristic
Dimensions: 296(W) x 305.5(H) x 97.5(D)mm
Weight: Approx. 2.9kg(without Battery)
Display
320x240 STN Color LCD(4.7 inch)
Printer
Print Method : Thermal Array Type
Resolution: 8(vertical)/10(horizontal) dot/mm
Print Speed : 1, 2, 3 cm/min
High Speed print : 50 cm/min
Battery(Option)
Ni-MH Battery
12V, 2600mA
External Link
RS232C : Program Down Load, Central (Option)
Product Warranty
Approval No.
Approval Date
Serial No.
Date of purchase
Hospital :
Address :
Customer
Name :
TEL :
Sales Agency
Manufacturer
Manufactured by -
Bionet Co., Ltd :
#1101 11F, E&C Venture Dream Tower, 3 3821,
DigitalRo 31Gil, GuroGu, Seoul, 152719, ROK
Tel : +82-2-6300-6418 / Fax : +82-2-6300-6425
Distributor by -
GIMA SPA
VIA MARCONI, 1 – 20060 GESSATE (MI) ITALY
Tel : ++39 02 953854209 Fax : ++39 02 95380056
E-mail : export@gimaitaly.com
EC Representative -
MGB Endoskopische Gerate GmbH Berlin
Schwarzschildstr. 6, 12489 Berlin, Germany
TEL : +49(30)63927010 / FAX : +49(30)63927011