H50 QC and Calibration SOP v1 1
H50 QC and Calibration SOP v1 1
℃~ ℃ ~
*If the room temperature changes more than 10°C during the working day, the analyzer must be calibrated.
Login the system with account name: ADMIN, password: 123456. Go to SERVICE>>SETTINGS>>LAB PARAMETERS >> CALIBRATION
FACTORS and the values shown here is old factors and it should be
changed to new factors.
14 ©2018 EDAN Confidential
Calibrate Methods – Manual calibration
Apply the equation below to calculate the new HGB calibration factor.
Where A= Fresh blood test result on standard machine, B= Fresh blood test result on H50.
Obtain the new HGB calibration factor and then replace the previous HGB calibration factor with the new HGB calibration factor.
1. Click CHANGE
to activate all these
2. Input a valid
changes.
date
www.edan.com.cn