You are on page 1of 9

RAPID COMMUNICATIONS IN MASS SPECTROMETRY

Rapid Commun. Mass Spectrom. 12, 456–464 (1998)

Investigation of Spectral Reproducibility in


Direct Analysis of Bacteria Proteins by Matrix-
assisted Laser Desorption/Ionization Time-of-
flight Mass Spectrometry
Zhengping Wang,1 Larry Russon,1 Liang Li,1* Dennis C. Roser2 and S. Randolph Long3*
1
Department of Chemistry, University of Alberta, Edmonton, Alberta, Canada T6G 2G2
2
GeoCenters, Gunpowder Branch, Aberdeen Proving Ground, MD 21010, USA
3
Edgewood RDE Center, SCBRD-RTE, Aberdeen Proving Ground, MD 21010, USA

Matrix-assisted laser desorption/ionization (MALDI) time-of-flight mass spectrometry (TOFMS) can be


used for rapid detection of bacteria proteins in a crude mixture. It can potentially be used as a tool for
bacterial identification based on the mass spectral patterns or the appearance of some characteristic mass
peaks. However, there are many experimental parameters that can potentially have a strong effect on the
observed mass spectra. The objective of this work is to address the mass spectral reproducibility issue.
Several experimental parameters that may affect the MALDI spectra are systematically investigated.
Results of spectral comparison from two laboratories with different operators and instrumentation are
presented. It is demonstrated that minor variations in the sample/matrix preparation procedures for
MALDI and in the experimental conditions used for bacterial protein extraction can result in a significant
change in the observed spectra, though a number of peaks are conserved in the spectra obtained under
different experimental conditions from the same bacterial sample. These conserved peaks may potentially
be used as the biomarkers for bacterial identification. It is stressed that this type of investigation on spectral
reproducibility should be carried out for different bacterial species in order to identify the mass spectral
peaks that are consistently detected regardless of operator and nominal variations in sample preparation
approach. # 1998 John Wiley & Sons, Ltd.
Received 5 December 1997; Revised 5 February 1998; Accepted 9 February 1998

Rapid and unambiguous identification of bacterial species identification of bacteria. A potential advantage of the
and strains has a broad appeal to in-field hazard assessment MALDI technique is its ability to analyze crude protein
as well as clinical and environmental applications.1 One mixtures without the need of extensive separation.4 Several
possible approach to identifying bacteria is to examine the researchers have shown the possibility of using MALDI
protein content and seek protein biomarkers that are unique time-of-flight mass spectrometry (TOFMS) for rapid
to specific bacteria. The type of proteins present in bacterial identification of bacteria based on mass spectral patterns
cells is immense and their amounts vary greatly. Two- derived either from protein extracts5,6 or from whole cells7,8
dimensional gel electrophoresis analysis of proteins has incorporated into the MALDI matrix. One or more peaks
revealed that at least 1000 different proteins are present in detected in the mass spectrum, which appeared to be
Escherichia coli cell extracts.2 The estimated total number conserved for a certain type of bacteria, were proposed for
of individual proteins expressed from the Escherichia coli identification. Within each of these studies, the mass
genome is well over 4000.3 One or several proteins among spectral patterns associated with different bacteria were
such a large pool of proteins may be used as the protein different, providing evidence that protein analysis by mass
biomarkers for bacteria identification. The challenge is to spectrometry can serve as a basis for bacterial discrimina-
identify these biomarkers. The selection of suitable tion. Identification of specific mass peaks as genus- and
biomarkers may well be dictated by the detection method even strain- specific ‘biomarkers’ has been attempted,6,8
used. The protein biomarkers identified should be readily though the database among which these identifications are
analyzed and less prone to interference from other species in made is limited in scope.
bacteria cells and from background species and molecules For general use of the MALDI method (as well as any
present in real world samples. other mass spectrometric methods such as electrospray
With the recent advances in ionization and detection ionization) in bacterial identification, there are several
methods, mass spectrometry has become a powerful tool for important questions that need to be addressed. Most
protein analysis. Both matrix-assisted laser desorption/ important is the level of reproducibility that can be obtained
ionization (MALDI) and electrospray ionization (ESI) mass when different investigators use nominally the same
spectrometry can potentially be used as a means of rapid methods for generating spectra from bacterial samples. An
essential related question is the extent to which experi-
*Correspondence to: Liang Li, Department of Chemistry, University of
Alberta, Edmonton, Alberta, Canada T6G 2G2 and S. R. Long, mental conditions affect the mass spectra observed. When
Edgewood RDE Center, SCBRD-RTE, Aberdeen Proving Ground, these issues are addressed and understood, it will become
MD 21010, USA. possible to systematically seek and identify those bacterial

CCC 0951–4198/98/080456–09 $17.50 # 1998 John Wiley & Sons, Ltd.


MALDI-TOFMS FOR BACTERIA PROTEIN ANALYSIS 457

Figure 1. MALDI spectra of Escherichia coli (ATCC 9637) or EC9637 obtained at the University
of Alberta lab. by using different solvents for preparing the second-layer solution in a two-layer
sample preparation method. The second-layer solution consists of saturated HCCA in (a) 33%
acetonitrile/67% water and the sample solution (1:1) (all by volume), (b) 17% formic acid/33%
isopropanol/50% water and the sample solution (1:1), and (c) 17% formic acid/33% methanol/
50% water (FMW) and the sample solution (1:1). The sample solution is the bacteria extract using
0.1% TFA as the extraction solvent.

peptide/protein masses which have validity as ‘biomarkers’ suspended in 250 mL solvent (100 mL for the two basic
useful for discrimination of bacteria. In this work, we solvents). The cell suspension was vortexed for about 2 min
attempt to address the spectral reproducibility issue as well and then centrifuged. The supernatant solution was taken for
as to examine the effects of sample/matrix preparation and MALDI analysis.
protein extraction method on mass spectral patterns. The two-layer method was used for matrix/sample
preparation.9 Both a-cyano-4-hydroxycinnamic acid
(HCCA) and sinapinic acid were initially examined as the
EXPERIMENTAL
matrices for bacterial protein analysis and it was found that
The studies reported here utilized as samples two strains HCCA provided better sensitivity. Consequently, HCCA
each of Escherichia coli (ATCC 9637 and 11175) and was used throughout this study. In the two-layer method, the
Bacillus thuringiensis (ATCC 10792 and 19267). The first layer was formed by placing 1 mL of 100 mM HCCA in
bacteria were grown overnight (18–24 hrs) in nutrient broth 99% acetone and 1% water (v/v) on the MALDI probe tip
at ambient temperature with shaking. Cells were harvested, and allowing it to dry in air. For the second-layer solution, a
washed with several volumes of sterile water, lyophilized to saturated solution of HCCA was prepared in a solvent
dryness, and stored at 0 °C until use. Bacterial samples were mixture. A variety of solvent mixtures were used; their
grown at Edgewood RDE Center (ERDEC). compositions are described in the results and discussion
The type of suspension solvent tested for dissolving section. The saturated solution was routinely mixed with the
proteins from the cells include water, 0.1% trifluoroacetic sample solutions in a volume ratio of 4:1 in the cases of
acid (TFA), isopropanol, 50 mM ammonium bicarbonate, using methanol, isopropanol, ammonium bicarbonate, and
10 mM Tris-HCl buffer, and a solvent mixture consisting of Tris-HCl buffer as the extraction solvent or 1:1 in the cases
10% formic acid, 45% methanol, and 45% water (by involving other extraction solvents. One mL of this solution
volume). About 1 to 1.5 mg of lyophilized bacteria were was placed on top of the first layer and allowed to dry. The

# 1998 John Wiley & Sons, Ltd. Rapid Commun. Mass Spectrom. 12, 456–464 (1998)
458 MALDI-TOFMS FOR BACTERIA PROTEIN ANALYSIS

tip was then dipped in pure water for approximately 10 s and detected in the three spectra. However, the relative
the excess water was shaken off. Variations from these intensities of these peaks are different. Similar observations
procedures are detailed as appropriate in the results and were obtained from the other bacterial samples examined.
discussion section. The salient feature is that a number of common peaks, albeit
In the University of Alberta experiments, MALDI mass with different relative intensities, are observed under
spectra were recorded in a time-lag focusing linear time-of- different sample/matrix preparation conditions. It appears
flight mass spectrometer with a 1m flight tube.10 A nitrogen that, with the two-layer method, variation of the solvent
laser (337 nm) was used for desorption/ionization. Mass conditions for matrix/sample preparation has a much greater
spectra shown in this work represent the sum of 20 to 100 effect on the relative intensities of the peaks than on the
individual spectra. Spectra were calibrated externally with actual number of proteins detected.
the use of bradykinin, bovine insulin chain B, and equine The observation of differences in overall detection
cytochrome C as the calibrants. sensitivity and relative peak intensity under different sample
In the ERDEC experiments, spectra were recorded on a preparation conditions is not surprising.9,13–19 The analyte
Vestec MALDI-TOF in linear mode using a nitrogen laser incorporation and distribution in the matrix crystals can be
(337 nm) for desorption/ionization, with typically 40 to 100 affected by sample preparation.9 The results shown in Fig. 1
individual scans accumulated per spectrum. The spectra illustrate the need to be careful in controlling the solvent
were externally calibrated using a matrix dimer ion peak and conditions used for preparing the samples in order to
cytochrome C, or calibration mixture 2 (angiotensin I, generate reproducible results. On the other hand, the ability
ACTH [1-17 clip], ACTH [18-39 clip], ACTH [7-38 clip], of varying spectral patterns by changing solvent conditions
and bovine insulin) from the Sequazyme Peptide Mass can be advantageous. Since bacterial identification in the
Standards Kit (PerSeptive Biosystems). MALDI method is based on one or a set of characteristic
peaks in the spectrum, once the biomarkers are identified for
a particular bacteria, optimal sample/matrix preparation
RESULTS AND DISCUSSION methods can be designed for the sensitive detection of these
In the process of generating the MALDI spectra from a biomarkers. While a universal sample preparation method
bacterial sample, there are many experimental factors that for detecting all biomarkers for many different bacterial
can influence the final appearance of the spectra. In order to strains and/or species is desirable, such a preparation
obtain reproducible spectra, these factors must be carefully protocol may be difficult to find. The utility of multiple
identified and controlled. Our approach of identifying these sample preparation protocols with each optimized for a
factors was through systematic investigation using different small number or a group of biomarkers may be a sensible
experimental conditions, with an emphasis on interlabora- approach. Multiple sample handling is a standard feature in
tory comparison. Comparison of experimental results commercial MALDI instruments. Automation of the sample
obtained from two laboratories has greatly facilitated the preparation step is possible. With this in mind, we can again
identification of sources of spectral discrepancy that may examine the MALDI spectra of Escherichia coli shown in
arise in analyzing the same samples. In both labs, sample Fig. 1. The peaks shown around m/z 7709 and 3854 are the
preparation was based on the two-layer matrix/analyte dominant peaks in Fig. 1(b). If these peaks were to be used
preparation method. This method was found to be for identification of Escherichia coli, the experimental
particularly useful for detecting mixtures of peptides and conditions used for generating Fig. 1(b) are preferred to
proteins covering a broad mass range.11,12 For the majority those used for Fig. 1(a) or (c), from the MALDI detection
of this work, a simple solvent suspension method was used point of view.
for extracting peptides and proteins from the bacteria
samples. Effect of the salt content
Another important factor influencing the detection of
Effect of the sample solvent
peptides and proteins is the salt content of the bacterial
In the two-layer method, the first layer is the matrix crystals samples. Fig. 2(a), (b) show the mass spectra of Bacillus
and the second layer is a mixture of the analyte and matrix thuringiensis (ATCC 19267) or BT19267 obtained with and
solution. The chemical composition or the type of solvent without the use of membrane filtration, respectively. (For
used to prepare the second layer solution can have a membrane desalting, Millipore cellulose ester filters having
significant effect on the mass spectral pattern. Figure 1 0.025 mm pore size were used.) It can be seen that a greater
shows the MALDI mass spectra of Escherichia coli (ATCC number of peaks are detected in Fig. 2(a), demonstrating
9637) or EC9637 obtained using different solvents for that salt content in the sample can have a significant effect
preparing the second layer. In all three spectra, the cell on signal detection. This is not surprising in light of the fact
suspension was prepared in 0.1% TFA and the first matrix that the presence of a large amount of salt in a sample can
layer on the probe tip was prepared using HCCA. The affect the efficiency of analyte incorporation in crystals
second-layer solution used for generating Fig. 1(a) consisted during sample preparation, the analyte desorption proper-
of saturated HCCA in 33% acetonitrile/67% water and the ties, and/or the ionization efficiency for certain peptides and
sample solution (1:1) (all by volume). For Fig. 1(b), the proteins.
second-layer solution consisted of saturated HCCA in 17% In Fig. 2 the peaks with similar masses found in both
formic acid/33% isopropanol/50% water and the sample spectra are labeled with m/z values. Several pairs of peaks at
solution (1:1) (all by volume). For Fig. 1(c), the solution m/z 2788/2793, 3683/3688, 5580/5588, 7185/7189, 7231/
consisted of saturated HCCA in 17% formic acid/33% 7236, 7372/7375, and 9521/9523 are likely from the same
methanol/50% water (FMW) and the sample solution (1:1) peptides and proteins. The mass discrepancy observed in the
(all by volume). The m/z values of major peaks in each corresponding peaks in the two spectra can be attributed to
spectrum are shown in the figures. Many common peaks are the difficulty of defining the peak centroid due to the low

Rapid Commun. Mass Spectrom. 12, 456–464 (1998) # 1998 John Wiley & Sons, Ltd.
MALDI-TOFMS FOR BACTERIA PROTEIN ANALYSIS 459

Figure 2. MALDI spectra of Bacillus thuringiensis (ATCC 19267) or Figure 3. MALDI spectra of EC9637 obtained at the University of
BT19267 obtained at the ERDEC lab. (a) The bacteria extract was Alberta lab. under a controlled experimental condition to illustrate the
desalted by using a membrane filter, prior to the MALDI sample effect of the type of extraction solvent on mass spectral pattern. (a)
preparation and (b) no desalting step was used. A 50 mM ammonium 0.1% TFA was used as the extraction solvent and (b) a mixture
bicarbonate solution was used for protein extraction. containing 17% formic acid, 33% methanol, and 50% water (by
volume) was used as the extraction solvent (see text for details on
performing this extraction). The final solvent composition used for
preparing the second-layer solution in MALDI was adjusted to be the
resolving capability of the instrument used, as well as mass same in both cases.
calibration (external calibration was used). Note that all the
peaks detected in Fig. 2(b) are observed in the spectrum
shown in Fig. 2(a), but the relative intensities of these peaks that the final composition of the solution is the same as that
are different. This finding suggests that it is very likely that used for Fig. 3(a). Thus, any mass spectral pattern
there are many more peptides and proteins present in the differences observed under these conditions are solely due
extract than those detected in these MALDI spectra. The to the effect of the type of extraction solvent.
variation of salt content from sample to sample can result in Figure 3 clearly shows that the extraction solvent can
the detection of some peptides and proteins preferentially have a significant effect on the spectral patterns. There are
over the others, resulting in different mass spectral patterns. several common peaks present in the spectra shown in Fig.
This underscores the importance of searching for unique 3(a) and 3(b). They are likely to be from the same peptides
biomarkers via their observed masses, instead of relying on and proteins dissolved in both solvents. Comparing these
the spectral pattern difference for bacterial identification two spectra also reveals that different proteins are extracted
and differentiation. from the two solvent systems. This example illustrates that
the selection of the type of extraction solvent is important
for detecting the potential biomarkers. In addition to TFA
Effect of the extraction solvent
and the FMW solvent mixture, several other extraction
In addition to the aforementioned issues related to sample solvents were tested. Figure 4 shows the mass spectra of
preparation in MALDI, the method of protein extraction EC9637 obtained using various solvents for extraction. Pure
from bacterial samples can also have a major impact on methanol (Fig. 4(a)) and isopropanol (Fig. 4(b)) are found to
mass spectral patterns. Using the solvent suspension method be poor solvents for extraction. Water, ammonium bicarbo-
for protein extraction, it is found that mass spectral patterns nate, and Tris-HCl buffer are found to provide similar
can be significantly affected by the type of the suspension spectra for EC9637, as shown in Fig. 4(c)–(e).
solvent used. This may be due to the solubility difference of In using any aqueous solution for extraction, it was found
the peptides and proteins in different solvents and/or to the that the change in solution pH can have a profound impact
differing extents to which the suspension solution may lyse on mass spectral pattern. This is illustrated in Fig. 5 for three
or perforate the cells. Figure 3 shows the mass spectra of spectra of BT19267 obtained by using ammonium bicarbo-
EC9637 obtained with two different extraction solvents, but nate solutions with different pHs for extraction. In this
under the same matrix/sample preparation condition. Figure experiment, the pH of the 50 mM ammonium bicarbonate
3(a) was obtained by using 0.1% TFA as the suspension solution was adjusted by adding a few drops of 3 M HCl.
solvent for protein extraction. The second-layer solution Figure 5 shows that only a small change in pH can result in
consists of the FMW solvent mixture and the sample very different mass spectra. The extent of the pH effect is
solution in 0.1% TFA (1:1) (all by volume). For Fig. 3(b), found to be sample dependent. For example, Fig. 6 shows
the FMW solvent mixture was used for extraction. The the mass spectra of Bacillus thuringiensis (ATCC 10792) or
second-layer solution for Fig. 3(b) was prepared by adding BT10792, a different strain from BT19267. In this case, the
an appropriate amount of 0.1% TFA to the FMW extract so overall pH effect is not as dramatic as those shown in Fig. 5.

# 1998 John Wiley & Sons, Ltd. Rapid Commun. Mass Spectrom. 12, 456–464 (1998)
460 MALDI-TOFMS FOR BACTERIA PROTEIN ANALYSIS

Figure 4. MALDI spectra of EC9637 obtained at the University of


Alberta lab. by using (a) methanol, (b) isopropanol, (c) water, (d)
ammonium bicarbonate, and (e) Tris-HCl buffer as the extraction Figure 5. MALDI spectra of BT19267 obtained at the University of
solvent. The second-layer solution in the two-layer sample preparation Alberta lab. by using ammonium bicarbonate solutions with different
consisted of 17% formic acid/33% isopropanol/50% water and the pH’s: (a) pH = 7.6, (b) pH = 8.0, and (c) pH = 8.5.
bacteria extract.

However, a significant change in mass spectral pattern is centrifuged at low speed for 10 minutes, and the clear
noted when the extraction solution pH is changed from pH supernatant was transferred to a clean microcentrifuge tube.
8.5 (Fig. 6(c)) to pH 8.0 (Fig. 6(b)). The observed pH effect The soluble proteins were precipitated by addition of 1 mL
is likely a reflection of the variation of peptide/protein of cold methanol followed by storage at 0 °C for 20 minutes.
solubilities in different solutions during extraction. The proteins were pelleted by centrifugation at high speed
It is clear that the type of extraction solvent and, in the for 10 minutes. The supernatant was removed and the
case of aqueous solution, the pH of the solution can have a precipitate was air dried for 10 minutes. Eight samples were
major impact on the mass spectral pattern. These two reconstituted with 75 mL Tris (1 mM, pH 8.0), combined and
parameters should be well controlled to obtain reproducible dried using a Centrivap. For MALDI analysis the dried
results. From the method development point of view, one precipitate was reconstituted in 100 mL water. The resulting
can potentially vary these two parameters to achieve solution was centrifuged at low speed to remove any
optimal conditions for sensitive and selective detection of particulates before use. The MALDI samples were prepared
biomarkers. by adding 1 mL of the extract solution to 9 mL of matrix
solution (saturated a-cyano-4-hydroxycinnamic acid solu-
tion in 1:1 acetonitrile/0.3% TFA by volume). This solution
Effect of the extraction method
was vortexed for 30 seconds and 1 mL was immediately
In the reported studies,4–8 different protein extraction spotted onto a clean MALDI sample pin and allowed to air
methods were used. We have compared the results obtained dry.
from one of the reported extraction methods6 with the As Fig. 7 shows, most peaks detected in Fig. 7(b) are also
solvent suspension method described in this work. Figure 7 found in Fig. 7(a) despite the fact that completely different
shows the mass spectra of Escherichia coli (ATCC 11775) sample preparation procedures were followed and different
or EC11775, a different strain from EC9637, obtained by MALDI instruments were used to obtain the two spectra.
the two extraction methods. Figure 7(a) was obtained at the Peaks observed in common between the two approaches are
University of Alberta lab using the same experimental labeled by the m/z values and they comprise a very large
condition as that used in Fig. 1(b) for EC9637. Figure 7(b) fraction of the observed peaks, particularly above m/z 4500.
shows the mass spectrum obtained at the ERDEC lab using a It is interesting to make a comparison of spectral patterns
modification of the chemical lysis based extraction obtained for the two Escherichia coli strains. The mass
procedure of Ref6. Dry, lyophilized bacteria (2–3 mg) were spectral patterns shown in Fig. 7 for EC11175 are quite
placed in a microcentrifuge tube to which was added 75 mL different (e.g. many more peaks observed) from those of
each of the following solutions: 10mM Tris buffer (pH 8.0), EC9637 shown in Fig. 1 and in Fig. 3(a), where nominally
1% SDS, and 0.1 mM b-mercaptoethanol. This mixture was the same sets of peaks are observed though intensities vary
vortexed, then incubated at 95 °C for 20 minutes. The due to sample preparation conditions. This comparison
cloudy solutions were held at room temperature, 10 mL would suggest that the mass spectral patterns reveal a
DNAse(I) (1 mg/mL) was added, and the solutions were difference between strains. However, when Fig. 3(b) is
held at room temperature for 20 minutes. The mixture was considered, one finds a considerable similarity between the

Rapid Commun. Mass Spectrom. 12, 456–464 (1998) # 1998 John Wiley & Sons, Ltd.
MALDI-TOFMS FOR BACTERIA PROTEIN ANALYSIS 461

prominent mass spectral peaks for EC11775 (Fig. 7) and for


EC9637 extracted by the FMW formulation (Fig. 3(b)).
These comparisons suggest that, while the application of a
given extraction approach may appear to reveal differences
due to strains, the application of alternative extraction
approaches may cause additional peaks to appear. When one
considers all the peaks appearing from different extraction
approaches, the (initially) apparent strain-based spectral
differences may become difficult to discern and certainly
difficult to verify as strain-based differences, given the
dependence of spectra on extraction approach, sample
preparation, and content of salt and other common
biological solution constituents.

Interlaboratory comparison
Although the experimental conditions can have significant
effects on mass spectral patterns, it was found that
reproducible results can be readily obtained within a
laboratory with appropriate controls on conditions applied.
We also examined the interlaboratory reproducibility.
Figure 8 shows the mass spectrum of BT10792 obtained
at the University of Alberta and ERDEC using the
ammonium bicarbonate suspension. The common peaks
Figure 6. MALDI spectra of Bacillus thuringiensis (ATCC 10792) or
BT10792 obtained at the University of Alberta lab. by using
detected in the two panels of Fig. 8 are marked with solid
ammonium bicarbonate extraction solutions with different pH’s: (a) circles. It can be seen that most of the observed peaks are
pH = 7.6, (b) pH = 8.0, and (c) pH = 8.5. replicated in the two studies for BT10792. In this case, even
the relative intensities of many of these peaks are similar in
both spectra. However, we also found that, for some
bacteria, the relative intensities can be quite different for the

Figure 7. MALDI spectra of Escherichia coli (ATCC 11775) or EC11775 obtained by


using different extraction methods: (a) using the solvent suspension method at the same
condition as that of Fig. 1(b) (collected at the University of Alberta lab) and (b) using the
enzyme-based extraction method (collected at the ERDEC lab).

# 1998 John Wiley & Sons, Ltd. Rapid Commun. Mass Spectrom. 12, 456–464 (1998)
462 MALDI-TOFMS FOR BACTERIA PROTEIN ANALYSIS

Figure 8. MALDI mass spectra of BT10792 obtained at University of


Alberta (a) and at ERDEC (b) labs Common peaks are designated by
solid circles. Ammonium bicarbonate solution was used for extraction
and the two-layer method was used for sample/matrix preparation.

spectra obtained in two labs using the same protocol for Figure 9. MALDI mass spectra of BT19267 obtained at University of
sample extraction and MALDI preparation. An example is Alberta (a) and at ERDEC (b). Common peaks are marked with solid
shown in Fig. 9(a) and 9(b) for BT19267 using the circles. Ammonium bicarbonate solution was used for extraction and
ammonium bicarbonate suspension. For comparison, Fig. the two-layer method was used for sample/matrix preparation. In (c) is
9(c) shows the spectrum of BT19267 obtained by using an shown the mass spectrum obtained at ERDEC using the FMW
extraction, wherein the major peaks are essentially replicates of those
extraction method involving sequential addition of formic in (a) and (b). To clarify, the approach used for (c) involved sequential
acid, methanol, and water (see Fig. caption). As Fig. 9(a) addition of the solution components in order F, M, W with vortexing
and 9(b) show, the spectral patterns at the m/z region between additions of components. If the FMW solution is prepared
between 1000 and 4000 are quite different; but by far most first, then added to the sample with vortexing, a significantly different
intensity distribution is observed.
of the peaks observed are common between the two spectra,
as marked by solid circles. In light of the above discussion
considering the factors affecting the spectral patterns, it is
not totally surprising that varying intensities are observed patterns. The intensity of the laser radiation at the sample is
among the peaks recorded in different laboratories. The also difficult to duplicate in different labs.
solvents, matrices, glassware and the sample probe can also The ability of achieving reproducible results that can be
introduce salts or impurities that may affect the spectral replicated in different laboratories is certainly important to
establish the validity of the MALDI technique for bacterial
identification. This work illustrates that control of the
Table 1. The m/z values of the common peaks observed for experimental parameters that can be readily controlled does
BT10792 from two labs permit a substantial, though not complete, replication of
Fig. 8(a) Fig. 8(b) bacterial peptide/protein extract spectra. Common masses
2418 2419 observed by the two labs are provided in Table 1 for
2460 2457 BT10792 and in Table 2 for BT19267.
3216 3216
3234 3237
3345 3344 Comparison with related efforts
3479 3485
3626 3629
The notion of species, genus and strain specific protein
3688 3685 biomarkers has been invoked to utilize protein MALDI
3898 3897 mass spectral profiles for bacterial differentiation/identifi-
4646 4647 cation. This notion is certainly well-founded in principle
4824 4822 due to the fact that the genomic structures of organisms
5017 5016 determine the differences between these organisms and their
5267 5269 classification according to species, genus and strain.20 Since
5598 5598 the genomic sequences code for the production of proteins
7230 7223
in the organism, proteins reflect the genomic differences
7249 7256
7373 7368
between organisms, and mass spectral examinations of
7517 7512
proteins should in principle serve as a basis for differentia-
9648 9652 tion. Specifically, among members of a species such as
9743 9746 Bacillus, some common masses (species-specific biomar-
10 038 10 036 kers) should be observed. Similarly, different strains of a
11 388 11 385 genus, such as Bacillus thuringiensis, should have common
masses, not observed for other Bacillus, that are genus-

Rapid Commun. Mass Spectrom. 12, 456–464 (1998) # 1998 John Wiley & Sons, Ltd.
MALDI-TOFMS FOR BACTERIA PROTEIN ANALYSIS 463

Table 2. The m/z values of the common peaks observed for and optimization of sample treatment/analysis protocols is
BT19267 from two labs clearly required. Lubman and co-workers22 demonstrated
Fig. 9(a) Fig. 9(b)
that application of a separation step such as liquid
1477 1477
chromatography prior to off-line MALDI analysis generates
1624 1627 an overall larger number of peptide/protein masses
1681 1684 observed. This can be attributed to fewer analytes in each
1828 1832 fraction, hence, less ion suppression in MALDI. Such an
1986 1990 approach should enhance the probability of identifying
2835 2831 common masses where expected; unfortunately, the mass
2867 2873 resolution applicable to that effort22 does not permit truly
3077 3083 valid mass matches to be identified.
3166 3171
The significance of the present work is the demonstration
3192 3196
3235 3241
of a capability to attain substantial replication of results by
3347 3347
different laboratories when common sample treatment
3616 3616 approaches are used, in addition to the identification of
3688 3688 some factors that have the potential to cause variability in
3920 3928 observed spectra. When optimization of sample treatment/
4659 4759 analysis methodologies is accomplished, it will become
4825 4826 possible to begin to identify with confidence species-,
5018 5020 genus- and strain-specific protein biomarkers through the
5582 5588 development of sufficiently large libraries of bacterial
6678 6682
protein mass spectral data.
7183 7189
7232 7236
7374 7375
9521 9523
9650 9646
10 036 10 053
CONCLUSIONS
Spectral reproducibility in MALDI analysis of peptides and
proteins directly from bacterial extracts can be influenced
specific. In addition, different strains of the same genus by a number of experimental factors. In this work we have
should have some different masses among them. A identified factors related to two steps in the process, namely
significant degree of similarity should therefore be ex- sample preparation and peptide/protein extraction. With
pected, certainly among members of a species, if a regard to sample preparation, it has been demonstrated that
‘statistically significant’ number of the proteins present in the solvent composition in preparing the MALDI matrix/
a bacterium are observed. sample solution and the salt content in the sample can have a
With this background, it should be possible to make significant effect on mass spectral pattern. On the issue of
comparisons of the results presented in this work with and extraction approach, we find that the solvent suspension
among the results reported by other workers. Several method provides a rapid means of extracting peptides and
previous studies addressing MALDI of Bacillus species proteins from bacterial samples. However, different mass
proteins have been conducted, using either direct analysis of spectra may be obtained using different protein extraction
extracted proteins6 or of whole cells mixed with matrix8,21 processes. The type of extraction solvent and the pH of an
as well as chromatographic separation of extracted proteins aqueous solution used for extraction can have a major
followed by off-line MALDI analysis of the chromato- impact on observed spectra.
graphic fractions.22 We believe that the application of By controlling such factors in interlaboratory compar-
variant experimental conditions, in extraction and/or sample ison, we find there is a much higher level of agreement than
handling/preparation, accounts for the considerably higher of difference in spectra generated by different operators in
level of disagreement than of agreement among these two labs, suggesting that the technique has the potential to
studies, which include several Bacillus species as well as be a valuable bacterial differentiation/identification tool,
different studies of Bacillus thuringiensis. Specifically, no once optimum sample extraction/preparation strategies have
common prominent ions were observed (by different been developed that translate well from one laboratory to
laboratories) to which attribution as species- or genus- another. It can be concluded that despite the significant
specific biomarkers can be made with confidence. Simi- variation of mass spectral patterns that may result from
larly, a comparison among the results of this study and minor changes in experimental conditions, there are many
others on various strains of Escherichia coli7,22,23 presents peaks whose detection seems assured even with some
difficulties with regard to finding commonality among variation in specific processes used. These ‘conserved’
prominent mass peaks. These comparisons can not by any peaks represent the ones that have the highest potential for
means suggest that any observations to date are ‘correct’ or use as biomarkers for bacterial identification.
‘incorrect’. The capability to attain good reproducibility Future work will address the identification of such peaks
within each laboratory is evident in each study. Rather, this and relating their appearance and the structures of
comparison suggests that application of variant experimen- correspondent peptides/proteins to the gene sequences that
tal approaches results in the observation of different subsets code for their production. To this end, an integrated
of the total set of proteins available in the sample. It further approach that combines separation techniques with MALDI
demonstrates that the mass spectrometric analysis of and electrospray ionization, enzyme chemistry, tandem MS,
bacterial proteins is indeed a complex analytical challenge and peptide database is being developed.

# 1998 John Wiley & Sons, Ltd. Rapid Commun. Mass Spectrom. 12, 456–464 (1998)
464 MALDI-TOFMS FOR BACTERIA PROTEIN ANALYSIS

Acknowledgment 12. R. M. Whittal and L. Li, American Laboratory, 29, 30 (1997).


13. K. Strupat, M. Karas and F. Hillenkamp, Int. J. Mass Spectrom. Ion
The authors express their appreciation to Ms. Karen Heroux for Processes 111, 89 (1991).
provision of the bacteria studied in this effort. 14. S. R. Weinberger, K. O. Boernsen, J. W. Finchy, V. Robertson and
B. D. Musselman, Proc. 41st ASMS Conference on Mass
Spectrometry and Allied Topics, San Francisco, CA (1993), pp
REFERENCES 775a–b.
15. F. Xiang and R. C. Beavis, Rapid Commun. Mass Spectrom. 8, 199
1. C. Fenselau, In Mass Spectrometry for Characterization of Micro-
(1994).
organisms, ACS Symposium Series, vol. 541, C. Fenselau (Ed.),
16. O. Vorm, P. Roepstorff and M. Mann, Anal. Chem. 66, 3281
American Chemical Society, Washington, DC (1993), pp 1–7.
(1994).
2. P. O’Farrell, J. Biol. Chem. 250, 4007 (1975).
17. T. M. Billeci and J. T. Stults, Anal. Chem. 65, 1709 (1993).
3. S. R. Pennington, M. R. Wilkins, D. F. Hochstrasser and M. J.
18. A. I. Gusev, W. R. Wilkinson, A. Proctor and D. M. Hercules,
Dunn, Trends in Cell Biology 7, 168 (1997).
Anal. Chem. 67, 1034 (1995).
4. F. Hillenkamp, M. Karas, R. C. Beavis and B. T. Chait, Anal.
19. S. L. Cohen and B. T. Chait, Anal. Chem. 68, 31 (1996).
Chem. 63, 1193A (1991).
20. B. Alberts, D. Bray, J. Lewis, M. Raff, K. Roberts and J. D.
5. T. C. Cain, D. M. Lubman and W. J. Weber Jr., Rapid Commun.
Watson, Molecular Biology of the Cell, 3rd Edn, Garland
Mass Spectrom. 8, 1026 (1994).
Publishing Inc., New York (1994).
6. T. Krishnamurthy, P. L. Ross and U. Rajamani, Rapid Commun.
21. K. L. Wahl, N. B. Valentine, S. L. Gantt, A. Saenz, S. A. Clauss
Mass Spectrom. 10, 883 (1996).
and M. T. Kingsley, In Proc. Joint Services Workshop on
7. R. D. Holland, J. G. Wilkes, F. Rafii, J. B. Sutherland, C. C.
Biological Mass Spectrometry, ERDEC Special Publication, in
Persons, K. J. Voorhees and J. O. Lay Jr., Rapid Commun. Mass
press.
Spectrom. 10, 1227 (1996).
22. X. L. Liang, K. F. Zheng, M. G. Qian and D. M. Lubman, Rapid
8. T. Krishnamurthy and P. L. Ross, Rapid Commun. Mass Spectrom.
Commun. Mass Spectrom. 10, 1219 (1996).
10, 1992 (1996).
23. R. Arnold, C. Houston and J. P. Reilly, In Proc. 45th ASMS
9. Y. Q. Dai, R. M. Whittal and L. Li, Anal. Chem. 68, 2494 (1996).
Conference on Mass Spectrometry and Allied Topics, Palm
10. R. M. Whittal and L. Li, Anal. Chem. 67, 1950 (1995).
Springs, CA (1997), p. 31.
11. R. M. Whittal, L. M. Russon, S. R. Weinberger and L. Li, Anal.
Chem. 69, 2147 (1997).

Rapid Commun. Mass Spectrom. 12, 456–464 (1998) # 1998 John Wiley & Sons, Ltd.

You might also like