Professional Documents
Culture Documents
Technology Behind
Diagnostic Reagent Strips
92 L A B O R ATO RY M E D I C I N E VO L U M E 3 1 , N U M B E R 2 F E B R U A RY 2 0 0 0
Instrumentation
Instrumental reading arose from the need for
accuracy and data documentation. It eliminated
the subjectivity of visual interpretation and Glucose
increased convenience. The earliest instrument
was the Ames Reflectance Meter (ARM; Miles
Bilirubin
Laboratories), introduced in 1966 for measuring
blood glucose concentration. During the next 30
years a series of instruments was developed to read Ketone
dry reagent strips. These instruments translated
the reagent colors into numerical output by mea- Urobilinogen
Scientific Communications
Reflolux, Boehringer Mannheim, Mannheim,
Germany).
Dry reagent strip technology involves the con- Fig 1. Multiple panels
trolled application of reagents in the manufacture with 4 reagent areas (Fig 2). The first area at the tip of dry reagents are
of the reagent pads. These reagents contain indica- of the strip is for sample application and overlaps used for urinalysis.
tor dyes, metals, enzymes, polymers, antibodies, the next reagent area, providing for transfer of the Each reagent detects
and various other chemicals dried onto carriers.3 patient sample (urine) to the third reagent area. a specific analyte
through the use of
Carriers commonly used are papers, membranes, The treated sample then migrates across a third colorimetric
or polymers with various sample uptake and layer, where reactants are immobilized for color reactions.
transporting properties. development. This migration is driven by a fourth
4
area that takes up excess specimen.
Section
Reagent Strip Design The chromatography reaction takes place in the
In the 1980s, reagent strip design began to include third area, called the test or capture zone, typically
multiple reagent layers to measure 1 analyte. This a nitrocellulose membrane. In the first and second
allowed chemical reagent systems to be placed in areas, an analyte-specific antibody reacts with the
distinct reagent layers and provided for reaction analyte in the specimen and is chromatographi-
separation steps such as chromatography and fil- cally transferred to the nitrocellulose membrane.
tration. Immunochromatography strips are con- The antibody is bound to colored latex particles as
structed so that chemical reactions occur in a label.4 If the sample contains the analyte, it reacts
distinct areas of reagents. with the labeled antibody. In the capture zone a
The Clinitek hCG strip test (Bayer) for human second antibody is immobilized in a band and
chorionic gonadotropin is a dry reagent strip test
F E B R U A RY 2 0 0 0 VO L U M E 3 1 , N U M B E R 2 L A B O R ATO RY M E D I C I N E 93
Fig 2. The immuno-
chromatography strip
has 4 distinct areas:
sample application
area; reagent area; Antibody-latex
capture zone; and conjugate
absorbent area.
Analyte
Test band
Control band
captures particles when analyte is present. A col- dyes can be covalently bound to the analyte (dia-
ored test line is formed (Fig 3). A second band of zonium compounds binding bilirubin) or tightly
reagent is also immobilized in the capture zone to associated with the analyte (sodium-sensing
allow a control line to react with particles, forming indicators). Enzymatic reactions can be used to
color. Color at the control line is always formed detect enzymes at micromolar levels through
when the test system is working properly, even in reactions with color-forming substrates. Enzy-
the absence of hCG in the patient sample (Fig 3). matic reactions can also be used to detect mole-
Whole blood glucose strips often use multiple cules, such as glucose, through reactions with
reagents to trap intact RBCs that interfere with the enzymes to yield colored end products. Particle-
color generation area. For example, the Encore labeled antibodies are the primary reagents that
glucometer (Bayer) uses a trapping area placed provide for the detectable reaction of immuno-
directly over the color-generating area. The color logic strips based on chromatography. Redox
is read from the bottom of the strip through a catalysis uses metal chelates to oxidize or reduce
Test Your transparent window. Other designs allow the sam- indicators in the presence of specific analytes
Knowledge! ple to migrate to a color-generating area aside such as hemoglobin, and can detect molecules
Look for the CE from the trapping area, and color is read from the down to the nanomolar level.
Update exam on
top of the strip. Whole blood test strips often
Instrumentation (002)
in the April issue of
make use of plastic cassettes to hold the reaction Reagent Strip Performance
Laboratory Medicine. areas in place. Reagent strip technology has been used to provide
Participants will earn Multiple layers of reagent have also been either threshold or quantitative results for many
4 CMLE credit hours. applied to film slides, such as the reagent system analytes. Both applications are well described for
used with the Ektachem analyzer (Vitros) devel- many clinical uses. Threshold tests are applied to
oped by Eastman Kodak in 1980. Slides enabled analytes with well-established clinical decision
multiple separating, spreading, and color-forming limits, based on large outcomes studies (eg, albu-
areas to enhance colors. minuria, hCG), and have well-defined differences
Typical chemical reactions occurring in dry between positive and negative test results. Quanti-
reagent strips can be grouped as dye binding, tative tests are more often applied to analytes that
enzymatic, immunologic, and redox catalysis.3 require measuring of the specific levels present
Dye binding to analytes such as albumin leads to
color changes at micromolar levels. Indicator
94 L A B O R ATO RY M E D I C I N E VO L U M E 3 1 , N U M B E R 2 F E B R U A RY 2 0 0 0
over a continuous range of values (eg, glucose or
therapeutic drug monitoring) or do not have
well-established clinical outcomes or have smaller
differences between decision levels (eg, cholesterol
risk assessment).
Reagent strips that provide threshold results
Control band
are usually calibrated by the manufacturer, based
on agreements with quantitative reference meth-
ods and accounting for as many sources of errors Test band
as possible. The dynamic range of the reagent sys-
tem is centered on the clinical decision threshold
Scientific Communications
changes in color
However, the operating instructions on dipping requires greater analysis, characterization, and
intensity at the test
and placing the strip on the analyzer must be fol- process control by the manufacturer. band.
lowed. Read times are typically set for conve- The challenge of providing quantitative results
nience. Reading at incorrect times can cause errors with reagent strip technology has prompted sev-
in some systems, as not all tests are read eral manufacturers to introduce liquid systems for
time–independent. Samples are usually applied by quantitative applications. This is particularly true
dipping the strip into the specimen with immedi- with analytes that require measurement over a
ate removal or by placing a drop of sample on the wide continuous range or without large degrees of
dry reagent pad. Prolonged exposure to the sam- difference between clinical results (positive and
ple can leach chemical from the reagent pad and negative). In many cases, point-of-care blood
4
alter results. Although most potential interfer- chemistry analyzers also require CLIA [Clinical
Section
ences are known and mentioned in the labeling, Laboratory Improvement Amendments of 1988]
contamination of specimens with any additive waived status (highest degree of accuracy with
may affect results and should be avoided. essentially error-free use). Liquid reagent systems
Quantitative results with reagent strips have can be factory calibrated yet automatically treat
required more carefully controlled calibration specimens with the diluents needed to reduce
processes, either through use of calibration solu- matrix affects and improve accuracy. One point-
tions or a lot-specific calibration number or cod- of-care system is the hemoglobin A1c assay system
ing. Calibration numbers are factors used in the on the DCA 2000+ Analyzer (Bayer).
standard curve to increase accuracy, and are The use of reagent strips for threshold results
derived from characterization of manufacturing has required products to be more specific to dis-
ease conditions. Manufacturers have addressed
F E B R U A RY 2 0 0 0 VO L U M E 3 1 , N U M B E R 2 L A B O R ATO RY M E D I C I N E 95
COMING UP IN MARCH
Feature
interference through development of new chem-
istry tests. Greater sensitivity is also being
addressed with tests that detect new analytes that
increase earlier in disease progression. For example,
Market Your Laboratory for urinary protein assessment, the move is away
Experienced laboratory leaders cite customer service, monitoring from using the presence of Ä300 mg/L protein as a
costs, and keeping abreast of volume demands as key marketing decision limit and toward using trace levels of albu-
musts for a successful outreach program. Review how marketing min (eg, Ä30 mg/L, or microalbuminuria) for
impacts outreach business in the next feature article to appear in detection of nephropathy risk. The Clinitek
Laboratory Medicine. Microalbumin Reagent Strip (Bayer) is an example
of a more sensitive and specific technology that
96 L A B O R ATO RY M E D I C I N E VO L U M E 3 1 , N U M B E R 2 F E B R U A RY 2 0 0 0