Designed for hospitals with surgical activities and

OXYGEN GENERATOR intensive care units, PREMIUM oxygen generators


guarantee a high oxygen concentration and an

PREMIUM RX 540 exceptional stability. They integrate advanced


features of oxygen quality analysis, control and
remote monitoring.
Medical oxygen - up to 95% +/- 1%

Rev A 02/2021

Key features :
n High and stable oxygen concentration
n Medical Device CE Class IIb
n Fully automatic operation 24h/24
n Touch screen control panel
n Remotely manageable

Designed for :
n Hospitals and clinics Medical Gas Piping Network
supply
n Replacement of liquid oxygen storage tanks and
cylinders

www.novairmedical.com
Premium RX 540
Oxygen generator 95% +/- 1%

Technology & Design


PREMIUM oxygen generators by NOVAIR are based on the Pressure Swing
Adsorption (PSA) process. This technology is a static separation of air
gases through a specific molecular sieve capable to adsorb nitrogen under
pressure. NOVAIR R&D has optimized the process to obtain an oxygen
concentration particularly stable and exceeding the highest quality
standards.

Designed for a simple and quick installation, this model includes:


n 2 PSA sieve bed columns* ( 1 module of 2 )
n 1 monitoring system Visio 2 with digital colour touch screen
n 2 AIR inlet filters 0.01 mg/m3 and 0.1 mg/m3
n 1 OXYGEN outlet filter ( final filter )
*The number of columns may vary in case of container installation / height constraints .

Medical quality
PREMIUM oxygen generators are CE marked as Medical Devices Class IIb
and comply with all medical oxygen standards and regulations:
ISO 7396-1, Oxygen 93% Monograph of European and United States
VISIO2 ® control panel
Pharmacopoeias, HTM 02-01.

Intelligent control
Technical data of the generator 1 VISIO2 is an advanced control device which automatically manages and
monitors the oxygen generator operation. It features a 5.7’’ HD colour touch
Continuous flow2 Nm3/h 24 screen displaying the production parameters and providing a diagnostic tool
Maximum flow Nm3/h 28 in case of default. VisiO2 capabilities include:
n Real times synoptics and curves display,
Outlet pressure barg 4.5 to 6 n Maintenance management,
Air inlet - Oxygen outlet diameter "G 1"1/2- 1/2 " n Alarms management,
n Remote alarm transfer,
Air inlet Pressure requirement barg >7 n Automatic permutation for installations including two oxygen generators,
Air inlet Quality requirement class 1.4.1 n RS 232 - RS 485 Modbus RTU communication protocols.

Dimensions ( W x D x H )
3
mm 1650 x 1200 x 23503 With its recording capabilty, VISIO2 provides healthcare facilities a complete
Weight3 kg 15503 traceability and a continuous quality control of the oxygen produced.

Power supply 120/230 V - 50/60 Hz - 2A


Available options
n SMS / EMAIL module
1
Performances values refer to the following conditions : 0°C and 1 atm
n TCP/IP communication protocol © 2021 All rights reserved
2
Continuous flow is a guaranteed oxygen flow on a continuous operation of the unit
3
Dimensions and weight are indicative and may vary n Medgas Flow® : measure of the oxygen flow consumed by
the healthcare facility
n Pharmagas® : CO/CO2 analysers, dew point, NO/NO2, SO2
Oxygen quality
n Automatic shut-off valve system ( ASV ) meant to stop production
Purity up to 95% +/- 1% when the oxygen purity goes under the required level
Dew point <-50 °C (< 67 ppm ) n Uninterruptible Power Supply ( UPS ) device against micro-power cuts
n Automatic Voltage Stabilizer ( AVS ) to cope with voltage hits and drops
Non contractual document

CO < 2 ppm
n Automatic Change Over System ( ACO ) for perfect switch with the backup
CO2 < 150 ppm system
SO2 0 ppm n Final Pressure Regulator System ( PRS ) and extra Filtration System
NO2 0 ppm n MedGas Bank Manifold Systems for back-up ( MGB )
Oil < 0.05 mg / m3 n Condensates Drain System ( CDS )
n 10 bar outlet pressure booster ( MP )
n High Pressure Cylinder Filling System ( HP )
n Turnkey skid container or cabin

: 139, rue de la Belle Etoile - 95700 Roissy-en-France - FRANCE : +33 1 56 46 02 03 : contact@novair.fr

OXYPLUS Technologies is a brand of


PSA OYXGEN PLANT ROOM PREREQUISITES

Prerequisites for installation of OXYPLUS on-site PSA oxygen production units

The room provided for the installation of the on-site oxygen 4/ Ventilation:
production unit should be dedicated to this use. It shall contain no • Fresh Air Intake minimum 0.10 m3 per Kw equipment output
storage or warehousing of any product whatsoever. In particular, power. Please ask NOVAIR for proper sizing and positionning of
the presence of flammable materials is strictly prohibited the ventilation grids.
(cardboard box, grease...) • Hot air evacuation at the same side as the entry of cold air must
be avoided (ideally = opposite side).
1/ Main construction specifications: • Hot air extraction through mechanical extractor at the top
• Minimum dimensions: according to the equipment. Must be position of the room.
approved by NOVAIR. • Warm air exhaust from compressor must be wrapped and
• Ground: concrete slab to level, overload calculated for a vented outside through duct (ask NOVAIR for design and sizing).
minimum of 1 000 kg/m2.
• Floor: smooth screed with floor washable paint or hard 5/ Power Cabinet:
coating quartz type. Floor siphon to flow into waste water should The power cabinet must be installed inside the room. The
be installed.(see evacuation of condensates section) details will be given by NOVAIR for each facility. It must include a
• Wall: painted or tiled so as to prevent from dust. ground line and other specifications in accordance with the
• Door : size of door must be large enough to allow equipment to regulations in force.
go through.
6/ Departure to the hospital oxygen network:
2/ Evacuation of condensates: The starting point to the oxygen network of the hospital has to be
A connection point for the evacuation of condensed water linked installed within the room.
with waste water network is required. Ideally a Pressure Control System has to be installed prior to
Ideally install a Condensates Separation System (CDS). network departure.

3/ Temperature: 7/ Lighting:
• The room temperature must be above +5°C and below the The room should be well lit, especially around compressors and
maximum temperature mentioned in the equipment specification generators.
(30°C as a standard). If necessary, appropriate means (heating, air
conditioning) must be installed to maintain proper temperature. 8/ Fire safety:
Sprinklers must be installed and security measures must be taken
and displayed in the room :
- Smoking ban
- Open fire ban

The local and international standards and guidelines must be


respected, including EN ISO 7396-1. © 2021 All rights reserved

Final O2 Filtration Automatic Changeover Pressure To Hospital


Electrical Box Filtration System System System Regulator Piping
System Network
Non contractual document

Feed Air Unit Oxygen Generator Cylinders Back-up Cylinders


Filling System
OXYPLUS oxygen plant

: 139, rue de la Belle Etoile - 95700 Roissy-en-France - FRANCE : +33 1 56 46 02 03 : contact@novair.fr


Premium RX 540
Oxygen generator 95% +/- 1%

Configuration examples
In plant room

Dimensions: 5500 x 5500 x 2500 mm (Optimal recommended dimensions – for any specific configuration, please contact us )

Air exhaust fan (not provided)

Ventilation duct
(not provided)

m j
Oil water separator (option)
n
l

5500

500
500

500
Air inlet grid
(not provided)
Evacuation of waste water 5500
Electrical cabinet (not provided)
Door dimension: 2000 x 2500 mm

Installation example. Tailored layout plan supplied to Oxyplus partners.


1 Oxygen generator Premium RX 540
2 Air compressor 37 KW
3 Air dryer
4 Air tank 1500 L
5 Oxygen tank 1000 L

Skid mounted In container equipped with power cabinet, ventilation


and lighting

© 2021 All rights reserved


Non contractual document

: 139, rue de la Belle Etoile - 95700 Roissy-en-France - FRANCE : +33 1 56 46 02 03 : contact@novair.fr