You are on page 1of 5

CE UPDATE—INSTRUMENTATION II

Michael J. Pugia, PhD

Technology Behind
Diagnostic Reagent Strips

Downloaded from https://academic.oup.com/labmed/article/31/2/92/2657056 by guest on 23 May 2021


ABSTRACT Diagnostic reagent strips are commonly used The routine urinalysis panel used today tests for
in clinical analysis of urine and blood, in particular for glucose (1956), protein (1957), ketone (1957), pH
monitoring glucose concentration. Results are obtained (1959), occult blood (1961), bilirubin (1969), uro-
instrumentally or visually as thresholds and quantitative bilinogen (1969), nitrite (1972), specific gravity
(1981), and WBCs (1984) (Fig 1). Multistix 10 SG
outputs. Dry reagents are applied in the construction of (Bayer Diagnostics, Elkhart, IN; 1984) was an early
strips in a variety of ways. The mechanism used to make multiple panel with 10 reagents included on 1
strips operational and practices used by manufacturers to reagent strip.
establish performance are reviewed, and limitations and Aside from urinalysis, strip technology was also
benefits of reagent strips are assessed. applied to blood analysis with a series of glucose-
monitoring products, starting with Dextrostix
This is the second article in a 4-part continuing education series on instrumentation.
(Miles Laboratories; now Bayer; 1966) and leading
On completion of this article, the reader will be able to identify the types of
to the current glucose methods. In addition, strip
technology used in reagent strip products.
technology was used for analysis of other blood
From Bayer
analytes, such as lipids, hemoglobin, alanine
Diagnostics, History of Reagent Strips aminotransferase (ALT), aspartate aminotrans-
Elkhart, IN. The diagnostic reagent strip, with 1 or more ferase (AST), bilirubin, and uric acid, and for ther-
Reprint requests to reagent pads adhered to a plastic handle, is one of apeutic drug monitoring. Instrumentation for
Dr Pugia, Bayer the most common testing technologies in routine these uses included the Reflotron system
Diagnostics, 1884 clinical use. The key reason for its acceptance is (Boehringer Mannheim, Indianapolis, IN; 1985)
Miles Ave, Elkhart, IN
ease of use. One of the first reagent strips contained and the Seralyzer system (Bayer; 1981).
46515-0070.
a reagent pad for glucose (Clinistix; Ames Co, Early dipsticks relied solely on colorimetric
Elkhart, IN; 1956) that could be dipped into urine, reagents, and even today most products are col-
was allowed to react for a minute, and read.1 This orimetric. The reagents change color, with the
diagnostic method eliminated the necessity of intensity of the color proportional to the concen-
preparing liquid reagents and was easier to use tration of the analyte measured in the clinical
than tablet reagents. During the next 40 years many specimen. For visual interpretation, the color of
products were developed that offered the same ease the reagent is matched against 2 or more blocks
of use and simplicity, benefits still valued today. on a carefully developed color chart. Each block
After the urine glucose strips, a series of reagents represents an assigned analyte concentration
were developed, initially with diabetes mellitus as range. Assignments are based on agreement of
the disease focus. As dry reagent pad tests for more reagent strip results with reference methods. The
analytes became available, testing panels were reagent is read at a specified time after application
defined to assist diagnosis in the areas of kidney of the specimen.
function, liver function, urinary tract infections,
and carbohydrate metabolism.2

92 L A B O R ATO RY M E D I C I N E VO L U M E 3 1 , N U M B E R 2 F E B R U A RY 2 0 0 0
Instrumentation
Instrumental reading arose from the need for
accuracy and data documentation. It eliminated
the subjectivity of visual interpretation and Glucose
increased convenience. The earliest instrument
was the Ames Reflectance Meter (ARM; Miles
Bilirubin
Laboratories), introduced in 1966 for measuring
blood glucose concentration. During the next 30
years a series of instruments was developed to read Ketone
dry reagent strips. These instruments translated
the reagent colors into numerical output by mea- Urobilinogen

Downloaded from https://academic.oup.com/labmed/article/31/2/92/2657056 by guest on 23 May 2021


suring reflectance. The first urine strip instrument
was the Clinilab (Bayer), introduced in 1972. Protein
Today analyzers range from high throughput, fully
automated systems (Clinitek Atlas; Bayer; 1994) to pH
single urine strip analyzers for use in the physi-
cian’s office (Clinitek 50, Bayer, 1996; Urilux S,
Occult blood
Boehringer Mannheim, 1998). Microscopic ana-
lyzers have also made use of reagent strips (Yellow
Iris, Iris, Chatsworth, CA). Specific gravity
Glucose meters were the first instruments with
low enough cost and ease of operation for con- Nitrite
sumer use. With the advance of low-cost digital
processing, in 1981 Bayer introduced its Glucome- WBCs
ter, and today a number of companies offer glu-
cose meters to meet a variety of individual needs Negative Positive
(eg, One Touch II, Lifescan, Milpitas, CA; Com-
panion II, Medisense, Cambridge, England;

Scientific Communications
Reflolux, Boehringer Mannheim, Mannheim,
Germany).
Dry reagent strip technology involves the con- Fig 1. Multiple panels
trolled application of reagents in the manufacture with 4 reagent areas (Fig 2). The first area at the tip of dry reagents are
of the reagent pads. These reagents contain indica- of the strip is for sample application and overlaps used for urinalysis.
tor dyes, metals, enzymes, polymers, antibodies, the next reagent area, providing for transfer of the Each reagent detects
and various other chemicals dried onto carriers.3 patient sample (urine) to the third reagent area. a specific analyte
through the use of
Carriers commonly used are papers, membranes, The treated sample then migrates across a third colorimetric
or polymers with various sample uptake and layer, where reactants are immobilized for color reactions.
transporting properties. development. This migration is driven by a fourth

4
area that takes up excess specimen.
Section
Reagent Strip Design The chromatography reaction takes place in the
In the 1980s, reagent strip design began to include third area, called the test or capture zone, typically
multiple reagent layers to measure 1 analyte. This a nitrocellulose membrane. In the first and second
allowed chemical reagent systems to be placed in areas, an analyte-specific antibody reacts with the
distinct reagent layers and provided for reaction analyte in the specimen and is chromatographi-
separation steps such as chromatography and fil- cally transferred to the nitrocellulose membrane.
tration. Immunochromatography strips are con- The antibody is bound to colored latex particles as
structed so that chemical reactions occur in a label.4 If the sample contains the analyte, it reacts
distinct areas of reagents. with the labeled antibody. In the capture zone a
The Clinitek hCG strip test (Bayer) for human second antibody is immobilized in a band and
chorionic gonadotropin is a dry reagent strip test

F E B R U A RY 2 0 0 0 VO L U M E 3 1 , N U M B E R 2 L A B O R ATO RY M E D I C I N E 93
Fig 2. The immuno-
chromatography strip
has 4 distinct areas:
sample application
area; reagent area; Antibody-latex
capture zone; and conjugate
absorbent area.

Analyte
Test band

Control band

Downloaded from https://academic.oup.com/labmed/article/31/2/92/2657056 by guest on 23 May 2021


Blood Absorbent/ Nitrocellulose Absorbent Polystyrene
separation conjugate area membrane area backing
area

captures particles when analyte is present. A col- dyes can be covalently bound to the analyte (dia-
ored test line is formed (Fig 3). A second band of zonium compounds binding bilirubin) or tightly
reagent is also immobilized in the capture zone to associated with the analyte (sodium-sensing
allow a control line to react with particles, forming indicators). Enzymatic reactions can be used to
color. Color at the control line is always formed detect enzymes at micromolar levels through
when the test system is working properly, even in reactions with color-forming substrates. Enzy-
the absence of hCG in the patient sample (Fig 3). matic reactions can also be used to detect mole-
Whole blood glucose strips often use multiple cules, such as glucose, through reactions with
reagents to trap intact RBCs that interfere with the enzymes to yield colored end products. Particle-
color generation area. For example, the Encore labeled antibodies are the primary reagents that
glucometer (Bayer) uses a trapping area placed provide for the detectable reaction of immuno-
directly over the color-generating area. The color logic strips based on chromatography. Redox
is read from the bottom of the strip through a catalysis uses metal chelates to oxidize or reduce
Test Your transparent window. Other designs allow the sam- indicators in the presence of specific analytes
Knowledge! ple to migrate to a color-generating area aside such as hemoglobin, and can detect molecules
Look for the CE from the trapping area, and color is read from the down to the nanomolar level.
Update exam on
top of the strip. Whole blood test strips often
Instrumentation (002)
in the April issue of
make use of plastic cassettes to hold the reaction Reagent Strip Performance
Laboratory Medicine. areas in place. Reagent strip technology has been used to provide
Participants will earn Multiple layers of reagent have also been either threshold or quantitative results for many
4 CMLE credit hours. applied to film slides, such as the reagent system analytes. Both applications are well described for
used with the Ektachem analyzer (Vitros) devel- many clinical uses. Threshold tests are applied to
oped by Eastman Kodak in 1980. Slides enabled analytes with well-established clinical decision
multiple separating, spreading, and color-forming limits, based on large outcomes studies (eg, albu-
areas to enhance colors. minuria, hCG), and have well-defined differences
Typical chemical reactions occurring in dry between positive and negative test results. Quanti-
reagent strips can be grouped as dye binding, tative tests are more often applied to analytes that
enzymatic, immunologic, and redox catalysis.3 require measuring of the specific levels present
Dye binding to analytes such as albumin leads to
color changes at micromolar levels. Indicator

94 L A B O R ATO RY M E D I C I N E VO L U M E 3 1 , N U M B E R 2 F E B R U A RY 2 0 0 0
over a continuous range of values (eg, glucose or
therapeutic drug monitoring) or do not have
well-established clinical outcomes or have smaller
differences between decision levels (eg, cholesterol
risk assessment).
Reagent strips that provide threshold results
Control band
are usually calibrated by the manufacturer, based
on agreements with quantitative reference meth-
ods and accounting for as many sources of errors Test band
as possible. The dynamic range of the reagent sys-
tem is centered on the clinical decision threshold

Downloaded from https://academic.oup.com/labmed/article/31/2/92/2657056 by guest on 23 May 2021


to allow optimal clinical decision making. There-
after, the instrument automatically self-adjusts
during first-time power-up and during each assay.
In the event the system is unable to make appro-
priate internal adjustments, an error message is
displayed.
Before the manufacturer releases reagent strips,
each lot of reagent paper and finished reagent
strip package undergoes thorough analysis and
Negative Low Medium High
characterization. Reagents are released with fac- positive positive positive
tory calibration after meeting requirements that
account for as many sources of variation as possi-
Fig 3. The
ble. When reagents are stored and used properly, lots. Calibration solutions reduce ease of use and
progression of
acceptable performance is ensured until the expi- provide additional sources of analytical error (the results for an
ration date. process of doing calibration), but provide greater immuno-
Major manufacturers ensure performance with accuracy potential. Lot-specific calibration cannot chromatography
tight process controls specific for their analyzers. offer as much user control of accuracy and strip occurs through

Scientific Communications
changes in color
However, the operating instructions on dipping requires greater analysis, characterization, and
intensity at the test
and placing the strip on the analyzer must be fol- process control by the manufacturer. band.
lowed. Read times are typically set for conve- The challenge of providing quantitative results
nience. Reading at incorrect times can cause errors with reagent strip technology has prompted sev-
in some systems, as not all tests are read eral manufacturers to introduce liquid systems for
time–independent. Samples are usually applied by quantitative applications. This is particularly true
dipping the strip into the specimen with immedi- with analytes that require measurement over a
ate removal or by placing a drop of sample on the wide continuous range or without large degrees of
dry reagent pad. Prolonged exposure to the sam- difference between clinical results (positive and
ple can leach chemical from the reagent pad and negative). In many cases, point-of-care blood

4
alter results. Although most potential interfer- chemistry analyzers also require CLIA [Clinical
Section
ences are known and mentioned in the labeling, Laboratory Improvement Amendments of 1988]
contamination of specimens with any additive waived status (highest degree of accuracy with
may affect results and should be avoided. essentially error-free use). Liquid reagent systems
Quantitative results with reagent strips have can be factory calibrated yet automatically treat
required more carefully controlled calibration specimens with the diluents needed to reduce
processes, either through use of calibration solu- matrix affects and improve accuracy. One point-
tions or a lot-specific calibration number or cod- of-care system is the hemoglobin A1c assay system
ing. Calibration numbers are factors used in the on the DCA 2000+ Analyzer (Bayer).
standard curve to increase accuracy, and are The use of reagent strips for threshold results
derived from characterization of manufacturing has required products to be more specific to dis-
ease conditions. Manufacturers have addressed

F E B R U A RY 2 0 0 0 VO L U M E 3 1 , N U M B E R 2 L A B O R ATO RY M E D I C I N E 95
COMING UP IN MARCH
Feature
interference through development of new chem-
istry tests. Greater sensitivity is also being
addressed with tests that detect new analytes that
increase earlier in disease progression. For example,
Market Your Laboratory for urinary protein assessment, the move is away
Experienced laboratory leaders cite customer service, monitoring from using the presence of Ä300 mg/L protein as a
costs, and keeping abreast of volume demands as key marketing decision limit and toward using trace levels of albu-
musts for a successful outreach program. Review how marketing min (eg, Ä30 mg/L, or microalbuminuria) for
impacts outreach business in the next feature article to appear in detection of nephropathy risk. The Clinitek
Laboratory Medicine. Microalbumin Reagent Strip (Bayer) is an example
of a more sensitive and specific technology that

Downloaded from https://academic.oup.com/labmed/article/31/2/92/2657056 by guest on 23 May 2021


CE Updates indicates clinically significant analyte levels through
Instrumentation—Hematology Analyzers new chemistry tests.5
Offer a User-Friendly New Technology
This third article in a 4-part continuing education series on Conclusion
instrumentation incorporates the principles, applications, and Diagnostic reagent strips have been used for more
user-friendliness of hematology analyzers. The once labor-intensive than 40 years, and they continue to serve clini-
manual procedure for blood-cell analysis can now be performed cians because of their ease of use and conve-
with highly automated instruments capable of measuring many nience. Manufacturers have continued to apply
disease-associated parameters, while incorporating nontraditional and improve the technology to adapt to market
flow cytometry, robotics, and expert system technologies. needs. Because of ease of use and acceptable reli-
ability, reagent strip technology remains the pre-
Waste—Medical Waste Management: ferred choice for many applications, particularly
Where Does the Garbage Go? for threshold tests (eg, urinalysis) and tests per-
The modern landfill is designed to control emissions and minimize formed in the physician’s office (eg, hCG,
adverse effects on the environment. This first article in a 3-part microalbuminuria, drugs of abuse, group A Strep-
continuing education series on current waste management tococcus, Helicobacter pylori, Chlamydia, and C-
practices addresses how careful management of medical waste reactive protein).l
minimizes the spread of infection and what happens to medical
waste once it is placed in a municipal solid waste landfill. References
1. Free A, Free H. Urinalysis in Clinical Laboratory Practice.
Boca Raton, FL: CRC Press; 1975:217-224.
Special Report 2. Henry JB, ed. Clinical Diagnosis and Management by Labo-
Medical Technology and Clinical Laboratory Science Faculty Survey ratory Methods, 19th ed. Philadelphia, PA: Saunders;
A questionnaire regarding salary compensation for those teaching 1996:164;411-456.
3. Burtis CA, Ashwood ER. Reference information for the
in medical technology (MT) and clinical laboratory science (CLS) clinical laboratory. In Tietz NW, ed. Tietz’ Textbook of Clinical
was sent to program directors at 126 university-based MT and CLS Chemistry, 3rd ed. Philadelphia, PA: Saunders; 1999:483-484,
programs to determine if there have been changes since previous 798-801, 1218, 1799-1839.
4. Bangs LB. New developments in particle-based immunoas-
surveys. Find out the results compiled by academic rank according says: introduction. Pure Appl Chem. 1996;68:1873-1879.
to geographic location and type of program in the March issue of 5. Pugia MJ, Lott JA, Luke KE, et al. Comparison of instru-
Laboratory Medicine. ment-read reagent strips for albumin and creatinine in urine
with visual results and quantitative methods. J Clin Lab Anal.
1998;12:280-284.
Science
Central Laboratory Analysis
of Intact PTH Using Intraoperative Samples
A more convenient and less expensive alternative to point-of-care
methods has been developed to test chemiluminescent assays of
plasma intact parathyroid hormone (PTH). Evaluated by collecting
venous blood samples in EDTA tubes, transporting them to the
laboratory by pneumatic tube or courier, and processing and
assaying them as stat tests, the central laboratory method achieved
results within a time interval equivalent to previous point-of-care
tests. In addition, the new procedure did so at a fraction of the cost
of bedside procedures and proved diagnostically valuable.l

96 L A B O R ATO RY M E D I C I N E VO L U M E 3 1 , N U M B E R 2 F E B R U A RY 2 0 0 0

You might also like