Professional Documents
Culture Documents
net/publication/268821995
CITATION READS
1 2,715
2 authors:
Some of the authors of this publication are also working on these related projects:
Digital Image Processing and Analysis of Complex Materials using Micro-CT View project
All content following this page was uploaded by Geoff Dougherty on 02 February 2018.
Radiography
journal homepage: www.elsevier.com/locate/radi
a r t i c l e i n f o a b s t r a c t
Article history: Purpose: Energy resolution is one of the major limitations of gamma camera performance, mainly
Received 2 August 2014 affecting image contrast and resolution. There is a need for a simple method of measuring gamma
Received in revised form camera energy resolution, which is practical for technology students as well as for routine quality
29 September 2014
control.
Accepted 17 October 2014
Available online 2 November 2014
Materials and methods: A 37 MBq (1 mCi) MBq 99Tcm point source was prepared and positioned 1.5 m
away from the gamma camera. Eleven static images were acquired with the same acquisition time (60 s),
using 1 keV windows at intervals of 5 keV from 115 to 165 keV. The counts for each image were recorded
Keywords:
Energy resolution
and plotted graphically (counts vs. energy). Gaussian fitting was used to estimate the full-width-at-half-
Imaging physics laboratory maximum height (FWHM), and the energy resolution (FWHM%) was calculated as a percentage of the
Dual-head gamma camera photopeak (~140 keV).
Image quality Results: The FWHM% of the energy peak was measured to be ~7%. Most values for the energy resolution
Gaussian fitting (FWHM%) of our system were significantly lower (i.e., higher energy resolution) than the commissioning
measurement, and were comparable to the recent preventive maintenance values. There were no sig-
nificant energy resolution differences between the two detectors of the dual head gamma camera.
Conclusions: This simple method for the evaluation of the energy resolution of a gamma camera system
can be easily implemented within routine quality assurance. The measure will help prevent image
deterioration through early detection of serious energy resolution issues which can be resolved in
routine gamma camera corrective maintenance.
© 2014 The College of Radiographers. Published by Elsevier Ltd. All rights reserved.
http://dx.doi.org/10.1016/j.radi.2014.10.007
1078-8174/© 2014 The College of Radiographers. Published by Elsevier Ltd. All rights reserved.
Table 1
Counts acquired at different energy windows using a99Tcm point source (The resulting FWHM and energy resolution, FWHM%, from Gaussian fitting is also tabulated).
Energy window Set#1 Counts/minute Set#2 Counts/minute Set#3 Counts/minute Set#4 Counts/minute Set#5 Counts/minute Average
Table 2 Table 3
Counts acquired from the two detectors of a dual-head imaging system at different Counts acquired at different energy windows using a 131I point source.
energy windows using a99Tcm point source. (The resulting FWHM and energy res- (The resulting FWHM and energy resolution, FWHM%, from Gaussian
olution, FWHM%, from Gaussian fitting is also tabulated). fitting is also tabulated).
Discussion
photomultiplier(s) failure or malfunction, crystal defects, physical
The ability of the scintillation camera to distinguish between separation of the photomultiplier-light guide assembly from the
photons of different energies is best expressed by energy resolu- crystal, and high count rate.3 Changes in gamma camera sensitivity
tion. It is conventionally quantified as the percentage ratio of the could indicate incorrect energy calibration of the PHA or impaired
full width at half-maximum height, FWHM, of the primary radia- energy resolution.3 An immediate evaluation of the gamma camera
tion to the photopeak energy, FWHM%.3 In normal operation the energy resolution is recommended in such an eventuality to
photopeak profile and its peak energy are not readily accessed by determine which is the case. Such an evaluation can be confirmed
the operator. In this work, we have demonstrated how they may be in an effortless and expeditious way.
obtained simply. In order to improve the overall energy resolution, gamma
In the camera system it is possible to discriminate against cameras have electronic methods of energy correction that align
scattered photons using pulse height analysis. The window setting the photopeaks produced by each of the photomultiplier tubes. This
of the pulse height analyzer (PHA) determines the amount of is usually done by energy correction tables which are periodically
scattering registered in the image; a narrower window setting (annually) recollected over the life of the camera, ensuring opera-
dramatically eliminates scattered photons, and consequently im- tional improvement.3 Our straightforward method of determining
proves image quality in terms of contrast and (spatial) resolution.1 the energy resolution can indicate whether a manufacturer service
The energy resolution of the detector is the most important feature or update of the proper energy correction is warranted.
that determines the efficiency with which this can be accom- Energy resolution is dependent on detector crystal thickness.
plished. With a smaller detector energy resolution (FWHM%), it is Better energy resolution can be obtained at thin crystal size, (3/8”
possible to set a narrower window setting and achieve better image (9.5 mm) compared to 5/8” (15.8 mm)). Energy resolution is more
quality. For example, with a FWHM% of ~10% the window can be set variable among different gamma cameras using a thick crystal.
to 130e150 keV (~15%), which is a common setting for clinical However, energy resolution should not be variable in multiple
studies. detectors within the same imaging system.11 In this experiment we
Many factors degrade the gamma camera energy resolution, found the energy resolution of all our departmental gamma cam-
such as poor matching of the gains of the photomultipliers, eras using 99Tcm to be similar within NEMA specifications (FWHM
% < 10%).
Figure 2. Gaussian fitting of the counts acquired at the two detectors (detector #1 in
black; detector #2 in red) of a dual-head imaging system versus energy window. (For Figure 3. Counts acquired at one detector versus energy window (black), and fitted to
99 m
Tc point source). a Gaussian (red). (For 131I point source).
Since the number of photoelectrons released from the photo- resolutions of the gamma cameras in the department were
cathode increases with increasing photon energy, energy resolu- confirmed to be 10% FWHM.
tion should become better with increasing gamma-ray energy.
However, most of the functions of a gamma camera are optimized Conflict of interest
for 99Tcm. This explains why we obtained the same energy resolu-
tion for 131I (364 keV). (This was also within NEMA specifications). None.
The increased gain used at higher energy is probably the main
reason for the elevated center value obtained for 131I, viz. 370 keV
References
rather than the expected 364 keV, although the arbitrariness of the
positioning of the energy windows will also contribute to a dif- 1. Simon RC, Sorenson JA, Phelps ME. The gamma camera: performance charac-
ference between the center value obtained and the expected value. teristics [chapter 14]. In: Physics in nuclear medicine. Philadelphia: Elsevier/
The latter reason also holds for the 99Tcm results. Saunders; 2012.
2. IAEA-TECDOC-602. Quality control of nuclear medicine instruments 1991.
Although dual- or triple-head gamma cameras present more Vienna: International Atomic Energy Agency; 1991.
quality control concerns, similar performance characteristics 3. IAEA-TECDOC-xxxx. Quality control of nuclear medicine instruments 2001.
among detectors within the same multi-head gamma camera Vienna: International Atomic Energy Agency; 2000.
4. NEMA Standards Publication NU1-2007. Performance measurements of gamma
would be reassuring for qualitative and quantitative imaging pro-
cameras. Rosslyn, VA: National Electrical Manufacturers Association; 2007.
cedures.10,11,16 For the dual-head gamma cameras tested here, the 5. Hines H, Kayayan R, Colsher J, Hashimoto D, Schubert R, Fernando J, et al.
energy resolution was the same for both detectors. The energy Recommendations for implementing SPECT instrumentation quality control.
Nuclear medicine SectioneNational Electrical Manufacturers Association
resolution measurement method which had been suggested pre-
(NEMA). Eur J Nucl Med 1999 May;26(5):527e32.
viously by Elliot10 did not result in any experimental data, so that 6. Groch M, Erwin WD. Single-photon emission computed tomography in the
we were unable to compare the method with ours. year 2001: instrumentation and quality control. J Nucl Med Technol 2001;29:
Continuous gamma camera quality control tests are essential. 9e15.
7. Bolster A, editor. IPEM Report 86, Quality control of gamma camera systems.
Proper testing can determine whether performance is acceptable. York: Institute of Physics and Engineering in Medicine; 2003.
Measurement of energy resolution at commissioning, and during 8. Rova A, Celler A, Hamarneh G. Development of NEMA-based software for
preventive maintenance and essential quality control times must gamma camera quality control. J Digit Imaging 2008;21(2):243e55.
9. AAPM Report No. 6, Scintillation camera acceptance testing and performance
be performed. This measurement is a task for the medical engineer evaluation. New York: The American Association of Physicists in Medicine;
and/or medical physicist, although a professional technologist 1980.
should also be familiar with the methodology and take turns on 10. Elliott AT. Quality assurance. In: Sharp PF, Gemmell HG, Murray AD, editors.
Practical nuclear medicine. Springer; 2005. p. 78 [chapter 6].
such a task. In certain countries the status of professional tech- 11. Ballani NS, Sukkar I. Medical imaging physics teaching to radiologic technol-
nologist would require energy resolution measurement to be part ogists in Kuwait. Radiography 2005;11(1):67e70.
of his/her role.11 12. Grosev D, Loncari c S, Vandenberghe S, Dodig D. Triple-head gamma camera
PET: system overview and performance characteristics. Nucl Med Commun
2002 Aug;23(8):809e14.
13. American Society of Nuclear Cardiology. Updated imaging guidelines for nu-
Conclusion clear cardiology procedures, part 1. J Nucl Cardiol 2001;8(1):G5e58.
14. Kappadath SC, Erwin WD, Wendt 3rd RE. Observed inter-camera variability of
clinically relevant performance characteristics for Siemens Symbia gamma
Experimentally, we have demonstrated a simple method and cameras. J Appl Clin Med Phys 2006;7(4):74e80.
procedures for measuring the energy resolution of a gamma cam- 15. Bugby SL, Lees JL, Bhatia BS, Perkins AC. Characterisation of a high resolution
era. It gives technologists, both students and practitioners, the small field of view portable gamma camera. Phys Medica 2014;30:331e9.
16. Blagosklonov O, Sabbah A, Verdenet J, Baud M, Cardot JC. Poststress motionlike
opportunity to optimize the gamma camera detector for clinical use artifacts caused by the use of a dual-head gamma camera for (201)Tl
before imaging human subjects. Using this method, the energy myocardial SPECT. J Nucl Med 2002;43(3):285e91.