Professional Documents
Culture Documents
Please read and understand the service manual carefully to ensure correct servicing of the Analyzer.
After reading, please keep the service manual properly and place it at an accessible position for lookup at any
time.
Intellectual Property
Dirui Industrial Co., Ltd. (hereinafter referred to as Dirui Company) reserves the intellectual property of the
service manual and related products.
Without the written approval of Dirui Company, any individuals or organizations shall not copy, alter or
translate any parts of the manual.
Statement
Dirui Company declares that it will be responsible for the safety, reliability and performances of the product
only if all following requirements are met.
(1) The assembly, readjustment, improvement and repair of the Analyzer are undertaken by personnel
approved by Dirui Company.
Repair Service
Charge-free services:
Products within the scope of warranty of Dirui Company can enjoy charge-free services.
Chargeable services:
(1) Services for products beyond the range of the warranty of Dirui Company.
(2) Repair of products caused by following factors, even within the warranty period.
Artificial damages, improper use, network voltage beyond the regulated range, irresistible natural disasters,
accessories or consumables not approved by Dirui Company, or repair of the product by personnel not authorized by
Dirui Company.
Change/ Refund
Please follow steps below for change/ refund from Dirui Company.
(1) Get the right of change/ refund: Contact the after-sales service personnel of Dirui Company, inform the
serial number of the product, indicate the product model and describe the reason of refund in brief.
(2) Transportation expense: The user shall bear the transportation expenses (including customs fees) to Dirui
Company.
Address: 3333 Yiju Road, New & High Tech. Development Zone, Changchun Post Code: 130103
Domestic Customer Service Hotline: 400 811 6695 400 811 6605
The readers of the service manual are Dirui Company and the maintenance personnel authorized by Dirui
Company.
Symbols
1. Symbols
Symbols Meaning
AC
Lot code
Service life
Serial number
Manufacturing date
Protective earthing
Manufacturer
The equipment complies with the requirements of European Union for in vitro diagnostic
medical devices.
Classification No.
On (power)
Off (power)
Symbols Meaning
The symbol of the crossed out wheeled bin indicates that the product (electrical and
electronic equipment) should not be placed in municipal waste. Please check local
regulations for disposal of electronic products.
Temperature limit
Humidity limitation
Network interface
The symbols above are also applicable to the Analyzer, reagent, QC object and calibration object.
2. Identifications
(1)
Safety Precautions
● If the hospital or organization in charge of the instrument can not formulate a satisfactory servicing/
repair plan, the instrument may not operate normally and personal health may be affected.
● Combustible gas (such as anesthetic) or combustible liquid (such as ethyl alcohol) shall not be used near
the product to prevent explosion.
● Power supply shall be cut off when the product is repaired. Repair of the instrument when it is turned on
may cause electric shock and damages of electrical components.
● Please connect the Analyzer to a socket with an independent fuse and protective switch. If the Analyzer
shares a fuse and protective switch with other equipment (such as life support equipment), trip may occur in
case of a fault, over current or impact current at the moment of startup.
● The clothes, hair and hands of the servicing personnel shall be away from moving components such as
sampling probe to prevent pinching or piercing.
● As the part with warning signs may have mechanical action, human body injuries such as pinching and
stabbing may be caused in normal operation, disassembly or servicing.
● The operator is obligated to follow national and local regulations on discharge and treatment of expired
reagent, waste liquid, waste sample and consumables.
● The reagent will irritate eyes, skin and mucosa. The operator shall abide by laboratory operation safety
regulations and wear personal protective appliances (such as laboratory clothes and gloves) when touching
reagent related articles in the laboratory.
● Once the reagent touches with skin, use a large amount of water to flush the skin, or send the injured to a
doctor if necessary. Once the reagent touches with eyes, use a large amount of water to flush the eyes and
send the injured to a doctor.
● The Analyzer shall be used by professional medical examination personnel or trained doctors, nurses or
testers.
● The samples, quality control objects, calibration objects and waste liquid have potential biological
infectivity. The servicing person shall abide by laboratory operation safety regulations and wear personal
protective appliances (such as laboratory clothes and gloves) when touching and handling relevant articles in
the laboratory.
● As all components and surface of the Analyzer have potential infectivity, safety protection measures shall
be taken in operation and servicing.
● Do not touch the tip of the sampling probe as the tip is sharp and it may have blood sample, quality control
object and calibration object that have potential biological infectivity.
Contents
III
9.4.11 Syringe unit ............................................................................................................................................................ 177
9.4.12 Waste liquid bottle unit .......................................................................................................................................... 179
9.4.13 Waste liquid pump unit ......................................................................................................................................... 181
9.4.14 Diluent assembling components ........................................................................................................................... 183
9.4.15 Waste liquid assembly .......................................................................................................................................... 185
9.4.16 FBA hemolysin assembly....................................................................................................................................... 187
9.4.17 FDO hemolysin assembly ...................................................................................................................................... 189
9.4.18 FDT hemolysin assembly ....................................................................................................................................... 191
9.4.19 CRP lyse assembly ................................................................................................................................................. 193
9.4.20 Residual volume assembly ..................................................................................................................................... 195
9.4.21 Reagent detection assembly ................................................................................................................................... 197
9.5 Liquid circuit diagram and spare parts list ............................................................................................. 199
9.6 Accessories case spare parts list ............................................................................................................... 207
IV
Service Manual of Automatic Hematology Analyzer
2
7
3 4
8
1
Service Manual of Automatic Hematology Analyzer
5
6
7
1
4 8
2
3
2
Service Manual of Automatic Hematology Analyzer
8
4
3
Service Manual of Automatic Hematology Analyzer
1 4
5
2
6
7
3
8
1
2
4
Service Manual of Automatic Hematology Analyzer
5
Service Manual of Automatic Hematology Analyzer
Dispense BF-FDO I
lyse 500μL
Dispense BF-FDT I
lyse 120μL
1:104
Sample 8μL
6
Service Manual of Automatic Hematology Analyzer
7
Service Manual of Automatic Hematology Analyzer
8
Serrvice Manual off Automatic Heematology Analyyzer
BF-6900CRP P has one DIFFD scatterrgram, whicch is obtaineed from low w-angle forw ward scattered light and d
middle-anglle forward sccattered lightt.
BF-6960CRP P has threee DIFF scatttergrams, which w are obbtained fromm low-angle forward sca attered lightt,
middle-anglle forward sccattered lightt and high-aangle forward d scattered liight respectiively.
From thhe BASO chhannel scatterrgram, the peercentage of white blood cell, basophhil and eosin nophil in totaal
number of white
w blood ceell can be obtaained.
1.3.3.2 Whitee blood cell parameters
Throughh the analysiss of BASO channel
c scatteergram and itts BASO reggion, the num mber of WBC and basophiil
(BASO#) is got
g and then the percentagge of basophiil (BASO%) can c be calcullated. The Annalyzer gets th he percentagee
of lymphocyyte (LYM%), neutrophil (N NEU%), monnocyte (MON N%) and eosinnophil (EOS% %) through th he analysis of
DIFF channeel scattergram m and the LY YM zone, NEU U zone, MON N zone and EOS
E zone in iit. Based on the
t number of
white blood cells acquireed with electtrical impedaance method,, the numberr of lymphoccytes (LYM#), neutrophils
(NEU#), monnocytes (MO ON#) and eosiinophils (EOS S#) is obtaineed. The countting unit of thhe cells abovee is 109/L.
(1)White bloood cell
WBC=Tootal number of all particless in the BASO O channel, exxcept from ghhost region.
(2)Basophil
BAASO#=The number
n of parrticles in BAS
SO region in BASO channnel.
(3)Percentagge of basophill
9
Service Manual of Automatic Hematology Analyzer
(4)Percentage of lymphocyte
(5)Percentage of neutrophil
(6)Percentage of monocyte
(7)Percentage of eosinophil
(8)Lymphocyte
(9)Neutrophil
(10)Monocyte
(11)Eosinophil
10
Service Manual of Automatic Hematology Analyzer
After the collected electric pulse is amplified, it is compared with the channel voltage threshold corresponding
to the volume range of normal red blood cell/ platelet. And then, the number of electric pulses of pulse amplitude in
red blood cell/ platelet channel will be calculated. The collected electric pulses are classified according to different
channel voltage threshold and the number of electric pulses in red blood cell/ platelet channel is just the number of
red blood cells/ platelets. The number of cells in each channel divided according to the pulse voltage range
determines the volume distribution of cells. The abscissa shows the volume of cell and the ordinate shows the 2D
diagram of cell quantity, i.e. the histogram reflecting the cell group distribution.
1.3.5.2 Red blood cell parameters
(1)Number of red blood cells
The Analyzer gets the number of red blood cells (RBC) by the direct measurement of the number of electric pulses
of red blood cell, and the unit is 1012/L.
RBC n 1012 / L
HGB
MCH =
Mean hemoglobin content RBC
HGB
MCHC 100
Mean hemoglobin concentration HCT
The unit of RBC is 1012/L, the unit of MCV is fL and the unit of HGB is g/L.
(4)Red blood cell distribution width variation coefficient
The red blood cell distribution width variation coefficient (RDW-CV) is obtained from the red blood cell
distribution histogram. It is a variation coefficient showing volume distribution in percentage.
(5)Standard deviation of red blood cell distribution width
The deviation limit of red blood cell distribution width (RDW-SD) is the width at peak 20% in histogram of red
blood cell distribution and the unit is fL, as shown in the figure.
11
Service Manual of Automatic Hematology Analyzer
12
Service Manual of Automatic Hematology Analyzer
13
Service Manual of Automatic Hematology Analyzer
To ensure the normal operation of the Analyzer after the installation, the Analyzer shall be installed by the
authorized operators of the manufacturer at the delivery.
If the operating environment or power supply of the Analyzer do not meet the requirements above, the
accuracy and precision of test results of the Analyzer may be affected, the Analyzer may be damaged or
personal injury may be caused.
14
Serrvice Manual off Automatic Heematology Analyyzer
(2)Safety connditions
a)Indoor use;
u
b)Altitude shall not excceed 2000m.
c)Ambientt temperaturee: 5℃~40℃.
d)The maxximum relativve humidity isi 80% whenn the temperaature is lowerr than 31℃; tthe relative hu
umidity has a
linear decrrease by 50%
% if the temperature is 40℃
℃.
i not greater than ±10% of
e)The flucctuation of poower voltage is o nominal vooltage.
f)The transsient overvoltage is installlation categorry (overvoltage category) II.
g)The polllution is in Cllass 2.
h)Waterprroof grade: IP
PX0.
i)Material grade: III a.
2.2 Unpacking
2.2.1 Unpaacking stepss
After the insttrument arrivves, please folllow the stepss below to un
npack:
(1) As show wn in the figuure below, makem the antti-tilt mark on
o the box upright.
up firm that the packaging is
Confi
complete, annd the appearrance of the package
p is noot allowed to have deformmation resulting from sharrp penetrationn
injury and heeavy extrusioon that cause the package to lose its pro otective funcction. In case of damage, please
p contacct
the manufactturer or local agent.
(2) Remove thet film wrappped around the t outer casinng of the macchine and cutt the packing strap.
Fig. 2-2-1
1
15
Serrvice Manual off Automatic Heematology Analyyzer
Fig. 2-2-2
Fig. 2-2-3
16
Serrvice Manual off Automatic Heematology Analyyzer
Fig. 2-2-4
b) Instrum
ment unpackinng steps withoout optional sample
s feederr:
Fig. 2-2-5
17
Serrvice Manual off Automatic Heematology Analyyzer
Fig. 2-2-6
b) Open thhe left door off the instrumeent and unscrrew the fixing
g screws on thhe left side off the front do
oor as shown:
Fig. 2-2-7
Fig. 2-2-8
18
Serrvice Manual off Automatic Heematology Analyyzer
Fig. 2-2-9
9
Fig. 2-2-10
19
Serrvice Manual off Automatic Heematology Analyyzer
Fig. 2-2-11
(11)Reinstalll the right dooor of the instrrument.
2.2.2 Hand
dling methood
(1)Use transpport machine such as utilitty trolley for the stable traansportation of
o short distannce.
(2)Prevent thhe aspiration probe
p from other objects anda being dam maged duringg handling andd transportatiion processess.
(3)Keep the Analyzer
A verttical, not tilting or side layying when mo
oving and carrrying.
(4)Try to avooid vibration when handlinng.
2.3 Instru
ument insstallation
Do not disasssemble or in
nstall the insttrument except for the normal
n mainttenance.
1 Waste
W liquid leveel sensor port 2 BF-5D diiluent port 3 Waste liquidd port
Figg. 2-3-1 Connecction of the instrrument and reaggent
20
Serrvice Manual off Automatic Heematology Analyyzer
Fig. 2-3-2
2 Connectioon with reagent inside the instrrument
● The receptacle conneccted with the power cablee shall be relliably ground
ded.
● Do not put the instrum
ment at a plaace where thee disconnectiing device is hard to be ooperated.
(5)Connectioon of bar codee reader
Connect anotther end of acccompanyingg bar code reaader with any " " intterface on thee instrument.
21
Serrvice Manual off Automatic Heematology Analyyzer
Fig. 2-4-1
22
Serrvice Manual off Automatic Heematology Analyyzer
Fig. 2-4-2
● The blankk result obtaiined during the blank tesst is beyond the range seet by the softtware and thhe instrument
will perform
m the blank test
t again. If the result is still not qua
alified after tested
t 3 timees, "blank tesst error" willl
be promptedd.
● The instru
ument will noot have High
h valve, Low value or Susspicious alarrms for the b
blank test ressults.
● When thee power-on self-check alarm
a triggerrs, the instrument enterrs the main interface. Click
C the red
d
exclamation
n mark to ch heck corresp
ponding alarrm informattion. After the
t alarm is cleared, thee instrument
performs sellf-check agaiin.
After the insttrument finishhes the self-ccheck, enter thhe main wind
dow of the software, as shoown in the fig
gure below:
Fig. 2-4-3
23
Serrvice Manual off Automatic Heematology Analyyzer
Fig. 2-4-4
Fig. 2-4-5
(1)Total num
mber of locatioons: Display the current saample serial number
n and total
t sample ssize
(2)Current sttatus: Displayy the analysis mode of the sample.
(3)Test statuss: Indicate the current stattus of the insttrument.
2.4.2.3 Statu
us indication area
On the right side
s of Fig. 2--4-5 is the staatus indicationn area. The arrea displays inn sequence thhe X-B QC sw
witch status, U
disk connectiion status, LIIS system connnection statuus, and printerr status, described as folloows:
(1)X-B QC switch
s status:
: The X-B
X quality coontrol is on.
24
Serrvice Manual off Automatic Heematology Analyyzer
: The X-B
X quality coontrol is not turned
t on.
(2)USB statuus:
: The USB
U storage device
d is connnected.
: The USB
U storage device
d is not yet
y connectedd.
(3)LIS system
m connectionn status:
: The LIS
L is connectted.
: The LIS
L is not yet connected.
(4)Printer staatus:
Fig. 2-4-6
(1)Spl. analyysis: Finish saample test andd the analysiss and display of test resultss on the interrface.
(2)History: The
T instrumeent will automatically savve the analy ysis results inn the databasse after execcuting samplee
analyzing eacch time. The operator can query all sam
mple results stored
s in the database.
d
(3)QC: Enterr the QC interrface to realizze the basic setting of QC and QC counnting.
(4)Add dilueent: Dispense diluent for thhe preparationn of pre-diluttion sample under
u the pre-dilution mod
de.
(5)Rinse: Maanually clean equipment.
(6)Print: Prinnt the reports of selected saamples.
(7)Fault information area::
In case of faiilures of the instrument,
i reelevant failure level inform
mation will bee displayed inn the area. Att the momentt,
click the areaa to pop up a failure displaay dialog boxx, as shown inn the figure beelow:
25
Serrvice Manual off Automatic Heematology Analyyzer
Fig. 2-4-7
I
Interface locatioon Prrompt Autoo-rinse times
Fig. 2-4-8
(1)Interface location:
l dispplay the curreent interface.
(2)Prompt message:
m instruument abnorm
mal info displlay area.
(3)Auto-rinsee times: whetther the instruuments reachees the numbeer of times forr rinse.
When the insstrument has abnormal inffo, it will be displayed
d in th
he area. Clickk the prompt m
message and the followingg
prompt will pop
p up:
26
Serrvice Manual off Automatic Heematology Analyyzer
Fig. 2-4-9
27
Service Manual of Automatic Hematology Analyzer
4
3
3 Rack-feeding Assembly
6
3 1
28
Service Manual of Automatic Hematology Analyzer
1
3
29
Service Manual of Automatic Hematology Analyzer
3.1.3.2 Diagram
3 1
30
Service Manual of Automatic Hematology Analyzer
1
5
4
2
3
31
Service Manual of Automatic Hematology Analyzer
6 1
4
3
32
Service Manual of Automatic Hematology Analyzer
2
3
4
5
6
33
Service Manual of Automatic Hematology Analyzer
3 4
5
6
7
Fig. 3-3-1 Mixing unit
4 Swinger
34
Service Manual of Automatic Hematology Analyzer
Fig. 3-4-1 RBC-PLT reaction cell unit
35
Service Manual of Automatic Hematology Analyzer
2 P2R06-1350-2\ Tube
36
Service Manual of Automatic Hematology Analyzer
5 1
4 3 2
37
Service Manual of Automatic Hematology Analyzer
1
6
38
Service Manual of Automatic Hematology Analyzer
Optocoupler Optocoupler
detection point is detection point is
sheltered not sheltered
Rotate in
Rotate in clockwise
anticlockwise
direction and detect
direction and detect
the boundary
the boundary
If the boundary is If the boundary is not If the boundary is If the boundary is not
detected, determine detected, determine detected, determine detected, determine
current position and current position and current position and current position and
rotate for fixed steps to rotate for fixed steps to rotate for fixed steps to rotate for fixed steps to
return to resetting point return to resetting point return to resetting point return to resetting point
39
Service Manual of Automatic Hematology Analyzer
Fig. 3-11-1 CRP syringe unit
40
Service Manual of Automatic Hematology Analyzer
41
Service Manual of Automatic Hematology Analyzer
4
Fig. 3-14-1 Waste liquid pump unit
42
Service Manual of Automatic Hematology Analyzer
4.2 Schematic diagram of liquid circuit and gas circuit of the Analyzer
43
Service Manual of Automatic Hematology Analyzer
Fig. 4-2-1 Liquid circuit and gas circuit of the Analyzer (model with CRP)
44
Service Manual of Automatic Hematology Analyzer
Fig. 4-2-2 Liquid circuit and gas circuit of the Analyzer(model without CRP)
45
Service Manual of Automatic Hematology Analyzer
46
Service Manual of Automatic Hematology Analyzer
Fig. 4-3-3 Schematic diagram of liquid circuit of red blood cell/ platelet detection module
47
Service Manual of Automatic Hematology Analyzer
48
Service Manual of Automatic Hematology Analyzer
Fig. 4-3-7 Schematic diagram of liquid circuit of waste liquid collection module
Fig. 4-3-8 Schematic diagram of liquid circuit of residual reagent detection module
49
Service Manual of Automatic Hematology Analyzer
4.4 Work process of the gas circuit and liquid circuit of the Analyzer
4.4.1 Sample aspirating process:
The probe unit controls the puncture probe to move to the puncture position and then punctures the test tube. When
the puncture probe moves downwards to the bottom of the test tube, it will control the sample aspirating syringe to
move downwards and start sample aspirating into the puncture probe. The probe unit will then control the puncture
probe to pull out the test tube and then open SV4 and SV16 valves and waste liquid pump, and then control the
syringe to move upwards and rinse the outer wall of puncture probe.
50
Service Manual of Automatic Hematology Analyzer
The probe operates to CRP1 cell first. Then, open SV27 valve, control CRP syringe to move upwards, dispense
BF-CRP lyse to CRP1 cell, control the puncture probe to move downwards below the liquid surface of CRP1 cell,
control the sample aspirating syringe pump to move upwards and dispense the blood sample. Control CRP syringe to
move upwards, dispense residual BF-CRP lyse, close valve SV27 and make the probe reset. Open valve SV24 and
control CRP syringe to have aspirating, dispensing and mixing in CRP1 cell.
The probe operates to W-B cell first. Open valves SV12 and SV29, control the sheath syringe M3 to move
downwards to have BF-FDT I aspirating, close SV29 and control the sheath syringe M3 to move upwards and
dispense reagent to W-B reaction cell. Close valve S12, control the puncture probe to move downwards below the
liquid surface of W-B reaction cell, control the sample aspirating syringe to move upwards to dispense DIFF sample
and reset the probe. Open valve SV20 to have air bubble mixing of DIFF sample. After the mixing, open valves SV3,
SV11 and SV13 and control the rinsing syringe M2 to move downwards, aspirate the sample to the tubing below the
optical unit for storage for optical test after DIFF mixing.
51
Service Manual of Automatic Hematology Analyzer
The probe operates to RBC cell first. Then, open SV8 valve, control rinsing syringe to move upwards, dispense
reagent to RBC cell, control the puncture probe to move downwards below the liquid surface of RBC cell, control the
sample aspirating syringe pump to dispense sample to RBC cell. Reset the probe. Make the syringe M2 move
upwards, dispense residual diluent, close valve SV8, open valve SV28 and enable the air bubble mixing function.
The probe operates to W-B cell first. Open valve SV10, control the rinsing syringe M2 to move upwards to dispense
reagent to W-B reaction cell. Control the puncture probe to move downwards below the liquid surface of W-B
reaction cell, control the sample aspirating syringe pump to move upwards to dispense BASO sample and reset the
probe. Control the rinsing syringe M2 to move upwards to dispense residual BF-FBH reagent to W-B reaction cell,
open valve SV20 to have air bubble mixing of BASO sample. Then, keep it still for HGB test.
52
Service Manual of Automatic Hematology Analyzer
Rinse the inner wall of puncture probe first and then control the puncture probe to move downwards below the liquid
surface of RBC cell, control the sample aspirating syringe pump to move downwards and aspirate the diluted blood
sample. Reset the puncture probe, open valves SV4 and SV16 and waste liquid pump and control the rinsing syringe
to move upwards to rinse the outer wall of the puncture probe. Open valve SV17, drain RBC cell, open valve SV8,
control rinsing syringe M2 to move upwards to dispense diluent to RBC cell for RBC rinsing, and then dispense
diluent to RBC cell, control the puncture probe to move downwards below the liquid surface of RBC cell, open valve
SV3, control the rinsing syringe M2 to move upwards to have secondary sample dispensing to RBC cell. Close valves
SV3 and SV8. Rest the puncture probe, open valves SV4 and SV16 and waste liquid pump, control the rinsing
syringe to move upwards to rinse the outer wall of the puncture probe.
Control the puncture probe to operate to the CRP reagent aspirating position, control the puncture probe to move
downwards to CRP reagent chamber, open valve SV3 and control the rinsing syringe M2 to move downwards.
Aspirate CRP reagent. Make the puncture move upwards and reset it, open valves SV4 and SV16 and waste liquid
pump and control the rinsing syringe to move upwards to rinse the outer wall of the puncture probe. Control the
puncture probe to operate above CRP1 cell and then move downwards to CRP1 cell. Control the rinsing syringe M2
to move upwards and dispense the aspirated CRP reagent to CRP1 cell and close valve SV3. Rest the puncture probe,
open valves SV4 and SV16 and waste liquid pump, control the rinsing syringe to move upwards to rinse the outer
wall of the puncture probe.
53
Service Manual of Automatic Hematology Analyzer
After DIFF dispensing is finished, open valve SV9 and control the sheath syringe M3 to move upwards, open valve
SV11 and control sample aspirating syringe pump to move upwards, dispense mixed DIFF sample stored in tubing to
optical unit to realize classified test of DIFF.
After HGB test is finished, open valves SV3, SV11 and SV13 and control the rinsing syringe M2 to move downwards,
aspirate mixed BASO sample to the tubing below optical unit for storage, close valves SV3, SV11 and SV13, open
valve SV9 and control the sheath syringe M3 to move upwards, open valve SV11 and control sample aspirating
syringe pump to move upwards, dispense mixed BASO sample stored in tubing to optical unit to realize BASO test.
54
Service Manual of Automatic Hematology Analyzer
Start waste liquid pump P1, open valve SV31, pump the negative pressure in waste liquid bottle and then close valve
SV31 after the pressure reaches -40KPa. Open valve SV15 and realize the counting of red blood cell and platelet.
After CRP reagent has full reaction, open valve SV24 and control CRP syringe M4 to move downwards, pump the
mixed sample in CRP1 cell to the cuvette to have data collection and test.
55
Service Manual of Automatic Hematology Analyzer
Label: L1509
Label: L1501
Label: N1509
RV0.75re d
RV1.5red
Terminal block
RV0.75blue
Label: L1508
RV0.75red
Label: N1507 Label: N1506
RV1.5blue
Terminal block
RV0.75red
Label: N1508
RV0.75blue
PWR01 04black GN D 03blue -12V 03blue -12V 06blue -12V Analog signal
06blue -12V 04red +5V 04red +5V 02red +5V collection board
03red L 02red +5V 05black GN D 05black GN D 01black GN D
02 03black GN D Label: J0102 Label: P0102 03
01bl ue N Label: P1501 Label: B20-P11
Label: P1504
Fan
01red +24V
05red L 02black GN D
04 Label: JF1501-2
03bl ue N Sample feeder control board
02 01green +24V 01green +24V 02green +24V 01black BGND 01black BGND SocketJ17
0 1yel low gree n PGN D 06black GN D 06black GN D 01black D GND 02green +24V 02green +24V
RV0.75yellow green Label: P1505
AC-DC power module 01green/red +24V 04 04 Label: P0203 Label: J0203 Label: B05-P17
PWR02 04bl ack GND 05 05
02green +24V 02green +24V 09
05bl ack GND 07black GN D 02green +24V 01green +24V SocketJ27
Label: G1505 03green +24V 03green +24V 07black GN D 02black D GND 01blue PHC_OUT Reagent detection
SocketJ1
06bl ack GND 08black GN D 03green +24V 03 Label: B06-P09 02red +5V board
Label: P1502 09 08black GN D 04 SocketJ9 01red +5V 03black DGND
FBA
Shel l gr ound wi re scr ew Label: J0105 Label: P0105 Label: B06-P27 02blue PHC_OUT1 Label: B08-P01-FBA
03black DGND
Label: G1506 04black DGND
05red +5V
Main control board 06blue PHC_OUT2
07red +5V 01blue PHC_OUT Reagent detection
SocketJ1
08blue PHC_OUT3 02red +5V board
Label: G1502 09black DGND 03black DGND
FDO
10black DGND Label: B08-P01-FD O
11red +5V
RV1.5yellow green 12blue PHC_OUT4
Fig. 5-1-1 Electrical wiring diagram of the Analyzer (model with CRP)
56
Service Manual of Automatic Hematology Analyzer
Lab el:L1 50 1
R V1.5r ed
Terminal block
Lab el:L1 50 8
RV0.7 5red
Lab el:N1 50 7 Lab el:N1 50 6
R V1.5b lu e
Terminal block
RV0.7 5red
Lab el:N1 50 8
RV0.7 5blu e
0 1r ed +1 2V
0 2b la c k G N D
RV0.7 5blu e 0 3b lu e -1 2V
L abe l:J010 3
01 0 1r ed +1 2V 0 1r ed +1 2V 0 4r ed +1 2V S oc k et J1 1
05r ed +1 2V 0 2b la c k G N D 0 2b lac k G N D 0 5b la c k G ND
AC-DC power modulePWR01
04b lac k G N D 0 3b lu e -1 2V 0 3b lue -1 2V 0 6b lu e -12V Analog signal
06b lue -1 2V 0 4r ed +5V 0 4r ed +5V 0 2r ed +5V collection board
0 3r ed L 02r ed +5V 0 5b la c k G N D 0 5b lac k G N D 0 1b la c k G ND
02 03b lac k G N D L abe l:J010 2 L ab e l:P0 10 2 03
0 1b lu e N Lab el:P1 50 1 L ab e l:B2 0-P11
L abe l:P150 4
R ad i at o r f an of pow er
0 1r ed +2 4V 0 1red +24V r ed +2 4V
u ni t
0 2b la c k G ND 0 2bla ck GN D b lac k G N D
L ab e l:J F1 50 1-2 Lab el:P F150 1
0 5r ed L Fa n
04
0 3b lu e N Sample feeder co ntrol board
02 01g re e n +24V 0 1gre en +2 4V 0 2gre en +2 4V 0 1b lac k B G ND 0 1b la c k B GND S oc k et J1 7
0 1y el l ow g r ee n
R V0.7 5y el l ow g ree n
P GND
L abe l:P150 5
AC-DC power module 01g ree n/r ed +2 4V
06b la c k G N D
04
0 6bla ck GN D
04
0 1bla ck DG ND
L abe l:P020 3
0 2g ree n +2 4V
L ab e l:J0 20 3
0 2g re e n +2 4V
L abe l:B05-P1 7
PWR02 04b lac k G N D 05 05
02g ree n +2 4V 02g re e n +24V 09
05b lac k G N D 07b la c k G N D 0 2gre en +2 4V 0 1g re e n +24V S oc k et J2 7
L ab e l:G1 505 03g ree n +2 4V 03g re e n +24V 0 7bla ck GN D 0 2b la c k D G ND 0 1b lu e P H C_O UT Reagent detectio n
S oc k et J1
06b lac k G N D 08b la c k G N D 0 3gre en +2 4V 03 Lab el:B0 6-P0 9 0 2r ed +5V board
Lab el:P1 50 2 09 0 8bla ck GN D 04 S ock etJ9 01r ed +5V 0 3b la c k D GND FBA
S he l l g rou nd wir e
L abe l:J010 5 L ab e l:P0 10 5 L abe l:B06-P2 7 02b lue PH C_O UT 1 L ab e l:B0 8-P01-F BA
scr ew
03b lac k D G ND
04b lac k D G ND
05r ed +5V
Main control board 06b lue PH C_O UT 2
07r ed +5V 0 1b lu e P H C_O UT Reagent detectio n
S oc k et J1
08b lue PH C_O UT 3 0 2r ed +5V board
L ab e l:G1 502 09b lac k D G ND 0 3b la c k D GND FDO
10b lac k D G ND L ab e l:B0 8-P01-F DO
11r ed +5V
RV1.5yel low gr een 12b lue PH C_O UT 4
Fig. 5-1-2 Electrical wiring diagram of the Analyzer (model without CRP)
57
Service Manual of Automatic Hematology Analyzer
01re d C AN_H
S ocket J11
02b la ck CA N_L
03
Label: B02-P11
03re d C AN_H
S ocket P7
01b la ck CA N_L
02
04
Label: B14-P07
03re d C AN_H
S ocket J37
01b la ck CA N_L
02
04
Label: B06-P37
Fig. 5-1-3 Communication connection diagram of the Analyzer (model with CRP)
58
Service Manual of Automatic Hematology Analyzer
03red CAN_H
Socket J37
01blac k CAN_L
02
04
Label: B06-P37
Fig. 5-1-4 Communication connection diagram of the Analyzer (model without CRP)
59
Service Manual of Automatic Hematology Analyzer
(1) Finish red blood cell and platelet test data collection (with electric impedance
Analog signal collection method) and hemoglobin test data collection (with colorimetric method)
board (2) Finish white blood cell test data collection (with laser scattering method)
(3) Read the output data of one thermal resistor
Large angle
(1) Finish laser scattering large angle optical signal collection and transmit to
optoelectronic receiving
simulation signal collection board
board
Small angle
(1) Finish laser scattering small angle optical signal collection and transmit to
optoelectronic receiving
simulation signal collection board
board
60
Service Manual of Automatic Hematology Analyzer
Optical (1) Finish laser scattering 45° optical signal collection and transmit to simulation
pre-amplification board signal collection board
(1) Drive laser tube to transmit laser light
Laser tube drive board
(2) Have constant control of laser power
45° photocell mounting
(1) Fix and install photocell
plate
FPC Socket
Socket J1 01 DGND
02 +3.3V
03 TFT_CLKOUT+
04 +3.3V
05 TFT_CLKOUT-
06 +3.3V
07 DGND
08 +3.3V
09 TFT_Y2+
10 DGND
11 TFT_Y2-
12 DGND
13 DGND LCD touch screen
14 DGND
15 TFT_Y1+
16 DGND
17 TFT_Y1-
Middle computer 18 DGND
board
19 DGND
20 DGND
21 TFT_Y0+
22 DGND
23 TFT_Y0-
24 DGND
25 DGND
26 TFT_BIST
Socket J2 01 +12V
02 TFT_BL_EN
03 +12V
04 TFT_BL_PWM
05 +12V
06 DGND
Socket J19 03red SIN_2 01red SIN_2 01red SIN_2 01red SIN_2 01red SIN_2 01red SIN_2 STAT but ton switch
04black GND 02black GND 02black GND 02black GND 02black GND 02black GND
01 Label: P0201 Label: J0201 Label: PSW0208 Label: JSW0208 Button Swit ch
02
Label: B01-P19
61
Service Manual of Automatic Hematology Analyzer
05 05 Label: B05-P15 12black DGND 02black DGND 02black DGND black DGND 2 1
Step Motor Label: JM0202 Label: PM0202 13 Label: PSW0203 Label: JSW0203
14
Co de sc anning
moto r yellow OU T7_Y- 01yellow OUT7_Y- 01red OUT7_Y- 01red OUT7_Y- 15red SIN_4 01red SIN_4 01blue SIN_4 blue SIN_4 ST AT i n-po sition swi tch
Socket J14
red OUT7_Y+ 02red OUT7_Y+ 02white OUT7_Y+ 02white OUT7_Y+ 16black DGND 02black DGND 02black DGND black DGND 2 1
orange OUT7_X- 03orange OUT7_X- 03blue OUT7_X- 03blue OUT7_X- Label: B05-P19 Label: PSW0204 Label: JSW0204
brown OUT7_X+ 04brown OUT7_X+ 04black OUT7_X+ 04black OUT7_X+
Rec laim ing a re a ful l
05 05 Label: B05-P14 01red +5V 01red +5V 01 red +5V o pto coupl er
Step Motor Label: JM0203 Label: PM0203 01red +5V 02black DGND 02black DGND 03 black DGND
Socket J18
02black DGND 03blue SIN_5 03blue SIN_5 02 blue SIN_5 Photosensor
Fe eding m oto r yellow OU T5_Y- 01yellow OUT5_Y- 01red OUT5_Y- 01red OUT5_Y- 03blue SIN_5 Label: PPS0201 Label: JPS0201
Socket J13
red OUT5_Y+ 02red OUT5_Y+ 02white OUT5_Y+ 02white OUT5_Y+ 04
orange OUT5_X- 03orange OUT5_X- 03blue OUT5_X- 03blue OUT5_X- 05 01red SIN_6 01blue SIN_6 blue SIN_6 Co unt bal l 1 (rig ht )
brown OUT5_X+ 04brown OUT5_X+ 04black OUT5_X+ 04black OUT5_X+ 06 02black DGND 02black DGND black DGND 2 1
05 05 Label: B05-P13 07red SIN_6 Label: PSW0206 Label: JSW0206
Step Motor Label: JM0204 Label: PM0204 08black DGND
09 01red SIN_7 01blue SIN_7 blue SIN_7 Co unt bal l 2 (le ft)
Unlo a ding
moto r yellow OU T3_Y- 01yellow OUT3_Y- 01red OUT3_Y- 01red OUT3_Y- 10 02black DGND 02black DGND black DGND 2 1
Socket J12
red OUT3_Y+ 02red OUT3_Y+ 02white OUT3_Y+ 02white OUT3_Y+ 11red SIN_7 Label: PSW0207 Label: JSW0207
orange OUT3_X- 03orange OUT3_X- 03blue OUT3_X- 03blue OUT3_X- 12black DGND
Ra ck-fee di ng in-
brown OUT3_X+ 04brown OUT3_X+ 04black OUT3_X+ 04black OUT3_X+ 13red +5V 01red +5V 01red +5V 01 red +5V po sitio n o ptoc oupl er
05 05 Label: B05-P12 14black DGND 02black DGND 02black DGND 03 black DGND
Step Motor Label: JM0205 Label: PM0205 15blue SIN_8 03blue SIN_8 03blue SIN_8 02 blue SIN_8 Photosensor
16 Label: PPS0202 Label: JPS0202
Pho toe le ctri c
switc h black PH 01black PH 01blue PH 01blue PH Label: B05-P18
Socket J9
brown +24V 02brown +24V 02red +24V 02red +24V
Ra ck-co nv ey ing z ero-po int
Photo Switch blue DGND 03blue DGND 03black DGND 03black DGND 01red +5V 01brown +5V brown +5V o pto el ec tr oni c switch
Label: JSW0202 Label: PSW0202 Label: B05-P09 02black DGND 02black DGND black DGND
01red +5V 03blue SIN_9 03white SIN_9 white SIN_9
Bar code reader wh ite TXD232_2 01white TXD232_2 01red TXD232_2 01red TXD232_2
Sample feeder Socket J8
02black DGND 04white DGND 04blue GND blue GND Photo Switch
purple DGND 02purple DGND 02black DGND 02black DGND
Socket J3 control board 03blue SIN_9 Label: PSW0205 Label: JSW0205
green RXD232_2 03green RXD232_2 03blue RXD232_2 03blue RXD232_2 04white DGND
Co de sc anni ng zer o-
black CC_IN 04black CC_IN 04red CC_IN 04red CC_IN 05red +5V 01red +5V 01red +5V white +5V po int o ptoc oupl er
red +5V 05red +5V 05blue +5V 05blue +5V 06black DGND 02black DGND 02black DGND black DGND
Barcode Reader orange CC_OUT 06orange CC_OUT 06black CC_OUT 06black CC_OUT 07blue SIN_10 03blue SIN_10 03blue SIN_10 blue SIN_10 Photosensor
Label: JBR0201 Label: PBR0201 Label: B05-P03 08 Label: PPS0203 Label: JPS0203
09red +5V
Ra ck-withdra wing
01red OUT1_Y- 10black DGND ze ro-po int o ptoc oupl er
Socket J16
02white OUT1_Y+ 11blue SIN_11 01red +5V
03blue OUT1_X- 12 02blue SIN_11
04black OUT1_X+ 13red +5V 03black DGND
Clamping
motor blue OUT1_Y- 01blue OUT1_Y- 01red OUT1_Y- 01red OUT1_Y- 01red OUT1_Y- 01red OUT1_Y- 01red OUT1_Y- Label: B05-P16 14black DGND Label: PPS0206 Photosensor
red OUT1_Y+ 02red OUT1_Y+ 02yellow OUT 1_Y+ 02yellow OU T1_Y+ 02white OUT1_Y+ 02white OUT1_Y+ 02white OUT1_Y+ 15blue SIN_12
green OUT1_X- 03green OUT1_X- 03blue OUT1_X- 03blue OUT1_X- 03blue OUT1_X- 03blue OUT1_X- 03blue OUT1_X- 01red OUT4_Y- 16
Socket J10 Ra ck-co nvey ing i n-
black OUT1_X+ 04black OUT1_X+ 04green OUT1_X+ 04green OUT1_X+ 04black OUT1_X+ 04black OUT1_X+ 04black OUT1_X+ 02white OUT4_Y+ Label: B05-P08 01red +5V 01red +5V 01 red +5V positio n o ptoc o upl er
05 05 05red +5V 05red +5V 09red +5V 05red OUT4_Y- 03blue OUT4_X- 02black DGND 02black DGND 03 black DGND
Step Motor Label: JM0401 Label: PM0401 06yellow DGND 06white DGND 10white DGND 06white OUT4_Y+ 04black OUT4_X+ 03blue SIN_12 03blue SIN_12 02 blue SIN_12 Photosensor
07blue SIN_16 07blue SIN_16 11blue SIN_16 07blue OUT4_X- Label: B05-P10 Label: PPS0204 Label: JPS0204
Clamping zero-point
optocoup ler red A 01red A 01red +5V 08green DGND 08black DGND 12black DGND 08black OUT4_X+ 01
Socket J2
bla ck K 02black K 02yellow DGN D Label: J0402 Label: P0402 05red OUT4_Y- 09red +5V 01red +5V 02
Socket J7
whi te C 03wh ite C 03blue SIN_16 06white OUT4_Y+ 10black DGND 02black DGND 03
Photosensor green E 04green E 04green DGND 07blue OUT4_X- 11white SIN_16 03white SIN_16 04
ST AT tube judg ing
Lab el: JPS0401 Label: PPS0401 08black OUT4_X+ 12blue DGND 04blue DGND 05red +5V 01red +5V 01red +5V 01 red +5V o ptoco upl er
17red +24V 13red +5V 05red +5V 06black DGND 02black DGND 02black DGND 03 black DGND
Liftin g motor black OUT4_Y- 01black OUT4_Y- 01red OUT4_Y- 18black C 14black DGND 06black DGND 07blue SIN_18 03blue SIN_18 03blue SIN_18 02 blue SIN_18 Photosensor
green OUT4_Y+ 02green OUT4_Y+ 02white OUT4_Y+ 13red +5V 15white SIN_15 07white SIN_15 08 Label: PPS0205 Label: JPS0205
red OUT4_X- 03red OUT4_X- 03blue OUT4_X- 14black DGND 16blue DGND 08blue DGND 09
blue OUT4_X+ 04blue OUT4_X+ 04black OUT4_X+ 15white SIN_15 17red +24V 09 10
white +24V 05white +24V 05red +24V 16blue DGND 18black C 10 11
Step Motor white C 06white C 06black C 19 19 11 12
Label: JM0402 Label: PM0402 20 20 12 13
Label: P0401 Label: J0401 13 14
Liftin g zero-point
optocoup ler re d A 01red A 01red +5V 14 15
bla ck K 02black K 02black DGND 15 16
whi te C 03wh ite C 03white SIN_15 16 Label: B05-P02
Photosensor gree n E 04green E 04blue DGND Label: B05-P07
Label: JPS0402 Label: PPS0402
01red +24V
Socket J21
03black C
02
Label: B05-P21
5.3.3 Electrical wiring diagram of main control board 01Red +24V 01Red +24V Red +24V No. 1 sole noi d
Overheat va lve
12
protector 02Black OUT_L36 02Black OUT_L36 Black OUT_L36 Valve
Heat ing rod
Label: JV0801 Label: PV0801
Whi te gre en Wh ite +24V Wh ite +24V
+2 4V 01Red +24V 01Red +24V 01Red +24V Socket J28
Whi te gre en Black D GND No. 2 sole noi d
GND 0 2B la ck DGN D 0 2B lack DGND 02B lack D GND 01Red +24V 01Red +24V Red +24V va lve
12
Label: JH1001 Label: PH1001 Label: B06-P28 01Black OUT_L36 02Black OUT_L35 02Black OUT_L35 Black OUT_L35 Valve
Socket J21
02Red +24V Label: JV0802 Label: PV0802
Black +5V 01Red +5V 01Red +5V 01Red +5V 03Black OUT_L35
Thermistor Socket J15 No. 3 sole noi d
Black TEMP_IN1 02Black TEMP_IN1 02Black TEMP_IN1 02Black TEMP_IN1 04Red +24V 01Red OUT_H36 01Red OUT_H36 Red OUT_H36 va lve
12
Label: JS1001 Label: PS1001 03 05Red OUT_H36 02Black DGND 02Black DGND Black DGND Valve
Label: B06-P15 06Black DGND Label: JV0803 Label: PV0803
07Red OUT_H35
Lif ting motor Red OUT5_Y- 01Red OUT5_Y- 01Red OUT5_Y- 01Red OUT5_Y- 08Black DGND 01Red OUT_H35 01Red OUT_H35 Red OUT_H35 No. 10 sole noid valve
123
Socket J22 Valve
RedWh ite OUT5_Y+ 02RedWh ite OUT5_Y+ 02Blue OUT5_Y+ 02Blue OUT5_Y+ 09Black OUT_L34 02Black DGND 02Black DGND Black DGND
Green white OUT5_X- 03Green w hite OUT5_X- 03White OUT5_X- 03White OUT5_X- 10Red +24V Label: JV0810 Label: PV0810
Green OUT5_X+ 04Green OUT 5_X+ 04Black OUT5_X+ 04Black OUT5_X+ 11Red OUT_H34
05 05 Label: B06-P22 12Black DGND 01Red +24V 01Red +24V Red +24V No. 11 sole noid valve
123
Step Motor Label: JM0301 Label: PM0301 13Black OUT_L33 02Black OUT_L34 02Black OUT_L34 Black OUT_L34 Valve
14Red +24V Label: JV0811 Label: PV0811
Rotating motor Red OUT4_Y- 01Red OUT4_Y- 01Red OUT4_Y- 01Red OUT4_Y- Label: B06-P21
Socket J23
Blue OUT4_Y+ 02Blue OUT4_Y+ 02Blue OUT4_Y+ 02Blue OUT4_Y+ 01Red OUT_H34 01Red OUT_H34 Red OUT_H34 No. 15 sole noid valve
12
Green OUT 4_X- 03White OUT4_X- 03White OUT4_X- 03White OUT4_X- 02Black DGND 02Black DGND Black DGND Valve
Black OUT4_X+ 04Black OUT4_X+ 04Black OUT4_X+ 04Black OUT4_X+ Label: JV0815 Label: PV0815
05 05 Label: B06-P23
Step Motor Label: JM0302 Label: PM0302 01Red +24V 01Red +24V Red +24V No. 16 sole noid valve
12
02Black OUT_L33 02Black OUT_L33 Black OUT_L33 Valve
Radiator fan of rack Red +24V 01Red +24V 01Red +24V 01Red +24V Label: JV0816 Label: PV0816
body unit Socket J32
Black D GN D 0 2B la ck DGN D 0 2B lack DGND 02B lack D GND
Label: JF0101 Label: PF0101 03 01Red OUT_H33 01Red OUT_H33 Red OUT_H33 No. 12 sole noid valve
123
Label: B06-P32 02Black DGND 02Black DGND Black DGND Valve
Label: JV0812 Label: PV0812
Fan 01 01Red OUT_H33
S ocket J8 Socket J20
02 02Black DGND 01Red OUT_H32 01Red OUT_H32 Red OUT_H32 No. 13 sole noid valve
Connect with
12
03 03Red OUT_H32 02Black DGND 02Black DGND Black DGND Valve
waste liquid tank 04 04Black DGND Label: JV0813 Label: PV0813
float switch 01Red SIN_3 01Red SIN_3 01Red SIN_3 05Red SIN_3 05Black OUT_L32
02Black DGND 02Black DGND 1 3 3 1 02Black DGND 06Black DGND 06Red +24V 01Red +24V 01Red +24V Red +24V No. 14 sole noid valve
Label: PSW0101 03 2 03 07 07Red OUT_H31 02Black OUT_L32 02Black OUT_L32 Black OUT_L32 Valve 12
2
Label: P0101 Label: J0101 08 08Black DGND Label: JV0814 Label: PV0814
Label: B06-P08 09Black OUT_L31
10Red +24V 01Red OUT_H31 01Red OUT_H31 Red OUT_H31 No. 28 sole noid valve
12
Black +5V 01Black +5V Label: B06-P20 02Black DGND 02Black DGND Black DGND Valve
Ther mistor Socket J13
Black TEMP_IN3 02Black TEMP_IN3 Label: JV0828 Label: PV0828
03
Label: B06-P13 01Red +24V 01Red +24V Red +24V No. 31 sole noid valve
12
Label: JM1401 Label: PM1401 02Black OUT_L31 02Black OUT_L31 Black OUT_L31 Valve
2.5 and 10ML duplex syringe Red OUT1_Y- 01Red OUT1_Y- 01Red OUT1_Y- 01Red OUT1_Y- Label: JV0831 Label: PV0831
assembly Socket J26
Yellow OUT1_Y+ 02Yellow OUT1_Y+ 02Blue OUT1_Y+ 02Blue OUT1_Y+
No. 4 sole noi d
Brown OUT1_X- 03Brown OUT1_X- 03White OUT1_X- 03White OUT1_X- 01Red +24V 01Red +24V Red +24V va lve
12
Orange OUT1_X+ 04Oran ge OU T1_X+ 04Black OUT1_X+ 04Black OUT1_X+ 02Black OUT_L26 02Black OUT_L26 Black OUT_L26 Valve
05 05 Label: B06-P26 Label: JV0804 Label: PV0804
Step Motor 06 06
No. 5 sole noi d
01Black OUT_L26 01Red +24V 01Red +24V Red +24V va lve
Socket J19
12
Red A 01Red A 01Red A 02Red +24V 02Black OUT_L25 02Black OUT_L25 Black OUT_L25 Valve
Black K 02Black K 02Black K 03Black OUT_L25 Label: JV0805 Label: PV0805
White C 03White C 03White C 04Red +24V
No. 7 sole noi d
Green E 04Green E 04Blue E 05 01Red OUT_H25 01Red OUT_H25 Red OUT_H25 va lve
12
Photosensor Label: JPS1401 Label: PPS1401 01Red +5V 06 02Black DGND 02Black DGND Black DGND Valve
S ocket J5
02Black DGND 07Red OUT_H25 Label: JV0807 Label: PV0807
03White S IN_13
Main control board 08Black DGND
Zero optocoupler of 04Blue DG ND 09Black OUT_L24 01Red +24V 01Red +24V Red +24V No. 8 sole noi d
lifting motor va lve
12
01Red +5V 01Red +5V 01Red +5V 05Red +5V 10Red +24V 02Black OUT_L24 02Black OUT_L24 Black OUT_L24 Valve
02Blue SIN_6 02Blue SIN_6 02Blue SIN_6 06Black DGND 11Red OUT_H24 Label: JV0808 Label: PV0808
03Black DGND 03Black DGND 03Black DGND 07Blue SIN_6 12Black DGND
No. 9 sole noi d
Photosensor Label: PPS0301 04 04 08 13Black OUT_L23 01Red OUT_H24 01Red OUT_H24 Red OUT_H24 va lve
12
Label: J0301 Label: P0301 09 14Red +24V 02Black DGND 02Black DGND Black DGND Valve
10 Label: B06-P19 Label: JV0809 Label: PV0809
Position judging 11
optocoupler of rotating 12 01Red +24V 01Red +24V Red +24V No. 2 9 soleno id v alv e
123
motor Valve
01Red +5V 01Red +5V 01Red +5V 13Red +5V 02Black OUT_L23 02Black OUT_L23 Black OUT_L23
02Blue SIN_8 02Blue SIN_8 02Blue SIN_6 14Black DGND Label: JV0829 Label: PV0829
03Black DGND 03Black DGND 03Black DGND 15Blue SIN_8
Photosensor Label: PPS0302 04 04 16 01Red OUT_H23 01Red OUT_H23 Red OUT_H23 No. 17 soleno id v alv e
12
Label: J0302 Label: P0302 Label: B06-P05 02Black DGND 02Black DGND Black DGND Valve
Label: JV0817 Label: PV0817
Label: JM0901 Label: PM0901 01Red OUT_H23
Socket J18
Dispensing syringe Brown OUT3_Y- 01Brown OUT3_Y- 01Red OUT3_Y- 01Red OUT3_Y- 02Black DGND 01Red OUT_H22 01Red OUT_H22 Red OUT_H22 No. 18 soleno id v alv e
Socket J24
12
Orange OUT3_Y+ 02Oran ge OU T3_Y+ 02Blue OUT3_Y+ 02Blue OUT3_Y+ 03Red OUT_H22 02Black DGND 02Black DGND Black DGND Valve
Red OUT3_X- 03Red OUT3_X- 03White OUT3_X- 03White OUT3_X- 04Black DGND Label: JV0818 Label: PV0818
Yellow OUT3_X+ 04Yellow OUT3_X+ 04Black OUT3_X+ 04Black OUT3_X+ 05Black OUT_L22
05 05 Label: B06-P24 06Red +24V 01Red +24V 01Red +24V Red +24V No. 19 soleno id v alv e
12
Step Motor 06 06 07Red OUT_H21 02Black OUT_L22 02Black OUT_L22 Black OUT_L22 Valve
08Black DGND Label: JV0819 Label: PV0819
Red A 01Red A 01Red +5V 01Red +5V 09Black OUT_L21
S ocket J7
Black K 02Black K 02Black DGND 02Black DGND 10Red +24V 01Red OUT_H21 01Red OUT_H21 Red OUT_H21 No. 2 0 soleno id v alv e
12
White C 03White C 03White SIN_5 03White SIN_5 Label: B06-P18 02Black DGND 02Black DGND Black DGND Valve
Green E 04Green E 04Blue DGND 04Blue DG ND Label: JV0820 Label: PV0820
Photosensor Label: JPS0901 Label: PPS0901 05Red +5V
06Black DGND 01Red +24V 01Red +24V Red +24V No. 3 0 soleno id v alv e
12
62
Service Manual of Automatic Hematology Analyzer
5.3.4 Electrical wiring diagram of CRP test unit board (model with CRP) 01Red +24V 01Red +24V Black +24V
No. 23 solenoid
valve
123
02Black OUT_L6 02Black OUT_L6 Black OUT_L6 Valve
CRP syringe 01Red OUT2_Y- 01Red OUT2_Y- 01Red OUT2_Y- 01Black OUT_L6 Label: JV0823 Label: PV0823
So ck et J4 So ck et J6
Red OUT2_Y- 02Yellow OUT2_Y+ 02Blue OUT2_Y+ 02Blue OUT2_Y+ 02Red +24V
No. 24 solenoid
Yellow OUT2_Y+ 03Brown OUT2_X- 03White OUT2_X- 03White OUT2_X- 03Black OUT_L5 01Red +24V 01Red +24V Black +24V valve
123
Brown OUT2_X- 04Oran ge OU T2_X+ 04Black OUT2_X+ 04Black OUT2_X+ 04Red +24V 02Black OUT_L5 02Black OUT_L5 Black OUT_L5 Valve
Orange OU T2_X+ 05 05 Label: B02-P04 05 Label: JV0824 Label: PV0824
Step Motor 06 06 06
No. 26 solenoid
Label: JM1701 Label: PM1701 07Red OUT_H5 01Red OUT_H5 01Red OUT_H5 B lack O UT _H5 valve
123
08Black D GN D 02Black DGN D 02Black D GND Black DGN D Valve
Red A 01Red A 01Red +5V 01Red +5V 09Black OUT_L4 Label: JV0826 Label: PV0826
So ck et J15
Black K 02Black K 02Black DGND 02Black DGND 10Red +24V
No. 27 solenoid
White C 03White C 03White SIN_4 03White SIN_4 11 01Red +24V 01Red +24V Black +24V valve
123
Photosensor Green E 04Green E 04Blue DGND 04Blue DGND 12 02Black OUT_L4 02Black OUT_L4 Black OUT_L4 Valve
Label: JPS1701 Label: PPS1701 05Red +5V 13 Label: JV0827 Label: PV0827
06Black DGND 14
Refri geratio n c ham ber
cover ope ning optoc oupl er Red A 07Blue SIN_6 Label: B02-P06
Bl ac k K 01Red A 07Black DGND 08
White C 02Black K 06Blue SIN_6 09Red +5V 09
So ck et J5
Photosensor Green E 03White C 05Red +5V 10Black DGND 10
04Green E 01Red A 11White SIN_2 01Red OUT_H3
No. 21 solenoid
Hall switch Red +5V 05Red +5V 02Black K 12Blue DGND 02Black D GN D 01Red OUT_H3 01Red OUT_H3 B lack O UT _H3 valve
123
Yellow S IN_6 06Yellow SIN_6 03White C 13 03Red OUT_H2 02Black DGN D 02Black D GND Black DGN D Valve
Hall Switch Black DGN D 07Black DGND 04Blue E 14 04Black D GN D Label: JV0821 Label: PV0821
Label: JPS1101 Label: PPS1101 15 05
No. 22 solenoid
16 06 01Red OUT_H2 01Red OUT_H2 B lack O UT _H2 valve
123
Label: B02-P15 07 02Black DGN D 02Black D GND Black DGN D Valve
08 Label: JV0822 Label: PV0822
Black +5V 01Black +5V 01Red +5V 01Red +5V Label: B02-P05
Thermistor So ck et J10
Black TEMP_IN2 02Black TEMP_IN2 02Black TEMP_IN2 02Black TEMP_IN2 3# pump
Label: JS1201 Label: PS1201 03
Label: B02-P10 CRP test unit board Pump
Overheat 01Red +24V 01Red +24V 01Red +24V Red +
Heating tape So ck et J18
protector 02B lack DGND 02B lack DGND 02B lack DGND Black -
Label: B02-P18 Label: PP0803 Label: JP0803
Red +24V White +24V White +24V 01Red +24V 01Red +24V 01Red +24V So ck et J1
Red GND Red DGND 02B lack DGND 02B lack DGND 02B lack DGND Red +24V Fan
Label: JH1201 Label: PH1201 Label: B02-P01 Black D GN D
Overheat
Heating tape
protector 01Red +24V
So ck et J17
02B lack DGND
Red +24V White +24V White +24V 01Red +24V 01Red +24V 01Red +24V So ck et J23 Label: B02-P17 01Red +24V 01Red +24V Fan
Red GND Red DGND 02B lack DGND 02B lack DGND 02B lack DGND 02Black DGND 02Black DGND
Label: JH1202 Label: PH1202 Label: B02-P23 01Red +5V 03Red +5V 03Black +5V Black +5V
So ck et J9 Thermistor
02Black TEMP_IN1 04Black TEMP_IN1 04Black TEMP_IN1 Black TEMP_IN1
Black +5V 01Black +5V 01Red +5V 01Red +5V 03 05Red OUT1_Y- 05Blue OUT1_Y-
Thermistor So ck et J21 Refri g era tion c ham ber
Black TEMP_IN3 02Black TEMP_IN3 02Black TEMP_IN3 02Black TEMP_IN3 Label: B02-P09 06Blue OUT1_Y+ 06Red OUT1_Y+ Blue OUT1_Y- cov er ope ning m otor
Label: T1 03Black +5V 03Blue +5V 03 07White OUT1_X- 07Green OUT1_X- Red OUT1_Y+
04Black TEMP_IN4 04White TEMP_IN4 Label: B02-P21 01Red OUT1_Y- 08Black OUT1_X+ 08Black OUT1_X+ Green OUT1_X-
So ck et J3
Black +5V Label: JS1202 Label: PS1202 02Blue OUT1_Y+ Label: PF1101 Label: JF1101 Black OUT1_X+
Thermistor
Black TEMP_IN4 01Blue +5V 03White OUT1_X-
So ck et J22
Label: T2 02White TEMP_IN4 04Black OUT1_X+ Step Motor
03 Label: B02-P03
Label: B02-P22
Peltier (1) Peltier (2)
Peltier Peltier
So ck et J19 01Red +24V 01Red +24V 01Red +24V Red +24V Black +24V Red +24V
02B lack DGND 02B lack DGND 02B lack DGND Black GND Black GND
Label: B02-P19 Label: PPR1101 Label: JPR1101
Fig. 5-3-4 Electrical wiring diagram of CRP test unit board (model with CRP)
5.3.5 Electrical wiring diagram of CRP unit collection board (model with CRP )
01black GND(shielded) yellow A
SocketP1
02green PD_K
03yellow PD_A Optical receiving
Label:B14-P01 mounting plate
green K
01red LED_A
SocketP2
02
03
04black LED_K
Label:B14-P02
01red A
02black K
01red A
02black K
Label:G1201 Label:G1202
RV1.5yellowgreen
Fig. 5-3-5 Electrical wiring diagram of CRP unit collection board (model with CRP)
63
Service Manual of Automatic Hematology Analyzer
02Red A
03Black K
Red A 04Green LED-
05Yellow LED+
1
Label: B20-P03
Yell ow A
Light emittin g
diode
Green K
64
Service Manual of Automatic Hematology Analyzer
Sample-aspirating
syringe Tightening screw
Sheath syringe
Rinsing Syringe
CRP Syringe
Tightening screw
Fig. 6-1-1
6.1.2 Waste Pump Replacement
Since waste liquid pump and positive-pressure pump may be worn by testing samples and diluent to some
degree during the test, the waste liquid pump shall be replaced when damaged after being used for a long time. The
operation is described as follows:
(1) As shown in Fig. 6-1-2, the waste liquid assembly can be taken down by removing the four fastening screws.
65
Service Manual of Automatic Hematology Analyzer
Tightening
screw
Fig. 6-1-2
6.1.3 Replacement of sampling probe
During the instrument testing, the probe tip will get worn by the puncturing movement to some degree, so that it
is recommended to replace the sampling probe when the sampling tests are up to 30,000 times as per the methods
below:
(1) First unscrew the two screws 1 & 2 of the sampling probe and draw out the sampling probe upward vertically
as shown in the figure below:
Swab
Fig. 6-1-3
Screw off the connectors on the upper part of the sampling probe to replace with a new sampling probe as shown
in the following figure:
Fig. 6-1-4
Insert the new sampling probe vertically into the swab (as shown in Fig. 6-1-4) and fix the screws 1 & 2 (as
shown in Fig. 6-1-3).
66
Service Manual of Automatic Hematology Analyzer
Tightening screw of
probe
Mounting
seat
Sampling
probe
The sampling probe is fixed on the mounting base and it can only be adjusted up and down by adjusting the
position of probe optocoupler partition, as shown in Fig. 6-2-2.
.
67
Service Manual of Automatic Hematology Analyzer
Fastening screw of
optocoupler
Fig. 6-2-2
Vertically insert the new sampling probe in the swab and observe the position of aspiration hole of the sampling
probe. When it is connected with power, the position of the sampling probe when Y-direction slide block stops
moving upward shall be that the water outlet of the sampling probe shall be above the center of the upper outlet of the
swab (as shown in Fig. 6-2-3). If the position is wrong, adjust the relative position of the optocoupler and the partition
and then fix the optocoupler fastening screws.
Fig. 6-2-3
68
Service Manual of Automatic Hematology Analyzer
Adjusting button
Fig. 6-2-4
Blade
Fig 6-2-5
(2) As shown in Figure 6-2-6, insert the leaf spring into the front end of the reagent bottle holder; as shown in
Figure 6-2-7, close the refrigeration chamber door and observe the state of the refrigeration chamber door in the
state of the display switch. At this time, if "refrigeration chamber door closed" appears, then it is qualified.
Otherwise, adjust the screw in the position in Figure 6-2-8 until it is qualified; check twice according to the above
steps.
69
Service Manual of Automatic Hematology Analyzer
Leaf Spring
Adjusting Screws
Fig 6-2-8
6.2.4 Refrigeration chamber cover optocoupler encoding disk adjustment
If the position of the refrigeration chamber cover’s optocoupler encoding disk changes, then the position of
optocoupler encoding disk needs to be adjusted. As shown in Figure 6-2-9, tighten the outermost top line of the
small corner of the optocoupler encoding disk to the bottom surface of the optocoupler encoding disk installing
block, and then tighten the two screws fixing the encoding disk.
encoding disk
installing block
Adjusting
Screws
Fig 6-2-9
70
Service Manual of Automatic Hematology Analyzer
Chapter7 Maintenance
7.1 Overview
To guarantee the effective operation and lengthen the service life of the instrument, users shall have routine
maintenance for the instrument according to requirements. Although the instrument will automatically give prompt
to remind users to execute corresponding maintenance after testing a certain amount of samples or continuously
operating for a certain time, and the software system is provided with a special service menu which will show the
measures of routine maintenance or in case of failure, users shall make appropriate maintenance and servicing plan
as per specific laboratory conditions, such as the daily sample volume, operating environment and operating time of
the instrument, to reduce factors that affect analysis accuracy and to guarantee the normal operation of instrument
system.
● Don't spill reagent, sample or waste liquid on the mechanical or electrical parts of the instrument in case
of any damages.
● During operation, operators shall take preventive measures like wearing protective gloves and working
clothes, or they may be infected after touching the polluted area or solution. Please wash with water and take
disinfection measures if touching any polluted liquid.
● During rinsing and maintenance, please check whether parts containing liquid lose efficiency as they may
cause hazards.
● The instrument is installed with a laser. Operator shall check the warning labels during maintenance and
servicing process, but not stare forward the light beam directly or through the optical instrument. Please
wear protective glasses when looking straight at laser source.
● Improper maintenance may cause instrument damage. Operators shall maintain the instrument according
to the user manual.
● If the instrument is installed with accessories which are not provided or recommended by the
manufacturer or the instrument is used otherwise specified by the manufacturer, the relating protection may
be weakened.
● During maintenance process, please check whether there are hazards caused by inefficiency of hoses or
parts filled with solution.
● For the problems without solutions in the User Manual, please contact the after-sales service department
of the manufacturer to obtain suggestions from professional personnel.
71
Service Manual of Automatic Hematology Analyzer
If the instrument is shut down or opened normally, execute [Rinse sample-aspirating probe] every month (execute
through selecting [Menu]→[Maint.]→[Service]→[Rinse]).
7.2.2 Timely maintenance
Users can set automatic rinsing interval in "Setting" (setting range: 10-200 times), and the instrument will
automatically execute rinsing after reaching the set test number.
If sheath flow cell is polluted, execute [Maint.]- [Service] - [Rinse] - [Sheath flow cell] to execute the rinsing
operation.
When instrument is left unused for over 2 weeks, replace reagent with distilled water, and click
[Menu]→[Maint.]→[Service]→[Maintenance] - [Package rinsing], then interrupt distilled water supply, and
execute [Package draining], after finishing the two operation above, place the instrument at a clean and dry
position.
When instrument gives prompt "blocking", press [Menu]→[Maint.]→[Service]→[Maintenance] - [Unblock] to
execute unblocking operation or execute [Cauterize] and [Back washing].
When instrument is left unused for a long time, after startup, select →[Maint.]→[Service]→[Filling]
[Menu]
successively and execute the operation of [BF-5D diluent], [BF-FDOⅠ lyse], [BF-FBH lyse], [BF-FDT I lyse], and
[BF-CRP lyse].
When the instrument works under a normal working environment and state, the analysis result is reliable.
If other failure prompt is given, please refer to troubleshooting table.
7.3 Maintenance
7.3.1 Basic status
Click [Menu] - [Maint.] - [Basic status], as shown in the figure below:
Fig. 7-3-1
The followings are the description about the basic state in the figure above:
(1)Temperature: display the ambient temperature, pre-heating cell temperature, CRP refrigeration temperature,
CRP reaction cell temperature, colorimetric ware temperature and the normal range of each temperature.
(2)Pressure: display the positive pressure, negative pressure, the pressure of sheath flow cell and normal pressure
range.
72
Service Manual of Automatic Hematology Analyzer
(3)Voltage: display the voltage value in small holes of red blood cell and the normal voltage range.
(4)Current: display the laser current and normal current range.
7.3.2 System version
Select [Menu]-[Maint.]-[SY Ver.] to view the version information of each circuit board. The interface is as shown
in the figure below:
Fig. 7-3-2
Version number in the figure is for reference only, the version number of software will change with the
upgrading of the program.
7.3.3 Service
The software provides many maintenance functions to guarantee the accurate and effective operation.
7.3.3.1 Replace/ fill
Conduct background test after replacing diluent and lyse, and execute sample test when background test result is
within the normal range. Click [Menu] - [Maint.] - [Services] and select the [Replace]/[Fill] tab.
(1)When it is necessary to replace some reagent, proceed as follows:
73
Service Manual of Automatic Hematology Analyzer
Fig. 7-3-3
Fig. 7-3-4
74
Service Manual of Automatic Hematology Analyzer
When software gives prompt of no BF-FDO I, the instrument will automatically fill BF-FDO I lyse after
corresponding failure information in "Fault information area" is eliminated.
7.3.3.2 Rinse
Click [Menu] - [Maint.] - [Service] and select the [Rinse] tab, as shown in the figure below:
Fig. 7-3-5
(1)If WBC, HGB or WBC classification related parameters are found abnormal in background test, execute "Rinse
WBC cell" operation. Click [WBC cell] and it will prompt "Rinse WBC cell". The rinsing is finished when the
progress bar comes to an end.
(2)When abnormal parameter value of background test relevant to RBC and PLT is founded, execute "Rinse RBC
cell" operation. Click [RBC cell] and it will prompt "Rinse RBC cell". The rinsing is finished when the progress
bar comes to an end.
(3)To avoid the interference of count result by the pollution of sampling parts, rinse the sampling probe and swab
after every month of instrument operation. Click "Sampling probe" to execute the "Rinse sampling probe"
operation and it will prompt "Rinsing sampling probe". The rinsing is finished when the progress bar comes to an
end.
The swab rinsing steps are same with that for sampling probe.
(4)When abnormally expanded cell population appears in scattergram, and the value of parameter background test
relevant to WBC is high, there may be air bubbles in flow cell. Execute the "Rinse sheath flow cell" operation,
click [Sheath flow cell] and it will prompt "Rinse sheath flow cell". The rinsing is finished when the progress bar
comes to an end.
(5)If CRP parameters are found abnormal in test, execute "Rinse CRP1 cell" or "Rinse CRP2 cell" operation, click
"CRP1 cell" and it will prompt "Rinse CRP1 cell". The rinsing is finished when the progress bar comes to an end.
75
Service Manual of Automatic Hematology Analyzer
The CRP2 cell rinsing steps are same with that for CRP1 cell.
(6)If the instrument is not shut down for over 24 hours and "Rinse tubing" is required to be conducted every day,
execute "Rinse tubing" at this moment.
Click [Tubing] and it will prompt "Rinse tubing". The rinsing is finished when the progress bar comes to an end.
7.3.3.3 Maintenance
Click [Menu] - [Maint.] - [Service] - [Maintenance], as shown in the figure below:
Fig. 7-3-6
(1)Drainage
During the maintenance of pipeline system of the instrument, execute drainage before maintenance to prevent the
spill of liquid, the followings are the specification, taking draining waste liquid disk as example:
a)Draining waste liquid bottle can drain the liquid in waste liquid bottle and tubing. Click [Waste bottle] and it
will prompt "Drain waste liquid bottle". The draining is finished when the progress bar comes to an end.
The operation to drain "WBC cell", "RBC cell", "CRP1 cell" and "CRP2 cell" is the same as above.
b)If the instrument will not be used for 3 days- 2 weeks, the tubing should be drained. Click [Tubing] and it will
prompt "Drain tubing" query information, as shown in figure below.
Fig. 7-3-7
Click [OK] and a prompt box shown below will pop up:
76
Service Manual of Automatic Hematology Analyzer
Fig. 7-3-8
After removing the reagent, click [OK] and it will prompt "Drain tubing". The draining is finished when the
progress bar comes to an end. After the drainage of liquid in tubing, place the instrument in a clean and dry
place.
c)If the instrument will not be used for over 2 weeks, replace the reagent with distilled water and execute
"Package rinsing" operation. Click [Package rinsing] and it will prompt "Package rinsing". The package rinsing
of tubing is finished when the progress bar comes to an end.
Then, cut off distilled water and execute package draining. The operation steps are the same with steps for
draining tubing. Then, the tubing draining operation is finished. After finishing the two operations above, place
the instrument at a clean position.
(2)Aperture
As sundries will be attached in aperture after long time of using, unblock, cauterize and back wash the aperture to
clear the attached sundries.
a)Click [Cauterize] to cauterize the two ends of the aperture with high voltage DC, and meanwhile, pop up the
progress bar "Cauterize".
Clear sundries blocked in aperture to prevent and eliminate RBC blocking failure.
b)Click [Back washing] to flush the aperture, and meanwhile, pop up the progress bar "Back washing".
Assistance in cauterization can prevent and eliminate RBC hole blockage.
c)Click [Unblocking], a progress bar "Unblocking" pops up, then the instrument will execute cauterization and
back washing. The progress bar ends and the task is complete.
(3)Soak
To guarantee the accuracy of the testing result of the instrument, soak WBC cell, RBC cell and sampling probe
with probe detergent under the following situation:
a)The instrument is blocked;
b)The instrument shows abnormal classification.
(1)Soak RBC cell:
Soak and rinse RBC cell every other week, and the operation steps are as below:
Click [RBC cell] and a dialog box as shown below will pop up:
Fig. 7-3-9
If the instrument has a sample feeder, place the probe detergent at the STAT position and click Count, and the
interface below will appear:
77
Service Manual of Automatic Hematology Analyzer
Fig. 7-3-10
Fig. 7-3-11
Click [Next] to start rinsing RBC cell. When the progress bar comes to an end, it shows the end of the process.
Fig. 7-3-12
It the instrument does not have a sample feeder, the prompt box below will pop up.
Fig. 7-3-13
Put the probe detergent below the sampling probe according to prompt and the steps are same with that with a
sample feeder.
(2)Soak WBC cell:
Soak WBC cell every two weeks, and the operation steps are same with the soaking of RBC cell.
(3)Soak sample aspirating probe:
Soak rinsing sample-aspirating probe every month, and the operation steps are as below:
Click [Maintenance] - [Sampling probe] to pop up the prompt box as shown below:
78
Service Manual of Automatic Hematology Analyzer
Fig. 7-3-14
Place the Probe cleanser at the STAT position and click Count, and the interface below will appear:
Fig. 7-3-15
Fig. 7-3-16
Click [Verification] in the figure, and HGB verification value will be displayed in the box following "Verification".
7.3.5 Mechanical detection
7.3.5.1 Whole machine
Click [Menu] - [Maint.] - [Mechanical inspection], as shown in the figure below:
79
Service Manual of Automatic Hematology Analyzer
Fig. 7-3-17
(1)Motor:
When some part of the instrument breaks down, the failure can be judged by motor detection. This program sets the
following motors: lifting motor, rotary motor and rinsing motor. Click [Detection] after the corresponding motor
and the test result will be displayed on the screen ("Success" or "Failure").
(2)Valve:
If the instrument works abnormally due to valve failure, valve detection can serve as an important method to
eliminate liquid path failure.
Valve status after the test is completed: Yellow indicates successful detection and red indicates failed detection.
The operation process is as follows:
Click the corresponding number of the valve (numbers in Fig. 7-3-17 refer to the valves), and the instrument will
automatically detect the valve. Click the valve number and the sound of opening/closing the valve can be heard,
which suggests that the valve is in a normal state.
Click [Detection] after automatic detection, and the instrument will automatically detect the valves in order.
Meanwhile, the valve opening and closing sounds should be able to hear.
(3)Pump:
Click "Positive pressure" or "Negative pressure" to check whether the pump is under normal status.
After detection, click [Sample analysis] at the upper part of the screen to exit mechanical detection interface and
return to main interface.
Valve detection and positive/negative pressure pump detection cannot be executed at the same time.
7.3.5.2 Sample feeder
Click [Menu] - [Maint.] - [Mech.test] and select the [Sample feeder] tab, as shown in the figure below:
80
Service Manual of Automatic Hematology Analyzer
Fig. 7-3-18
When some part of moving parts of the sampler breaks down, the failure can be judged by motor detection. The
program of motor detection includes: loading motor, rack-conveying motor, emergency feeder motor, unloading
motor, shaking & lifting motor, pinch motor, bar code scanning motor click [Detection] on the back of
corresponding motor, and the test result will be displayed on the screen ("Succeeded" or "Failed").
The interface will not display the tab if no sample feeder is set.
Fig. 7-3-18(a)
81
Service Manual of Automatic Hematology Analyzer
7.3.5.3 CRP
Click [Menu] - [Maint.] - [Mech. test] and select the [CRP] tab, as shown in the figure below:
Fig. 7-3-19
The bar code reader will emit light harmful to people's eyes, please do not look straight at the light when the
Analyzer is working.
82
Service Manual of Automatic Hematology Analyzer
Fig. 7-3-20
Enter the instrument registration code or click Import (import the registration code in the USB flash drive), enter
the registration code and click Register. Registration success is prompted and the registration is completed.
7.3.7 Data backup
Click [Menu] - [Maint.] - [Data backup], as shown in the figure below:
Fig. 7-3-21
After inserting the U disk in the USB interface behind the instrument, click [Backup] on the interface to export the
83
Service Manual of Automatic Hematology Analyzer
Data should be backed up regularly to prevent data loss when hardware or software fails.
7.3.8 Counter
Click [Menu] - [Maint.] - [Counter], as shown in the figure below:
Fig. 7-3-22
(1)The sampling syringe, rinsing syringe and sheath syringe operation times, as well as the operating time of the
positive pressure pump and negative pressure pump (in unit of s) and probe puncturing times can be viewed.
(2)Users can view the operating times of the CRP syringe and the operating time of CRP negative pressure pump
(in unit of s).
84
Service Manual of Automatic Hematology Analyzer
Swab
Fig. 7-4-1
Screw off the connectors on the upper part of the sampling probe to replace with a new sampling probe as shown in
the following figure:
Fig. 7-4-2
Insert the new sampling probe vertically into the swab (as shown in Fig. 7-4-2) and fix the screws 1 & 2 (as shown
in Fig. 7-4-1).
85
Service Manual of Automatic Hematology Analyzer
(2)Don't use the detergent or sanitizer with possibility of danger due to chemical reaction with parts or materials in
equipment. (Soft cloth dipped with a small amount of 75% ethanol can be used for disinfection)
(3)In case of doubts about the compatibility between sanitizer or detergent and parts or materials in equipment,
please consult the manufacturer or its agencies.
Please handle the waste fluid of the Analyzer to prevent potential biological and chemical pollution.
86
Service Manual of Automatic Hematology Analyzer
87
Service Manual of Automatic Hematology Analyzer
Alarm code Alarm level Fault description Fault cause analysis Solutions
1. The middle computer board is
incompatible with the edition of CRP 1. Enter system version interface and check if the CRP test unit board version can be read;
test unit board; 2. If it can be read, please confirm if the compatibility of middle computer board and CRP test unit
2. The CAN communication interface
(socket J4) of middle computer board
board is correct and execute [Clear alarm];
fails; 3. If it cannot be read, check if version number of other lower computer boards can be normally
3. Harness (with sample feeder harness: read;
B01-P04-1/without sample feeder 3.1 If it cannot be read, check the CAN communication interface (socket J4) on middle computer
Abnormal CRP test unit
01-201-02 Stop level harness: B01-P04-2) or harness board and check the harness (with sample feeder check harness: B01-P04-1/without sample
board communication (B02-P11) fails; feeder check harness: B01-P04-2); if it fails, repair the interface or harness;
4. CAN communication interface 3.2 If only the version of CRP test unit board cannot be read, replace the PCB;
(socket J17) on analog signal collection 3.3 If the version of CRP test unit board, CRP unit collection board and main control board can all
board (B20) fails;
5. The CAN communication interface
not be read, check the CAN communication interface (socket J11) on CRP test unit board, check
(socket J11) on CRP test unit board the harness (harness: B02-P11) and check the CAN communication interface (socket J17) on
fails; analog signal collection board (B20), if it fails, repair the interface or harness.
6. CRP test unit board fails.
1. The middle computer board is
incompatible with the edition of analog 1. Enter system version interface and check if the analog signal collection board version can be
signal collection board; read;
2. The CAN communication interface 2. If it can be read, please confirm if the compatibility of middle computer board and analog signal
(socket J4) of middle computer board collection board is correct and execute [Clear alarm];
fails; 3. If it cannot be read, check if version number of other lower computer boards can be normally
3. Harness (with sample feeder harness: read;
Abnormal communication B01-P04-1/without sample feeder
3.1 If it cannot be read, check the CAN communication interface (socket J4) on middle computer
01-201-03 Stop level of analog signal collection harness: B01-P04-2) or harness
(B20-P16) fails; board and check the harness (with sample feeder check harness: B01-P04-1/without sample feeder
board
4. The communication interface (label: check harness: B01-P04-2), if it fails, repair the interface or harness;
P0204) on sample feeder control board 3.2 If only the version of analog signal collection board cannot be read, replace the PCB;
fails ; 3.3 If only the version of sample feeder control board can be read, check the CAN communication
5. The CAN communication interface interface (socket J16) of analog signal collection board, check the CAN communication interface
(socket J16) on analog signal collection (label: P0204) of sample feeder control board and check the harness (harness: B20-P16); if it fails,
board fails; repair the interface or harness.
6. Analog signal collection board fails.
1. The middle computer board is
incompatible with the edition of sample
feeder control board; 1. Enter system version interface and check if the sample feeder control board version can be read;
2. The CAN communication interface 2. If it can be read, please confirm if the compatibility of middle computer board and sample feeder
(socket J4) of middle computer board control board is correct and execute [Clear alarm];
Abnormal communication fails;
3. If it cannot be read, check if version number of other lower computer boards can be normally read;
01-201-05 Stop level of sample feeder control 3. Harness (with sample feeder harness:
B01-P04-1/without sample feeder 3.1 If it cannot be read, check the CAN communication interface (socket J4) of middle computer
board
harness: B01-P04-2) fails; board and the CAN communication interface (label: p0202) of sample feeder control board, and
4. The communication interface (label: check the harness (harness: b01-p04-1), if it fails, repair the interface or harness;
P0202) on sample feeder control board 3.2 If only the version of sample feeder control board cannot be read, replace the PCB.
fails;
5. Sample feeder control board fails.
88
Service Manual of Automatic Hematology Analyzer
Alarm code Alarm level Fault description Fault cause analysis Solutions
1. The middle computer board is
incompatible with the edition of main
control board; 1. Enter system version interface and check if the main control board version can be read;
2. The CAN communication interface 2. If it can be read, please confirm if the compatibility of middle computer board and main control
(socket J4) of middle computer board board is correct and execute [Clear alarm];
fails; 3. If it cannot be read, check if version number of other lower computer boards can be normally
3. Harness (with sample feeder harness: read;
B01-P04-1/without sample feeder
Abnormal communication 3.1 If it cannot be read, check the CAN communication interface (socket J4) on middle computer
01-201-06 Stop level harness: B01-P04-2) or harness
of main control board (B14-P08) fails; board and check the harness (with sample feeder check harness: B01-P04-1/without sample feeder
4. The CAN communication interface check harness: B01-P04-2), if it fails, repair the interface or harness;
(socket J37) on main control board 3.2 If only the version of main control board cannot be read, check the CAN communication
fails; interface (socket J37) of main control board, check the CAN communication interface (B14) of
5. CAN communication interface CRP unit collection board (socket J8), check harness (harness: B14-P08) ,if it fails, repair the
(socket J8) on CRP unit collection interface or harness, or replace the PCB.
board (B14) fails;
6. Main control board fails.
1. The middle computer board is
incompatible with the edition of CRP
unit collection board; 1. Enter system version interface and check if the CRP unit collection board version can be read;
2. The CAN communication interface 2. If it can be read, please confirm if the compatibility of middle computer board and CRP unit
(socket J4) of middle computer board collection board is correct and execute [Clear alarm];
fails; 3. If it cannot be read, check if version number of other lower computer boards can be normally read;
3. Harness (with sample feeder harness: 3.1 If it cannot be read, check the CAN communication interface (socket J4) on middle computer
B01-P04-1/without sample feeder
Abnormal communication board and check the harness (with sample feeder check harness: B01-P04-1/without sample feeder
01-201-07 Stop level harness: B01-P04-2) or harness
of CRP unit collection board (B02-P12) fails; check harness: B01-P04-2), if it fails, repair the interface or harness;
4. The CAN communication interface 3.2 If only the version of CRP unit collection board cannot be read, replace the PCB;
(socket J7) on CRP unit collection 3.3 If the version of CRP unit collection board and main control board can all not be read, check the
board fails; CAN communication interface (socket P7) on CRP unit collection board, check the harness
5. CAN communication interface (harness: B02-P12) and check the CAN communication interface (socket J12) on CRP test unit
(socket J12) on CRP test unit board board (B02) , if it fails, repair the interface or harness.
(B02) fails;
6. CRP unit collection board fails.
89
Service Manual of Automatic Hematology Analyzer
Alarm code Alarm level Fault description Fault cause analysis Solutions
1. Plug of the sensor is loose; 1. Enter [Menu] - [System maintenance] - [Basic status] interface and observe if the displayed value of ambient
Caution Ambient temperature is 2. Actual ambient temperature is temperature is below 10℃;
06-91-05
level lower than 10℃ lower than 10℃; 2. If the displayed temperature is below 10℃, please test again after ambient temperature rises;
3. The sensor is damaged. 3. If the temperature fluctuates, check if the plug J13 on main control board is loose.
1. Plug of the sensor is loose; 1. Enter [Menu] - [System maintenance] - [Basic status] interface and observe if the displayed value of ambient
Caution Ambient temperature is 2. Actual ambient temperature is temperature is higher than 35℃;
06-91-06
level higher than 35℃ higher than 35℃; 2. If the displayed temperature is higher than 35℃, please test again after ambient temperature decreases;
3. The sensor is damaged. 3. If the temperature fluctuates, check if the plug J13 on main control board is loose.
1. If the diluent tank is empty, please replace with new reagent and then execute alarm clearing operation;
1. Diluent tank is empty; 2. If the diluent tank is not empty but there are a large amount of air bubbles in the tubing connecting the diluent
2. There are a large amount of air tank and complete machine, please check if the diluent tank tubing is damaged at the connector and causes air
06-111-01 Stop level Insufficient BF-5D diluent bubbles at the inlet of diluent tubing; leakage;
3. The residual diluent detection 3. If it cannot be restored, enter [Menu] - [System maintenance] - [Advanced] - [Switch status] - [Complete
optocoupler is damaged. machine] and observe the status of diluent air bubble optocoupler. If it is not sheltered, please check if the
connecting cable is loose or if there are air bubbles in the tested object.
1. If the reagent bottle is empty, please replace with new reagent and then execute alarm clearing operation;
1. Empty reagent bottle; 2. If the reagent bottle is not empty but there are a large amount of air bubbles in the tubing connecting the
Partial 2. There are a large amount of air reagent bottle and the tested object, please check if the reagent bottle tubing is damaged at the connector and
Insufficient BF-FDOI
06-111-02 function bubbles in reagent tubing; causes air leakage;
reagent in test process 3. If it cannot be restored, enter [Menu] - [System maintenance] - [Advanced] - [Switch status] - [Complete
stop level 3. The residual reagent detection
optocoupler is damaged. machine] and observe the status of FDOI air bubble optocoupler. If it is not sheltered, please check if the
connecting cable is loose or if there are air bubbles in the tested object.
1. If the reagent bottle is empty, please replace with new reagent and then execute alarm clearing operation;
1. Empty reagent bottle; 2. If the reagent bottle is not empty but there are a large amount of air bubbles in the tubing connecting the
Partial 2. There are a large amount of air reagent bottle and the tested object, please check if the reagent bottle tubing is damaged at the connector and
Insufficient BF-FBH
06-111-03 function bubbles in reagent tubing; causes air leakage;
reagent in test process 3. If it cannot be restored, enter [Menu] - [System maintenance] - [Advanced] - [Switch status] - [Complete
stop level 3. The residual reagent detection
optocoupler is damaged. machine] and observe the status of FBH air bubble optocoupler. If it is not sheltered, please check if the
connecting cable is loose or if there are air bubbles in the tested object.
1. If the reagent bottle is empty, please replace with new reagent and then execute alarm clearing operation;
1. Empty reagent bottle; 2. If the reagent bottle is not empty but there are a large amount of air bubbles in the tubing connecting the
Partial 2. There are a large amount of air reagent bottle and the tested object, please check if the reagent bottle tubing is damaged at the connector and
Insufficient BF-FDTI
06-111-04 function bubbles in reagent tubing; causes air leakage;
reagent in test process 3. If it cannot be restored, enter [Menu] - [System maintenance] - [Advanced] - [Switch status] - [Complete
stop level 3. The residual reagent detection
optocoupler is damaged. machine] and observe the status of FDTI air bubble optocoupler. If it is not sheltered, please check if the
connecting cable is loose or if there are air bubbles in the tested object.
Partial 1. Enter [Menu] - [System maintenance] - [Basic status] interface and observe if the displayed value of
WBC test unit preheating 1. Plug of the sensor is loose; preheating tank temperature is normal;
06-91-01 function
temperature sensor failure 2. The sensor is damaged. 2. If the displayed temperature is always 0℃, check if the plug J15 on main control board is loose;
stop level 3. If it is not loose, please replace the sensor.
1. Enter [Menu] - [System maintenance] - [Basic status] interface and observe if the displayed value of
preheating tank temperature is within the range;
Partial 2. If the temperature is always below the lower limit of the allowable fluctuation range, check if the heating
WBC test unit preheating 1. The heading rod is damaged;
06-91-02 function indicator lamp DS8 near the heating rod plug J28 on the main control board flickers or is normally on;
temperature is too low 2. Temperature sensor is damaged 3. If the heating rod always works but the displayed temperature is still low, please replace the temperature
stop level
sensor on J15 socket or the heating rod;
4. If the heating rod does not work, please check if the U43 chip on main control board has poor soldering.
90
Service Manual of Automatic Hematology Analyzer
Alarm code Alarm level Fault description Fault cause analysis Solutions
1. Enter [Menu] - [System maintenance] - [Basic status] interface and observe if the displayed value of
Partial preheating tank temperature is within the range;
WBC test unit preheating 1. The heading rod is damaged; 2. If the temperature is always above the upper limit of the allowable fluctuation range, check if the heating
06-91-03 function
temperature is too high 2. Temperature sensor is damaged indicator lamp DS8 near the heating rod plug J28 on the main control board flickers or is normally on;
stop level 3. If the heating rod works at all times and does not stop, please check if the U43 chip on main control board is
damaged.
1. Execute [Clear alarm], observe if the probe lifting motor has operation action and if the fault can be removed.
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
The probe lifting motor does 1. Poor soldering or damage of [Complete machine], click [Lifting motor] and observe the operating status of probe lifting motor.
3. If the motor does not operate, please check if the motor connecting cable and corresponding socket have poor
not leave the resetting connecting cable of motor;
06-101-01 Stop level soldering or are damaged; check if the U30 chip .U1(MCU) of main control board has poor soldering or is
position in the process of 2. Motor zero-point optocoupler damaged.
moving downwards signal line broken or damaged; 4. Shelter the slot of motor zero-point optocoupler with an object and observe the DS58 signal lamp on main
control board near U33. When it is not sheltered, DS58 will be on, or DS58 will be off. If motor resetting
optocoupler has status error, please replace the resetting optocoupler.
1. Execute [Clear alarm], observe if the probe lifting motor has operation action and if the fault can be removed.
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
1. Motor zero-point optocoupler is [Complete machine], click [Lifting motor] and observe the operating status of probe lifting motor.
Resetting optocoupler signal 3. If the motor does not operate, please check if the motor connecting cable and corresponding socket have poor
damaged;
06-101-02 Stop level not detected after the probe soldering or are damaged; check if the U30 chip .U1(MCU) of main control board has poor soldering or is
2. Poor soldering or damage of damaged.
lifting motor resets
connecting cable of motor; 4. Shelter the slot of motor zero-point optocoupler with an object and observe the DS58 signal lamp on main
control board near U33. When it is not sheltered, DS58 will be on, or DS58 will be off. If motor resetting
optocoupler has status error, please replace the resetting optocoupler.
1. Execute [Clear alarm], observe if the probe lifting motor has operation action and if the fault can be removed.
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
[Complete machine], click [Lifting motor] and observe the operating status of probe lifting motor.
1. Poor soldering or damage of 3. If the motor does not operate, please check if the motor connecting cable and corresponding socket have poor
Step loss of probe lifting connecting cable of motor; soldering or are damaged; check if the U30 chip .U1(MCU) of main control board has poor soldering or is
06-101-03 Stop level
motor 2. The mechanism operates unstably damaged.
and causes step loss. 4. Shelter the slot of motor zero-point optocoupler with an object and observe the DS58 signal lamp on main
control board near U33. When it is not sheltered, DS58 will be on, or DS58 will be off. If motor resetting
optocoupler has status error, please replace the resetting optocoupler.
5. If the structure of probe lifting unit is loose, please fix the unit again.
1. Execute [Clear alarm], observe if the probe lifting motor has operation action and if the fault can be removed.
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
1. Poor soldering or damage of [Complete machine], click [Rotating motor] and observe the operating status of rotating motor.
Probe rotating motor does 3. If the motor does not operate, please check if the motor connecting cable and corresponding socket have poor
connecting cable of motor;
06-101-04 Stop level not leave the coded disk soldering or are damaged; check if the U28 chip .U1(MCU) of main control board has poor soldering or is
2. Motor zero-point optocoupler damaged.
positioning hole
signal line broken or damaged; 4. Rotate the probe rotating motor manually and observe the DS60 signal lamp on main control board near U33.
When it is not sheltered, DS60 will be on, or DS60 will be off. If motor resetting optocoupler has status error,
please replace the resetting optocoupler.
91
Service Manual of Automatic Hematology Analyzer
Alarm code Alarm level Fault description Fault cause analysis Solutions
1. Execute [Clear alarm], observe if the probe lifting motor has operation action and if the fault can be removed.
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
1. Poor soldering or damage of [Complete machine], click [Rotating motor] and observe the operating status of rotating motor.
Probe rotating motor does connecting cable of R-axis motor; 3. If the motor does not operate, please check if the motor connecting cable and corresponding socket have poor
06-101-05 Stop level not reach the coded disk 2. The probe fails to operate to the soldering or are damaged; check if the U28 chip .U1(MCU) of main control board has poor soldering or is
positioning hole appointed position due to damaged.
obstruction in its operating scope. 4. Rotate the probe rotating motor manually and observe the DS60 signal lamp on main control board near U33.
When it is not sheltered, DS60 will be on, or DS60 will be off. If motor resetting optocoupler has status error,
please replace the resetting optocoupler.
1. Execute [Clear alarm], observe if the sample aspirating motor has operation action and if the fault can be
removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
The sample aspirating motor 1. Poor soldering or damage of [Complete machine], click [Sample aspirating motor] and observe the operating status of the motor;
does not leave the resetting connecting cable of motor; 3. If the motor does not operate, please check if the motor connecting cable and corresponding socket J24 have
06-101-07 Stop level poor soldering or are damaged; check if the U31 chip .U1(MCU) of main control board has poor soldering or is
position in the process of 2. Motor zero-point optocoupler
moving downwards signal line broken or damaged; damaged;
4. Shelter the slot of motor zero-point optocoupler with an object and observe the DS49 signal lamp on main
control board near U32. When it is not sheltered, DS49 will be off, or DS49 will be on. If motor resetting
optocoupler has status error, please replace the resetting optocoupler.
1. Execute [Clear alarm], observe if the sample aspirating motor has operation action and if the fault can be
removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
1. Motor zero-point optocoupler is [Complete machine], click [Sample aspirating motor] and observe the operating status of the motor;
Resetting optocoupler signal
damaged; 3. If the motor does not operate, please check if the motor connecting cable and corresponding socket J24 have
06-101-08 Stop level not detected after the sample poor soldering or are damaged; check if the U31 chip .U1(MCU) of main control board has poor soldering or is
2. Poor soldering or damage of
aspirating motor resets damaged;
connecting cable of motor;
4. Shelter the slot of motor zero-point optocoupler with an object and observe the DS49 signal lamp on main
control board near U32. When it is not sheltered, DS49 will be off, or DS49 will be on. If motor resetting
optocoupler has status error, please replace the resetting optocoupler.
1. Execute [Clear alarm], observe if the sample aspirating motor has operation action, if it operates smoothly and
if the fault can be removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
1. Poor soldering or damage of [Complete machine], click [Sample aspirating motor] and observe the operating status of the motor;
Step loss of sample connecting cable of motor; 3. If the motor does not operate, please check if the motor connecting cable and corresponding socket J24 have
06-101-09 Stop level poor soldering or are damaged; check if the U31 chip .U1(MCU) of main control board has poor soldering or is
aspirating motor 2. The mechanism operates unstably
and causes step loss. damaged;
4. Shelter the slot of motor zero-point optocoupler with an object and observe the DS49 signal lamp on main
control board near U32. When it is not sheltered, DS49 will be off, or DS49 will be on. If motor resetting
optocoupler has status error, please replace the resetting optocoupler.
1. Execute [Clear alarm], observe if the rinsing motor has operation action and if the fault can be removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
The rinsing motor does not 1. Poor soldering or damage of [Complete machine], click [Rinsing motor] and observe the operating status of the motor;
3. If the motor does not operate, please check if the motor connecting cable and corresponding socket J25 have
leave the resetting position connecting cable of motor;
06-101-10 Stop level poor soldering or are damaged; check if the U29 chip .U1(MCU) of main control board has poor soldering or is
in the process of moving 2. Motor zero-point optocoupler damaged;
downwards signal line broken or damaged; 4. Shelter the slot of motor zero-point optocoupler with an object and observe the DS50 signal lamp on main
control board near U32. When it is not sheltered, DS50 will be off, or DS50 will be on. If motor resetting
optocoupler has status error, please replace the resetting optocoupler.
92
Service Manual of Automatic Hematology Analyzer
Alarm code Alarm level Fault description Fault cause analysis Solutions
1. Execute [Clear alarm], observe if the rinsing motor has operation action and if the fault can be removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
1. Motor zero-point optocoupler is [Complete machine], click [Rinsing motor] and observe the operating status of the motor;
Resetting optocoupler signal 3. If the motor does not operate, please check if the motor connecting cable and corresponding socket J25 have
damaged;
06-101-11 Stop level not detected after the rinsing poor soldering or are damaged; check if the U29 chip .U1(MCU) of main control board has poor soldering or is
2. Poor soldering or damage of damaged;
motor resets
connecting cable of motor; 4. Shelter the slot of motor zero-point optocoupler with an object and observe the DS50 signal lamp on main
control board near U32. When it is not sheltered, DS50 will be off, or DS50 will be on. If motor resetting
optocoupler has status error, please replace the resetting optocoupler.
1. Execute [Clear alarm], observe if the rinsing motor has operation action, if it operates steadily and if the fault
can be removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
1. Poor soldering or damage of [Complete machine], click [Rinsing motor] and observe the operating status of the motor;
connecting cable of motor; 3. If the motor does not operate, please check if the motor connecting cable and corresponding socket J25 have
06-101-12 Stop level Step loss of rinsing motor poor soldering or are damaged; check if the U29 chip .U1(MCU) of main control board has poor soldering or is
2. The mechanism operates unstably
and causes step loss. damaged;
4. Shelter the slot of motor zero-point optocoupler with an object and observe the DS50 signal lamp on main
control board near U32. When it is not sheltered, DS50 will be off, or DS50 will be on. If motor resetting
optocoupler has status error, please replace the resetting optocoupler.
1. Execute [Clear alarm], observe if the rinsing motor has operation action and if the fault can be removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
The sheath flow motor does 1. Poor soldering or damage of [Complete machine], click [Sheath flow motor] and observe the operating status of the motor;
3. If the motor does not operate, please check if the motor connecting cable and corresponding socket J26 have
not leave the resetting connecting cable of motor;
06-101-13 Stop level poor soldering or are damaged; check if the U27 chip .U1(MCU) of main control board has poor soldering or is
position in the process of 2. Motor zero-point optocoupler damaged;
moving downwards signal line broken or damaged; 4. Shelter the slot of motor zero-point optocoupler with an object and observe the DS57 signal lamp on main
control board near U33. When it is not sheltered, DS57 will be off, or DS57 will be on. If motor resetting
optocoupler has status error, please replace the resetting optocoupler.
1. Execute [Clear alarm], observe if the rinsing motor has operation action and if the fault can be removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
1. Motor zero-point optocoupler is [Complete machine], click [Sheath flow motor] and observe the operating status of the motor;
Resetting optocoupler signal 3. If the motor does not operate, please check if the motor connecting cable and corresponding socket J26 have
damaged;
06-101-14 Stop level not detected after the sheath poor soldering or are damaged; check if the U27 chip .U1(MCU) of main control board has poor soldering or is
2. Poor soldering or damage of damaged;
flow motor resets
connecting cable of motor; 4. Shelter the slot of motor zero-point optocoupler with an object and observe the DS57 signal lamp on main
control board near U33. When it is not sheltered, DS57 will be off, or DS57 will be on. If motor resetting
optocoupler has status error, please replace the resetting optocoupler.
1. Execute [Clear alarm], observe if the rinsing motor has operation action, if it operates steadily and if the fault
can be removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
1. Poor soldering or damage of [Complete machine], click [Sheath flow motor] and observe the operating status of the motor;
Step loss of sheath flow connecting cable of motor; 3. If the motor does not operate, please check if the motor connecting cable and corresponding socket J26 have
06-101-15 Stop level poor soldering or are damaged; check if the U27 chip .U1(MCU) of main control board has poor soldering or is
motor 2. The mechanism operates unstably
and causes step loss. damaged;
4. Shelter the slot of motor zero-point optocoupler with an object and observe the DS57 signal lamp on main
control board near U33. When it is not sheltered, DS57 will be off, or DS57 will be on. If motor resetting
optocoupler has status error, please replace the resetting optocoupler.
93
Service Manual of Automatic Hematology Analyzer
Alarm code Alarm level Fault description Fault cause analysis Solutions
1. Execute [Clear alarm], observe if the fault can be removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
[Complete machine], click [SV1] and observe if [SV1] button turns yellow and if the DS43 near J21 is on with
1. Solenoid valve is not connected; "KADA" sound;
06-131-01 Stop level SV1 fault
2. Solenoid valve has short circuit. 3. If the lamp is on but generates no "KADA" sound, please check if the solenoid valve connecting cable and
corresponding socket have poor soldering or are damaged;
4. If the lamp is not on, please check if the U14 chip .U1(MCU) of main control board has poor soldering or is
damaged.
1. Execute [Clear alarm], observe if the fault can be removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
[Complete machine], click [SV2] and observe if [SV2] button turns yellow and if the DS41 near J21 is on with
1. Solenoid valve is not connected; "KADA" sound;
06-131-02 Stop level SV2 fault
2. Solenoid valve has short circuit. 3. If the lamp is on but generates no "KADA" sound, please check if the solenoid valve connecting cable and
corresponding socket have poor soldering or are damaged;
4. If the lamp is not on, please check if the U14 chip .U1(MCU) of main control board has poor soldering or is
damaged.
1. Execute [Clear alarm], observe if the fault can be removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
[Complete machine], click [SV3] and observe if [SV3] button turns yellow and if the DS44 near J21 is on with
1. Solenoid valve is not connected; "KADA" sound;
06-131-03 Stop level SV3 fault
2. Solenoid valve has short circuit. 3. If the lamp is on but generates no "KADA" sound, please check if the solenoid valve connecting cable and
corresponding socket have poor soldering or are damaged;
4. If the lamp is not on, please check if the U14 chip .U1(MCU) of main control board has poor soldering or is
damaged.
1. Execute [Clear alarm], observe if the fault can be removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
[Complete machine], click [SV4] and observe if [SV4] button turns yellow and if the DS31 near J19 is on with
1. Solenoid valve is not connected; "KADA" sound;
06-131-04 Stop level SV4 fault
2. Solenoid valve has short circuit. 3. If the lamp is on but generates no "KADA" sound, please check if the solenoid valve connecting cable and
corresponding socket have poor soldering or are damaged;
4. If the lamp is not on, please check if the U13 chip .U1(MCU) of main control board has poor soldering or is
damaged.
1. Execute [Clear alarm], observe if the fault can be removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
[Complete machine], click [SV5] and observe if [SV5] button turns yellow and if the DS29 near J19 is on with
1. Solenoid valve is not connected; "KADA" sound;
06-131-05 Stop level SV5 fault
2. Solenoid valve has short circuit. 3. If the lamp is on but generates no "KADA" sound, please check if the solenoid valve connecting cable and
corresponding socket have poor soldering or are damaged;
4. If the lamp is not on, please check if the U13 chip .U1(MCU) of main control board has poor soldering or is
damaged.
1. Execute [Clear alarm], observe if the fault can be removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
[Complete machine], click [SV7] and observe if [SV7] button turns yellow and if the DS30 near J19 is on with
1. Solenoid valve is not connected; "KADA" sound;
06-131-07 Stop level SV7 fault
2. Solenoid valve has short circuit. 3. If the lamp is on but generates no "KADA" sound, please check if the solenoid valve connecting cable and
corresponding socket have poor soldering or are damaged;
4. If the lamp is not on, please check if the U13 chip .U1(MCU) of main control board has poor soldering or is
damaged.
94
Service Manual of Automatic Hematology Analyzer
Alarm code Alarm level Fault description Fault cause analysis Solutions
1. Execute [Clear alarm], observe if the fault can be removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
[Complete machine], click [SV8] and observe if [SV8] button turns yellow and if the DS27 near J19 is on with
1. Solenoid valve is not connected; "KADA" sound;
06-131-08 Stop level SV8 fault
2. Solenoid valve has short circuit. 3. If the lamp is on but generates no "KADA" sound, please check if the solenoid valve connecting cable and
corresponding socket have poor soldering or are damaged;
4. If the lamp is not on, please check if the U13 chip .U1(MCU) of main control board has poor soldering or is
damaged.
1. Execute [Clear alarm], observe if the fault can be removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
[Complete machine], click [SV9] and observe if [SV9] button turns yellow and if the DS28 near J19 is on with
1. Solenoid valve is not connected; "KADA" sound;
06-131-09 Stop level SV9 fault
2. Solenoid valve has short circuit. 3. If the lamp is on but generates no "KADA" sound, please check if the solenoid valve connecting cable and
corresponding socket have poor soldering or are damaged;
4. If the lamp is not on, please check if the U13 chip .U1(MCU) of main control board has poor soldering or is
damaged.
1. Execute [Clear alarm], observe if the fault can be removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
[Complete machine], click [SV10] and observe if [SV10] button turns yellow and if the DS42 near J21 is on with
1. Solenoid valve is not connected; "KADA" sound;
06-131-10 Stop level SV10 fault
2. Solenoid valve has short circuit. 3. If the lamp is on but generates no "KADA" sound, please check if the solenoid valve connecting cable and
corresponding socket have poor soldering or are damaged;
4. If the lamp is not on, please check if the U14 chip .U1(MCU) of main control board has poor soldering or is
damaged.
1. Execute [Clear alarm], observe if the fault can be removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
[Complete machine], click [SV11] and observe if [SV11] button turns yellow and if the DS39 near J21 is on with
1. Solenoid valve is not connected; "KADA" sound;
06-131-11 Stop level SV11 fault
2. Solenoid valve has short circuit. 3. If the lamp is on but generates no "KADA" sound, please check if the solenoid valve connecting cable and
corresponding socket have poor soldering or are damaged;
4. If the lamp is not on, please check if the U14 chip .U1(MCU) of main control board has poor soldering or is
damaged.
1. Execute [Clear alarm], observe if the fault can be removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
[Complete machine], click [SV12] and observe if [SV12] button turns yellow and if the DS38 near J20 is on with
1. Solenoid valve is not connected; "KADA" sound;
06-131-12 Stop level SV12 fault
2. Solenoid valve has short circuit. 3. If the lamp is on but generates no "KADA" sound, please check if the solenoid valve connecting cable and
corresponding socket have poor soldering or are damaged;
4. If the lamp is not on, please check if the U14 chip .U1(MCU) of main control board has poor soldering or is
damaged.
1. Execute [Clear alarm], observe if the fault can be removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
[Complete machine], click [SV13] and observe if [SV13] button turns yellow and if the DS36 near J20 is on with
1. Solenoid valve is not connected; "KADA" sound;
06-131-13 Stop level SV13 fault
2. Solenoid valve has short circuit. 3. If the lamp is on but generates no "KADA" sound, please check if the solenoid valve connecting cable and
corresponding socket have poor soldering or are damaged;
4. If the lamp is not on, please check if the U14 chip .U1(MCU) of main control board has poor soldering or is
damaged.
95
Service Manual of Automatic Hematology Analyzer
Alarm code Alarm level Fault description Fault cause analysis Solutions
1. Execute [Clear alarm], observe if the fault can be removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
[Complete machine], click [SV14] and observe if [SV14] button turns yellow and if the DS35 near J20 is on with
1. Solenoid valve is not connected; "KADA" sound;
06-131-14 Stop level SV14 fault
2. Solenoid valve has short circuit. 3. If the lamp is on but generates no "KADA" sound, please check if the solenoid valve connecting cable and
corresponding socket have poor soldering or are damaged;
4. If the lamp is not on, please check if the U14 chip .U1(MCU) of main control board has poor soldering or is
damaged.
1. Execute [Clear alarm], observe if the fault can be removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
[Complete machine], click [SV15] and observe if [SV15] button turns yellow and if the DS40 near J21 is on with
1. Solenoid valve is not connected; "KADA" sound;
06-131-15 Stop level SV15 fault
2. Solenoid valve has short circuit. 3. If the lamp is on but generates no "KADA" sound, please check if the solenoid valve connecting cable and
corresponding socket have poor soldering or are damaged;
4. If the lamp is not on, please check if the U14 chip .U1(MCU) of main control board has poor soldering or is
damaged.
1. Execute [Clear alarm], observe if the fault can be removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
[Complete machine], click [SV16] and observe if [SV16] button turns yellow and if the DS37 near J21 is on with
1. Solenoid valve is not connected; "KADA" sound;
06-131-16 Stop level SV16 fault
2. Solenoid valve has short circuit. 3. If the lamp is on but generates no "KADA" sound, please check if the solenoid valve connecting cable and
corresponding socket have poor soldering or are damaged;
4. If the lamp is not on, please check if the U14 chip .U1(MCU) of main control board has poor soldering or is
damaged.
1. Execute [Clear alarm], observe if the fault can be removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
[Complete machine], click [SV17] and observe if [SV17] button turns yellow and if the DS26 near J18 is on with
1. Solenoid valve is not connected; "KADA" sound.
06-131-17 Stop level SV17 fault
2. Solenoid valve has short circuit. 3. If the lamp is on but generates no "KADA" sound, please check if the solenoid valve connecting cable and
corresponding socket have poor soldering or are damaged;
4. If the lamp is not on, please check if the U13 chip .U1(MCU) of main control board has poor soldering or is
damaged.
1. Execute [Clear alarm], observe if the fault can be removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
[Complete machine], click [SV18] and observe if [SV18] button turns yellow and if the DS24 near J18 is on with
1. Solenoid valve is not connected; "KADA" sound;
06-131-18 Stop level SV18 fault
2. Solenoid valve has short circuit. 3. If the lamp is on but generates no "KADA" sound, please check if the solenoid valve connecting cable and
corresponding socket have poor soldering or are damaged;
4. If the lamp is not on, please check if the U13 chip .U1(MCU) of main control board has poor soldering or is
damaged.
1. Execute [Clear alarm], observe if the fault can be removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
[Complete machine], click [SV19] and observe if [SV19] button turns yellow and if the DS23 near J18 is on with
1. Solenoid valve is not connected; "KADA" sound;
06-131-19 Stop level SV19 fault
2. Solenoid valve has short circuit. 3. If the lamp is on but generates no "KADA" sound, please check if the solenoid valve connecting cable and
corresponding socket have poor soldering or are damaged;
4. If the lamp is not on, please check if the U13 chip .U1(MCU) of main control board has poor soldering or is
damaged.
96
Service Manual of Automatic Hematology Analyzer
Alarm code Alarm level Fault description Fault cause analysis Solutions
1. Execute [Clear alarm], observe if the fault can be removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
[Complete machine], click [SV20] and observe if [SV20] button turns yellow and if the DS22 near J18 is on with
1. Solenoid valve is not connected; "KADA" sound.
06-131-20 Stop level SV20 fault
2. Solenoid valve has short circuit. 3. If the lamp is on but generates no "KADA" sound, please check if the solenoid valve connecting cable and
corresponding socket have poor soldering or are damaged;
4. If the lamp is not on, please check if the U13 chip .U1(MCU) of main control board has poor soldering or is
damaged.
1. Execute [Clear alarm], observe if the fault can be removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
[Complete machine], click [SV28] and observe if [SV28] button turns yellow and if the DS34 near J20 is on with
1. Solenoid valve is not connected; "KADA" sound;
06-131-28 Stop level SV28 fault
2. Solenoid valve has short circuit. 3. If the lamp is on but generates no "KADA" sound, please check if the solenoid valve connecting cable and
corresponding socket have poor soldering or are damaged;
4. If the lamp is not on, please check if the U14 chip .U1(MCU) of main control board has poor soldering or is
damaged.
1. Execute [Clear alarm], observe if the fault can be removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
[Complete machine], click [SV29] and observe if [SV29] button turns yellow and if the DS25 near J19 is on with
1. Solenoid valve is not connected; "KADA" sound;
06-131-29 Stop level SV29 fault
2. Solenoid valve has short circuit. 3. If the lamp is on but generates no "KADA" sound, please check if the solenoid valve connecting cable and
corresponding socket have poor soldering or are damaged;
4. If the lamp is not on, please check if the U13 chip .U1(MCU) of main control board has poor soldering or is
damaged.
1. Execute [Clear alarm], observe if the fault can be removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
[Complete machine], click [SV30] and observe if [SV30] button turns yellow and if the DS21 near J18 is on with
1. Solenoid valve is not connected; "KADA" sound;
06-131-30 Stop level SV30 fault
2. Solenoid valve has short circuit. 3. If the lamp is on but generates no "KADA" sound, please check if the solenoid valve connecting cable and
corresponding socket have poor soldering or are damaged;
4. If the lamp is not on, please check if the U13 chip .U1(MCU) of main control board has poor soldering or is
damaged.
1. Execute [Clear alarm], observe if the fault can be removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] -
[Complete machine], click [SV31] and observe if [SV31] button turns yellow and if the DS33 near J20 is on with
1. Solenoid valve is not connected; "KADA" sound;
06-131-31 Stop level SV31 fault
2. Solenoid valve has short circuit. 3. If the lamp is on but generates no "KADA" sound, please check if the solenoid valve connecting cable and
corresponding socket have poor soldering or are damaged;
4. If the lamp is not on, please check if the U14 chip .U1(MCU) of main control board has poor soldering or is
damaged.
The sample dispensing 1. CAN communication cable is 1. Please execute alarm clearing operation after the version is checked as normal.
06-181-01 Stop level command of CRP unit not damaged; 2. If it cannot be restored or there are other faults, please confirm if the version is compatible and upgrade to the
received in given time limit 2. CRP test unit board is damaged. newest compatible version.
The C-reactive protein
1. CAN communication cable is 1. Please execute alarm clearing operation after the version is checked as normal.
reagent dispensing
06-181-02 Stop level damaged; 2. If it cannot be restored or there are other faults, please confirm if the version is compatible and upgrade to the
command of CRP unit not newest compatible version.
2. CRP test unit board is damaged.
received in given time limit
97
Service Manual of Automatic Hematology Analyzer
Alarm code Alarm level Fault description Fault cause analysis Solutions
The refrigeration chamber
1. CAN communication cable is 1. Please execute alarm clearing operation after the version is checked as normal.
cover opening completion
06-181-03 Stop level damaged; 2. If it cannot be restored or there are other faults, please confirm if the version is compatible and upgrade to the
command of CRP unit not newest compatible version.
2. CRP test unit board is damaged.
received in given time limit
1. Enter the basic status interface to observe if the sheath pressure is within normal range;
2. If it is not within the normal range, please pull off the sheath pressure sensor plug J2 on main control board and
observe if the sheath pressure approaches to 0Kpa or not;
3. If the sheath pressure does not approach to 0Kpa after the plug is pulled off, adjust the slide rheostat RP2 to
Sheath pressure is higher 1. Optical sheath flow unit is adjust the pressure around 0Kpa;
06-92-01 Stop level than the normal range in blocked; 4. If the sheath pressure under basic status is still not within the normal range after the adjustment, please pull off
sheath flow test process 2. Hydraulic sensor is damaged. the connector of tubing at upper terminal of the sheath syringe and enter the basic status interface to observe if the
sheath pressure approaches to 0Kpa;
5. If the sheath pressure is abnormal, please replace the sheath pressure sensor;
6. If the sheath pressure approaches to 0Kpa, please execute sheath flow cell soaking operation again after
turn-off and restart.
1. Sheath flow tubing has air
Sheath pressure is lower bubbles; 1. Enter the basic status interface to observe if the sheath pressure is within normal range;
2. If the pressure is below the lower limit of the normal range, please check if there are air bubbles in sheath
06-92-02 Stop level than the normal range in 2. Diluent is empty; tubing or if the sheath tubing falls off;
sheath flow test process 3. Hydraulic sensor is damaged or 3. If there are no the phenomena above, please replace the sheath pressure sensor.
connection is insecure.
1. Drive part of main control board 1. Enter the basic status interface to observe if the negative air pressure is within normal range;
Pressure in waste liquid negative pressure pump is damaged 2. If the negative air pressure is above the normal range, enter [Menu] - [System maintenance] - [Mechanical
bottle is higher than the and it leads to the negative pressure detection] - [Complete machine], click [SV31] and observe if [SV31] button turns yellow and if the DS33 near
06-92-03 Stop level
normal range in RBC test pumps works continuously; J20 is on with "KADA" sound;
process 2. Negative pressure sensor on main 3. If SV31 is normal but the negative air pressure under basic status is above the normal range, please replace the
control board is damaged. main control board.
1. Enter the basic status interface to observe if the negative air pressure is within normal range;
2. If the negative air pressure is below the normal range, enter [Menu] - [System maintenance] - [Mechanical
Pressure in waste liquid detection] - [Complete machine], click [SV31] and observe if [SV31] button turns yellow and if the DS33 near
1. Abnormal liquid drainage of RBC J20 is on with "KADA" sound.
bottle is lower than the
06-92-04 Stop level cell; 3. If SV31 is normal but the negative air pressure under basic status is still below the normal range, please replace
normal range in RBC test the negative pressure pump.
2. Diluent is empty.
process 4. If the negative air pressure under basic status is within the normal range, please execute RBC cell rinsing
operation and observe if RBC cell is blocked in liquid drainage. If the RBC cell is blocked in liquid drainage,
please check if the SV17 valve and filter screen of RBC cell waste liquid buffer bottle is blocked.
1. Drive part of main control board 1. Enter the basic status interface to observe if the positive air pressure is within normal range;
positive pressure pump is damaged 2. If positive pressure is higher than 15Kpa, enter [Menu] - [System maintenance] - [Mechanical detection] -
Pressure in air bottle is [Complete machine], turn on and off [Positive pressure pump] repeatedly and observe if the positive pressure
and it leads to the positive pressure
06-92-05 Stop level higher than the normal range pump is under control;
pumps works continuously; 3. If the positive pressure pump is not under control, please replace the main control board;
in mixing process
2. Positive pressure sensor on main 4. If the positive pressure pump is under control but the negative pressure under basic status is still high, please
control board is damaged. replace the main control board.
98
Service Manual of Automatic Hematology Analyzer
Alarm code Alarm level Fault description Fault cause analysis Solutions
1. Enter the basic status interface to observe if the positive air pressure is within normal range;
1. Positive pressure bottle has air 2. If positive pressure is lower than 15Kpa, enter [Menu] - [System maintenance] - [Mechanical detection] -
leakage; [Complete machine], turn on and off [Positive pressure pump] repeatedly and observe if the positive pressure
Pressure in air bottle is 2. Positive pressure pump is pump is under control;
06-92-06 Stop level lower than the normal range damaged; 3. If the positive pressure pump is not under control, please replace the positive pressure pump;
in mixing process 3. There is no water in mixing 4. If the positive pressure decreases quickly after it becomes stable under the basic status, it means the tubing of
process of RBC cell or WBC cell in the positive pressure bottle has air leakage. Please find out the air leakage point and make connection again.;
test process. 5. If there are no phenomena above, please execute WBC cell and RBC cell operations and observe if there is
water leakage.
99
Service Manual of Automatic Hematology Analyzer
Alarm code Alarm level Fault description Fault cause analysis Solutions
FPGA does not enter data 1. Execute [Clear alarm], observe if the fault can be removed;
1. Poor soldering or damage of
03-183-01 Stop level receiving state when the 2. If the fault cannot be removed, please check if the analog board U30 chip has poor soldering or is damaged;
FPGA chip. 3. If the chip is well welded, please replace the analog signal collection board.
system is powered or reset
FPGA entering resetting 1. Execute [Clear alarm], observe if the fault can be removed;
1. Poor soldering or damage of
03-183-02 Stop level status failed when the 2. If the fault cannot be removed, please check if the analog board U30 chip has poor soldering or is damaged;
FPGA chip. 3. If the chip is well welded, please replace the analog signal collection board.
system is powered or reset
1. Execute [Clear alarm], observe if the fault can be removed;
1. FPGA program is not 2. If the fault cannot be removed, after the instrument is powered on, insert the U disk with analog board FPGA
FPGA program guidance upgrading file to the USB interface of the instrument, enter [Menu] - [System maintenance] - [System version]
programmed;
03-183-03 Stop level failure when the system is interface, click the Upgrade button behind FPGA version number to upgrade analog board FPGA program again,
2. poor soldering or damage of and power off and restart after the upgrading.
powered or reset
FPGA chip. 3. If the fault cannot be removed, please check if the analog board U30 chip has poor welding or is damaged;
4. If the chip is well welded, please replace the analog signal collection board.
1. Solenoid valve V18 has fault; 1. Execute [Clear alarm], observe if the fault can be removed;
Liquid drainage failure of
03-201-22 Stop level 2. Waste liquid bottle in WBC cell is 2. If the fault cannot be removed, please check if solenoid valve V18 has fault;
WBC cell 3. If solenoid valve V18 is in good condition, please check if WBC waste liquid buffer bottle is blocked.
blocked.
100
Service Manual of Automatic Hematology Analyzer
Alarm code Alarm level Fault description Fault cause analysis Solutions
1. Check if the diluent is sufficient;
2. Check if CRP tubing has air bubbles;
3. Check if the tubing connection is correct, if there is liquid leakage and if the tubing has normal flow;
4. Enter [Menu] - [System maintenance] - [Service] - [Rinsing] and execute [CRP2], execute blank test for
reagent in corresponding channel after it is confirmed the tubing is full filled with diluent, check if [CRP blank
1. The cuvette is dirty and it causes test result] in debug information is within the range of CRP spotlight value (default 3000)±1000;
poor transmittance; 5. If [CRP blank test result] is correct, execute [Clear alarm]; if [CRP blank test result] is incorrect, execute [CRP
verification];
Caution Abnormal data in CRP 2. The cuvette has air bubbles and it
07-51-02 6. After successful [CRP verification], execute [Clear alarm]; if [CRP verification] fails, disconnect the power
level blank test causes light scattering. supply for CRP unit acquisition board. Then, use a multimeter at diode gear to measure if there is reading
3.Results of CRP verification error between B02-P04 red and black wires. If it is open circuit, it means the light-emitting diode is damaged;
7. The light-emitting diode is not damaged, pull out B02-P03 and power on the CRP unit acquisition board again
and then use a multimeter at mA gear (it should be noted to put the pen at the current gear position) to measure if
there is current between the yellow and green wires;
8. If there is current, it means the photocell is not damaged. If the light-emitting diode and photocell are measured
as normal, replace the electric circuit board of CRP unit acquisition board and execute [CRP verification];
9. After successful [CRP verification], execute [Clear alarm] and conduct blank test of reagent.
1. Check if the latex reagent is normal and if the reagent volume is sufficient;
2.Check if CRP lyse is sufficient and if the tubing has air bubbles;
3. Check if the tubing connection is correct, if there is liquid leakage and if the tubing has normal flow;
4. Check the first parameter of the alarm in fault log. If the first parameter is 1, it means the test data of CRP
channel 1 is abnormal; if the first parameter is 2, it means the test data of CRP channel 2 is abnormal;
1. C-reactive protein reagent used 5. After the channel with abnormal test data is confirmed, check the debug information of the last test of
up or error; corresponding channel and check if [CRP blank test result] in debug information is within the range of CRP
2. The cuvette has air bubbles and it spotlight value (default 3000)±1000; if the value is correct, it means there is no problem with the circuit and unit
causes light scattering; and the alarm may be caused by air in the reaction solution;
Caution Abnormal data in CRP 6. If the value of [CRP blank test result] is incorrect and it equals to [CRP-Start] and [CRP-End] as 65535, check
07-51-03 3. Signal line (including light
level sample test if the diluent is sufficient and execute [Clear alarm] if the diluent is sufficient;
emitting and receiving elements) 7. If the alarm cannot be cleared, disconnect the power supply for CRP unit acquisition board. Then, use a
circuit board is damaged or signal multimeter at diode gear to measure if there is reading between B02-P02 red and black wires for channel 1
line and circuit board have poor (B02-P04 for channel 2). If it is open circuit, it means the light-emitting diode is damaged;
contact. 8. If the light-emitting diode is not damaged, pull out B02-P01 for channel 1 (B02-P03 for channel 2) and power
on the CRP unit acquisition board again and then use a multimeter at mA gear (it should be noted to put the pen at
the current gear position) to measure if there is current between the yellow and green wires;
9. If there is current, it means the photocell is not damaged. If the light-emitting diode and photocell are measured
as normal, replace the electric circuit board of CRP unit acquisition board and execute [CRP verification];
10. After successful [CRP verification], execute [Clear alarm] and conduct blank test of reagent.
101
Service Manual of Automatic Hematology Analyzer
Alarm code Alarm level Fault description Fault cause analysis Solutions
1. Check if the diluent is sufficient;
2. Check if CRP tubing has air bubbles;
3. Check if the tubing connection is correct, if there is liquid leakage and if the tubing has normal flow;
4. Execute [Clear alarm];
1. There is no diluent in tubing; 5. If the alarm cannot be cleared, check the parameters of the alarm in fault log. If the first parameter is 1, it means
2. The cuvette has air bubbles and it successful verification of CRP channel 1, or verification of channel 1 fails; if the second parameter is 1, it means
causes light scattering; successful verification of CRP channel 2, or verification of channel 2 fails;
Caution 3. Signal line (including light 6. Disconnect the power supply for CRP unit acquisition board. Then, use a multimeter at diode gear to measure
07-51-04 CRP verification failed
level emitting and receiving elements) if there is reading between B02-P02 red and black wires for channel 1 (B02-P04 for channel 2). If it is open
circuit board is damaged or signal circuit, it means the light-emitting diode is damaged.
line and circuit board have poor 7. If the light-emitting diode is not damaged, pull out B02-P01 for channel 1 (B02-P03 for channel 2) and power
on the CRP unit acquisition board again and then use a multimeter at mA gear (it should be noted to put the pen at
contact.
the current gear position) to measure if there is current between the yellow and green wires.
8. If there is current, it means the photocell is not damaged. If the light-emitting diode and photocell are measured
as normal, replace the electric circuit board of CRP unit acquisition board and execute [CRP verification].
9. After successful [CRP verification], execute [Clear alarm] and conduct blank test of reagent.
102
Service Manual of Automatic Hematology Analyzer
Alarm code Alarm level Fault description Fault cause analysis Solutions
1. Execute [Clear alarm], observe if the fault can be removed;
1. The temperature sensor is 2. Power on the instrument, enter [Menu] - [System maintenance] - [Basic status] interface and observe if CRP
Partial The CRP reaction cell
damaged or poorly connected; reaction tank temperature rises;
02-91-17 function temperature is lower than
2.The heating band is poorly 3. Check if wire B02-P10 is correctly connected on J01 interface of CRP test unit board;
stop level the lower temperature limit 4. Check if wire PH1201 is correctly connected on the CRP unit;
connected or damaged.
5. Ensure the resistance of heating band is correct (smaller than 50Ω, theoretical value is around 24Ω).
1. Execute [Clear alarm], observe if the fault can be removed;
1. The temperature sensor is 2. Power on the instrument, enter [Menu] - [System maintenance] - [Basic status] interface and observe if CRP
Partial The CRP reaction cell
damaged or poorly connected; reaction tank temperature decreases;
02-91-18 function temperature is lower than
2. Heating band connecting cable 3. Check if wire B02-P01 is correctly connected on J01 interface of CRP test unit board;
stop level the upper temperature limit 4. Check if wire PH1201 is correctly connected on the CRP unit;
has short circuit.
5. Ensure the resistance of heating band is correct (smaller than 50Ω, theoretical value is around 24Ω).
1. Execute [Clear alarm], observe if the fault can be removed;
1. The temperature sensor is 2. Power on the instrument, enter [Menu] - [System maintenance] - [Basic status] interface and observe if cuvette
Temperature of cuvette is
Caution damaged or poorly connected; temperature rises;
02-91-19 lower than the lower
level 2. The heating band is poorly 3. Check if wire B02-P23 is correctly connected on J23 interface of CRP test unit board;
temperature limit 4. Check if wire PH1202 is correctly connected on the CRP unit;
connected or damaged.
5. Ensure the resistance of heating band is correct (smaller than 50Ω, theoretical value is around 41Ω).
1. Execute [Clear alarm], observe if the fault can be removed;
1. The temperature sensor is 2. Power on the instrument, enter [Menu] - [System maintenance] - [Basic status] interface and observe if cuvette
Temperature of cuvette is
Caution damaged or poorly connected; temperature decreases;
02-91-20 lower than the upper
level 2. Heating band connecting cable 3. Check if wire B02-P10 is correctly connected on J10 interface of CRP test unit board;
temperature limit 4. Check if wire PH1202 is correctly connected on the CRP unit;
has short circuit.
5. Ensure the resistance of heating band is correct (smaller than 50Ω, theoretical value is around 41Ω).
1. Execute [Clear alarm], observe if the fault can be removed;
2. Power on the instrument, enter [Menu] - [System maintenance] - [Basic status] interface and observe if cuvette
Caution Cuvette temperature sensor 1. The temperature sensor is temperature changes;
02-91-21
level fault damaged or not connected 3. Check if wire B02-P21 is correctly connected on J21 interface of CRP test unit board;
4. Check if wire PS1202 is correctly connected on the refrigeration chamber unit;
5. Replace the temperature sensor. (Wire B02-P22 can be plugged in J21 interface on CRP control board for use).
1. Execute [Clear alarm], observe if the CRP syringe has operation action and if the fault can be removed.
1. Poor soldering or damage of 2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Advanced] - [Optocoupler
connecting cable of CRP syringe detection], click [Mechanical resetting] and observe the operating status of CRP syringe.
motor; 3. If the CRP syringe does not operate, please check if the motor connecting cable and corresponding socket have
The syringe does not detach
2. CRP syringe motor zero-point poor soldering or are damaged; check if the U7 chip .U1(MCU) of sample feeder control board has poor
02-101-01 Stop level from the sensor position in
optocoupler signal line broken or soldering or is damaged.
the pulling down process 4. Shelter the slot of CRP syringe zero-point optocoupler with an object and observe the DS22 signal lamp on
damaged;
3. Poor soldering or damage of drive CRP test unit circuit board above U5. When it is not sheltered, DS22 will be on, or DS22 will be off. If CRP
syringe zero-point optocoupler has status error, please replace the zero-point optocoupler.
chip of CRP syringe motor.
5. If CRP syringe has too loose jackscrew, please tighten the jackscrew.
103
Service Manual of Automatic Hematology Analyzer
Alarm code Alarm level Fault description Fault cause analysis Solutions
1. Execute [Clear alarm], observe if the CRP syringe has operation action and if the fault can be removed.
1. Poor soldering or damage of 2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Advanced] - [Optocoupler
connecting cable of CRP syringe detection], click [Mechanical resetting] and observe the operating status of CRP syringe.
motor; 3. If the CRP syringe does not operate, please check if the motor connecting cable and corresponding socket have
The syringe does not reach
2. CRP syringe motor zero-point poor soldering or are damaged; check if the U7 chip .U1(MCU) of sample feeder control board has poor
02-101-02 Stop level sensor when returning to
optocoupler signal line broken or soldering or is damaged.
zero point 4. Shelter the slot of CRP syringe zero-point optocoupler with an object and observe the DS22 signal lamp on
damaged;
3. Poor soldering or damage of drive CRP test unit circuit board above U5. When it is not sheltered, DS22 will be on, or DS22 will be off. If CRP
syringe zero-point optocoupler has status error, please replace the zero-point optocoupler.
chip of CRP syringe motor.
5. If CRP syringe has too loose jackscrew, please tighten the jackscrew.
1. Execute [Clear alarm], observe if the CRP syringe has operation action and if the fault can be removed.
1. Poor soldering or damage of 2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Advanced] - [Optocoupler
connecting cable of CRP syringe detection], click [Mechanical resetting] and observe the operating status of CRP syringe.
motor; 3. If the CRP syringe does not operate, please check if the motor connecting cable and corresponding socket have
Syringe has step loss in 2. CRP syringe motor zero-point poor soldering or are damaged; check if the U7 chip .U1(MCU) of sample feeder control board has poor
02-101-03 Stop level
operating process optocoupler signal line broken or soldering or is damaged.
damaged; 4. Shelter the slot of CRP syringe zero-point optocoupler with an object and observe the DS22 signal lamp on
3. Poor soldering or damage of drive CRP test unit circuit board above U5. When it is not sheltered, DS22 will be on, or DS22 will be off. If CRP
syringe zero-point optocoupler has status error, please replace the zero-point optocoupler.
chip of CRP syringe motor.
5. If CRP syringe has too loose jackscrew, please tighten the jackscrew.
1. Execute [Clear alarm], observe if the CRP syringe has operation action and if the fault can be removed.
1. Poor soldering or damage of 2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Advanced] - [Optocoupler
connecting cable of refrigeration detection], click [Mechanical resetting] and observe the operating status of refrigeration chamber cover.
chamber cover motor; 3. If the refrigeration chamber cover does not operate, please check if the motor connecting cable and
The motor does not leave
2. Refrigeration chamber cover corresponding socket have poor soldering or are damaged; check if the U7 chip .U1(MCU) of sample feeder
sensor after being driven or
02-101-04 Stop level zero-point optocoupler signal line control board has poor soldering or is damaged.
fails to be closed in cover 4. Lift the refrigeration chamber cover and observe if the DS20 signal lamp on CRP test unit circuit board
broken or damaged;
closing process. above U5 is on, close the refrigeration chamber cover and observe if the DS20 signal lamp on circuit board
3. Poor soldering or damage of drive
chip of refrigeration chamber cover above U5 is off. If refrigeration chamber cover zero-point optocoupler has status error, please replace the
zero-point optocoupler.
motor. 5. If refrigeration chamber cover has too loose jackscrew, please tighten the jackscrew.
1. Execute [Clear alarm], observe if the CRP syringe has operation action and if the fault can be removed.
1. Poor soldering or damage of 2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Advanced] - [Optocoupler
connecting cable of refrigeration detection], click [Mechanical resetting] and observe the operating status of refrigeration chamber cover.
chamber cover motor; 3. If the refrigeration chamber cover does not operate, please check if the motor connecting cable and
The motor reaches the 2. Refrigeration chamber cover corresponding socket have poor soldering or are damaged; check if the U7 chip .U1(MCU) of sample feeder
02-101-05 Stop level maximum steps but not zero-point optocoupler signal line control board has poor soldering or is damaged.
detect sensor signal broken or damaged; 4. Lift the refrigeration chamber cover and observe if the DS20 signal lamp on CRP test unit circuit board
3. Poor soldering or damage of drive above U5 is on, close the refrigeration chamber cover and observe if the DS20 signal lamp on circuit board
chip of refrigeration chamber cover above U5 is off. If refrigeration chamber cover zero-point optocoupler has status error, please replace the
zero-point optocoupler.
motor. 5. If refrigeration chamber cover has too loose jackscrew, please tighten the jackscrew.
104
Service Manual of Automatic Hematology Analyzer
Alarm code Alarm level Fault description Fault cause analysis Solutions
1. Execute [Clear alarm], observe if the CRP syringe has operation action and if the fault can be removed;
2. If the alarm cannot be cleared, pull out test tube from CRP lyse bottle, enter [Menu] - [System maintenance] -
Too much air bubbles are [Advanced] - [Mechanical detection] interface, input the speed and steps of CRP syringe motor in [CRP], click
Partial 1. Insufficient CRP lyse; [Down] until the air bubbles are detected at the test tube air bubble optocoupler. Observe if DS27 is on, click [Up]
detected in reagent tubing
02-111-30 function 2. Poor soldering or damage of CRP until it is liquid at the test tube air bubble optocoupler position and observe if DS27 is off;
by the air bubble 3. If the lamp is not on, please check if the air bubble optocoupler connecting cable and corresponding socket
stop level lyse air bubble optocoupler.
optocoupler have poor soldering or are damaged; check if the U1(MCU) chip of CRP test unit board has poor soldering or is
damaged;
4. If the air bubble optocoupler has error in on and off, please replace the air bubble optocoupler.
1. C-reactive protein reagent 1. Place CRP reagent chamber back in position and ensure the reagent chamber has been installed in place;
Partial chamber is opened; 2. Execute [Clear alarm], observe if the fault can be removed;
CRP reagent chamber door
02-111-31 function 2. Poor soldering or damage of 3. Take out C-reactive protein reagent chamber and observe if the DS21 signal lamp on circuit board above U5 is
is not closed off, and observe if it is on after it is put back. If C-reactive protein reagent chamber door optocoupler has status
stop level C-reactive protein reagent chamber
door optocoupler. error, please replace the optocoupler.
105
Service Manual of Automatic Hematology Analyzer
Alarm code Alarm level Fault description Fault cause analysis Solutions
1. Execute [Clear alarm], observe if the fault can be removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] - [CRP],
click [CSV26] and observe if [CSV26] button turns yellow and if the DS16 near J5 is on with "KADA" sound.
1. Solenoid valve is not connected;
02-131-10 Stop level SV26 fault 3. If the lamp is on or generates no "KADA" sound, please check if the solenoid valve connecting cable and
2. Solenoid valve has short circuit. corresponding socket have poor soldering or are damaged;
4. If the lamp is not on , check if the U4 chip .U1(MCU) chip of sample feeder control board has poor soldering
or is damaged.
1. Execute [Clear alarm], observe if the fault can be removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Mechanical detection] - [CRP],
click [CSV27] and observe if [CSV27] button turns yellow and if the DS13 near J5 is on with "KADA" sound.
1. Solenoid valve is not connected;
02-131-11 Stop level SV27 fault 3. If the lamp is on or generates no "KADA" sound, please check if the solenoid valve connecting cable and
2. Solenoid valve has short circuit. corresponding socket have poor soldering or are damaged;
4. If the lamp is not on, check if the U4 chip .U1(MCU) chip of sample feeder control board has poor soldering or
is damaged.
Sample dispensing to 1. Sample dispensing to CRP
02-94-51 Stop level 1. Please execute alarm clearing operation.
channel 1 timeout channel 1 timeout.
C-reactive protein reagent 1. C-reactive protein reagent
02-94-52 Stop level dispensing to channel 1 dispensing to CRP channel 1 1. Please execute alarm clearing operation.
timeout timeout.
Sample dispensing to 1. Sample dispensing to CRP
02-94-53 Stop level 1. Please execute alarm clearing operation.
channel 2 timeout channel 2 timeout.
C-reactive protein reagent 1. C-reactive protein reagent
02-94-54 Stop level dispensing to channel 2 dispensing to CRP channel 2 1. Please execute alarm clearing operation.
timeout timeout.
106
Service Manual of Automatic Hematology Analyzer
Alarm code Alarm level Fault description Fault cause analysis Solutions
1. Execute [Clear alarm], observe if the rack-conveying motor has operation action and if the fault can be
1. Poor soldering or damage of removed;
The rack-conveying connecting cable of rack-conveying 2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Advanced] - [Optocoupler
mechanism does not operate motor; detection], click [Rack-conveying] and observe the operating status of rack-conveying motor.
Partial
to the rack-conveying 2. Rack-conveying zero-point 3. If the rack-conveying motor does not operate, please check if the motor connecting cable and corresponding
05-101-02 function
zero-point optoelectronic optoelectronic switch signal line socket have poor soldering or are damaged; check if the U6 chip .U1(MCU) of sample feeder control board and
stop level control part of the rack-conveying motor have poor soldering or are damaged.
switch detection position in broken or damaged;
specified steps 3. Poor soldering or damage of drive 4. Shelter the rack-conveying zero-point optoelectronic switch with an object and observe the DS13 signal
lamp on circuit board near J8. When it is not sheltered, DS13 will be on, or DS13 will be off. If rack-conveying
chip of rack-conveying motor.
zero-point optoelectronic switch has status error, please replace the zero-point optoelectronic switch.
1. Execute [Clear alarm], observe if the scanning motor has operation action and if the fault can be removed;
1. Poor soldering or damage of 2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Advanced] - [Optocoupler
Zero-point signal error of
connecting cable of scanning motor; detection], click [Bar code detection] and observe the operating status of loading motor.
scanning motor before
Partial 2. Scanning motor zero-point 3. If the loading motor does not operate, please check if the motor connecting cable and corresponding socket
operating or scanning motor
05-101-03 function optocoupler signal line broken or have poor soldering or are damaged; check if the U9 chip .U1(MCU) of sample feeder control board and control
does not reach scanning part of the loading motor have poor soldering or are damaged.
stop level damaged;
zero-point optocoupler 4. Shelter the slot of scanning motor zero-point optocoupler with an object and observe the DS14 signal lamp
3. Poor soldering or damage of drive
position after operating on circuit board near J8. When it is not sheltered, DS14 will be on, or DS14 will be off. If scanning motor
chip of scanning motor.
zero-point optocoupler has status error, please replace the zero-point optocoupler.
1. Poor soldering or damage of 1. Execute [Clear alarm], observe if the unloading motor has operation action and if the fault can be removed.
The unloading motor does connecting cable of unloading 2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Advanced] - [Optocoupler
not reach the motor; detection], click [Unload] and observe the operating status of unloading motor.
Partial 3. If the unloading motor does not operate, please check if the motor connecting cable and corresponding socket
rack-withdrawing 2. Unloading motor zero-point
05-101-04 function have poor soldering or are damaged; check if the U7 chip .U1(MCU) of sample feeder control board and control
zero-point optocoupler optocoupler signal line broken or part of the unloading motor have poor soldering or are damaged.
stop level
position within the specified damaged; 4. Shelter the slot of rack-withdrawing zero-point optocoupler with an object and observe the DS15 signal lamp
steps 3. Poor soldering or damage of drive on circuit board near J8. When it is not sheltered, DS15 will be on, or DS15 will be off. If rack-withdrawing
chip of unloading motor. zero-point optocoupler has status error, please replace the zero-point optocoupler.
1. Execute [Clear alarm], observe if the mixing motor has operation action and if the fault can be removed.
1. Poor soldering or damage of 2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Advanced] - [Optocoupler
connecting cable of mixing motor; detection], click [Up], [Down] and observe the operating status of mixing motor.
The mixing motor does not
Partial 2. Lifting motor zero-point 3. If the mixing motor does not operate, please check if the motor connecting cable and corresponding socket
reach the lifting zero-point
05-101-05 function optocoupler signal line broken or have poor soldering or are damaged; check if the U4 chip .U1(MCU) of sample feeder control board and control
optocoupler position within part of the mixing motor have poor soldering or are damaged.
stop level damaged;
the specified steps 4. Shelter the slot of lifting zero-point optocoupler with an object and observe the DS19 signal lamp on circuit
3. Poor soldering or damage of drive
chip of mixing motor. board near J19. When it is not sheltered, DS19 will be on, or DS19 will be off. If lifting zero-point optocoupler
has status error, please replace the zero-point optocoupler.
1. Execute [Clear alarm], observe if the clamping motor has operation action and if the fault can be removed.
1. Poor soldering or damage of 2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Advanced] - [Optocoupler
The clamping zero-point connecting cable of clamping motor; detection], click [Mechanical resetting] and observe the operating status of clamping motor.
Partial 3. If the clamping motor does not operate, please check if the motor connecting cable and corresponding socket
optocoupler does not detect 2. Clamping zero-point optocoupler
05-101-06 function have poor soldering or are damaged; check if the U3 chip .U1(MCU) of sample feeder control board and control
signal change within signal line broken or damaged; part of the clamping motor have poor soldering or are damaged.
stop level
specified steps 3. Poor soldering or damage of drive 4. Shelter the slot of clamping zero-point optocoupler with an object and observe the DS20 signal lamp on
chip of clamping motor. circuit board near J8. When it is not sheltered, DS20 will be on, or DS20 will be off. If clamping zero-point
optocoupler has status error, please replace the zero-point optocoupler.
107
Service Manual of Automatic Hematology Analyzer
Alarm code Alarm level Fault description Fault cause analysis Solutions
1. Execute [Clear alarm], observe if the STAT motor has operation action and if the fault can be removed.
1. Poor soldering or damage of 2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Advanced] - [Optocoupler
The signal that the STAT connecting cable of STAT motor; detection], click [Feed sample], [Withdraw sample] and observe the operating status of STAT motor.
Partial 3. If the STAT motor does not operate, please check if the motor connecting cable and corresponding socket have
zero-point switch is pressed 2. STAT zero-point switch signal
05-101-07 function poor soldering or are damaged; check if the U5 chip .U1(MCU) of sample feeder control board and control part
down is not detected within line broken or damaged; of the STAT motor have poor soldering or are damaged.
stop level
specified steps 3. Poor soldering or damage of 4. Press the STAT motor zero-position switch with objects to observe the DS7 signal light besides J19 on the
drive chip of STAT motor. circuit board, when the switch is closed, DS7 is on, otherwise, DS7 is off. If the zero position switch of STAT
motor is error, replace the zero position switch.
1. Execute [Clear alarm], observe if the STAT motor has operation action and if the fault can be removed.
1. Poor soldering or damage of 2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Advanced] - [Optocoupler
The signal that the STAT connecting cable of STAT motor; detection], click [Feed sample], [Withdraw sample] and observe the operating status of STAT motor.
Partial 3. If the STAT motor does not operate, please check if the motor connecting cable and corresponding socket have
in-position switch is pressed 2. STAT in-position switch signal
05-101-08 function poor soldering or are damaged; check if the U5 chip .U1(MCU) of sample feeder control board and control part
down is not detected within line broken or damaged; of the STAT motor have poor soldering or are damaged.
stop level
specified steps 3. Poor soldering or damage of 4. Press the STAT motor in-position switch with objects to observe the DS8 signal light besides J19 on the
drive chip of STAT motor. circuit board, when the switch is closed, DS8 is on, otherwise, DS8 is off. If the STAT in-position switch has
status error, replace the zero position switch.
1. Poor soldering or damage of 1. Execute [Clear alarm], observe if the mixing motor has operation action and if the fault can be removed.
connecting cable of mixing motor; 2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Advanced] - [Optocoupler
The steps difference of the 2. Mixing zero-point optocoupler detection], click [Up], [Down] and observe the operating status of mixing motor.
Partial 3. If the mixing motor does not operate, please check if the motor connecting cable and corresponding socket
mixing mechanism is signal line broken or damaged;
05-101-09 function have poor soldering or are damaged; check if the U4 chip .U1(MCU) of sample feeder control board and control
greater than theoretical 3. Poor soldering or damage of drive part of the mixing motor have poor soldering or are damaged.
stop level
value chip of mixing motor; 4. Shelter the slot of lifting zero-point optocoupler with an object and observe the DS19 signal lamp on circuit
4. Check if the lifting mechanism board near J19. When it is not sheltered, DS19 will be on, or DS19 will be off. If lifting zero-position
has lubricant. optocoupler has status error, please replace the zero-point optocoupler.
1. Execute [Clear alarm], observe if the clamping motor has operation action and if the fault can be removed.
1. Poor soldering or damage of 2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Advanced] - [Optocoupler
Extending or clamping steps connecting cable of clamping motor; detection], click [Mechanical resetting] and observe the operating status of clamping motor.
Partial 3. If the clamping motor does not operate, please check if the motor connecting cable and corresponding socket
of clamping mechanism are 2. Clamping zero-point optocoupler
05-101-10 function have poor soldering or are damaged; check if the U3 chip .U1(MCU) of sample feeder control board and control
inconsistent with the signal line broken or damaged; part of the clamping motor have poor soldering or are damaged.
stop level
theoretical operating steps 3. Poor soldering or damage of drive 4. Shelter the slot of clamping zero-point optocoupler with an object and observe the DS20 signal lamp on
chip of clamping motor. circuit board near J8. When it is not sheltered, DS20 will be on, or DS20 will be off. If clamping zero-point
optocoupler has status error, please replace the zero-point optocoupler.
1. Mechanical structure of count ball
Partial The count ball is in lifted is loose; 1. If the count ball is loose, use tool to fix the count ball;
2. If the scanning motor causes movement of sample rack, detect if the scanning motor is loose or has position
05-101-11 function status before executing 2. Sample rack is moved when error. If it is loose, please fix the scanning motor. If it is position error, please restore the position of scanning
stop level mixing scanning motor has bar code motor.
detection.
108
Service Manual of Automatic Hematology Analyzer
Alarm code Alarm level Fault description Fault cause analysis Solutions
1. Execute [Clear alarm], observe if the clamping motor has operation action and if the fault can be removed.
1. The spring of clamping motor is 2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Advanced] - [Optocoupler
too tight or too loose; detection], click [Mechanical resetting] and observe the operating status of clamping motor.
When the mixing
2. Poor soldering or damage of 3. If the clamping motor does not operate, please check if the motor connecting cable and corresponding socket
Partial mechanism is operating,
connecting cable of clamping motor; have poor soldering or are damaged; check if the U3 chip .U1(MCU) of sample feeder control board and control
05-101-12 function abnormal clamping motor
3. Clamping zero-point optocoupler part of the clamping motor have poor soldering or are damaged.
stop level zero-point signal change is 4. Shelter the slot of clamping zero-point optocoupler with an object and observe the DS20 signal lamp on
signal line broken or damaged;
detected circuit board near J8. When it is not sheltered, DS20 will be on, or DS20 will be off. If clamping zero-point
4. Poor soldering or damage of drive
optocoupler has status error, please replace the zero-point optocoupler;
chip of clamping motor.
5. If the spring of clamping motor is too tight or too loose, please replace the spring.
1. Execute [Clear alarm], observe if the loading motor has operation action and if the fault can be removed.
1. Poor soldering or damage of 2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Advanced] - [Optocoupler
connecting cable of loading motor; detection], click [Load] and observe the operating status of loading motor.
Partial Loading motor operation 2. Rack-feeding in-position 3. If the loading motor does not operate, please check if the motor connecting cable and corresponding socket
05-101-17 function completion mark not optocoupler signal line broken or have poor soldering or are damaged; check if the U8 chip .U1(MCU) of sample feeder control board and control
stop level detected in specified time damaged; part of the loading motor have poor soldering or are damaged;
3. Poor soldering or damage of drive 4. Shelter the rack-feeding in-position optocoupler with an object and observe the DS12 signal lamp on circuit
chip of loading motor. board near J19. When it is not sheltered, DS12 will be on, or DS12 will be off. If the rack-feeding in-position
optocoupler has status error, replace the optocoupler.
1. Execute [Clear alarm], observe if the rack-conveying motor has operation action and if the fault can be
1. Poor soldering or damage of removed;
connecting cable of rack-conveying 2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Advanced] - [Optocoupler
Rack-conveying motor motor; detection], click [Rack-conveying] and observe the operating status of rack-conveying motor.
Partial
operation completion mark 2. Rack-conveying zero-point 3. If the rack-conveying motor does not operate, please check if the motor connecting cable and corresponding
05-101-18 function
not detected in specified optoelectronic switch signal line socket have poor soldering or are damaged; check if the U6 chip .U1(MCU) of sample feeder control board and
stop level control part of the rack-conveying motor have poor soldering or are damaged.
time broken or damaged;
3. Poor soldering or damage of 4. Shelter the rack-conveying zero-point optoelectronic switch with an object and observe the DS13 signal lamp
on circuit board near J8. When it is not sheltered, DS13 will be on, or DS13 will be off. If rack-conveying
drive chip of rack-conveying motor.
zero-point optoelectronic switch has status error, please replace the zero-point optocoupler.
1. Execute [Clear alarm], observe if the scanning motor has operation action and if the fault can be removed;
1. Poor soldering or damage of 2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Advanced] - [Optocoupler
connecting cable of scanning motor; detection], click [Bar code detection] and observe the operating status of scanning motor.
Partial Scanning motor operation 2. Scanning motor zero-point 3. If the scanning motor does not operate, please check if the motor connecting cable and corresponding socket
05-101-19 function completion mark not optocoupler signal line broken or have poor soldering or are damaged; check if the U9 chip .U1(MCU) of sample feeder control board and control
stop level detected in specified time damaged; part of the scanning motor have poor soldering or are damaged.
3. Poor soldering or damage of 4. Shelter the slot of scanning zero-point optocoupler with an object and observe the DS14 signal lamp on circuit
drive chip of scanning motor. board near J8. When it is not sheltered, DS14 will be on, or DS14 will be off. If scanning zero-point optocoupler
has status error, please replace the zero-point optocoupler.
109
Service Manual of Automatic Hematology Analyzer
Alarm code Alarm level Fault description Fault cause analysis Solutions
1. Execute [Clear alarm], observe if the unloading motor has operation action and if the fault can be
removed.
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Advanced] -
1. Poor soldering or damage of
[Optocoupler detection], click [Rack-conveying] and observe the operating status of unloading
connecting cable of unloading
motor.
motor;
Partial Unloading motor operation 3. If the unloading motor does not operate, please check if the motor connecting cable and
2. Unloading motor zero-point
05-101-20 function completion mark not corresponding socket have poor soldering or are damaged; check if the U7 chip .U1(MCU) of
optocoupler signal line broken or
stop level detected in specified time sample feeder control board and control part of the unloading motor have poor soldering or are
damaged;
damaged.
3. Poor soldering or damage of
4. Shelter the slot of rack-withdrawing zero-point optocoupler with an object and observe the DS15
drive chip of unloading motor.
signal lamp on circuit board near J8. When it is not sheltered, DS15 will be on, or DS15 will be off.
If rack-withdrawing zero-point optocoupler has status error, please replace the zero-point
optocoupler.
1. Execute [Clear alarm], observe if the STAT motor has operation action and if the fault can be
removed.
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Advanced] -
1. Poor soldering or damage of
[Optocoupler detection], click [Feed sample], [Withdraw sample] and observe the operating status
connecting cable of STAT motor;
Partial STAT motor operation of STAT motor.
2. STAT in-position switch signal
05-101-21 function completion mark not 3. If the STAT motor does not operate, please check if the motor connecting cable and corresponding
line broken or damaged;
stop level detected in specified time socket have poor soldering or are damaged; check if the U5 chip .U1(MCU) of sample feeder control
3. Poor soldering or damage of
board and control part of the STAT motor have poor soldering or are damaged.
drive chip of STAT motor.
4. Press the STAT motor in-position switch with objects to observe the DS8 signal light besides J19
on the circuit board, when the switch is closed, DS8 is on, otherwise, DS8 is off. If the zero position
switch of STAT motor is error, replace the zero position switch.
1. Execute [Clear alarm], observe if the mixing motor has operation action and if the fault can be
removed.
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Advanced] -
1. Poor soldering or damage of
[Optocoupler detection], click [Up], [Down] and observe the operating status of mixing motor.
connecting cable of mixing motor;
Partial Mixing motor operation 3. If the mixing motor does not operate, please check if the motor connecting cable and
2. Mixing zero-point optocoupler
05-101-22 function completion mark not corresponding socket have poor soldering or are damaged; check if the U4 chip .U1(MCU) of
signal line broken or damaged;
stop level detected in specified time sample feeder control board and control part of the mixing motor have poor soldering or are
3. Poor soldering or damage of drive
damaged.
chip of mixing motor.
4. Shelter the slot of lifting zero-point optocoupler with an object and observe the DS19 signal lamp
on circuit board near J19. When it is not sheltered, DS19 will be on, or DS19 will be off. If lifting
zero-point optocoupler has status error, please replace the zero-point optocoupler.
110
Service Manual of Automatic Hematology Analyzer
Alarm code Alarm level Fault description Fault cause analysis Solutions
1. Execute [Clear alarm], observe if the clamping motor has operation action and if the fault can be
removed.
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Advanced] -
1. Poor soldering or damage of [Optocoupler detection], click [Mechanical resetting] and observe the operating status of clamping
connecting cable of clamping motor; motor.
Partial Clamping motor operation
2. Clamping zero-point optocoupler 3. If the clamping motor does not operate, please check if the motor connecting cable and
05-101-23 function completion mark not
signal line broken or damaged; corresponding socket have poor soldering or are damaged; check if the U3 chip .U1(MCU) of
stop level detected in specified time
3. Poor soldering or damage of drive sample feeder control board and control part of the clamping motor have poor soldering or are
chip of clamping motor. damaged.
4. Shelter the slot of clamping zero-point optocoupler with an object and observe the DS20 signal
lamp on circuit board near J8. When it is not sheltered, DS20 will be on, or DS20 will be off. If
clamping zero-point optocoupler has status error, please replace the zero-point optocoupler.
1. Tube judging sensor signal line
1. Shelter the tube judging sensor with an object and observe the D3 signal lamp on circuit board
Partial broken or damaged;
Abnormal tube judging near J9. When it is not sheltered, D3 will be on, or D3 will be off. If the tube judging sensor has
05-111-01 function 2. The tube judging sensor is moved;
sensor signal status error, please replace the tube judging sensor.
stop level 3. The tube judging sensor is
2. If the position of tube judging sensor is incorrect, please correct its position.
damaged.
1. Execute [Clear alarm], observe if the rack-conveying motor has operation action and if the fault
can be removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Advanced] -
1. Poor soldering or damage of
[Optocoupler detection], click [Rack-conveying] and observe the operating status of rack-conveying
connecting cable of rack-conveying
The zero-point motor.
motor;
Partial optoelectronic switch is still 3. If the rack-conveying motor does not operate, please check if the motor connecting cable and
2. Rack-conveying zero-point
05-111-02 function sheltered after the corresponding socket have poor soldering or are damaged; check if the U6 chip .U1(MCU) of
optoelectronic switch signal line
stop level rack-conveying action is sample feeder control board and control part of the rack-conveying motor have poor soldering or are
broken or damaged;
executed damaged.
3. Poor soldering or damage of drive
4. Shelter the rack-conveying zero-point optoelectronic switch with an object and observe the DS13
chip of rack-feeding motor.
signal lamp on circuit board near J8. When it is not sheltered, DS13 will be on, or DS13 will be off.
If rack-conveying zero-point optoelectronic switch has status error, please replace the zero-point
optocoupler.
1. Execute [Clear alarm], observe if the scanning motor has operation action and if the fault can be
removed;
1. Poor soldering or damage of
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Advanced] -
connecting cable of scanning motor;
[Optocoupler detection], click [Bar code detection] and observe the operating status of scanning
2. Scanning motor zero-point
motor.
Partial The zero-point optocoupler optocoupler signal line broken or
3. If the scanning motor does not operate, please check if the motor connecting cable and
05-111-03 function is still sheltered after the damaged;
corresponding socket have poor soldering or are damaged; check if the U9 chip .U1(MCU) of
stop level scanning action is executed 3. Poor soldering or damage of drive
sample feeder control board and control part of the scanning motor have poor soldering or are
chip of scanning motor;
damaged.
4. The installation position of the
4. Shelter the slot of scanning zero-point optocoupler with an object and observe the DS14 signal
scanning motor is incorrect.
lamp on circuit board near J8. When it is not sheltered, DS14 will be on, or DS14 will be off. If
scanning zero-point optocoupler has status error, please replace the zero-point optocoupler.
111
Service Manual of Automatic Hematology Analyzer
Alarm code Alarm level Fault description Fault cause analysis Solutions
1. Execute [Clear alarm], observe if the unloading motor has operation action and if the fault can be
removed.
1. Poor soldering or damage of 2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Advanced] -
connecting cable of unloading [Optocoupler detection], click [Unload] and observe the operating status of unloading motor.
The zero-point optocoupler motor; 3. If the unloading motor does not operate, please check if the motor connecting cable and
Partial
is still sheltered after the 2. Unloading motor zero-point corresponding socket have poor soldering or are damaged; check if the U7 chip .U1(MCU) of
05-111-04 function
rack withdrawing action is optocoupler signal line broken or sample feeder control board and control part of the unloading motor have poor soldering or are
stop level
executed damaged; damaged.
3. Poor soldering or damage of drive 4. Shelter the slot of rack-withdrawing zero-point optocoupler with an object and observe the DS15
chip of unloading motor. signal lamp on circuit board near J8. When it is not sheltered, DS15 will be on, or DS15 will be off.
If rack-withdrawing zero-point optocoupler has status error, please replace the zero-point
optocoupler.
1. Execute [Clear alarm], observe if the mixing motor has operation action and if the fault can be
removed.
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Advanced] -
1. Poor soldering or damage of
[Optocoupler detection], click [Up], [Down] and observe the operating status of mixing motor.
The zero-point optocoupler connecting cable of mixing motor;
Partial 3. If the mixing motor does not operate, please check if the motor connecting cable and
is still sheltered after the 2. Mixing zero-point optocoupler
05-111-05 function corresponding socket have poor soldering or are damaged; check if the U4 chip .U1(MCU) of
mixing mechanism makes signal line broken or damaged;
stop level sample feeder control board and control part of the mixing motor have poor soldering or are
action 3. Poor soldering or damage of drive
damaged.
chip of mixing motor.
4. Shelter the slot of lifting zero-point optocoupler with an object and observe the DS19 signal lamp
on circuit board near J19. When it is not sheltered, DS19 will be on, or DS19 will be off. If lifting
zero-position optocoupler has status error, please replace the zero-point optocoupler.
1. Execute [Clear alarm], observe if the unloading motor has operation action and if the fault can be
removed.
1. Poor soldering or damage of
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Advanced] -
connecting cable of unloading
The rack [Optocoupler detection], click [Unload] and observe the operating status of unloading motor.
motor;
Partial conveying-to-position 3. If the unloading motor does not operate, please check if the motor connecting cable and
2. Unloading motor zero-point
05-111-06 function optocoupler is still sheltered corresponding socket have poor soldering or are damaged; check if the U7 chip .U1(MCU) of
optocoupler signal line broken or
stop level after the rack withdrawing sample feeder control board and control part of the unloading motor have poor soldering or are
damaged;
action is executed damaged.
3. Poor soldering or damage of drive
4. Shelter the slot of rack-conveying in-position optocoupler with an object and observe the DS16
chip of unloading motor.
signal lamp on circuit board near J8. When it is not sheltered, DS16 will be on, or DS16 will be off.
If the rack-conveying in-position optocoupler has status error, replace the in-position optocoupler.
112
Service Manual of Automatic Hematology Analyzer
Alarm code Alarm level Fault description Fault cause analysis Solutions
1. Execute [Clear alarm], observe if the rack-conveying motor has operation action and if the fault
can be removed;
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Advanced] -
1. Poor soldering or damage of
[Optocoupler detection], click [Rack-conveying] and observe the operating status of rack-conveying
connecting cable of rack-conveying
motor.
motor;
Partial Rack-conveying initial 3. If the rack-conveying motor does not operate, please check if the motor connecting cable and
2. Rack-feeding in-position
05-111-07 function sensor is sheltered in the corresponding socket have poor soldering or are damaged; check if the U6 chip .U1(MCU) of
optocoupler signal line broken or
stop level process of conveying back sample feeder control board and control part of the rack-conveying motor have poor soldering or are
damaged;
damaged.
3. Poor soldering or damage of drive
4. Shelter the rack-feeding in-position optocoupler with an object and observe the DS12 signal lamp
chip of rack-conveying motor.
on circuit board near J2. When it is not sheltered, DS12 will be on, or DS12 will be off. If the
rack-feeding in-position optocoupler has status error, please replace the rack-feeding in-position
optocoupler.
1. Execute [Clear alarm], observe if the clamping motor has operation action and if the fault can be
removed.
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Advanced] -
1. The spring of clamping motor is
[Optocoupler detection], click [Mechanical resetting] and observe the operating status of clamping
too tight or too loose;
motor.
2. Poor soldering or damage of
Zero-point optocoupler 3. If the clamping motor does not operate, please check if the motor connecting cable and
Partial connecting cable of clamping motor;
status of clamping corresponding socket have poor soldering or are damaged; check if the U3 chip .U1(MCU) of
05-111-08 function 3. Clamping zero-point optocoupler
mechanism is inconsistent sample feeder control board and control part of the clamping motor have poor soldering or are
stop level signal line broken or damaged;
with the theoretical status damaged.
4. Poor soldering or damage of drive
4. Shelter the slot of clamping zero-point optocoupler with an object and observe the DS20 signal
chip of clamping motor;
lamp on circuit board near J8. When it is not sheltered, DS20 will be on, or DS20 will be off. If
5. Incorrect installation of partition.
clamping zero-point optocoupler has status error, please replace the zero-point optocoupler;
5. If the spring of clamping motor is too tight or too loose, please replace the spring;
6. If the partition position is incorrect, change the partition position.
1. Execute [Clear alarm], observe if the STAT motor has operation action and if the fault can be
removed.
1. Poor soldering or damage of 2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Advanced] -
Abnormal tube connecting cable of STAT motor; [Optocoupler detection], click [Feed sample], [Withdraw sample] and observe the operating status
Partial holding-to-position 2. Tube holding-to-position of STAT motor.
05-111-09 function optocoupler signal after the optocoupler signal line broken or 3. If the STAT motor does not operate, please check if the motor connecting cable and corresponding
stop level STAT mechanism finishes damaged; socket have poor soldering or are damaged; check if the U5 chip .U1(MCU) of sample feeder control
movement 3. Poor soldering or damage of drive board and control part of the STAT motor have poor soldering or are damaged.
chip of STAT motor. 4. Shelter the tube holding-to-position optocoupler with objects to observe the DS8 signal light
besides J19 on the circuit board, when the switch is closed, DS6 is on, otherwise, DS6 is off. If the
tube holding-to-position optocoupler has status error, replace the in-position optocoupler.
113
Service Manual of Automatic Hematology Analyzer
Alarm code Alarm level Fault description Fault cause analysis Solutions
1. Execute [Clear alarm], observe if the loading motor has operation action and if the fault can be
removed.
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Advanced] -
1. Poor soldering or damage of
[Optocoupler detection], click [Load] and observe the operating status of loading motor.
Zero-point signal still exits connecting cable of loading motor;
Partial 3. If the loading motor does not operate, please check if the motor connecting cable and
after the rack feeding 2. Rack-feeding zero-point switch
05-151-01 function corresponding socket have poor soldering or are damaged; check if the U8 chip .U1(MCU) of
mechanism operates for signal line broken or damaged;
stop level sample feeder control board and control part of the loading motor have poor soldering or are
fixed steps 3. Poor soldering or damage of drive
damaged;
chip of loading motor.
4. Shelter the rack-feeding in-position optocoupler with an object and observe the DS12 signal lamp
on circuit board near J19. When it is not sheltered, DS12 will be on, or DS12 will be off. If the
rack-feeding in-position optocoupler has status error, replace the optocoupler.
1. Execute [Clear alarm], observe if the STAT motor has operation action and if the fault can be
removed.
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Advanced] -
1. Poor soldering or damage of
[Optocoupler detection], click [Feed sample], [Withdraw sample] and observe the operating status
connecting cable of STAT motor;
Partial of STAT motor.
Abnormal STAT chamber 2. STAT motor zero-point switch
05-151-02 function 3. If the STAT motor does not operate, please check if the motor connecting cable and corresponding
zero-point signal signal line broken or damaged;
stop level socket have poor soldering or are damaged; check if the U5 chip .U1(MCU) of sample feeder control
3. Poor soldering or damage of drive
board and control part of the STAT motor have poor soldering or are damaged.
chip of STAT motor.
4. Press the STAT motor zero-position switch with objects to observe the DS7 signal light besides
J19 on the circuit board, when the switch is closed, DS7 is on, otherwise, DS7 is off. If the zero
position switch of STAT motor is error, replace the zero position switch.
1. Execute [Clear alarm], observe if the STAT motor has operation action and if the fault can be
removed.
2. After the instrument is powered on, enter [Menu] - [System maintenance] - [Advanced] -
1. Poor soldering or damage of
[Optocoupler detection], click [Feed sample], [Withdraw sample] and observe the operating status
connecting cable of STAT motor;
Partial of STAT motor.
Abnormal STAT chamber 2. STAT motor in-position switch
05-151-03 function 3. If the STAT motor does not operate, please check if the motor connecting cable and corresponding
in-position switch signal signal line broken or damaged;
stop level socket have poor soldering or are damaged; check if the U5 chip .U1(MCU) of sample feeder control
3. Poor soldering or damage of drive
board and control part of the STAT motor have poor soldering or are damaged.
chip of STAT motor.
4. Press the STAT motor in-position switch with objects to observe the DS8 signal light besides J19
on the circuit board, when the switch is closed, DS8 is on, otherwise, DS8 is off. If the STAT
in-position switch has status error, replace the in-position switch.
1. There is foreign matter on rack 1. If there is foreign matter on rack surface, please remove the foreign matter and then operate;
Partial The left count ball does not surface; 2. If the count ball is loose, please fix the count ball;
05-151-04 function drop in the slot of test tube 2. The left count ball is loose; 3. Please click the rack conveying button through optocoupler detection interface and observe if the
stop level rack after rack conveying 3. Rack-conveying motor shifting shifting fork operates synchronously. If the shifting fork of rack conveying motor does not operate
fork is in non-synchronous state. synchronously, adjust the shifting fork to synchronous status.
1. There is foreign matter on rack 1. If there is foreign matter on rack surface, please remove the foreign matter and then operate;
Partial The right count ball does not surface; 2. If the count ball is loose, please fix the count ball;
05-151-05 function drop in the slot of test tube 2. The right count ball is loose; 3. Please click the rack conveying button through optocoupler detection interface and observe if the
stop level rack after rack conveying 3. Rack-conveying motor shifting shifting fork operates synchronously. If the shifting fork of rack conveying motor does not operate
fork is in non-synchronous state. synchronously, adjust the shifting fork to synchronous status.
114
Service Manual of Automatic Hematology Analyzer
Alarm code Alarm level Fault description Fault cause analysis Solutions
1. Press the count ball microswitch with objects to observe the DS11 signal light besides J19 on the
Partial The left count ball does not 1. Count ball microswitch signal line
circuit board, when the switch is closed, DS11 is on, otherwise, DS11 is off. If the zero position
05-151-06 function detect signal change after broken or damaged;
switch of rack-feeding motor is error, replace the zero position switch of rack-feeding motor;
stop level rack conveying 2. The left count ball is loose.
2. If the count ball is loose, please fix the count ball.
1. Press the count ball microswitch with objects to observe the DS10 signal light besides J19 on the
Partial The right count ball does not 1. Count ball microswitch signal line
circuit board, when the switch is closed, DS10 is on, otherwise, DS10 is off. If the zero position
05-151-07 function detect signal change after broken or damaged;
switch of rack-feeding motor is error, replace the zero position switch of rack-feeding motor;
stop level rack conveying 2. The right count ball is loose.
2. If the count ball is loose, please fix the count ball.
1. Press the count ball microswitch with objects to observe the DS10 or DS11 signal light besides
J19 on the circuit board, when the switch is closed, DS10 or DS11 is on, otherwise, DS10 or DS11 is
1. Count ball microswitch signal line
Partial The count ball detects signal off. If the zero position switch of rack-feeding motor is error, replace the zero position switch of
broken or damaged;
05-201-01 function change abnormally for rack-feeding motor;
2. Sample rack is moved during the
stop level non-rack conveying action 2. If the count ball is loose, please fix the count ball;
scanning process of scanning motor.
3. If the scanning motor moves the sample rack, check if the scanning motor is at a correct position.
If the position is incorrect, please correct its position.
1. Check if the optocoupler in the reclaiming area is sheltered and if there is small object in front of
1. Reclaiming area full judging
Partial the optocoupler in the reclaiming area. If yes, remove the object;
Reclaiming area full judging optocoupler signal line broken or
05-201-02 function 2. Shelter the reclaiming area full judging optocoupler with an object and observe the DS9 signal
sensor is sheltered damaged;
stop level lamp on circuit board near J19. When it is not sheltered, DS9 will be on, or DS9 will be off. If the
2. Reclaiming area is full.
reclaiming area full optocoupler has error, replace the optocoupler.
Partial Rack feeding-to-position 1. Rack-feeding in-position 1. Shelter the rack-feeding in-position optocoupler with an object and observe the DS12 signal lamp
05-201-03 function optocoupler detects no test optocoupler signal line broken or on circuit board near J19. When it is not sheltered, DS12 will be on, or DS12 will be off. If the
stop level tube rack damaged; conveying initial status has error, please replace the optocoupler.
1. Check if the bar code pasted on test tube is wrinkled;
2. If the bar code is not wrinkled, observe if the test tube rack is at a correct position; 1) If the test
tube rack is not at a correct position, put it at a correct position first and then rotate the roller of
Partial
Failure of the bar code reader to scan scanning motor. If the roller interferes with the test tube, please adjust the position of the roller. In
05-201-04 function Bar code scanning failed
the bar code case of too large resistance in rotation, please adjust the position of the roller when it is rotated. 2) If
stop level
the test tube rack is at a correct position, please check if the count ball is in a correct status. Check the
count ball according to signal lamp DS11 and DS10 and if it is in an incorrect status, rectify the
position of the count ball.
115
Chapter 9 List of Spare Parts
116
17 Tube 3 2014038 1
18 Tube 4 2013949 1
19 Tube 5 1044514 1
20 Tube 6 2013834 1
21 Tube 7 2013799 1
22 Tube 8 2016199 1
23 Tube 9 2013837 1
24 Tube 10 2013943 1
25 Tube 11 2014006 1
26 Tube 12 2013840 1
27 Tube 13 2013055 1
28 Tube 14 2013800 1
29 Tube 16 2013802 1
30 Tube 17 2013803 1
31 Tube 19 2013886 1
32 Tube 20 2013805 1
33 Tube 21 2013849 1
34 Tube 22 2013865 1
35 Tube 24 2013852 1
36 Tube 25 2013807 1
37 Tube 26 2013854 1
38 Tube 27 2013945 1
39 Tube 28 2013832 1
40 Tube 29 2013857 1
41 Tube 30 2013950 1
42 Tube 31 2013859 1
43 Tube 32 2013860 1
44 Tube 33 2014205 1
45 Tube 34 2013808 1
46 Tube 35 2012700 1
47 Tube 36 2013863 1
48 Tube 37 2013836 1
49 Tube 38 2013866 1
50 Tube 39 2014715 1
51 Tube 40 2013868 1
52 Tube 41 2013944 1
53 Tube 42 2013869 1
54 Tube 43 2013870 1
55 Tube 44 2014148 1
117
56 Tube 45 2014716 1
57 Tube 46 2013810 1
58 Tube 47 2013874 1
59 Tube 48 2013811 1
60 Tube 49 2016226 1
61 Tube 50 2012699 1
62 Tube 51 2014008 1
63 Tube 52 2014009 1
64 Tube 53 2014010 1
65 Tube 54 2013815 1
66 Tube 55 2014719 1
67 Tube 56 2013009 1
68 Tube 57 2013883 1
69 Tube 58 2013816 1
70 Tube 59 2013885 1
71 Tube 60 2013946 1
72 Tube 61 2013886 1
73 Tube 62 1044515 1
74 Tube 63 2013887 1
75 Tube 64 2013888 1
76 Tube 65 2011989 1
77 Tube 66 2011994 1
78 Tube 67 2013890 1
79 Tube 68 2012717 1
80 Tube 69 2016147 1
81 Tube 70 2013947 1
82 Tube 71 2013447 1
83 Tube 72 2013895 1
84 Tube 73 2013948 1
85 Tube 74 2012395 1
86 Tube 77 2013817 1
87 Tube 78 2013818 1
88 Tube 81 2013904 1
89 Tube 82 2013905 1
90 Tube 83 2013857 1
91 Tube 84 2013907 1
92 Tube 85 2013908 1
93 Tube 86 2013909 1
94 Tube 89 2013874 1
118
95 Tube 90 2013945 1
96 Tube 91 2013061 1
97 Tube 92 2013915 1
98 Tube 95 2013918 1
99 Tube 96 2014720 1
100 Tube 97 2013920 1
101 Tube 98 2014140 1
102 Tube 99 2012962 1
103 Tube 100 2012959 1
104 Tube 101 2012960 1
105 Tube 102 2012961 1
106 Tube 103 2013926 1
107 Tube 104 2013927 1
108 Tube 105 2013928 1
109 Tube 106 2013929 1
110 Tube 107 2013930 1
111 Tube 108 2013931 1
112 Tube 109 2014143 1
113 Tube 110 2014144 1
114 Tube 111 2014145 1
115 Tube 112 2014146 1
116 Tube 113 2014036 1
117 Tube 114 2013936 1
118 Tube 115 2013937 1
119 Tube 116 2013938 1
120 Tube 117 2013939 1
121 Tube 118 2012645 1
122 Tube 119 2013940 1
123 Tube 120 2013941 1
124 Tube 121 2011849 1
125 Tube 122 2017056 1
126 Tube 123 2013887 1
127 Tube 124 2013909 1
128 π Connector 1020205 5
129 T-Connector 1009673 4
130 Quick Bulkhead Union 1009567 1
131 T-Connector 1009631 1
132 L-Reducer 1020207 5
133 T-Reducer 1032593 3
119
134 Y-Connector 1009549 2
135 Value Connector 1009580 1
136 Straight-cutting Reducer 1009628 5
137 Connector 1009629 1
138 L-Quick Connector 1009620 1
139 Tube Inversion Connector 1021518 1
140 Tube Connector 1009590 5
141 Tube Connector 1029144 3
142 Clamp 1008072 5
143 Clamp 1008074 4
144 L-Connector 1009674 1
145 Tube Connector 1009564 2
146 Nut 1009563 2
147 Diaphragm-replaceable Probe Filter 1009881 5
18 Tube 1 2017241 1
19 Tube 2 2013830 1
120
20 Tube 3 2014038 1
21 Tube 4 2013949 1
22 Tube 5 1044514 1
23 Tube 6 2013834 1
24 Tube 7 2013799 1
25 Tube 8 2016199 1
26 Tube 9 2013837 1
27 Tube 10 2013943 1
28 Tube 11 2014006 1
29 Tube 12 2013840 1
30 Tube 13 2013055 1
31 Tube 14 2013800 1
32 Tube 16 2013802 1
33 Tube 17 2013803 1
34 Tube 19 2013886 1
35 Tube 20 2013805 1
36 Tube 21 2013849 1
37 Tube 22 2013865 1
38 Tube 24 2013852 1
39 Tube 25 2013807 1
40 Tube 26 2013854 1
41 Tube 27 2013945 1
42 Tube 28 2013832 1
43 Tube 29 2013857 1
44 Tube 30 2013950 1
45 Tube 31 2013859 1
46 Tube 32 2013860 1
47 Tube 33 2014205 1
48 Tube 34 2013808 1
49 Tube 35 2012700 1
50 Tube 36 2013863 1
51 Tube 37 2013836 1
52 Tube 38 2013866 1
53 Tube 39 2014715 1
54 Tube 40 2013868 1
121
55 Tube 41 2013944 1
56 Tube 42 2013869 1
57 Tube 43 2013870 1
58 Tube 44 2014148 1
59 Tube 45 2014716 1
60 Tube 46 2013810 1
61 Tube 47 2013874 1
62 Tube 48 2013811 1
63 Tube 49 2016226 1
64 Tube 50 2012699 1
65 Tube 51 2014008 1
66 Tube 52 2014009 1
67 Tube 53 2014010 1
68 Tube 54 2013815 1
69 Tube 55 2014719 1
70 Tube 56 2013009 1
71 Tube 57 2013883 1
72 Tube 58 2013816 1
73 Tube 59 2013885 1
74 Tube 60 2013946 1
75 Tube 61 2013886 1
76 Tube 62 1044515 1
77 Tube 63 2013887 1
78 Tube 64 2013888 1
79 Tube 65 2011989 1
80 Tube 66 2011994 1
81 Tube 67 2013890 1
82 Tube 68 2012717 1
83 Tube 69 2016147 1
84 Tube 70 2013947 1
86 Tube 72 2013895 1
122
90 Tube 78 - matched with CRP pipeline 2013818 1
91 Tube 81 2013904 1
92 Tube 82 2013905 1
97 Tube 89 2013874 1
98 Tube 90 2013945 1
123
125 Tube 119 2013940 1
124
6 Valve Unit VI 2011551
125
9.3.6 List of Optocoupler Spare Parts
No. Name SAP Code Remark
126
9.3.9 List of Pipeline System Function Spare Parts
No. Name SAP Code Remark
127
Two-position and Three-way
23 Wire PV0823 2013731
Solenoid Valve 1033553
Two-position and Three-way
24 Wire PV0824 2013732
Solenoid Valve 1033553
Two-position and Three-way
26 Wire PV0826 2013733
Solenoid Valve 1033553
Two-position and Three-way
27 Wire PV0827 2013734
Solenoid Valve 1033553
Two-position and Three-way
28 Wire PV0828 2013715
Solenoid Valve 1033553
Two-position and Three-way
29 Wire PV0829 2013722
Solenoid Valve 1009828
Two-position and Three-way
30 Wire PV0830 2013727
Solenoid Valve 1009828
Two-position and Two-way
31 Wire PV0831 2013716
Solenoid Valve 1009818
128
9.4 Unit Exploded Drawing and Spare Parts List
9.4.1 Cover Unit
9.4.1.1 Tank Assembly (BF-6900CRP-01-00)
SAP Code: ——
5 Rack Body ——
129
BF-6900CRP-01-00
5
6 2
7
8
3
9
10
4
11
130
9.4.1.2 Front Door Assembly (BF-6900CRP-01-01)
SAP Code: ——
2 LCD 1034823
131
BF-6900CRP-01-01
1
2
3
4
8 5
6
132
9.4.1.3 Tube Connector Assembly (BF-6900CRP-01-02)
SAP Code: ——
3 PMS230-11 1009564
4 PMSN-X0 1009563
133
BF-6900CRP-01-02
2
4 3
134
9.4.2 Sample Feeding Unit
9.4.2.1 Sample Feeding Unit Assembly (BF-6900CRP-02-00)
SAP Code: ——
135
BF-6900 CRP-02-00
10
5
4
136
9.4.2.2 Baseplate Assembly (BF-6900CRP-02-01-00)
SAP Code: ——
137
BF-6900CRP-02-01-00
3 2
4 1
138
9.4.2.3 Rack-feeding Assembly (BF-6900CRP-02-02-00)
SAP Code: ——
139
BF-6900CRP-02-02-00
5 4
6 3
7
2
8
10 1
16
11
12
13
14 15
140
9.4.2.4 Rack-conveying Assembly (BF-6900CRP-02-03-00)
SAP Code: ——
9 JSW0206 2013661
10 Counting Unit I ——
12 Counting Unit II ——
13 JSW0207 2013662
141
BF-6900 CRP-02-03-00
10 9 8 7 6 5 4 3 2
11 12 13 14 15 16 1
142
9.4.2.5 Rack-withdrawing Assembly (BF-6900CRP-02-04-00)
SAP Code: ——
143
BF-6900 CRP-02-04-00
4 3 2 1
5 6
144
9.4.2.6 Scanning Assembly (BF-6900CRP-02-05-00)
SAP Code: ——
145
BF-6900CRP-02-05-00
6 5
4
7
2
10
11
12
13
14
15 1
16 3
146
9.4.2.7 STAT Assembly (BF-6900CRP-02-06-00)
SAP Code: ——
147
BF-6900 CRP-02-06-00
12
11
13 10
14
9
8
15
7
16
6
17
5
18 19 1 2
148
9.4.2.8 Working Panel Assembly (BF-6900CRP-02-07-00)
SAP Code: ——
149
BF-6900 CRP-02-07-00
150
9.4.2.9 Outer Cover Assembly (BF-6900CRP-02-08-00)
SAP Code: ——
151
BF-6900 CRP-02-08-00
2
1
152
9.4.3 Probe Unit (BF-6900CRP-03-00)
SAP Code: 2013098
153
BF-6900CRP-03-00
1 36
2
3 35
4 34
27 33
5 26 32
25
24
31
23
6
30
37
22 29
28
7
9
21
10
19 20
11
18
12
13
14
15 16 17
154
9.4.4 Shaking Unit (BF-6900CRP-04-00)
SAP Code: 2013097
21 Magnet 1007943
24 Washer 1026605
155
BF-6900CRP-04-00
1 39 40
2 38
3
37
4
36
5
35
34
32 33
31
30
6
7
8 29
41
9 28 10
42
27 43
10
44
11 26
24 25
21 23
20 22
19
18
17
12 16 44
45
13 46
45
47
14 15 15
48
156
9.4.4 Shaking Unit (Continued) (BF-6900CRP-04-00)
40 Bearing 1008856
41 Swinger 1023806
157
BF-6900CRP-04-00
1 39 40
2 38
3
37
4
36
5
35
34
32 33
31
30
6
7
8 29
41
9 28 10
42
27 43
10
44
11 26
24 25
21 23
20 22
19
18
17
12 16 44
45
13 46
45
47
14 15 15
48
158
9.4.5 Optical unit (BF-6900CRP-05-00)
SAP Code: 2009465
7 Collecting Assembly ——
159
BF-6900CRP-05-00
12 1
11 2
3
10
8
7 6
160
9.4.6 RBC-PLT Reaction Cell Unit
9.4.6.1 RBC-PLT Reaction Cell Unit (BF-6900CRP-06-00)
SAP Code: 2013094
161
BF-6900CRP-06-00
1 2 3 4 5 6 7 8
10
5
13 12 11
162
9.4.6.2 RBC Count Cell Assembly (BF-6900CRP-06-01-00)
SAP Code: 2017072
3 Electrode 1004203
5 O-ring 1008743
163
BF-6900CRP-06-01-00
11
1,2
4
5
6
7
8
7
9
10
164
9.4.7 W-B Reaction Cell Unit
9.4.7.1 W-B Reaction Cell Unit (BF-6900CRP-07-00)
SAP Code: 2013093
165
BF-6900CRP-07-00
1 2 3 4 5 6 7 8 9
10
11 15 14 13 12 11
166
9.4.7.2 HGB Assembly (BF-6900CRP-07-01-00)
SAP Code: 2017073
1 Photodiode 1000041
3 O-ring 1008707
Luminescent Tube Colorimeter Cell
4 1007461
(Transmitting Terminal)
5 LED 1000022
167
BF-6900CRP-07-01-00
1 2 3 3 4 5 6
168
9.4.8 Preheat Unit (BF-6900CRP-10-00)
SAP Code: 2013092
169
BF-6900CRP-10-00
10
11
12
170
9.4.9 CRP Refrigeration Unit (BF-6900CRP-11-00)
SAP Code: 2013091
3 Optocoupler 1013339
4 Peltier 1023908
171
BF-6900CRP-11-00
2-01
2-02
1
2-03
172
9.4.10 CRP Reaction Cell Unit (BF-6900CRP-12-00)
SAP Code: 2013095
No. Name SAP Code Remark
173
BF-6900CRP-12-00
25
24
23
22
31
31-1 1
32 28 2
2 3
28-1
21
20 4
8
19 30-1
30-2 5
38 6
7
27 29-1
30
27-3 30-1
29
27-2 18
17 8
16
15 14
13
28-1
12
33 39
34 16 10 11 9
35
36
37
27-1 12
27-4
26
174
9.4.10 CRP Reaction Cell Unit (Continued) (BF-6900CRP-12-00)
No. Name SAP Code Remark
To be welded, recommended to
30-01 Heating Protector 1013321
replace the whole wire
33 Support 1046390
175
BF-6900CRP-12-00
25
24
23
22
31
31-1 1
32 28 2
2 3
28-1
21
20 4
8
19 30-1
30-2 5
38 6
7
27 29-1
30
27-3 30-1
29
27-2 18
17 8
16
15 14
13
28-1
12
33 39
34 16 10 11 9
35
36
37
27-1 12
27-4
26
176
9.4.11 Syringe Unit
9.4.11.1 Rinsing Syringe Unit (BF-6900CRP-13-00) SAP Code: 2013102
9.4.11.2 Reagent Syringe Unit (BF-6900CRP-14-00) SAP Code: 2013103
9.4.11.3 CRP Syringe Unit (BF-6900CRP-17-00) SAP Code: 2013104
177
BF-6900CRP-13-00
BF-6900CRP-14-00
BF-6900CRP-17-00
4 3 2 1
178
9.4.12 Waste Liquid Bottle Unit (BF-6900CRP-08-07-00)
SAP Code: 2011552
7 O-ring 1008737
8 O-ring 1008743
9 Plug 1032587
179
BF-6900CRP-08-07-00
1 2 3 4 5 6
10
11
180
9.4.13 Waste Pump Unit (BF-6900CRP-08-08-00)
SAP Code: 2011553
181
BF-6900CRP-08-08-00
1 2 3
6
4
5
5
182
9.4.14 Diluent Assembling Components (BF-5180-14-07-00)
SAP Code: 2002041
5 Nut 1009592
7 Connector 1007164
183
BF-5180-14-07-00
184
9.4.15 Waste Liquid Assembly (BF-6500-09-13-00)
SAP Code: 2002091
4 O-ring 1008738
5 Connector 1007093
185
BF-6500-09-13-00
186
9.4.16 FBA Haemolysin Assembly (BF-6900CRP-08-06-00)
SAP Code: 2011897
3 Connector 1009520
5 Nut 1009542
187
BF-6900CRP-08-06-00
188
9.4.17 FDO Haemolysin Assembly (BF-6900CRP-08-09-00)
SAP Code: 2015048
3 Connector 1009520
5 Nut 1009513
189
BF-6900CRP-08-09-00
190
9.4.18 FDT Haemolysin Assembly (BF-6900CRP-08-10-00)
SAP Code: 2015049
3 Connector 1009520
5 Nut 1009543
191
BF-6900CRP-08-10-00
192
9.4.19 CRP Lyse Assembly (BF-6900CRP-08-14-00)
SAP Code: 2015050
3 Connector 1009520
5 Nut 1031314
193
BF-6900CRP-08-14-00
194
9.4.20 Residual Volume Assembly (BF-6900CRP-08-11-00)
SAP Code: 2011626
195
BF-6900CRP-08-11-00
2 1
196
9.4.21 Reagent Test Assembly (BF-6900CRP-08-16-00)
SAP Code: 2017052
197
BF-6900CRP-08-16-00
2 1
198
9.5 Liquid Circuit Diagram and Spare Parts List
199
No. Code Name SAP Code Remark
200
No. Code Name SAP Code Remark
201
No. Code Name SAP Code Remark
103 Tube 105 CR78-280\ Tube 2013928 Matched with CRP pipeline
202
No. Code Name SAP Code Remark
203
No. Code Name SAP Code Remark
153 F29 P315\ Tube Connector 1009590 Matched with CRP pipeline
154 F30 ABR007-6005-1\ Tube Connector 1029144 Matched with CRP pipeline
204
BF-FBH BF-FDT I BF-FDO I
Silicon tube
Silicon tube
Silicon tube
Gas pump
H
205
BF-CRP Lyse BF-FDO I
BF-FBH BF-FDT I
Silicon tube
Silicon tube
206
9.6 Accessories Case Spare Parts List
No. Name SAP Code Remark
1039472
1 User’s Manual
1039480 CRP
1039474
2 Installation Acceptance Report
1039482 CRP
1039478
3 Maintenance Procedure
1039486 CRP
1039476
4 Operation Procedure
1039484 CRP
5 Certificate 1034308
207
No. Name SAP Code Remark
24 Tube 1 2001669
25 Tube 2 2001670
26 Tube 3 2001671
1043586
29 On-line Software Manual
1043588 CRP
208
Service Manual of Automatic Hematology Analyzer
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
209