You are on page 1of 2

Model: CS-T240Plus

Auto-Chemistry Analyzer Operation Procedure


This operation procedure is applicable to the Auto-Chemistry Analyzer, In the “Item Parameters” window in “System Setting”, select the ion item and set
hereinafter referred to as the Analyzer. the ISE reagent volume. After the Analyzer uses ISE reagent, the remaining
amount is automatically decremented and displayed in the ISE reagent volume
1 Power on setting workspace.
1.1 Check before power on (4)In the “System Maintenance” window, execute rinse of all ISE pipelines.
(1)Sample adding system: check whether the probe and the mixing bar are
covered with water droplets and dirt; whether it is bent or blocked; whether each
2 Sample collection and pretreatment
rinsing bath is dirty or blocked. If the above situation occurs, please refer to The analyzer is mainly used for the clinical biochemical tests of blood (serum,
“Auto-Chemistry Analyzer Maintenance Procedure” for maintenance. plasma), urine, cerebrospinal fluid, and pleural effusion. As the anticoagulant in
the plasma interferes with the test results, serum samples are mostly used for
(2)Detergent: Before the test, check the detergent on the left side of the Analyzer
clinical purpose. Accurate collection, preservation, and separation of samples are
and add it when it is insufficient.
an important part of clinical biochemical tests. Improper sample collection and
pretreatment may affect the accuracy of test results.
Please use the original detergent. 2.1 Sample collection
(3)Waste liquid tank: Ensure that there is enough space in the waste liquid tank to (1)Blood samples
hold the waste. When the tank is full, please drain the waste and clean the waste Sample container: Biochemical vacuum blood collection tube.
liquid tank immediately. Collection site: Elbow veins are often involved.
(4)Printer: Check whether the printer is properly installed and whether the paper Collection volume: 3mL~5mL.
is sufficient.
(2)Collection of other body fluids
(5)Analyzer Worktable: Check whether the Analyzer worktable is clean and free
of debris. Pleural fluid, ascites, cerebrospinal fluid, etc., should be collected using sterile
puncture, and the amount collected should not be less than 1mL.
(6)Power supply: Check that the UPS power switch is in the ON state.
(7)Connection: The analyzer is correctly connected to the computer host with a 2.2 Pretreatment of serum samples
communication cable. (1)Preparation of serum sample: After blood coagulation, centrifuge at a speed of
3000R/min for 5 minutes to 10 minutes. In summer, the serum precipitates faster.
It is advisable to separate the serum about half an hour after the blood coagulation.
The new cuvettes should be immersed in 2% CS-Anti-Bacterial In winter, the clot shrinks slowly. The blood can also be placed in a 37℃ water
Phosphor-Free Detergent for more than 8 hours before installation, then bath to promote the serum to precipitate. When the blood is left for too long, the
rinsed repeatedly with tap water, then rinsed with pure water, dried in air, biochemical components are prone to change. If the time is too short, the serum is
and then installed on the reaction disk. not completely separated. The fibrin suspended in the serum will block the probe
and affect the normal operation of the test. (After centrifugation, make sure that
1.2 Power on no fibrin filaments are floating in each sample, otherwise the probe will become
(1)Turn on the main power switch above the power socket on the right side of the clogged).
Analyzer. (2)Preparation of body fluid samples: clear and transparent body fluid samples
can be directly loaded for use. Turbid or flocculent ones should be centrifuged
and then the supernatant should be taken. The supernatant should be free from
In order to ensure the refrigerating effect of the sample reagent hemolysis, fibrin, clots, blood cells and flocculent suspensions, and the amount of
compartment, when the reagent is stored, the main power switch should be supernatant should not be less than 1 mL.
in the ON state.
(2)Turn on the power switch (analysis part power supply) next to the main power
switch on the right side of the Analyzer. ● Samples with severe hemolysis and a sample volume of less than 1 mL are
unqualified serum samples and must be collected and centrifuged again.
(3)Turn on the computer, enter the “Auto-Chemistry Analyzer” operating
software, the Analyzer enters the standby state before proceeding to conduct the ● Samples of bodily fluids that have been stored for more than 4 hours may
next step. have been contaminated with bacteria, and they are unacceptable samples.
● Samples should be submitted immediately after collection.
1.3 Reagent preparation after power on
● The amount of serum used for testing should be greater than 200μL.
3 General operating procedures
To ensure accurate test results, please use the original reagents.
3.1 Calibration process
1.3.1 Colorimetric item (1)Place the calibrator in the set position.
(1)Click “Reagent” on the main interface of the software to check the remaining (2)Calibration of colorimetric item: After entering the “Calibration Info” window,
amount of each reagent. click the type of calibration you want to select in the “Calibration Info” window,
(2)Reagent replacement: Replacing the reagents in time with the expected amount select the item to be calibrated, click [Register], and the calibration item will be
of measurement on the day (reagents of different lot numbers cannot be mixed), saved in the calibration list. In the calibration registration interface, set the
the reagent position is displayed on the screen, and attention should be paid to the calibrator position and save it.
presence of air bubbles in the reagent bottle. (3)ISE Calibration: After entering the “Calibration Info” window, on the “ISE
(3)Reagent remaining volume scan: Select the item to be scanned, click the Calibration” interface, select the ISE item from the pull-down menu, enter the
[Volume Scan] button, and the Analyzer automatically performs the remaining concentration, disk number and position of the item, click [Save Settings], and
reagent scan of the selected reagent position. check “Execute ISE Calibration.”
1.3.2 Preparation of ISE item
(1)Activate ISE The sample volume of ISE is a fixed value of 45 μL and does not require the
Before performing the ISE test, you must first select “Use ISE” in the “Other operator to set it.
Info” window in the “System Setting”. All settings on ISE in the software are
activated. (4)Click [Send Test], select “Calibration Test”, and click [Start Test] to execute
the calibration test.

Only when the Analyzer is equipped with an ISE module and ISE is in
standby, “Use ISE” can be selected; otherwise this option is gray. Before the test, confirm whether the status of the item to be calibrated in the
calibration list is execution. During routine testing, the calibration test is
(2)Place reagent performed first.
Place the internal standard solution, diluent, and reference solution in the
electrolyte reagent rack on the left side of the analysis unit.
(3)Input the remaining amount of reagent

1/2 REV. 09-2020


Model: CS-T240Plus

Auto-Chemistry Analyzer Operation Procedure


3.2 QC process number and analysis item to be added from “Analysis Items” and click [Add] to
save adding info; or input the sample number in “Manual Items”, select the item
3.2.1 Prepare QC serum info, and click [Add] to save the adding info. After registration, click [Send Test],
Place the control serum in the set position. enter the starting sample number and ending sample number, and then click [Start
Test] to send the test adding command.
3.2.2 QC test
(1)Register QC item: Enter the “QC Registration” interface in the “QC
Management” window. Enter or select the Cup type, QC name, QC validity
As for the addition of the manual item results, you must first register the
period, QC lot number, QC item, sample type, target value, and standard
manual item info in the “Item Maintenance” window of the “System Setting”
deviation on the right side of the interface. Click [Add QC]. The QC item is saved
screen; then you can add them to test results.
to the left QC item list. Click [Edit Position], select the registered QC item from
the list, set the place for placing the QC serum, click the [Update] button, and the 3.7 Recheck samples
QC location is set successfully. (1)Automatic registration of recheck samples
(2)QC Test: Click to select the “Execute” check box before the QC list, click On the “Item Retest” window of the “Test Result” interface, click [Condition
[Send Test], select “QC Test” on the “Start Test” interface, and click [Start Test] Setting] to set the recheck conditions. After the sample test is completed, the
to execute the QC test. After the test, check the QC results on the “Daily QC” Analyzer automatically registers the sample satisfying the conditions in the
screen (Daily QC results can be accumulated and viewed in the monthly QC). recheck list and tests according to the set recheck conditions.
(2)Manual registration of recheck samples
After the sample test is completed, if there is a sample that needs to be rechecked,
The QC test can be performed only when the QC item status is in execution.
select “Recheck” before the corresponding item on the test result screen. This
During routine testing, QC tests are performed first.
item is registered in the recheck list. Under the “Item Retest” interface, you can
3.3 Sample registration select the dilution, increment, and decrement tests for the recheck item. After
confirming, click [Send Test], select “Rerun Test”, and click [Start Test].
3.3.1 Single sample registration
In the main interface, click the “Sample” option, and enter the following content: 3.8 Confirm and print results
(1)Sample number, disk number, and position number. (1)Results audit: After confirming the results on the “Result Query” interface of
(2)Barcode ID number (You may not enter barcode). “Test Result”, click [Audit], enter the sample number to be audited in “Sample
No. Range” and click the “Audit” button. Then the audit of the sample results can
(3)Sample types: serum, urine, plasma, cerebrospinal fluid, and pleural effusion. be completed.
(4)Sample cup types: test tubes, standard cups, and micro cups.
(5)Sample volume: normal volume, decrement, and increment.
(6)Dilution: Dilution. The sample of the audit must have the patient’s name and result. The sample
that has been audited cannot be deleted. Only the administration authority
(7)Select a test item or a combination of items. can modify the result of the audit sample.
After entering the above contents, click [Register] to register.
(2)Result printing: After confirming the result, click [Print] and enter the sample
(8)Registering patient info: After registering the sample info, register the sample number to be printed in “Sample No. Range” to print the sample result.
number, medical record number, name, gender, age, patient type, test sending
department, test sending doctor, nationality, examiner, inpatient area, ward, bed
number, test sending date, etc. on the “Patient Info” interface. After registering,
click [Register], the sample number will be automatically incremented to the next If a result is to be rechecked, the rechecked results must be audited to
number, and repeat the above operation. cover the initial test results. If not, the initial test results will be printed.

3.3.2 Batch sample registration 4 Shutdown procedure


Perform batch registration if multiple sample test items are the same: (1)After the test is completed on a daily basis, click [System Offline] on the main
In the “Sample Registration” interface, according to the content of the “3.3.1 screen, and then click [Log out], turn off the computer host and the monitor.
Single sample registration” section, register the initial sample info of this batch, (2)Turn off the Analyzer’s power supply.
and click [Batch Register] to enter the end sample number or the batch sample
number, and click [Register] to complete the batch sample registration. (3)Turn off the printer.

5 Precautions during testing


When conducting routine sample batch registration, you must first register (1)The detergent should be regarded as a corrosive liquid. Once in contact with
sample info and then register patient info. your skin or eyes, rinse with plenty of water.
(2)Some items, due to the reagent formulation, may affect the results of analysis
3.4 Routine sample test
of other items in the analysis process. The degree of contamination varies
(1)According to the registered position, place the sample in the appropriate according to the reagent. For details, contact the manufacturer or distributor of the
position on the sample reagent disk. reagent. It is recommended that operators set up cross-contamination reagents
(2)Click [Send Test], select “Routine” on the “Start Test” screen, enter the separately. If they cannot be completely separated, cross-contamination between
starting sample number and ending sample number, and click [Start Test]. items can be reduced by using the cross-contamination avoidance function.
3.5 Emergency sample test (3)When the Analyzer reads the barcode, it must perform a reset operation first,
and the probe mechanism, mixing mechanism, etc. will be started. Therefore, the
(1)Emergency sample registration: Select “Emergency”, register according to cover of the sample reagent disk should be closed. Do not reach into the motion
steps (1) to (6) in 3.3.1, and click [Register]. Other operation is the same as that of range of the moving parts of the Analyzer. Doing so may cause injury.
routine sample registration.
(4)Samples of hemolysis, lipemia, and jaundice will affect the test results.
(2)After registration is completed, click [Send Test], select “Emergency” on the
“Start Test” screen, enter the starting sample number and ending sample number, (5)It should be ensured that the sample does not contain clots and is not turbid.
and then click [Start Test]. The emergency sample will be tested first. Otherwise, the probe will be blocked, which will seriously affect the test results.
(6)Some substances in the sample, such as drugs, anticoagulants, preservatives,
3.6 Add samples etc., may interfere with certain test results.
According to your needs, choose one of the following adding methods: (7)Don’t leave the sample open for a long time, otherwise the sample will
(1)Add a new sample: During the testing process, register a new sample (sample volatilize and affect the test result.
info, patient info) on the “Sample Registration” screen, either by routine or (8)Failure to perform “system maintenance” according to the User Manual may
emergency sample registration. After registration, click [Send Test] again. After cause the Analyzer to become dirty or damaged and affect the test results.
selecting “Routine” or “Emergency” on the “Start Test” screen, enter the starting
(9)If system sleep is performed, it is not necessary to turn off the power of the
sample number and ending sample number, and then click [Start Test] to send the
Analyzer and the power of the host computer or the monitor.
test adding command.
(2)Add other test items to the samples that have been tested: Click [Add Item] in
the “Result Query” window of the “Test Result” window, and select the sample

2/2

You might also like