SWEN KONFORMITY – ISO 17025:2017 QUALITY MANUAL

ISSUE NO. 05 REV NO. 00


L-01 LISTS LIST OF TESTS CARRIED OUT BY LABORATORY ISSUE DATE: REV. DATE: -
01.01.2019

L-01 LIST OF TESTS CARRIED OUT BY LABORATORY


Specification,
Specific tests /
Equipment Used standard (method) or Range of testing/
Sl. types of tests
for Testing technique used and referring Limit of detection
Performed
collateral standards
Ground Bond Test
( On-Site and at Permanent 3Aac-33Aac
1 Hipot Tester
facility) 10mΩ-650mΩ

Insulation Resistance
( On-Site and at Permanent 50V – 1kVdc
2 Hipot Tester
facility) 1MΩ - 9999 MΩ

Di-electric Withstand Test


A.C Withstand Test
D.C Withstand Test Vdc: 0.100 kV – 6 kVdc
3 Power frequency Test Hipot Tester Vac: 0.100 kV – 5kVac
( On-Site and at Permanent 3.00 A – 33.00 A
facility) BS EN 60204 – 1 : 2006 + A1: 2009,
BS EN 61010 - 1 : 2010
Temperature Rise Test IEC 60601 – 1 : 2012 Edition 3.1,
Infrared
( On-Site and at Permanent IEC 61439-1 Ed. 2.0 B:2011 IR:- -18 °C to +450 °C
4 Thermometer&
facility) IEC 60335 - 1 : 2013 : Edition 5.1, Data Logger:- 0 - 100°C
Temp. Data Logger
EN 60950 - 1 : 2006 + A2 : 2013.
Leakage current Hi- IEC 62040-1:2013 DC / AC :50 μA/500 μA/5
5 Leakage Current Test IEC 62109-1:2010
tester mA/50 mA
IS 302-1:2008
IS 1356-1:1972 Temperature : (-10 to
IS 13450-1:2008 150) °C,
IS 13252-1:2010 Humidity : (10 to 96)
6 Humidity Pre-conditioning Humidity Chamber ISO 11197:2016 %RH
Size : 500 mm * 600 mm
* 620 mm (L*B*H)

Petroleum Spirit
No Equipment Isopropyl Alcohol
7 Durability Of Marking
Required Methylated Spirit

Lp : 30 ~ 130 dB (A)
8 Sound Level Measurement dB Meter 30 ~ 130 dB (F)
35 ~ 130 dB (C)
Digital Vernier
9 Creepage & Clearance 0 – 150 mm
Calliper
Impact Test
10 Impact Test 500 g ± 25 gm Ball
Apparatus
BS EN 61010 - 1 : 2010,
EN 60950 - 1 : 2006 + A2 : 2013
IEC 60335 - 1 : 2013 : Edition 5.1
Ball Pressure
11 Ball Pressure Test IEC 60601 – 1 : 2012 Edition 3.1 500 g
Apparatus
IS 302-1:2008
IS 13252-1:2010
IS 13450-1:2008
BS EN 61010 - 1 : 2010,
EN 60950 - 1 : 2006 + A2 : 2013
IEC 62040-1:2013
Glow Wire Test IEC 60335 - 1 : 2013 : Edition 5.1 (1-960)°C
12 Glow Wire Test
Apparatus IEC 60601 – 1 : 2012 Edition 3.1, 10N
IEC 61439-1:2011
IEC 62109-1:2010
IS 302-1:2008

Issued by: Quality Assurance Manager


Page
1 of 3
Approved by: Managing Partner
SWEN KONFORMITY – ISO 17025:2017 QUALITY MANUAL
ISSUE NO. 05 REV NO. 00
L-01 LISTS LIST OF TESTS CARRIED OUT BY LABORATORY ISSUE DATE: REV. DATE: -
01.01.2019

IS 13252-1:2010
BS EN 61010 - 1 : 2010,
φ 12mm
Accessibility Test with Accessibility Probes EN 60950 - 1 : 2006 + A2 : 2013
13 Length : 80mm ; Force :
Jointed Finger Probe & Force Gauge IEC 60601 – 1 : 2012 Edition 3.1
10 N
IEC 60335 - 1 : 2013 : Edition 5.1
BS EN 61010 - 1 : 2010,
φ 1.0 mm
Accessibility Test with Accessibility Probes EN 60950 - 1 : 2006 + A2 : 2013
14 Length : 100mm ; Force :
Test Pin Probe & Force Gauge IEC 60601 – 1 : 2012 Edition 3.1,
1N
BS EN 61010 - 1 : 2010,
Accessibility Test with Accessibility Probes EN 60950 - 1 : 2006 + A2 : 2013
15 Force : 50 N
Rigid Finger Probe & Force Gauge IEC 60601 – 1 : 2012 Edition 3.1

φ 2.5 mm
Accessibility Test with Accessibility Probes
16 IEC 60601 – 1 : 2012 Edition 3.1 Length : 100mm ; Force :
Test Rod & Force Gauge
3N
Accessibility Test with Accessibility Probes
17 IEC 60601 – 1 : 2012 Edition 3.1 Force : 50 N
Test Hook & Force Gauge
BS EN 60204 – 1 : 2006 + A1: 2009
IEC 62040-1:2013
IEC 60601-1:2005+A1:2012
IEC 60335-1:2010+A1 2013+A2
2016
Angle Measurement
18 Angle Measurement BS EN 61010 - 1 : 2010 0-90°
Instrument
EN 60950 - 1 : 2006 + A2 : 2013
IS 302-1:2008
IS 13252-1:2010
IS 13450-1:2008
IS 13450-1:2008
BS EN 61010-1,Ed 3 : 2010
EN 60950-1 : 2006 + A1 : 2006 +
Hardwood
A2 : 2013
50mm Thick
IEC 60601-1 : 2012 Edition 3.1
19 Drop test
IEC 62109-1:2010
Density :> 800 Kg/m3
IS 13252-1:2010
IS 13450-1:2008
ISO 11197:2016
Cross Hatch Cutter
20 Paint Adhesive Test ASTM D3359 -
Kit
Salt Spray Corrosion ASTM B117
21 Corrosion Test -
Test Chamber IEC 60068-2-11 ED 3 : 1981
EN 60950-
1:2006+A1:2006+A2:2013
22 Steady Force Test Force Gauge 1N to 50N
IEC 62040-1:2013
IS 13252-1:2010
EN 60950-
1:2006+A1:2006+A2:2013
IEC 62040-1:2013
Temperature IEC 60601-1:2005+A1:2012
23 Mould Stress relief Test (-10 to 150)°C
chamber IS 13450-1:2008
ISO 11197:2016
IEC 62109-1:2010
IEC 62109-1:2010
Cylindrical pin 4mm
IEC 62109-1:2010 Force : 1N – 50 N/
24 Access probe Test Conical pin 3mm
IEC 62109-2:2011 0V - 50 Vac
Rod 2.5mm
IEC 60601-1:2005+A1:2012
0-600vac
25 Power Input Digital Multimeter IEC 62109-1
0-10Aac
IEC 62109-2

Issued by: Quality Assurance Manager


Page
2 of 3
Approved by: Managing Partner
SWEN KONFORMITY – ISO 17025:2017 QUALITY MANUAL
ISSUE NO. 05 REV NO. 00
L-01 LISTS LIST OF TESTS CARRIED OUT BY LABORATORY ISSUE DATE: REV. DATE: -
01.01.2019

26 Push Test Standard CI weight ISO 11197 Up to 281 Kg


Needle Flame Test
27 Needle Flame Test IEC 60695-11-5 -
Apparatus
IEC 61439-1,
IEC 61010-1,
28 Impulse Withstand Test Impulse Tester ( 0 to 12) kV
IEC 60335-1,
IEC 60730 etc.
IEC 60601-1,
29 Weight/ Load Application Test Standard CI weight Up to 281 Kg
IEC 60730-1 etc.
Digital Torque IEC 60335-1,
30 Torque Test ( 0.3 to 12) N/m
Wrench IEC 61010-1
31 Tensile safety Factor Test Standard CI weight IEC 60601-1 and collateral standards Up to 281 Kg

Issued by: Quality Assurance Manager