Rev A 02/2021
Key features :
n High and stable oxygen concentration
n Medical Device CE Class IIb
n Fully automatic operation 24h/24
n Touch screen control panel
n Remotely manageable
Designed for :
n Hospitals and clinics Medical Gas Piping Network
supply
n Replacement of liquid oxygen storage tanks and
cylinders
www.novairmedical.com
Premium RX 540
Oxygen generator 95% +/- 1%
Medical quality
PREMIUM oxygen generators are CE marked as Medical Devices Class IIb
and comply with all medical oxygen standards and regulations:
ISO 7396-1, Oxygen 93% Monograph of European and United States
VISIO2 ® control panel
Pharmacopoeias, HTM 02-01.
Intelligent control
Technical data of the generator 1 VISIO2 is an advanced control device which automatically manages and
monitors the oxygen generator operation. It features a 5.7’’ HD colour touch
Continuous flow2 Nm3/h 24 screen displaying the production parameters and providing a diagnostic tool
Maximum flow Nm3/h 28 in case of default. VisiO2 capabilities include:
n Real times synoptics and curves display,
Outlet pressure barg 4.5 to 6 n Maintenance management,
Air inlet - Oxygen outlet diameter "G 1"1/2- 1/2 " n Alarms management,
n Remote alarm transfer,
Air inlet Pressure requirement barg >7 n Automatic permutation for installations including two oxygen generators,
Air inlet Quality requirement class 1.4.1 n RS 232 - RS 485 Modbus RTU communication protocols.
Dimensions ( W x D x H )
3
mm 1650 x 1200 x 23503 With its recording capabilty, VISIO2 provides healthcare facilities a complete
Weight3 kg 15503 traceability and a continuous quality control of the oxygen produced.
CO < 2 ppm
n Automatic Change Over System ( ACO ) for perfect switch with the backup
CO2 < 150 ppm system
SO2 0 ppm n Final Pressure Regulator System ( PRS ) and extra Filtration System
NO2 0 ppm n MedGas Bank Manifold Systems for back-up ( MGB )
Oil < 0.05 mg / m3 n Condensates Drain System ( CDS )
n 10 bar outlet pressure booster ( MP )
n High Pressure Cylinder Filling System ( HP )
n Turnkey skid container or cabin
The room provided for the installation of the on-site oxygen 4/ Ventilation:
production unit should be dedicated to this use. It shall contain no • Fresh Air Intake minimum 0.10 m3 per Kw equipment output
storage or warehousing of any product whatsoever. In particular, power. Please ask NOVAIR for proper sizing and positionning of
the presence of flammable materials is strictly prohibited the ventilation grids.
(cardboard box, grease...) • Hot air evacuation at the same side as the entry of cold air must
be avoided (ideally = opposite side).
1/ Main construction specifications: • Hot air extraction through mechanical extractor at the top
• Minimum dimensions: according to the equipment. Must be position of the room.
approved by NOVAIR. • Warm air exhaust from compressor must be wrapped and
• Ground: concrete slab to level, overload calculated for a vented outside through duct (ask NOVAIR for design and sizing).
minimum of 1 000 kg/m2.
• Floor: smooth screed with floor washable paint or hard 5/ Power Cabinet:
coating quartz type. Floor siphon to flow into waste water should The power cabinet must be installed inside the room. The
be installed.(see evacuation of condensates section) details will be given by NOVAIR for each facility. It must include a
• Wall: painted or tiled so as to prevent from dust. ground line and other specifications in accordance with the
• Door : size of door must be large enough to allow equipment to regulations in force.
go through.
6/ Departure to the hospital oxygen network:
2/ Evacuation of condensates: The starting point to the oxygen network of the hospital has to be
A connection point for the evacuation of condensed water linked installed within the room.
with waste water network is required. Ideally a Pressure Control System has to be installed prior to
Ideally install a Condensates Separation System (CDS). network departure.
3/ Temperature: 7/ Lighting:
• The room temperature must be above +5°C and below the The room should be well lit, especially around compressors and
maximum temperature mentioned in the equipment specification generators.
(30°C as a standard). If necessary, appropriate means (heating, air
conditioning) must be installed to maintain proper temperature. 8/ Fire safety:
Sprinklers must be installed and security measures must be taken
and displayed in the room :
- Smoking ban
- Open fire ban
Configuration examples
In plant room
Dimensions: 5500 x 5500 x 2500 mm (Optimal recommended dimensions – for any specific configuration, please contact us )
Ventilation duct
(not provided)
m j
Oil water separator (option)
n
l
5500
500
500
500
Air inlet grid
(not provided)
Evacuation of waste water 5500
Electrical cabinet (not provided)
Door dimension: 2000 x 2500 mm