Test record
TEST PASSED
Test performed Ansur components used
Date: 07/08/2014 Ansur Version 2.9.7
ECOGRAFO
Record: Plug-In: AVPI Version 2.4.10
ESAOTE.mtr
Plug-In: ESA612 Version 1.1.2
ECOGRAFO
Template:
ESAOTE.mtt
Template version: 1.2.5
Test setup
Selections
Service events performed Standards performed
User defined
MTI Data
Test instrument Serial number Firmware version
ESA612B 2540003 v2.03
Signatures
Page 1 of 3
Date: 07/08/2014 ECOGRAFO ESAOTE.mtr
Fluke Biomedical AnsurTest Report TECNOIMAGEN S.A. 07/08/2014
Test result
Test element Test type Fail
Visual test
Visual Test - Pass/Fail Pass/Fail
Procedure:
Visual test with result that can be True or False
Result: Recorded value
Exterior of equipment is
without damage or missing Pass
hardware.
Mains Voltage
Live to Neutral Live to Neutral
Result: Value Unit High limit Low limit Standard
Live to Neutral 221,2 V User defined
Mains Voltage
Neutral to Earth Neutral to Earth
Result: Value Unit High limit Low limit Standard
Neutral to Earth 221,5 V User defined
Mains Voltage
Live to Earth Live to Earth
Result: Value Unit High limit Low limit Standard
Live to Earth 1,3 V User defined
Page 2 of 3
Date: 07/08/2014 ECOGRAFO ESAOTE.mtr
Fluke Biomedical AnsurTest Report TECNOIMAGEN S.A. 07/08/2014
Page 3 of 3
Date: 07/08/2014 ECOGRAFO ESAOTE.mtr