You are on page 1of 7

Journal of the Korean Physical Society, Vol. 60, No. 7, April 2012, pp.

1150∼1156

Estimates of the Image Quality and the Radiation Dose for Head and
Abdomen Phantom Image Acquisition by Using Dual-energy CT

Dae-Hong Kim, Hee-Joung Kim∗ and Chang-Lae Lee


Department of Radiological Science and Institute of Health Science,
College of Health Science, Yonsei University, Wonju 220-710, Korea

Pil-Hyun Jeon
Department of Radiological Science and Institute of Health Science,
College of Health Science, Yonsei University, Wonju 220-710, Korea and
Department of Radiology, Yonsei University Wonju Christian Hospital, Wonju 220-701, Korea

Won-Hyung Lee and Sung-Soo Jeon


Department of Radiology, Yonsei University Wonju Christian Hospital, Wonju 220-701, Korea

(Received 21 November 2011, in final form 9 February 2012)

Using dual-energy computed tomography (CT) scans, we obtained images from a raw data set
by using low- and high-energy scans (usually 80 and 140 kV, respectively). Generally, the head and
the abdomen examinations were performed using single-energy (120 kV) scanning with a routine
exposure protocol, and the images were used for diagnostic interpretation. The dual-energy CT
scans can also be used for diagnosis by using the linearly-mixed method with low- and high-energy
images. In the current study, we evaluated mixed images in dual-energy and single-energy scans for
image quality and radiation dose. The CT scan protocol for single energy was adopted from IEC
protocols, and the recommended dose from the EC. The dual-energy scan protocol was based on the
Siemens dual-energy CT scan protocol. The CT scan protocols were extended further in the mA
range for both scans to estimate the image quality corresponding to these dose alternations. The
results demonstrate that the Contrast-to-noise ratio (CNR) value of a dual-energy mixed image
is higher than that of a single-energy image in similar dose regions for both the head and the
abdomen phantoms. We observed that the dual-energy images could reduce the dose compared to
single-energy images. In the abdomen phantom study, the CNR of dual-energy images was even
higher than that of single-energy images with half the radiation dose of the single-energy scan.
Therefore, the dual-energy CT scan can accomplish a remarkable dose reduction while preserving
image quality for head and abdomen imaging.

PACS numbers: 87.59.Fm


Keywords: Linearly-mixed image, CTDIw, Contrast-to-noise ratios (CNR), Dual-energy CT
DOI: 10.3938/jkps.60.1150

I. INTRODUCTION be divided for projection-based and image-based ap-


proaches. A synthesized monochromatic image is used in
the projection-based approach, which can reduce beam-
Dual-energy computed tomography (CT) is commonly hardening artifacts in the CT image. The projection-
used to obtain information such as tissue segmenta- based approach is also able to produce a material se-
tion [1], plaque segmentation [2], angiography [3], and lective image. The image-based approach is involved in
lung perfusion [4]. Prior studies using dual-energy CT bone removal, effective atomic number and electron den-
involved measurements of both the effective atomic sity imaging, and linearly-mixed imaging for diagnos-
numbers and electron densities [5–7], and synthesized tic interpretations. Thus, dual-energy CT is widely used
monochromatic imaging [8]. In dual-energy CT imag- for a variety of purposes. Additionally, dual-energy CT
ing, the projection data require two different data sets images are also used in diagnostic interpretation with
having both low- and high-energy. These data sets can equations to generate linearly-mixed images. Yu et al.
described the optimized dose fraction for dual-energy
∗ E-mail: hjk1@yonsei.ac.kr; Tel:+82-33-760-2475; Fax: +82-33- CT scans when the total dose was the same as that
760-2562
-1150-
Estimates of the Image Quality and the Radiation Dose for Head · · · – Dae-Hong Kim et al. -1151-

in single-energy CT scans [9]. Generally, head and ab- Table 1. Diagnostic reference levels for CT examination of
domen examinations have been performed with single- adults (European Commission, 1999).
energy CT scans by using routine scan protocols. There-
fore, both single-energy and dual-energy mixed images Diagnostic reference level
Examination
have been used for diagnostic interpretations. As the CTDIw (mGy) DLP (mGy cm)
dual-energy scan includes low- and high-energy images, Head 60 1050
the image quality and the dose dependence of the phan- Sinuses 35 360
tom size need to be compared to those in single-energy Chest 30 650
images. Furthermore, dual-energy images must optimize Abdomen 35 780
the image quality within recommended dose limitations.
Pelvis 25 520
Hence, the purpose of the current study was to estimate
the image quality of head and abdomen phantoms with
single-energy and dual-energy mixed methods. The re-
sults from single- and dual-energy CT scans were com-
pared under routine clinical exposure conditions. The
scan protocol for tube current such as mA was alter-
nated within the recommended limited dose levels while
the other scan protocols were fixed in the single- and
the dual-energy scans. The dose ranges were extended
within the recommended dose levels for the head and the
abdomen.

II. MATERIAL AND METHODS Fig. 1. A PMMA phantom for CTDIw . The diameters
of the PMMA phantoms are 16 cm for the head and 32 cm
1. Data Acquisition for the abdomen. The ion chamber was inserted into the
hole at the center, 3, 6, 9, and 12 o’clock positions, and the
CTDI100 was calculated from the measured charges on the
The image quality and the radiation dose of single- ion chamber. CTDIw was obtained with weighting from the
energy and dual-energy scans were evaluated for head equation for CTDI100 .
and abdomen phantoms. We utilized the X-ray computed
tomography (CT) modality (Philips, Brilliance CT 64,
Eindhoven, Netherlands), which was originally designed The single-energy scan image was used as a refer-
to perform single-energy scans with exposure modes at ence for image quality and radiation dose for compar-
80, 120, and 140 kV. Typically, single-energy scans (usu- ison to dual-energy scans. We used X-ray tube voltages
ally at 120 kV) are performed for patient diagnostic in- of 80 and 140 kV to obtain results similar to those of
terpretations. Dual-energy scans (generally at 80 and dual-energy CT scans. The single-energy scans for the
140 kV) can also generate a single set of images for diag- head and the abdomen phantoms were performed with
nostic interpretation with linearly-mixed equations from scan protocols based on the International Electrotech-
two different energy images, termed the linearly-mixed nical Commission (IEC). The X-ray tube voltage was
image. The equation reported by Yu et al. for obtaining 120 kV with a rotation time of 1 s. The current was only
the linearly-mixed image is as follows [9]: varied up to the levels recommended for CT radiation
doses for both head and abdomen imaging. The diag-
I = wL IL + wH IH , (1) nostic reference levels for the CT radiation doses for the
head and the abdomen were 60 and 35 mGy, respectively,
where IL and IH denote the low- and high-energy scan
given as CTDIw in Table 1, referring to the European
images intensity,respectively, wL and wH are the weight-
Commission (EC) [10]. The image quality was estimated
ing factors for the low- and high-energy scan images, and
using the Philips system phantom (Philips, Sys. phan-
wL + wH = 1. wL is given by
tom kit, Eindhoven, Netherlands). Dual-energy scans
2 2 were also performed to assess the radiation dose and im-
CL (σH,s + σH,b )
wL = 2 2 ) + C (σ 2 + σ 2 ) , (2) age quality.
CL (σH,s + σH,b H L,s L,b

where CL and CH are the contrasts between the sig-


2. CTDIw
nal and the noise for low- and high-energy scans, σL,s
and σL,b are the noises expressed as standard deviations
at the signal and the background images in low-energy Prior to radiation exposure to image the quality phan-
scans, and σH,s and σH,b are the noises at the signal and tom, the CTDIw values of a single-energy scan and a
the background in high-energy scans, respectively. dual-energy scan were measured for a head (16 cm di-
-1152- Journal of the Korean Physical Society, Vol. 60, No. 7, April 2012

ameter) and an abdomen (32 cm diameter) Polymethyl


methacrylate (PMMA) phantom. The CTDIw can be
obtained from the CDTI100 . The CTDI100 was measured
using both head and abdomen phantoms with nine holes
and a pencil ion chamber for the CT (Radical Corpora-
tion, 20X6-3CT, CA, USA). The values of CTDI100 were
calculated with the following equation:
M × Chamberlength (cm) × F × tp × cf
CTDI100 = , (3)
n×T
where M is the measured value in Roentgen, F is
the exposure-to-dose conversion factor (0.78 cGy/R in
acryl), tp is the correction factor for both temperature Fig. 2. The image quality phantoms used in this study.
and pressure, cf is the chamber calibration factor, n is The diameters of the head and the abdomen phantoms are
the number of slices, and T is the slice thickness (cm). 20 cm and 30 cm, respectively. The circles on both phantoms
The dose measurements were estimated at the center and represent the signal, and the rectangles are noise.
four peripheral locations (3 o’clock, 6 o’clock, 9 o’clock,
and 12 o’clock), as shown in Fig. 1. The CTDIw can
be obtained by using a weighted CTDI100 from the cen-
ter and the peripheral locations and the following the
equation:
1 2
CTDIw = CTDIC + CTDIP , (4)
3 3
where CTDIC is the CTDI100 measured in the center,
and CTDIP is the mean of the CTDI100 values measured
at the 3 o’clock, 6 o’clock, 9 o’clock, and 12 o’clock po-
sitions.

3. Image Evaluation cmcm


Fig. 3. The CNR evaluations of single-energy (SE) and
The image quality metric for medical images is di- dual-energy (DE) scans were performed using the head phan-
vided into quantitative methods and qualitative meth- tom in similar dose regions. The DE scan outperformed at
ods. We chose the image metric as the contrast-to-noise CNR values in similar CTDIw regions.
ratio (CNR) for the quantitative method for the head
and the abdomen phantoms. The CNR is calculated as
III. RESULTS
|Ss − Sb |
CNR =  , (5)
σs2 + σb2 1. Estimates of CNR and CTDIw in Single- and
Dual-energy Scans
where Ss and Sb are the signal intensities of the signal
and background, respectively, and σs and σb are the noise
values for the signal and background in Region of interest In single scans for the head phantom, the CTDIw
(ROI). ranged from 29.75 to 58.75 mGy. The CTDIw of 58.75
The head and abdomen phantom kits were provided by mGy was within the EC recommendation for the head
the Philips CT system. The phantom was constructed in region. The maximum CNR value for the head phantom
two sections: one with a diameter of 20 cm for the head was 142.28, and the minimum was 107.53. In abdomen
section consisting of water, polyethylene, Nylon, Lexan, phantom imaging, the CTDIw ranged from 15.55 to
acrylic, and Teflon materials, and one with a diameter 31.94 mGy, the maximum CNR was 49.34, and the mini-
of 30 cm for the abdomen section composed of water, mum was 39.73. An evaluation of image quality and radi-
Nylon, and Teflon materials. The phantom was used to ation dose between single-energy and dual-energy imag-
carry out a quality-control analysis of the Philips CT ing was performed for both head and abdomen phantom
modality. In head phantom imaging, the signal and the imaging.
noise are set as Teflon and water, respectively, as shown Figures 3 and 4 give the CNR values for single- and
in Fig. 2, while the signal and the noise for abdomen dual-energy imaging of the head and the abdomen phan-
phantom imaging are Teflon and Nylon, respectively. toms for similar dose regions. The maximum values of
Estimates of the Image Quality and the Radiation Dose for Head · · · – Dae-Hong Kim et al. -1153-

cmcm
cmcm
Fig. 6. The CNR values for single-energy (SE) scans using
Fig. 4. The CNR values of dual-energy scans were higher
abdomen phantoms were 10.62% higher than those of dual-
than those of single-energy scans in similar dose regions when
energy (DE) scans at 17.32 vs. 8.36 at one-half the radiation
using abdomen phantoms.
dose of the SE scans. With increasing radiation dose, the
CNR values of the DE scan increased from 2.51% to 13.42%.

cmcm
Fig. 5. Comparisons of CNR values in head phantoms cmcm
between single-energy (SE) and dual-energy (DE) scans at
one-half the radiation dose of SE scans. The CNR values Fig. 7. The dose evaluations at similar dose points for
were 16.27, 13.69, and 8.18% higher for head phantoms when single-energy (SE) and dual-energy (DE) scans in head phan-
using single-energy scans. As a point of comparison between toms. The radiation dose was higher for the SE scans at every
the 32.85- (single-energy) and the 16.05-mGy (dual-energy) point. The point of 126.82 vs. 127.42 was chosen to estimate
scans, the CNR of the single-energy scan was 11.89% higher the effects in clinical radiation exposure conditions. The dose
whereas the CNR values for dual-energy scans were 2.45, 4.29, was reduced up to 40.86% when the DE scan was performed
and 11.84% higher. at that point.

the CNR for single-energy and dual-energy scans were and 129.26 at 32.85 mGy, 47.98 mGy, and 56.25 mGy, re-
115.50 and 133.45 in similar dose regions for the head spectively, and the CNR values for the dual-energy scans
phantom, respectively. The minimum values of CNR were 92.48, 108.70, and 119.49 at 16.05 mGy, 23.92 mGy,
for single-energy and dual-energy scans were 107.53 and and 27.80 mGy, respectively. In the abdomen image eval-
119.49, respectively. In the abdomen phantom analy- uation, the CNR values for the single-energy scans were
sis, the maximum values of the CNR single-energy and 41.32, 41.78, 44.80, and 44.47 at 17.32 mGy, 20.65 mGy,
dual-energy scans were 41.83 and 54.58 in similar dose 25.67 mGy, and 28.85 mGy, respectively. The CNR val-
regions, respectively. The minimum values of the CNR ues for the dual-energy scans were 36.93, 42.83, 46.81,
values for single-energy and dual-energy scans were 39.73 and 50.44 at 8.36 mGy, 10.48 mGy, 12.67 mGy, and 14.41
and 50.44, respectively. mGy, respectively.
One-half of the doses for single-energy scans were cho- Figures 7 and 8 show the dose evaluations between
sen for dual-energy scans to validate dose reduction, as single- and dual-energy scans at similar CNRs. The doses
described in Figs. 5 and 6. In the single-energy scans for between the single- and the dual-energy scans were eval-
the head phantom, the CNR values were 107.53, 123.58, uated when the image qualities were similar. With the
-1154- Journal of the Korean Physical Society, Vol. 60, No. 7, April 2012

Yu et al. evaluated the image quality of linearly-mixed


images when the total radiation dose was constant and
the dose fraction was varied in dual-energy CT imaging
[9]. This study evaluated the optimal weighting factors
for generating linearly mixed images. In this study, CNR
values heavily depended on the phantom size. However,
the CNR values and the noise levels were altered accord-
ing to changes in weighting factors for the linear combi-
nation at the same phantom size, except for 70% of dose
partitioning at 80 kV. Therefore, we did not consider op-
timal weighting factors when the maximum CNRs were
calculated, and the mixed images were produced by us-
ing the weighting factors in Eqs. (1) and (2).
cmcm As seen in Figs. 3 and 4, the CNR values for the
dual-energy scans were higher than these for the single-
Fig. 8. The radiation dose required to obtain similar CNRs energy scans for both head and abdomen phantoms in
is increased for single-energy (SE) scans at every data point.
similar dose regions. The CNR improvement of dual-
The point of 41.83 vs. 42.83 was used to evaluate the dose and
the image quality for clinical exposure conditions. The dose
energy images was remarkable, with maxima of 18.50
was reduced up to 45.90% when a dual-energy (DE) scan was and 30.48% for the head and the abdomen phantoms,
used at that point. respectively, relative to single-energy images. These re-
sults contributed to the dose reduction observed using
dual-energy CT.
head phantom, the doses in the single-energy scans were When comparing the dose reduction effect, Figs. 5
32.85, 38.70, and 54.19 mGy at CNR values of 107.53, and 6 demonstrate that about one-half the dose could
120.98, and 126.82, respectively. The doses in the dual- be used for the head and the abdomen regions when
energy scans were 23.92, 27.80, and 32.05 mGy at CNR dual-energy scans. The single-energy scan was outper-
values of 108.70, 119.49, and 127.42, respectively. For formed in terms of CNR values for head phantoms at ev-
the abdomen phantoms, the doses were 19.37, 29.39, and ery dose comparison. The CNR values were 16.27, 13.69,
31.94 mGy at CNR values of 41.83, 47.59, and 49.34 for and 8.18% higher for head phantoms when using single-
single-energy scans, and the dose for dual-energy scans energy scans. The CNR results for the head phantom in-
were 10.48, 12.67, and 14.41 mGy at CNR values of dicate that the dose reduction effect was approximately
42.83, 46.81, and 50.44, respectively. 50% for similar CNR values. The CNR difference be-
tween single- and dual-energy scans decreased when the
current was increased. Relative to the single-energy scan,
the dual-energy scan demonstrated higher CNR values
2. Estimates of CNR and CTDIw in Clinical than the single-energy scans, except for one point for
Radiation Exposure Conditions
the abdomen phantoms at one-half the dose. As a point
of comparison between the 32.85- (single-energy) and the
In single-energy scans, the routine radiation exposure 16.05-mGy (dual-energy) scans, the CNR of the single-
condition for the current on the head ranged from ap- energy scan was 11.89% higher whereas the CNR values
proximately 300 to 360 mAs at Yonsei University Wonju for dual-energy scans were 2.45, 4.29, and 11.84% higher.
Christian Hospital. The CNR value was 126.82 for this For the abdomen study, the CNR values for single-energy
range of currents for single-energy scans. The CNR of (SE) scans using abdomen phantoms were 10.62 % higher
127.42 in dual-energy CT scans is comparable to the than those of dual-energy (DE) scans at 17.32 vs. 8.36
CNR value for the single-energy scans for head phan- at one-half the radiation dose of the SE scans. With in-
toms; the associated doses for single- and dual-energy creasing radiation dose, the CNR values of the DE scan
scans were 54.19 and 32.05 mGy, respectively. The cur- increased from 2.51% to 13.42%. Thus, a dual-energy
rent for the abdomen imaging ranged from 250 to 300 scan of the abdomen can be used with one-half of the
mAs. The CNR values for single- and dual-energy scans radiation dose for a similar single-energy scan.
within this range were 41.83 and 42.83, and the radiation In Figs. 7 and 8, the dose evaluations at similar CNR
doses were 19.37 and 10.48 mGy, respectively. points are presented using single- and dual-energies for
head and abdomen phantoms. Greater dose increases
were required for single energy CTs to obtain CNRs sim-
ilar to the dual-energy CTs for both head and abdomen
IV. DISCUSSION phantoms. These results also included the dose reduction
aspect of dual-energy CT.
The effects of single- and dual-energy scans were eval- We estimated the image quality and dose between
uated for head and abdomen phantoms. In a prior study, single- and dual-energy scans in routine exposure condi-
Estimates of the Image Quality and the Radiation Dose for Head · · · – Dae-Hong Kim et al. -1155-

single-energy scans. We performed image quality analy-


sis using quantitative methods, taking into account the
contrasts and noise levels in the signal and the back-
ground images. The quantitative method, expressed as
the CNR, has a value for both single- and dual-energy
scans. However, it is necessary to carry out qualitative
evaluations, such as Receiver Operating Characteristic
(ROC), to determine the correlation between the quanti-
tative and the qualitative method, as the CNR values are
a relative metric in single-energy and dual-energy scans.
In the current study, the CNR values for the head phan-
tom were twice those for the abdomen for both single-
Fig. 9. (Color online) The X-ray spectra for single- and and dual-energy scans. A study by Siegel et al. improved
dual-energy scans used in this study. The total X-ray pho- the correlations of the radiation dose and image quality
ton fluences were 84,252,513 and 35,767,322 for single- and to the effects of phantom size [11]. In the study, noise
dual-energy scans, respectively. The X-ray photons of the increased at tube voltages of 80, 100, 120, and 140 kV
single-energy scan are approximately 2.35 times higher than when the phantom size was increased and iodine con-
those of the dual-energy scans. This implies that relative to trast decreased with increasing phantom size. The iodine
the typical single-energy method, the dual-energy method is contrast/noise function rapidly decreased with increas-
efficient for producing mixed CT images without increasing ing phantom size. In the current study, the CNR of the
the dose.
head was considerably higher than that of the abdomen
phantom due to our use of Eq. (2) to obtain a linearly-
mixed image. The dose measurements were performed
tions seen in clinical practice. Figures 7 and 8 show the using CTDI in this work because the measurements of
evaluation for two scans in which the dose was reduced organ dose were difficult for actual patient dose. There-
up to 40.86 and 45.90% for the head and the abdomen fore, previous studies included Monte-Carlo simulation
phantoms, respectively. method for the organ dose [12,13]. Especially, Christner
The trends of dual-energy CT scans for the current et al. estimated the organ dose between single- and dual-
study demonstrate that the dual-energy technique can energy CT imaging using Dose-Length Product (DLP)
reduce the radiation dose and maintain the image quality with both experimental and estimating methods. The
compared to single-energy scans because the dual-energy results showed that contained the effective dose using
scans (80 and 140 kV) used suitable energy spectra for dual-energy CT were reduced or similar to that using
generating dual-energy images characterized by the at- single-energy CT technique [14].
tenuation coefficients of the relevant materials. Dual- The current study suggests that dual-energy scans can
energy methods acquire their separated spectra while the contribute to a reduced radiation dose while maintain-
tube voltage of 120 kV generates X-ray spectra across ing image quality by using the only current alternation.
the energy band according to the attenuation coefficient The dual-energy scan achieved a higher CNR with one-
of a material. The mean energy of a spectrum was de- half the radiation dose of a single-energy scan of the ab-
fined as 1/3 of the maximum energy. Figure 9 describes domen.
the X-ray spectra for single-energy scans used in clini-
cal exposure conditions for abdomen phantoms, and the
spectra for dual-energy scans were added to analyze the
mean energy and photon numbers. The mean energies of ACKNOWLEDGMENTS
the dual-energy spectra were 64.0 and 97.6 keV at 80 and
140 kV, respectively, and the mean energy for 120 kV was This work was supported by the Basic Atomic Energy
87.6 keV as measured with a SRS-78 spectrum measure- Research Institute (BAERI) of the National Research
ment program. As seen in Fig. 9, the spectra at 140 kV Foundation of Korea (KRF) funded by the Ministry of
exhibited a higher mean energy than that at 120 kV. Education, Science & Technology (MEST, grand code:
This phenomenon could be differentiated from the low- 2011-0006368).
kV energy. The total numbers of X-ray photons were
84,252,513 and 35,767,322 for single- and dual-energy
scans, respectively. The number of X-ray photons for
a single-energy scan is approximately 2.35 times greater REFERENCES
than those for a dual-energy scan. From these data, the
dual-energy method is efficient for producing mixed CT [1] M. Bazalova, J.-F. Carrier, L. Beaulieu and F. Verhae-
images without increased dose compared to the typical gen, Radiother. Oncol. 86, 93 (2008).
single-energy method. Consequently, the energy separa- [2] D. T. Boll, E. M. Merkle, E. K. Paulson, R. A Mirza and
tion for dual-energy scans is more effective than it is for T. R. Fleiter, Radiology 249, 119 (2008).
-1156- Journal of the Korean Physical Society, Vol. 60, No. 7, April 2012

[3] Y. Watanabe, K. Uotani, T. Nakazawa, M. Higashi, N. [9] L. Yu, A. N. Primak, X. Liu and C. H. McCollough, Med.
Yamada, Y. Hori, S. Kanzaki, T. Fukuda, T. Itoh and Phys. 36, 1019 (2009).
H. Naito, Eur. Radiol. 19, 1019 (2009). [10] European Commission: European Guidelines on Quality
[4] L.-J. Zhang, Y.-E Zhao, S.-Y. Wu, B. M. Yeh, C.-S. Criteria for Computed Tomography (EUR 16262 EN).
Zhou, X.-B. Hu, Q.-J. Hu and G.-M. Lu, Radiology 252, European Commission, Luxembourg, 1999.
61 (2009). [11] M. J. Siegel, B. Schmidt, D. Bradley, C. Suess and C.
[5] T. Tsunoo, M. Torikoshi, M. Sasaki, M. Endo, N. Yagi Hildebolt, Radiology 233, 515 (2004).
and K. Uesugi, IEEE Trans. Nucl. Sci. 5, 1678 (2003). [12] G. Jarry, J. J. DeMarco, U. Beifuss, C. H. Cagnon and
[6] M. Torikoshi, T. Tsunoo, M. Sasaki, M. Endo, Y. Noda, M. F. McNitt-Gray, Phys. Med. Biol. 48, 2645 (2003).
Y. Ohno, T. Kohno, K. Hyodo, K. Uesugi and N. Yagi, [13] J. J. DeMarco, C. H. Cagnon, D. D. Cody, D. M. Stevens,
Phys. Med. Biol. 48, 673 (2003). C. H. McCollough, J. O’Daniel and M. F. McNitt-Gray,
[7] M. Bazalova, J.-F. Carrier, L. Beaulieu and F. Verhae- Phys. Med. Biol. 50, 3989 (2005).
gen, Phys. Med. Biol. 53, 2439 (2008). [14] J. A. Christner, J. M. Kofler and C. H. McCollough, Am.
[8] M. M. Goodsitt, E. G. Christodoulou and S. C. Larson, J. Roentgenol. 194, 886 (2010).
Med. Phys. 38, 2222 (2011).

You might also like