You are on page 1of 12

Trends Trends in Analytical Chemistry, Vol.

40, 2012

Seawater-pH measurements for


ocean-acidification observations
Victoire M.C. Rérolle, Cedric F.A. Floquet, Matt C. Mowlem,
Richard R.G.J. Bellerby, Douglas P. Connelly, Eric P. Achterberg

The uptake of anthropogenic CO2 by the oceans since the onset of the industrial revolution is considered a serious challenge to
marine ecosystems due to ensuing carbonate-chemistry changes (ocean acidification). Furthermore, the CO2 uptake is reducing
the oceanÕs capacity to absorb future CO2 emissions. In order to follow the changes in the oceanÕs carbonate system, high-quality
analytical measurements with good spatial and temporal resolution are necessary. High-precision and accurate pH measurements
are now possible, and allow us to determine the progression of ocean acidification. The spectrophotometric pH technique is now
widely used and capable of the required high-quality measurements. Spectrophotometric pH systems are deployed on ships and
in situ on remote platforms. Smaller and more rugged instruments are nevertheless required for more widespread in situ
application to allow routine high-resolution measurements, even in the most remote regions.
We critically review oceanic pH measurements, and focus on state-of-the-art spectrophotometric pH measurement techniques
and instrumentation. We present a simple microfluidic design integrated in a shipboard instrument featuring high accuracy and
precision as a key step towards a targeted pH microsensor system.
ª 2012 Elsevier Ltd. All rights reserved.

Keywords: Carbonate system; Colorimetry; Microfluidic pH sensor; Ocean acidification; Oceanic pH; pH microsensor system; Seawater; Sensor;
Spectrophotometric pH; Spectrophotometry

Victoire M.C. Rérolle, 1. Introduction biodiversity and ecosystem functioning,


Cedric F.A. Floquet, Matt particularly for calcifying organisms [6].
C. Mowlem, Douglas 1.1. Impact of atmospheric CO2 emissions The changes in the marine-carbonate
P. Connelly, The increasing concentration of CO2 in the system also reduce the oceanÕs capacity to
Eric P. Achterberg*
atmosphere as a result of fossil-fuel absorb future atmospheric CO2 emissions
University of Southampton,
National Oceanography Centre, burning and land-use changes has signif- [7]. This decline in the efficiency of the
Southampton, icant impacts on the global climate, and oceanic CO2 sink contributes to accelera-
Southampton SO14 3ZH, UK one of the most noticeable effects is global tion of the rate of increase of atmospheric
warming. However, only 43% of the CO2 CO2 [1].
Richard R.G.J. Bellerby
emitted between 1959 and 2006 is still High-quality carbonate-chemistry
Norwegian Institute for Water
Research,
present in the atmosphere, the remainder measurements are required in order to
Thormøhlensgate 53D, having been taken up by land vegetation understand fully the dynamics of the
N-5006 Bergen, Norway (29%) and the oceans (28%) [1]. Without oceanic carbonate system [8]. The preci-
these carbon sinks, climate change would sion of seawater-pH measurements needs
probably be more pronounced. to be better than 0.002 pH units in order
The accumulation of CO2 in the ocean to allow detection of the average annual
results in a shift in the carbonate-system decrease in surface-ocean pH [3–5]. Sea-
equilibrium, with a reduction in seawater water-pH data with good spatial and
pH [2–5]. Since the onset of the industrial temporal coverage are particularly critical
revolution, average surface-ocean pH has to appreciate ocean-acidification phenom-
decreased by 0.1 pH units [2], and, ena and their consequences.
currently, the rate of decrease is 0.002 pH
units per year [3–5]. This phenomenon is 1.2. Marine carbonate-chemistry
called ocean acidification and is a major variables
*
Corresponding author. topic in oceanographic research. Signifi- Four variables of the marine carbonate-
Tel.: +44 (0)23 80593199; cant shifts in the carbonate system form a chemistry system can be directly deter-
Fax: +44 (0)23 80593059;
serious cause of concern for marine mined in an accurate manner: total
E-mail: eric@noc.soton.ac.uk

146 0165-9936/$ - see front matter ª 2012 Elsevier Ltd. All rights reserved. doi:http://dx.doi.org/10.1016/j.trac.2012.07.016
Trends in Analytical Chemistry, Vol. 40, 2012 Trends

dissolved inorganic carbon (DIC), partial pressure of CO2 whereby the hydrogen-ion concentration is expressed in
(pCO2), total alkalinity (TA) and pH. If two of the car- moles per kilogram (mol/kg) of solution.
bonate-system variables are determined, along with In chemical oceanography, different pH scales have
in situ temperature, salinity and macronutrients, it is been used, with pH buffers for the various pH scales
possible to calculate the other two. With the spectro- made up in artificial seawater. The free pH scale only
photometric pH method, pH can be measured with a takes into account free hydrogen ions [14], whereas the
sufficiently high precision and accuracy to allow com- total pH scale also considers sulfate [15], and the sea-
parison with the other thermodynamic variables that water-pH scale includes sulfate and fluoride complexes
characterize the oceanic carbonate system [9]. Never- [16,17]. The IUPAC NBS pH scale is based on low ionic
theless, inconsistencies have been observed when using strength buffer standards from the US National Bureau
different pairs of variables to determine the carbonate of Standards (NBS) and is universally used in laborato-
system [10]. The sources of these inconsistencies are ries. The NBS and free pH scales are often used in
uncertain, but may be due to inaccuracies in the physiological studies [18].
analytical measurements, inaccuracies in the thermo- The difference between the ionic strengths of the NBS
dynamic constants used for the calculations or presence pH buffers and seawater samples results in significant
of dissolved organic acid/base compounds, which are not changes between calibration and sample measurements,
taken into account in the estimation of alkalinity from and particularly affects the liquid-junction potential
titrations using the Gran method [11]. This highlights when using potentiometric pH systems. This pH scale is
the need to measure more than two carbonate-chemistry hence deemed inappropriate for seawater applications
variables. and should not be used in marine-carbonate chemistry.
The spatial and temporal resolution of carbonate- In oceanic carbonate-chemistry research, the total
system measurements is currently still insufficient, hydrogen-ion scale is most commonly used and is the
despite significant efforts to improve analytical methods, recommended pH scale [19]. The use of the total pH
to develop new instruments and to coordinate interna- scale avoids problems associated with uncertainties in
tional monitoring activities [12]. The deployment of the stability constants for fluoride complexes, which are
appropriate sensors on moorings, drifters, or profiling required for the seawater-pH scale. Importantly, a cer-
floats would enable routine, high-resolution measure- tified pH buffer based on the total pH scale is now
ments in even the most remote regions [13]. There is available from the laboratory of Professor A. Dickson
therefore a growing need for autonomous in situ (Scripps Institution of Oceanography, San Diego, CA,
instruments that measure carbonate chemistry on re- USA). This buffer is made of 2-amino-2-methyl-1,3-
mote platforms. Miniaturized pH systems with high propanediol (Tris) in artificial seawater. There is no
analytical precision and accuracy would be optimal for certified reference material available based on the other
deployment on oceanic moorings and small vehicles to seawater-pH scales.
allow large-scale monitoring of ocean acidification.
Widespread use, particularly on small platforms as part 2.2. Potentiometric method
of a global observing system measuring ocean-carbonate 2.2.1. Glass electrode. The classical potentiometric
chemistry, will only be possible with low-cost sensors method for the determination of pH in seawater consists
featuring low power and resource consumption (e.g., of the measurement of the electromotive force (EMF) of a
reagents). The simplicity and the precision of the spec- cell, most commonly composed of a silver/silver chloride
trophotometric seawater-pH method make it a good electrode and a glass pH electrode. The pH of a sample is
candidate for miniaturization and deployment. defined in terms of EMF measured in the sample (Esample)
This article critically reviews oceanic pH measure- and a standard buffer solution (Ebuffer) of assigned pH
ments, and focuses on state-of-the art spectrophotomet- (pHbuffer):
ric pH techniques and instrumentation. We then present pHsample ¼ pHbuffer þ ðEbuffer  Esample ÞF=ðRTln10Þ ð2Þ
a simple microfluidic design integrated in a shipboard pH
instrument with high accuracy and precision as a key where R is the gas constant, T is the measurement
step towards the targeted in situ pH micro-sensor. temperature and F is the Faraday constant.
The advantages of the potentiometric method are that
2. Analytical techniques for seawater-pH it is non-destructive, the equipment is relatively cheap
measurements and it can achieve a high reading rate (> 1 Hz). Never-
theless it requires regular calibration due to electrode
2.1. Definition of pH and pH scales drift (up to 0.02 pH units per day [20]), making
pH is a measure of the hydrogen-ion concentration in a autonomous measurements problematic. Furthermore,
solution and defined as: potentiometric sensors are susceptible to electromagnetic
interferences and there are various problems with both
pH ¼ log10 ð½Hþ Þ ð1Þ residual liquid-junction potentials and variations in

http://www.elsevier.com/locate/trac 147
Trends Trends in Analytical Chemistry, Vol. 40, 2012

asymmetry potential [21]. As a result of the ubiquitous on the variation of the fluorescence of the indicator dye
experimental problems, potentiometric pH measure- with hydrogen-ion concentration. The most promising
ments rarely achieve an accuracy of better than 0.01 pH approach for seawater-pH measurement uses time-
units [21]. Potentiometric pH sensors are therefore an domain dual-lifetime referencing (t-DLR), which
option when studying pH variations larger than 0.01 pH combines ratiometric and lifetime measurements [25]. It
units (e.g., [22]), but are unsuitable for ocean-acidifi- uses the fluorescence-intensity ratio of two luminescent
cation observations. indicators: one indicator is pH sensitive and has a short
luminescence lifetime, and the other is not pH sensitive
and has a longer lifetime. The advantages of the fluor-
2.2.2. ISFET sensors. A new type of Nernstian sensor
escence method are its high sensitivity and selectivity.
using ion-sensitive field-effect transistors (ISFETs) could
Fluorosensors can also be used for simultaneous detec-
overcome some of the issues encountered with tradi-
tion of several analytes (e.g. oxygen) [26].
tional potentiometric pH measurements, and be appli-
Despite important improvements in analytical lumi-
cable to in situ pH and pCO2 measurements [23,24]. The
nescent techniques, current pH luminescence-based
ISFET is a semiconductor made of p-type silicon coated
sensors are still subject to a range of issues (e.g., cross
with SiO2 without a metal gate electrode over the con-
sensitivity to ionic strength and temperature, fluores-
duction channel. The ion-sensing layer is a thin Si3N4
cence foil bleaching and sensitivity to ambient light). The
coating on the conduction channel. The surface charge
best precision obtained with a pH optode is 0.00064 pH
at the interface of this coating with the solution is con-
units at pH 7.9 in 35& artificial seawater containing
trolled by the solution pH and determines the strength of
10 mM Tris-HCl buffer [27], with the system operational
the electric field in the conduction channel of the FET.
in the range pH 6–9. This pH optode sensor was cali-
The pH is therefore obtained by the measurement of the
brated against a potentiometric pH electrode, which had
voltage between the reference electrode and the sensing
an accuracy of ± 0.02 pH units after calibration with
layer. A Nernstian slope is verified using standard pH
standard buffers (pH 4, 7, and 10, NBS) [27].
buffers in a manner similar to the classical glass-
electrode-based sensor.
The advantage of the ISFET sensor includes reduced 2.4. Spectrophotometric method
drift and noise due to stray currents as a result of the The spectrophotometric pH determination is based on
lower impedance compared to a glass electrode. Martz the addition to a seawater sample of an indicator dye,
and co-workers [24] tested the commercially-available which changes color with pH. The dye is an acid-base
Honeywell Durafet system for pH measurements of sur- compound with a second dissociation constant that
face seawater. They obtained very promising results with should be close to the pH of seawater. The solution pH
a stability better than 0.005 pH units over periods of (i.e. indicator and sample mix) can be expressed as a
weeks to months, and a short-term precision of ±0.0005 function of the acid-dissociation constant
pH units over several hours. These workers showed the (pKind = log10(KHI)) and the concentration of the
great potential of the ISFET-based sensor for in situ protonated (HI) and the unprotonated (I2) forms of
applications at 1 atm. the indicator:
An oceanographic instrument is now commercially pH ¼ pKind þ log10 ð½I2 =½HI Þ ð3Þ
available (SeaFET Ocean pH Sensor, Satlantic LP), and
the ISFET sensors are becoming more commonly used in The acid and base forms of the indicator have different
experimental laboratory studies and on moorings. absorption spectra. This allows the determination of the
However, further work is required on the performance of ratio of concentration of the indicator species by multi-
the reference electrode and the chip packaging for wavelength absorption measurements:
deployment at depth. In addition, the stability of the
½I2  A1 =A2  e1 ðHI Þ=e2 ðHI Þ
sensor during long-duration oceanic deployments needs  ¼ ð4Þ
to be assessed. ½HI  e1 ðI Þ=e2 ðHI Þ  A1 =A2 e2 ðI2 Þ=e2 ðHI Þ
2

In future, an ideal in situ system for long-term


deployment may be formed by a combined ISFET and A1 and A2 refer to the absorbance at wavelengths 1
spectrophotometric sensor system, which provides and 2, and ek(I2) and ek(HI) are the extinction coeffi-
high-frequency ISFET measurements with additional cients at wavelength k of the base (I2) and acid (HI)
lower frequency, high-precision and accurate spectro- forms of the dye, respectively.
photometric analyses. The pH can therefore be calculated from the absor-
bance ratio of the unprotonated and protonated forms
using Equation (5) [28]:
2.3. Luminescence method  
Luminescence-based sensors use optodes with indicator A1 =A2  e1 ðHI Þ=e2 ðHI Þ
pH ¼ pKind þ log10 ð5Þ
dyes fixed in a polymer. The pH fluorosensors are based e1 ð1 Þ=e2 ðHI Þ  A1 =A2 e2 ðI2 Þ=e2 ðHI Þ
2

148 http://www.elsevier.com/locate/trac
Trends in Analytical Chemistry, Vol. 40, 2012 Trends

which can be rewritten as: the reported instrument with a measurement by a


pH ¼ pKind þ log10 ððR  e1 Þ=ðe2  Re3 ÞÞ ð6Þ bench-top spectrophotometer [30,36,38,40–43], the
square root of the sum of the squares of the estimated
  2
with R = A1/A2, e1 = e1(HI )/e2(HI ), e2 = e1(I )/ inaccuracies [29] or the discrepancy between measured
e2(HI) and e3 = e2(I2)/e2(HI)). and certified values established by the analysis of certi-
The extinction coefficients and the pKind are functions fied Tris buffer (underlined values [30]).
of temperature and salinity.
The advantage of the spectrophotometric pH method 3.2. Fluid propulsion
is that it is not prone to drift, as is the case with the Most of the reported pH systems use two pumps: one for
potentiometric method. Because the ratio of the con- the seawater sample and the other for the indicator
centrations of the two indicator forms is used to calculate solution [31,35,37]. Alternatively, one single pump can
pH, the exact total concentration of the indicator is not be used with a system of valves controlling the selection
required to determine the pH of the sample mixed with of liquids [39]. Pumps have also been replaced by a
the indicator. In case the exact indicator concentration is gravity-based fluid-propulsion system [29]. Valve control
needed, this can be obtained from the absorbance mea- simplifies the routing and pumping systems, but may
sured at the isosbestic wavelength (kiso) at which the result in lower accuracy of the pumped volumes and an
extinction coefficients for both protonated and unproto- increase in dead volume in the flow system. A range of
nated forms are equal. Absorbance at the isosbestic pump types has been reported, including solenoid [41],
wavelength depends on the total indicator concentration peristaltic [38], piston [43] and syringe pumps [32]. The
only. The spectrophotometric pH measurements are choice of pump will depend on sample volume, accuracy
therefore inherently calibrated. of the required volume delivery and whether the flow
The pH-indicator dye needs to be characterized for has to be continuous. If continuous flow is required,
dependence on temperature, pressure and ionic strength. peristaltic pumps provide a constant mean flow with a
The characterization should take place in the laboratory superimposed near-sinusoidal variation, and two pumps
prior to instrument deployment. Variations in the (sample and indicator pumps) can readily be synchro-
background absorption of seawater samples, and nized. Syringe pumps provide the best accuracy in
changes to the optical cell, the strength and the spec- volume delivery, and can also be used for mixing the
trum of the light source, and the sensitivity of the sample and indicator dye [32]. Continuous fluid flow is
detector can all lead to drift and inaccuracies in pH possible using syringe pumps with two parallel pumping
measurements. However, these effects are typically small set-ups [44]. A key consideration for in situ pH systems is
(<0.004 pH [29]) and can be mitigated by periodic the size of the pump and its power consumption. Differ-
reference measurements of the light in the absence of the ent types of micro-scale pumps were recently developed
indicator (see sub-section 5.5. Baseline correction). The and allow miniaturization of pH systems [45]. At pres-
precision and long-term reproducibility of spectrophoto- ent, small peristaltic pumps are certainly most appro-
metric pH measurements are potentially better than priate for in situ applications.
0.001 pH units (see Table 1).
3.3. The absorption cell
Most bench-top pH systems use 100-mm path-length
3. Spectrophotometric pH systems cuvettes and relatively low indicator concentrations of
1 · 106–1 · 105 mol/kg, in order to minimize sample-
3.1. pH systems pH perturbations due to indicator addition [34,43,46].
Fig. 1 shows the different components of an automated Systems with shorter path-lengths of 50 mm [29] or
spectrophotometric pH system. The seawater sample and 10 mm [35] have been used successfully but require a
the indicator dye are pumped into the system and mixed higher indicator concentration to achieve a precision
before detection in the absorption cell. The light trans- similar to the 100-mm path-length systems. Increasing
mitted in the absorption cell is recorded by a photo- the indicator concentration requires consideration of the
detector and the data obtained are converted into associated pH perturbations as discussed in Section 5.
absorption values that are used for the pH calculations. The minimum path-length required to obtain a per-
Most automated spectrophotometric pH systems in use turbation smaller than one milli-pH depends on the pH
by researchers are deployed at sea on ships for surface- difference between indicator and seawater sample, and
water measurements [29–36], but, recently, successful on the indicator-to-sample mixing ratio. The latter is
in situ deployments were reported [37–41]. Table 1 determined by the apparent extinction coefficients of the
summarizes some of the automated spectrophotometric pH system, and the concentration of the stock solution of
systems currently used, and we review them in the fol- indicator required to obtain appropriate absorption
lowing paragraphs. The accuracy reported in Table 1 readings for high-quality pH measurements. The flow-
corresponds to the discrepancy of a pH measurement by cells and cuvettes used in pH systems include standard

http://www.elsevier.com/locate/trac 149
Trends Trends in Analytical Chemistry, Vol. 40, 2012

Table 1. Figures of merit of spectrophotometric pH systems. Underlined values correspond to the discrepancy between measured and certified
values for the analysis of certified TRIS buffer. Precision and accuracy are given in pH units

Systems Bellerby et al. 2002 [29] Friis et al. 2004 [43] Wang et al. 2007 [36] Aßman et al. 2011 [30]
A. Automated ship-board pH systems
Precision <0.001 0.0032 0.008 0.0012/0.0007
Accuracy 0.004 not reported 0.0012 ± 0.0042 0.0018/0.0076
Frequency (sample/h) not reported not reported 7 60

Systems SEAS-pH (Liu et al. 2006 [38]) (Nakano et al. 2006 [40]) SAMI-pH (Seidel et al. 2008 [41])
B. Automated in situ pH systems
Application Deployed to 250 m on rosette frame Deployed to 1000 m on rosette frame Moored/ deployed at 4 m deep
Precision 0.0014 0.002 ±0.0007
Accuracy not reported not reported +0.0017
Frequency (sample/h) 1800 360 15

Sea water To waste


pH dye Pump Mixer Absorption cell

Light source Photodetector

Instrument control
Data acquisition

Figure 1. Automated spectrophotometric pH system.

cylindrical glass cells, custom-made polyvinyl chloride samples are readily flushed through the system and
(PVC), polyethylene, polyether ether ketone (PEEK) or cause only temporary disturbances.
other polymeric cells. Some reported systems use liquid-
core waveguides (LCWs) [38], which display excellent 3.4. Mixing indicator with seawater sample
optical properties and can be as long as 2 m. However, The reported seawater-to-indicator mixing ratios vary
high precision can be achieved with 100-mm path- between systems from 10:1 to 1000:1; an optimal sen-
length cells [32], and the standard polymeric cells are sitivity is obtained with absorbance readings in the
more rugged, less expensive and less prone to lining with range 0.4–1 [19]. Mixing can be achieved by active or
bubbles or contaminants than the LCWs. The PEEK flow- passive approaches. Active methods include the use of
cells are also easier to mount and to maintain compared syringe pumps [46], or a magnetic stirrer to mix the
with the LCWs, and do not require the use of surfactants indicator with the sample [29]. Passive methods use a
to minimize bubble problems [38]. Polymeric cells are mixing coil or an in-line static mixer, and mixing is often
therefore attractive for in situ applications. enhanced by injecting the indicator into the sample
The presence of bubbles is a recurrent issue in spec- stream [31,36,38,39,41,43]. Specifically designed mix-
trophotometric flow systems. Bubbles can cause signifi- ing joints enhance the mixing process [34,35], but may
cant interferences in fluid flow, mixing and result in longer flushing times due to dead volumes in
spectrophotometric measurements. Care should be taken corners and joints. The advantage of the passive mixing
to limit potential sources of bubbles in the system (e.g., systems is that they require only a negligible increase in
fluid inlets and connectors). As bubbles often power consumption (due to additional pressure drops
inadvertently occur in flow systems, it is key to design across the mixers). However, homogenization of the
the systems so as to minimize the possibility for bubbles sample-indicator mixture can be a challenge, given the
to get stuck, which often takes place at sharp angles and high mixing ratios required for spectrophotometric pH
rough surfaces. The bubble problems are greatly mini- measurements.
mized when channel dimensions are reduced, internal
surface roughness is reduced, and system integration is 3.5. Light source
increased in order to reduce the number of connections. High intensities of light are required at the absorbance
These features are typically found in lab-on-chip sensors maxima of the indicator. High intensities at the isosbestic
and lead to prevention of bubble formation. In these point and a non-absorbing wavelength will allow cor-
systems, the bubbles acquired from environmental rections for pH perturbations and light variations (see

150 http://www.elsevier.com/locate/trac
Trends in Analytical Chemistry, Vol. 40, 2012 Trends

Section 5. Corrections). The most commonly used light example, Seidel et al. [41] used three silicon photodi-
source is a tungsten-halogen lamp, which has low odes to detect and to discriminate absorption spectra
intensity in the wavelength region of 430 nm, required from a pH indicator. The transmitted light was divided
for all spectrophotometric pH determinations. It is pos- and then filtered to isolate optical power in three nar-
sible to improve the light spectrum with a combination row wavelength bands of interest.
of optical filters [24], but this leads to a more complex Photodiodes are extremely versatile and low cost, and
system. Another approach to improve the spectrum at have been employed in various configurations (e.g., in
low wavelengths involves the use of a deuterium lamp in detectors of flow-injection-analysis systems, separation
addition to the tungsten lamp. Combined halogen and systems and probe photometers) [49]. They are robust,
deuterium light sources are commercially available, but and have a rapid response and wide linear range (3–4
are expensive. Furthermore, halogen lamps are fragile, orders of magnitude better than a phototransistor).
demanding in power and relatively large, so making Simple spectral separation and photodiode detection is
them less suitable for shipboard or in situ applications. therefore attractive for miniaturized in situ devices.
Averaging several spectra constitutes a simple way to
improve the signal-to-noise ratio and mitigates the ef- 3.7. Temperature
fects of low power at the lower wavelength. However, Seawater-pH analysis is very sensitive to temperature
bias in the mean absorbance due to potentially strong variations. Sample temperature must therefore be mea-
absorbance fluctuations can induce error in the final pH sured. pH measurements are often made under carefully-
value. controlled temperature conditions (typically 20–25C)
Light-emitting diodes (LEDs) provide an excellent [43]. Alternatively, pH is measured at in situ tempera-
alternative to the tungsten-halogen lamp. LEDs are ture [40], or simply at measurement temperature (i.e.
widely used in modern spectrometry as the light source without temperature control) [29,31]. The advantage of
because they are cheap, compact, extremely stable, measuring pH at a fixed temperature is that the indicator
typically monochromatic, and require little power and solution needs to be characterized at only one tempera-
have a long life-time with minimal intensity degradation ture. The instrument components (in particular, the
over their life span. The light spectrum of LEDs has a full detector) may also be more stable when operated at a
width at half maximum (FWHM) of typically 25 nm, constant temperature. However, enhanced bubble for-
which is comparable to bandwidths achieved by the mation may occur in the sample stream if cold seawater
combination of wideband sources and interference-type is warmed and this will then require the inclusion of a
optical filters. It is therefore possible to carry out pho- debubbler upstream of the optical cell [43].
tometric measurements without using wavelength- In addition to flow and measurement problems, bubble
discriminating devices [47]. formation may also cause outgassing of CO2, which
would change the sample pH. The thermostatting of the
3.6. The signal detector pH measurements leads to a more complex, expensive
Light absorbances at multiple wavelengths can be system with higher consumption of power, but it also
determined using a scanning spectrophotometer [48], or requires correction of the measured pH back to in situ
a diffractive element (e.g., an optical grating) and a seawater temperatures. The best option is to measure
detector array [e.g., a charge-coupled device (CCD) directly at in situ temperatures and to avoid errors in-
array], as used in CCD spectrophotometers. Double-beam duced by temperature corrections.
spectrophotometers are recommended because of the We should note that, despite the high sensitivity of pH
high sensitivity and precision [32] obtained by the measurements to temperature, the accuracy of the
simultaneous measurement of both source and trans- temperature measurement is not so critical. An error in
mitted light. These devices are expensive and have a the temperature measurement of a sample is partly
high consumption of power, whilst in situ instruments cancelled by the temperature dependency of the indica-
for pH measurements need to be rugged and compact. tor parameters. Indeed, an unrecognized temperature
The photodiode-array spectrophotometer is more error of 0.1C will lead to an error of only about 0.001
appropriate for in situ operations and its rapid acquisi- pH units.
tion of a broad wavelength region allows repeated
spectral measurements, which improve the signal-to-
noise ratio. The limited wavelength resolution (because 4. Sulfonephthalein indicator
of the large pixel size), high dark current and poor
signal-to-noise ratio at low optical powers are the The optimum pH range for an indicator dye is deter-
principal limitations to the portable versions of photo- mined by its pK value, and the extinction-coefficient
diode-array systems. It is possible to perform multiple ratios of the acidic and basic indicator forms [19,50].
wavelength measurements of pH-indicator absorption The pK of the indicator should be comparable to the
spectra without using diffractive spectrometry. For expected pH of the indicator-seawater solution, so that

http://www.elsevier.com/locate/trac 151
Trends Trends in Analytical Chemistry, Vol. 40, 2012

pK-16pH6pK [51]. The pH indicator Thymol Blue (TB) range of conditions, it is advisable to archive the fol-
{pK  8.6 [52]} has been proposed as optimal for the lowing data [46]:
narrow pH range encountered in surface seawaters (1) raw absorbance measurements;
{open-ocean surface-water-pH range 7.95–8.35 [53]}, (2) volume of indicator added;
whereas meta-cresol purple (mCP) {pK  8.0 [46]} has (3) seawater salinity;
been proposed for the pH range encountered in typical (4) temperature of measurements;
surface-to-deep-ocean profiles. (5) indicator concentration;
Table 2 shows the equations for pH calculations (6) absorption ratio;
involving the different indicator dyes. We recommend (7) pH perturbation;
that every new batch of indicator is characterized, (8) algorithms used to calculate R (corrected) and
despite the availability in the literature of the sample pH; and,
indicator-extinction coefficients as a function of tem- (9) quality-control documentation.
perature, salinity [33,46,54–56] and even pressure [57].
This is particularly important in case the indicator has
not been purified to remove impurities [58]. Liu and 5. Corrections
co-workers [41] observed discrepancies between
measurements with purified and unpurified indicator of Several corrections need to be applied to pH values ob-
as much as 0.010 pH units at pH 7.4 and 0.018 pH tained using spectrophotometric techniques.
units at pH 8.2.
The apparent extinction coefficients also depend on
5.1. Baseline correction
the detection system and the measurement procedure, so
Changes in the instrumental response due to optical
they need to be determined for each instrument. The
source fluctuations and movement of optics can be cor-
apparent extinction coefficients are influenced by the
rected using parallel measurements of the light signal on
wavelength resolution of the detector and the spectral
a second (slave or reference) channel without cuvette
slope of the light source within the resolution window.
[43]. However, a second channel is unnecessary if reg-
Published values are generally obtained with a scanning
ular reference-light measurements are made before
double-beam spectrometer using a detector with a rela-
adding the indicator at a higher rate than any periodic
tively narrow slit-width. The use of the published values
fluctuation or variation in the source or optics.
obtained with instruments using larger wavelength
resolutions and light sources with shaped spectra will
lead to inaccuracies in the measurements. In case the 5.2. Transparency correction
indicator cannot be characterized for a particular sys- Light scatter due to the presence of particles in seawater
tem, it is desirable to purify the indicator and thus ensure is detected through intensity measurements at a non-
that the indicator used is accurately characterized by absorbing wavelength of the indicator dye, usually 700–
literature coefficients [41]. Inaccuracies due to source 750 nm. Some systems also use the transparency cor-
spectral slope and finite detector resolution can then be rection to correct for baseline fluctuations, assuming
reduced by calculating their impact through simple that instrument-response fluctuations are the same at
absorption models [59]. every wavelength [60].
The extinction coefficients of the indicator are deter-
mined by absorption measurements in acidic and basic 5.3. Corrections of pH to in situ conditions
solutions, conditions that ensure that only one form of When measurements are conducted at a convenient
the indicator (HI or I2) is present. The coefficients at constant standard temperature and pressure, it is nec-
specific wavelengths are obtained using the Beer-Lam- essary to correct the pH data back to the in situ condi-
bert law from absorption data and known indicator tions, at which the sample was collected. The most
concentrations. reliable way to correct for temperature changes is to use
The acid-formation constant of the indicator is deter- the measured value of pH with another carbonate vari-
mined by absorption measurements in a solution of able {e.g., TA (or DIC) [61,62] to calculate DIC (or TA)}.
known pH (e.g., Tris buffer). Liu and co-workers [55] DIC and TA are not temperature dependent and can be
rearranged Equation (6) in order to reduce the number used to calculate pH at the in situ oceanic temperature
of parameters to be measured in the basic buffer to one using thermodynamic relationships.
(e3/e2) and in the acid buffer to one (e1). This approach If no second carbonate chemistry variable has been
reduces the number of parameters to be determined in determined, TA can also be calculated from salinity data
the indicator-characterization process. using the equations from Bradshaw and co-workers [63]
To allow future corrections of pH data when refined (TA = 660 + 47.6S). Hunter [61] obtained a resulting
indicator characterizations are available for a wider error smaller than 0.0002 pH units using this method.

152 http://www.elsevier.com/locate/trac
Trends in Analytical Chemistry, Vol. 40, 2012
Table 2. Equations for pH calculations (pK and extinction coefficient) reported in literature

Publication Indicator Extinction coefficient ratio pK2 Operating ranges


max max
kHIind kIind- kiso e1 e2 e3 Salinity Temperature
Robert-Baldo Phenol Red 433 558 479.7 0.0038 2.6155 0.1234 4054.8374/T + 116.6037 336S637 06T(C)635
et al. 1985 38.6645log(T) + 0.004(35-S)
log(10.00106S)
Zhang & Thymol Blue 435 596 495 0.00132 + 7.2326  0.0223 + 4.706S/T + 26.3300  306S640 56T(C)635
Byrne 1996 1.6x105T 0.0299717T + 0.0003917T 7.17218log(T)  0.017316
4.6 · 105T2
Mosley Thymol Blue 9.1427  0.3105S1/2 + 0.0783S 06S640 T = 25C
et al. 2004  0.01135S3/2 + 0.000684S2
Mosley meta-Cresol 8.6353 – 0.3238S1/2 + 0.0807S 06S640 T=25C
et al. 2004 Purple  0.01157S3/2 + 0.000694S2
Hopkins Thymol Blue 435 596 0.0035 2.386  0.139 + -log(pK20)  (2.99(±0.02) 104)P 06P640 bar T=25C
et al. 2000 (2.7 106)P (6.6x106)P + (3.3(±0.3) 108)P2
e1 e3/e2 log(K2e2) = a + (b/T) + cln(T)  dT
4
Liu meta-Cresol 434 578 0.007762 + 0.020813 + 2.60262 · 10 T a = 246.64209 + 0.315971S + 206S640 256T(C)635
et al. 2011 Purple 4.5174 · 105T + 1.0436 · 104(S-35) 2.8855 x 104S2;
b = 7229.23864  7.098137S
 0.057034S2;
c = 44.493382  0.052711S;
d = 0.0781344;
http://www.elsevier.com/locate/trac

Trends
153
Trends Trends in Analytical Chemistry, Vol. 40, 2012

5.4. Correction for indicator-induced pH perturbation work (i.e. 1:50 [64] or 1:100 and 1:500 [65]). Fur-
The correction for indicator-induced pH perturbation is thermore, an accurate knowledge of the dilution factors
made by measurements of pH at several indicator con- is required and that can be problematic in miniaturized
centrations. The indicator-concentration range is ob- systems using less accurate pumps.
tained through sequential addition of the indicator to the The pH of the stock solution of indicator should be
sample [29,43,46,56] or a dilution/dispersion curve of adjusted to the expected seawater pH in order to mini-
the indicator aliquot in the sample after passing through mize the indicator-induced perturbations. Sample dilu-
a static mixer [39,41]. A linear relationship is observed tion by the indicator solution could also alter the sample
between pH and the indicator concentration in the pH [39], as it decreases DIC and TA. The stock solution
sample, and the correct sample pH is obtained by linear of indicator should therefore be prepared at a sufficiently
extrapolation to an indicator concentration equal to high concentration (typically 2 mM) in order to mini-
zero. mize sample-dilution effects and to optimize the optical
Similarly, Clayton and Byrne proposed an empirical detection.
approach, where a pair of additions of dye is only made
once per series of seawater samples with different pH
[19,46]. A relationship is obtained for the change in the 6. Precision and accuracy
measured absorbance ratio, DR, between the two addi-
tions, the volume added, V, and the absorbance ratio, R1, 6.1. Precision
measured after the first addition: The precision reported in literature corresponds to the
standard deviations of repeated measurements. For
DR=V ¼ a þ bR1 ð7Þ
ocean-acidification observations, the precision should be
The absorbance ratio R is then corrected using the better than the average annual decrease of 0.002 pH
following equation: units in surface oceans [3–5].
Rcorrected ¼ R  Vða þ bRÞ ð8Þ
6.2. Accuracy
Hunter and Macaskill, and Chiercici and co-authors The accuracy of pH measurements is more difficult to
[64,65] proposed theoretical methods to correct for dye determine, as the pH method determines the absorption
additions but this requires the use of the same seawater- ratio of the acidic and basic forms of the indicator dye
to-indicator dye mixing ratio as the one used in their and thus requires no calibration. The accuracy of the

Figure 2. Microfluidic pH system. The seawater sample (400 lL) and the indicator (5 lL) are pumped by two individual syringe pumps and mixed
in the flow cell using a static mixer prior to reaching the absorption cell. Light from LEDs is transmitted through the absorption cell and measured
by a miniature spectrophotometer. The measurement is made close to in situ temperature (+0.2C) in a sampling chamber that has a continuous
flow of the shipÕs underway seawater supply. Processing and analysis of one sample takes 6 min.

154 http://www.elsevier.com/locate/trac
Trends in Analytical Chemistry, Vol. 40, 2012 Trends

method depends on the accuracy of the indicator pK and dissociation constants chosen and the observed DIC/TA
molar absorptivity values, the accuracy of the tempera- ratios [62].
ture and salinity measurements, and the calibration and
sensitivity of the spectrometers.
The main sources of inaccuracy involve the uncer- 7. Microfluidic pH system
tainty in the pK value of the indicator and the error in
the absorption measurements at the wavelength corre- Current pH systems are unsuitable for widespread
sponding to the absorption of the basic form of the application on in situ platforms, so smaller, more rugged
indicator. Tris buffer is used to determine pK, so the instruments are required. The next-generation pH sys-
uncertainty in the pK value cannot be smaller than the tems will use simple microfluidic designs with low con-
uncertainty in the Tris buffer used. The accuracy of the sumption of power and reagents. Such a pH microsensor
Tris buffer can be as good as 0.002 pH units if prepared featuring high precision and accuracy is being developed
carefully [66]. Certified Tris buffers are currently pro- at the National Oceanography Centre, Southampton,
vided by the laboratory of Prof. Dickson (as above), and UK. The implementation of the spectrophotometric
have an overall uncertainty of 0.004 (Certificate of method on a robust microfluidic flow cell is a key step
Analysis, e.g., Batch 8). towards the targeted microsensor (Fig. 2).
Care should be taken when comparing performance The microfluidic flow cell comprising the absorption
metrics from the literature. Often, the accuracies re- cell and the static mixer is manufactured in poly(methyl
ported are the discrepancies observed between direct methacrylate) (PMMA) [59,67]. Fluidics are controlled
instrument measurements and discrete samples mea- by four micro-inert valves (LFNA1250125H, Lee Prod-
sured with a bench-top spectrophotometer ucts Ltd., UK) directly mounted on the chip, and fluid
[36,38,40,41,43]. For example, inaccuracy due to the propulsion is conducted using two syringe pumps (Har-
use of biased extinction coefficients resulting from vard Apparatus Nanomite, Kent, UK). The 10-mm
impurities in the indicator will not be noted by such a absorption cell has a volume of 5 lL and is connected to
comparison. Accuracy can be assessed by measuring the the light source and detector by two optical fibers
pH of Tris buffer and comparing the results to the the- (600 lm diameter, Thorlabs, USA). The light source is a
oretical pH of Tris calculated with an accuracy of 0.005 tri-colored LED with wavelengths of 435 nm and
pH units from Equation (9) [66]:

pHðTÞ ¼ ð11911:08  18:2499S


 0:039336S2 Þ=T þ ð366:27059
þ 0:53993607S þ 0:00016329S2 Þ
þ ð64:52243  0:084041SÞlog10 ðTÞ
 0:11149858T ð9Þ

where T is the temperature (in Kelvin) and S is the


salinity.
Certified Tris buffer should be used in order to assess
sensor performance. This is critical for comparison of data
obtained by different sensors. However, the use of Tris
buffer does not enable coverage of the pH and tempera-
ture range encountered in the oceans. A combination of
Tris buffer measurements along with an estimation of the
system inaccuracies covering the natural range of tem-
perature and pH would therefore be advisable.
We should note that the Tris buffer is less affected by
pH perturbations due to indicator additions than sea-
water, so it does not allow us to check the adequacy of
the correction method for indicator pH. Comparisons
with pH calculated from other carbonate parameters will
provide further information on the system accuracy,
particularly on its temperature dependence and on the
correction applied to the indicator-dye pH. However, Figure 3. Map of surface pH determined during the first week of
these comparisons depend on the carbonate-system cruise D366 (6–13 June 2011).

http://www.elsevier.com/locate/trac 155
Trends Trends in Analytical Chemistry, Vol. 40, 2012

596 nm corresponding to the absorption maximum of development of a pH microsensor with high accuracy
the TB indicator forms HI and I2, and 750 nm to and precision is therefore possible. The next generation
monitor the sample turbidity. A linear array photodiode of spectrophotometric pH systems will use simple
spectrophotometer (HR4000, Ocean Optics) is used as microfluidic designs coupled with LEDs and photodiodes.
detector. The change in the indicator concentration ob-
served after the indicator aliquot in the sample stream
passed through a static mixer was used to quantify the Acknowledgements
perturbation of the sample pH by the indicator [39,41]. This work is supported by SENSEnet, a Marie Curie Ini-
The performance of the system has been evaluated by tial Training Network (ITN) funded by the European
analyzing certified Tris buffer at 25.02 ± 0.01C. The Commission Seventh Framework Programme, Contract
short-term precision was 0.001 pH unit (n = 20) and a Number PITN-GA-2009-237868. The Discovery D366
discrepancy of 0.0035 pH unit was observed between cruise was funded by the Natural Environment Research
measured and certified values. The long-term precision Council as part of the UK Ocean Acidification Pro-
of the system has not yet been evaluated. gramme (Grant no. NE/H017348/1).
The pH sensor has been deployed on RRS Discovery
cruise D366 as part of the UK Ocean Acidification Re-
References
search Programme. The automated pH system was
[1] J.G. Canadell, C. Le Quéré, M.R. Raupach, C.B. Field, E.T.
operated continuously over a period of a month on an
Buitenhuis, P. Ciais, T.J. Conway, N.P. Gillett, R. Houghton, G.
underway seawater supply. Measurements were only Marland, Proc. Natl. Acad. Sci. USA 104 (2007) 18866.
interrupted for system-performance checks and mainte- [2] K. Caldeira, M.E. Wickett, J. Geophys. Res. 110 (2005) 1.
nance. Fig. 3 shows seawater-surface pH data collected [3] M. González-Dávila, Biogeosciences 7 (2010) 1995.
during the first week of this cruise. [4] J.E. Dore, R. Lukas, D.W. Sadler, M.J. Church, D.M. Karl, Proc.
Natl. Acad. Sci. USA 106 (2009) 12235.
The pH data will be corrected to in situ temperature
[5] R.H. Byrne, S. Mecking, R.A. Feely, X. Liu, Geophys. Res. Lett. 37
using total alkalinity data [61], but this correction will (2010) L02601.
be minor as the difference between measurement and [6] S.C. Doney, Science (Washington, DC) 328 (2010) 1512.
temperature was only +0.2C. The pH observed during [7] C.L. Sabine, R.A. Feely, N. Gruber, R.M. Key, K. Lee, J.L. Bullister,
the first week of the cruise was in the range 7.974– R. Wanninkhof, C.S. Wong, D.W.R. Wallace, B. Tilbrook, Science
(Washington, DC) 305 (2004) 367.
8.118. The highest pH values were observed to the north
[8] R.G.J. Bellerby, G.E. Millward, D.R. Turner, P.J. Worsfold, Trends
west of Ireland, whilst sailing through waters with high Anal. Chem. 12 (1993) ix.
phytoplankton biomass and primary productivity, [9] S.E. Cullison-Gray, M.D. DeGrandpre, T.S. Moore, T.R. Martz, G.E.
resulting in the uptake of DIC. The lowest pH values Friederich, K.S. Johnson, Mar. Chem. 125 (2011) 82.
were observed in the northern Irish Sea as a result of DIC [10] C. Hoppe, G. Langer, S. Rokitta, D. Wolf-Gladrow, B. Rost,
Biogeosciences (2012).
production through remineralization of organic matter.
[11] H. Kim, K. Lee, Geophys. Res. Lett. 36 (2009) L20603.
[12] S.C. Doney, B. Tilbrook, S. Roy, N. Metzl, C. Le Quéré, M. Hood,
R.A. Feely, D. Bakker, Deep-Sea Res., Part II 56 (2009) 504.
8. Conclusions and future direction [13] K.S. Johnson, W.M. Berelson, E.S. Boss, Z. Chase, H. Claustre, S.R.
Emerson, N. Gruber, A. Kortzinger, M.J. Perry, S.C. Riser,
Oceanography 22 (2009) 216.
The spectrophotometric pH technique is capable of the
[14] R. Bates, C. Culberson, in: N.R. Andersen, A. Malahoff (Editors),
high-quality measurements required for ocean-acidifi- The Fate of Fossil Fuel CO2 in the Oceans, Plenum, New York,
cation observations. Spectrophotometric pH systems USA, 1977, p. 45.
with high precision and accuracy have been deployed on [15] I. Hansson, Acta Chem. Scand. 27 (1973) 931.
ships and recently also in situ on moorings. Systems for [16] A. Dickson, F. Millero, Deep-Sea Res., Part A 34 (1987) 1733.
[17] A. Dickson, J. Riley, Mar. Chem. 7 (1979) 89.
pH measurements have improved in recent years and
[18] A. Dickson, in: U. Riebesell, V.J. Fabry, L. Hansson, J.-P. Gattuso
production costs have come down, but they remain (Editors), Guide to Best Practices for Ocean Acidification Research
complex and relatively expensive, particularly the optical and Data Reporting, Publications Office of the European Union,
components. Luxembourg, 2010, p. 17.
The most suitable components for autonomous in situ [19] A.G. Dickson, C.L. Sabine, J.R. Christian, PICES special publication
3 (2007) 1.
instruments are a polymeric cell, a static mixer, micro-
[20] J. Seiter, M. DeGrandpre, Talanta 54 (2001) 99.
pumps, LEDs and photodiodes. Ideally, the absorption [21] A. Dickson, Mar. Chem. 44 (1993) 131.
cell should be 100 mm long in order to minimize indi- [22] N. Le Bris, M. Zbinden, F. Gaill, Deep-Sea Res., Part I 52 (2005)
cator-dye addition and consequent sample-pH pertur- 1071.
bation, but mixing achievable with the passive mixer [23] K. Shitashima, Y. Koike, M. Kyo, A. Hemmi, in: 2008 IEEE
Sensors, IEEE, Piscataway, NJ, USA, p. 1414.
will ultimately determine the required path-length.
[24] T.R. Martz, J.G. Connery, K.S. Johnson, Limnol. Oceanogr.
Readily available lab-on-chip technologies have al- Methods 8 (2010) 172.
lowed the development of robust, low-cost sensors with [25] G. Liebsch, I. Klimant, C. Krause, O.S. Wolfbeis, Anal. Chem. 73
low consumption of power and reagents. The (2001) 4354.

156 http://www.elsevier.com/locate/trac
Trends in Analytical Chemistry, Vol. 40, 2012 Trends

[26] M.I.J. Stich, L.H. Fischer, O.S. Wolfbeis, Chem. Soc. Rev. 39 [47] P.C. Hauser, T.W.T. Rupasinghe, N.E. Cates, Talanta 42 (1995)
(2010) 3102. 605.
[27] M. Larsen, S.M. Borisov, B. Grunwald, I. Klimant, R.N. Glud, [48] S. McElligott, R. Byrne, K. Lee, R. Wanninkhof, F. Millero, R.
Limnol. Oceanogr. Methods 9 (2011) 348. Feely, Mar. Chem. 60 (1998) 63.
[28] G. Robert-Baldo, M. Morris, R. Byrne, Anal. Chem. 57 (1985) [49] M. OÕToole, D. Diamond, Sensors 8 (2008) 2453.
2564. [50] A.G. Dickson, C. Goyet, DOE Handbook of methods for the analysis
[29] R.G.J. Bellerby, A. Olsen, T. Johannessen, P. Croot, Talanta 56 of the various parameters of the carbon dioxide system in sea
(2002) 61. water, ORNL/CDIAC-74, Second Edition, 1994 (http://cdi-
[30] S. Aßmann, C. Frank, A. Körtzinger, Ocean Science 7 (2011) 597. ac.ornl.gov/ftp/cdiac74/0content.pdf).
[31] R. Bellerby, D. Turner, G. Millward, P. Worsfold, Anal. Chim. Acta [51] R.H. Byrne, G. Robert-Baldo, S.W. Thompson, C.T.A. Chen, Deep-
309 (1995) 259. Sea Res., Part A 35 (1988) 1405.
[32] T.A. DelValls, Ciencias Marinas 25 (1999) 345. [52] R.H. Byrne, Anal. Chem. 59 (1987) 1479.
[33] L.M. Mosley, S.L.G. Husheer, K.A. Hunter, Mar. Chem. 91 (2004) [53] R.A. Feely, S.C. Doney, S.R. Cooley, Oceanography 22 (2009) 36.
175. [54] M.D. Gabriel, J.M. Forja, J.A. Rubio, A. Gomez-Parra, Ciencias
[34] S.M. Ohline, M.R. Reid, S.L.G. Husheer, K.I. Currie, K.A. Hunter, Marinas 31 (2005) 309.
Mar. Chem. 107 (2007) 143. [55] X. Liu, M.C. Patsavas, R.H. Byrne, Environ. Sci. Technol. 45
[35] M. Tapp, K. Hunter, K. Currie, B. Mackaskill, Mar. Chem. 72 (2011) 4862.
(2000) 193. [56] H. Zhang, R.H. Byrne, Mar. Chem. 52 (1996) 17.
[36] Z.A. Wang, X. Liu, R.H. Byrne, R. Wanninkhof, R.E. Bernstein, [57] A.E. Hopkins, K.S. Sell, A.L. Soli, R.H. Byrne, Mar. Chem. 71
E.A. Kaltenbacher, J. Patten, Anal. Chim. Acta 596 (2007) 23. (2000) 103.
[37] E. Kaltenbacher, E. Steimle, R. Byrne, in: 2000 IEEE Sensors, IEEE, [58] W. Yao, X. Liu, R.H. Byrne, Mar. Chem. 107 (2007) 167.
Piscataway, NJ, USA, p. 45. [59] C.F.A. Floquet, V.J. Sieben, A. Milani, E.P. Joly, I.R.G. Ogilvie, H.
[38] X. Liu, Z.A. Wang, R.H. Byrne, E.A. Kaltenbacher, R.E. Bernstein, Morgan, M.C. Mowlem, Talanta 84 (2011) 235.
Environ. Sci. Technol. 40 (2006) 5044. [60] S.L.G. Husheer, MasterÕs Thesis in Chemistry, University of Otago,
[39] T.R. Martz, J.J. Carr, C.R. French, M.D. DeGrandpre, Anal. Chem. New Zealand, 2001.
75 (2003) 1850. [61] K.A. Hunter, Deep-Sea Res., Part I 45 (1998) 1919.
[40] Y. Nakano, H. Kimoto, S. Watanabe, K. Harada, Y.W. Watanabe, [62] F.J. Millero, Chem. Rev. 107 (2007) 308.
J. Oceanogr. 62 (2006) 81. [63] A. Bradshaw, P. Brewer, D. Shafer, R. Williams, Earth Planet. Sci.
[41] M.P. Seidel, M.D. DeGrandpre, A.G. Dickson, Mar. Chem. 109 Lett. 55 (1981) 99.
(2008) 28. [64] K.A. Hunter, B. Macaskill, 2nd International Symposium CO2 in
[42] S. Aßmann, C. Frank, A. Kötzinger, Ocean Sci. Discuss. 8 (2011) the, Oceans (1999) 471.
1367. [65] M. Chierici, A. Fransson, L.G. Anderson, Mar. Chem. 65 (1999)
[43] K. Friis, A. Körtzinger, D.W.R. Wallace, Limnol. Oceanogr. 281.
Methods 2 (2004) 136. [66] T.A. DelValls, A.G. Dickson, Deep-Sea Res., Part I 45 (1998)
[44] I. Ogilvie, V. Sieben, M. Mowlem, H. Morgan, Anal. Chem. 83 1541.
(2011) 4814. [67] I. Ogilvie, V. Sieben, C. Floquet, R. Zmijan, M. Mowlem, H.
[45] D. Laser, J. Santiago, J. Micromech. Microeng. 14 (2004) R35. Morgan, J. Micromech. Microeng. 20 (2010) 065016.
[46] T. Clayton, R. Byrne, Deep-Sea Res., Part 1 40 (1993) 2115.

http://www.elsevier.com/locate/trac 157

You might also like