You are on page 1of 30

DATA ENTRY SHEET

Doc. # QC/DES/E/01 Issue Date: Dec 24, 2013 Issue # 01 Rev. # 0 Page 1 of 7
Job # & Certificate Details
DATE OF ISSUE 5-Jan-2021
REQUEST # By mail
JOB # QC/JN/21/00005
CERTIFICATE # QC210105-13
Customer Details
CUSTOMER NAME AL HARITH COMPANY IRAQ
ADDRESS North Rumaila Oilfield, Basra, Iraq
UNIT ----
Details of UUC
DESCRIPTION Multifunction Tester
MANUFACTURER Fluke, USA
MODEL 1653B
ASSEST NUMBER N/A
SERIAL NUMBER 1352029
CALIBRATED RANGE
VOLTAGE (DC) 100 mV to 900 V
VOLTAGE (AC) @ 50 Hz 100 mV to 900 V
VOLTAGE (AC) @ 1 kHz 100 mV to 900 V
CURRENT (DC) 100 mA to 10 A
CURRENT (AC) @ 50 Hz 100 mA to 10 A
CURRENT (AC) @ 1 kHz 100 mA to 10 A
RESISTANCE 1W to 100 MW
Capacitance 1 pF to 100 mF
Inductance 1 mH to 100 H
READABILITY / RESOLUTION
VOLTAGE (DC) 0.1 mV 0.001 mV 0.001 V
VOLTAGE (AC) @ 50 Hz 0.1 mV 0.001 mV 0.001 V
VOLTAGE (AC) @ 1 kHz 0.1 mV 0.001 mV 0.001 V
CURRENT (DC) 0.1 mA 0.001 mA 0.001 A
CURRENT (AC) @ 50 Hz 0.1 mA 0.001 mA 0.001 A
CURRENT (AC) @ 1 kHz 0.1 mA 0.001 mA 0.001 A
RESISTANCE 0.1 W 0.001 kW 0.001 MW
Capacitance 0.1 pF 0.01 nF 0.01 mF 0.01 mF
Inductance 0.1 mH 0.01 mH 0.01 H
TYPE OF INDICATION Digital <IMPORTANT: DON'T FORGET TO LOG THE "TYPE OF INDICATION" OF THE INSTRUMENT.
ASSET NUMBER --- <A SELECTION OF EITHER "ANALOGUE" OR "DIGITAL" ONLY.
DATE OF CALIBRATION 5-Jan-2021
DATE OF RE-CALIBRATION Recommended Validity : one ( 1 ) year from the date of calibration, (Where Required) 05-03-2021
CALIBRATED BY Abdulrahman Loay / Shakeel Ahmed
LOCATION QC Labs North Rumailah, Iraq
ENVIRONMENTAL CONDITIONS C
o
%RH
Start 21.3 Co
Start 46.5 %RH
Temperature: End 21.6 oC Humidity : End 44.9 %RH
Avg. 21.5 oC + 0.2 Avg. 45.7 %RH + 0.8
DETAILS OF REF. STD USED

Description Make Model Serial No. ID Calibration Date Certificate # Traceability from (Lab Name)
Multifunction Calibrator Fluke, USA 5522A 2806902 QC-E-01 02/09/2020 F61180119 NIST, USA (Fluke Cal)

Ref Mulitmeter Fluke, USA 8508A 276568089 QC-E-02 02/09/2020 046461 NIST, UK (Fluke Cal)

Decade Resistance
Box Corpico RBB6-B 18F-1093 QC-E-03 02/09/2020 1353397 NPL, UK (Rs Calibration UK)

Programmable 20140627-
Inductance substitutor IET Labs PLS-1492 J1-1419517 QC-E-04 02/09/2020 42142
NIST, USA (IET Labs)
Programmable 20140627-
Inductance substitutor IET Labs PLS-1492 J1-1419517 QC-E-04 02/09/2020 42142
NIST, USA (IET Labs)

DETAILS OF CALIBRATION PROCEDURE(S) USED


Procedure Number Description Type
QC/CP/E/01 Calibration of V, I, R, H & C Lab Developed

Measurement Data for DC Voltage

OFFSET / ZERO SETTING OFFSET / ZERO SETTING OFFSET / ZERO SETTING


Before After Before After Before After
Adjustment Adjustment Adjustment Adjustment Adjustment Adjustment
mV mV mV mV V V
0 0 0 0 0 0

Values observed on the UUC Values obs


Corrected Value
of Ref Std
Error of Std
Rf.Std - E from
on Ref Certificate
x1 x2 x3 x4 x5 Mean Error
mV mV mV mV mV mV mV mV mV mV
99.9 99.9 99.8 99.9 99.9 99.9 100.00 100.00 -0.10 0.00075
mV mV mV mV mV mV mV mV mV mV
0.9 0.9 1.0 0.9 0.9 0.9 1.000 1.000 -0.10 0.0003
9.8 9.9 9.8 9.8 9.9 9.8 10.000 10.000 -0.20 0.0003
100.0 99.9 100.0 100.0 99.9 100.0 100.000 100.000 0.00 0.0003
500.0 499.9 500.0 500.0 499.9 500.0 500.000 499.998 0.00 0.0017
900.0 899.9 900.0 900.0 899.9 900.0 900.000 899.998 0.00 0.0017
V V V V V V V V V V
1.002 1.001 1.002 1.002 1.001 1.002 1.000000 0.999998 0.00200 0.000002
-1.002 -1.001 -1.002 -1.002 -1.001 -1.002 -1.000000 -0.999995 -0.00201 -0.000005
10.01 10.00 10.01 10.01 10.00 10.01 10.00000 9.99998 0.01002 0.000020
100.0 100.1 100.0 100.0 100.1 100.0 100.0000 100.0001 -0.0001 -0.000100
300.1 300.0 300.0 300.1 300.0 300.0 300.000 300.000 0.000 0.0002
501 500 500 501 500 500 500.000 500.000 0.000 0.0002
901 901 900 901 901 901 900.00 900.00 1.00 0.0000
-901 -902 -901 -902 -902 -902 -900.00 -900.00 -2.00 -0.0020000
x1 to x3 are in ascending order x4 and x5 are in decending order E = Error from Cal. Certificate
Page 2 of 7
Measurement Data for AC Voltage @ 50Hz
OFFSET / ZERO SETTING OFFSET / ZERO SETTING OFFSET / ZERO SETTING
Before After Before After Before After
Adjustment Adjustment Adjustment Adjustment Adjustment Adjustment
mV mV mV mV V V
0 0 0 0 0 0

Values observed on the UUC Values obs


Corrected Value
of Ref Std
Error of Std
Rf.Std - E from
on Ref Certificate
x1 x2 x3 x4 x5 Mean Error
mV mV mV mV mV mV mV mV mV mV
99.9 99.9 99.8 99.9 99.9 99.9 100.00 100.00 -0.10 0.00000
mV mV mV mV mV mV mV mV mV mV
0.9 0.9 1.0 0.9 0.9 0.9 1.000 1.000 -0.10 0.0000
9.8 9.9 9.8 9.8 9.9 9.8 10.000 10.000 -0.20 0.0000
100.0 99.9 100.0 100.0 99.9 100.0 100.000 100.000 0.00 0.0000
500.0 499.9 500.0 500.0 499.9 500.0 500.000 499.998 0.00 0.0020
900.0 899.9 900.0 900.0 899.9 900.0 900.000 899.998 0.00 0.0020
V V V V V V V V V V
1.002 1.001 1.002 1.002 1.001 1.002 1.000000 0.999980 0.00202 0.000020
10.01 10.00 10.01 10.01 10.00 10.01 10.00000 9.99998 0.01002 0.000020
100.0 100.1 100.0 100.0 100.1 100.0 100.0000 100.0130 -0.0130 -0.013000
300.1 300.0 300.0 300.1 300.0 300.0 300.000 300.013 -0.013 -0.013000
501 500 500 501 500 500 500.000 500.013 -0.013 -0.013000
901 901 900 901 901 901 900.00 900.08 0.92 -0.080000

x1 to x3 are in ascending order x4 and x5 are in decending order E = Error from Cal. Certificate

Measurement Data for AC Voltage @ 1 kHz


OFFSET / ZERO SETTING OFFSET / ZERO SETTING OFFSET / ZERO SETTING
Before After Before After Before After
Adjustment Adjustment Adjustment Adjustment Adjustment Adjustment
mV mV mV mV V V
0 0 0 0 0 0

Values observed on the UUC Values obs


Corrected Value
of Ref Std
Error of Std
Rf.Std - E from
on Ref Certificate
x1 x2 x3 x4 x5 Mean Error
mV mV mV mV mV mV mV mV mV mV
99.9 99.9 99.8 99.9 99.9 99.9 100.00 100.00 -0.10 0.00075
mV mV mV mV mV mV mV mV mV mV
0.9 0.9 1.0 0.9 0.9 0.9 1.000 1.001 -0.10 -0.0010
9.8 9.9 9.8 9.8 9.9 9.8 10.000 10.001 -0.20 -0.0010
100.0 99.9 100.0 100.0 99.9 100.0 100.000 100.001 0.00 -0.0010
500.0 499.9 500.0 500.0 499.9 500.0 500.000 500.000 0.00 0.0000
900.0 899.9 900.0 900.0 899.9 900.0 900.000 900.000 0.00 0.0000
V V V V V V V V V V
1.002 1.001 1.002 1.002 1.001 1.002 1.000000 1.000010 0.00199 -0.0000100
10.01 10.00 10.01 10.01 10.00 10.01 10.00000 10.00001 0.00999 -0.0000100
100.0 100.1 100.0 100.0 100.1 100.0 100.0000 99.9870 0.0130 0.0130
300.1 300.0 300.0 300.1 300.0 300.0 300.000 299.987 0.013 0.0130
501 500 500 501 500 500 500.000 499.987 0.013 0.01300
901 901 900 901 901 901 900.00 899.93 1.07 0.0720

x1 to x3 are in ascending order x4 and x5 are in decending order E = Error from Cal. Certificate

Page 3 of 7
Measurement Data for DC CURRENT
OFFSET / ZERO SETTING OFFSET / ZERO SETTING OFFSET / ZERO SETTING

Before After Before After Before After


Adjustment Adjustment Adjustment Adjustment Adjustment Adjustment
mA mA mA mA A A
0 0 0 0 0 0

Values observed on the UUC Values obs


Corrected Value
of Ref Std
Error of Std
Rf.Std - E from
on Ref Certificate
x1 x2 x3 x4 x5 Mean Error
mA mA mA mA mA mA mA mA mA mA
99.9 99.9 99.8 99.9 99.9 99.9 100.00 100.00 -0.10 0.00075
mA mA mA mA mA mA mA mA mA mA
0.9 0.9 1.0 0.9 0.9 0.9 1.000 1.000 -0.10 0.00003
9.8 9.9 9.8 9.8 9.9 9.8 10.000 10.000 -0.20 0.00004
100.0 99.9 100.0 100.0 99.9 100.0 100.000 99.999 0.00 0.0010
500.0 499.9 500.0 500.0 499.9 500.0 500.000 499.994 0.01 0.0060
900.0 899.9 900.0 900.0 899.9 900.0 900.000 900.010 -0.01 -0.0100
mA mA mA mA mA mA mA mA mA mA
1.002 1.001 1.002 1.002 1.001 1.002 1.000000 1.000010 0.00199 -0.0000100
-1.002 -1.001 -1.002 -1.002 -1.001 -1.002 -1.000000 -1.000030 -0.00197 0.0000300
3.002 3.002 3.002 3.002 3.001 3.002 3.000000 2.999950 0.00205 0.000050
5.002 5.003 5.003 5.003 5.002 5.003 5.000000 4.999950 0.00305 0.000050
9.004 9.003 9.003 9.004 9.003 9.003 9.000000 8.999800 0.00320 0.0002
x1 to x3 are in ascending order x4 and x5 are in decending order E = Error from Cal. Certificate

Measurement Data for AC CURRENT @ 50 Hz


OFFSET / ZERO SETTING OFFSET / ZERO SETTING OFFSET / ZERO SETTING

Before After Before After Before After


Adjustment Adjustment Adjustment Adjustment Adjustment Adjustment
mA mA mA mA A A
0 0 0 0 0 0

Values observed on the UUC Values obs Corrected Value


of Ref Std
Error of Std
Rf.Std - E from
on Ref Certificate
x1 x2 x3 x4 x5 Mean Error
mA mA mA mA mA mA mA mA mA mA
99.9 99.9 99.8 99.9 99.9 99.9 100.000 99.999 -0.099 0.00075
mA mA mA mA mA mA mA mA mA mA
0.9 0.9 1.0 0.9 0.9 0.9 1.000 1.000 -0.10 0.0000
9.8 9.9 9.8 9.8 9.9 9.8 10.000 10.000 -0.20 0.0000
100.0 99.9 100.0 100.0 99.9 100.0 100.000 100.000 0.00 0.0000
500.0 499.9 500.0 500.0 499.9 500.0 500.000 500.030 -0.03 -0.0300
900.0 899.9 900.0 900.0 899.9 900.0 900.000 899.900 0.10 0.1000
mA mA mA mA mA mA mA mA mA mA
1.002 1.001 1.002 1.002 1.001 1.002 1.000000 1.000100 0.00190 -0.00010
3.002 3.002 3.002 3.002 3.001 3.002 3.000000 3.000040 0.00196 -0.00004
5.002 5.003 5.003 5.003 5.002 5.003 5.000000 5.000040 0.00296 -0.00004
9.004 9.003 9.003 9.004 9.003 9.003 9.000000 9.001000 0.00200 -0.0010
x1 to x3 are in ascending order x4 and x5 are in decending order E = Error from Cal. Certificate

Page 4 of 7
Measurement Data for AC CURRENT @ 1 kHz
OFFSET / ZERO SETTING OFFSET / ZERO SETTING OFFSET / ZERO SETTING

Before After Before After Before After


Adjustment Adjustment Adjustment Adjustment Adjustment Adjustment
mA mA mA mA A A
0 0 0 0 0 0

Values observed on the UUC Values obs


Corrected Value
of Ref Std
Error of Std
Rf.Std - E from
x1 x2 x3 x4 x5 Mean on Ref Error Certificate
mA mA mA mA mA mA mA mA mA mA
99.9 99.9 99.8 99.9 99.9 99.9 100.00 100.00 -0.10 0.00075
mA mA mA mA mA mA mA mA mA mA
0.9 0.9 1.0 0.9 0.9 0.9 1.000 1.000 -0.10 0.0000
9.8 9.9 9.8 9.8 9.9 9.8 10.000 10.000 -0.20 0.0000
100.0 99.9 100.0 100.0 99.9 100.0 100.000 100.020 -0.02 -0.0200
500.0 499.9 500.0 500.0 499.9 500.0 500.000 500.040 -0.04 -0.0400
900.0 899.9 900.0 900.0 899.9 900.0 900.000 900.050 -0.05 -0.0500
mA mA mA mA mA mA mA mA mA mA
1.002 1.001 1.002 1.002 1.001 1.002 1.000000 0.999950 0.00205 0.0000500
3.002 3.002 3.002 3.002 3.001 3.002 3.000000 2.999780 0.00222 0.000220
5.002 5.003 5.003 5.003 5.002 5.003 5.000000 4.999780 0.00322 0.000220
9.004 9.003 9.003 9.004 9.003 9.003 9.000000 9.001000 0.00200 -0.0010
x1 to x3 are in ascending order x4 and x5 are in decending order E = Error from Cal. Certificate

Measurement Data for Resistance


OFFSET / ZERO SETTING OFFSET / ZERO SETTING OFFSET / ZERO SETTING

Before After Before After Before After


Adjustment Adjustment Adjustment Adjustment Adjustment Adjustment
W W kW kW MW MW
0 0 0 0 0 0

Values observed on the UUC Values obs


Corrected Value
of Ref Std
Error of Std
Rf.Std - E from
x1 x2 x3 x4 x5 Mean on Ref Error Certificate
W W W W W W W W W W
1.1 1.1 1.1 1.1 1.0 1.1 1.00000 1.00010 0.09990 -0.00010
10.1 10.1 10.2 10.0 10.1 10.1 10.00000 9.99990 0.10010 0.0001
100.1 100.1 100.2 100.1 100.2 100.1 100.0000 99.9994 0.1006 0.0006
900.1 900.2 900.1 900.2 900.2 900.2 900.0000 900.0000 0.2000 0.0000
kW kW kW kW kW kW kW kW kW kW
1.002 1.001 1.002 1.002 1.001 1.002 1.000000 1.000000 0.00200 0.0000000
10.002 10.002 10.002 10.001 10.002 10.002 10.00000 9.99996 0.00204 0.0000400
100.0 100.1 100.0 100.1 100.1 100.1 100.0000 100.0006 0.0994 -0.0006
500.1 500.1 500.2 500.1 500.2 500.1 500.0000 499.9993 0.1007 0.0007
900.1 900.2 900.3 900.1 900.2 900.2 900.000 899.999 0.201 0.0010
MW MW MW MW MW MW MW MW MW MW
1.001 1.000 1.000 1.001 1.001 1.001 1.000000 1.000001 0.000999 -0.000001
10.001 10.000 10.000 10.000 10.001 10.000 10.00000 10.00003 -0.00003 -0.00003
30.002 30.002 30.002 30.001 30.001 30.002 30.0000 29.9990 0.0030 0.00096
50.003 50.002 50.003 50.001 50.002 50.002 50.0000 49.9990 0.0030 0.00096
90.00 90.00 90.00 89.99 90.00 90.00 90.0000 89.9710 0.0290 0.0290
x1 to x3 are in ascending order x4 and x5 are in decending order E = Error from Cal. Certificate Page 5 of 7

Measurement Data for Capacitance

OFFSET / ZERO SETTING

Before After Before After Before Before After


After Adjustment
Adjustment Adjustment Adjustment Adjustment Adjustment Adjustment Adjustment
pF pF nF nF mF mF mF mF
0 0 0 0 0 0 0 0

Values observed on the UUC Values obs


Corrected Value
of Ref Std
Error of Std
Rf.Std - E from
on Ref Certificate
x1 x2 x3 x4 x5 Mean Error
pF pF pF pF pF pF pF pF pF pF
299.9 299.9 299.8 299.9 299.9 299.9 300.00 299.97 -0.07 0.03
nF nF nF nF nF nF nF nF nF nF
1.00 1.00 1.00 1.00 1.00 1.00 1.0000 1.0009 -0.0009 -0.0009
10.00 10.00 10.00 10.00 10.00 10.00 10.0000 10.0004 -0.0004 -0.00040
100.0 100.1 100.0 100.1 100.1 100.1 100.000 99.971 0.129 0.029
500.1 500.1 500.2 500.1 500.2 500.1 500.000 499.820 0.280 0.180
900.1 900.2 900.3 900.1 900.2 900.2 900.000 899.720 0.480 0.280
mF mF mF mF mF mF mF mF mF mF
1.00 1.00 1.00 1.00 1.00 1.00 1.0000 0.9997 0.0003 0.00028
10.00 10.00 10.00 10.00 10.00 10.00 10.0000 9.9965 0.0035 0.0035
100.0 100.1 100.0 100.1 100.1 100.1 100.000 100.071 0.029 -0.071
500.1 500.1 500.2 500.1 500.2 500.1 500.000 500.538 -0.438 -0.538
900.1 900.2 900.3 900.1 900.2 900.2 900.000 899.000 1.200 1.000
mF mF mF mF mF mF mF mF mF mF
1.00 1.00 1.00 1.00 1.00 1.00 1.0000 1.0010 -0.0010 -0.0010
10.00 10.00 10.00 10.00 10.00 10.00 10.0000 10.0070 -0.0070 -0.0070
50.0 50.1 50.0 50.1 50.1 50.1 50.000 50.023 0.077 -0.023
100.1 100.1 100.2 100.1 100.1 100.1 100.000 100.095 0.005 -0.095

x1 to x3 are in ascending order x4 and x5 are in decending order E = Error from Cal. Certificate
Page 6 of 7

Measurement Data for Inductance

OFFSET / ZERO SETTING OFFSET / ZERO SETTING OFFSET / ZERO SETTING

Before After Before After Before After


Adjustment Adjustment Adjustment Adjustment Adjustment Adjustment
mH mH H H kH kH
0 0 0 0 0 0
Corrected Value Error of Std
Values observed on the UUC Values obs of Ref Std
Rf.Std - E from
Certificate
x1 x2 x3 x4 x5 Mean on Ref Error
mH mH mH mH mH mH mH mH mH mH
1.002 1.001 1.002 1.002 1.001 1.002 1.000000 1.000003 0.00200 -0.0000032
10.002 10.002 10.002 10.001 10.002 10.002 10.00000 10.00000 0.00200 -0.0000032
100.0 100.1 100.0 100.1 100.1 100.1 100.0000 100.0012 0.0988 -0.0012
500.1 500.1 500.2 500.1 500.2 500.1 500.0000 500.0015 0.0985 -0.0015
900.1 900.2 900.3 900.1 900.2 900.2 900.000 900.002 0.198 -0.0018
H H H H H H H H H H
1.002 1.001 1.002 1.002 1.001 1.002 1.000000 1.000003 0.00200 -0.0000032
10.002 10.002 10.002 10.001 10.002 10.002 10.00000 10.00000 0.00200 -0.0000032
100.0 100.1 100.0 100.1 100.1 100.1 100.0000 100.0012 0.0988 -0.0012
500.1 500.1 500.2 500.1 500.2 500.1 500.0000 500.0015 0.0985 -0.0015
900.1 900.2 900.3 900.1 900.2 900.2 900.000 900.002 0.198 -0.0018
kH kH kH kH kH kH kH kH kH kH
1.002 1.001 1.002 1.002 1.001 1.002 1.000000 1.000003 0.00200 -0.0000032
-1.002 -1.001 -1.002 -1.002 -1.001 -1.002 -1.000000 -0.999997 -0.00200 -0.0000032
10.01 10.00 10.01 10.01 10.00 10.01 10.00000 10.00002 0.00998 -0.000016
100.0 100.1 100.0 100.0 100.1 100.0 100.0000 100.0021 -0.0021 -0.00209
300.1 300.0 300.0 300.1 300.0 300.0 300.000 300.007 -0.007 -0.0065
501 500 500 501 500 500 500.000 500.011 0.389 -0.0105
x1 to x3 are in ascending order x4 and x5 are in decending order E = Error from Cal. Certificate
Page 7 of 7
of 7
of 7
of 7
of 7
of 7
of 7
of 7
UNCERTAINTY EVALUATION SHEET
Doc. # QC/US/E/01 Issue Date: Dec 24, 2013 Issue # 01 Rev. # 0 Page 1 of 9

Uncertainty Evaluation for DC Voltage


Standard Uncertainty Budget
Deviation
Uncertainty Uncertainty
S.D. Combined Uncertainty
Uncertainty Due to Drift of the due to Lead Uncertainty due Expanded
Uncertainty due to Combined
Type A resolution of Standard Error or to Hysteresis uncertainty
of Standard
UUC
Accuracy of
Offset of uncertainty (uc)
(uA) UUC (uB3)
(uB4) UUC(uB6) (U)
(uB1) (uB2) UUC(uB5)

 ( x  x) 2

2uc
UC from cal (Res / 2) / √3 Accuracy /√3
n 1 SD /√5 U / √3 Offset/√3 ** (DX/2)/√3 *
certificate 95% C.L.
U of Std from
Coeffient of -- 1 1 1 1 1 1 -- -- Certificate of the
sensitivity
Ref. Std

-- --
Distribution Normal (K=1) Normal (K=1) Rectangular Rectangular Rectangular Rectangular Rectangular
(mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV)
0.040824829 0.018257 0.0006450 0.02886836 0.057737 0.0007448 0.0000 0.0096 0.06800 0.1 0.0013
(mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV)
0.04472136 0.020000 0.0001950 0.000288684 0.057737 0.0002252 0.0000 0.0096 0.06200 0.1 0.0004
0.054772256 0.024495 0.0001950 0.000288684 0.057737 0.0002252 0.0000 0.0048 0.06300 0.1 0.0004
0.054772256 0.024495 0.0001950 0.000288684 0.057737 0.0002252 0.0000 0.0048 0.06300 0.1 0.0004
0.054772256 0.024495 0.0008500 0.000288684 0.057737 0.0009815 0.0000 0.0048 0.06300 0.1 0.0017
0.054772256 0.024495 0.0010000 0.000288684 0.057737 0.0011547 0.0000 0.0048 0.06300 0.1 0.0020
(V) ± (V) ± (V) ± (V) ± (V) ± (V) ± (V) ± (V) ± (V) ± (V) ± (V)
0.000547723 0.000245 0.0000010 0.000288684 0.057737 0.0000012 0.0000 0.0000 0.05800 0.1 0.000002
0.000547723 0.000245 -0.0000025 0.000288684 0.057737 -0.0000029 0.0000 0.0000 0.05800 0.1 -0.000005
0.005477226 0.002449 0.0000100 0.002886836 0.057737 0.00001 0.0000 0.0005 0.05800 0.1 0.00002
0.054772256 0.024495 -0.0000500 0.02886836 0.057737 -0.00006 0.0000 0.0048 0.06900 0.1 -0.0001
0.054772256 0.024495 0.0001000 0.02886836 0.057737 0.00012 0.0000 0.0048 0.06900 0.1 0.0002
0.547722558 0.244949 0.0001000 0.288683603 0.057737 0.00012 0.0000 0.0481 0.38600 1.0 0.0002
0.447213595 0.200000 0.0000000 0.288683603 0.057737 0.00000 0.0000 0.0962 0.36900 1.0 0.000
0.547722558 0.244949 -0.0010000 0.288683603 0.057737 -0.00115 0.0000 0.1925 0.42900 1.0 -0.002

3 5
n
 xi x
u A   u Bi
2 2 i
X  i 1
 i 4
i 1 3 2
3 5
n
 xi x
u A   u Bi
2 2 i
X  i 1
 i 4
* uc = ± i 1 ** 3 2
Uncertainty Evaluation for AC Voltage @ 50 Hz
Standard Uncertainty Budget
Deviation

Uncertainty Uncertainty
S.D. Uncertainty Combined Due to Uncertainty
Drift of the due to Lead Uncertainty due Expanded
Uncertainty due to Combined
Type A resolution of Standard Error or to Hysteresis uncertainty
of Standard Accuracy of uncertainty (uc)
(uA) UUC
UUC (uB3)
(uB4) Offset of UUC(uB6) (U)
(uB1) (uB2) UUC(uB5)

 ( x  x) 2

2uc
UC from cal (Res / 2) / √3 Accuracy /√3
n 1 SD /√5 U / √3 Offset/√3 ** (DX/2)/√3 *
certificate 95% C.L.
U of Std from
Coeffient of
sensitivity -- 1 1 1 1 1 1 -- -- Certificate of the
Ref. Std
-- --
Distribution Normal (K=1) Normal (K=1) Rectangular Rectangular Rectangular Rectangular Rectangular
(mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV)
0.04472136 0.020000 0.0006450 0.02886836 0.057737 0.0007448 0.0000 0.0096 0.06800 0.1 0.0013
(mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV)
0.04472136 0.020000 0.0005500 0.000288684 0.057737 0.0006351 0.0000 0.0096 0.06200 0.1 0.0011
0.054772256 0.024495 0.0005500 0.000288684 0.057737 0.0006351 0.0000 0.0048 0.06300 0.1 0.0011
0.054772256 0.024495 0.0010500 0.000288684 0.057737 0.0012125 0.0000 0.0048 0.06300 0.1 0.0021
0.054772256 0.024495 0.0044500 0.000288684 0.057737 0.0051386 0.0000 0.0048 0.06300 0.1 0.0089
0.054772256 0.024495 0.0044500 0.000288684 0.057737 0.0051386 0.0000 0.0048 0.06300 0.1 0.0089
(V) ± (V) ± (V) ± (V) ± (V) ± (V) ± (V) ± (V) ± (V) ± (V) ± (V)
0.000547723 0.000245 0.0000165 0.000288684 0.057737 0.0000191 0.0000 0.0000 0.05800 0.1 0.000033
0.005477226 0.002449 0.0000365 0.002886836 0.057737 0.00004 0.0000 0.0005 0.05800 0.1 0.00007
0.054772256 0.024495 0.0004050 0.02886836 0.057737 0.00047 0.0000 0.0048 0.06900 0.1 0.0008
0.054772256 0.024495 0.0055000 0.02886836 0.057737 0.00635 0.0000 0.0048 0.07000 0.1 0.011
0.547722558 0.244949 0.0055000 0.288683603 0.057737 0.00635 0.0000 0.0481 0.38600 1.0 0.011
0.447213595 0.200000 0.0175000 0.288683603 0.057737 0.02021 0.0000 0.0962 0.37000 1.0 0.035
#DIV/0! #DIV/0! 0.0007326 0.288683603 0.057737 0.00085 0.0000 0.0000 #DIV/0! #DIV/0! 0.001

3 5
n
 xi x
u A   u Bi
2 2 i
X  i 1
 i 4
* uc = ± i 1 ** 3 2
u  uA Bi X  i 1
 i 4
i 1 3 2

Page 2 of 9

Uncertainty Evaluation for AC Voltage @ 1 kHz


Standard Uncertainty Budget
Deviation

Uncertainty Uncertainty
S.D. Uncertainty
Combined
Due to Uncertainty
Drift of the due to Lead Uncertainty due Expanded
Uncertainty due to Combined
Type A resolution of Standard Error or to Hysteresis uncertainty
of Standard Accuracy of uncertainty (uc)
(uA) UUC
UUC (uB3)
(uB4) Offset of UUC(uB6) (U)
(uB1) (uB2) UUC(uB5)

 ( x  x) 2

2uc
UC from cal (Res / 2) / √3 Accuracy /√3
n 1 SD /√5 U / √3 Offset/√3 ** (DX/2)/√3 *
certificate 95% C.L.
U of Std from
Coeffient of
sensitivity -- 1 1 1 1 1 1 -- -- Certificate of the
Ref. Std
Distribution Normal (K=1) Normal (K=1) Rectangular Rectangular Rectangular Rectangular Rectangular -- --
(mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV)
0.04472136 0.020000 0.0006450 0.02886836 0.057737 0.0007448 0.0000 0.0096 0.06800 0.1 0.0013
(mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV) ± (mV)
0.04472136 0.020000 0.0010500 0.000288684 0.057737 0.0012125 0.0000 0.0096 0.06200 0.1 0.0021
0.054772256 0.024495 0.0010500 0.000288684 0.057737 0.0012125 0.0000 0.0048 0.06300 0.1 0.0021
0.054772256 0.024495 0.0010500 0.000288684 0.057737 0.0012125 0.0000 0.0048 0.06300 0.1 0.0021
0.054772256 0.024495 0.0039500 0.000288684 0.057737 0.0045612 0.0000 0.0048 0.06300 0.1 0.0079
0.054772256 0.024495 0.0039500 0.000288684 0.057737 0.0045612 0.0000 0.0048 0.06300 0.1 0.0079
(V) ± (V) ± (V) ± (V) ± (V) ± (V) ± (V) ± (V) ± (V) ± (V) ± (V)
0.000547723 0.000245 0.0000330 0.000288684 0.057737 0.0000381 0.0000 0.0000 0.05800 0.1 0.000066
0.005477226 0.002449 0.0000330 0.002886836 0.057737 0.00004 0.0000 0.0005 0.05800 0.1 0.000066
0.054772256 0.024495 0.0003800 0.02886836 0.057737 0.00044 0.0000 0.0048 0.06900 0.1 0.0008
0.054772256 0.024495 0.0041000 0.02886836 0.057737 0.00473 0.0000 0.0048 0.06900 0.1 0.008
0.547722558 0.244949 0.0041000 0.288683603 0.057737 0.00473 0.0000 0.0481 0.38600 1.0 0.008
0.447213595 0.200000 0.0175000 0.288683603 0.057737 0.02021 0.0000 0.0962 0.37000 1.0 0.035
#DIV/0! #DIV/0! 0.0007326 0.288683603 0.057737 0.00085 0.0000 0.0000 #DIV/0! #DIV/0! 0.001

3 5
n

u  u
2
A
2
Bi
x x i i

** X  
i 1 i 4
* uc = ± i 1
3 2
X  i 1
 i 4

3 2

Page 3 of 9

Uncertainty Evaluation for DC Current


Standard Uncertainty Budget
Deviation

Uncertainty Uncertainty
S.D. Uncertainty
Combined
Due to Uncertainty
Drift of the due to Lead Uncertainty due Expanded
Uncertainty due to Combined
Type A resolution of Standard Error or to Hysteresis uncertainty
of Standard Accuracy of uncertainty (uc)
(uA) UUC
UUC (uB3)
(uB4) Offset of UUC(uB6) (U)
(uB1) (uB2) UUC(uB5)

 ( x  x) 2

2uc
UC from cal (Res / 2) / √3 Accuracy /√3
n 1 SD /√5 U / √3 Offset/√3 ** (DX/2)/√3 *
certificate 95% C.L.
U of Std from
Coeffient of
sensitivity -- 1 1 1 1 1 1 -- -- Certificate of the
Ref. Std
-- --
Distribution Normal (K=1) Normal (K=1) Rectangular Rectangular Rectangular Rectangular Rectangular
(mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA)
0.04472136 0.020000 0.0006450 0.02886836 0.057737 0.0007448 0.0000 0.0096 0.06800 0.1 0.0013
(mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA)
0.04472136 0.020000 0.0000110 0.000288684 0.005774 0.0000127 0.0000 0.0096 0.02300 0.0 0.000022
0.054772256 0.024495 0.0000170 0.000288684 0.005774 0.0000196 0.0000 0.0048 0.02600 0.1 0.000034
0.054772256 0.024495 0.0025500 0.000288684 0.005774 0.0029446 0.0000 0.0048 0.02600 0.1 0.0051
0.054772256 0.024495 0.0043500 0.000288684 0.005774 0.0050231 0.0000 0.0048 0.02600 0.1 0.0087
0.054772256 0.024495 0.0245000 0.000288684 0.005774 0.0282910 0.0000 0.0048 0.04500 0.1 0.049000
(A) ± (A) ± (A) ± (A) ± (A) ± (A) ± (A) ± (A) ± (A) ± (A) ± (A)
0.000547723 0.000245 0.0000245 0.000288684 0.005774 0.0000283 0.0000 0.0000 0.00600 0.01 0.000049
0.000547723 0.000245 0.0000270 0.000288684 0.005774 0.0000312 0.0000 0.0000 0.00600 0.01 0.000054
0.000447214 0.000200 0.0000650 0.002886836 0.005774 0.00008 0.0000 0.0001 0.00600 0.01 0.00013
0.000547723 0.000245 0.0000650 0.02886836 0.005774 0.00008 0.0000 0.0000 0.02900 0.06 0.00013
0.000547723 0.000245 0.0006500 0.02886836 0.005774 0.00075 0.0000 0.0000 0.02900 0.06 0.0013

n 3 5

u
2
A
  u Bi
2
x x i i
i 1 X  i 1
 i 4
n 3 5

u
2
A
  u Bi
2
x x i i
* uc = ± i 1 ** X  i 1
 i 4

3 2

Page 4 of 9

Uncertainty Evaluation for AC Current @ 50 Hz


Standard Uncertainty Budget
Deviation

Uncertainty Uncertainty
S.D. Uncertainty Combined Due to Uncertainty
Drift of the due to Lead Uncertainty due Expanded
Uncertainty due to Combined
Type A resolution of Standard Error or to Hysteresis uncertainty
of Standard Accuracy of uncertainty (uc)
(uA) UUC
UUC (uB3)
(uB4) Offset of UUC(uB6) (U)
(uB1) (uB2) UUC(uB5)

 ( x  x) 2

2uc
UC from cal (Res / 2) / √3 Accuracy /√3
n 1 SD /√5 U / √3 Offset/√3 ** (DX/2)/√3 *
certificate 95% C.L.
U of Std from
Coeffient of
sensitivity -- 1 1 1 1 1 1 -- -- Certificate of the
Ref. Std
Normal (K=1) Normal (K=1) Rectangular Rectangular Rectangular Rectangular Rectangular -- --
Distribution

(mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA)
0.04472136 0.020000 0.0006450 0.02886836 0.057737 0.0007448 0.0000 0.0096 0.06800 0.1 0.0013
(mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA)
0.04472136 0.020000 0.0001150 0.000288684 0.005774 0.0001328 0.0000 0.0096 0.02300 0.0 0.00023
0.054772256 0.024495 0.0001150 0.000288684 0.005774 0.0001328 0.0000 0.0048 0.02600 0.1 0.00023
0.054772256 0.024495 0.0017000 0.000288684 0.005774 0.0019630 0.0000 0.0048 0.02600 0.1 0.0034
0.054772256 0.024495 0.0140000 0.000288684 0.005774 0.0161663 0.0000 0.0048 0.03300 0.1 0.0280
0.054772256 0.024495 0.0600000 0.000288684 0.005774 0.0692841 0.0000 0.0048 0.09500 0.2 0.12000
(A) ± (A) ± (A) ± (A) ± (A) ± (A) ± (A) ± (A) ± (A) ± (A) ± (A)
0.000547723 0.000245 0.0000600 0.000288684 0.005774 0.0000693 0.0000 0.0000 0.00600 0.01 0.000120
0.000447214 0.000200 0.0001350 0.002886836 0.005774 0.00016 0.0000 0.0001 0.00600 0.01 0.00027
0.000547723 0.000245 0.0001350 0.02886836 0.005774 0.00016 0.0000 0.0000 0.02900 0.06 0.00027
0.000547723 0.000245 0.0008500 0.02886836 0.005774 0.00098 0.0000 0.0000 0.02900 0.06 0.0017
3 5
n

u
2
A
  u Bi
2
 xi x i

** X  
i 1 i 4
* uc = ± i 1
3 2

Page 5 of 9

Uncertainty Evaluation for AC Current @ 1 kHz


Standard Uncertainty Budget
Deviation

Uncertainty Uncertainty
S.D. Uncertainty Combined Due to Uncertainty
Drift of the due to Lead Uncertainty due Expanded
Uncertainty due to Combined
Type A resolution of Standard Error or to Hysteresis uncertainty
of Standard Accuracy of uncertainty (uc)
(uA) UUC
UUC (uB3)
(uB4) Offset of UUC(uB6) (U)
(uB1) (uB2) UUC(uB5)

 ( x  x) 2

2uc
UC from cal (Res / 2) / √3 Accuracy /√3
n 1 SD /√5 U / √3 Offset/√3 ** (DX/2)/√3 *
certificate 95% C.L.
U of Std from
Coeffient of
sensitivity -- 1 1 1 1 1 1 -- -- Certificate of the
Ref. Std
Distribution Normal (K=1) Normal (K=1) Rectangular Rectangular Rectangular Rectangular Rectangular -- --
(mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA)
0.04472136 0.020000 0.0006450 0.02886836 0.057737 0.0007448 0.0000 0.0096 0.06800 0.1 0.0013
(mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA) ± (mA)
0.04472136 0.020000 0.0001000 0.000288684 0.005774 0.0001155 0.0000 0.0096 0.02300 0.0 0.0002
0.054772256 0.024495 0.0019000 0.000288684 0.005774 0.0021940 0.0000 0.0048 0.02600 0.1 0.0038
0.054772256 0.024495 0.0110000 0.000288684 0.005774 0.0127021 0.0000 0.0048 0.03100 0.1 0.0220
0.054772256 0.024495 0.0135000 0.000288684 0.005774 0.0155889 0.0000 0.0048 0.03300 0.1 0.0270
0.054772256 0.024495 0.0700000 0.000288684 0.005774 0.0808314 0.0000 0.0048 0.11000 0.2 0.1400
(A) ± (A) ± (A) ± (A) ± (A) ± (A) ± (A) ± (A) ± (A) ± (A) ± (A)
0.000547723 0.000245 0.0000700 0.000288684 0.005774 0.0000808 0.0000 0.0000 0.00600 0.01 0.000140
0.000447214 0.000200 0.0001600 0.002886836 0.005774 0.00018 0.0000 0.0001 0.00600 0.01 0.00032
0.000547723 0.000245 0.0001600 0.02886836 0.005774 0.00018 0.0000 0.0000 0.02900 0.06 0.00032
0.000547723 0.000245 0.0009500 0.02886836 0.005774 0.00110 0.0000 0.0000 0.02900 0.06 0.0019
3 5
n

u  u
2
A
2
Bi
x x i i

** X  
i 1 i 4
* uc = ± i 1
3 2

Page 6 of 9

Uncertainty Evaluation for Resistance


Standard Uncertainty Budget
Deviation

Uncertainty Uncertainty
S.D. Uncertainty Combined Due to Uncertainty
Drift of the due to Lead Uncertainty due Expanded
Uncertainty due to Combined
Type A resolution of Standard Error or to Hysteresis uncertainty
of Standard Accuracy of uncertainty (uc)
(uA) UUC
UUC (uB3)
(uB4) Offset of UUC(uB6) (U)
(uB1) (uB2) UUC(uB5)

 ( x  x) 2

2uc
UC from cal (Res / 2) / √3 Accuracy /√3
n 1 SD /√5 U / √3 Offset/√3 ** (DX/2)/√3 *
certificate 95% C.L.
U of Std from
Coeffient of
sensitivity -- 1 1 1 1 1 1 -- -- Certificate of the
Ref. Std
Distribution Normal (K=1) Normal (K=1) Rectangular Rectangular Rectangular Rectangular Rectangular -- --
(W) ± (W) ± (W) ± (W) ± (W) ± (W) ± (W) ± (W) ± (W) ± (W) ± (W)
0.04472136 0.020000 0.0000365 0.02886836 0.005774 0.0000421 0.0000 0.0144 0.04 0.08 0.000073
0.070710678 0.031623 0.0001100 0.02886836 0.005774 0.0001270 0.0000 0.0241 0.05 0.10 0.00022
0.054772256 0.024495 0.0003700 0.02886836 0.005774 0.0004273 0.0000 0.0048 0.04 0.08 0.00074
0.054772256 0.024495 0.0022500 0.02886836 0.005774 0.0025982 0.0000 0.0192 0.04 0.08 0.0045000
kW ± (kW) ± (kW) ± (kW) ± (kW) ± (kW) ± (kW) ± (kW) ± (kW) ± (kW) ± (kW)
0.000547723 0.000245 0.0000023 0.000288684 0.005774 0.0000026 0.0000 0.0000 0.006 0.01 0.0000045
0.000447214 0.000200 0.0000270 0.000288684 0.005774 0.0000312 0.0000 0.0001 0.006 0.01 0.000054
0.054772256 0.024495 0.0002400 0.000288684 0.005774 0.0002771 0.0000 0.0192 0.032 0.06 0.00048
0.054772256 0.024495 0.0008000 0.000288684 0.005774 0.0009238 0.0000 0.0048 0.026 0.05 0.0016
0.083666003 0.037417 0.0043500 0.000288684 0.005774 0.0050231 0.0000 0.0144 0.041 0.08 0.0087
MW ± (MW) ± (MW) ± (MW) ± (MW) ± (MW) ± (MW) ± (MW) ± (MW) ± (MW) ± (MW)
0.000547723 0.000245 0.0000044 0.000288684 0.057737 0.00001 0.0000 0.0002 0.058 0.12 0.0000087
0.000547723 0.000245 0.0000700 0.000288684 0.005774 0.00008 0.0000 0.0000 0.006 0.01 0.00014
0.000547723 0.000245 0.0005500 0.000288684 0.005774 0.00064 0.0000 0.0003 0.010 0.02 0.0011
0.00083666 0.000374 0.0005500 0.000288684 0.005774 0.00064 0.0000 0.0003 0.010 0.02 0.0011
0.004472136 0.002000 0.0070000 0.000288684 0.005774 0.00808 0.0000 0.0014 0.010 0.02 0.0140

3 5
n
x x
u A   u Bi
2 2 i i
X  i 1
 i 4

* uc = ± i 1 ** 3 2

Page 7 of 9

Uncertainty Evaluation for Capacitance


Standard Uncertainty Budget
Deviation

Uncertainty Uncertainty
S.D. Uncertainty Combined Due to Uncertainty
Drift of the due to Lead Uncertainty due Expanded
Uncertainty due to Combined
Type A resolution of Standard Error or to Hysteresis uncertainty
of Standard Accuracy of uncertainty (uc)
(uA) UUC
UUC (uB3)
(uB4) Offset of UUC(uB6) (U)
(uB1) (uB2) UUC(uB5)

 ( x  x) 2

2uc
UC from cal (Res / 2) / √3 Accuracy /√3
n 1 SD /√5 U / √3 Offset/√3 ** (DX/2)/√3 *
certificate 95% C.L.
U of Std from
Coeffient of
sensitivity -- 1 1 1 1 1 1 -- -- Certificate of the
Ref. Std
-- --
Distribution Normal (K=1) Normal (K=1) Rectangular Rectangular Rectangular Rectangular Rectangular
(pF) ± (pF) ± (pF) ± (pF) ± (pF) ± (pF) ± (pF) ± (pF) ± (pF) ± (pF) ± (pF)
0.04472136 0.020000 0.0010500 0.02886836 0.057737 0.0012125 0.0000 0.0096 0.10000 0.2 0.0021
(nF) ± (nF) ± (nF) ± (nF) ± (nF) ± (nF) ± (nF) ± (nF) ± (nF) ± (nF) ± (nF)
0.000547723 0.000245 0.0020000 0.002886836 0.057737 0.0023095 0.0000 0.0000 0.10000 0.2 0.0040
0.000447214 0.000200 0.0028500 0.002886836 0.057737 0.0032910 0.0000 0.0001 0.05797 0.1 0.0057
0.054772256 0.024495 0.0380000 0.002886836 0.057737 0.0438799 0.0000 0.0192 0.08551 0.2 0.0760
0.054772256 0.024495 0.1050000 0.002886836 0.057737 0.1212471 0.0000 0.0048 0.17224 0.3 0.2100
0.083666003 0.037417 0.3750000 0.002886836 0.057737 0.4330254 0.0000 0.0144 0.57696 1.2 0.7500
(mF) ± (mF) ± (mF) ± (mF) ± (mF) ± (mF) ± (mF) ± (mF) ± (mF) ± (mF) ± (mF)
0.000547723 0.000245 0.0003750 0.002886836 0.057737 0.0004330 0.0000 0.0000 0.05781 0.1 0.000750
0.000447214 0.000200 0.0038000 0.002886836 0.057737 0.0043880 0.0000 0.0001 0.05810 0.1 0.007600
0.054772256 0.024495 0.0750000 0.002886836 0.057737 0.0866051 0.0000 0.0192 0.13064 0.3 0.1500
0.054772256 0.024495 0.0650000 0.002886836 0.057737 0.0750577 0.0000 0.0048 0.11748 0.2 0.1300
0.083666003 0.037417 0.1250000 0.002886836 0.057737 0.1443418 0.0000 0.0144 0.20298 0.4 0.2500
(mF) ± (mF) ± (mF) ± (mF) ± (mF) ± (mF) ± (mF) ± (mF) ± (mF) ± (mF) ± (mF)
0.000547723 0.000245 0.0001900 0.002886836 0.057737 0.00022 0.0000 0.0000 0.05781 0.1 0.00038
0.000447214 0.000200 0.0018500 0.002886836 0.057737 0.00214 0.0000 0.0001 0.05788 0.1 0.0037
0.054772256 0.024495 0.0050000 0.002886836 0.057737 0.00577 0.0000 0.0192 0.06325 0.1 0.0100
0.04472136 0.020000 0.0180000 0.002886836 0.057737 0.02079 0.0000 0.0096 0.06707 0.1 0.0360
0.001

3 5
n
2
  u Bi
2  xi x
* uc = ±
u A
**
X  i 1
 i 4
i

i 1 3 2
Page 8 of 9

Uncertainty Evaluation for Inductance


Standard Uncertainty Budget
Deviation

Uncertainty Uncertainty
S.D. Uncertainty Combined Due to Uncertainty
Drift of the due to Lead Uncertainty due Expanded
Uncertainty due to Combined
Type A resolution of Standard Error or to Hysteresis uncertainty
of Standard Accuracy of uncertainty (u c)
(uA) UUC
UUC (uB3)
(uB4) Offset of UUC(uB6) (U)
(uB1) (uB2) UUC(uB5)

 ( x  x) 2

2uc
UC from cal (Res / 2) / √3 Accuracy /√3
n 1 SD /√5 U / √3 Offset/√3 ** (DX/2)/√3 *
certificate 95% C.L.
U of Std from
Coeffient of
sensitivity -- 1 1 1 1 1 1 -- -- Certificate of the
Ref. Std
-- --
Distribution Normal (K=1) Normal (K=1) Rectangular Rectangular Rectangular Rectangular Rectangular
(mH) ± (mH) ± (mH) ± (mH) ± (mH) ± (mH) ± (mH) ± (mH) ± (mH) ± (mH) ± (mH)
0.000547723 0.000245 0.0005500 0.02886836 0.057737 0.0006351 0.0000 0.0000 0.10000 0.2 0.0011
0.000447214 0.000200 0.0006450 0.02886836 0.057737 0.0007448 0.0000 0.0001 0.06456 0.1 0.0013
0.054772256 0.024495 0.0006450 0.02886836 0.057737 0.0007448 0.0000 0.0192 0.06905 0.1 0.0013
0.054772256 0.024495 0.0006450 0.02886836 0.057737 0.0007448 0.0000 0.0048 0.06905 0.1 0.0013
0.083666003 0.037417 0.0008000 0.02886836 0.057737 0.0009238 0.0000 0.0144 0.07462 0.1 0.0016
H ± (H) ± (H) ± (H) ± (H) ± (H) ± (H) ± (H) ± (H) ± (H) ± (H)
0.000547723 0.000245 0.0000045 0.002886836 0.057737 0.0000052 0.0000 0.0000 0.05781 0.1 0.000009
0.000447214 0.000200 0.0000045 0.002886836 0.057737 0.0000052 0.0000 0.0001 0.05781 0.1 0.000009
0.054772256 0.024495 0.0008790 0.002886836 0.057737 0.0010150 0.0000 0.0192 0.06280 0.1 0.0018
0.054772256 0.024495 0.0008790 0.002886836 0.057737 0.0010150 0.0000 0.0048 0.06280 0.1 0.0018
0.083666003 0.037417 0.0008790 0.002886836 0.057737 0.0010150 0.0000 0.0144 0.06887 0.1 0.0018
kH ± (kH) ± (kH) ± (kH) ± (kH) ± (kH) ± (kH) ± (kH) ± (kH) ± (kH) ± (kH)
0.000547723 0.000245 0.0008027 0.002886836 0.057737 0.00093 0.0000 0.0000 0.05782 0.1 0.0016
0.000547723 0.000245 0.0003347 0.002886836 0.057737 0.00039 0.0000 0.0000 0.05781 0.1 0.001
0.005477226 0.002449 0.0007744 0.002886836 0.057737 0.00089 0.0000 0.0005 0.05787 0.1 0.002
0.054772256 0.024495 0.0027815 0.002886836 0.057737 0.00321 0.0000 0.0048 0.06293 0.1 0.006
0.054772256 0.024495 0.0007326 0.002886836 0.057737 0.00085 0.0000 0.0048 0.06279 0.1 0.001
0.547722558 0.244949 0.0000000 0.002886836 0.057737 0.00000 0.0000 0.0481 0.25168 0.5

3 5
n
2
  u Bi
2  xi x
* uc = ±
u A
**
X  i 1
 i 4
i

i 1 3 2
Page 9 of 9
Calibration Certificate
Al Takamul Yard North Rumailah, Iraq
●Phone : +964 7810009138 ● www.qualitycontrol-iraq.com ● E-mail: op@qualitycontrol-iraq
Date of Issue: 1/5/2021 Page 1 of 5
REQUEST NUMBER : By mail APPROVED BY LAB INCHARGE QC
JOB NUMBER : QC/JN/21/00005
CERTIFICATE NUMBER : QC210105-13 ABDULRAHMAN LOAY

CUSTOMER DETAILS

Name : AL HARITH COMPANY IRAQ


Address : North Rumaila Oilfield, Basra, Iraq

EQUIPMENT IDENTIFICATION AND SPECIFICATIONS

Description : Multifunction Tester


Type of Indication : Digital
Manufacturer : Fluke, USA
Model : 1653B
Asset No. : N/A
Serial Number : 1352029
Calibrated Range:
Voltage (AC) @ 50 Hz 100 mV to 500 V
Voltage (AC) @ 1 kHz 100 mV to 500 V
Current 3 mA to 550 mA
Current (AC) @ 50 Hz 3 mA to 550 mA
Current (AC) @ 1 kHz 100 mA to 550 mA
Resistance 10 W to 2000 W
Resolution:
Voltage (DC) 0.1 mV 0.001 mV 0.001 V
Voltage (AC) @ 50 Hz 0.1 mV 0.001 mV 0.001 V
Voltage (AC) @ 1 kHz 0.1 mV 0.001 mV 0.001 V
Current (DC) 0.1 mA 0.001 mA 0.001 A
Current (AC) @ 50 Hz 0.1 mA 0.001 mA 0.001 A
Current (AC) @ 1 kHz 0.1 mA 0.001 mA 0.001 A
Resistance 0.1 W 0.001 W 0.001 W
Location : QC Labs North Rumailah, Iraq
Calibrated By : Abdulrahman Loay / Shakeel Ahmed
Calibration Date : January 5, 2021
Calibration Due : Recommended Validity : one ( 1 ) year from the date of calibration, (Where Required): 04-01-2022
ENVIRONMENTAL CONDITIONS DURING TEST
Ambient Temperature : AL HARITH CO°C ± 0 °C
Relative Humidity : 0 %RH ± 0 %RH
CALIBRATION METHOD
The above equipment has been calibrated in accordance with QC Calibration Procedure # CURRENT (AC) @ 50 Hz
The deviations of the measurements obtained from UUC with respect to reference standards are determined to obtain the error.
TRACEABILITY
The measurements made by Quality Control Labs are traceable to NIST USA, vide Fluke Cal Cert # 0
The measurements made by Quality Control Labs, realize the physical units of measurements (SI), through its state of
the art calibration standards that are controlled and maintained by QC.
REFERENCE EQUIPMENT USED :

DESCRIPTION MAKE MODEL # SERIAL # CALIBRATION DATE TRACEABILITY


Multifunction Calibrator Fluke, USA 5522A 2806902 02/09/2020 NIST, USA (Fluke Cal)
Ref Mulitmeter Fluke, USA 8508A 276568089 02/09/2020 NIST, UK (Fluke Cal)
Decade Resistance Box Corpico RBB6-B 18F-1093 02/09/2020 NPL, UK (Rs Calibration UK)
Programmable Inductance
IET Labs PLS-1492 J1-1419517 02/09/2020 NIST, USA (IET Labs)
substitutor
CERTIFICATE OF CALIBRATION CERTIFICATE NUMBER

QC210105-13
REQUEST NUMBER: By mail Page 2 of 5
JOB NUMBER: QC/JN/21/00005

CALIBRATION TEST RESULTS

Measurement Data for DC Voltage


Zero or Offset Readings of UUC
Before Before Before
After Adjustment After Adjustment After Adjustment
Adjustment Adjustment Adjustment
mV mV mV mV V V
0 0 0 0 0 0

Readings on UUC Readings on Ref. Standard Error Uncertainty (95 % C.L)


mV mV mV ± (mV)
99.9 100.00 -0.10 0.1
mV mV mV ± (mV)
0.9 1.00 -0.10 0.1
9.8 10.00 -0.20 0.1
100 100.00 0.00 0.1
500 500.00 0.00 0.1
900 900.00 0.00 0.1
V V V ± (V)
1.002 1.00 0.00 0.1
-1.002 -1.00 0.00 0.1
10.01 10.00 0.01 0.1
100 100.00 0.00 0.1
300 300.00 0.00 0.1
500 500.00 0.00 1

Measurement Data for AC Voltage @ 50 Hz


Zero or Offset Readings of UUC
Before Before Before
After Adjustment After Adjustment After Adjustment
Adjustment Adjustment Adjustment

mV mV mV mV V V
0 0 0 0 0 0

Readings on UUC Readings on Ref. Standard Error Uncertainty (95 % C.L)


mV mV mV ± (mV)
99.9 100.000 -0.100 0.1
mV mV mV ± (mV)
0.9 1.000 -0.100 0.1
9.8 10.000 -0.200 0.1
100 100.000 0.000 0.1
500 499.998 0.002 0.1
900 899.998 0.002 0.1
V V V ± (V)
1.002 0.999980 0.002020 0.1
10.01 9.99998 0.01002 0.1
100.0 100.0130 -0.0130 0.1
300.0 300.013 -0.013 0.1
500 500.013 -0.013 1
CERTIFICATE OF CALIBRATION CERTIFICATE NUMBER

QC210105-13
REQUEST NUMBER: By mail Page 3 of 5
JOB NUMBER: QC/JN/21/00005

Measurement Data for AC Voltage @ 1 kHz


Zero or Offset Readings of UUC
Before Before Before
After Adjustment After Adjustment After Adjustment
Adjustment Adjustment Adjustment

mV mV mV mV V V
0 0 0 0 0 0

Readings on UUC Readings on Ref. Standard Error Uncertainty (95 % C.L)


mV mV mV ± (mV)
99.9 100.00 -0.10 0.1
mV mV mV ± (mV)
0.9 1.001 -0.101 0.1
9.8 10.001 -0.201 0.1
100 100.001 -0.001 0.1
500 500.000 0.000 0.1
900 900.000 0.000 0.1
V V V ± (V)
1.002 1.000010 0.001990 0.1
10.01 10.00001 0.00999 0.1
100.0 99.9870 0.0130 0.1
300.0 299.987 0.013 0.1
500 499.987 0.013 1

Measurement Data for DC Current


Zero or Offset Readings of UUC
Before Before Before
After Adjustment After Adjustment After Adjustment
Adjustment Adjustment Adjustment

mA mA mA mA mA mA
0 0 0 0 0 0

Readings on UUC Readings on Ref. Standard Error Uncertainty (95 % C.L)


mA mA mA ± (mA)
99.9 99.999 -0.099 0.1
mA mA mA ± (mA)
0.9 1.000 -0.100 0
9.8 10.000 -0.200 0.1
100 99.999 0.001 0.1
500 499.994 0.006 0.1
900 900.010 -0.010 0.1
mA mA mA ± (A)
1.002 1.000010 0.001990 0.01
-1.002 -1.000030 -0.001970 0.01
3.002 2.999950 0.002050 0.01
5.003 4.999950 0.003050 0.06
9.003 8.999800 0.003200 0.06
CERTIFICATE OF CALIBRATION CERTIFICATE NUMBER

QC210105-13
REQUEST NUMBER: By mail Page 4 of 5
JOB NUMBER: QC/JN/21/00005

Measurement Data for AC Current @ 50 Hz


Zero or Offset Readings of UUC
Before Before Before
After Adjustment After Adjustment After Adjustment
Adjustment Adjustment Adjustment

mA mA mA mA mA mA
0 0 0 0 0 0

Readings on UUC Readings on Ref. Standard Error Uncertainty (95 % C.L)


mA mA mA ± (mA)
99.9 99.999 -0.099 0.1
mA mA mA ± (mA)
0.9 1.000 -0.100 0
9.8 10.000 -0.200 0.1
100 100.000 0.000 0.1
500 500.030 -0.030 0.1
900 899.900 0.100 0.2
mA mA mA ± (A)
1.002 1.000100 0.001900 0.01
3.002 3.000040 0.001960 0.01
5.003 5.000040 0.002960 0.06
9.003 9.001000 0.002000 0.06

Measurement Data for AC Current @ 1 kHz


Zero or Offset Readings of UUC
Before Before Before
After Adjustment After Adjustment After Adjustment
Adjustment Adjustment Adjustment

mA mA mA mA mA mA
0 0 0 0 0 0

Measurement Data for AC Current @ 1 kHz

Readings on UUC Readings on Ref. Standard Error Uncertainty (95 % C.L)


mA mA mA ± (mA)
99.9 99.999 -0.099 0.1
mA mA mA ± (mA)
0.9 1.000 -0.100 0
9.8 10.000 -0.200 0.1
100 100.020 -0.020 0.1
500 500.040 -0.040 0.1
900 900.050 -0.050 0.2
mA mA mA ± (A)
1.002 0.999950 0.002050 0.01
3.002 2.999780 0.002220 0.01
5.003 4.999780 0.003220 0.06
9.003 9.001000 0.002000 0.06
CERTIFICATE OF CALIBRATION CERTIFICATE NUMBER

QC210105-13
REQUEST NUMBER: By mail Page 5 of 5
JOB NUMBER: QC/JN/21/00005

Measurement Data for Resistance


Zero or Offset Readings of UUC
Before Before Before
After Adjustment After Adjustment After Adjustment
Adjustment Adjustment Adjustment

W W kW kW MW MW
0 0 0 0 0 0

Readings on UUC Readings on Ref. Standard Error Uncertainty (95 % C.L)


W W W ± (W)
1.1 1.00010 0.09990 0.08
10.1 9.99990 0.10010 0.10
100.1 99.9994 0.1006 0.08
900.2 900.0000 0.2000 0.08

The reported expanded uncertainty is based on a standard uncertainty multiplied by a coverage factor k=2, providing a level of confidence of
approximately 95%. The uncertainty evaluation has been carried out in accordance with international practice.

DEVIATION FROM STANDARD METHOD : None

REMARK (S) :
The results are as found (no adjustment done).
The results are post adjustment.

This certificate is issued in accordance with the laboratory accreditation requirements as per ISO/IEC17025:2005.
This certificate may not be reproduced other than in full, unless permission for the publication of an approved extract has been obtained
in writing from the director of QC. It does not of itself impute to the subject of the calibration any attributes beyond those shown by
the data contained herein.

~ END OF DOCUMENT ~