Vibrating Wire Piezometer

Customer: SUPERTEC SAS.


Sales Order: 229356
Customer ID:
Model: VW2100-0.7
Serial Number: VW135287
Mfg Number: P135287
Range: 700 kPa
Cable Length: 50 meters
Cable Marking: 101489 m to 101540 m
Cable Type: EL380004
Cable Colour Code: Red/Black (Coil) Green/White (Thermistor)
Thermistor Type: 3K

Applied First Second Average Calculated Linearity Calculated Polynomial


Pressure Reading Reading Reading Linear Error Polynomial Error
(kPa) (B units) (B units) (B units) (kPa) (%FS) (kPa) (%FS)
0.0 9769 9769 9769 0.8 0.12 0.1 0.01
140.0 9128 9128 9128 139.7 -0.04 139.8 -0.02
280.0 8483 8483 8483 279.4 -0.09 279.9 -0.01
420.0 7836 7836 7836 419.5 -0.07 420.1 0.01
560.0 7187 7187 7187 560.0 -0.00 560.1 0.02
700.0 6537 6538 6538 700.6 0.09 699.9 -0.01
Max Error (%) 0.12 0.02

Linear Calibration Factor: CF = 2.1655e-01 kPa/B unit


Temperature Correction Factor: Tk = 1.7757e-01 kPa/°C rise

Polynomial Gauge Factor:


A = -5.0934e-07 kPa/(B unit)² B = -2.0825e-01 kPa/B unit C = calculate (see below) kPa

Users must establish site zero readings for calculation purposes


Polynomial C = -[A(L0²) + B(L0)]

Pressure is calculated with the following equations:


Linear: P = CF(L0 - L) - Tk(T0 - T) + (S0 - S)
Polynomial: P = A(L²) + B(L) + C -Tk(T0 - T) + (S0 - S)

L0, L = initial (installation) and current readings, in B units


T0, T = initial (installation) and current temperature, in °C
S0, S = initial (installation) and current barometric pressure readings, in kPa
B units = Hz²/1000 ie: 1700 Hz = 2890 B units

Date VW Reading (B Units) Temperature (°C) Baro (mbar)


Shipped Zero Readings:
23 Sep 2021 9770 21.0 1025.6

This instrument has been calibrated using standards traceable to the NIST in compliance with ANSI Z540-1

Technician: Kailah Toews Date: 23/09/2021

Approved: Sarah Scott Date: 23/09/2021