Professional Documents
Culture Documents
Related Publications:
“Caution” notes alert the user to the possibility of damage to the equipment.
CAUTION
NOTE This manual contains proprietary information of Getinge USA, Inc. It shall
not be reproduced in whole or in part without the written permission of
Getinge USA, Inc.
ii
Table of Contents
61301609701 iii
Section 3 Trouble Analysis
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–1
Service Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–1
TROUBLE ANALYSIS TOOLS . . . . . . . . . . . . . . . . . . . . . . 3–2
Viewing the Error Log . . . . . . . . . . . . . . . . . . . . . . . . . 3–3
Viewing the Analog and Digital Inputs . . . . . . . . . . . . . 3–5
Operating the Digital Outputs Manually. . . . . . . . . . . . 3–7
Viewing the System and User Flags . . . . . . . . . . . . . 3–10
TROUBLE ANALYSIS TABLES. . . . . . . . . . . . . . . . . . . . 3–12
Printer Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . 3–14
iv
Section 5 Maintenance and Repair
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–1
Service Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–1
PERIODIC MAINTENANCE SCHEDULE—STANDARD . . 5–2
CONTROL SYSTEM MAINTENANCE AND REPAIR . . . . 5–3
Replacing the Temperature Sensors (RTDs). . . . . . . . 5–3
Replacing the Pressure Transducer . . . . . . . . . . . . . . 5–5
Replacing the Flash EPROM . . . . . . . . . . . . . . . . . . . 5–11
PIPING SYSTEM MAINTENANCE AND REPAIR . . . . . . 5–23
Check the Pressure Relief Valve . . . . . . . . . . . . . . . . 5–23
Clean the Water and Steam Strainers . . . . . . . . . . . . 5–23
Clean the Water Solenoid Valve . . . . . . . . . . . . . . . . 5–23
Replace the Air Filter . . . . . . . . . . . . . . . . . . . . . . . . . 5–24
DOOR MAINTENANCE AND REPAIR. . . . . . . . . . . . . . . 5–25
Opening the Door Manually (Manual Gasket Retract) 5–25
Door Gasket Replacement. . . . . . . . . . . . . . . . . . . . . 5–27
Replace And Adjust the Door Cable . . . . . . . . . . . . . 5–30
Replacing the Door Spacers . . . . . . . . . . . . . . . . . . . 5–31
61301609701 v
Selecting a Cycle. . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–14
PARAMETER SOFTKEY . . . . . . . . . . . . . . . . . . . . . . . . . 7–15
Editing the Cycle Parameters (Quick Edit Mode) . . . 7–15
Editing the Cycle Parameters (Full Edit Mode) . . . . . 7–17
SYSTEM MENU. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–19
TIME SETTINGS MENU . . . . . . . . . . . . . . . . . . . . . . . . . 7–20
Setting the Clock and Calendar. . . . . . . . . . . . . . . . . 7–21
Setting the Alarm Clock . . . . . . . . . . . . . . . . . . . . . . . 7–22
CALIBRATION MENU . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–23
Calibrating the Temperature Sensors and Pressure
Transducer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–24
Entering the Bridge Compensation Table Data . . . . . 7–26
DOCUMENTATION MENU . . . . . . . . . . . . . . . . . . . . . . . 7–27
Printing a Program Listing . . . . . . . . . . . . . . . . . . . . . 7–28
Setting Up the Printer Log Interval . . . . . . . . . . . . . . 7–29
SettingUp the Printer Log Values Definition . . . . . . . 7–30
CONFIGURATION MENU . . . . . . . . . . . . . . . . . . . . . . . . 7–32
Changing the Cycle Name and Number . . . . . . . . . . 7–33
Assigning the Sterilizer Name . . . . . . . . . . . . . . . . . . 7–34
Selecting the Language, Date Format, and Units . . . 7–36
Setting Up the Display Panel. . . . . . . . . . . . . . . . . . . 7–38
PASSWORDS MENU . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–40
Configuring Password Access for Individual Users . . 7–40
DIPSWITCHES MENU. . . . . . . . . . . . . . . . . . . . . . . . . . . 7–42
SERVICE MENU . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–43
Viewing the Error Log . . . . . . . . . . . . . . . . . . . . . . . . 7–44
Viewing the Analog and Digital Inputs . . . . . . . . . . . . 7–45
Operating the Digital Ouptuts . . . . . . . . . . . . . . . . . . 7–48
Viewing the User and System Flags . . . . . . . . . . . . . 7–49
Testing the Printer and LEDs . . . . . . . . . . . . . . . . . . 7–51
SAVE RAM TO FLASH . . . . . . . . . . . . . . . . . . . . . . . . . . 7–52
vi
WIRING DIAGRAM, CONTROL BOX 2A. . . . . . . . . . . . . . .9–4
WIRING DIAGRAM, OPTIONAL EQUIPMENT . . . . . . . . . .9–5
ELECTRICAL SCHEMATIC, 433HCMC & 533HCMC . . . . .9–6
ELECTRICAL SCHEMATIC, POWER BOX 1A . . . . . . . . . .9–7
ELECTRICAL SCHEMATIC, CONTROL OUTPUT . . . . . . .9–8
WIRING DIAGRAM, MP-129A WATER-SAVER . . . . . . . .9–10
WIRING DIAGRAM, MP-129F WATER SAVER . . . . . . . .9–11
Index
61301609701 vii
viii
DESCRIPTION OF SYMBOLS ON THE EQUIPMENT
Switches
Open Door — Unseals the Door
Clear Alarm
!
Start
Enter
Up Arrows (for selecting softkey options (i.e. Save, Cancel, OK, and so
forth)
Indicators
Door(s) Closed
Door(s) Sealed
In Process
Process Complete
Process Failure
!
61301609701 ix
Labels The following label on the sterilizer alerts personnel to possible hazards:
Hot Surfaces
The following labels on the Control Box and Power Box alert service
personnel to possible hazards:.
High Voltage
CAUTION: To reduce the risk of electrical shock, do not remove cover. Refer
servicing to qualified service personnel.
Attention
OBSERVE PRECAUTIONS
FOR HANDLING
SENSITIVE
DEVICES
x
DESCRIPTION OF ICONS USED IN THIS MANUAL
SERVICE This procedure requires a password. See page 1-14 for a list of password
access areas.
ACCESS REQUIRED
61301609701 xi
Manual Conventions Before you begin using this manual, it is important to understand the
conventions used. These conventions are established for visual ease of use.
ALL CAPS Switches and indicators on the control panel, such as ENTER.
> The > symbol leads you through nested menu items and dialog box options
to a final action. For example: SYSTEM MENU > enter password >
CONFIGURATION > EDIT CYCLES.
Italic Upper/Lower Case Cross-reference to another manual or guide, such as see EDIT CYCLE
PARMAMETERS on page 7-15.
xii
Section 1 General Information
INTRODUCTION
The 433HCMC and 533HCMC Vacuum Steam Sterilizers are intended for
use by health care facilities and to be used to sterilizer wrapped and
unwrapped surgical instruments, linens, and liquids (liquids not intended for
patient contact) by means of pressurized steam.
Six cycle types are available: Prevacuum, Gravity, Flash 3+, Flash 10+,
Liquids, and Leak Test. The factory set parameters are listed in Table 1–1.
The cycle parameters can be adjusted by a user with parameter password
access (see page 1–14).
Gravity cycles are intended for wrapped goods, while the Flash 3+ cycle is
intended for unwrapped non-porous items and the Flash 10+ cycle for
unwrapped porous and/or non-porous items.
NOTES
• See “CYCLE TYPES” on page 1–15 for cycle details.
• Refer to the Load Chart in the Operator Manual for load composition and
maximum number of items.
• Gravity cycles include five conditioning pulses; Flash 10+ cycles, two
pulses.
• The PREVAC 3 cycle may be used to process unwrapped porous and
non-porous items.
• Vacuum leak test cycle parameters are not adjustable.
61301609701 1–1
General Information
1–2
433HCMC & 533HCMC Steam Sterilizers
STERILIZER OPERATION
NOTE
Refer to Section 7, “Service Menu” for an in-depth description of the user
interface.
CONTROLS
ON / OFF
A09701-M
61301609701 1–3
General Information
Controls and Indicators The sterilizer controls and indicators are shown in Figure 1–2.
8 6
5
7 CONTROLS
1
ON / OFF
!
4
!
2
9 3 WS-0047A
1–4
433HCMC & 533HCMC Steam Sterilizers
WS-0091
• If the switch is illuminated but the display screen is blank, the controls are
NOTES
in screensaver (sleep) mode. Press any button on the control panel to
wake up the controls.
• The CONTROLS OFF/ON switch does not turn off 115 VAC power to the
sterilizer. To remove all power, turn OFF the sterilizer at the mains
disconnect switch.
• If the controls are accidentally turned OFF, press CONTROLS OFF/ON
immediately to turn the controls back ON. The cycle will not be aborted if
the controls are turned back ON within 60 seconds.
61301609701 1–5
General Information
0
REMAINING TIME
• Press [SETUP] to access the setup and service menus. See “SETUP
SOFTKEY” on page 7–9 for a detailed description.
• Press [SELECT CYCLE] to access the cycle list. See “SELECT CYCLE
SOFTKEY” on page 7–14 for a detailed description.
• Press [PARAMETER] to view and change the cycle parameter settings.
See “PARAMETER SOFTKEY” on page 7–15 for a detailed description.
• Press [UNSEAL] to unseal the door.
1–6
433HCMC & 533HCMC Steam Sterilizers
(7) PRINTER
Printer components are shown in Figure 1–4. Refer to the Operator Manual
for a detailed description of the printer.
Roll
Cover Status
STATUS
Lever Copy
Feed
Retainer
Shaft
Clear
Cover
Tab
00-01A
Cover Slot
Slot Notch
Core
61301609701 1–7
General Information
1–8
433HCMC & 533HCMC Steam Sterilizers
61301609701 1–9
General Information
Fault and Failure Conditions If a fault or failure condition occurs, PROCESS FAILURE lights and an
alarm sounds. A diagnostic message will be displayed and printed out. See
Section 2, “Diagnostic Messages” for a description of the condition and
corrective actions.
Press CLEAR/ALARM to silence the alarm.
!
2. Press [ABORT].
WS-0342
1–10
433HCMC & 533HCMC Steam Sterilizers
Advancing a Cycle Phase This feature may be used to terminate a timed phase and advance to the
(Manual Phase Advance) next phase. Timed phases include PURGE, LIQ DWELL, EXPOSURE, LIQ
EXPOSURE, DRYING, EQUALIZATION, and LEAK TEST.
NOTE
The cycle advance feature is intended for use by qualified service personnel
only. Improper use of this feature can compromise the integrity of the
processing cycle.
To advance a cycle:
1. Press [MORE], then
SETUP PARA- MORE
METER
WS-0344
WS-0343
The CYCLE ADVANCED message displays for 5 seconds and is printed out.
After a cycle phase is advanced, the cycle continues through the air-in
phase but the controls enter STANDBY mode, bypassing the CYCLE
COMPLETE phase.
To unseal the door and remove the load, press UNSEAL while the controls
are in standby mode.
NOTE
If the phase is manually advanced, it may be necessary to manually retract
the gasket in order to open the door. See “Opening the Door Manually
(Manual Gasket Retract)” on page 5–25.
61301609701 1–11
General Information
CONFIGURATIONS
Door Configurations All models may be either a single door or a double door unit.
NOTE
The door will not unseal if the pressure in the chamber is not within 2 PSI of
atmospheric pressure.
• Abbreviations used for the doors in the text, drawings, and diagnostic
NOTES
messages are as follows:
• CE = Control End — The end of the unit where most programming and
settings occur during use; usually the “load end” of a double door unit.
The printer is at this end.
• RE= Remote End — Usually the “unload end” of a double door unit.
1–12
433HCMC & 533HCMC Steam Sterilizers
Loading Equipment 400HCMC Series units (16" x 16" chamber) are equipped with a rack and
extendible shelves.
500HCMC Series units (20" x 20" chamber) are either equipped with a rack
and extendible shelves or furnished with a track assembly designed to
accommodate a loading car and transfer carriage.
61301609701 1–13
General Information
Password Access The password system is designed to allow individual users to have various
levels of access privileges as listed in Table 1–2.
A Parameter Settings Edit the Cycle Parameters (Quick Edit Mode) 7-15
Edit the Cycle Parameters (Full Edit Mode) 7-17
J Password Setup Configure the Password Access for Individual Users 7-39
1–14
433HCMC & 533HCMC Steam Sterilizers
CYCLE TYPES
Gravity Cycles Gravity cycles use the gravity (downward displacement) method of air
removal. The Conditioning phase of these cycles includes pressure pulses
to aid in air removal.
Refer to Figure 1–6 for the adjustable parameters and their maximum,
minimum, and factory settings.
Exposure
Temp
Drying
Time Dry 1=vac, 2=grav
Atmospheric
Gravity Dry
Exposure
Time Time
Vacuum Dry
1 2 3 4 5 6 7 8 9
1 Cycle Start
2 Purge
3 Conditioning
Gravity 1 Gravity 2
4 Heat-up
Exposure Time 00:01:00 – 06:00:00 1 sec. 00:30:00 00:10:00
5 Exposure Exposure Temp 230.0 – 275.0˚F 0.1˚F (0.1˚C) 250.0˚F 275.0˚F
(110.0 – 135.0˚C)
6 Exhaust
Drying Time 00:00:10 – 12:00:00 1 sec. 00:30:00 00:30:00
7 Drying
Dry 1=vac, 2=grav 1=VAC 2=GRAV 00001 – 00002 1 00002 00002
8 Air-in
9 Cycle Complete
WS-0160 A
DRYING METHOD
61301609701 1–15
General Information
DRYING RECOMMENDATIONS
The actual drying time and load size may need to be adjusted based on a
variety of conditions including:
• load composition
• wrapping technique
• ambient conditions
• quality of utilities
Flash 3+ Cycle Flash 3+ cycles use the gravity (downward displacement) method of air
removal. The Conditioning phase of these cycles does not include pressure
pulses to aid in air removal.
Refer to Figure 1–7 for the adjustable parameters and their maximum,
minimum, and factory settings.
Exposure
Temp
Pressure
Drying
Time Dry 1=vac,2=grav
Gravity Dry
Atmospheric
Exposure
Time Vacuum Dry
Time
1 2 3 4 5 6 7 8 9
2 Purge
3 Conditioning
4 Heat-up
9 Cycle Complete
WS-0161 A
1–16
433HCMC & 533HCMC Steam Sterilizers
Flash 10+ Cycle Flash 10+ cycles use the gravity (downward displacement) method of air
removal. The Conditioning phase of these cycles includes two pressure
pulses to aid in air removal.
Refer to Figure 1–8 for the adjustable parameters and their maximum,
minimum, and factory settings.
Exposure
Temp
Pressure
Drying
Time Dry 1=vac, 2=grav
Gravity Dry
Atmospheric
Exposure
Time Vacuum Dry
Time
1 2 3 4 5 6 7 8 9
2 Purge
3 Conditioning
4 Heat-up
9 Cycle Complete
WS-0162 A
61301609701 1–17
General Information
Liquids Cycle Liquids cycles use the gravity method of air removal. The cooldown rate
during the Liquid Exhaust phase is computer-controlled.
Refer to Figure 1–9 for the adjustable parameters and their maximum,
minimum, and factory settings.
Atmospheric
Exposure Time
Time
1 2 3 4 5 6 7 8
2 Purge
3 Heat-up
Liquids 1 Liquids 2
4 Liquid Dwell 00:00:00 – 00:30:00 1 sec. 00:08:00 00:08:00
5 Liquid Exposure 219.0 – 275.0˚F
Exposure Temp 0.1 250.0 250.0
(104.0 – 135.0˚C)
Exposure Time 00:01:00 – 06:00:00 1 sec. 00:30:00 00:45:00
6 Liquid Exhaust
Chamber Cooldown 0.00 – 5.00 PSIA/min. 0.01 0.75 0.75
7 Air-in
8 Cycle Complete
WS-0163 A
1–18
433HCMC & 533HCMC Steam Sterilizers
Prevacuum Cycles Prevacuum cycles use pressure/vacuum pulses to remove air from the load
during the Conditioning phase. During the drying phase, vacuum in the
chamber reduces the time required to dry the load.
Refer to Figure 1–10 for the adjustable parameters and their maximum,
minimum, and factory settings.
Exposure
Temp
Pressure
Atmospheric
Exposure
Drying
Time
Time
Time
1 2 3 4 5 6 7 8 9
2 Purge
3 Conditioning
7 Drying
Drying Time 00:00:00 – 12:00:00 1 sec. 00:16:00 00:03:00 00:00:00 00:00:00
8 Air-in
9 Cycle Complete
WS-0164 A
DRYING RECOMMENDATIONS
The actual drying time and load size may need to be adjusted based on a
variety of conditions including:
• load composition
• wrapping technique
• ambient conditions
• quality of utilities
61301609701 1–19
General Information
Leak Test Cycle The Vacuum Leak Test cycle parameters are not adjustable.
Atmospheric
Time
1 2 3 4 5 6 7 8 9 10 11
4 Heat-up
10 Air-in
11 Cycle Complete
WS-0165 A
1–20
Section 2 Diagnostic Messages
INTRODUCTION
Failure Conditions Failure conditions are indicated by a flashing Process Failure light and
pulsing alarm tone (exceptions are noted in diagnostic message table).
Failure messages appear in the lower left of the display screen and are
highlighted in red (see Figure 2–1). When a failure occurs, the controls
automatically cancel the cycle.
66
REMAINING TIME
ws-0043
Fault Conditions Fault conditions are less severe than a process failure and are indicated by
a steady alarm tone (exceptions are noted in diagnostic message table).
Fault messages appear in the lower left of the display screen and are
highlighted in red. The sterilizer is usually able to complete the cycle if a
fault occurs.
61301609701 2–1
Diagnostic Messages
grvGRAVITY1
P10 grv GRAVITY1 00:30:14
00:30:14
01
02 CYCLE
HEAT-UPSTART 00:00:05
00:24:15
66
REMAINING TIME
ws-0044
2–2
433HCMC & 533HCMC Steam Sterilizers
Explanation of the failure or fault condi- Level 1 troubleshooting: These procedures should be performed before check-
tion. ing piping or electrical components.
(Type of alarm) Level 2 troubleshooting: These procedures involve checking piping or electrical
components and cleaning, repairing, or replacing them as necessary.
To clear the message and silence the alarm, press the CLEAR/
!
ALARM switch.
61301609701 2–3
Diagnostic Messages
2–4
433HCMC & 533HCMC Steam Sterilizers
ABORT-AIR REMOVAL
ABORT-DOOR SW
ABORT-EXH SLOWED
61301609701 2–5
Diagnostic Messages
ABORT-HEAT-UP
ABORT-HIGH TEMP
ABORT-LIQexh SLOW
2–6
433HCMC & 533HCMC Steam Sterilizers
ABORT-SLOW STEAM
AIR-IN SLOW
BATTERY LOW
61301609701 2–7
Diagnostic Messages
CE NOT SEALED
Control end door did not seal before Check the door gasket and replace if damaged.
door seal timer expired (12 sec.).
Digital input DI 02 CE (RE) Sealed is If the problem persists, proceed as follows:
OFF. • Check chamber pressure switch 4S.
• Check CM1 PC Board 2A2U.
• Clean or replace Door Seal valve 10SV.
• Steam supply or strainer clogged.
CE NOT UNSEALED
2–8
433HCMC & 533HCMC Steam Sterilizers
During standby, the chamber pressure Compare the values on the BRIDGE COMPENSATION screen
exceeded 72.5 PSI (the highest value (Figure 7–24 on page 7–26) with the values on the pressure transducer
the transducer can read) and a data card (Figure 5–2 on page 5–5). Enter the correct values as
PRESS ERROR did not occur. necessary.
61301609701 2–9
Diagnostic Messages
Doors did not seal before door seal Check the door gasket and replace it if damaged.
timer expired (12 sec.).
If the problem persists, proceed as follows:
Digital input DI 02 CE (RE) Sealed is • Check pressure switch 4S.
OFF. • Check CM1 Board 2A2U.
• Check Display Panel 5A.
(steady alarm) • Check Door Seal valve 10SV (standard double door) or 11SV (unidirectional
doors).
2–10
433HCMC & 533HCMC Steam Sterilizers
JACKET OVER-TEMP
61301609701 2–11
Diagnostic Messages
433/533HCMC: The unit failed the vac- Clean or replace the door gasket.
uum leak test.
If the problem persists, proceed as follows:
Possible reasons for failure: • Clean or repair Steam to Chamber valve 1SV.
• Leak rate greater than 0.78 psi/15 • Clean or repair Air-In valve 3SV.
minute (40.34 mm Hg/min). • Clean or repair chamber drain check valve(s) 2CV and 5CV.
• Chamber pressure at end of test • Check Chamber Pressure transducer 1PT.
exceeded 4.75 psi. • Check CM1 Board 2A2U.
(pulsing alarm)
POWER FAILED
PRESS ERROR
2–12
433HCMC & 533HCMC Steam Sterilizers
PRESS IN CHAMBER
(steady alarm)
RESET-TEMP LOW
WATER IN DRAIN
PROBLEM CORRECTIVE ACTION
61301609701 2–13
Diagnostic Messages
INFORMATIONAL MESSAGES
66
REMAINING TIME
CYCLE ADVANCED 2-15
WAIT FOR JACKET
SETUP PARA-
METER
ABORT ENTER TO EDIT 2-15
ws-0044
ABORT MANUAL
Operator aborted the cycle by pressing Message will clear automatically when the abort sequence is completed.
the [ABORT] softkey.
CE DOOR OPEN
Operator pressed START with the con- Message will clear when door is closed.
trol end door was opened.
2–14
433HCMC & 533HCMC Steam Sterilizers
CYCLE ADVANCED
Operator pressed [PHS ADV] softkey to Message will clear when cycle has advanced through all the phases.
manually advance the cycle.
ENTER TO EDIT
Operator selected cycle parameters Message on for 5 seconds after a cycle with low cycle values is saved.
that were below the parameter check
limit. Correct by adjusting the exposure time or temperature.
Settings for selected cycle are below Select a cycle with settings that are equal to or above the default factory
the parameter check limit. settings.
The LOW CYCLE VALUES message Message will clear when a cycle that meets the parameter check limit is
does not prevent the cycle from being selected or when START is pressed.
started.
Printer message. Occasionally, with the introduction of AC power to the sterilizer, the
message "LOW POWER BROWN OUT" may occur on the printout. This
message should be disregarded.
61301609701 2–15
Diagnostic Messages
MANUAL MODE
One or more digital outputs are set to Check that all outputs on the TEST DIGITAL OUT screen are set to
manual mode. AUTO (See “Operating the Digital Outputs Manually” on page 3–7).
RE DOOR OPEN
Operator pressed START with the Message will clear when door is closed.
remote end door was opened.
UTILITY SHUTDOWN
The sterilizer is in Utility Shutdown Message will clear when the Utilities On alarm clock setting is activated.
mode.
Utility Shutdown mode can be bypassed to run one cycle (see “Bypassing Utility
Shutdown Mode” on page 6–4).
The jacket is heating up or cooling Message will clear when the jacket reaches the temperature required to run the
down. cycle.
2–16
Section 3 Trouble Analysis
INTRODUCTION
NOTE
Trouble analysis information for optional equipment is in Section 6, Optional
Equipment & Features.
Descriptions of the system and user flags may be found in the Appendix.
61301609701 3–1
Trouble Analysis
The SERVICE menu includes several diagnostic screens that allow service
technicians to monitor the status of the sterilizer control system:
• ERROR LOG — displays error (diagnostic) messages in chronological
order and provides details about individual error conditions.
• TEST ANALOG IN — displays the current value of the temperature and
pressure inputs.
• TEST DIGITAL IN — displays the current status (on/off) of the switch
inputs.
• TEST DIGITAL OUT — displays the current status (on/off) of the control
outputs (e.g., solenoid valves) and allows the user to turn individual
outputs ON and OFF manually.
• TEST SYSTEM FLAG — displays the current status (on/off) of the
software flags associated with the C language portion (operating system)
of the control software.
• TEST USER FLAG — displays the current status (on/off) of the software
flags associated with the Boolean language portion (application program)
of the control software.
• TEST PRINTER — prints out a test string to verify that the printer is
functioning.
• TEST LED/BUZZER — lights the display panel LEDs and sounds the
alarm buzzer.
NOTE
Refer to Section 7 for detailed instructions on using the SERVICE menu.
3–2
433HCMC & 533HCMC Steam Sterilizers
Viewing the Error Log The error log lists the type of error, date, time, and cycle number during
which it occurred. The error log contains information about the last 20
errors. When the error log is full, the oldest message is erased each time a
new message is stored.
1. Press [SETUP].
SERVICE 2. Select SYSTEM MENU > enter password > SERVICE >
ERROR LOG.
ACCESS REQUIRED The ERROR LOG screen appears.
3. Select the desired error from the list.
ERROR LOG
CHAMBER TEMP
DESCRIPTION CHAMBER PSIA
DATE TIME CYCLE NO.
0
REMAINING TIME
61301609701 3–3
Trouble Analysis
CHAMBER TEMP
ERROR. WATER IN DRAINCHAMBER PSIA
DATE 03/16/03
TIME 37.3 C 22:10 14.68 PSI
CYCLE P10 grv GRAVITY 1
PHASE PH01 STANDBY
CYCLE TIME
CYCLE COUNTER 43 0
00:00:00
REMAINING TIME
CANCEL OK
WS-0074
5. Press [OK].
6. Press [PRINT] to print out the list.
7. Press [OK] to return to the SERVICE menu, or press
[HOME] to return to the process screen.
3–4
433HCMC & 533HCMC Steam Sterilizers
1. Press [SETUP].
SERVICE 2. Select SYSTEM MENU > enter password > SERVICE >
DIAGNOSTIC > ANALOG INPUT.
ACCESS REQUIRED The TEST ANALOG IN screen appears.
0
REMAINING TIME
CANCEL HOME OK
WS-0076A
61301609701 3–5
Trouble Analysis
DIGITAL INPUTS
1. Press [SETUP].
SERVICE 2. Select SYSTEM MENU > enter password > SERVICE >
DIAGNOSTIC > DIGITAL INPUT.
ACCESS REQUIRED The TEST DIGITAL IN screen appears.
TEST DIGITAL IN
INPUT
CHAMBER TEMP CHAMBER PSIA
MODE
DI00 CE Open 0
37.3 C 14.68 PSI
DI01 CE Not Closed 0
DI02 CE (RE) Sealed 0
DI04
DI05
DI08
RE Open
RE Not Closed
H20 Not In Drain
0
REMAINING TIME
0
0
0
CANCEL HOME OK
WS-0077A
The black bar on the right indicates that there are more inputs than can be
NOTE viewed on the screen at one time. Use the up and down arrows to view
inputs not shown.
3–6
433HCMC & 533HCMC Steam Sterilizers
Operating the Digital Most sterilizer malfunctions will result in a specific diagnostic message. The
Outputs Manually diagnostic message tables in Section 2 may recommend testing specific
solenoid valves to isolate the problem. Individual valves may be tested by
using the TEST DIGITAL OUT screen.
1. Press [SETUP].
SERVICE 2. Select SYSTEM MENU > enter password > SERVICE >
DIAGNOSTIC > DIGITAL OUTPUT
ACCESS REQUIRED The TEST DIGITAL OUT screen appears.
0
03 Secondary Drain AUT 0 0
04 Secondary Buzzer AUT 0 0
05 Water to Ejector AUT 0 0
REMAINING TIME
The black bar on the right indicates that there are more inputs than can be
NOTE viewed on the screen at one time. Use the up and down arrows to view
inputs not shown.
61301609701 3–7
DI 00 1 JI CE Open
3–8
D32
D33 DI 01 2 JI CE Not Closed
D34 DI 02 3 JI CE (RE) Sealed
D35 DI 03 4 JI
D36 DI 04 5 JI RE Open
Trouble Analysis
2A2U
DI 08 9 JI H20 Not in Drain
CM1 Board
D56
D57 DI 09 10 JI Controls ON
D58 DI 10 11 JI Adequate Air Press.
3.
D59 DI 11 12 JI
D60 DI 12 13 JI
D55 DI 13 14 JI
Figure 3–6.
D61
DI 14 15 JI
Input LEDs
D62
DI 15 16 JI
D45 DO 06 7 JO Boiler ON
D46 DO 07 8 JO Steam to Jacket
D4 DO 08 9 JO CE, Open
D5 DO 09 10 JO CE, Close
D6 DO 10 11 JO CE (RE) Dr Seal
D7 DO 11 12 JO CE (RE) Unseal
D8 DO 12 13 JO RE, Open
D9 DO 13 14 JO RE, Close
Output LEDs
Compare the value (1 = ON, 0 = OFF) in the ACT. VAL.
D11 DO 15 16 JO
WS-0114 A
433HCMC & 533HCMC Steam Sterilizers
61301609701 3–9
Trouble Analysis
1. Press [SETUP].
SERVICE 2. Select SYSTEM MENU > enter password > SERVICE >
DIAGNOSTIC >SYSTEM FLAG
ACCESS REQUIRED The TEST SYSTEM FLAG screen appears.
CANCEL HOME OK
WS-0080
NOTE
The black bar on the right indicates that there are more flags than can be
viewed on the screen at one time. Use the up and down arrows to view flags
not shown.
NOTE
System flags do NOT provide a real-time indication of the control system
status.
3–10
433HCMC & 533HCMC Steam Sterilizers
USER FLAGS
1. Press [SETUP].
SERVICE 2. Select SYSTEM MENU > enter password > SERVICE >
DIAGNOSTIC > USER FLAG
ACCESS REQUIRED The TEST USER FLAG screen appears.
CANCEL HOME OK
WS-0079
NOTE
The black bar on the right indicates that there are more flags than can be
viewed on the screen at one time. Use the up and down arrows to view flags
not shown.
NOTE User flags provide a real-time indication of the program status. They update
immediately when an event occurs.
61301609701 3–11
Trouble Analysis
Random or erratic action of signals on Power line malfunction. Turn power OFF, wait one minute, turn
the control panel. power ON.
Door does not rise easily when clos- Door cable restricted by object. Remove object.
ing.
Door cable broken. Replace cable.
Door does not latch when fully closed. Object under counterbalance weight. Remove object.
3–12
433HCMC & 533HCMC Steam Sterilizers
2A
DOOR
CONFIGURATION INPUT INPUT
SWITCHES
JACK
2JS Jumper
2JS Jumper
NOTES:
1. Interlock 1JS supplies power for the Steam to Chamber output when
the control end (CE) door is sealed. This interlock works in series with
2JS (power has to be supplied through both interlocks).
2. Interlock 2JS supplies power for the Steam to Chamber output when
the control end (RE) door is sealed. This interlock works in series with
1JS (power has to be supplied through both interlocks).
3. Interlock 3JS supplies power to the CE (RE) Steam to Gasket output
when the door(s) are closed. If the door(s) are not closed, then power
is not supplied and the door gasket(s) cannot be sealed. On standard
double door units, both doors must be closed.
61301609701 3–13
Trouble Analysis
Printer Troubleshooting The printer has a status indicator that flashes when the printer requires
attention.
OFF = NO POWER
ON = READY
Control Panel is ON but printer status No power to printer. Check fuse 4F in Power Box 1A.
indicator is OFF.
Status Indicator is blinking. Paper jam Open the clear cover and paper roll
cover and check for paper jam. Rein-
sert the paper roll and close the paper
roll cover.
Paper roll cover not closed. Close the paper roll cover.
LOW POWER BROWN OUT mes- Message may occur when power is No action required.
sage printed out. applied to the sterilizer.
3–14
Section 4 Calibration and Adjustments
INTRODUCTION
For detailed drawings of the assemblies, refer to the Parts Catalog (Pub. No.
61301609702).
61301609701 4–1
Calibration and Adjustments
CALIBRATION
Overview There are two methods of calibration available through the service menu:
automatic and manual.
The procedures in this section use the automatic method (see page 7–24),
in which low and high reference values for each sensor are entered.
When the manual method (see page 7–25) is used, the gain and offset
values for each sensor are entered directly.
Pressure
Transducer
Control
Display
Box
Chamber Compare
Readings
Pressure
Meter
4–2
433HCMC & 533HCMC Steam Sterilizers
Compare
Readings
Test Fixture
Pressure
Meter
Jacket RTD
Control Display
Box
Chamber
Test Plug
(Precision Resistance)
Drain
Chamber RTD
WS-0089
61301609701 4–3
Calibration and Adjustments
General Check To check the calibration of the unit, following the procedure below.
If the General Check shows that it is required, recalibrate the unit. See the
following procedures:
• “Calibrating the Pressure Transducer (In Situ Method)” on page 4–8.
• “Calibrating the Pressure Transducer (Using Test Fixture)” on page 4–13.
• “Calibrating the Temperature Sensors” on page 4–17.
SPECIAL TOOLS
• pressure Meter, 0–50 psia, accurate to ±0.05 psia.
(must be an absolute pressure meter)
Wear a grounding strap at all times when inside the control box.
Otherwise, damage to the electronic components can result.
OBSERVE PRECAUTIONS
FOR HANDLING
SENSITIVE
DEVICES
SETUP
4–4
433HCMC & 533HCMC Steam Sterilizers
NOTE
Be sure the pressure meter is at the same level or higher than the pressure
meter connection on the unit, so that water does not build up in the tubing to
the pressure meter. Also be sure the tubing does not form a loop below the
pressure meter. If standing water collects in the tubing, extra static pressure
from the water column can result in faulty meter readings.
Pressure Meter
Connection
61301609701 4–5
Calibration and Adjustments
4–6
433HCMC & 533HCMC Steam Sterilizers
61301609701 4–7
Calibration and Adjustments
SETUP
• Wear a grounding strap at all times when inside the control box.
CAUTIONS Otherwise, damage to the electronic components can result.
• During calibration of any circuit, an appropriate input must be inserted on
all temperature and pressure circuits. The absence of an input on any
circuit may result in an improper calibration.
• When calibrating temperature and pressure signals, all sensor inputs
must be within range. Before turning on power, ensure that all sensors
are connected or that a calibration test plug is attached.
NOTES
• Pressure readings must agree to within ±0.2 psi.
• Each calibration may be done either as an independent procedure or as
part of a series (see “Calibrating the Temperature Sensors and Pressure
Transducer” on page 7–24).
PULL A VACUUM
SERVICE 3. Select SYSTEM MENU > enter password > SERVICE >
DIAGNOSTICS > TEST DIGITAL OUT.
ACCESS REQUIRED 4. Turn ON the following outputs:
• Steam to Chamber (DO 00)
• Chamber Drain (DO 01)
• Water to Ejector (DO 06)
5. Purge the chamber for 3 minutes.
6. Turn OFF the Steam to Chamber (DO 00) output.
7. When the pressure is below 2 psia on the pressure meter,
turn OFF the Chamber Drain (DO 01) and Water to Ejector
(DO 06) outputs.
8. Allow the vacuum to stabilize.
4–8
433HCMC & 533HCMC Steam Sterilizers
0
REMAINING TIME
CANCEL HOME OK
WS-0066A
11. Verify Chamber PSIA sensors are set to GAIN = 1.000 and
OFFSET = 0.000.
12. Adjust the GAIN and OFFSET as necessary. (See
“RECORD CALIBRATION DATA” on page 5–15).
61301609701 4–9
Calibration and Adjustments
AUTOMATIC CALIBRATION
CHAMBER TEMP
SENSOR CHAMBER PSIA
CALIBRATE
0
REMAINING TIME
WS-0063A
4–10
433HCMC & 533HCMC Steam Sterilizers
CANCEL HIGH
SAVE
WS-0065
61301609701 4–11
Calibration and Adjustments
4–12
433HCMC & 533HCMC Steam Sterilizers
SETUP
• Wear a grounding strap at all times when inside the control box.
CAUTIONS Otherwise, damage to the electronic components can result.
• During calibration of any circuit, an appropriate input must be inserted on
all temperature and pressure circuits. The absence of an input on any
circuit may result in an improper calibration.
• When calibrating temperature and pressure signals, all sensor inputs
must be within range. Before turning on power, ensure that all sensors
are connected or that a calibration test plug is attached.
NOTES
• Pressure readings must agree to within ±0.2 psi.
• Each calibration may be done either as an independent procedure or as
part of a series (see “Calibrating the Temperature Sensors and Pressure
Transducer” on page 7–24).
0
REMAINING TIME
CANCEL HOME OK
WS-0066A
61301609701 4–13
Calibration and Adjustments
PULL A VACUUM
5. Apply vacuum to the text fixture until the pressure is below
2 psia.
6. Allow the vacuum to stabilize.
7. Wait for the reading on the pressure meter to stabilize.
AUTOMATIC CALIBRATION
CHAMBER TEMP
SENSOR CHAMBER PSIA
CALIBRATE
0
REMAINING TIME
WS-0063A
4–14
433HCMC & 533HCMC Steam Sterilizers
12. Enter the pressure meter reading in the LOW REF: field.
CANCEL HIGH
SAVE
WS-0065
61301609701 4–15
Calibration and Adjustments
4–16
433HCMC & 533HCMC Steam Sterilizers
SETUP
• Wear a grounding strap at all times when inside the control box.
CAUTIONS Otherwise, damage to the electronic components can result.
• During calibration of any circuit, an appropriate input must be inserted on
all temperature and pressure circuits. The absence of an input on any
circuit may result in an improper calibration.
• When calibrating temperature and pressure signals, all sensor inputs
must be within range. Before turning on power, ensure that all sensors
are connected or that a calibration test plug is attached.
0
REMAINING TIME
CANCEL HOME OK
WS-0066A
61301609701 4–17
Calibration and Adjustments
The steam supply can be turned off by using Manual Mode to turn off the
NOTE
Steam to Chamber and Steam to Jacket outputs.
J4
R R W W R R W W
R R W W R G Bu Bk W
4–18
433HCMC & 533HCMC Steam Sterilizers
7. Replace the sensor with the LOW Test Plug (15°C (59°F)).
8. Turn ON the circuit breaker on Power Box 1A.
9. Press [SETUP].
10. Select SYSTEM MENU > enter password > CALIBRATION
> AUTOMATIC.
The AUTOMATIC CALIBRATION screen appears.
11. Set the CALIBRATE field for the 00 Chamber Temp input to
YES.
0
REMAINING TIME
CANCEL HOME
WS-0128A
61301609701 4–19
Calibration and Adjustments
AUTOMATIC CALIBRATION
CHAMBER TEMP
SENSOR GAIN CHAMBER
OFFSET VALUEPSIA
00 Chamber
37.3 C Temp 1.000 0.0 15.0
14.68 PSI
AUTOMATIC CALIBRATION
CHAMBER TEMP
SENSOR GAIN CHAMBER
OFFSET VALUEPSIA
03 Chamber
37.3 C Temp 1.000 0.0 143.0
14.68 PSI
CANCEL HIGH
SAVE
WS-0130
4–20
433HCMC & 533HCMC Steam Sterilizers
61301609701 4–21
Calibration and Adjustments
• The steam supply can be turned off by using Manual Mode to turn off the
NOTES
Steam to Chamber and Steam to Jacket outputs.
• If the steam supply is not turned off, the jacket pressure may increase,
possibly venting steam into the room.
00 Chamber Temp NO
37.3 C 14.68 PSI
01 Jacket Temp YES
03 Chamber PSIA NO
0
REMAINING TIME
CANCEL HOME
WS-0131A
4–22
433HCMC & 533HCMC Steam Sterilizers
AUTOMATIC CALIBRATION
CHAMBER TEMP
SENSOR GAIN CHAMBER
OFFSET VALUEPSIA
AI01 Jacket
37.3 CTemp 1.000 0.0 15.0
14.68 PSI
CANCEL HIGH
SAVE
WS-0133
61301609701 4–23
Calibration and Adjustments
4–24
433HCMC & 533HCMC Steam Sterilizers
61301609701 4–25
Calibration and Adjustments
STERILIZER CONFIGURATION
The door configuration and sterilizer options are selected from the
DIPSWITCHES menu.
0 TIME
WS-0071A
4–26
433HCMC & 533HCMC Steam Sterilizers
61301609701 4–27
Calibration and Adjustments
The pressure switch detects the pressure applied to the chamber door
gasket. When the gasket pressure exceeds 18 - 20 PSIG, two internal
switches are activated (closed) and the chamber door is considered sealed.
The other switch is used by the control system to sense when the door is
sealed/unsealed. When it is closed, it activates DI 02, the control system
considers the door to be sealed, and a cycle can then be started.
Interlock
NO
C Pressure
NC
NO Switch
Door Sealed (DI 02) C 4S
NC
A08215-AE
12SV
Vacuum/Drain
4–28
433HCMC & 533HCMC Steam Sterilizers
VERIFYING OPERATION
Test Lamp
DI 02 Input
(2A3JI)
NC
NO
NC
NO
Ground
Interlock
(2A1JS)
A08215-AF
6. Turn ON the circuit breaker and open the control box cover
to view the digital input and output LEDs.
a. Confirm the test lamp is OFF (interlock open).
b. Confirm DI 02 LED is OFF (door unsealed).
7. Using manual mode (see “Operating the Digital Outputs
Manually” on page 3–7), attempt to pressurize the
chamber by turning ON digital output DO 00 (Steam to
Chamber).
61301609701 4–29
Calibration and Adjustments
The DOOR SEALED LED will illuminate, but there should be no power to
NOTE
1SV.
4–30
433HCMC & 533HCMC Steam Sterilizers
ADJUSTMENT
The pressure switch has two internal switches. Although each switch is
capable of being adjusted at different settings, both are set to activate
(close) at 20 PSIG and de-activate (open) at 18 PSIG (deadband of 2 PSI)
To adjust the setting, remove the switch cover and locate the adjustment
screw that protrudes from the vessel module. Adjusting it clockwise
increases the pressure setpoint and counterclockwise decreases it.
The secondary adjustment screw that is between one of the switches and
NOTE
the transfer pivot arm assures that both switches activate/de-activate at the
same time and should not require adjustment.
Secondary Adjustment
(DO NOT ADJUST)
Adjustment
Screw
A09701-K
61301609701 4–31
Calibration and Adjustments
ABOUT
CHAMBER TEMP CHAMBER PSIA
Getinge PACS 3000
PANEL Version: 3.05
37.3 C 14.68 PSI
PACS Version: 3.08
PROM Update: 9/1/2004
SDN:
Model No.:
0 61301609674
400-500 HCMC
REMAINING TIME
Adjusting the Secondary The secondary alarm is located on the back of Control Box 2A.
Alarm Buzzer The volume level is adjusted by opening or closing the shutter
as shown in Figure 4–26.
Baffle Plate
Alarm
Shutter
Increase Decrease
Volume Volume
99-06C
4–32
433HCMC & 533HCMC Steam Sterilizers
The chamber and jacket bleed valves are closed during the normal
operation of the sterilizer. They are NOT used to adjust the bleed rate of the
chamber or jacket pressure.
The chamber bleed valve should only be open when the chamber pressure
needs to be drained for service. The jacket bleed valve should only be open
when the jacket pressure needs to be drained prior to service. Keep these
valves closed all other times.
The drain cooling water valve should be adjusted to ensure that the sterilizer
discharge at the drain funnel does not exceed 140°F.
Jacket Bleed
Needle Valve
VACUUM TO DOOR(S)
Ball Valve
STEAM TO DOOR(S)
Ball Valve A09701-H
61301609701 4–33
Calibration and Adjustments
Hand-Operated Ball Valves There are two hand-operated ball valves on the sterilizer:
• STEAM TO DOOR(S) valve 1HV
• VACUUM TO DOOR(S) valve 2HV.
The STEAM TO DOOR(S) valve is normally OPEN and is only closed during
emergency opening of the door.
4–34
433HCMC & 533HCMC Steam Sterilizers
DOOR ADJUSTMENTS
The door will travel fully upward and downward with no binding, and with
minimum friction and noise, when properly adjusted.
Remove the lower face panel for access to the door for adjustments.
Door Closed
Switch
Door Guide
Channel
Channel
Bracket
Door Cover
Door Lug
Frame
Weight
Guide
Weight
Rod Locknuts
A09701-C
61301609701 4–35
Calibration and Adjustments
The door latch will not support the weight of the door. Support
WARNING the weight of the door with blocking under it before repair to
avoid injury from a free-falling door.
NOTE Make door adjustments with a hot chamber and door so that adjustments
are made under normal heat expansion conditions.
Door Counterbalance The door counterbalance weight should just touch the frame at the same
Weight Adjustment time that the door latches, and have tension on the door cables. The top of
the door should be even with the top of the machined surface of the
headring.
4–36
433HCMC & 533HCMC Steam Sterilizers
Door Catch Adjustment Check and adjust the door catch as follows:
Engagement
Roller
Adjustment
Screw
Upper Leaf
Forward Slope
Door Catch
Bottom of
Door Plate
A09701-D
61301609701 4–37
Calibration and Adjustments
Door Switch Adjustment The door switch is located above the door opening. The top of the door
engages the switch plunger. Door switch operation is identified by the proper
LED on CM1 Board (2U).
Switch Bracket
Slotted Hole
Door Plate
Door Lug
CL
Door Guide Adjustment The door guides are the cylindrical rods below the door. They control the
position of the door during door travel.
4–38
433HCMC & 533HCMC Steam Sterilizers
Door Cover Adjustment The door cover should be centered in the lower face panel with an alignment
and centering clearance of 1/16” (1.2 mm) on all sides.
1. Loosen the two cover screws and center the lower door
edge in the face panel.
2. Loosen the cover retaining plate screws and move the
cover plate to center the upper edge of the door cover.
Ensure the door lug bolts are secure. If loose, apply Loctite
WARNING #242 to the bolts and torque to 75 ft-lbs to ensure the integrity
of the door, maintain a sealed system, and prevent steam
burns.
61301609701 4–39
Calibration and Adjustments
SPECIAL TOOLS—SPECIFICATIONS
4–40
433HCMC & 533HCMC Steam Sterilizers
Temperature Test Plugs For the calibration procedures, temperature test plugs are used with adapter
plugs (see Calibrating the Temperature Sensors on page 4–17).
8 Pin Male
Adaptor Plug
Connector 9 Pin Male
Test Plug
Connector
Female
Connector
Chamber Temperature Chamber Temperature
Adaptor Plug (8-Pin) Adaptor Plug (9-Pin)
A00299-R
TEST PLUGS
ADAPTER PLUGS
61301609701 4–41
Calibration and Adjustments
Pin 1
Test Plug
Combination
Adaptor Plug
Pin 1 Pin 1
J4
J3
Chamber Jacket
Temperature Temperature
Test Plug Test Plug
A00299-S
4–42
Section 5 Maintenance and Repair
INTRODUCTION
For detailed drawings of the assemblies, refer to the Parts Catalog (Pub. No.
61301609702).
61301609701 5–1
Maintenance and Repair
RECOMMENDED
MAINTENANCE ITEM SEE PAGE
INTERVAL
5–2
433HCMC & 533HCMC Steam Sterilizers
Replacing the Temperature Two temperature sensors and one pressure transducer are standard on this
Sensors (RTDs) equipment:
• Chamber RTD 1RT senses the chamber (drain) temperature.
• Jacket RTD 2RT senses the jacket temperature.
• Pressure transducer 1PT senses chamber pressure.
Be sure to insert the temperature sensors to the correct depth (see Figure
5–1. on page 5-4) so that they will give accurate temperature readings.
• BURN HAZARD: All steam lines will be hot! Shut off the main
WARNINGS
steam supply and depressurize steam in the system before
removing transducer probes. Ensure that all liquids are
drained from the system.
• SHOCK HAZARD: Turn OFF the mains circuit breaker before
replacing the transducer probes.
Wear a grounding strap at all times when inside the control box.
Otherwise, damage to the electronic components can result.
OBSERVE PRECAUTIONS
FOR HANDLING
SENSITIVE
DEVICES
61301609701 5–3
Maintenance and Repair
CHAMBER
DRAIN
5.00 ± .12"
5.00 ± .12"
Jacket RTD
Chamber RTD
JACKET
A09699-AN
CALIBRATION
5–4
433HCMC & 533HCMC Steam Sterilizers
Replacing the Pressure Each pressure transducer comes with a unique resistor set and a matching
Transducer data card.
When replacing the transducer, install the matching resistor set and enter
the data from its data card. Each data card includes the serial number (S–N)
of its matching pressure transducer, and bridge compensation values to
match it to the resistor set if required (see Figure 5–2).
Installation Sequence:
• REPLACE THE RESISTOR SET ALONG WITH THE PRESSURE
TRANSDUCER
• ENTER NEW VALUES FROM THE DATA CARD
• CALIBRATE THE PRESSURE TRANSDUCER
• SAVE RAM TO THE EPROM
ELLER
ckmesstechnik
PE PAA-15/
NGE 0...5 bar
rie-Nr. BJ417
Pressure
Transducer
Data Card
Resistor
Set
WS-0137
61301609701 5–5
Maintenance and Repair
Each pressure transducer comes with a unique resistor set. In the rare
NOTE
event that a resistor set is separated from its pressure transducer, it is
important that the resistor set is rejoined with the same pressure transducer.
EXAMPLE
As an example, assume the resistor set is separated from the first pressure
transducer shown in Figure 5–2 on page 5–5. Proceed as follows:
1. Check that the serial number on the first data card, BJ417, matches the
one on the pressure transducer. (See Figure 5–2 on page 5–5.)
2. In the data card, values from under the heading COMP are R1=931
kOhm, R4=4.02 Ohm.
3. Measure the values of the resistors in positions R1 and R4 of the resistor
set. (See Figure 5–3 on page 5–7 for positions of resistors.) If the
measured values equal the COMP values from the data card, the resistor
set is the correct one for this pressure transducer.
NOTES
• A COMP value listed as 999.9 kOhms means there is no resistor in this
location.
• Other resistors may be in the resistor set. For this test, measure the
values of only those resistors listed on the data card under the heading
COMP.
5–6
433HCMC & 533HCMC Steam Sterilizers
Pressure Transducer
Resistor Set
(R1 R2 R3 R4)
Can have any number of resistors
(R1–R4).
Align the rounded notch with the
notch in the socket.
Must be installed correctly for the
Notch unit to run properly.
Resistor
Set
A00299-F
1. Step A 1st
61301609701 5–7
Maintenance and Repair
BRIDGE COMPENSATION
CHAMBER TEMP CHAMBER PSIA
TRANSDUCER SN J123 TABLE 1
RB GAIN OFFSET
5–8
433HCMC & 533HCMC Steam Sterilizers
Pressure
Transducer
Resistor
Set
P10
0 grv GRAVITY 1 00:00:00
01 S
STANDBY 00:24:15
BRIDGE
R COMPENSATION
TRANSDUCER
RA SN BJ417 TABLE 1
RB GAIN OFFSET
3680 85.40 3.83
3836 85.20 3.96
4010 85.30 3.95
4198 85.70 3.86
4399 86.30 3.59
CANCEL SAVE
WS-0138
61301609701 5–9
Maintenance and Repair
0
REMAINING TIME
WS-0081
5. Press [YES].
• The loading process takes approximately 10 to 15
seconds.
• During the loading process, a communication error
message appears.
6. If the screen displays an error message:
a. Verify that the EPROM is properly seated.
b. If necessary, replace the EPROM (see “Replacing the
Flash EPROM” on page 5–11).
5–10
433HCMC & 533HCMC Steam Sterilizers
INSTALLATION SEQUENCE
1. “RECORD THE CUSTOMER’S SETTINGS” on page 5–12
2. “RECORD PRESSURE TRANSDUCER DATA” on page 5–14
3. “RECORD CALIBRATION DATA” on page 5–15
4. “RECORD THE STERILIZER CONFIGURATION” on page 5–16 (DIP
switch settings)
5. “REMOVE THE OLD EPROM” on page 5–16
6. “CLEAR THE CPU BOARD MEMORY (Cold Start)” on page 5–17
7. “INSTALL THE NEW EPROM” on page 5–18
8. “ENTER TRANSDUCER DATA AND CONFIGURATION SETTINGS”
on page 5–19
9. “ENTER THE CUSTOMER’ SETTINGS” on page 5–20
10. “SAVE RAM TO THE EPROM” on page 5–21
11. “VERIFY STERILIZER OPERATION” on page 5–22
61301609701 5–11
Maintenance and Repair
NOTE
Check if the customer has already written down the settings using the cycle
worksheets included in the 400/500HC Operator Manual.
Printing out all the cycle programs will take more than an hour. It may be
NOTE
more efficient to write down the settings using the cycle worksheets (see
“CYCLE TYPES” on page 1–15).
5–12
433HCMC & 533HCMC Steam Sterilizers
CONFIGURATION Menu
MACHINE NAME NAME
61301609701 5–13
Maintenance and Repair
5–14
433HCMC & 533HCMC Steam Sterilizers
Before starting, be sure the sterilizer door is open. This will ensure that the
NOTE
chamber pressure is at atmospheric.
5. Record the GAIN and OFFSET values for all the sensors:
a. Select SYSTEM MENU > CALIBRATION > MANUAL.
The MANUAL CALIBRATION screen appears.
0
REMAINING TIME
CANCEL HOME OK
WS-0066A
61301609701 5–15
Maintenance and Repair
0 TIME
WS-0071A
Only half of the switch settings are displayed at a time. Be sure to scroll
NOTE
down to view the remaining settings.
5–16
433HCMC & 533HCMC Steam Sterilizers
Cold Start
Socket A Push Button
A B C CPU Board
2A1U
Pin 1
Flash EPROM
WS-0194
The next step removes the existing program from the CPU Board memory.
NOTE
61301609701 5–17
Maintenance and Repair
5–18
433HCMC & 533HCMC Steam Sterilizers
61301609701 5–19
Maintenance and Repair
EDIT CYCLES
CHAMBER TEMP CHAMBER PSIA
1 vac PREVAC1
2 vac37.3
PREVAC1
C 14.68 PSI
3 vac PREVAC1
4 vac PREVAC2
5
6
7
vac PREVAC2
vac PREVAC3 0
vac B-D TEST REMAINING TIME
CANCEL HOME OK
ws-0148
5–20
433HCMC & 533HCMC Steam Sterilizers
61301609701 5–21
Maintenance and Repair
5–22
433HCMC & 533HCMC Steam Sterilizers
Check the Pressure Relief Follow the directions on the tag attached to the pressure relief valve. When
Valve the jacket pressure exceeds 33 PSIG, operate the pressure release handle
of the pressure relief valve to ensure functioning of all moving parts.
AVOID BURNS: (1) Keep away from the outlet (discharge) end
WARNINGS
of the pressure relief valve when operating the pressure
release handle. (2) The handle on the pressure release valve
may be hot. Wear protective gloves when operating the valve.
Clean the Water and Steam Remove the screen from the strainer. Clean it with a stiff brush or suitable
Strainers instrument.
OUT
IN
Screen
Cap
350470JD
Clean the Water Solenoid Disassemble and clean the interior parts of the water solenoid valve.
Valve
61301609701 5–23
Maintenance and Repair
Replace the Air Filter The air filter provides a filtration rate of 99.97% efficiency for 0.3 micron
particles and droplets. Replace the complete air filter after an AIR IN SHOW
diagnostic message. The filter has a threaded NPT fitting and only needs to
be hand-tightened.
A00298QR
5–24
433HCMC & 533HCMC Steam Sterilizers
Opening the Door Manually Manual gasket retraction may be necessary if access to the chamber is
(Manual Gasket Retract) required and UNSEAL DOOR is inoperative.
PREPARATION
1. Turn OFF the steam supply.
2. Turn OFF power to the sterilizer at the power source (circuit
breaker on sterilizer power box).
Water to Ejector
Solenoid Valve
VACUUM TO DOOR(S)
Manual Ball Valve
Operation Rod
STEAM TO DOOR(S)
Ball Valve
A09700-J
61301609701 5–25
Maintenance and Repair
5–26
433HCMC & 533HCMC Steam Sterilizers
Door Gasket Replacement 1. Swing open the front panel. Remove the hex nut and
rubber door stop bumper from the lower frame cross
member.
LOWER
DOOR STOP
BUMPER
DOOR OPEN
SWITCH
(POWER DOOR ONLY)
Do not use sharp instruments or tools that may scratch the headring when
CAUTION removing the gasket.
61301609701 5–27
Maintenance and Repair
A00297-F
5–28
433HCMC & 533HCMC Steam Sterilizers
If using a U-shaped gasket, be sure that the gasket is not twisted and that
CAUTION
the open end is facing into the unit. If using a round gasket, gasket
orientation is not required.
CORRECT ORIENTATION OF
U-SHAPED GASKET
FRONT OF
STERILIZER
A0 0619-A
The door latch will not support the weight of the door. Support
WARNING the weight with blocking under it or secure the door casting to
the sterilizer body bottom engagement lug with the shipping
spacer, washer, lockwasher and 1/4” screw. This will avoid
injury or sterilizer damage from a free-falling door.
61301609701 5–29
Maintenance and Repair
Door Pulley
Door Cover
Screw and Washer
Cable Clamp
Spacer 90921
Cable
Shipping Flat Washer E6974
Hardware
Lockwasher E4671
Turnbuckle
Screw E0108
Weight
5–30
433HCMC & 533HCMC Steam Sterilizers
Replacing the Door Spacers In order to replace the door spacers, the door plate is removed and the
counter-weight is disengaged. The following instructions provide the
necessary steps to safely replace the door spacers.
1. Remove the lower front trim panel from the sterilizer. Lift
and pull on the panel for removal.
2. Remove the hex nut and rubber door stop bumper from the
lower frame cross member
LOWER
DOOR STOP
BUMPER
DOOR OPEN
SWITCH
(POWER DOOR ONLY)
61301609701 5–31
Maintenance and Repair
The door latch will not support the weight of the door. Support
WARNING
the weight of the door with blocking under it or secure the door
casting to the sterilizer body bottom engagement lug with the
shipping spacer, washer, lockwasher and 1/4" screw. This will
avoid injury or sterilizer damage from a free falling door.
Door Pulley
Door Cover
Screw and Washer
Cable Clamp
Spacer 90921
Cable
Shipping Flat Washer E6974
Hardware
Lockwasher E4671
Turnbuckle
Screw E0108
Weight
5–32
433HCMC & 533HCMC Steam Sterilizers
7. Remove the hex nut, lock washer and washer holding the
door track mounting bracket to the frame.
Door Spacer
Door Channel
Frame
A00298qs
61301609701 5–33
Maintenance and Repair
If the door channel is removed or adjusted, check the door plate to gasket
NOTE
ring gap at the four corners of the chamber. Refer to the section entitled
Align the Door in the 400/500 Series Steam Sterilizers Service Manual, Part
No. 61301608209.
5–34
Section 6 Optional Equipment & Features
INTRODUCTION
This section includes service information for the following sterilizer options:
• Utilities Control Feature (see page 6–2)
• Parameter Check Feature (see page 6–5)
Utility control and parameter check are built-in software features that are
enabled or disabled from the DIPSWITCHES menu.
61301609701 6–1
Optional Equipment & Features
The Utilities Control feature turns the steam supply to the jacket ON and
OFF at scheduled times each day. If a cycle is in progress at the alarm time,
utilities shutdown will occur when the cycle ends.
NOTE
Utility S.D. switch DS08 must be set to 1 for the utilities control feature to
operate (see “Enabling the Sterilizer Options” on page 4–27). The Utilities
Control feature does not require any additional hardware; the timer is built
into the control software but must be enabled before it can be used.
Set Utility Control OFF Time and Set Utility Control ON Time and
Days of Week on ALARM CLOCK Screen Days of Week on ALARM CLOCK Screen
00:30:00 121.1 C 00:30:00 00:30:00 121.1 C 00:30:00
Exposure Time Exposure Temp Drying Time Exposure Time Exposure Temp Drying Time
37.3 C
Auto Blow-down
Utilities Off
Utilities On
17:00
19:30
06:00
14.68 PSI
1010100
1111100
0111110
+ 37.3 C
Auto Blow-down
Utilities Off
Utilities On
17:00
19:30
06:00
14.68 PSI
1010100
1111100
0111110
0
REMAINING TIME
0
REMAINING TIME
ws-0024 ws-0024
A08209-C
6–2
433HCMC & 533HCMC Steam Sterilizers
Setting the Utilities Control To set the utilities control ON and OFF times:
ON and OFF Times
1. Press [SETUP].
CALENDAR 2. Select SYSTEM MENU > enter password > TIME
SETTINGS > ALARM CLOCK.
ACCESS REQUIRED • The ALARM CLOCK screen appears.
ALARM CLOCK
CHAMBER TEMP CHAMBER PSIA
(HH:MM) MTWTFSS
37.3 C
Utilities Off 19:30 14.68 PSI
1111100
Utilities On 03:00 1111100
0
REMAINING TIME
CANCEL HOME OK
WS-0023
61301609701 6–3
Optional Equipment & Features
There can be only be one Utilities Off and one Utilities On time each day.
NOTE
Once Utilities Off is activated, Utilities On cannot be activated until the next
day.
Bypassing Utility Shutdown When Utility Shutdown is activated, the control panel displays UTILITY
Mode SHUTDOWN. The operator can bypass this mode in order to run a single
cycle.
6–4
433HCMC & 533HCMC Steam Sterilizers
Each cycle has a minimum parameter check value based on the factory set
parameters (see Table 6–1 on page 6–5). If a user selects cycle settings
that are below this value, a red LOW CYCLE VALUES and a blue ENTER
TO EDIT message appear.
The parameter check value is derived from a theoretical value based on the
factory set exposure time and temperature. For example, the parameter
check value for a Prevac 1 cycle is 73.4; that is, a 3 minute exposure at
275°F is equivalent to 73.4 minutes (73:24) at 250°F.
NOTES
• Parameter Check switch DS11 must be set to 1 for the parameter check
feature to operate (see “Enabling the Sterilizer Options” on page 4–27).
• The LOW CYCLE VALUES condition does not prevent the cycle from
starting. If the operator chooses to run the cycle without adjusting the
cycle parameters, LOW CYCLE VALUES will be printed. The message
will disappear while in cycle. After the cycle, when the controls return to
standby mode, the message will reappear.
PREVAC1 73.4
PREVAC3
GRAVITY1 29.9
GRAVITY2 245.1
FLASH 3+ 73.4
LIQUIDS1 N/A
61301609701 6–5
Optional Equipment & Features
Estimating Valid Cycle The graphs in this section provide a visual means of determining whether a
Parameters specific cycle setting is above the parameter check limit.
Note that the y-axis (exposure time) is a logarithmic scale. The divisions on
NOTE
the y-axis are not evenly spaced. For example, the number 15 is not exactly
between 10 and 20, but closer to 20 than to 10.
100
Exposure Time (minutes)
20 min. @ 255˚F
(above limit)
10 min. @ 260˚F
10 (above limit)
10 min. @ 255˚F
(below limit)
GRAVITY 1
(Parameter Check
1 Limit = 29.9)
230 235 240 245 250 255 260 265 270 275
A09133-D2
20
Exposure Temperature (˚F)
19
18
17
16
15
14
13
12
11
10
DETAIL OF
LOGARITHMIC SCALE
6–6
433HCMC & 533HCMC Steam Sterilizers
100
Exposure Time (minutes)
10
GRAVITY 1
(Parameter Check
1 Limit = 29.9)
230 235 240 245 250 255 260 265 270 275
Exposure Temperature (˚F) A09133-E1
61301609701 6–7
Optional Equipment & Features
100
Exposure Time (minutes)
10 GRAVITY 2
(Parameter Check
Limit = 245.1)
1
230 235 240 245 250 255 260 265 270 275
Exposure Temperature (˚F) A09133-E2
100
Exposure Time (minutes)
10
FLASH 3+
(Parameter Check
Limit = 73.4)
1
230 235 240 245 250 255 260 265 270 275
Exposure Temperature (˚F) A09701-G
6–8
433HCMC & 533HCMC Steam Sterilizers
100
Exposure Time (minutes)
FLASH 10+
10 (Parameter Check
Limit = 245.1)
1
230 235 240 245 250 255 260 265 270 275
Exposure Temperature (˚F) A09701-F
100
Exposure Time (minutes)
10
PREVAC
(Parameter Check
Limit = 73.4)
1
230 235 240 245 250 255 260 265 270 275
Exposure Temperature (˚F) A09133-H1
61301609701 6–9
Optional Equipment & Features
100
Exposure Time (minutes)
10
B-D TEST
(Parameter Check
Limit = 49.0)
230 235 240 245 250 255 260 265 270 275
Exposure Temperature (˚F) A09133-H2
6–10
433HCMC & 533HCMC Steam Sterilizers
WATER-SAVER
Both the MP-129A (designed for the M/C 3500 series) and MP-129F
(designed for the 400/500 series) water-savers are compatible with the
433HCMC and 533HCMC control system.
• If the water-saver is an MP-129A, the Water to Ejector output (DO 05) is
routed to the water-saver and the water-saver is operated by a pressure
switch located in the sterilizer piping.
• If the water-saver is an MP-129F, the controls monitor the chamber
pressure and operate the Water to Ejector (DO 05) and Water-Saver ON
(DO 14) outputs directly.
Refer to the User Manual supplied with the water-saver for maintenance
information.
61301609701 6–11
Optional Equipment & Features
6–12
Section 7 Service Menu
INTRODUCTION
The display panel consists of the LCD screen, five function buttons
(softkeys), directional arrows, and ENTER button.
LCD Screen
Softkeys
!
WS-0341
61301609701 7–1
Service Menu
What You Can Do From the Tasks that can be performed from the display panel, along with the menu
Display Panel path are listed in Table 7–1.
NOTE Most tasks require a password with a specific access area enabled. See
“Password Access” on page 1–29 for a list of access areas.
Setting the Clock and Calendar [SETUP] SYSTEM > TIME SETTINGS > CALENDAR 7-20
Setting the Alarm Clock SYSTEM > TIME SETTINGS > ALARM CLOCK
Calibrating the Temperature Sensors [SETUP] SYSTEM > CALIBRATION > AUTOMATIC 7-22
and Pressure Transducer (Automatic)
Calibrating the Temperature Sensors SYSTEM > CALIBRATION > MANUAL 7-23
and Pressure Transducer (Manual)
Entering the Bridge Compensation SYSTEM > CALIBRATION > BRIDGE TABLES 7-24
Table Data
Printing a Program Listing [SETUP] SYSTEM > DOCUMENTATION > PRINT CYCLE 7-26
SETUP
Setting Up the Printer Log Interval SYSTEM > DOCUMENTATION > PRINTER 7-27
SettingUp the Printer Log Values Defi- SYSTEM > DOCUMENTATION > PRINT LOG VAL- 7-28
nition UES
7–2
433HCMC & 533HCMC Steam Sterilizers
Changing the Cycle Name and Num- [SETUP] SYSTEM > CONFIGURATION > EDIT CYCLES 7-31
ber
Assigning the Sterilizer Name SYSTEM > CONFIGURATION > NAME 7-32
Select the Language [SETUP] SYSTEM > CONFIGURATION > LANGUAGE DATE 7-32
UNIT
Select the Date Format 7-34
SettingUp the Display Panel [SETUP] SYSTEM > CONFIGURATION > PANEL SETUP 7-36
Configuring Password Access for Indi- [SETUP] SYSTEM > PASSWORD 7-39
vidual Users
Viewing the Error Log [SETUP] SYSTEM > SERVICE > ERROR LOG 7-43
View the Analog Inputs [SETUP] SYSTEM > SERVICE > DIAGNOSTIC > ANALOG 7-45
INPUT
View the Digital Inputs SYSTEM > SERVICE > DIAGNOSTIC > DIGITAL 7-46
INPUT
Operating the Digital Ouptuts SYSTEM > SERVICE > DIAGNOSTIC > DIGITAL 7-47
OUTPUT
View the User Flags SYSTEM > SERVICE > DIAGNOSTIC > USER 7-48
FLAGS
View the System Flags SYSTEM > SERVICE > DIAGNOSTIC > SYSTEM 7-49
FLAGS
Test the Printer SYSTEM > SERVICE > DIAGNOSTIC > PRINTER 7-50
Test the LEDs SYSTEM > SERVICE > DIAGNOSTIC > LED/ 7-50
BUZZER
Save RAM to the EPROM [SETUP] SYSTEM > SAVE RAM TO FLASH 7-51
61301609701 7–3
Service Menu
NAVIGATION BASICS
2. Press ENTER.
CANCEL
WS-0039
NOTE If all the items in the list do not fit on one screen, a black bar will appear to
the right of the list. Use the down arrow to view the additional items.
7–4
433HCMC & 533HCMC Steam Sterilizers
Entering a Password (or When a password is required (or an alphanumeric value needs to be
Alphanumeric Value) entered), a screen similar to the password entry screen appears.
1 2 3 4 5 6 7 8 9 0
A B C D E F G H I J
K L M N O P Q R S T
U V W X Y Z Å Ä Ö Ü
! " # % & ( ) \ , ;
. : + SPACE - / =
CANCEL DELETE OK
WS-0017
2. Press ENTER.
The password character is indicated by a pound sign (#).
3. Repeat steps 1 and 2 for additional characters.
NOTE If you make a mistake, press [DELETE] to erase the last character.
61301609701 7–5
Service Menu
Entering a Data Value When a data value (e.g., cycle setting, calibration value, etc.) needs to be
entered, a popup menu appears.
CANCEL
WS-0018 HC
1. Press ENTER.
The last digit of the value will blink.
2. Use the UP and DOWN arrows to increase or decrease the
digit.
3. Press ENTER.
4. Press [SAVE].
The CONFIRM SAVE? YES OR NO prompt appears.
NOTE If desired, the CONFIRM SAVE? prompt can be turned off by changing the
CONFIRM SAVE/CANCEL setting on the PANEL SETUP screen (see page
7–36.)
5. Press [YES].
6. Press [OK].
7–6
433HCMC & 533HCMC Steam Sterilizers
Selecting a Setup Option Some fields have multiple choice setup options (e.g., temperature and
pressure units setup).
0
REMAINING TIME
CANCEL HOME OK
WS-0019
2. Press ENTER.
4. Press ENTER.
5. Press [SAVE].
The [CONFIRM SAVE? YES OR NO] prompt appears.
6. Press [YES].
7. Repeat steps 1 thru 6 for other setup options.
8. Press [OK] when finished.
61301609701 7–7
Service Menu
Exiting a Display Screen There are three commands that are used to exit a display screen:
[CANCEL], [HOME], and [OK].
CANCEL HOME OK
WS-0158
7–8
433HCMC & 533HCMC Steam Sterilizers
SETUP SOFTKEY
When the SETUP softkey is pressed, the SETUP MENU appears. The
SETUP MENU (see Figure 7–6) has five items:
• DETAILS — see page 7–10.
• PLOT GRAPH — see page 7–10.
• BAR GRAPH — see page 7–11.
• SYSTEM MENU — see page 7–12.
• ABOUT — see page 7–13.
CANCEL
WS-0039
61301609701 7–9
Service Menu
Selecting the Process Screen The user has the option of selecting one of three process screens: the
normal process screen (DETAILS), a screen with a graphical display (PLOT
GRAPH) or a screen with an enlarged timer (BAR GRAPH). A password is
not required to select the process screen.
DETAILS
1. Press [SETUP].
2. Select DETAILS.
3. Press ENTER.
The Details screen (Figure 7–7) appears.
PLOT GRAPH
1. Press [SETUP].
2. Select PLOT GRAPH.
3. Press ENTER.
The Plot Graph screen appears (Figure 7–8).
7–10
433HCMC & 533HCMC Steam Sterilizers
Time
01:40
BAR GRAPH
1. Press [SETUP]
2. Select BAR GRAPH.
3. Press ENTER.
The Bar Graph screen (Figure 7–9) appears.
0
REMAINING TIME
61301609701 7–11
Service Menu
Accessing the System Menu The SYSTEM MENU lists the service features of the control software. A
password (with specific area access) is required to use these features.
1. Press [SETUP]
2. Select SYSTEM MENU.
3. Press ENTER.
The ENTER PASSWORD screen appears.
4. Enter your password (see page 7–5).
The SYSTEM MENU (Figure 7–10) appears.
0
PASSWORDS
ABOUT
DIP SWITCHES
REMAINING TIME
CANCEL HOME
WS-0032A
7–12
433HCMC & 533HCMC Steam Sterilizers
Viewing the About Screen The ABOUT screen lists the software version and other pertinent
information about the sterilizer.
1. Press [SETUP].
2. Select ABOUT.
3. Press ENTER.
The ABOUT screen (Figure 7–11) appears.
ABOUT
CHAMBER TEMP CHAMBER PSIA
Getinge PACS 3000
PANEL Version: 3.05
37.3 C 14.68 PSI
PACS Version: 3.08
PROM Update: 9/1/2004
SDN:
Model No.:
0 61301609674
400-500 HCMC
REMAINING TIME
NOTE The ABOUT screen is also used to adjust the brightness of the display (see
page 4–31).
61301609701 7–13
Service Menu
SELECT CYCLE
CHAMBER TEMP CHAMBER PSIA
P08 grv GRAVITY1
P09 grv37.3
GRAVITY1
C 14.68 PSI
P10 grv GRAVITY1
P11 grv GRAVITY2
P12
P13
P14
grv GRAVITY2
grv GRAVITY2 0
f 3 FLASH 3+ REMAINING TIME
CANCEL OK
ws-0016 400_500HC
7–14
433HCMC & 533HCMC Steam Sterilizers
PARAMETER SOFTKEY
The parameters display is configured for either Quick Edit or Full Edit.
• Quick Edit allows the user to change the exposure time, exposure tem-
perature, and drying time without using the EDIT PARAMETERS
screen.
• Full Edit provides access to all adjustable parameters.
To set up the display for Quick Edit or Full Edit, see “Setting Up the Display
NOTE
Panel” on page 7–36.
0
REMAINING TIME
CANCEL MORE OK
PARAMS
WS-0033 733HC
61301609701 7–15
Service Menu
CANCEL
WS-0018 HC
7–16
433HCMC & 533HCMC Steam Sterilizers
0
REMAINING TIME
CANCEL EDIT OK
WS-0037 733HC
3. Press [EDIT].
The EDIT PARAMETERS screen (Figure 7–16) appears.
0
REMAINING TIME
CANCEL EDIT OK
WS-0015A 733HC
61301609701 7–17
Service Menu
7–18
433HCMC & 533HCMC Steam Sterilizers
SYSTEM MENU
0
PASSWORDS
ABOUT
DIP SWITCHES
REMAINING TIME
CANCEL HOME
WS-0032A
61301609701 7–19
Service Menu
0.00 PSI
DETAILS
CHAMBER TEMP CHAMBER PSIA
TIME SETTINGS C
PLOT CALENDAR
CALIBRATION
ALARM CLOCKC
BAR G DOCUMENTATION
37.3 C 14.68 PSI
I
SYSTEMCONFIGURATION
0.05 PSI
ABOUTPASSWORDS
0
0.00 C
DIP SWITCHES
134.3 C
REMAINING TIME
CANCEL HOME OK
WS-0062
7–20
433HCMC & 533HCMC Steam Sterilizers
CALENDAR
CHAMBER TEMP CHAMBER PSIA
TIME 17:18:27 HH:MM:SS
0
REMAINING TIME
CANCEL HOME OK
ws-0021
To change the time and date formats, see “Selecting the Language, Date
NOTE
Format, and Units” on page 7–34.
61301609701 7–21
Service Menu
Setting the Alarm Clock The alarm clock function is used to control the Utility Control option. Alarm
clock times can be enabled on a day-to-day basis.
1. Press [SETUP].
CALENDAR 2. Select SYSTEM MENU > enter password > TIME
SETTINGS > ALARM CLOCK.
ACCESS REQUIRED The ALARM CLOCK screen (Figure 7–35) appears.
ALARM CLOCK
CHAMBER TEMP CHAMBER PSIA
(HH:MM) MTWTFSS
37.3 C
Utilities Off 19:30 14.68 PSI
1111100
Utilities On 03:00 1111100
0
REMAINING TIME
CANCEL HOME OK
WS-0023
7–22
433HCMC & 533HCMC Steam Sterilizers
CALIBRATION MENU
0.00 PSI
DETAILS
CHAMBER TEMP CHAMBER PSIA
TIME SETTINGS C
PLOT MANUAL
CALIBRATION
AUTOMATIC C
BAR G DOCUMENTATION
37.3 C 14.68 PSI
0.05 PSI
BRIDGE TABLES
SYSTEMCONFIGURATION
0.05 PSI
ABOUTPASSWORDS
0
0.00 C
DIP SWITCHES
134.3 C
REMAINING TIME
CANCEL HOME OK
WS-0062
When automatic calibration is selected, the user enters high and low
reference values for the sensor and the software recalculates the gain and
offset value. This is the usual method of calibration.
When manual calibration is selected, the user directly enters the gain and
offset values. This method is useful when reprogramming the values after
replacing the flash EPROM.
61301609701 7–23
Service Menu
1. Press [SETUP].
CALIBRATION 2. Select SYSTEM MENU > enter password > CALIBRATION
> AUTOMATIC.
ACCESS REQUIRED The AUTOMATIC CALIBRATION (Figure 7–20) screen
appears.
AUTOMATIC CALIBRATION
CHAMBER TEMP
SENSOR CHAMBER PSIA
CALIBRATE
0
REMAINING TIME
WS-0063A
7–24
433HCMC & 533HCMC Steam Sterilizers
MANUAL CALIBRATION
1. Press [SETUP].
CALIBRATION 2. Select SYSTEM MENU > enter password > CALIBRATION
> MANUAL.
ACCESS REQUIRED The MANUAL CALIBRATION screen (Figure 7–21)
appears.
0
REMAINING TIME
CANCEL HOME OK
WS-0066A
61301609701 7–25
Service Menu
BRIDGE COMPENSATION
CHAMBER TEMP CHAMBER PSIA
TRANSDUCER SN J123 TABLE 1
RB GAIN OFFSET
7–26
433HCMC & 533HCMC Steam Sterilizers
DOCUMENTATION MENU
0.00 PSI
DETAILS
CHAMBER TEMP CHAMBER PSIA
CALENDAR C
PRINT CYCLE SETUP
PLOT CALIBRATION
PRINTER C
BAR G DOCUMENTATION
37.3 C 14.68 PSI
PRINTER 0.05 PSI
LOG VALUES
SYSTEMCONFIGURATION
0.05 PSI
ABOUTPASSWORDS
DIP SWITCHES
0
0.00 C
134.3 C
REMAINING TIME
CANCEL HOME OK
WS-0069
61301609701 7–27
Service Menu
1. Press [SETUP].
DOCUMENTATION 2. Select SYSTEM MENU > enter password >
DOCUMENTATION > PRINT CYCLE SETUP.
ACCESS REQUIRED The PRINTOUT CYCLE screen (Figure 7–24) appears.
PRINTOUT CYCLE
CHAMBER TEMP CHAMBER PSIA
P08 grv GRAVITY1
P09 grv GRAVITY1
37.3 C 14.68 PSI
P10 grv GRAVITY1
P11 grv GRAVITY1
P12
P13
P14
grv GRAVITY1
liq LIQUIDS1
liq LIQUIDS1
0
REMAINING TIME
7–28
433HCMC & 533HCMC Steam Sterilizers
Setting Up the Printer Log The printer records data at two rates: a normal rate (fast log interval) during
Interval the exposure phase and a slower rate (slow log interval) during the other
cycle phases.
1. Press [SETUP].
DOCUMENTATION 2. Select SYSTEM MENU > enter password >
DOCUMENTATION > PRINTER.
ACCESS REQUIRED The PRINTER screen (Figure 7–25) appears.
PRINTER
CHAMBER TEMP CHAMBER PSIA
MODE 1
SLOW LOG INTERVAL 01:00
37.3 C 14.68 PSI
FAST LOG INTERVAL 00:30
0
REMAINING TIME
CANCEL HOME OK
WS-0029
61301609701 7–29
Service Menu
SettingUp the Printer Log The printout can be set up to display up to four values. The values are
Values Definition identified by name and value type/item (e.g., AI01 CHAMBER TEMP is
analog input (AI) number 1, chamber temperature).
1 Chamber Temp
2 37.3
Jacket C
Temp 14.68 PSI
4 Chamber PSIA
0
REMAINING TIME
7–30
433HCMC & 533HCMC Steam Sterilizers
AI ANALOG INPUT
AO 37.3 C OUTPUT
ANALOG 14.68 PSI
TP TP CONVERTER
FC FUNCTION CALCULATION
CP COMPARATOR
0
REMAINING TIME
CANCEL
WS-0027
5. Press ENTER.
The log value item list (Figure 7–28) appears.
6. Select the input source from the list (e.g., Chamber RTD).
0
REMAINING TIME
CANCEL
WS-0026 833HCMC
7. Press [SAVE].
The CONFIGURE PRINTER LOG VALUES screen (Figure
7–26) appears again.
8. Repeat steps 2 thru 8 for other column(s).
9. Press [OK] when finished.
61301609701 7–31
Service Menu
CONFIGURATION MENU
0
ABOU EXTENDED PARAMETERS
DIPSWITCHES
REMAINING TIME
WS-0031 633HC
7–32
433HCMC & 533HCMC Steam Sterilizers
Changing the Cycle Name To change the cycle number (P number) and name:
and Number
1. Press [SETUP].
CONFIGURATION 2. Select SYSTEM MENU > enter password >
CONFIGURATION > EDIT CYCLES.
ACCESS REQUIRED The EDIT CYCLES screen (Figure 7–31) appears.
EDIT CYCLES
CHAMBER TEMP CHAMBER PSIA
1 vac PREVAC1
2 vac37.3
PREVAC1
C 14.68 PSI
3 vac PREVAC1
4 vac PREVAC2
5
6
7
vac PREVAC2
vac PREVAC3 0
vac B-D TEST REMAINING TIME
CANCEL HOME OK
ws-0148
NOTE If two or more numbers are equal when SAVE is pressed, TWO OR MORE
SEQ. NUMBERS EQUAL will be displayed. Before pressing OK, be sure to
change one of the duplicate numbers.
61301609701 7–33
Service Menu
Cycle names can be up to eight characters long. The three letter prefix (e.g.,
NOTE
grv) indicates the cycle type and cannot be changed.
grv GRAVITY 1
MACHINE NAME
CHAMBER TEMP CHAMBER PSIA
NAME STERILIZER No. 1
37.3 C 14.68 PSI
0
REMAINING TIME
CANCEL HOME OK
WS-0030
7–34
433HCMC & 533HCMC Steam Sterilizers
1 2 3 4 5 6 7 8 9 0
A B C D E F G H I J
K L M N O P Q R S T
U V W X Y Z Å Ä Ö Ü
! " # % & ( ) \ , ;
. : + SPACE - / =
61301609701 7–35
Service Menu
Selecting the Language, Date To select the language, date format, and pressure and
Format, and Units temperature units:
1. Press [SETUP].
CONFIGURATION 2. Select SYSTEM MENU > enter password >
CONFIGURATION > LANGUAGE DATE UNITS.
ACCESS REQUIRED The LANGUAGE DATE UNITS screen (Figure 7–33)
appears.
0
REMAINING TIME
CANCEL HOME OK
WS-0019
LANGUAGE
DATE FORMAT
3. Select the DATE FORMAT field.
4. Press ENTER.
5. Use the UP or DOWN arrow to select from the available
date formats: MM/DD/YYYY, DD/MM/YYYY or
YYYY/MM/DD.
6. Press [SAVE].
7–36
433HCMC & 533HCMC Steam Sterilizers
PRESSURE UNIT
7. Select the PRESSURE UNITS field.
8. Press ENTER.
9. Use the UP or DOWN arrow to select PSI, KPA, or BARS.
10. Press [SAVE].
TEMPERATURE UNIT
11. Select the TEMP UNITS field.
12. Press ENTER.
13. Use the UP or DOWN arrow to select F or C.
14. Press [SAVE].
15. Press [OK].
61301609701 7–37
Service Menu
1. Press [SETUP].
CONFIGURATION 2. Select SYSTEM MENU > enter password >
CONFIGURATION > PANEL SETUP.
ACCESS REQUIRED The PANEL SETUP screen (Figure 7–34) appears.
PANEL SETUP
CHAMBER TEMP CHAMBER PSIA
DEFAULT MENU: DETAILS
PARAMETER
37.3EDIT:
C QUICK 14.68 PSI
SCREEN SAVER: 99 (MIN)
CONFIRM SAVE/CANCEL: YES
0
REMAINING TIME
CANCEL HOME OK
WS-0025
7–38
433HCMC & 533HCMC Steam Sterilizers
SCREEN SAVER
11. Select the SCREEN SAVER field.
12. Press ENTER.
13. Use the arrows to select the screen saver delay value (in
minutes).
14. Press [SAVE].
CONFIRM SAVE/CANCEL
15. Select the CONFIRM SAVE/CANCEL field.
16. Press ENTER.
17. Use the UP or DOWN arrow to select YES or NO.
18. Press [SAVE].
19. Press [OK] when finished.
61301609701 7–39
Service Menu
PASSWORDS MENU
Configuring Password Access The PASSWORDS menu is used to assign passwords and enable levels of
for Individual Users access. See Table 1–7 on page 1–29 for a list of access areas and
privileges.
1. Press [SETUP].
PASSWORD SETUP 2. Select SYSTEM MENU > enter password > PASSWORDS.
The PASSWORDS (Figure 7–36) screen appears.
ACCESS REQUIRED
Figure 7–37. PASSWORDS SCREEN
0
REMAINING TIME
3. Select the user name from the list. To add a new user, see
“ADD USER” on page 7–40.
4. Select the PASSWORD field for the desired user.
5. Press ENTER.
6. Assign the password (see “Entering a Password (or
Alphanumeric Value)” on page 7–5).
7. Press [OK].
NOTE Press ACCESS AREAS to view the access area letter assignments.
7–40
433HCMC & 533HCMC Steam Sterilizers
ADD USER
1. Press [ADD/DELETE].
2. Press [ADD USER]
3. Assign the user name (see “Entering a Password (or
Alphanumeric Value)” on page 7–5).
4. Select the PASSWORD field.
5. Press ENTER.
6. Assign the password (see “Entering a Password (or
Alphanumeric Value)” on page 7–5).
7. Select the ACCESS AREA field.
8. Select the desired access area options one at a time. (See
step 8 of previous procedure).
9. Press [HOME] when finished (or [CANCEL] to return to the
SYSTEM MENU).
DELETE USER
1. Press [ADD/DELETE].
2. Press [DELETE USER]
3. Select the user name from the list.
4. Press ENTER.
5. Press [SAVE]
6. Press [HOME] when finished (or [CANCEL] to return to the
SYSTEM MENU).
61301609701 7–41
Service Menu
DIPSWITCHES MENU
1. Press [SETUP].
DIP SWITCHES 2. Select SYSTEM MENU > enter password > DIP
SWITCHES.
ACCESS REQUIRED The DIP SWITCHES screen (Figure 7–37) appears.
0 TIME
WS-0071A
7–42
433HCMC & 533HCMC Steam Sterilizers
SERVICE MENU
0.00 PSI
CHAMBER TEMP
DETAILS CHAMBER PSIA
CONFIGURATION
ERROR LOG C
PLOT DOCUMENTATYION
DIAGNOSTIC C
BAR G PASSWORDS
37.3 C 14.68 PSI
0.05 PSI
PACS ADDRESSES
SYSTEMDIP
M SWITCHES
0.05 PSI
SERVICE
ABOUTENU
0.00 C
SAVE RAM TO FLASH
0
134.3 C
REMAINING TIME
CANCEL HOME
WS-0072
61301609701 7–43
Service Menu
1. Press [SETUP].
SERVICE 2. Select SYSTEM MENU > enter password > SERVICE >
ERROR LOG.
ACCESS REQUIRED The ERROR LOG screen (Figure 7–39) appears.
ERROR LOG
CHAMBER TEMP
DESCRIPTION CHAMBER PSIA
DATE TIME CYCLE NO.
0
REMAINING TIME
7–44
433HCMC & 533HCMC Steam Sterilizers
Viewing the Analog and The DIAGNOSTICS menu (Figure 7–40) has eight options:
Digital Inputs • ANALOG INPUT — see page 7–45
• ANALOG OUTPUT (not used)
• DIGITAL INPUT — see page 7–46
• DIGITAL OUTPUT — see page 7–47
• SYSTEM FLAG — see page 7–49
• USER FLAG — see page 7–48
• PRINTER — see page 7–48
• LED/BUZZER — see page 7–50
0.00 PSI
CHAMBER TEMP
DETAILS CHAMBER PSIA
CONFIGURATION C INPUTOG
ERROR LANALOG
PLOT SYSTEM SETUPION ANALOG INPUT
DIAGNOSTICA C
37.3 C
BAR G PASSWORDS ANALOG 14.68 PSI
0.05 PSI OUTPUT
PACS ADDRESSES
SYSTEMDIP SWITCHES DIGITAL
0.05 PSIINPUT
ABOUTSERVICE
SYSTEM
134.3 0
0.00 C OUTPUT
DIGITAL
SAVE RAM TO FLASH
C FLAG
REMAINING TIME
CANCEL HOME
WS-0075
NOTE This sterilizer does not use analog outputs (all control valves are on/off
rather than the proportional [throttling] type).
61301609701 7–45
Service Menu
ANALOG INPUTS
1. Press [SETUP].
SERVICE 2. Select SYSTEM MENU > enter password > SERVICE >
DIAGNOSTIC > ANALOG INPUT.
ACCESS REQUIRED The TEST ANALOG IN screen (Figure 7–41) appears.
0
REMAINING TIME
CANCEL HOME OK
WS-0076A
7–46
433HCMC & 533HCMC Steam Sterilizers
DIGITAL INPUTS
1. Press [SETUP].
SERVICE 2. Select SYSTEM MENU > enter password > SERVICE >
DIAGNOSTIC > DIGITAL INPUT.
ACCESS REQUIRED The TEST DIGITAL IN screen (Figure 7–42) appears.
TEST DIGITAL IN
INPUT
CHAMBER TEMP CHAMBER PSIA
MODE
DI00 CE Open 0
37.3 C 14.68 PSI
DI01 CE Not Closed 0
DI02 CE (RE) Sealed 0
DI04
DI05
DI08
RE Open
RE Not Closed
H20 Not In Drain
0
REMAINING TIME
0
0
0
CANCEL HOME OK
WS-0077A
61301609701 7–47
Service Menu
0
03 Secondary Drain AUT 0 0
04 Secondary Buzzer AUT 0 0
05 Water to Ejector AUT 0 0
REMAINING TIME
NOTE
If any outputs are left in the manual mode, a MANUAL MODE message is
displayed. All outputs must be in AUTO mode before a cycle can be run.
7–48
433HCMC & 533HCMC Steam Sterilizers
1. Press [SETUP].
SERVICE 2. Select SYSTEM MENU > enter password > SERVICE >
DIAGNOSTIC > USER FLAG.
ACCESS REQUIRED The TEST USER FLAG screen (Figure 7–44) appears.
CANCEL HOME OK
WS-0079
61301609701 7–49
Service Menu
SYSTEM FLAGS
The system flags indicate the status of the C language portion of the control
software.
1. Press [SETUP].
SERVICE 2. Select SYSTEM MENU > enter password > SERVICE >
DIAGNOSTIC > SYSTEM FLAG.
ACCESS REQUIRED The TEST SYSTEM FLAG screen (Figure 7–45) appears.
CANCEL HOME OK
WS-0080
7–50
433HCMC & 533HCMC Steam Sterilizers
1. Press [SETUP].
SERVICE 2. Select SYSTEM MENU > enter password > SERVICE >
DIAGNOSTIC > PRINTER.
ACCESS REQUIRED The printer outputs a test string.
LEDs
1. Press [SETUP].
SERVICE 2. Select SYSTEM MENU > enter password > SERVICE >
DIAGNOSTIC > LED/BUZZER.
ACCESS REQUIRED All control panel LEDs light and the alarm sounds for five
seconds.
61301609701 7–51
Service Menu
1. Press [SETUP].
SERVICE 2. Select SYSTEM MENU > enter password > SAVE RAM TO
FLASH.
ACCESS REQUIRED Figure 7–47. SAVE RAM TO FLASH SCREEN
0
REMAINING TIME
WS-0081
3. Press [YES].
4. Press ENTER.
7–52
Figure 7–48. MENU TREE (sheet 1 of 4)
0
REMAINING TIME
Setup Menu
00:30:00 121.1 C 00:30:00 00:30:00 121.1 C 00:30:00 00:30:00 121.1 C 00:30:00 00:30:00 121.1 C 00:30:00 00:30:00 121.1 C 00:30:00
Exposure Time Exposure Temp Drying Time Exposure Time Exposure Temp Drying Time Exposure Time Exposure Temp Exposure Time Exposure Temp Drying Time
Exposure Time Exposure Temp Drying Time Drying Time
P10 grv GRAVITY 1 01:36:36 P10 grv GRAVITY 1 01:36:36 P10 grv GRAVITY1 00:00:00 P10 grv GRAVITY1 00:00:00 P10 grv GRAVITY1 00:00:00
02 CYCL COMPLETE 00:04:34 02 CYCL COMPLETE 00:03:42 01 STANDBY 00:24:15 01 STANDBY 00:24:15 01 STANDBY 00:24:15
ChamberTemp 81.6 C
Chamber Temp 61.2 C ABOUT
Chamber PSIA 14.45 PSI CHAMBER TEMP CHAMBER PSIA CHAMBER TEMP CHAMBER PSIA CHAMBER TEMP CHAMBER PSIA
Chamb Press/PSIG 0.00 PSI DETAILS
TIME SETTINGS Getinge PACS 3000
Jacket Temp 122.3 C PLOT GRAPH
CALIBRATION PANEL Version: 3.05
37.3 C 14.68 PSI 37.3 CTEMP
CHAMBER 14.68 PSI 37.3 C 14.68 PSI
Atmosphere PSIA 14.37 PSI DOCUMENTATION
BAR GRAPH PACS Version: 3.08
SYSTEMCONFIGURATION
MENU
Chamber PSIA 14.35 PSI PROM Update: 2/19/2004
0 0 0
PASSWORDS
Exp. Temp Max 121.7 C ABOUT SDN: 6130160xxxx
DIP SWITCHES
Exp. Temp Min 121.4 C Model No.: 3.00 400-500 HCMC
Time REMAINING TIME REMAINING TIME REMAINING TIME
01:40
SETUP SELECT PARA- UNSEAL SETUP PARA- UNSEAL SETUP SELECT PARA- UNSEAL CANCEL HOME BRIGHT BRIGHT OK
CANCEL HOME
CYCLE METER METER CYCLE METER DEC INC
WS-0012 WS-0013 800
WS-0014 WS-0035 400-500 HCMC
WS-0032A
Details Screen Plot Graph Screen Bar Graph Screen About Screen
System Menu
61301609701 7–53
Figure 7–49. MENU TREE (sheet 2 of 4)
0
PASSWORDS
ABOUT
DIP SWITCHES
REMAINING TIME
CANCEL HOME
WS-0032A
System Menu
00:30:00 121.1 C 00:30:00 00:30:00 121.1 C 00:30:00 00:30:00 121.1 C 00:30:00
Exposure Time Exposure Temp Drying Time Exposure Time Exposure Temp Drying Time Exposure Time Exposure Temp Drying Time
P10 grv GRAVITY1 00:00:00 P10 grv GRAVITY1 00:00:00 P10 grv GRAVITY1 00:00:00
01 STANDBY 00:24:15 01 STANDBY 00:24:15 01 STANDBY 00:24:15
0
REMAINING TIME
DS13 T-Doc
0 0
REMAINING TIME
0
REMAINING TIME
CANCEL HOME PRINT ACCESS ADD/ CANCEL HOME PRINT OK CANCEL HOME NO YES
AREA DELETE
WS-0052 WS-0071 400-500 HCMC WS-0081
0 0
0.00 C PASSWORDS SERVICE
0 0
0.00 C ABOUTENU
0
DIP SWITCHES DIP SWITCHES ABOU 0.00 C
134.3 C DIPSWITCHES SAVE RAM TO FLASH
134.3 C 134.3 C
REMAINING TIME REMAINING TIME REMAINING TIME REMAINING TIME REMAINING TIME
CANCEL HOME OK CANCEL HOME CANCEL HOME OK CANCEL MORE CANCEL HOME
WS-0062 WS-0063 833HCMC
WS-0069 WS-0031
WS-0072
Time Settings Menu Calibration Menu Documentation Menu Configuration Menu Service Menu
See Sheet 3 See Sheet 4
WS-0198B 400-500 HCMC
7–54
Figure 7–49. MENU TREE (sheet 3 of 4)
00:30:00 121.1 C 00:30:00 00:30:00 121.1 C 00:30:00 00:30:00 121.1 C 00:30:00 00:30:00 121.1 C 00:30:00 00:30:00 121.1 C 00:30:00
Exposure Time Exposure Temp Drying Time Exposure Time Exposure Temp Drying Time Exposure Time Exposure Temp Drying Time Exposure Time Exposure Temp Drying Time
Exposure Time Exposure Temp Drying Time
P10 grv GRAVITY1 00:00:00 P10 grv GRAVITY1 00:00:00 P10 grv GRAVITY1 00:00:00 P10 grv GRAVITY1 00:00:00 P10 grv GRAVITY1 00:00:00
01 STANDBY 00:24:15 01 STANDBY 00:24:15 01 STANDBY 00:24:15 01 STANDBY 00:24:15 01 STANDBY 00:24:15
0
REMAINING TIME
0
REMAINING TIME
P12
P13
P14
grv GRAVITY1
liq LIQUIDS1
liq LIQUIDS1
0
REMAINING TIME
0
REMAINING TIME
0
REMAINING TIME
CANCEL HOME CANCEL HOME OK CANCEL HOME PRINT PRINT OK CANCEL HOME OK CANCEL HOME NEXT DELETE OK
OK
ALL LIST
WS-0023 WS-0070 WS-0029 WS-0028
ws-0021
Calendar Screen Alarm Clock Screen Printout Cycle Screen Printer Screen Configure Printer Log Values Screen
0 0 0
vac PREVAC1
0 0 0
3233 79.20 4.31
WS-0019 WS-0025
WS-0148 800HC WS-0030
Automatic Calibration Screen Manual Calibration Screen Bridge Compensation Screen Edit Cycles Screen Machine Name Screen Language Date Units Screen Panel Setup Screen
WS-0199 833HCMC
61301609701 7–55
D32 DI 00 1 JI CE Open
D33 DI 01 2 JI CE Not Closed
D34 DI 02 3 JI CE (RE) Sealed
D35 DI 03 4 JI
D36 DI 04 5 JI RE Open
D31 DI 05 6 JI RE Not Closed
D37 DI 06 7 JI
D38
DI 07 8 JI
2A2U
D56
D57 DI 09 10 JI Controls ON
D58 DI 10 11 JI Adequate Air Press.
D59 DI 11 12 JI
7–56
D60 DI 12 13 JI
D55 DI 13 14 JI
D61
DI 14 15 JI
Input LEDs
D62
DI 15 16 JI
D4 DO 08 9 JO CE, Open
D5 DO 09 10 JO CE, Close
D6 DO 10 11 JO CE (RE) Dr Seal
D7 DO 11 12 JO CE (RE) Unseal
D8 DO 12 13 JO RE, Open
D9 DO 13 14 JO RE, Close
Output LEDs
0.00 PSI
CHAMBER TEMP
DETAILS CHAMBER PSIA
CONFIGURATION
ERROR LOG C
PLOT DOCUMENTATYION
DIAGNOSTIC C
BAR G PASSWORDS
37.3 C 14.68 PSI
0.05 PSI
PACS ADDRESSES
SYSTEMDIP
M SWITCHES
0.05 PSI
SERVICE
ABOUTENU
0.00 C
SAVE RAM TO FLASH
134.3 C 0
REMAINING TIME
CANCEL HOME
WS-0072
Service Menu
0
ABOUT
REMAINING TIME
DIGITAL
0.00
SAVE RAM TO FLASH
SYSTEM
134.3 0
C OUTPUT
C FLAG
REMAINING TIME
Diagnostic Menu
00:30:00 121.1 C 00:30:00 00:30:00 121.1 C 00:30:00 00:30:00 121.1 C 00:30:00 00:30:00 121.1 C 00:30:00 00:30:00 121.1 C 00:30:00 00:30:00 121.1 C 00:30:00 00:30:00 121.1 C 00:30:00
Exposure Time Exposure Temp Drying Time Exposure Time Exposure Temp Drying Time Exposure Time Exposure Temp Exposure Time Exposure Temp Drying Time
Drying Time Exposure Time Exposure Temp Drying Time Exposure Time Exposure Temp Drying Time Exposure Time Exposure Temp Drying Time
P10 grv GRAVITY1 00:00:00 P10 grv GRAVITY1 00:00:00 P10 grv GRAVITY1 00:00:00 P10 grv GRAVITY1 00:00:00 P10 grv GRAVITY1 00:00:00 P10 grv GRAVITY1 00:00:00 P10 grv GRAVITY1 00:00:00
01 STANDBY 00:24:15 01 STANDBY 00:24:15 01 STANDBY 00:24:15 01 STANDBY 00:24:15 01 STANDBY 00:24:15 01 STANDBY 00:24:15 01 STANDBY 00:24:15
TEST ANALOG IN TEST DIGITAL IN TEST DIGITAL OUT TEST USER FLAG PRINTOUT CYCLE TEST PRINTER TEST LEDs AND BUZZER
INPUT CHAMBER TEMP CHAMBER
ACT. VAL PSIA INPUT
CHAMBER TEMP CHAMBER PSIA
MODE CHAMBER TEMP
OUTPUT MODE CHAMBER
SET PSIA
ACT. VAL FLAG CHAMBER TEMP CHAMBER PSIA
MODE CHAMBER TEMP CHAMBER PSIA CHAMBER
STRING TO BE TEMP
PRINTED CHAMBER PSIA 5 SECCHAMBER
LED ANDTEMP
BUZZER TEST CHAMBER PSIA
P08 grv GRAVITY1 <ASCII TABLE>
UF000 Process Failure 1
AI00 Chamber Temp 88.0 C DI00 CE Open 0 00 Steam to Chamber AUT 0 0 grv GRAVITY1
37.3 C 14.68 PSI 37.3 C 14.68 PSI 37.3 C 14.68 PSI 37.3 C 14.68 PSI P09 37.3 C 14.68 PSI 37.3 C
UF001 Start OK 0 14.68 PSI 37.3 C 14.68 PSI
AI01 Jacket Temp 87.5 C DI01 CE Closed 0 01 Chamber Drain AUT 0 0 grv GRAVITY1
P10
UF002 Door Unseal OK 0
AI03 Chamber PSIA 11.85 PSI DI02 CE (RE) Sealed 0 02 Chamber Air-In AUT 0 0 grv GRAVITY1
P11
UF004 In Progress 0
0
REMAINING TIME
DI04
DI05
DI09
RE Open
RE Closed
Controls ON
0
REMAINING TIME
0
0
0
03
05
07
Drain Cooling Water
Water to Ejector
Steam to Jacket
0
AUT
AUT
AUT
REMAINING TIME
0
0
0
0
0
0
UF005
UF006
UF007
Cycle Good
Always True Flag
0 1
Uni RE-EnabledREMAINING TIME0
0
P12
P13
P14
grv GRAVITY1
liq LIQUIDS1
liq LIQUIDS1
0
REMAINING TIME
0
REMAINING TIME
0
REMAINING TIME
CANCEL HOME OK CANCEL HOME OK CANCEL AUTO/ 1/0 PULSE ACTI- CANCEL HOME OK CANCEL HOME PRINT PRINT OK CANCEL HOME OK
TEST CANCEL HOME TEST OK
MAN VATE ALL
WS-0076 833 HCMC WS-0078 833HCMC WS-0079
WS-0077 833HCMC WS-0202
WS-0070 WS-0203
Test Analog In Screen Test Digital In Screen Test Digital Out Screen Test User Flag Screen Test System Flag Screen Test Printer Screen Test LEDs & Buzzer Screen
WS-0200 400-500 HCMC
61301609701 7–57
61301609701 7–58
Section 8 Piping Schematics
61301609701 8–1
Figure 8–1. PIPING COMPONENTS, 433HCMC & 533HCMC
6SV JACKET
CHAMBER CHAMBER BLEED
JACKET PRESSURE GAUGE PRESSURE GAUGE
This illustration is intended to show PRESSURE GAUGE (CONTROL END) (REMOTE END)
NOTE component locations only. Refer to (REMOTE END) 1PT
RE DOOR 3ST
Parts Catalog 61301609702 for an GASKET
illustrated parts breakdown of this RETRACT 1TR
JACKET
assembly PRESSURE GAUGE CHAMBER C 5SV
(CONTROL END) 6CV
CE DOOR H
GASKET
G to 14SV E
RETRACT to 3CV
PRESSURE METER
CONNECTION
D
8SV
JACKET A
Drain piping is
C C rotated 180˚ on
JACKET
PRESSURE GAUGE
CE DOOR A 16x16 DD Units
2P 6SV GASKET RETRACT G
(CONTROL END)
1SV 5S
Air-In
Filter 8SV 5SV DRAIN
JACKET
BLEED
1ST H Ejector
3SV Vintage 2000 Units
14SV VACUUM TO DOOR(S)
STEAM 3CV HAND VALVE
SUPPLY 2ST
CHAMBER G
PRESSURE GAUGE
1RV (CONTROL END) 2FCV P
2BV E
B A 3ST
4CV F
CHAMBER 1TR
DRAIN D
WATER
SUPPLY
2RT D 2SV H
5CV
1RT 1BV B 6CV
E F
1PT
2
I
3FCV 2
A 1FCV I
12SV
10SV
A STEAM TO DOOR(S)
HAND VALVE
Drain piping is
rotated 180˚ on
16x16 DD Units
8–2
Figure 8–2. INSTRUMENTATION DIAGRAM, 433HCMC & 533HCMC
61301609701 8–3
Figure 8–3. PROCESS DIAGRAM, 433HCMC & 533HCMC (sheet 1 of 4)
Getinge Sourcing, LLC Doc. Name: Process Diagram, 400/500HC Doc. No.: PP0006 Rev.: C
1777 East Henrietta Road Created By: SEB 07/27/01 File Name: PP0006 Rev C Page 1-2.xls Product: 400/500HC Latest ECO: 121644
Rochester, NY 14623-3133 Approved: SEB 07/27/01 File Type: MS Excel 97 Class Code: 8104 Page: 1 of 5
PH1200
PH1200
PH1200
PH1200
PH0000
PH0500
PH1100
PH1800
PH2000
PH3100
PH3110
PH3800
PH4000
PH0000
PH0500
PH1100
PH1800
PH2000
PH3100
PH3110
PH3800
PH4000
Cycle Phase No.: Cycle Phase No.:
PH3190
PH1200
PH1200
PH1200
PH0000
PH0500
PH1100
PH1890
PH1990
PH2090
PH3800
PH4000
PH0000
PH0500
PH1100
PH1800
PH2000
PH3100
PH3110
PH3700
PH3710
PH3800
PH4000
Cycle Phase No.: Cycle Phase No.:
This document contains proprietary information of Getinge Sourcing, LLC. It shall not be reproduced in whole or in part without the authorization of Getinge. Any authorized reproduction must bear this entire legend.
8–4
Figure 8–3. PROCESS DIAGRAM, 433HCMC & 533HCMC (sheet 2 of 4)
Getinge Sourcing, LLC Doc. Name: Process Diagram, 400/500HC Doc. No.: PP0006 Rev.: C
1777 East Henrietta Road
Rochester, NY 14623-3133 Page: 2 of 5
PH1200
PH3100
PH1200
PH1200
PH1200
PH3100
PH0000
PH0500
PH1100
PH1800
PH2000
PH3110
PH3800
PH4000
PH0000
PH0500
PH1100
PH1800
PH2000
PH3110
PH3800
PH4000
Cycle Phase No.: Cycle Phase No.:
NOTES: 1) X* - For Vac. Dry: DO14 is ON, DO01 & DO05 are OFF.
For Grav. Dry: DO14 is OFF, DO01 & DO05 are ON.
2) Refer to pg 3 for charts, keys & other notes.
Pressure T(exp) 3) Refer to pg 4-5 for cycle phase descriptions.
lkt LEAK TEST Cycle P(+pul)
with water saver
Atm (Atm-5.0)
PH1200
PH1200
PH3100
PH0000
PH0500
PH1100
PH1800
PH2000
PH3110
PH3700
PH3710
PH3800
PH4000
Cycle Phase No.:
For units with WATER SAVER PKG ONLY
This document contains proprietary information of Getinge Sourcing, LLC. It shall not be reproduced in whole or in part without the authorization of Getinge. Any authorized reproduction must bear this entire legend.
61301609701 8–5
Figure 8–3. PROCESS DIAGRAM, 433HCMC & 533HCMC (sheet 3 of 4)
Getinge Sourcing, LLC Doc. Name: Process Diagram, 400/500HC Doc. No.: PP0006 Rev.: C
1777 East Henrietta Road
Rochester, NY 14623-3133 Page: 3 of 5
Cycle Exposure Drying 1) Refer to Instrumentation Diagram PI0012 for piping schematics
T(exp) Time P(dry) Time and hardware references.
No. Type Type °F (°C) (min.) Type (PSIA) (mm:ss)
* P1 vac PREVAC 1 275.0 (135.0) 3 vac 4.75 16:00 2) Pressure vs. Time cycle phase diagrams are not to scale
* P2 vac PREVAC 1 275.0 (135.0) 3 vac 4.75 16:00
* P3 vac PREVAC 1 275.0 (135.0) 3 vac 4.75 16:00 3) Drain Cooling (5SV) is activated/controlled by 2S at all times while unit is ON.
* P4 vac PREVAC 2 275.0 (135.0) 3 vac 4.75 03:00
* P5 vac PREVAC 2 275.0 (135.0) 3 vac 4.75 03:00 4) Factory Setting for Liquid Dwell Time = 8 minutes
* P6 vac PREVAC 3 275.0 (135.0) 3 vac 4.75 00:00
* P7 vac B-D TEST 273.0 (133.9) 3.5 vac 4.75 00:00 5) 1SV, 2SV & 14SV in some phases are controlled with a fixed limit ramp rate of
P8 grv GRAVITY 1 250.0 (121.1) 30 grv atm 30:00 72.5 PSI/min (see Factory Ramp Rate Settings Table). Normally during these
P9 grv GRAVITY 1 250.0 (121.1) 30 grv atm 30:00 phases, the SV's energize fully and are indicated as such in the phase diagrams
P10 grv GRAVITY 1 250.0 (121.1) 30 grv atm 30:00 and the Cycle Phase Descriptions.
P11 grv GRAVITY 2 275.0 (135.0) 10 grv atm 30:00
P12 grv GRAVITY 2 275.0 (135.0) 10 grv atm 30:00 6) Additional outputs applicable to units with steam generators:
P13 grv GRAVITY 2 275.0 (135.0) 10 grv atm 30:00 a) "Boiler ON" (DO06) turns ON the steam generator while controls
P14 f3 FLASH 3+ 275.0 (135.0) 3 grv atm 00:30 are ON (except during auto blowdown or utility shut-off condition).
P15 f3 FLASH 3+ 275.0 (135.0) 3 grv atm 00:30 b) "Boiler Blowdown" (DO115) turns ON during an auto blowdown event.
P16 f3 FLASH 3+ 275.0 (135.0) 3 grv atm
P17 f3 FLASH 3+ 275.0 (135.0) 3 grv atm 00:30
P18 f10 FLASH 10+ 275.0 (135.0) 10 grv atm 00:30
P19 f10 FLASH 10+ 275.0 (135.0) 10 grv atm 00:30
P20 liq LIQUIDS 1 250.0 (121.1) 30 n/a n/a n/a
P21 liq LIQUIDS 2 250.0 (121.1) 45 n/a n/a n/a
* P22 lkt LEAKTEST 268.0 (131.1) 3 vac 4.75 n/a LEGEND for Cycle Phase Diagrams:
De-energized
Cycle Phase Ramp Rate S Energized per selection (e.g. Gravity Dry vs. Vac Dry)
No. Description SV No. (PSI/Min.) Comments
PH1200 Conditioning (Pos) 1SV 72.5 non-adjustable D Energized per door status
PH1200 Conditioning (Neg) 2SV 72.5 non-adjustable
PH1800 Heat Up 1SV, 14SV varying non-adjustable SV Solenoid Valve
PH1890 Liquid Heat Up 1SV, 14SV varying non-adjustable
PH3100 Exhaust 2SV 72.5 non-adjustable DO Digital Output (Software reference)
PH3190 Liquid Exhaust 14SV 0.75 Supervisor adjustable
This document contains proprietary information of Getinge Sourcing, LLC. It shall not be reproduced in whole or in part without the authorization of Getinge. Any authorized reproduction must bear this entire legend.
8–6
Figure 8–3. PROCESS DIAGRAM, 433HCMC & 533HCMC (sheet 4 of 4)
Doc. Name: Process Diagram, 400/500HC Doc. No.: PP0006 Rev.: C Doc. Name: Process Diagram, 400/500HC Doc. No.: PP0006 Rev.: C
Created By: SEB 7/27/01 Product: 400/500HC Class Code: 8104 Created By: SEB 7/27/01 Product: 400/500HC Class Code: 8104
Approved By: SEB 7/27/01 Doc. File Name: PP0006 Rev C Page 4-5 Latest ECO: 121644 Approved By: SEB 7/27/01 Doc. File Name: PP0006 Rev C Page 4-5 Latest ECO: 121644
Approved By: Doc. File Type: MS Word 97 Page: 4 of 5 Approved By: Doc. File Type: MS Word 97 Page: 5 of 5
_____________________________________________________________________________________________________ _____________________________________________________________________________________________________
PH3800 – AIR-IN: 3SV & 14SV energize and, for Liquid cycles, 8SV remains de-energized. When chamber pressure is within 2 PSI of
Cycle Phase Descriptions: atmosphere, 3SV & 14SV de-energizes and the phase is complete.
PH0000 – STANDBY: Jacket temperature control is enabled. Jacket RTD maintains temperature by energizing and de-energizing 8SV. 8SV PH4000 – CYCLE COMPLETE: 2SV & 3SV energize and, for Liquid cycles, 8SV remains de-energizes. The cycle complete buzzer sounds for 1
maintains steam to jacket throughout the cycle unless noted in the phase description. Also throughout the cycle, drain temperature switch 2S minute or until clear button is pressed. The door(s) automatically unsealed. When either door is opened, the buzzer will turn OFF (if not already
maintains the drain temperature below 60°C (140°F) by energizing and de-energizing 5SV (water to drain). OFF) and the phase is complete. Note: 2SV & 3SV de-energizes when door is unsealed (sensed by DI02).
PH0500 – CYCLE START: Cycle Start button is pressed, door(s) seals per PH110/PH130 described below. Once jacket temperature is within SEALING DOOR(S): PH110/130 – CE SEALING/DOORS SEALING: 10SV energizes and 12SV de-energizes. Gasket groove is
+4.4 / -2.2°C (+8.0 / -4°F) of T(exp), the phase is complete. pressurized, sealing the chamber.
PH01100 – PURGE: 1SV & 2SV energize. Steam enters the chamber through 1SV and purges chamber of air through 2SV. 6SV energizes if UNSEALING DOOR(S): PH111/131 – CE UNSEALING/DOORS UNSEALING: 10SV remains energized. 12SV & 6SV energize. After 4
chamber pressure exceeds 2 PSI above atmosphere and de-energizes if pressure falls 1 PSI below atmosphere. Cooling water from 6SV seconds, 10SV de-energizes. Water to the ejector creates a vacuum in the gasket groove retracting the door gasket.
condenses steam at ejector and flows to drain. After 90 seconds (for 422/433HC models) or 90 seconds (for 522/533HC models), 1SV, 2SV & After 30 seconds, 6SV de-energizes and the phase is complete.
6SV de-energize and the phase is complete.
PH1200 – CONDITIONING: Note: This phase may be repeated based on the quantity of pulses programmed.
O
x Steam pressurizes chamber through energized 1SV. When chamber pressure reaches P(+pul) (T(exp) – 2.2°C (or 4.0 F) converted to steam
pressure, but no greater than 32.0 PSIA), 1SV de-energizes and 2SV & 6SV energize. Chamber steam is exhausted to ejector, condensed
and drained.
x For Gravity or Flash 10+ Cycles: When chamber pressure reaches 0.5 PSI above atmosphere, 2SV & 6SV de-energizes and the phase is
complete.
st
x For PreVac, B-D TEST or LEAK TEST Cycles – 1 Pulse only: When chamber pressure reaches 9.75 PSIA, 2SV & 6SV de-energizes and
the phase is complete.
nd rd
x For PreVac, B-D TEST or LEAK TEST Cycles – 2 & 3 Pulses only: When chamber pressure reaches 4.75 PSIA, 2SV & 6SV de-
energizes and the phase is complete.
x For PreVac , B-D TEST or LEAK TEST w/ water saver: When chamber pressure reaches 5.0 PSI below atmosphere, 2SV & 6SV de-
energizes. 15SV energizes until 4.75 psia is reached and the phase is complete.
PH1800 – HEAT-UP: 1SV & 14SV energize and steam pressurizes the chamber. As the chamber temperature approaches the programmed
exposure temperature, 1SV & 14SV are controlled (via chamber pressure transducer) to slow down the ascent. When the programmed exposure
temperature +0.4°C (0.7°F) is reached (sensed by the chamber RTD), the phase is then complete.
PH1890 – LIQUID HEAT-UP: 1SV & 14SV energize and steam pressurizes the chamber. As the chamber temperature approaches the
programmed exposure temperature, 1SV & 14SV are controlled (via chamber pressure transducer) to slow down the ascent. When the
programmed exposure temperature is reached (sensed by the chamber RTD), the phase is then complete.
PH1990 – LIQUID DWELL: 14SV is energized & 1SV is controlled to maintain the selected exposure temperature. After a programmed time
period, the phase is complete.
PH2000 – EXPOSURE: The programmed exposure timing begins. 1SV & 14SV are controlled by the chamber RTD to maintain the selected
exposure temperature. After the programmed exposure time expires, 1SV & 14SV de-energizes and the phase is complete.
PH2090 – LIQUID EXPOSURE: The programmed exposure timing begins. 1SV & 14SV are controlled by the chamber RTD to maintain the
selected exposure temperature. After the programmed exposure time expires, 1SV & 14SV de-energizes and the phase is complete.
PH3100 – EXHAUST: 2SV, 6SV, & 14SV energize. Chamber steam is exhausted to ejector, condensed and drained.
x For Gravity Drying: When chamber pressure reaches atmosphere, 2SV, 6SV, & 14SV de-energizes and the phase is complete.
x For Vacuum Drying (including B-D Test & Leak Test Cycles): When chamber pressure reaches 4.75 PSIA, 2SV, 6SV, & 14SV de-
energizes and the phase is complete.
x For Vacuum Drying w/ water saver (including B-D Test & Leak Test Cycles): When chamber pressure reaches 5.0 PSI below
atmosphere, 2SV & 6SV de-energize. 15SV energizes until 4.75 psia is reached and the phase is complete.
PH3190 – LIQUID EXHAUST: 8SV de-energizes. Chamber pressure is decreased at a programmed ramp rate by energizing and de-energizing
14SV. When chamber pressure reaches 6 PSI above atmosphere, 6SV energizes. When the chamber temp reaches 100.0qC / 212.0qF (HCMC:
chamber pressure must also be <= atm pressure), the chamber pressure is decreased at half the programmed ramp rate. When chamber
temperature (sensed by the chamber RTD) cools to <= Atm – 3.5 psia, the phase is complete.
PH3110 – DRYING: The programmed dry timing begins. For unit with ejector only, 2SV & 6SV energize. For unit with a water saver, 14SV and
15SV energize. If Gravity Drying was selected, 3SV also energizes. When the time period elapses, 2SV, 3SV, 6SV, 14SV & 15SV de-energize
and the phase is complete. Note: For Leak Test cycle, the time in this phase is fixed at 15 minutes.
PH3700 – EQUALIZATION: 14SV is energized. All other SV’s, except for door and jacket control, are de-energized. After a dwell period of 5
minutes, the phase is complete.
PH3710 – LEAK TEST: 14SV is energized. All other SV’s, except for door and jacket control, are de-energized. Chamber pressure is initially
checked. After 15 minutes, chamber pressure is checked again to determine leak rate and the phase is complete.
___________________________________________________________________________________________________________________ ___________________________________________________________________________________________________________________
Getinge Sourcing, LLC. 1777 East Henrietta Road, Rochester NY 14623-3133 Getinge Sourcing, LLC. 1777 East Henrietta Road, Rochester NY 14623-3133
This document contains proprietary information of Getinge Sourcing, LLC. and is subject to being returned upon request. It shall not be This document contains proprietary information of Getinge Sourcing, LLC. and is subject to being returned upon request. It shall not be
reproduced in whole or in part without the authorization of Getinge. Any authorized reproduction must bear this entire legend. reproduced in whole or in part without the authorization of Getinge. Any authorized reproduction must bear this entire legend.
61301609701 8–7
Figure 8–4. WATER SAVER PACKAGE, MP-129A
61301609701 8–8
Figure 8–5. PIPING DIAGRAM, MP-129F WATER-SAVER
61301609701 8–9
61301609701 8–10
Section 9 Electrical Schematics
61301609701 9–1
Figure 9–1. WIRING DIAGRAM, SYSTEM, 400/500 HCMC
9–2
Figure 9–2. WIRING DIAGRAM, POWER BOX 1A
61301609701 9–3
Figure 9–3. WIRING DIAGRAM, CONTROL BOX 2A
9–4
Figure 9–4. WIRING DIAGRAM, OPTIONAL EQUIPMENT
1A3TB3 (BK)
1A3TB4 (W)
1A3TB5 (G)
1A3TB3 (BK)
1A3TB4 (W)
1A3TB5 (G)
17A1TB1 (BK)
17A1TB2 (W)
17AGND (G)
61301609701 9–5
Figure 9–5. ELECTRICAL SCHEMATIC, 433HCMC & 533HCMC
9–6
Figure 9–6. ELECTRICAL SCHEMATIC, POWER BOX 1A
61301609701 9–7
Figure 9–7. ELECTRICAL SCHEMATIC, CONTROL OUTPUT (sheet 1 of 2)
9–8
Figure 9–7. ELECTRICAL SCHEMATIC, CONTROL OUTPUT (sheet 2 of 2)
61301609701 9–9
Figure 9–8. WIRING DIAGRAM, MP-129A WATER-SAVER
61301609701 9–10
Figure 9–9. WIRING DIAGRAM, MP-129F WATER SAVER
61301609701 9–11
61301609701 9–12
Section 10 Roughing-In Drawings
DRAWING
FIGURE TITLE PAGE
NUMBER
61301609701 10–1
Roughing-In Drawings
10–2
433HCMC & 533HCMC Steam Sterilizers
61301609701 10–3
Roughing-In Drawings
10–4
433HCMC & 533HCMC Steam Sterilizers
61301609701 10–5
Roughing-In Drawings
10–6
433HCMC & 533HCMC Steam Sterilizers
61301609701 10–7
Roughing-In Drawings
10–8
433HCMC & 533HCMC Steam Sterilizers
61301609701 10–9
Roughing-In Drawings
10–10
433HCMC & 533HCMC Steam Sterilizers
61301609701 10–11
Roughing-In Drawings
10–12
433HCMC & 533HCMC Steam Sterilizers
61301609701 10–13
Roughing-In Drawings
10–14
433HCMC & 533HCMC Steam Sterilizers
61301609701 10–15
Roughing-In Drawings
10–16
433HCMC & 533HCMC Steam Sterilizers
61301609701 10–17
Roughing-In Drawings
10–18
433HCMC & 533HCMC Steam Sterilizers
61301609701 10–19
Roughing-In Drawings
10–20
Appendix A System and User Flags
SYSTEM FLAGS
SF000 INT I2C ERROR This is an overall I/O error flag indicating communication error with digital
input modules, digital output modules, analog input modules, and analog out-
put modules over the I2C bus.
SF006 PRINT COM. ERROR Communication error with printer connected to PACS 3000.
SF011 OP21 MASTER First panel is master. SF11 – SF15 are used for handling the master/slave
concept for operator panels and are controlled by the Boolean program. One
of the panels is defined as the master panel, all others are slaves. The master
panel is accepted to start a program and to change the program parameters.
SF016 PIT EXCEEDED This flag is set if the power supply has been interrupted longer than the Power
Interrupt Time set in the system configuration.
61301609701 A–1
System and User Flags
SF020 BLANK TEXT Resets all information and error text if set. The system resets this flag.
SF021 CYCLE EXECUTED This flag should be set for every executed sterilizer cycle in order to increment
the service interval cycle counter. The system resets this flag.
SF022 START LOG This flag is used for starting and stopping the PACS 3000 and panels logging.
It also starts and stops the graphics in process pictures.
SF023 SLOW LOG Activates the long logging interval set in the printer definition menu (non-
exposure phases when in cycle).
SF024 FAST LOG Activates the short logging interval set in the printer definition menu (expo-
sure and liquid exposure phases only).
SF025 ALARM PHASE Alarm phase PH9998 is activated when this flag is set.
SF026 BUZZER All panel buzzers are activated when this flag is set.
SF027 BCKUP PROM ERROR Not used for this sterilizer model.
SF028 PC PRINT LOG
SF029 PC DISK LOG The log sent to the CS100/CS1000 is stored on disk if this flag is set.
SF032 AI. CH 00 ERROR Indicates an error at the analog input channel 0. The flag is reset if the input is
resumed.
SF033 AI. CH 01 ERROR Indicates an error at the analog input channel 1. The flag is reset if the input is
resumed.
SF035 AI. CH 03 ERROR Indicates an error at the analog input channel 3. The flag is reset if the input is
resumed.
SF064 P01 Program selection flags. 433HCMC & 533HCMC has 22 cycles.
thru
SF087 P22
A–2
433HCMC & 533HCMC Steam Sterilizers
SF128 DO MOD. 0 ERROR This flag is set if communication with Digital Output module 0 has failed 10
times in sequence. Resets at power up.
SF129 DO MOD. 1 ERROR This flag is set if communication with Digital Output module 1 has failed 10
times in sequence. Resets at power up.
SF136 DI MOD. 0 ERROR This flag is set if communication with Digital Input module 0 has failed 10
times in sequence. Resets at power up.
SF137 DI MOD. 1 ERROR This flag is set if communication with Digital Input module 1 has failed 10
times in sequence. Resets at power up.
SF151 MANUAL MODE SET This flag is set if any digital output is in manual mode.
SF162 LOG VAR 1 – 20 This flag indicates which log group is printed.
SF165 BATTERY LOW This flag is set when the battery power is low.
61301609701 A–3
System and User Flags
SF179 CORRECT PROCESS Set after a correct process and remaining time is updated.
SF181 MORE BUTTONS Forces the display into “More Buttons” mode.
SF182 SCREEN ON/OFF Associated with the CONTROLS ON/OFF button. Set when the controls are
OFF. Resets when the controls are ON.
SF183 CYCLE SELECTED Sets after the ENTER button is pressed when selecting a new cycle. Resets
after one second.
A–4
433HCMC & 533HCMC Steam Sterilizers
USER FLAGS
User flags indicate events that occur in the application program (Boolean
language) portion of the control software.
UF001 Start OK Set when cycle can be started (no failure conditions, doors OK).
UF002 Door Unseal OK Set when chamber pressure is within 2 psi of atmosphere.
UF010 Negative Pulse 1 Set when negative pulse 1 is in progress (first pulse).
UF011 Negative Pulse 2 Set when negative pulse 2 is in progress (subsequent pulses).
UF012 Chamber Temp Low Set when chamber temperature is below allowable range.
UF016 Loss of Dr Sw Set when door switch or pressure switch input is lost.
UF017 Start phs Set when the controls are in the Start phase.
UF018 Purge phs Set when controls are in the Purge stage.
UF019 Standby phs Set when the controls are in the Standby phase.
UF020 Cycle Complete Set when the controls are in the Cycle Complete phase.
61301609701 A–5
System and User Flags
UF024 End Liq Exh Service diagnostic flag. If the door has been manually retracted during a cycle,
this flag is set in Liq Exhaust so the controls have a way to exit the phase.
UF025 Liq Exh Too Fast Set when the Liquid Exhaust phase exceeds the programmed rate.
UF030 Liq Lev – 1sec Set when liquid is detected in the drain for 1 second. Secondary drain valve
14SV opens.
UF031 Liq Lev – 6sec Set when liquid is detected in the drain for 6 seconds. Ejector and chamber
drain valve 2SV opens.
UF032 Liq Lev – 11sec Set when liquid is detected in the drain for 11 seconds. WATER IN DRAIN
message occurs.
UF033 Fault non cyc/dr Set when a fault occurs when not in cycle.
UF034 Press Low in Tnk Not Used for this sterilizer model.
UF035 D/D Doors Closed Set when both doors are closed (double door units).
UF036 S/D Door Closed Set when door is closed (single door units).
UF040 Chk for Temp Low When set, the controls check for low temperature during the Exposure phase.
UF041 Reached Neg Pul Set when negative pulse setpoint is reached.
UF043 Dr Sw Fail T/O Set when there has been an automatic ABORT - DR SW while in cycle.
UF044 Buff TI Complete Set when a non-RTD related error is seen and has been present for more than
1 second.
UF045 Start Buff TI Set when a non-RTD related fault or failure occurs. Starts a 1 second buffer
which prevents momentary conditions from causing a fault or failure.
UF046 WS - Config Set when the unit is configured with the water saver option.
A–6
433HCMC & 533HCMC Steam Sterilizers
UF049 S/Fair Tnk - Config Not used for this sterilizer model.
UF051 USD – Config Set when unit is configured with the utility shutdown option.
UF052 Vac Pmp – Config Not used for this sterilizer model.
UF053 Jkt Byps Config Set when unit is configured with the jacket bypass option.
UF054 Already Jumped Set after the CYCLE RESULTS section has been printed following a good
cycle.
UF055 Set MORE This is set to allow a phase to be manually advanced or aborted.
UF057 Set Phs Adv BT This is set to allow a phase to be manually advanced.
UF060 Start RTD Buf TI Set when an RTD related fault or failure occurs. Starts a 3 second buffer which
prevents momentary conditions from causing a fault or failure.
UF061 RTD Buf TI Cmplt Set when an RTD related error is seen and has been present for more than 3
seconds.
UF062 Manual Abort Act Set when a cycle has been manually aborted.
UF065 At Temp Set when the chamber temperature is greater than the exposure setpoint.
Reset when the temperature is below the setpoint.
UF066 Timer Started Set when the chamber temperature is greater than the exposure setpoint for
the first time. Stays set until the cycle is complete.
UF069 Fault in cyc/dr Set when a fault or failure occurs when in cycle.
UF070 Open CE Door Set when appropriate condition is satisfied that tells the controls to open the
CE Door.
UF071 Close CE Door Set when appropriate condition is satisfied that tells the controls to close the
CE Door.
UF072 Seal CE Door Set when appropriate condition is satisfied that tells the controls to seal the CE
Door.
61301609701 A–7
System and User Flags
UF073 Unseal CE Door Set when appropriate condition is satisfied that tells the controls to unseal the
CE Door.
UF074 Open RE Door Set when appropriate condition is satisfied that tells the controls to open the
RE Door.
UF075 Close RE Door Set when appropriate condition is satisfied that tells the controls to close the
RE Door.
UF079 Exp Dropout RESET - TEMP LOW situation. Controls exit Exposure and go back to Heat-
Up to heat the chamber.
UF080 Liq Exp Dropout RESET - TEMP LOW situation. Controls exit Liquid Exposure and go back to
Liquid Heat-Up to heat the chamber.
UF083 Jkt<= 126.6 liq Set when the jacket temperature at the start of purge is less than or equal to
126.6°C (liquid cycle only).
UF084 Mt DD-S Seal Set to maintain a door seal (hold DO10 on). (standard double door unit)
UF085 Mt DD-S Unseal Set to maintain a door unseal (hold DO11 on). (standard double door unit)
UF086 Mt CE Seal Set to maintain a control end (CE) door seal (hold DO10 on). (single door
units)
UF088 Mt CE Unseal Set to maintain a control end (CE) door unseal (hold DO011 on). (single door
units)
A–8
433HCMC & 533HCMC Steam Sterilizers
UF102 Seal Other Door Not used for this sterilizer model.
UF107 CC: Print TOD Set after time of day has been printed following a complete cycle.
UF108 Process Alarm Set when a phase-specific error condition occurs during cycle.
UF109 System Alarm Set when an error condition occurs during cycle.
UF110 Alarm Silence Set when CLEAR/ALARM is pressed to silence the alarm tone.
UF114 P-meters Low Set when the controls have read a low Parameter Check value.
UF115 Prt Actual Time Set when controls have been turned OFF, then ON while in cycle.
UF116 Controls Off Set when the controls have been turned OFF.
UF117 Text on scrn -usd Set if CE DOOR OPEN, RE DOOR OPEN, BATTERY LOW, or MANUAL
MODE info text is visible and the controls are in Utility Shutdown mode.
UF118 Utility Shutdown Set when the controls are in Utility Shutdown mode.
UF119 Manual Phs Adv Set when the PHS ADV softkey has been pressed. Resets after the advance
has been executed.
61301609701 A–9
System and User Flags
UF123 Man Phs Adv Code Set when a cycle has been manually advanced.
UF126 USD Pause Set when Utility Shutdown mode has been paused.
A–10
Numerics Alarm Silence flag (UF110) A–9
Already Jumped flag (UF054) A–7
400 series seismic anchorage drawing 10–10, 10–11, 10–12, Always True Flag flag (UF006) A–5
10–13
analog inputs, viewing 3–5, 7–46
400/500 series
piping package 8–2 analog outputs (not used) 7–45
roughing-in drawings 10–2 assigning
system wiring diagram 9–2 passwords 7–40
500 series seismic anchorage drawing 10–14, 10–15, 10–16, sterilizer name (machine name) 7–34
10–17 At Temp flag (UF065) A–7
AUTOMATIC CALIBRATION - HIGH REF screen 4–11, 4–15,
A 4–20
AUTOMATIC CALIBRATION - LOW REF screen 4–11, 4–15,
abort (manual) 1–10 4–20, 4–23
ABORT MANUAL message 2–14 AUTOMATIC CALIBRATION screen 4–10, 4–14, 4–19, 4–22,
ABORT softkey 1–10 7–23
ABORT-AIR REMOVAL message 2–5
ABORT-DOOR SW message 2–5
B
ABORT-EXH SLOWED message 2–5 BAR GRAPH screen 1–6, 7–11
ABORT-HEAT-UP message 2–6 BATTERY LOW flag (SF165) A–3
ABORT-HIGH TEMP message 2–6 BATTERY LOW message 2–7
ABORT-LIQexh SLOW message 2–6 battery replacement
ABORT-SLOW STEAM message 2–7 control board 2–7
ABOUT screen 4–32, 7–13 BLANK TEXT flag (SF020) A–2
accessing menu tree BOOLEAN CODE access 1–14
CALENDAR screen 7–20 BOOLEAN PASSWORD access 1–14
CALIBRATION menu 7–20, 7–23
CONFIGURATION menu 7–32 BRIDGE COMPENSATION screen 5–8, 7–26
DIPSWITCHES screen 7–42 bridge compensation table 2–9, 5–8, 5–9
DOCUMENTATION menu 7–27
PASSWORDS screen 7–40 brightness of display, adjusting 4–32
SAVE RAM TO FLASH screen 7–52 Buff TI Complete flag (UF044) A–6
SERVICE menu 7–43
SYSTEM menu 7–12 BUZZER flag (SF026) A–2
adjustments
brightness of display 4–32 C
buzzer volume 4–32
pressure switch (door gasket) 4–28 CALENDAR access 1–14, 7–21
AI. CH 01 ERROR flag (SF033) A–2 CALENDAR screen 7–19, 7–21
AI. CH 02 ERROR flag (SF034) A–2 calendar, setting 7–21
AI. CH 03 ERROR FLAG (SF035) A–2 calibration
automatic method 4–2, 7–24
AIR-IN SLOW message 2–7
chamber RTD 4–18
alarm buzzer, secondary 4–32 checking 4–4
ALARM CLOCK screen 6–3, 7–22 jacket RTD 4–22
manual method 4–2, 7–25
Alarm Enable flag (UF106) A–9 pressure transducer 4–2, 4–3, 4–8, 7–24
ALARM PHASE flag (SF025) A–2 special tools 4–17
61301609701 I–1
temperature sensors 7–24 LANGUAGE DATE UNITS screen 7–36
temperature sensors (using test plugs) 4–2, 4–17 MACHINE NAME screen 7–34
PANEL SETUP screen 7–38
CALIBRATION access 1–14, 4–8, 4–10, 4–13, 4–14, 4–17, 4–
22, 5–8, 5–15, 5–19, 7–24, 7–25, 7–26 CONFIGURE PRINTER LOG VALUES screen 7–30, 7–31
calibration check control box (2A)
pressure and temperature verification 4–5 electrical schematic 9–8
set up 4–4 wiring diagram 9–4
special tools 4–4
control system
CALIBRATION menu wiring diagram 9–2
AUTOMATIC CALIBRATION screen 7–24
control system adjustments
BRIDGE COMPENSATION screen 7–26
buzzer volume 4–32
MANUAL CALIBRATION screen 7–25
display brightness 4–32
CC Print TOD flag (UF107) A–9
control system maintenance
CE DOOR OPEN message 2–14 replacing flash EPROM 5–10
replacing pressure transducer 5–5
CE NOT SEALED message 2–8
replacing temperature sensors (RTDs) 5–3
CE NOT UNSEALED message 2–8
controls and indicators
CHAM RTD ERROR message 2–8 CLEAR/ALARM switch 1–5
CHAMBER pressure gauge 1–7 CLOSE DOOR switch 1–5
CONTROLS OFF/ON switch 1–5
chamber RTD cycle switches 1–5
calibrating 4–18 directional arrows 1–5
insertion depth 5–4 DOOR(S) CLOSED indicator 1–6
replacing 5–4 DOOR(S) SEALED indicator 1–6
Chamber Temp Low flag (UF012) A–5 ENTER switch 1–5
IN PROCESS indicator 1–6
changing OPEN DOOR switch 1–5
cycle name 7–33 operator panel 1–4
cycle number 7–33 PROCESS COMPLETE 1–6
CHECK BCT VALUES message 2–9 PROCESS FAILURE indicator 1–6
programming switches 1–5
Chk for Temp Low flag (UF040) A–6 START switch 1–5
Chmbr TmpLow Blk flag (UF039) A–6 status indicators 1–6
UP ARROW softkey 1–5
Clear Pressed flag (UF014) A–5
Controls Off flag (UF116) A–9
CLEAR/ALARM switch 1–5, 2–3
CONTROLS OFF/ON switch 1–5
clock, setting 7–21
CORRECT PROCESS flag (SF179) A–4
Close CE Door flag (UF071) A–7
Cycle Aborted flag (UF059) A–7
CLOSE DOOR switch 1–5
CYCLE ADVANCED message 2–15
Close RE Door flag (UF075) A–8
Cycle Complete flag (UF020) A–5
closing door
manual door 1–8 CYCLE EXECUTED flag (SF021) A–2
cold start button (on CPU Board) 5–16 Cycle Good flag (UF005) A–5
cold start procedure 5–16 cycle name, changing 7–33
component locations cycle number, changing 7–33
piping 8–2 cycle parameter popup screen 7–16
CONFIGURATION access 1–14, 6–3, 7–22, 7–33, 7–34, 7– cycle parameters (HC)
36, 7–38 flash 10+ 1–17
CONFIGURATION menu flash 3+ 1–16
ALARM CLOCK screen 7–22 gravity 1–15
EDIT CYCLES screen 7–33 leak test 1–20
I–2
liquids 1–18 CE NOT SEALED 2–8
prevacuum 1–19 CE NOT UNSEALED 2–8
CHAM RTD ERROR 2–8
CYCLE SELECTED flag (SF183) A–4
CHECK BCT VALUES 2–9
cycle types (HC) DI 00-07 COM ERR 2–9
flash 10+ 1–17 DI 08-15 COM ERR 2–9
FLASH 3+ 1–17 DO 00-07 COM ERR 2–9
flash 3+ 1–16 DO 08-15 COM ERR 2–10
gravity 1–15 DO 16-23 COM ERR 2–10
leak test 1–20 DOOR SWITCH FAIL 2–10
liquids 1–18 DOORS NOT SEALED 2–10
prevacuum 1–19 DOORS NOT UNSLD 2–11
cycle worksheets (HC) failure conditions 1–10, 2–1
flash 10+ 1–17 fault conditions 1–10, 2–1
flash 3+ 1–16 informational messages 2–2
gravity cycle 1–15 JACKET OVER-TEMP 2–11
leak test 1–20 JKT RTD ERROR 2–11
liquids 1–18 LEAK TEST FAIL 2–12
prevacuum 1–19 level 1 troubleshooting 2–3
level 2 troubleshooting 2–3
POWER FAILED 2–12
D PRESS ERROR 2–12
PRESS IN CHAMBER 2–13
D/D Doors Closed flag (UF035) A–6 RESET-TEMP LOW 2–13
WATER IN DRAIN 2–13
data entry popup screen 7–6
digital inputs, viewing 3–6
date, setting 7–21
digital outputs, operating manually 3–7
DD P-Interlock flag (UF097) A–9
DIPSWITCHES
DD P-Standard (UF093) A–9 400 series/500 series 4–26
DD P-Unidirect flag (UF095) A–9 DS01 SD=0 DD=1 (single door/double door) 4–26
DS08 Utility S.D. 4–27
deleting user from password list 7–41
DS11 Parameter Check 4–27
DETAILS screen 1–6, 7–10
DIPSWITCHES screen 5–16, 7–42
DI 00-07 COM ERR message 2–9 door configuration 4–26
DI 08-15 COM ERR message 2–9 directional arrows 1–5
DI MOD. 0 ERROR flag (SF136) A–3 display brightness adjustment 4–32
DI MOD. 1 ERROR flag (SF137) A–3 display screens
ABOUT 4–32, 7–13
DIAGNOSTIC menu
ALARM CLOCK 6–3, 7–22
LED/BUZZER screen 7–51
AUTOMATIC CALIBRATION 4–10, 4–14, 4–19, 4–22, 7–
PRINTER screen 7–51
23
TEST ANALOG IN screen 7–46
AUTOMATIC CALIBRATION - HIGH REF 4–11, 4–15, 4–
TEST DIGITAL IN screen 7–47
20
TEST DIGITAL OUT screen 7–48
AUTOMATIC CALIBRATION - LOW REF 4–11, 4–15, 4–
TEST SYSTEM FLAG screen 7–50
20, 4–23
TEST USER FLAG screen 7–49
BAR GRAPH 1–6, 7–11
diagnostic messages BRIDGE COMPENSATION 5–8, 7–26
ABORT-AIR REMOVAL 2–5 CALENDAR 7–21
ABORT-DOOR SW 2–5 CALIBRATION menu 7–20, 7–23
ABORT-EXH SLOWED 2–5 CONFIGURATION menu 7–32
ABORT-HEAT-UP 2–6 CONFIGURE PRINTER LOG VALUES 7–30, 7–31
ABORT-HIGH TEMP 2–6 cycle parameter popup 7–16
ABORT-LIQexh 2–6 data entry popup 7–6
ABORT-SLOW STEAM 2–7 DETAILS 1–6, 7–10
AIR-IN SLOW 2–7 DIAGNOSTIC menu 7–45
BATTERY LOW 2–7
61301609701 I–3
DIPSWITCHES 4–26, 5–16, 7–42 DOORS NOT SEALED message 2–10
DOCUMENTATION menu 7–27
DOORS NOT UNSLD message 2–11
EDIT CYCLES 5–20, 7–33
EDIT PARAMETERS (full edit) 7–17 Dr Sw Fail T/O flag (UF043) A–6
EDIT PARAMETERS (quick edit) 7–15 drying method, gravity cycle 1–15
ENTER PASSWORD 7–5
ERROR LOG 3–2, 3–3, 7–44 drying recommendations 1–16, 1–19
ERROR LOG -MORE INFO 3–4
full edit 7–17 E
LANGUAGE DATE UNITS 7–36
MACHINE NAME 7–34
MACHINE NAME ENTRY 7–35 EDIT CYCLES screen 5–20, 7–33
MANUAL CALIBRATION 5–15, 7–25 EDIT PARAMETERS screen
PANEL SETUP 7–38 full edit 7–17
password entry 7–5
EDIT PARAMETERS screen (quick edit) 7–15
PASSWORDS 7–40
PLOT GRAPH 1–6, 7–11 editing
PRINTER 7–29 cycle parameters (full edit mode) 7–17
PRINTOUT CYCLE 7–28 cycle parameters (quick edit mode) 7–15
quick edit 7–15
electrical schematics
SAVE RAM TO FLASH 5–10, 7–52
control box wiring 9–4
SELECT CYCLE 7–14
control outputs 9–8
SERVICE menu 7–43
power box (1A) 9–7
SETUP menu 7–9
power box wiring 9–3
SYSTEM menu 7–12, 7–19
system wiring 9–2
TEST ANALOG IN 3–2, 3–5, 7–46
TEST DIGITAL IN 3–2, 3–6, 7–47 enabling optional equipment
TEST DIGITAL OUT 3–2, 3–7, 7–48 jacket bypass 4–27
TEST LED/BUZZER 3–2 liquid RTD 4–27
TEST PRINTER 3–2 T-Doc 4–27
TEST SYSTEM FLAG 3–2, 3–10, 7–50 vacuum pump 4–27
TEST USER FLAG 3–2, 3–11, 7–49 water-saver 4–27
VIEW PARAMETERS (full edit) 7–17 enabling software features
VIEW PARAMETERS (quick edit) 7–15 parameter check 4–27
DO 00-07 COM ERR message 2–9 utilities control 4–27
DO 08-15 COM ERR message 2–10 End Liq Exh flag (UF024) A–6
DO 16-23 COM ERR message 2–10 ENTER PASSWORD screen 7–5
DO MOD. 0 ERROR flag (SF128) A–3 ENTER switch 1–5
DO MOD. 1 ERROR flag (SF129) A–3 ENTER TO EDIT message 2–15
DOCUMENTATION access 1–14, 7–28, 7–29, 7–30 entering
alphanumeric value 7–5
DOCUMENTATION menu
data value 7–6
CONFIGURE PRINTER LOG VALUES screen 7–30
password 7–5
PRINTER screen 7–29
PRINTOUT CYCLE screen 7–28 ERROR LOG - MORE INFO screen 3–4
door configurations ERROR LOG screen 3–2, 3–3, 7–44
double door (standard) 1–12 error log, viewing 3–3
single door 1–12
exiting display screen 7–8
door interlock switches 3–13
Exp Dropout flag (UF079) A–8
DOOR SWITCH FAIL message 2–10
Door Unseal OK flag (UF002) A–5 F
DOOR(S) CLOSED indicator 1–6
DOOR(S) SEALED indicator 1–6 factory settings (HC)
B-D TEST 1–2, 1–19
I–4
FLASH 10+ 1–2, 1–17 cycle worksheet 1–16
FLASH 3+ 1–2, 1–16 factory settings 1–16
GRAVITY 1 1–2, 1–15
flash EPROM
GRAVITY 2 1–2, 1–15
clearing CPU Board memory 5–16
LEAK TEST 1–20
entering configuration settings 5–19
LEAKTEST 1–2
entering the customer’s settings 5–20
LIQUIDS 1 1–18
entering transducer data 5–19
LIQUIDS 2 1–18
installing 5–18
LIQUIDS1 1–2
recording calibration data 5–10
LIQUIDS2 1–2
recording sterilizer configuration 5–15
PREVAC 1 1–2, 1–19
recording the customer’s settings 5–12
PREVAC 2 1–2
removing 5–16
PREVAC3 1–2
replacing 5–10
failure conditions 1–10 special tools 5–10
ABORT HEAT-UP 2–6
Fo Auto Selected flag (UF026) A–6
ABORT-AIR REMOVAL 2–5
ABORT-EXH SLOWED 2–5 full edit mode
ABORT-HIGH TEMP 2–6 editing parameters 7–17
ABORT-LIQexh 2–6 selecting 7–39
ABORT-SLOW STEAM 2–7 viewing parameters 7–17
AIR-IN SLOW 2–7
ANORT-DOOR SW 2–5 G
CHAM RTD ERROR 2–8
DI 00-07 COM ERR 2–9
DO 00-07 COM ERR 2–9 gravity cycle (HC)
DO 08-15 COM ERR 2–10 cycle parameters 1–15
DO 16-23 COM ERR 2–10 cycle worksheet 1–15
JACKET OVER-TEMP 2–11 drying method 1–15
JKT RTD FAILURE 2–11 drying recommendation 1–16
POWER FAILED 2–12 factory settings 1–15
PRESS ERROR 2–12 Gravity Dry Jump flag (UF038) A–6
WATER IN DRAIN 2–13
FAST LOG flag (SF024) A–2 H
fast log interval 7–29
fault conditions 1–10 hardware configuration (DIPSWITCHES)
AIR-IN SLOW 2–7 door configuration 4–26
BATTERY LOW 2–7 model configuration (400/500) 4–26
CE NOT SEALED 2–8 optional equipment 4–27
CE NOT UNSEALED 2–8 HARDWARE SETUP access 4–26, 4–27, 5–16, 7–42
CHECK BCT VALUES 2–9
DI 08-15 COM ERR 2–9
DOOR SWITCH FAIL 2–10
I
DOORS NOT SEALED 2–10
DOORS NOT UNSLD 2–11 IN PROCESS indicator 1–6
LEAK TEST FAIL 2–12 In Progress flag (UF004) A–5
PRESS IN CHAMBER 2–13
RESET-TEMP LOW 2–13 informational messages
WATER IN DRAIN 2–13 ABORT MANUAL 2–14
CE DOOR OPEN 2–14
Fault in cyc/dr flag (UF069) A–7 CYCLE ADVANCED 2–15
Fault non cyc/dr flag (UF033) A–6 ENTER TO EDIT 2–15
LOW CYCLE VALUES 2–15
flash 10+ cycle
LOW POWER BROWN OUT 2–15
cycle worksheet 1–17
MANUAL MODE 2–16
factory settings 1–17
RE DOOR OPEN 2–16
FLASH 3+ cycle 1–17 UTILITY SHUTDOWN 2–16
WAIT FOR JACKET 2–16
flash 3+ cycle 1–16
61301609701 I–5
input/output LEDs LOG flag (SF031) A–2
on CM1 Board 2U 3–7
log value
insertion depth item list 7–31
chamber RTD 5–4 type list 7–30
jacket RTD 5–4
LOG VAR 1 – 20 flag (SF162) A–3
INT I2C ERROR flag (SF000) A–1
Loss of Dr Sw flag (UF016) A–5
LOW CYCLE VALUES message 2–15
J
LOW POWER BROWN OUT message 2–15, 3–14
JACKET OVER-TEMP message 2–11
JACKET pressure gauge 1–7
M
jacket RTD
MACHINE NAME ENTRY screen 7–35
calibrating 4–22
insertion depth 5–4 MACHINE NAME screen 7–34
replacing 5–4 machine name, assigning 7–34
Jkt Byps Config flag (UF053) A–7 maintenance
JKT RTD ERROR message 2–11 control system 5–3
door 5–25
Jkt<= 126.6 liq flag (UF083)> A–8
piping system 5–23
Man Phs Adv Code flag (UF123) A–10
L
manual abort
during a timed phase 1–10
LANGUAGE DATE UNITS screen 7–36
during an untimed phase 1–10
leak test cycle (HC)
Manual Abort Act flag (UF062) A–7
cycle parameters 1–20
cycle worksheet 1–20 MANUAL CALIBRATION screen 5–15, 7–25
factory settings 1–20
manual door
LEAK TEST FAIL message 2–12 closing 1–8
opening 1–8
LED/BUZZER screen 7–51
manual mode 3–7
lift/swing door
adjustments 4–35 MANUAL MODE message 2–16
maintenance 5–25
MANUAL MODE SET flag (SF151) A–3
Liq Exh Too Fast flag (UF025) A–6
manual operation of solenoid valves 3–7
Liq Exp Droput flag (UF080) A–8
Manual Phs Adv flag (UF119) A–9
Liq Lev – 11sec flag (UF032) A–6
menu tree
Liq Lev – 1sec flag (UF030) A–6 CALENDAR screen 7–20
CALIBRATION menu 7–20, 7–23
Liq Lev – 6sec flag (UF031) A–6
CONFIGURATION menu 7–32
Liquid Cycle flag (UF029) A–6 DIPSWITCHES screen 7–42
liquids cycle (HC) DOCUMENTATION menu 7–27
cycle parameters 1–18 PASSWORDS screen 7–40
cycle worksheet 1–18 SAVE RAM TO FLASH screen 7–52
factory settings 1–18 SYSTEM MENU 7–12
Load Probe Instr flag (UF028) A–6 menu tree diagram 7–53
I–6
selecting a setup option 7–7 Mt RE Unseal flag (UF089) A–8
messages
ABORT MANUAL 2–14 N
ABORT-AIR REMOVAL 2–5
ABORT-DOOR SW 2–5 naming a cycle 7–33
ABORT-EXH SLOWED 2–5
ABORT-HEAT-UP 2–6 Negative Pulse 1 flag (UF010) A–5
ABORT-HIGH TEMP 2–6 Negative Pulse 2 flag (UF011) A–5
ABORT-LIQexh SLOW 2–6
ABORT-SLOW STEAM 2–7
AIR-IN SLOW 2–7 O
BATTERY LOW 2–7
CE DOOR OPEN 2–14 ON OFF FLAG flag (SF019) A–2
CE NOT SEALED 2–8
OP21 MASTER flag (SF011) A–1
CE NOT UNSEALED 2–8
CHAM RTD ERROR 2–8 OP22 MASTER flag (SF012) A–1
CHECK BCT VALUES 2–9
Open CE Door flag (UF070) A–7
CYCLE ADVANCED 2–15
DI 00-07 COM ERR 2–9 OPEN DOOR switch 1–5
DI 08-15 COM ERR 2–9 Open RE Door flag (UF074) A–8
DO 00-07 COM ERR 2–9
DO 08-15 COM ERR 2–10 opening
DO 16-23 COM ERR 2–10 manual door 1–8
DOOR SWITCH FAIL 2–10 operating digital outputs manually 3–7, 7–48
DOORS NOT SEALED 2–10
DOORS NOT UNSLD 2–11 optional features
ENTER TO EDIT 2–15 parameter check 6–5
JACKET OVER-TEMP 2–11 utilities control 6–2
JKT RTD ERROR 2–11
LEAK TEST FAIL 2–12 P
LOW CYCLE VALUES 2–15
LOW POWER BROWN OUT 2–15, 3–14
P01 thru P22 flags (SF064 thru SF087) A–2
MANUAL MODE 2–16
POWER FAILED 2–12 PANEL 1 ERROR flag (SF001) A–1
PRESS ERROR 2–12
PANEL 2 ERROR flag (SF002) A–1
PRESS IN CHAMBER 2–13
RE DOOR OPEN 2–16 PANEL SETUP screen 7–38
RESET-TEMP LOW 2–13 parameter check feature
UTILITY SHUTDOWN 2–16 description 6–5
WAIT FOR JACKET 2–16 enabling 4–27
WATER IN DRAIN 2–13
PARAMETER SETTINGS access 1–14, 7–15, 7–17
model 433HC
cycle types 1–1 PARAMETER softkey 1–6
factory set parameters 1–1 PASSWORD access 1–14
model 533HC password access levels
cycle types 1–1 BOOLEAN CODE 1–14
factory set parameters 1–1 BOOLEAN PASSWORD 1–14
MORE BUTTONS flag (SF181) A–4 CALENDAR 1–14
CALIBRATION access 1–14
MS250 EXCEEDED flag (SF030) A–2 CONFIGURATION 1–14
Mt CE Seal flag (UF086) A–8 DOCUMENTATION 1–14
PARAMETER SETTINGS 1–14
Mt CE Unseal flag (UF088) A–8
PASSWORD SETUP 1–14
Mt DD-S Seal flag (UF084) A–8 SERVICE 1–14
SYSTEM DEFINITION 1–14
Mt DD-S Unseal flag (UF085) A–8
password entry screen 7–5
Mt RE Seal flag (UF087) A–8
61301609701 I–7
PASSWORD SETUP access 1–14, 7–40 cycle worksheet 1–19
drying recommendations 1–19
passwords
factory settings 1–19
adding user to list 7–41
assigning 7–40 PRINT COM. ERROR flag (SF006) A–1
deleting user from list 7–41
PRINT REALTIME flag (SF167) A–3
entering 7–5
selecting access areas 7–41 printer 1–7
log interval 7–29
PASSWORDS screen 7–40
log values definition 7–30
PC DISK LOG flag (SF029) A–2 testing 3–14
trouble analysis 3–14
periodic maintenance
cleaning water and steam strainers 5–2 PRINTER BUSY flag (SF166) A–3
cleaning water solenoid valve 5–2
PRINTER screen 7–29
every six months 5–2
every three months 5–2 PRINTOUT CYCLE screen 7–28
monthly 5–2 Process Alarm flag (UF108) A–9
periodic maintenance schedule PROCESS COMPLETE indicator 1–6
sterilizer 5–2
Process Failure flag (UF000) A–5
PHS ADV softkey 1–11
PROCESS FAILURE indicator 1–6, 2–3
piping package
400/500 series 8–2 process screen
BAR GRAPH 1–6
piping system maintenance DETAILS 1–6
checking pressure relief valves 5–19 PLOT GRAPH 1–6
PIT EXCEEDED flag (SF016) A–1 selecting default 7–10
PLOT GRAPH screen 1–6, 7–11 program listing, printing 7–28
P-meters Low flag (UF114) A–9 Prt Actual Time flag (UF115) A–9
power box (1A)
electrical schematic 9–7 Q
wiring diagram 9–3
POWER FAILED message 2–12 quick edit mode
editing parameters 7–15
PRESS ERROR message 2–12 selecting 7–39
PRESS IN CHAMBER message 2–13 quick edit screen 7–15
pressure meter specification 4–40
pressure relief valves, checking 5–19 R
pressure switch adjustment 4–28
rack and shelves 1–13
pressure transducer
bridge compensation table 5–8, 5–9 RE DOOR OPEN message 2–16
calibration 4–8 Reached Neg Pul flag (UF041) A–6
data card 5–8
entering data from card 5–8 replacing battery
replacing 5–5 control board 2–7
replacing resistor set 5–7 RESET-TEMP LOW message 2–13
resistor set 5–7
roughing-in drawing
pressure transducer calibration 400/500 series 10–2
high reference value 4–11, 4–15 water treatment system 10–18
low reference value 4–10, 4–14 water-saver 10–18
special tools 4–8, 4–13
RS232 COM. ERROR flag (SF003) A–1
pressure unit, selecting 7–37
RS232 MASTER flag (SF013) A–1
prevacuum cycle (HC)
cycle parameters 1–19 RTD Buf TI Cmplt (UF061) A–7
I–8
S door configuration 4–26
printer log interval 7–29
printer log values definition 7–30
S/D Door Closed flag (UF036) A–6
SETUP menu 7–9
SAVE RAM TO FLASH screen 5–10, 7–52
SETUP softkey 1–6
saving RAM to EPROM 5–9, 7–52
ABOUT screen 7–13
SCREEN ON/OFF flag (SF182) A–4 BAR GRAPH screen 7–11
DETAILS screen 7–10
SD Power flag (UF091) A–8
PLOT GRAPH screen 7–10
Seal CE Door flag (UF072) A–7 SYSTEM MENU 7–12
Seal Other Door flag (UF102) A–9 SF000 INT I2C ERROR flag A–1
Seal RE Door flag (UF076) A–8 SF001 PANEL 1 ERROR flag A–1
secondary alarm buzzer, adjusting 4–32 SF002 PANEL 2 ERROR flag A–1
seismic anchorage drawings SF003 RS232 COM. ERROR flag A–1
400 series 10–10, 10–11, 10–12, 10–13
SF006 PRINT COM. ERROR flag A–1
500 series 10–14, 10–15, 10–16, 10–17
SF011 OP21 MASTER FLAG A–1
SELECT CYCLE screen 7–14
SF012 OP22 MASTER flag A–1
SELECT CYCLE softkey 1–6
SF013 RS232 MASTER A–1
selecting
confirm save/cancel 7–39 SF016 PIT EXCEEDED flag A–1
cycle 7–14
SF019 ON OFF FLAG flag A–2
date format 7–36
default edit mode 7–39 SF020 BLANK TEXT flag A–2
default process screen 7–38 SF021 CYCLE EXECUTED A–2
full edit mode 7–39
item from menu 7–4 SF022 START LOG flag A–2
language 7–36 SF023 SLOW LOG flag A–2
parameter edit mode 7–39
pressure unit 7–37 SF024 FAST LOG flag A–2
process screen 7–10 SF025 ALARM PHASE flag A–2
quick edit mode 7–39
screen saver 7–39 SF026 BUZZER flag A–2
setup option 7–7 SF029 PC DISK LOG flag A–2
temperature unit 7–37
SF030 MS250 EXCEEDED flag A–2
SERVICE access 1–14, 4–12, 4–16, 4–24, 5–10, 7–44, 7–46,
SF031 LOG flag A–2
7–47, 7–48, 7–49, 7–50, 7–51, 7–52
SF033 AI. CH 01 ERROR A–2
SERVICE menu 3–2
DIAGNOSTIC menu 7–45 SF033 AI. CH 01 ERROR flag A–2
ERROR LOG screen 7–44
SF034 AI. CH 02 ERROR flag A–2
SERVICE menu screen 7–43
SF035 AI. CH 03 ERROR flag A–2
Set Abort BT flag (UF056) A–7
SF064 thru SF087 P01 thru P22 flags A–2
Set MORE flag (UF055) A–7
SF128 DO MOD. 0 ERROR flag A–3
Set Phs Adv BT flag (UF057) A–7
SF129 DO MOD. 1 ERROR flag A–3
setting
SF136 DI MOD. 0 ERROR flag A–3
alarm clock 7–22
date 7–21 SF137 DI MOD. 1 ERROR flag A–3
time 7–21 SF151 MANUAL MODE SET flag A–3
utilities control OFF time 6–3, 7–22
utilities control ON time 6–3, 7–22 SF162 LOG VAR 1 – 20 flag A–3
61301609701 I–9
SF167 PRINT REALTIME flag A–3 BATTERY LOW (SF165) A–3
BLANK TEXT (SF020) A–2
SF179 CORRECT PROCESS flag A–4
BUZZER (SF026) A–2
SF181 MORE BUTTONS flag A–4 CORRECT PROCESS (SF179) A–4
SF182 SCREEN ON/OFF flag A–4 CYCLE EXECUTED (SF021) A–2
CYCLE SELECTED (SF183) A–4
SF183 CYCLE SELECTED flag A–4 DI MOD. 0 ERROR (SF136) A–3
SLOW LOG flag (SF023) A–2 DI MOD. 1 ERROR (SF137) A–3
DO MOD. 0 ERROR (SF128) A–3
slow log interval 7–29 DO MOD. 1 ERROR (SF129) A–3
softkeys FAST LOG (SF024) A–2
ABORT 1–10 INT I2C ERROR (SF000) A–1
PARAMETERS 7–15 LOG (SF031) A–2
PHS ADV 1–11 LOG VAR 1 – 20 (SF162) A–3
SELECT CYCLE 7–14 MANUAL MODE SET (SF151) A–3
SETUP 7–9 MORE BUTTONS (SF181) A–4
MS250 EXCEEDED (SF030) A–2
software configuration ON OFF FLAG (SF019) A–2
parameter check feature 4–27 OP21 MASTER (SF011) A–1
utilities control feature 4–27 OP22 MASTER (SF012) A–1
special tools P01 thru P22 (SF064 thru SF087) A–2
adapter plugs 4–41 PANEL 1 ERROR (SF001) A–1
combination adapter plug 4–42 PANEL 2 ERROR (SF002) A–1
flash EPROM replacement 5–10 PC DISK LOG (SF029) A–2
pressure meter 4–40 PIT EXCEEDED (SF016) A–1
pressure sensor calibration 4–8, 4–13 PRINT COM. ERROR (SF006) A–1
specifications 4–40 PRINT REALTIME (SF167) A–3
temperature sensor calibration (test plugs) 4–17 PRINTER BUSY (SF166) A–3
temperature test plugs 4–41 RS232 COM. ERROR (SF003) A–1
RS232 MASTER (SF013) A–1
Standby phs flag (UF019) A–5
SCREEN ON/OFF (SF182) A–4
Start Buff TI flag (UF045) A–6 SLOW LOG (SF023) A–2
START LOG (SF022) A–2
Start Buff TI flag UF045 A–6
viewing 3–10
START LOG flag (SF022) A–2
SYSTEM menu 7–12
Start OK flag (UF001) A–5 CALENDAR screen 7–19
Start phs flag (UF017) A–5 CALIBRATION menu 7–20, 7–23
CONFIGURATION menu 7–32
Start RTD Buf TI flag (UF060) A–7 DIPSWITCHES screen 7–42
START switch 1–5 DOCUMENTATION menu 7–27
PASSWORDS screen 7–40
Start TI Fo flag (UF113) A–9 SAVE RAM TO FLASH 7–52
status indicator (printer) 3–14 SERVICE menu 7–43
sterilizer configurations
door types 1–12 T
double door (standard) 1–12
single door 1–12 temperature sensor calibration
sterilizer name, assigning 7–34 chamber RTD 4–18
high reference value 4–20, 4–24
System Alarm flag (UF109) A–9 jacket RTD 4–22
SYSTEM DEFINITION access 1–14 low reference value 4–18, 4–22
special tools 4–17
system flags using test plugs 4–17
AI. CH 01 ERROR A–2
AI. CH 01 ERROR (SF033) A–2 temperature sensor calibration (overview) 4–3
AI. CH 02 ERROR A–2 temperature sensors
AI. CH 03 ERROR (SF035) A–2 insertion depth 5–4
ALARM PHASE (SF025) A–2
I–10
replacing RTDs 5–3 UF025 Liq Exh Too Fast flag A–6
temperature unit, selecting 7–37 UF026 Fo Auto Selected flag A–6
TEST ANALOG IN screen 3–2, 3–5, 7–46 UF027 LoadPb – Config flag A–6
TEST DIGITAL IN screen 3–2, 3–6, 7–47 UF028 Load Probe Instr flag A–6
TEST DIGITAL OUT screen 3–2, 3–7, 7–48 UF029 Liquid Cycle flag A–6
TEST LED/BUZZER screen 3–2 UF030 Liq Lev – 1sec flag A–6
test plugs 4–18 UF031 Liq Lev – 6sec flag A–6
TEST PRINTER screen 3–2 UF032 Liq Lev – 11sec flag A–6
TEST SYSTEM FLAG screen 3–2, 3–10, 7–50 UF033 Fault non cyc/dr flag A–6
TEST USER FLAG screen 3–2, 3–11, 7–49 UF035 D/D Doors Closed flag A–6
testing UF036 S/D Door Closed flag A–6
buzzer 7–51
UF038 Gravity Dry Jump flag A–6
LEDs 7–51
printer 7–51 UF039 Chmbr TmpLow Blk flag A–6
Text on scrn -usd flag (UF117) A–9 UF040 Chk for Temp Low flag A–6
time of day, setting 7–21 UF041 Reached Neg Pul flag A–6
Timer Started flag (UF066) A–7 UF043 Dr Sw Fail T/O flag A–6
transfer carriage 1–13 UF044 Buff TI Complete flag A–6
trouble analysis UF046 WS - Config A–6
control system 3–12 UF047 Unit – 800 flag A–6, A–7
door 3–12
printer 3–14 UF052 Vac Pmp – Config flag A–7
UF053 Jkt Byps Config flag A–7
U UF054 Already Jumped flag A–7
UF055 Set MORE flag A–7
UF000 Process Failure flag A–5
UF056 Set Abort BT flag A–7
UF001 Start OK flag A–5
UF057 Set Phs Adv BT flag A–7
UF002 Door Unseal OK flag A–5
UF059 Cycle Aborted flag A–7
UF004 In Progress flag A–5
UF060 Start RTD Buf TI flag A–7
UF005 Cycle Good flag A–5
UF061 RTD Buf TI Complt flag A–7
UF006 Always True Flag flag A–5
UF062 Manual Abort Act flag A–7
UF007 Uni RE-Enabled flag A–5
UF065 At Temp flag A–7
UF008 UniCE unsld Last flag A–5
UF066 Timer Started flag A–7
UF009 UniRE unsld Last flag A–5
UF069 Fault in cyc/dr flag A–7
UF010 Negative Pulse 1 flag A–5
UF070 Open CE Door flag A–7
UF011 Negative Pulse 2 flag A–5
UF071 Close CE Door flag A–7
UF012 Chamber Temp Low flag A–5
UF072 Seal CE Door flag A–7
UF014 Clear Pressed flag A–5
UF073 Unseal CE Door flag A–8
UF016 Loss of Dr Sw flag A–5
UF074 Open RE Door flag A–8
UF017 Start phs flag A–5
UF075 Close RE Door flag A–8
UF019 Standby phs flag A–5
UF076 Seal RE Door flag A–8
UF020 Cycle Complete flag A–5
UF077 Unseal RE Door flag A–8
UF024 End Liq Exh flag A–6
61301609701 I–11
UF079 Exp Dropout flag A–8 Alarm Silence (UF110) A–9
Already Jumped (UF054) A–7
UF080 Liq Exp Droput flag A–8
Always True Flag (UF006) A–5
UF083 Jkt<= 126.6 liq flag> A–8 At Temp (UF065) A–7
UF084 Mt DD-S Seal flag A–8 Buff TI Complete (UF044) A–6
CC Print TOD (UF107) A–9
UF085 Mt DD-S Unseal flag A–8 Chamber Temp Low (UF102) A–5
UF086 Mt CE Seal flag A–8 Chk for Temp Low (UF040) A–6
Chmbr TmpLow Blk (UF039) A–6
UF087 Mt RE Seal flag A–8 Clear Pressed (UF014) A–5
UF088 Mt CE Unseal flag A–8 Close CE Door (UF071) A–7
Close RE Door (UF075) A–8
UF089 Mt RE Unseal flag A–8 Controls Off (UF116) A–9
UF091 SD Power flag A–8 Cycle Aborted (UF059) A–7
Cycle Complete (UF020) A–5
UF093 DD P-Standard flag A–9
Cycle Good (UF005) A–5
UF095 DD P-Unidirect flag A–9 D/D Doors Closed (UF035) A–6
DD P-Interlock (UF097) A–9
UF097 DD P-Interlock flag A–9
DD P-Standard (UF093) A–9
UF102 Seal Other Door flag A–9 DD P-Unidirect (UF095) A–9
UF106 Alarm Enable flag A–9 Door Unseal OK (UF002) A–5
Dr Sw Fail T/O (UF043) A–6
UF107 CC Print TOD flag A–9 End Liq Exh (UF024) A–6
UF108 Process Alarm flag A–9 Exp Dropout (UF079) A–8
Fault in cyc/dr (UF069) A–7
UF109 System Alarm flag A–9 Fault non cyc/dr (UF033) A–6
UF110 Alarm Silence flag A–9 Fo Auto Selected (UF026) A–6
Gravity Dry Jump (UF038) A–6
UF113 Start TI Fo flag A–9 In Progress (UF004) A–5
UF114 P-meters Low flag A–9 Jkt A–8
Jkt Byps Config (UF053) A–7
UF115 Prt Actual Time flag A–9
Liq Exh Too Fast (UF025) A–6
UF116 Controls Off flag A–9 Liq Exp Droput (UF080) A–8
Liq Lev – 11sec (UF032) A–6
UF117 Text on scrn -usd flag A–9
Liq Lev – 1sec (UF030) A–6
UF118 Utility Shutdown flag A–9 Liq Lev – 6sec (UF031) A–6
UF119 Manual Phs Adv flag A–9 Liquid Cycle (UF029) A–6
Load Probe Instr (UF028) A–6
UF123 Man Phs Adv Code flag A–10 LoadPb – Config (UF027) A–6
UF126 USD Pause flag A–10 Loss of Dr Sw (UF016) A–5
Man Phs Adv Code (UF123) A–10
Uni RE-Enabled flag (UF007) A–5 Manual Abort Act (UF062) A–7
UniCE unsld Last flag (UF008) A–5 Manual Phs Adv A–9
Manual Phs Adv (UF119) A–9
UniRE unsld Last flag (UF009) A–5 Mt CE Seal (UF086) A–8
Unit – 800 flag (UF047) A–6, A–7 Mt CE Unseal (UF088) A–8
Mt DD-S Seal (UF084) A–8
Unseal CE Door flag (UF073) A–8
Mt DD-S Unseal (UF085) A–8
Unseal RE Door flag (UF077) A–8 Mt RE Seal (UF087) A–8
Mt RE Unseal (UF089) A–8
UNSEAL softkey 1–6
Negative Pulse 1 (UF010) A–5
UP ARROW softkey 1–5 Negative Pulse 2 (UF011) A–5
USD – Config flag (UF051) A–7 Open CE Door (UF070) A–7
Open RE Door (UF074) A–8
USD – Config flag UF051 A–7 P-meters Low (UF114) A–9
USD Pause flag (UF126) A–10 Process Alarm (UF108) A–9
Process Failure (UF000) A–5
user flags Prt Actual Time (UF115) A–9
Alarm Enable (UF106) A–9 Reached Neg Pul (UF041) A–6
I–12
RTD Buf TI Cmplt (UF061) A–7 vacuum leak test
S/D Door Closed (UF036) A–6 (HC) 1–20
SD Power (UF091) A–8
vacuum pump
Seal CE Door (UF072) A–7
trouble analysis 6–1
Seal Other Door (UF102) A–9
Seal RE Door (UF076) A–8 verifying
Set Abort BT (UF056) A–7 pressure 4–5
Set MORE (UF055) A–7 temperature 4–5
Set Phs Adv BT (UF057) A–7 VIEW PARAMETERS screen (full edit) 7–17
Standby phs (UF019) A–5
Start Buff TI (UF045) A–6 VIEW PARAMETERS screen (quick edit) 7–15
Start OK (UF001) A–5 viewing
Start phs (U017) A–5 analog inputs 3–5, 7–46
Start RTD Buf TI (UF060) A–7 digital inputs 3–6, 7–47
Start TI Fo (UF113) A–9 digital outputs 7–48
System Alarm (UF109) A–9 error log 3–3, 7–44
Text on scrn -usd (UF117) A–9 system flags 3–10, 7–50
Timer Started (UF066) A–7 user flags 3–11, 7–49
Uni RE-Enabled (UF007) A–5
UniCE unsld Last (UF008) A–5
UniRE unsld Last (UF009) A–5 W
Unit – 800 (UF047) A–6, A–7
Unseal CE Door (UF073) A–8 WAIT FOR JACKET message 2–16
Unseal RE Door (UF077) A–8
WATER IN DRAIN message 2–13
USD – Config (UF051) A–7
USD Pause (UF126) A–10 water treatment system
Utility Shutdown (UF118) A–9 roughing-in drawing 10–18
Vac Pmp – Config (UF052) A–7
water-saver
viewing 3–11
roughing-in drawing 10–18
WS - Config (UF046) A–6
wiring diagrams
using temperature test plugs 4–42
control box (2A) 9–4
utilities control feature power box (1A) 9–3
description 6–2 system 9–2
enabling 4–27
worksheets (HC)
setting on and off times 6–3
flash 10+ cycle 1–17
Utility Shutdown flag (UF118) A–9 flash 3+ cycle 1–16
gravity cycle 1–15
UTILITY SHUTDOWN message 2–16
leak test cycle 1–20
liquids cycle 1–18
V prevacuum cycle 1–19
WS - Config (UF046) A–6
Vac Pmp – Config flag (UF052) A–7
61301609701 I–13
I–14