UNIVERSAL INSPECTION CO. LTD.
‘eRe ror REED
INSTRUMENT & EQUIPMENT CALIBRATION DIVISION — race tors
CERTIFICATE ALIBR/ yt
CERTIFICATE NO: — UIC / 232706
his certifies that the instrument below has been tested/calibrated using
equipment traceable to National/International Standards or verified by
ratio type mei s using self calibrating techniques.
Information on the calibration process and the results are produced
below and given in the accompanying, sheets of this document
All related traceable documents are on file al GINAO7 ENE SAIC.
INSPECTION CO. LIED. and may be examined on request
Customer Tove Leeds Wanco Certification and Inspection
Address ‘Ar-Riyadh, Kingdom of Saudi Arabia
Description Digital Vernier Caliper (0 - 150 mm)
Manufacturer INSIZE
Code 1108-150 is
Serial No. 2601213307 26-Apr-22
ID No. NA li q 28-Apr-22 |
Location Calibration Lab (Jubail) i 27-Apr-23
The instrument was calibrated using the following reference, maintaived.at UIC Instrument & Equipment
Calibration division conforming to ISO/IEC 17025:2017 requirements
Reference Standard Equipment Used
Description Cert. No. Serial No Due Date
Caliper Checker ro9/200087 02-Sep-22,
Depth Cheker /09/200052 03-Sep-22
‘The measurement results pertain only to the instrument calibrated.
‘This certficate and the attached measurement result sheets shall not be reproduced except in ull without the prior permission
P.O. Box No. 2072, Jubail - 31951, K.S.A. Baap agp ASL P9401 aed VT GG
34d 8881 ANT PEL AMM) Gaal / AYU PPLE ANAT ASL
wy.com.sa AN FW 29) SLsUNIVERSAL INSPECTION CO, LID. Bagh pa! Glow bball AS yl
CERTIFICATE NO: UIC / 232706 PAGE ors
customer Tove Leeds Wanco Certification and Inspection
Address ‘Ar-Riyadh, Kingdom of Saudi Arabia
Description Digital Vernier Caliper (0-150 mm)
Manufacturer INSIZE Calibration Date 28-Apr22,
Model 1108-150, Calibration Due Date 27-Apr-23}
lSerial No. 2601213307 Lab Temperature aire]
1D No. A Relative Hum RH 53%
Calibration Test Results for Digital Vernier Caliper
Min, Range: 0.0 (ara) Max. Range: 150.0 (mm)
Res. 0.01 (mm)
standard Setting UUT Reading Error Expanded Uncertainty
(mm) (mm) (mm) (um)
25.0 25.02 0.02 1457
outside Dia 500 50.02 on 1457
0.50
(mm) 750 75.01 oot 1458
150.0 150.01 oot 14.62
250 25.02 oor 1457
Inside Dia 50.0 5001 oot 1457
0-150
(mm) 750 7499 0.01 1458
150.0 149.99 001 1482UNIVERSAL INSPECTION CO.LTD. Baystat! ul ae Ballad ea Ltd
CERTIFICATE NO: UIC / 232706 act sos
Standard Setting uurReading Exvor Expanded Uncertainty
(mm) (rom) (om) (um)
250 24.99 0.01 1438
Depth 50.0 50.00 0.09 1439
0-150
(mm) 750 74,99 0.01 1439
150.0 150.00 .00 15.89
UIC/P/DVC Rev. 02
Referer
Based on JIS 87507:2016 & ISO 13385-1
standards
Remarks / Deviation IF ANY:
The results stated in this certificate relate only to the item calibrated.
Physical Condition of the UUT
The Unit Under Testis in Goad Physical Condition.
Calibration uncertainty
The report expanded uncertainty of measurement is stated as the stadard uncertainty of measurement multiplied by the
coverage factor k=2, corresponding to a coverage prabability of approximately 95%,
calibration Engineer