You are on page 1of 9

Clinical Oral Investigations

https://doi.org/10.1007/s00784-020-03348-8

ORIGINAL ARTICLE

Effect of canal curvature location on the cyclic fatigue resistance


of reciprocating files
Tyler Sobotkiewicz 1 & Xiangya Huang 2 & Markus Haapasalo 1 & Christophe Mobuchon 3 & Ahmed Hieawy 1 &
Jinghao Hu 1 & Huimin Zhou 4 & Zhejun Wang 1 & Ya Shen 1

Received: 18 February 2020 / Accepted: 15 May 2020


# Springer-Verlag GmbH Germany, part of Springer Nature 2020

Abstract
Objectives To determine the effect of the location of the canal curvature on the fatigue resistance of WaveOne (WO), WaveOne
Gold (WOG), Reciproc (Rec), and Reciproc Blue (RecB) files, and to examine the phase transformation behaviors of the
reciprocating file systems.
Material and methods The instruments were subjected to fatigue testing in five artificial canals with a curvature of 60° angle and a
3-mm radius. The location of the curvature was unique for each canal. Each file was inserted 16 mm into the canal and operated until
fracture occurred. The time to fracture was recorded and the length of the fragment was measured. Differential scanning calorimetry
was used to characterize the thermal behavior of the files. The number of cycles to failure was analyzed using two-way analysis of
variance and the post hoc Tukey test. The Kruskal-Wallis test was used to compare the mean fragment lengths between groups.
Results The instruments had significantly lower fatigue resistance in canals with curvatures in the middle and coronal canals
compared with those with apical curvatures (p < 0.05). At all tested curvature locations, RecB had superior fatigue resistance
compared with WO and Rec (p < 0.05). There were no significant differences between WOG and Rec in canals with curvatures
in the middle and coronal canals. The DSC thermograms for RecB exhibit a single exothermic peak during cooling but double
endothermic peaks during heating indicating that a two-step phase transformation from martensite to R-phase to austenite takes place.
Conclusions The reciprocating instruments experience decreased cyclic fatigue resistance when operated in canals with coronal-
and middle-third curvatures when compared with curvatures in the apical-third. Instrumenting coronally positioned curvatures
with reciprocating files needs to be performed with caution.
Clinical relevance The location of the root canal’s curvature influences the fracture resistance of rotary files that are used with
reciprocating movements. Therefore, caution needs to be exercised when using reciprocating instruments in canals with coronal
or middle curvatures.

Keywords Curvature location . Fatigue resistance . Reciproc . Reciproc Blue . WaveOne . WaveOne Gold

* Ya Shen
Introduction
yashen@dentistry.ubc.ca
Endodontic nickel-titanium (NiTi) rotary instruments fracture
due to torsional overloading or cyclic fatigue [1]. The overall
1
Division of Endodontics, Faculty of Dentistry, The University of prevalence of instrument fracture during root canal treatment
British Columbia (UBC), 2199 Wesbrook Mall, has been found to be between 1.7 and 3.3%, with rotary files
Vancouver V6T1Z3, Canada
responsible for 70–85% of the fractured instruments [2–4].
2
Department of Conservative Dentistry & Endodontics, Guanghua Using microscopic and fractographic analysis, authors have
School of Stomatology, Guangdong Province Key Laboratory of
Stomatology, Sun Yat-Sen University Guangzhou, concluded that 66–93% of rotary instrument fractures were
Guangzhou, China due to cyclic fatigue failure while 7–34% were due to shear
3
Department of Materials Engineering, UBC, Vancouver, Canada failure [5, 6]. Fracture due to fatigue occurs when an instru-
4 ment freely rotates in a curved canal. Unlike torsional
Center for Biomedical Materials and Engineering, College of
Material Science and Chemical Engineering, Harbin Engineering overloading, files subject to cyclic fatigue will not demon-
University, Harbin, China strate visible signs of defects or changes in shape and the
Clin Oral Invest

fracture point will occur at the canal’s point of maximum Ozyurek et al. [14] that studied the fatigue resistance of recip-
curvature [1]. The risk of fatigue failure of a NiTi rotary in- rocating instruments were used to calculate effective sample
strument is affected by the anatomy of the root canal and the size while the significance level was set at 0.05 and power was
file’s cyclic fatigue resistance, which is dependent on its met- set at 0.80. The software calculation showed that 8 files per
allurgy and design. Geometric properties of canals that in- group were necessary, and the decision was made to include
crease the risk of fatigue failure include short radii of curva- 10 files in each group. As there are 20 groups, 200 total files
ture, large angles of curvature, and curvatures with long arc (50 of each file system) were used for the experiments.
lengths [1, 7, 8]. Additionally, canals with curvatures located
closer to the canal orifice and canals with double curvatures Cyclic fatigue test
(S-shaped canals) have been shown to increase the risk of
fatigue failure of rotary instruments [9, 10]. As evidence of Five ceramic artificial canals were milled in an InCoris ZI
fatigue cannot be gathered by visual inspection of the file zirconium oxide disc (Dentsply Sirona) using the inLab MC
during treatment, it is important for clinicians to choose a X5 Digital computer-aided design and computer-aided
rotary instrument system with an appropriate fatigue resis- manufacturing (CAD/CAM) System (Dentsply Sirona).
tance for the root canal anatomy at hand. Each artificial canal was 16 mm in length and formed a ta-
Reciprocating motion was applied to NiTi instruments in pered groove on one side of the zirconium oxide block [19].
2008. Yared [11] was the first author to report the use and effec- The most coronal extent of the canal, the canal orifice, was
tiveness of reciprocating movements for rotary files, as he dem- 1.31 mm in diameter and constantly tapered by 0.06 mm per
onstrated that a single reciprocating file was sufficient to replace every millimeter of canal length, resulting in a 0.35-mm di-
multiple files used in continuous rotation. Reciprocating files ameter for the apical extent of the canal. Each canal had a
engage and debride dentine when rotating in the cutting direction single curvature with a radius of curvature of 3 mm and an
and disengage in the opposite direction. Reciprocating move- angle of curvature of 60°. The five artificial canals differed
ments appear to decrease the risk of file fracture caused by fa- only in the location of their curvatures along the length of the
tigue failure, when compared with continuous rotation [12, 13]. canal. The distance between the canal orifice to the start of the
Reciprocating movements result in the file experiencing fewer curve will be abbreviated as DOC (distance from the orifice to
complete rotations and fewer tensile-compressive stress cycles the curvature). The canals were separated into 5 groups, based
compared with continuous rotation, which may account for the on the location of where the curvature began (Fig. 1); group 1:
increased cyclic fatigue resistance [12]. WaveOne (WO) DOC of 5 mm; group 2: DOC of 6 mm; group 3: DOC of 8
(Dentsply Sirona, York, PA, USA) and Reciproc (Rec) mm; group 4: DOC of 10 mm; group 5: DOC of 11 mm.
(VDW, Munich, Germany) are among the earliest manufactured Polyvinyl siloxane impression material was used to create a
NiTi rotary files intended for reciprocating movements. Years custom jig that provided reproducible staging for the artificial
later, heat-treated reciprocating files, WaveOne Gold (WOG) canals. The artificial canal was placed inside a glass container
(Dentsply Sirona) and Reciproc Blue (RecB) (VDW), were re- that was filled with deionized water. Within the glass contain-
leased and demonstrated increased flexibility compared with er, plastic plumber screws were used to stabilize the fiber glass
their predecessors. WOG and RecB have been shown to be a to the zirconium oxide block. The glass container was placed
safer alternative to WO and Rec in curved canals due to their on a hotplate (Radnor, PA, USA), which kept the deionized
relatively superior cyclic fatigue resistance [14–18]. While fa- water at a constant temperature of 37 ± 2 °C [19, 20]. The
tigue resistance studies of these reciprocating file systems are following NiTi instruments were subject to fatigue testing:
available, there is a gap in the literature concerning the effect WO Primary (25/0.08), WOG Primary (25/0.07), Rec-25
of the canal’s curvature location on the fatigue resistance of these (25/0.08), and RecB-25 (25/0.08).
instruments. Therefore, the aim of the study was to examine the A total of 50 files of each instrument system were used and
effect of the location of canal curvature on the fatigue resistance all files were 25 mm in length for the fatigue test. Each instru-
of reciprocating NiTi files, and analyze the thermal behavior and ment system was evenly distributed to the five groups of ca-
phase transformation temperatures of Rec, RecB, WO, and nals. The hotplate was connected to a stainless steel frame-
WOG using differential scanning calorimetry (DSC). work that included a fastened utility clamp, which was used to
secure the motor handpiece in position (Fig. 2). Rec and RecB
files were used with a 6:1 reduction handpiece connected to a
Materials and methods VDW Silver motor (VDW Munich). WO and WOG files were
used with an 8:1 reduction handpiece connected to a ProMark
Sample size calculation Endo motor (Dentsply Sirona). As per the manufacturer’s in-
structions, Rec and RecB were operated with “RECIPROC
Sample size was determined using G*Power version 3.1.9.4 ALL” settings while WO and WOG were operated with
software (Brunsbuttel, Germany). The results reported by “WAVEONE ALL” settings. The files were placed to the full
Clin Oral Invest

Fig. 1 Schematic diagrams of five curvature locations of the canals. The distance from the orifice to the curvature of 5 mm (a), 6 mm (b), 8 mm (c),
10 mm (d), and 11 mm (e)

length of the artificial canal and operated until fracture oc- DSC
curred. Each experiment was recorded with a video camera
that was capable of high-definition magnification (Panasonic, Eight brand new instruments, including two from each of the
Mississauga, Ontario, Canada); each video was reviewed to reciprocating instrument systems, were subjected for DSC
determine the time to fracture (TTF) in seconds. The number testing. The instruments were separated into two fragments
of cycles to fracture (Nf) was calculated by dividing the TTF using a diamond bur under copious water coolant. The apical
by 60 s/min and multiplying by the number of rotations per fragments, consisting of D0–D4 of the rotary file, and the
minute (rpm) that each file operates at. Rec and RecB operate coronal segments, consisting of D7–D11 of the rotary file,
at 300 rpm (via the “RECIPROC ALL” setting) while WO were assessed.
and WOG operate at 350 rpm (via the “WAVEONE ALL” The DSC2500 machine (TA Instruments, New Castle, USA)
setting) [14]. was used in this study. The reference sample was an empty Tzero
Each fractured fragment was collected and its length was pan sealed with a Tzero Hermetic Lid. With the use of liquid
measured under magnification by means of a surgical operat- nitrogen, the sample was initially cooled to – 75 °C. When the
ing microscope (Global, St. Louis, MO, USA). Two frag- sample reached – 75 °C, it was then heated at 10 °C/min until it
ments of each instrument system per group were randomly reached 100 °C. After a brief pause, the sample was then re-
chosen for fractographic examination. The fragments were cooled to − 75 °C at a rate of 10 °C/min. The data was transferred
submerged in absolute alcohol and ultrasonically cleaned to TRIOS software (TA Instruments, New Castle, USA), which
(Endo Ultra, MicroMega, Besancon, France) for 60 sec. The was used to analyze the thermal behavior of each specimen. The
fragments were mounted with the fractured surface facing austenitic transformation start and finish points (As and Af) and
upwards and were submitted for scanning electron microsco- the martensitic reverse transformation start and finish points
py (SEM) (Hitachi High-Technologies Canada Inc, Toronto, (Ms and Mf) were determined on the thermograms by using the
Canada) at a magnification of × 110–× 1000 using 3.0 kV. TRIOS software function “Onset Point” [21].

Fig. 2 The setup of the fatigue


test. a The water container and
thermometer were placed on the
heating device and the handpiece
secured in a utility clamp that was
secured to a stainless-steel frame-
work. b Arrangement of the arti-
ficial canals
Clin Oral Invest

Statistical analysis all groups (p < 0.05). WOG had significantly greater fatigue life
than WO when used in a canal with a DOC of 5 mm, 8 mm, 10
Statistical analysis was performed with SPSS software version mm, and 11 mm (p < 0.05) while no significant differences
3.1.1.0 (IBM, Armonk, NY, USA). The normality of the data existed between these files when the canal’s DOC was 6 mm
was determined by using the Kolmogorov-Smirnov test. The (p > 0.05). WOG had significantly greater fatigue resistance
equality of variance was assessed with Levene’s test. Two-way than Rec in canals with a DOC of 5 mm, 8 mm, and 10 mm
analysis of variance (ANOVA) was used to compare the aver- (p < 0.05) while there were no differences between these files
age Nf of the study groups. Post hoc multiple comparison with when used in canals with a DOC of 6 mm or 11 mm (p > 0.05).
the Tukey test was used to identify and confirm statistically Compared with WO, Rec had superior fatigue resistance in the
significant differences between groups. The Kruskal-Wallis test canal with a DOC of 11 mm (p < 0.05).
was used to compare the mean fragment lengths between The mean values of the length of the fractured frag-
groups. The statistical significance level, α, was set at 0.05. ments are displayed in Fig. 4. There were no significant
differences in length of the fractured fragments between
the reciprocating file systems (p > 0.05) except between
Results WOG and Rec in the canal with a DOC of 11 mm (p < 0.05).
Based on the mean length of the file fragments, the file
Comparisons of the Nf between the reciprocating file sys- fractures occurred at or near the point of maximum
tems are presented in Fig. 3. All the reciprocating files curvature of each canal. The surfaces of the fractured file
showed significantly lower Nf when they were used in fragments all showed evidence of fatigue failure. The
canals with a DOC of 5 mm, 6 mm, and 8 mm when crack initiation points were seen at stress concentration
compared with canals with a DOC of 10 mm or 11 mm areas (Figs. 5 and 6).
(p < 0.05). Additionally, WO, Rec, and RecB showed The DSC thermograms shown in this study include a
significantly higher fatigue resistance when used in a ca- cooling curve at the top of the figure and a heating curve
nal with a DOC of 11 mm when compared with a canal at the bottom of the figure (Fig. 7). The mean phase trans-
with a DOC of 10 mm (p < 0.05). formation temperatures for each file system were
The only significant difference in fatigue resistance between determined and are displayed in Table 1. The average aus-
WOG and RecB occurred when the DOC was 11 mm (p < tenite start temperatures (As) found for WO (26.22 °C),
0.05). RecB had significantly higher Nf than WO and Rec in WOG (35.4 °C), Rec (26.6 °C), and RecB (30.2 °C) were

Fig. 3 The Nf of reciprocating


instruments at different curvature
locations. Different letters
indicate statistically significant
differences within each group (p <
0.05)
Clin Oral Invest

Fig. 4 The mean length of the


fractured fragments of the
reciprocating instruments at
different curvature location.
Different letters indicate
statistically significant differences
within each group (p < 0.05)

all below body temperature (~ 37 °C). The average austen- martensite during cooling. The thermograms for RecB ex-
ite finish temperature (Af) of RecB (37.1 °C) approximated hibit a single exothermic peak in the cooling direction but
body temperature while the other files, WO (48.7 °C), double endothermic peaks in the heating direction (Fig. 7).
WOG (47.3 °C), and Rec (51.4 °C), exceeded body temper- It appears that RecB undergoes a one-step phase transfor-
ature. The thermograms for WO, WOG, and Rec exhibited mation from austenite to martensite when cooled while a
single peaks in both the heating and cooling directions. This two-step phase transformation from martensite to R-phase
suggests that WO, WOG, and Rec undergo a one-step phase to austenite takes place when the file is heated. The phase
transformation from martensite to austenite during heating transformation peaks associated with WO and Rec were
and a one-step phase transformation from austenite to broad while sharp peaks were noted with RecB.

Fig. 5 Scanning electron micrographs of fracture surfaces of Reciproc (Rec) and Reciproc Blue (RecB) files after fatigue tests at different curvature
locations (arrows point toward crack initiation points)
Clin Oral Invest

Fig. 6 The fracture surfaces of WaveOne (WO) and WaveOne Gold (WOG) files after fatigue tests in canals with a DOC of 5 mm, 8 mm, and 11 mm
(arrows point toward crack initiation points)

Discussion lacked taper) with an apical curvature versus a canal with a


mid-root curvature, in a dry environment at room temperature
Heat treatment is known to increase the flexibility and fatigue [8, 9]. These studies concluded that BioRace and Mtwo had
resistance of NiTi files; therefore, it is anticipated that WOG significantly reduced fatigue resistance in canals with a mid-
and RecB would take longer to fracture in the fatigue test root curvature than canals with an apical curvature [8, 9]. A
compared with their predecessors, WO and Rec [22, 23]. study by Arias et al. [24] compared the fatigue resistance of
While past studies have demonstrated improved fatigue resis- WO and Rec in non-tapered canals with a curvature located at
tance in WOG and RecB compared with WO and Rec, they 5 mm or 13 mm from the tip of the instrument and concluded
tend to use canals that have only apical curvatures [14–18]. that the more coronally located curvature (at 13 mm from the
There are few studies in the literature that compare the fatigue tip of the instrument) significantly decreased the fatigue resis-
resistance of files rotating in canals with different curvature tance of these reciprocating files. Apart from the three studies
locations. Two studies compared the fatigue resistance of con- mentioned, there appears to be an absence of literature regard-
tinuously rotating files (in a curved cylindrical canal that ing the effect of canal curvature location on the fatigue

Fig. 7 Differential scanning


calorimetry curves of Reciproc,
Reciproc Blue, WaveOne, and
WaveOne Gold instruments.
Cooling (upper) and heating
(lower) curves are shown
Clin Oral Invest

Table 1 Mean phase


transformation temperatures (± Cooling Heating
standard deviation) of the
reciprocating instrument systems File system Ms (°C) Mf (°C) Rs (°C) As (°C) Af (°C) ΔH (J/g)
when the coronal and apical
fragments were combined into WO 45.10 ± 1.38 22.54 ± 1.01 N/A 26.22 ± 1.82 48.70 ± 1.67 1.54 ± 0.36
one group WOG 44.40 ± 0.30 26.26 ± 1.00 N/A 35.36 ± 0.80 47.26 ± 1.02 2.31 ± 0.51
Rec 44.88 ± 0.64 23.72 ± 2.09 N/A 26.76 ± 1.23 51.42 ± 1.62 1.98 ± 0.32
RecB 32.55 ± 1.00 24.09 ± 1.23 12.71 ± 0.90 30.24 ± 0.22 37.08 ± 0.60 9.10 ± 1.08

resistance of new reciprocating files. In addition, recent re- fulfills this assumption as it should be almost completely in the
ports have shown that file fatigue tests done at room temper- martensitic phase with As and Af values of 35.4 °C and 47.3 °C,
ature do not predict file behavior at body temperature, espe- respectively, if intracanal temperatures of 35 ± 1 °C are to be
cially for heat-treated files [19, 20, 25–28]. In order to mimic assumed [29]. In a recent paper, Garcia et al. [31] used DSC
clinical situations where coronal, middle, and apical curved and X-ray diffraction to analyze WO and WOG files. In addi-
canals are encountered, the present study included five loca- tion to presenting similar phase transformation values for WO
tions of curvatures to evaluate fatigue resistance. The artificial and WOG as the current study, they also found the presence of
canals were customized to be similar in shape and taper to the austenite and R-phase in WO and WOG at room temperature
instrument to ensure a close fit without binding. The canals [32]. Unfortunately, due to peak superposition, Garcia et al.
were designed to have an angle of 60° and a radius of curva- were not able to determine the amount of each phase present
ture of 3 mm as this was the only possible combination that in these files at room temperature [31].
allowed the five different curvatures locations to share the While RecB had a higher As (30.2 °C) than WO and Rec, it
same radius, angle of curvature, and arc length. also had the lowest Af of all the tested files in this study with a
In this study, RecB had superior fatigue resistance com- value of 37.1 °C. Based on these results, it would seem likely that
pared with Rec in all of the artificial canals while WOG had RecB would be mostly austenite during clinical root canal treat-
superior fatigue resistance compared with WO in all canals ment, which is unexpected due to its superior fatigue resistance
except for the canal with a DOC of 6 mm. Differences in compared with WO and Rec, which showed higher Af values.
fatigue resistance between instruments may be due to differ- The thermogram for RecB in the current study consists of sharp
ences in diameter, taper, metallurgy, and cross-section of the peaks, which are characteristic of NiTi with a high R-phase com-
rotary file [1, 7, 23]. The reciprocating instruments in our position, whereas broad peaks, which are characteristic of a high-
study fractured significantly faster due to fatigue in canals austenite composition, can be seen with thermograms for WO,
with a DOC of 5 mm, 6 mm, and 8 mm compared with canals WOG, and Rec [31, 32]. With the use of X-ray diffraction,
with a DOC of 10 mm and 11 mm. Studies have shown that Generali et al. [32] found that the major phase composition of
continuous rotary files with a larger diameter and/or taper will RecB was R-phase and austenite while small amounts of mar-
have lower fatigue resistance than rotary files with smaller tensite were present at room temperature. In the same study, they
diameters and/or tapers [1, 7]. Therefore, caution needs to be found that Rec was primarily austenite while R-phase and mar-
exercised when using reciprocating instruments in canals with tensite could also be detected in the samples [32].
coronal or middle curvatures. According to the DSC results of the present study, the en-
The most important temperature-based characteristic of NiTi thalpy changes (ΔH) of RecB were much higher than that of the
rotary instruments during clinical endodontic treatment is the other instruments, indicating a greater amount of transformable
potential phase transformation of martensite to austenite. To martensite and austenite. During the machining of endodontic
achieve maximum possible fatigue resistance, the NiTi instru- files, the austenite and martensite are inevitably stabilized and
ment should have both the As and Af above 35 ± 1 °C, which is are unable to undergo transformation. Compared with Rec, the
the clinical intracanal temperature [29]. If the As and/or Af heat treatment of RecB helps to relieve stress and eliminate
values of the NiTi are between room temperature (approximate- physical defects arising during processing, which facilitate the
ly 23 °C) and the clinical intracanal temperature, the file will phase transformation. Generali et al. [32] attributed the higher
start phase transformation when used in the root canal. The fatigue resistance of RecB to the grain refinement and different
reciprocating files used in this study had mean As values that phase transformation temperatures. Combined with the results
ranged from 26.2 to 35.4 °C and Af values that ranged from of the present study, we can assume that recrystallization may
37.18 to 51.4 °C (Table 1). It would be reasonable to assume occur during heat treatment of RecB, which could produce
that the heat-treated files, WOG and RecB, have high As and Af small grains with less defects while lowering the material’s
values as their superior flexibility and fatigue resistance over strength and hardness. This was also verified by Generali
WO and Rec is suggestive of martensitic alloy [21, 30]. WOG et al. who showed that Blue heat-treated (RecB) samples had
Clin Oral Invest

a lower elastic modulus and a significantly lower hardness analysis of fractured surface in a cohort study. J Endod 35:133–136.
https://doi.org/10.1016/j.joen.2008.10.013
when compared with their M-wire (Rec) counterparts. While
7. Haikel Y, Serfaty R, Bateman G, Senger B, Allemann C (1999)
elevations in phase transformation temperatures are an obvious Dynamic and cyclic fatigue of engine-driven rotary nickel-titanium
benefit of heat-treatment of NiTi rotary instruments, these find- endodontic instruments. J Endod 25:434–440. https://doi.org/10.
ings suggest that alteration of the phase composition of these 1016/S0099-2399(99)80274-X
files may be the most significant benefit of heat treatment when 8. Lopes HP, Vieira MVB, Elias CN, Goncalves LS, Siqueira JF,
Moreira EJL, Vieira VTL, Souza LC (2013) Influence of the geom-
considering fatigue resistance. etry of curved artificial canals on the fracture of rotary nickel-
In conclusion, reciprocating instruments experience lower titanium instruments subjected to cyclic fatigue tests. J Endod 39:
fatigue resistance in canals with middle- or coronally located 704–707. https://doi.org/10.1016/j.joen.2012.12.027
curvatures when compared with canals with apical curvatures. 9. Lopes HP, Chiesa WMM, Correia NR, Navegante NCD, Elias CN,
Moreira EJL, Chiesa BEC (2011) Influence of curvature location
WOG and REB have superior CFR when compared with WO
along an artificial canal on cyclic fatigue of a rotary nickel-titanium
and RE. When heated, WO, WOG, and RE undergo a one- endodontic instrument. Oral Surg Oral Med Oral Pathol Oral Radiol
stage phase transformation from martensite to austenite, while Endod 111:792–796. https://doi.org/10.1016/j.tripleo.2010.12.006
REB undergoes a two-stage phase transformation from mar- 10. Al-Sudani D, Grande NM, Plotino G, Pompa G, Di Carlo S,
tensite to R-phase to austenite. Testarelli L, Gambarini G (2012) Cyclic fatigue of nickel-
titanium rotary instruments in a double (S-shaped) simulated cur-
vature. J Endod 38:987–989. https://doi.org/10.1016/j.joen.2012.
Acknowledgment The authors thank VDW for donating the files used in 03.025
this study. 11. Yared G (2008) Canal preparation using only one Ni-Ti rotary
instrument: preliminary observations. Int Endod J 41:339–344.
Author contribution T. S. and X. H. contributed equally to this work. https://doi.org/10.1111/j.1365-2591.2007.01351.x
12. De-Deus G, Moreira EJL, Lopes HP, Elias CN (2010) Extended
Funding information The work was supported by the Canada cyclic fatigue life of F2 ProTaper instruments used in reciprocating
Foundation for Innovation (CFI 32623) movement. Int Endod J 43:1063–1068. https://doi.org/10.1111/j.
1365-2591.2010.01756.x
13. Pedulla E, Grande NM, Plotino G, Gambarini G, Rapisarda E
Compliance with ethical standards (2013) Influence of continuous or reciprocating motion on cyclic
fatigue resistance of 4 different nickel-titanium rotary instruments. J
Conflict of interest The authors declare that they have no conflict of Endod 39:258–261. https://doi.org/10.1016/j.joen.2012.10.025
interest. 14. Ozyurek T (2016) Cyclic fatigue resistance of Reciproc, WaveOne,
and WaveOne Gold nickel-titanium instruments. J Endod 42:1536–
Ethical approval This article does not contain any studies with human 1539. https://doi.org/10.1016/j.joen.2016.06.019
participants or animals performed by any of the authors. 15. Keskin C, Ivan U, Demiral M, Keles A (2017) Cyclic fatigue resis-
tance of Reciproc Blue, Reciproc, and WaveOne Gold reciprocat-
Informed consent For this type of study, formal consent is not required. ing instruments. J Endod 43:1360–1363. https://doi.org/10.1016/j.
joen.2017.03.036
16. Al-Obaida MI, Merdad K, Alanazi MS, Altwaijry H, AlFaraj M,
Alkhamis AA, Al-Madi EM (2019) Comparison of cyclic fatigue
resistance of 5 heat-treated nickel-titanium reciprocating systems in
References canals with single and double curvatures. J Endod 45:1237–1241.
https://doi.org/10.1016/j.joen.2019.06.011
1. Pruett JP, Clement DJ, Carnes DL (1997) Cyclic fatigue testing of 17. Keles A, Ozyurek EU, Uyanik MO, Nagas E (2019) Effect of tem-
nickel-titanium endodontic instruments. J Endod 23:77–85. https:// perature of sodium hypochlorite on cyclic fatigue resistance of heat-
doi.org/10.1016/S0099-2399(97)80250-6 treated reciprocating files. J Endod 45:205–208. https://doi.org/10.
2. Spili P, Parashos P, Messer HH (2005) The impact of instrument 1016/j.joen.2018.11.003
fracture on outcome of endodontic treatment. J Endod 31:845–850. 18. Scott R, Arias A, Macorra JC, Govindjee S, Peters OA (2019)
https://doi.org/10.1097/01.don.0000164127.62864.7c Resistance to cyclic fatigue of reciprocating instruments determined
3. Iqbal MK, Kohli MR, Kim JS (2006) A retrospective clinical study at body temperature and phase transformation analysis. Aust Endod
of incidence of root canal instrument separation in an endodontics J. 45:400–406. https://doi.org/10.1111/aej.12374
graduate program: a PennEndo database study. J Endod 32:1048– 19. Huang X, Shen Y, Wei X, Haapasalo M (2017) Fatigue resistance
1052. https://doi.org/10.1016/j.joen.2006.03.001 of nickel-titanium instruments exposed to high-concentration hypo-
4. Tzanetakis GN, Kontakiotis EG, Maurikou DV, Marzelou MP chlorite. J Endod 43:1847–1851. https://doi.org/10.1016/j.joen.
(2008) Prevalence and management of instrument fracture in the 2017.06.033
postgraduate endodontic program at the dental school of athens: a 20. Shen Y, Huang X, Wang Z, Wei X, Haapasalo M (2018) Low
five-year retrospective clinical study. J Endod 34:675–678. https:// environmental temperature influences the fatigue resistance of
doi.org/10.1016/j.joen.2008.02.039 nickel-titanium files. J Endod 44:626–629. https://doi.org/10.
5. Cheung GSP, Peng B, Bian Z, Shen Y, Darvell BW (2005) Defects 1016/j.joen.2017.11.004
in ProTaper S1 instruments after clinical use: fractographic exami- 21. Miyai K, Ebihara A, Hayashi Y, Doi H, Suda H, Yoneyama T
nation. Int Endod J 38:802–809. https://doi.org/10.1111/j.1365- (2006) Influence of phase transformation on the torsional and bend-
2591.2005.01020.x ing properties of nickel-titanium rotary endodontic instruments. Int
6. Shen Y, Cheung GSP, Peng B, Haapasalo M (2009) Defects in Endod J 39:119–126. https://doi.org/10.1111/j.1365-2591.2006.
nickel-titanium instruments after clinical use. part 2: fractographic 01055.x
Clin Oral Invest

22. Kuhn G, Jordan L (2002) Fatigue and mechanical properties of 28. Hulsmann M, Donnermeyer D, Schafer E (2019) A critical apprais-
nickel-titanium endodontic instruments. J Endod 28:716–720. al of studies on cyclic fatigue resistance of engine-driven endodon-
https://doi.org/10.1097/00004770-200210000-00009 tic instruments. Int Endod J 52:1427–1445. https://doi.org/10.1111/
23. Yahata Y, Yoneyama T, Hayashi Y, Ebihara A, Doi H, Hanawa T, iej.13182
Suda H (2009) Effect of heat treatment on transformation temper- 29. de Hemptinne F, Slaus G, Vandendael M, Jacquet W, De Moor RJ,
atures and bending properties of nickel-titanium endodontic instru- Bottenberg P (2015) In vivo intracanal temperature evolution dur-
ments. Int Endod J 42:621–626. https://doi.org/10.1111/j.1365- ing endodontic treatment after the injection of room temperature or
2591.2009.01563.x preheated sodium hypochlorite. J Endod 41:1112–1115. https://doi.
24. Arias A, Perez-Higueras JJ, de la Macorra JC (2012) Differences in org/10.1016/j.joen.2015.02.011
cyclic fatigue resistance at apical and coronal levels of Reciproc and 30. Hieawy A, Haapasalo M, Zhou H, Wang ZJ, Shen Y (2015) Phase
WaveOne new files. J Endod 38:1244–1248. https://doi.org/10. transformation behavior and resistance to bending and cyclic fa-
1016/j.joen.2012.05.022 tigue of ProTaper Gold and ProTaper Universal instruments. J
25. de Vasconcelos RA, Murphy S, Carvalho CA, Govindjee RG, Endod 41:1134–1138. https://doi.org/10.1016/j.joen.2015.02.030
Govindjee S, Peters OA (2016) Evidence for reduced fatigue resis- 31. Garcia PR, Resende PD, Lopes NIA, Peixoto I, Buono VTL, Viana
tance of contemporary rotary instruments exposed to body temper- ACD (2019) Structural characteristics and torsional resistance eval-
ature. J Endod 42:782–787. https://doi.org/10.1016/j.joen.2016.01. uation of WaveOne and WaveOne Gold instruments after simulated
025 clinical use. J Endod 45:1041–1046. https://doi.org/10.1016/j.joen.
26. Arias A, Macorra JC, Govindjee S, Peters OA (2018) Correlation 2019.04.009
between temperature-dependent fatigue resistance and differential 32. Generali L, Puddu P, Borghi A, Brancolini S, Lusvarghi L, Bolelli
scanning calorimetry analysis for 2 contemporary rotary instru- G, Consolo U, Pedulla E (2019) Mechanical properties and metal-
ments. J Endod 44:630–634. https://doi.org/10.1016/j.joen.2017. lurgical features of new and ex vivo used Reciproc Blue and
11.022 Reciproc. Int Endod J. 53:250–264. https://doi.org/10.1111/iej.
27. Arias A, Hejlawy S, Murphy S, de la Macorra JC, Govindjee S, 13214
Peters OA (2019) Variable impact by ambient temperature on fa-
tigue resistance of heat-treated nickel titanium instruments. Clin
Oral Investig 23:1101–1108. https://doi.org/10.1007/s00784-018- Publisher’s note Springer Nature remains neutral with regard to jurisdic-
2543-6 tional claims in published maps and institutional affiliations.

You might also like