Professional Documents
Culture Documents
Only when the Analyzer is equipped with an ISE module and ISE is in
standby, “Use ISE” can be selected; otherwise this option is gray. Before the test, confirm whether the status of the item to be calibrated in the
calibration list is execution. During routine testing, the calibration test is
(2)Place reagent performed first.
Place the internal standard solution, diluent, and reference solution in the
electrolyte reagent rack on the left side of the analysis unit.
(3)Input the remaining amount of reagent
2/2