You are on page 1of 2

AC-1342

Certificate of Calibration
Manufacturer: Mitutoyo Report #: TDM-116-0000866
Description: Surface Plate P.O. #: N/A

Model #: 517-912-1MEX Procedure: CPM004(A)


Serial #: 12090486 Asset #: NAN

Calibration Lab.: Tijuana, BC


Customer: ALUMINIO DE BAJA CALIFORNIA SA DE CV
Carr. Tij-Ens Km 16.432
Tijuana, BC 22563

This certifies that the above instrument was calibrated in compliance with the Calibration System Requirements of ISO /IEC
17025:2017, ANSI/NCSL Z540-1-1994 in accordance with referenced procedures. Standards used to perform this calibration
are traceable to SI units; their source of traceability derives from a National Metrology Institute such as NIST, CENAM, NPL,
DIN, from natural physical constants, consensus standards or derived by the ratio type of calibrations. Collective
uncertainties are determined as required with a distribution that corresponds to a probability of approximately 95% (k=2).
Unless otherwise noted calibrations are performed to manufacturer's specifications. Compliance statements are in
conformance with ILAC-G8:2019 simple acceptance decision rule. This form shall not be reproduced, except in full, without
the expressed written consent of Techmaster. Contact our customer service representative for clarification of this document.

Test Result: PASSED Calibration Cycle: 12 Months


Temperature: 22ºC Calibration Date: Jun/27/2022
Relative Humidity: 44% Next Due Date: Jun/30/2023
Calibration Location: On-Site Technician: Abraham Arámburo

Standard # Manufacturer Model # Due Date Test Report #


147-0067 Mahr Supramess May 30, 2023 TDM-0-22593341
174-0024 Rahn Granite Repeat-O-Meter Sep 24, 2022 QRO-0-22519744-R1

Special Remarks or Requirements:

A. Arámburo
Ernesto Matamoros Ing. Mauricio Garayzar Abraham Arámburo
Quality Assurance Technical Manager Certified By

Tijuana, B.C. Calle Seminario No. 8610-11, Col. Niños Héroes Este, CP 22120, Tijuana, B.C. México Tels.: (664) 624-4444 / 624-3128 / 624-4445 / 624-0440

Mexicali, B.C. Blvd. Lázaro Cárdenas 543 F03, Ex ejido Coahuila, C.P. 21360, Mexicali, B.C. Tels.:(686) 555-1660 / 557-6117 / 555-1766

Cd. Juárez,Chih. Ave. Antonio J. Bermudez No 2151-5 Fracc. Parque Ind J.Bermudez, C.P32470 Cd. Juarez Chih. Tels.:(656) 687-2471 / 648-1181

Monterrey, N.L. Ave. Morones Prieto No. 914, Oriente Int. 112, Col. La Huerta, C.P. 67144. Tels.:(81)1334-0701 / 1334-0702

Queretaro, QRO. Ctra. Federal 57,Mex-Qro Lateral Norte Km 210+100,Advance Business Park Conin, int.26 CP.76240,Col. Cumbres de Conin, Tels.:(442)241 51 04
Issued on: 2022-06-27 11:38:24.3830000 -07:00 540.1 2101

Page 1 of 2
TDM-116-0000866 22
22
Calibration Results
Manufacturer: Mitutoyo Asset No. NAN

Description: Surface Plate Serial No. 12090486

Model #: 517-912-1MEX Report No. TDM-116-0000866

Step # Function Tested / Nominal Value As Found ± MU* In Tol As Left ± MU* In Tol Calibration Tolerance
1 Surface Deviation N/A N/A
2 Diagonal Line NW-SE 0 µin 20.00 ± 23 YES 20.00 ± 23 YES -150 to 150 µin
3 Diagonal Line NE-SW 0 µin 20.00 ± 23 YES 20.00 ± 23 YES -150 to 150 µin
4 North Perimeter Line 0 µin 0.00 ± 23 YES 0.00 ± 23 YES -150 to 150 µin
5 East Perimeter Line 0 µin 0.00 ± 23 YES 0.00 ± 23 YES -150 to 150 µin
6 West Perimeter Line 0 µin 20.00 ± 23 YES 20.00 ± 23 YES -150 to 150 µin
7 South Perimeter Line 0 µin 20.00 ± 23 YES 20.00 ± 23 YES -150 to 150 µin
8 Center Line E-W 0 µin 30.00 ± 23 YES 30.00 ± 23 YES -150 to 150 µin
9 Center Line N-S 0 µin 0.00 ± 23 YES 0.00 ± 23 YES -150 to 150 µin

*MU stands for Measurement Uncertainty at a 95% confidence level.