Prajyo Instrument Pvt. Ltd.

PIMS (Formally : Prajyo Instrument& Maintenance Services)

18. 19. 22. 24, Rachana Industrial Complex, Plot No.71/1B/14, Telco Road, Indrayani Corner,
General Block. MIDC, Bhosari, Pune -411026 Mob 8975896690,
Mail pims pune@gmail com, info@prajyo.com, Web. wwwprajyo.com
Calibration of Electro- Technical, Mechanical, Thermal, Mass, Torque, Power, HV, CT/PT

CC501

CALIBRATION CERTIFICATE
Job No. P1220830-15 Calibration Date:30/08/2022 Issue Date. 02/09/2022 Page No.:I of 2

Receipt Date.30/08/2022 Due Date.29 08/2023 Certilicate No. PI220830-15-15

1. CALIBRATED FOR
Gemini Instratech Lid Calibration Location Site
B-83. MIDC, Ambad Nashik Maharashtra India 422010 Cond. On Receipt Satisfactory

2. DETAILS OF UNIT UNDER CALIBRATION (UUC);:


Nomenclature Digital Multimeter Range Selectable
Make : Rishabh
Model : RISH multi.-14S Resolution/l.east Count As per manual

Accuracy(+) As per manual Serial No. 004532


ULR NO ULR-CC250322000051746F Identification No. GIL/ETE/O79

3. DETAILS OF STANDARD USED FOR CALIBRATION

NAME D.INO. CERTIFICATE NO. CALIBRATED BY VALID UPTo


Multi-ProduciCalibrator PV97 PI211016-2-1 CC-2503 15/10/2022
4. WORK INSTRUCTION NO.: CP-T/05 REFERANCE STANDARD NO.: NAA
5. ENVIRONMENTAL DETAILS:
Temperature: 25 + 10 °CC Relative Humidity: 50 + 20 %RH

6. CALIBRATION RESULTS:
CALIBRATION RESULTS:AC Voltage@50 Hz Discipline:Group - Electro-Technical: Alternative Voltage
Sr. No. Range Set Value on Average Reading Error Expanded Coverage Factor
Master On UUC Uncertainty (% (k)
Rdg)
N Rdg
0.3 @50Hz 0.301 0.001 +0.19
3V 1.5 @50Hz_ 1.497 -0.003 +0.054
3.0 @501Hz 2.996 -0.004 0.054
30 V 30@5011z 29.97 -0.03 +0.054
300V 300 @50Hz 299.6 -0.4 +0.054
1000V 1000@501Hz 998 2 +0.078
CALIBRATION RESULTS:DC Voltage Discipline:Group-Electro-Technical:Direct Voltage
Sr. No. Range Set Value on Average Reading Error Expanded Coverage Factor
Master On UUC Uncertainty (% (k)
Rdg)
3 mV 2.99mV -0.01mV t0.2

300 mV
15mV 15.02mV 0.02mV t0.04
30mV 29.95 mV -0.05mV_ +0.02
300 mV 299.91 mV 0.09 mV +0.004
3V 3V 2.998 V -0.002 V 0.019
30 V 30 V 29.97V -0.03 V +0.019
300 V 300 V 299.5V -0.5V +0.019
8 1000V 1000 998 -2 V +0.058

is
Fgauppa PIPL
Approved By: S.E.Shinde/
Calibrated By: Rahul Dora D.S.Deshpande / R.B.Kasar
Calibration Engineer d Technical Manager
F: CP-T/05
Prajyo Instrument Pvt. Ltd.
PIMS (Formally: Prajyo Instrument &Maintenance Services)
18, 19, 22, 24, Rachana Industrial Complex, Plot No. 71/1B/14, Telco Road, Indrayani Corner
General Block, MIDC, Bhosari, Pune -411026. Mob.: 8975896690,
Mail pimspune@gmail.com,info@prajyo.com,Web: www.prajyo.com
Calibration of: Electro - Technical, Mechanical, Thermal, Mass, Torque, Power, HV, CT/PT

CC
2503
Job No.:P1220830-15 Calibration Date:30/08/2022 Issue Date.:02/09/2022 Page No.:2 of 2

Receipt Date:30/08/2022 DueDate.:29/08/2023 Certificate No. P1220830-15-15


CALIBRATION RESULTS: DC Current Diseipline:Group Flectro-Technical: Direct Current
Coveragc Factor
Sr. No. Range Set Value on Average Reading Error Expanded
Master On UUC Uncertainty (% (k)
Rdg)
0.3 mA 0.298 mA -0.002mA +0.19
3 mA 1.5 mA 1.502 mA 0.002 mA 0.042
3.0mA 2.995mA -0.005mA 0.025
300mA 300 299.7 mA -0.3 mA 0.031
10 A
3 A 2.997 A -0.003 A 0.092
10A 9.96 A_ -0.04 A 0.098
CALIBRATION RESULTS: ACCurrent @50 Hz Discipline:Group - Electro-Technical:Alternative Current
r. No. Range Set Value on Average Reading Error Expandedd Coverage Factor
Master On UUC Uncertainty (% (k)
Rdg)
0.3 mA @50Hz 0.296 mA -0.004mA
3 mA
1.5mA @50Hz 1.495 mA -0.005 mA 0.45

3.0 mA @50Hz 2.994 mA -9.006mA 0.44


300mA 300 mA@5OHz 300.2 mA 0.2 mA_ +0.12
5 10A 10A@50Hz_ 9.95 A -0.05 A 0.21
CALIBRATION RESULTIS:Resistance Discipline:Group - Electro-Technical:Resistance
. No. Range Set Value on Average Reading Error Expanded Coverage Factor
Master On UUC Uncertainty (% (k)
Rdg)_
32 98 2 -0.022 +0.43
30 2 15 2 14.96 2 -0.04 2 +0.13
30 2 29.95 2 -0.05 2 +0.064
300 300 2 299.72 -0.3 +0.02.3
3 k2 3 k2 2.998 k2 -0.002k2 +0.023
30k2 30 k2 29.98 k2 -0.02 k2 t0.021
300 k2 300 k2_ 299.6k2 -0.4 k +0.022
8
3 M2 3 k2 2.998 k2 -0.002k2 +0.023
9
30M2 30 kS2 29.97 k2 -0.03 k2 +0.021
7. REMARK
1. UUC Stands for Unit Under Calibration.
2. A sticker indicating "CALIBRATION STATUS" has been
affixed on the UUC.
3. The standards used for
calibration
4. Calibralion points were selected as
were calibrated by
using reference standard traceable to National/International standard.
per customer's speciflications.
5. The Calibralion results reported in this certilicate are
related to this item only and valid at the time of and
under stated condition of the
measurement.
6. The certilicate should not be
reproduced expect in full without the prior permission of
Laboratory.
The reported expanded
uncertainty of measurement is caleulated at approximately 95.45% of confidence level

F (ohpau iP
Calibrated By: Rahul Dora
Calibration Engineer Approved By: S.E.Shinde/
D.S.Deshpande/ R.B.Kasar
Technical Manager
End of Cert1ficate---- F: CP-T/05