Certificate of Calibration: A. Arámburo
Certificate of Calibration: A. Arámburo
Certificate of Calibration
Manufacturer: Mitutoyo Report #: TDM-116-0000866
Description: Surface Plate P.O. #: N/A
This certifies that the above instrument was calibrated in compliance with the Calibration System Requirements of ISO /IEC
17025:2017, ANSI/NCSL Z540-1-1994 in accordance with referenced procedures. Standards used to perform this calibration
are traceable to SI units; their source of traceability derives from a National Metrology Institute such as NIST, CENAM, NPL,
DIN, from natural physical constants, consensus standards or derived by the ratio type of calibrations. Collective
uncertainties are determined as required with a distribution that corresponds to a probability of approximately 95% (k=2).
Unless otherwise noted calibrations are performed to manufacturer's specifications. Compliance statements are in
conformance with ILAC-G8:2019 simple acceptance decision rule. This form shall not be reproduced, except in full, without
the expressed written consent of Techmaster. Contact our customer service representative for clarification of this document.
A. Arámburo
Ernesto Matamoros Ing. Mauricio Garayzar Abraham Arámburo
Quality Assurance Technical Manager Certified By
Tijuana, B.C. Calle Seminario No. 8610-11, Col. Niños Héroes Este, CP 22120, Tijuana, B.C. México Tels.: (664) 624-4444 / 624-3128 / 624-4445 / 624-0440
Mexicali, B.C. Blvd. Lázaro Cárdenas 543 F03, Ex ejido Coahuila, C.P. 21360, Mexicali, B.C. Tels.:(686) 555-1660 / 557-6117 / 555-1766
Cd. Juárez,Chih. Ave. Antonio J. Bermudez No 2151-5 Fracc. Parque Ind J.Bermudez, C.P32470 Cd. Juarez Chih. Tels.:(656) 687-2471 / 648-1181
Monterrey, N.L. Ave. Morones Prieto No. 914, Oriente Int. 112, Col. La Huerta, C.P. 67144. Tels.:(81)1334-0701 / 1334-0702
Queretaro, QRO. Ctra. Federal 57,Mex-Qro Lateral Norte Km 210+100,Advance Business Park Conin, int.26 CP.76240,Col. Cumbres de Conin, Tels.:(442)241 51 04
Issued on: 2022-06-27 11:38:24.3830000 -07:00 540.1 2101
Page 1 of 2
TDM-116-0000866 22
22
Calibration Results
Manufacturer: Mitutoyo Asset No. NAN
Step # Function Tested / Nominal Value As Found ± MU* In Tol As Left ± MU* In Tol Calibration Tolerance
1 Surface Deviation N/A N/A
2 Diagonal Line NW-SE 0 µin 20.00 ± 23 YES 20.00 ± 23 YES -150 to 150 µin
3 Diagonal Line NE-SW 0 µin 20.00 ± 23 YES 20.00 ± 23 YES -150 to 150 µin
4 North Perimeter Line 0 µin 0.00 ± 23 YES 0.00 ± 23 YES -150 to 150 µin
5 East Perimeter Line 0 µin 0.00 ± 23 YES 0.00 ± 23 YES -150 to 150 µin
6 West Perimeter Line 0 µin 20.00 ± 23 YES 20.00 ± 23 YES -150 to 150 µin
7 South Perimeter Line 0 µin 20.00 ± 23 YES 20.00 ± 23 YES -150 to 150 µin
8 Center Line E-W 0 µin 30.00 ± 23 YES 30.00 ± 23 YES -150 to 150 µin
9 Center Line N-S 0 µin 0.00 ± 23 YES 0.00 ± 23 YES -150 to 150 µin
*********************************End of Report***********************************
CPM004(A)
540.6 1503
Page 2 of 2
TDM-116-0000866 22
22