Professional Documents
Culture Documents
Jkps 60 1150
Jkps 60 1150
1150∼1156
Estimates of the Image Quality and the Radiation Dose for Head and
Abdomen Phantom Image Acquisition by Using Dual-energy CT
Pil-Hyun Jeon
Department of Radiological Science and Institute of Health Science,
College of Health Science, Yonsei University, Wonju 220-710, Korea and
Department of Radiology, Yonsei University Wonju Christian Hospital, Wonju 220-701, Korea
Using dual-energy computed tomography (CT) scans, we obtained images from a raw data set
by using low- and high-energy scans (usually 80 and 140 kV, respectively). Generally, the head and
the abdomen examinations were performed using single-energy (120 kV) scanning with a routine
exposure protocol, and the images were used for diagnostic interpretation. The dual-energy CT
scans can also be used for diagnosis by using the linearly-mixed method with low- and high-energy
images. In the current study, we evaluated mixed images in dual-energy and single-energy scans for
image quality and radiation dose. The CT scan protocol for single energy was adopted from IEC
protocols, and the recommended dose from the EC. The dual-energy scan protocol was based on the
Siemens dual-energy CT scan protocol. The CT scan protocols were extended further in the mA
range for both scans to estimate the image quality corresponding to these dose alternations. The
results demonstrate that the Contrast-to-noise ratio (CNR) value of a dual-energy mixed image
is higher than that of a single-energy image in similar dose regions for both the head and the
abdomen phantoms. We observed that the dual-energy images could reduce the dose compared to
single-energy images. In the abdomen phantom study, the CNR of dual-energy images was even
higher than that of single-energy images with half the radiation dose of the single-energy scan.
Therefore, the dual-energy CT scan can accomplish a remarkable dose reduction while preserving
image quality for head and abdomen imaging.
in single-energy CT scans [9]. Generally, head and ab- Table 1. Diagnostic reference levels for CT examination of
domen examinations have been performed with single- adults (European Commission, 1999).
energy CT scans by using routine scan protocols. There-
fore, both single-energy and dual-energy mixed images Diagnostic reference level
Examination
have been used for diagnostic interpretations. As the CTDIw (mGy) DLP (mGy cm)
dual-energy scan includes low- and high-energy images, Head 60 1050
the image quality and the dose dependence of the phan- Sinuses 35 360
tom size need to be compared to those in single-energy Chest 30 650
images. Furthermore, dual-energy images must optimize Abdomen 35 780
the image quality within recommended dose limitations.
Pelvis 25 520
Hence, the purpose of the current study was to estimate
the image quality of head and abdomen phantoms with
single-energy and dual-energy mixed methods. The re-
sults from single- and dual-energy CT scans were com-
pared under routine clinical exposure conditions. The
scan protocol for tube current such as mA was alter-
nated within the recommended limited dose levels while
the other scan protocols were fixed in the single- and
the dual-energy scans. The dose ranges were extended
within the recommended dose levels for the head and the
abdomen.
II. MATERIAL AND METHODS Fig. 1. A PMMA phantom for CTDIw . The diameters
of the PMMA phantoms are 16 cm for the head and 32 cm
1. Data Acquisition for the abdomen. The ion chamber was inserted into the
hole at the center, 3, 6, 9, and 12 o’clock positions, and the
CTDI100 was calculated from the measured charges on the
The image quality and the radiation dose of single- ion chamber. CTDIw was obtained with weighting from the
energy and dual-energy scans were evaluated for head equation for CTDI100 .
and abdomen phantoms. We utilized the X-ray computed
tomography (CT) modality (Philips, Brilliance CT 64,
Eindhoven, Netherlands), which was originally designed The single-energy scan image was used as a refer-
to perform single-energy scans with exposure modes at ence for image quality and radiation dose for compar-
80, 120, and 140 kV. Typically, single-energy scans (usu- ison to dual-energy scans. We used X-ray tube voltages
ally at 120 kV) are performed for patient diagnostic in- of 80 and 140 kV to obtain results similar to those of
terpretations. Dual-energy scans (generally at 80 and dual-energy CT scans. The single-energy scans for the
140 kV) can also generate a single set of images for diag- head and the abdomen phantoms were performed with
nostic interpretation with linearly-mixed equations from scan protocols based on the International Electrotech-
two different energy images, termed the linearly-mixed nical Commission (IEC). The X-ray tube voltage was
image. The equation reported by Yu et al. for obtaining 120 kV with a rotation time of 1 s. The current was only
the linearly-mixed image is as follows [9]: varied up to the levels recommended for CT radiation
doses for both head and abdomen imaging. The diag-
I = wL IL + wH IH , (1) nostic reference levels for the CT radiation doses for the
head and the abdomen were 60 and 35 mGy, respectively,
where IL and IH denote the low- and high-energy scan
given as CTDIw in Table 1, referring to the European
images intensity,respectively, wL and wH are the weight-
Commission (EC) [10]. The image quality was estimated
ing factors for the low- and high-energy scan images, and
using the Philips system phantom (Philips, Sys. phan-
wL + wH = 1. wL is given by
tom kit, Eindhoven, Netherlands). Dual-energy scans
2 2 were also performed to assess the radiation dose and im-
CL (σH,s + σH,b )
wL = 2 2 ) + C (σ 2 + σ 2 ) , (2) age quality.
CL (σH,s + σH,b H L,s L,b
cmcm
cmcm
Fig. 6. The CNR values for single-energy (SE) scans using
Fig. 4. The CNR values of dual-energy scans were higher
abdomen phantoms were 10.62% higher than those of dual-
than those of single-energy scans in similar dose regions when
energy (DE) scans at 17.32 vs. 8.36 at one-half the radiation
using abdomen phantoms.
dose of the SE scans. With increasing radiation dose, the
CNR values of the DE scan increased from 2.51% to 13.42%.
cmcm
Fig. 5. Comparisons of CNR values in head phantoms cmcm
between single-energy (SE) and dual-energy (DE) scans at
one-half the radiation dose of SE scans. The CNR values Fig. 7. The dose evaluations at similar dose points for
were 16.27, 13.69, and 8.18% higher for head phantoms when single-energy (SE) and dual-energy (DE) scans in head phan-
using single-energy scans. As a point of comparison between toms. The radiation dose was higher for the SE scans at every
the 32.85- (single-energy) and the 16.05-mGy (dual-energy) point. The point of 126.82 vs. 127.42 was chosen to estimate
scans, the CNR of the single-energy scan was 11.89% higher the effects in clinical radiation exposure conditions. The dose
whereas the CNR values for dual-energy scans were 2.45, 4.29, was reduced up to 40.86% when the DE scan was performed
and 11.84% higher. at that point.
the CNR for single-energy and dual-energy scans were and 129.26 at 32.85 mGy, 47.98 mGy, and 56.25 mGy, re-
115.50 and 133.45 in similar dose regions for the head spectively, and the CNR values for the dual-energy scans
phantom, respectively. The minimum values of CNR were 92.48, 108.70, and 119.49 at 16.05 mGy, 23.92 mGy,
for single-energy and dual-energy scans were 107.53 and and 27.80 mGy, respectively. In the abdomen image eval-
119.49, respectively. In the abdomen phantom analy- uation, the CNR values for the single-energy scans were
sis, the maximum values of the CNR single-energy and 41.32, 41.78, 44.80, and 44.47 at 17.32 mGy, 20.65 mGy,
dual-energy scans were 41.83 and 54.58 in similar dose 25.67 mGy, and 28.85 mGy, respectively. The CNR val-
regions, respectively. The minimum values of the CNR ues for the dual-energy scans were 36.93, 42.83, 46.81,
values for single-energy and dual-energy scans were 39.73 and 50.44 at 8.36 mGy, 10.48 mGy, 12.67 mGy, and 14.41
and 50.44, respectively. mGy, respectively.
One-half of the doses for single-energy scans were cho- Figures 7 and 8 show the dose evaluations between
sen for dual-energy scans to validate dose reduction, as single- and dual-energy scans at similar CNRs. The doses
described in Figs. 5 and 6. In the single-energy scans for between the single- and the dual-energy scans were eval-
the head phantom, the CNR values were 107.53, 123.58, uated when the image qualities were similar. With the
-1154- Journal of the Korean Physical Society, Vol. 60, No. 7, April 2012
[3] Y. Watanabe, K. Uotani, T. Nakazawa, M. Higashi, N. [9] L. Yu, A. N. Primak, X. Liu and C. H. McCollough, Med.
Yamada, Y. Hori, S. Kanzaki, T. Fukuda, T. Itoh and Phys. 36, 1019 (2009).
H. Naito, Eur. Radiol. 19, 1019 (2009). [10] European Commission: European Guidelines on Quality
[4] L.-J. Zhang, Y.-E Zhao, S.-Y. Wu, B. M. Yeh, C.-S. Criteria for Computed Tomography (EUR 16262 EN).
Zhou, X.-B. Hu, Q.-J. Hu and G.-M. Lu, Radiology 252, European Commission, Luxembourg, 1999.
61 (2009). [11] M. J. Siegel, B. Schmidt, D. Bradley, C. Suess and C.
[5] T. Tsunoo, M. Torikoshi, M. Sasaki, M. Endo, N. Yagi Hildebolt, Radiology 233, 515 (2004).
and K. Uesugi, IEEE Trans. Nucl. Sci. 5, 1678 (2003). [12] G. Jarry, J. J. DeMarco, U. Beifuss, C. H. Cagnon and
[6] M. Torikoshi, T. Tsunoo, M. Sasaki, M. Endo, Y. Noda, M. F. McNitt-Gray, Phys. Med. Biol. 48, 2645 (2003).
Y. Ohno, T. Kohno, K. Hyodo, K. Uesugi and N. Yagi, [13] J. J. DeMarco, C. H. Cagnon, D. D. Cody, D. M. Stevens,
Phys. Med. Biol. 48, 673 (2003). C. H. McCollough, J. O’Daniel and M. F. McNitt-Gray,
[7] M. Bazalova, J.-F. Carrier, L. Beaulieu and F. Verhae- Phys. Med. Biol. 50, 3989 (2005).
gen, Phys. Med. Biol. 53, 2439 (2008). [14] J. A. Christner, J. M. Kofler and C. H. McCollough, Am.
[8] M. M. Goodsitt, E. G. Christodoulou and S. C. Larson, J. Roentgenol. 194, 886 (2010).
Med. Phys. 38, 2222 (2011).