Infant Warmer

There are three channel of temperature sensor in the sensor


box. One channel is for reading hood temperature to remain
the temperature at setting temperature.

Another temperature sensor is an individual circuit for prote-


ction only.

The third sensor is used to compare the reading with the


other two sensors.

If the reading of sensor is differ from 0.8 degree, the alarm


will be triggered

The incubator is equipped with both thermocouple and


mechanical temperature switch to protect the heating
source of heating system and evaporator.

Thermocouple is reading the real time temperature and


limits the temperature in safety range by feedback the
temperature value to control module.

While, mechanical temperature switch is in series circ-


uits, if the heating source temperature goes unnormal
level, temperature switch cuts off the circuit physically.