Professional Documents
Culture Documents
“analyzer”.
Instructions
Dear users, thanks for purchasing H Series Urine Analyzer (Model: H-800).
Please read this User Manual carefully for correct operation. Any incorrect operation may affect the precision and
accuracy of test results, may damage analyzer or injure human body.
Version: REV.06-2017.
● The protective measures may not be effective if the analyzer is not operated according to the user manual.
● Do not pull the power cable with wet hands, otherwise electric shock may happen.
● Do not use broken wires and connecting cables. Do not pull, stamp on, twist or drag the wires and cables.
Otherwise fire may be caused.
● Confirm that the analyzer switch is on [O] before power supply is connected.
Biohazard
● Users should comply with the local and national regulations when discharging and disposing of reagents, waste
solution, waste samples, and consumables, etc.
● Samples, Control and waste solution have potential biological infectivity to irritate eyes, skin and mucosa.
Users should refer to the safety regulations for lab operation. Protective measures should be taken for users (Such
as lab protective clothes and gloves).
● The analyzer should be operated by medical laboratorial specialists, or physicians, nurses and lab assistants who
are specially trained.
● The analyzer poses biological, chemical risk; the operator must receive training and use personal protection
articles to reduce the risk.
● The operator must be trained before performing operation with danger, such as moving parts.
● Hospitals or clinical institutions should work out a maintenance plan and strictly follow it. Otherwise instrument
malfunction may be caused.
● As for the usage and storage of reagent strips and Control, refer to the relevant instructions. Confirm that they
are within shelf life.
● Do not use turpentine, benzene, or other organic solvent to clean the surface of the analyzer because they may
cause the color or shape to change. Wipe with soft cloth or moistened cloth. In case of severe dirt, clean the
analyzer with diluted cleaning liquid or alcohol.
●When the transportation or storage temperature is fairly low or the relative humidity is more than 75%, keep the
analyzer in the normal work environment for 24 hours before power on and tests.
Statement
DIRUI is responsible for the security, reliability and performance of H Series Urine Analyzer (Model: H-800)
after the following requirements are met:
Contents
I
User Manual
II
User Manual
7.8 Clean the Worktable Panel and Stepping Gear Panel ................................................................................ 7-4
III
User Manual
IV
User Manual
Performance Specifications
Test Storage
Test Items Functions Dimensions Power N/W
Speed Capacity
1-1
User Manual
In equation:
R Test data of strip pad
Tr Reflection signal of testing light against strip reflection pad
Cr Reflection signal of reference light against strip reflection pad
Tm Reflection signal of testing light against compensation pad
Cm Reflection signal of reference light against compensation pad
The specific gravity of sample solution changes with the sample temperature, the relation temperature changes
3℃, the specific gravity changes 0.001.
1.3.3 The Turbidity Testing Principle
The LED light from turbidimeter goes through the sample, and detects how much light was scattered by the
granule in sample from the angle of 45° by the incident light. This is called scattering method. Turbidity contains
three levels: “clear”, “slight cloudy”, “cloudy”, and “heavy cloudy”. Corresponding descriptions are as follows:
Note:
“clear”: the test result<200NTU;
“slight cloudy”: the test result between 200NTU-600NTU;
“cloudy”: the test result between 600NTU-800NTU;
“heavy cloudy”: the test result >800NTU.
1-2
User Manual
1.3.4 Color
Color uses RGB color sensor to test. Using the white light LED irradiating the sample. After transmission, the
color sensor tests R, G, B value, and then based on R, G, and B value to get sample color.
1.4 Structure
This analyzer is composed of: control system, photic system, liquid path system, mechanical system and input &
output system.
Control System: includes main boards, drive boards, printing boards, monitoring modules and built-in softwares.
Liquid Path System: includes syringes, hydraulic pressure motors, pumps, solenoid valve and tubes to perform
sampling, allocation and cleaning.
Mechanical System: includes housing, frame, strip selection unit, strip conveying unit, sampling unit, gravimeter
unit(optional) and sample feeding unit.
Input & Output System: includes touch screen, LCD, thermal printer, external printer to control the commands
input, and test result output.
The figures below are the front view and back view of the analyzer (Figure 1-4-1, Figure 1-4-2).
6
2
3
4
1-3
User Manual
15
14
13
8
12
9 10 11
1.5 Symbol
Table 1-5-1
Symbol Meaning
BIOLOGICAL RISKS
ALTERNATING CURRENT
BATCH CODE
USE BY
SERIAL NUMBER
DATE OF MANUFACTURE
1-4
User Manual
Symbol Meaning
PROTECTIVE GROUND
MANUFACTURER
The symbol of the crossed out wheeled bin indicates that the product (electrical and electronic
equipment) should not be placed in municipal waste. Please check local regulations for disposal
of electronic products.
CATALOGUE NUMBER
REMINDER ID
Unprofessionals are prohibited to open
“ON”(POWER)
“OFF”(POWER)
The symbols above are involved in the analyzer, test strips, calibrators.
1-5
User Manual
1-6
User Manual
2.1 Precautions
(1)Installation of analyzer by professionals is advised.
(2)If installed by yourself, please read the user manual carefully and take appropriate actions.
(3)Put the analyzer on a stable, flat and vibration-free surface which can bear at least 56kg.
(4)Do not place the analyzer in a place where chemicals and caustic gas and strong electromagnet are located.
(5)Keep the analyzer away from direct sunlight, humidity or high temperature.
(6)Before move the analyzer, please disassemble it into host and sample conveyer.
(7)Keep the back of the analyzer 20cm away from the wall to make sure the ventilation of the analyzer works
well.
(8)Please do not dismantle or refit the analyzer.
(9)Please do not put any bottle with liquid in it on the analyzer in case of liquid leaking into the analyzer when the
bottle overturns.
(10)Temperature 15℃~35℃, Optimal environment temperature 20℃~25℃, relative humidity≤ 75%.
(11)Keep a good ventilation status. If need be, a ventilation device can be used. However, do not make the device
blow the analyzer directly. Otherwise, the test accuracy would be affected.
1 2
1 Host 2 Operation and maintenance illustration
Figure 2-2-1
2-1
User Manual
1 2
1 Sample conveyer 2 Barcode reader subassembly
Figure 2-2-2
1 2 3 4 5
6 7 8 9 10 11
12 13 14 15 16
1 waste liquid tank 2 reagent strip 3 printing paper 4 calibration strip box (prepared with 3 calibration strip)
5 serial port cable 6 fuse 7 brush 8 power wire 9 screw 10 CD
11 disposable plastic tube 12 liquid sensor port 13 routine test tube rack 14 cleaning liquid tank
15 barcode reader 16 user manual
Figure 2-2-3
2-2
User Manual
1-2 3-7 8
1,2 Cleaning liquid(condensed type) 3 Calibration liquid for specific gravity 4 Calibration liquid for Turbidity
5 Turbidity QC liquid 6 Color QC liquid 7 specific gravity QC liquid
8 Cleaning liquid for turbidimeter and Gravimeter
Figure 2-2-4
If analyzer does not install refractometer, turbidimeter and reflective color sensor, Items 3, 4, 5, 6, 7, 8 are
not included in package D.
2.3 Installation
2.3.1 Connection of Sample Conveyer
(1)Remove the four foot stands mounted under the wood boxes after take out the sample conveyer as Figure 2-3-1
shows:
Cover
Wooden box
Figure 2-3-1
(2)Remove the boxes of both sides, and dismantle upper cover components of sample conveyer (As Figure 2-3-1
shows, hold the left and right side of the cover with both hands to lift.);
(3)Mount the stand bolts on the installation holes of baseboard and screw the nut down to fix it firmly as Figure
2-3-2 shows:
Nut
Bolt
Figure 2-3-2
(4)Thereafter, remove two red nylon ties fixing the left and right pushers (as Figure 2-3-3 shows).
2-3
User Manual
Nylon tie
Figure 2-3-3
(5)Please fix the barcode reader subassembly with screws in the place illustrated as Figure 2-3-4.
Figure 2-3-4
(6)Put the conveyer in front of the analyzer and connect the two plugs to the host with sockets in the conveyer.
After install the sample conveyer, please take down the red ties between the left side of sample conveyer
and sample rack after take down sample assimilator on the host, or the analyzer will not operate
appropriately.
2.3.2 Connection of Waste Liquid Tank and Washing Liquid Tank
Before installation of waste liquid pipe and washing liquid pipe, please remove the waste liquid pipe and
washing liquid pipe from the back cover side of analyzer.
2.3.2.1 Connection of Waste Liquid Tank
Please connect the other side of waste liquid pipe with waste liquid tank cover and keep the inserted length of
waste liquid pipe as long as 40 to 50 mm. Then fix the cover to the tank. The waste liquid pipe is 2 M long. If the
waste liquid tank is near to the analyzer, operator could cut off certain waste liquid pipe to adjust the distance. Do
not twist the waste liquid pipe and insert it into tank, it may cause block. Put the tank vertically on the ground.
The difference in height between the waste liquid tank and analyzer waste liquid outlet should be no less
than 0.5m.
2.3.2.2 Connection of Cleaning Liquid Tank
Before connecting cleaning liquid tank, please remove the cover of cleaning liquid tank, replace with cover
assembly of the cleaning liquid tank carried in the package, and seal the cleaning liquid tank tightly. Connect the
cleaning liquid pipe (10 in Figure 1-4-2 ) to the joint on the cover.
2.3.3 Set Printing Paper
(1)Select thermal printing paper with width of 57mm and diameter less than 50mm.
2-4
User Manual
Figure 2-3-5
The analyzer can be connected with EPSON LQ1600K series, EPSON LQ300K series needle printers and
PANASONIC1121 series needle printers.
2.3.7 Connect with Barcode Reader
Take out the barcode reader from the box, connect the data cable of bar code reader(Figure 2-3-6), and connect the
other end of the data cable (RS-232 port) with the barcode reader port of the analyzer.(7 in Figure1-4-2)
2-5
User Manual
Figure 2-3-6
Fixed Screws
Figure 2-3-7
(2)Remove the feeder cover of FUS Series Urine Sediment Analyzer (as Figure 2-3-7 shows, hold the right and
left side of the cover with both hands and pull upward), place fixing block 1 under the feeder, align it with the
installing holes and fix the screws, as Figure 2-3-8 shows:
Fix Block 1
Figure 2-3-8
(3)Remove the fixing baffle on the left of the analyzer feeder’s cover, as Figure 2-3-9 shows:
2-6
User Manual
Figure 2-3-9
(4)Remove the sample feeder cover of the analyzer, place fixing block 4 under the sample feeder, align it with the
installing hole, and tighten the screws, as Figure 2-3-10 shows:
Fix Block 4
Figure 2-3-10
(5)Install the cover of feeder of Urine Sediment Analyzer and the analyzer respectively. Place the connecting
bridge between FUS Series Urine Sediment Analyzer and the analyzer (the side with switch should be placed
close to the operator). The load path on the left of the connecting bridge, and the installing column at the bottom
of the connecting bridge should be aligned with the gap on the right of the sample feeder and the installing hole at
the bottom of the sample feeder, as Figure 2-3-11 shows:
Figure 2-3-11
(6)Put down the left baffle of the analyzer sample feeder, and insert it into the gap on the right of the connecting
bridge load path, insert the locating pin on the right of the connecting bridge into the locating hole of the analyzer
sample feeder. Insert the connecting cable of FUS Series Urine Sediment Analyzer and the analyzer sample feeder
into the socket of the connecting bridge, as Figure 2-3-12 shows:
2-7
User Manual
Figure 2-3-12
(7)Place a tube rack on the sample feeder of H-800 Urine Analyzer, manually push the rack into the sample feeder
of analyzer through the connecting bridge. No clamping stagnation will be fine. If clamping stagnation happens,
remove the connecting bridge cover (the side with a switch), loosen the screws on the left and right of the fixing
block (as Figure 2-3-13 shows) and adjust the connecting bridge position back and forth until no clamping
stagnation exists. Install the connecting bridge cover to finish the connection between Urine Sediment Analyzer
and the analyzer.
Figure 2-3-13
2-8
User Manual
Figure 2-3-14
2-9
User Manual
Enter
Figure 2-4-1
(2)Register
R Code:
7 8 9
4 5 6
1 2 3
0
Figure2-4-1(a)
Please input the Registration No. provided by the supplier and press to confirm. After “Valid Input” pops
2-10
User Manual
FUS strips
Maintenance
QC and Calibration
Test Memory
Set Up
Figure 2-4-1(b)
Confirm Save?
Y N
Figure 2-4-1(c)
Press Y to confirm saving, then the instrument is registered successfully as the figure shows below; Press
N to cancel the registration.
H12-800MA
No. 0291
ID 0120
Menu
STAT
Start
26℃ 2010-11-20 14:46
Figure 2-4-1(d)
2-11
User Manual
If the registration number you input is wrong, “Invalid number” will pop up after pressing . Click the
confirming key to return to the previous page as Figure 2-4-1(a) shows.
(3)Unregister
As to the unregistered instrument, press the key before “Menu” to enter the menu screen (Figure 3-3-2) and
press the Set Up key to enter the “Set Option” screen (Figure 3-3-3), and then press to switch to the
screen as the figure shows:
Set Option
Set Scaler
Manufacturer Setup
Figure 2-4-1(e)
Press and the screen shows as the figure 2-4-1(a), the operation is the same as that above. After registration,
a)If the Register ID includes capitals, press in the right corner of the figure 2-4-1(a) and switch to figure
2-4-1(f):
A B C D E F
G H I J K L
M N O P Q R
S T U V W X
Y Z
Figure 2-4-1(f)
b)Press the corresponding ID numbers; if small letters are needed, press to switch keyboard as the figure
below shows:
2-12
User Manual
a b c d e f
g h i j k l
m n o p q r
s t u v w x
y z
Figure 2-4-1(g)
Press the corresponding small letters.
● The unregistered analyzer will prompt “instrument not register” after every 10 sample tests.
● “Informal result” will show on the report of unregistered analyzer.
2.4.2 Test Strip Security Barcode Instruction
After analyzer power on, press key, and a form will pop up as Figure 2-4-2 shows:
Enter ID:
7 8 9
4 5 6
1 2 3
0
Figure 2-4-2
Enter the valid barcode on the test strip canister manually or by scanning, and press key, in Figure 2-4-2.
After the analyzer displays “Input valid”, tests can be carried out normally.
The existence of the following situations may occur in the process of entering the barcode:
(1)If the barcode entered has been used, press key, and a prompt “Used Check Code” will pop up.
(2)When entering manually, if entering barcode error occurs, press key, and a prompt “Invalid Check
Code” will pop up.
2-13
User Manual
(3)If entered barcode is overdue (test strip is out of expiry date), press key, and a prompt “Expired Check
Code” will pop up.
of main menu, the number will change as the input strip number increases.
● Test cannot be carried out when the number under is 0000. Test can be recovered after one
anti-fake barcode input.
2.4.3 Usage of Power-off Button
Click in Figure 2-4-1(d) after daily tests. The screen will display “Washing……”. Cut off the power
supply when the screen displays “Please Turn Off The Power”.
2.4.4 Key Note in Using
(1)Place the strips in the corresponding place (Place the unused strips back into the strip bottle after daily tests and
tighten the bottle to ensure the test accuracy. The strips in the analyzer should not be placed more than 24 hours.)
(2)Place the tube with sample on the tube rack, and then place the tube rack on the feeder.
(3)Press “Start”, the analyzer will carry out test automatically, the printer will print the test report automatically
after test (Do Not touch the screen during test.)
(4)Some of the keys on the screen will be “Unavailable” during test.
(5)If the display is not consistent with the expected, or the value is not consistent with the expected, please refer to
relevant setting.
2-14
User Manual
Chapter 3 Setup
Icon:
Function: Return to upper menu or cancel current modification and return to upper menu.
Position: Right corner of screen.
Description: If there isn’t key, this key is used to roll back to the upper menu. If there is this key
, this key is to ignore the modification at current menu, and roll back to upper menu.
(2)Name: Enter key
Icon:
Function: To execute current modification and exit the current screen.
Position: Bottom of screen.
Description: This key is used together with this key . Press this key , to keep the modification (e.g.
serial number, date, time ), then exit the current menu. If press this key , it will ignore the modification for
the current menu, then, exit the current operation, the setup of analyzer does not change.
(3)Name: This key is rolling back pages or cursor left moving.
Icon:
Function: This is used to switch to the upper menu, and it is only valid when there is a previous operation, or it
will only flash the current menu, or it can move the cursor left for one step (only valid when user inputs numbers).
Position: Right side of screen.
(4)Name: This is used to move to next operation page or cursor right moving.
Icon:
Function: To move into next page; or it can move the cursor right for one step (only valid when user inputs
numbers). If cursor is at the very right side of screen, this key is invalid.
Position: Right side of screen.
(5)Name: Cursor up moving key
Icon:
Function: To move cursor one row up.
Note:
This key is used to check the last record in record menu. (e.g. the above record of current record. )
Position: Right side of screen.
3-1
User Manual
Icon:
Function: To move cursor one row down.
Note:
This key is used to check the next record in record menu. (e.g. the below record of current record.)
Position: Right side of screen.
(7)Name: Cursor left moving key.
Icon:
Function: To move the cursor left for one digit. This key will only show on the screen which user could input
Figures.
Position: Right side of screen.
(8)Name: Cursor right moving key
Icon:
Function: To move the cursor right for one digit. This key will only show on the screen which user could input
Figures.
Position: Right side of screen.
(9)Name: Delete key
Icon:
Function: Delete the numbers in the position of cursor.
Position: Bottom of screen
(10)Name: Number key
Icon: 0 1 2 3 4 5 6 7 8 9
Function: Input corresponding numbers.
(11)Name: Printing key
Icon:
Function: Print the current test record showed on the screen.
Position: Right side of screen.
(12)Name: Submenu key
Icon:
Function: Press this key, it will show the content of submenu corresponding to the current menu.
(13)Name: Cycling option key
Icon:
Function: Press this key, the corresponding setup will be cycled with several internal options.
(14)Name: Confirm
Icon: Y
Function: Confirm or cancel the current option it means “Yes”. When you operate some important operation. (e.g.
save the changed initial value, analyzer will prompt to remind user to confirm this operation.)
3-2
User Manual
Icon: N
Function: Cancel the current operation, it means “no”. This key appears with Y on the same menu, press this
key, it means that user canceled the current operation.
3.3 Setup
3.3.1 Set Sequence Number
Press the key “No.” on the main screen (Figure 2-4-1(d)) to enter into its corresponding submenu (Figure 3-3-1):
7 8 9
4 5 6
1 2 3
0
Figure 3-3-1
Press number key to input sequence number. Press to confirm the modification and return to upper menu in
Figure 2-4-1(d); the system starts using the new sequence number. Press to cancel the modification and
back to main menu (2-4-1(d)). The system keeps primary sequence number.
The sequence number displayed on the screen is for next ready-to-test sample. The sequence number will
automatically plus one after each successful test for reagent strip. (Note: The sequence number does not include
STAT test; it sets serial numbers separately).
the key in front of “Set up” to enter into “Set option” in Figure 3-3-3, Press the key in front of
“Date”, screen shows as Figure 3-3-4:
3-3
User Manual
FUS strips
Maintenance
QC and Calibration
Test Memory
Set Up
Figure 3-3-2
Set Option
Date
Time
Printer-Internal
Language-English
Figure 3-3-3
Set Date
Date Format-YY MM DD
Figure 3-3-4
(1)Set date
Press the key in front of “Set Date” in Figure 3-3-4, screen shows as Figure 3-3-5:
3-4
User Manual
7 8 9
4 5 6
1 2 3
0
Figure 3-3-5
The date “2010-11-20” showed on the top screen is the current date, operator could enter the relevant numbers by
pressing number keys to modify the date, or moving the cursor to the place need change to modify the date. Press
the key to confirm and return to upper menu. Press the key to quit and back to upper menu.
(2)Set date format
Press the key in front of “Date Format” in Figure 3-3-4, the date format switches according to the
following table.
Table 3-3-1
1 YY MM DD
2 MM DD YY
3 DD MM YY
Press the key in front of “Time” in Figure 3-3-3, screen shows as Figure 3-3-6:
Input Time:
hh: mm: ss
08:00:00
7 8 9
4 5 6
1 2 3
0
Figure 3-3-6
3-5
User Manual
The time “08:00:00” showed on the top screen is current time. Operator could input relevant numbers by pressing
the number keys to modify the time, or moving the cursor to the place need change to modify the time. Press the
key to confirm the modification and return to upper menu; Press the key to quit and back to upper
menu.
3.3.4 Set Printer
Press the key in front of: “printer” in Figure 3-3-3, change the printer type according to the following table.
Table 3-3-2
1 Internal
2 External
3 None
“Internal” means the result will be printed out with internal thermal printer; “External” means the result will be
printed out with external printer connect with analyzer; “None” means the result will not be printed out.
3.3.5 Set Language
Press the key in front of “Language” in Figure 3-3-3, refer to Table 3-3-3 to change language setup.
Table 3-3-3
1 English
2 Spanish
3 Hungarian
4 Portuguese
5 Russian
6 Polish
7 Italian
Press the key in Figure 3-3-3, the screen shows as Figure 3-3-7:
Set Option
Result Unit - SI
Plus System – ON
Abnormal Value - ON
Figure 3-3-7
Press the key in front of “Result unit”, select the unit according to the following table.
3-6
User Manual
Table 3-3-4
1 SI
2 Conventional
3 Plus System
4 DR200
5 DR800
The unit system can be selected as the following four types: international, routine, semiotic and DR200
when the analyzer incorporates FUS series Urine Analyzer. While unit system cannot be selected as DR800
because the communication format of DR800 is inconsistent with other unit system.
3.3.7 Plus System
Press the key in front of “Plus system” in Figure 3-3-7 to switch the “ON” and “OFF” of the system.
(1)When “Plus system” is “on”.
When “Result unit” is “Conventional”, the printed result is as Figure 3-3-8.
When “Result unit” is “SI”, the printed result is as Figure 3-3-9.
When “Result unit” is “Plus system”, the printed result is as Figure 3-3-10.
3-7
User Manual
Press the key in front of “Set Abnormal Value” in Figure 3-3-7 to switch “ON” and “OFF”.
When the testing value reaches or exceeds the critical value that set by operator, “*”will be marked before the
item. In Figure 3-3-8 to 3-3-12 when “Set Abnormal Value” is “Off”, there is no“*”in the printed test result in
Figure 3-3-13.
3.3.9 Set Abnormal Value
Press the key in front of “Set Abnormal Value” in Figure 3-3-7, screen shows as follows:
UBG 1+ 34μmol/L
BIL 1+ 17μmol/L
KET +- 0.5mmol/L
Figure 3-3-14
Press the key in Figure 3-3-14 to set the protein, nitrite, leucocyte and glucose, screen shows as Figure
3-3-15:
3-8
User Manual
PRO Trace
NIT Pos
LEU +- Ca15Leu/uL
GLU 1+ 5.6mmol/L
Figure 3-3-15
Press the key in front of each item, the value switchs according to the following table:
Table 3-3-5
Name
Urobilinogen Bilirubin Ketone Blood Protein Nitrite Leucocyte Glucose
No.
1 17μmol/L 17μmol/L 0.5mmol/L Ca10Ery/μL Trace Pos Ca15Leu/μL 5.6 mmol/L
When one or some item of test result reaches or exceeds the set value of 3-3-14 and 3-3-15, “*” will be displayed
in front of the item when output result. As Figure 3-3-16 shows.
When one or some item of test result reaches or exceeds the set value of 3-3-14 and 3-3-15, but “*” is not
displayed in front of the item when output result. Set “Abnormal value mark-OFF” to be “Abnormal value
mark-ON” in Figure 3-3-7 (print report is taken H11-800 as example.)
Figure 3-3-16
3-9
User Manual
Set Option
ID -ON
PC Interface –ON
Figure 3-3-17
Press the key in front of “ID”, change the “ON” or “OFF” of ID in output result. If the “ID” is “OFF” no
bar code will be printed in tested result (Figure 3-3-16). If the “ID” is “ON”, then the printed result will show the
bar code (Figure 3-3-8 to 3-3-13).
3.3.11 Set Test Tube Rack Conveying Mode
Press the key in front of “Rack Feed” in Figure 3-3-17, it switches according the following table.
Table 3-3-6
1 Offline
2 Online
All analyzer factory settings are “offline” mode. The supplier’s authorization is needed to make the
analyzer and FUS Series urine analyzer online.
3.3.12 Set Baud Rate
Press the key in front of “Baud Rate” in Figure 3-3-17, baud rate switches according to following table.
Table 3-3-7
1 9600
2 4800
3 1200
4 19200
Press the key in front of “PC Interface” in Figure 3-3-17 to switch the “ON” and “OFF”.
When the computer interface is “On”, analyzer could send data to computer through communication interface;
when the computer interface is “off”, analyzer does not send data to computer.
3-10
User Manual
Set Option
Set Scaler
Manufacturer Setup
Figure 3-3-18
Press the key in front of “Strip Type” in Figure 3-3-18, it switches according to the following table.
Table 3-3-8
1 12 MA PARAMETERS
2 11 PARAMETERS
3 10 PARAMETERS
4 11 MA PARAMETERS
The strip type has been set before leaving factory, and any change could be done after getting supplier’s
authorization.
3.3.15 Use Default Parameters
Press the key in front of “Use Default parameters” in Figure 3-3-18, screen shows as follows:
Y N
Figure 3-3-19
Press the key Y to confirm, all the inner parameters will recover to original setup and return to Figure 3-3-18;
3-11
User Manual
Confirm Save?
Y N
Figure 3-3-20
Press the key in front of “Set Scaler” in Figure 3-3-18, screen shows as Figure 3-3-21:
UBG
0700 0490 0340 0230
7 8 9
4 5 6
1 2 3
0
Figure 3-3-21
The Scaler has been set by manufacturer, please get supplier’s authorization if any change needed.
3.3.17 Manufacturer Setup
Press the key in front of “Manufacturer Setup” in Figure 3-3-18, screen shows as Figure 3-3-22:
3-12
User Manual
7 8 9
4 5 6
1 2 3
0
Figure 3-3-22
User cannot use or change the setting has been set by the manufacture (for manufacturer use only).
Press the key before “Menu” in the main screen (figure 2-4-1(d)), the screen will display as the figure
3-3-2 shows. If press , you can switch between “H strips” and “FUS strips”:
(1) H strips: H10-800, H11-800, H11-800MA, H12-800MA.
(2) FUS strips: FUS-10, FUS-11, FUS-11MA, FUS-12MA.
3-13
User Manual
3-14
User Manual
Press the key in front of “Test Memory” in Figure 3-3-2, screen shows as Figure 4-1-1:
Normal Result
Control Result
STAT Result
Delete Data
Figure 4-1-1
Operator can inquire Normal Result, Control Result and STAT Result, and could make record clearance
operation.
4.1.1 Normal Result
Press the key in front of “Normal result” in Figure 4-1-1, screen shows as Figure 4-1-2:
Recent Result
Specify Meas ID
Figure 4-1-2
(1)Recent result
Press the key in front of “Recent result” in Figure 4-1-2, screen shows the latest test record, as Figure 4-1-3
(take 11 reagent strip using for example to explain):
4-1
User Manual
Figure 4-1-3
Press the key to show the former record of current one (the sequence number is the present result on screen
minus one);
Press the key to show the later record of current one (the sequence number is the present result on screen
plus one);
Enter ID:
7 8 9
4 5 6
1 2 3
0
Figure 4-1-4
a)Press the Figure keys to modify current ID No.; If there is capital letters in the ID No., please press key
in the right bottom corner to enter into capital letter interface shown as Figure 4-1-5:
4-2
User Manual
A B C D E F
G H I J K L
M N O P Q R
S T U V W X
Y Z
Figure 4-1-5
b)Press corresponding capital letters directly to input ID No.; If there is small letters in the ID No., please press
key in the right bottom corner to enter into small letter interface shown as Figure 4-1-6:
a b c d e f
g h i j k l
m n o p q r
s t u v w x
y z
Figure 4-1-6
c)Press corresponding small letters directly to input ID No.; Press key to confirm the modification and
Press the key in front of “specify Meas Date” in Figure 4-1-2, screen shows as Figure 4-1-7.
4-3
User Manual
7 8 9
4 5 6
1 2 3
0
Figure 4-1-7
Press the number key to check out the specific date, press to cancel the modification and return to upper
menu.
(3)Specify Meas No.
Press the key in front of “Specify Meas No.” in Figure 4-1-2, screen shows as Figure 4-1-8:
Enter SEQ:0000-0000
7 8 9
4 5 6
1 2 3
0
Figure 4-1-8
When first run of this program, the SEQ will show 0000-0000; if not, it shows the last SEQ with last time
operation.
Operators could input corresponding numbers to check up. Press the enter key to confirm the modification
screen shows as Figure 4-1-9; Press the key to cancel the checkup and back to upper menu. Take 11
reagent strip using for example to explain and screen shows as follows.
4-4
User Manual
Figure 4-1-9
digit), press to confirm, then screen displays the record of input the first digit of sample, press key
to start batch printing. The range of printing is between the first digit and the last digit.
(4)Specify Meas ID
Press key in front of “Specify Meas ID” in Figure 4-1-2, screen shows as Figure 4-1-4.Input the potential
checking ID No. (please reference the steps in 4.1.1) press to confirm the checking, press to cancel
checking and back to upper menu.
4.1.2 Control Result
Press key in front of “Control Result” in Figure 4-1-1, screen shows as Figure 4-1-10:
Control Result-Neg
Control Result-Pos
Figure 4-1-10
4-5
User Manual
(1)Press key in front of “Control Result-Pos” in Figure 4-1-10, shown as Figure 4-1-11:
Recent Result
Specify Meas ID
Figure 4-1-11
Users could check positive QC results according to date, sequence number and ID No. as 4.1.1 shows specificly.
(2)Press key in front of “Control Result-Neg” in Figure 4-10, screen shows as Figure 4-1-11. User could
check negative QC results according to date, sequence number and ID No. Operation method shown as 4.1.1.
Press the key in front of “Delete data” in Figure 4-1-1, screen shows as follows:
Normal Result
Control Result
STAT Result
Delete All
Figure 4-2-1
(1)Delete normal result
Press the key in front of “normal result” in Figure 4-2-1, screen shows as follows:
4-6
User Manual
Confirm Delete?
Y N
Figure 4-2-2
Press the key Y , and system will clear the normal result and return to upper menu; Press the key N ,
system return to upper menu in execution of clearance.
(2)Delete control results
Press key in front of “Control Result” in Figure 4-2-1, screen shows as Figure 4-2-3:
Control Result-Neg
Figure 4-2-3
a)Press key in front of “Control Result-Pos” in Figure 4-2-3, screen shows as Figure 4-2-2. Press key
Y to delete history record and return to upper menu; Press key N to cancel the delete operation and
return to upper menu.
b)Press key in front of “Control Result-Neg” in Figure 4-2-3, screen shows as Figure 4-2-2.Users could
delete negative QC record.
(3)Delete STAT result
The same as “Delete normal result”.
(4)Delete all
The same as “Delete normal result”.
4-7
User Manual
4-8
User Manual
Press key “Menu” in main menu (Figure 2-4-1(d)) to enter into “Menu” interface (Figure 3-3-2). Press key
in front of “QC and Calibration” to enter into “set option”, shown as Figure 5-1-1:
Set Option
Check Calibration
Controls
Calibrate Refractometer
Calibrate Turbidimeter
Figure 5-1-1
Press the key in front of “Check Calibration”, and “Check Strip Testing…” will be displayed on the screen,
and the analyzer comes to calibrating status.
If it passed calibration, the test result is “Calibration OK”, or the result is “Calibration Not OK”.
If the calibration failed, please perform the calibration test again. Contact Dirui customer service dept or
distributor if it failed again.
5-1
User Manual
In Figure 5-1-1, press key in front of “Calibrate Refractometer”, and the screen shows as below:
High SG
1.040
7 8 9
4 5 6
1 2 3
0
Figure 5-2-1
When the screen shows the high specific gravity value 1.040 as Figure 5-2-1 (High specific gravity calibration
value should be input as the given value in the Calibrator instructions). After inputting the high specific gravity
value, press key to enter the low specific gravity value input interface, shown as Figure 5-2-2:
Low SG
1.000
7 8 9
4 5 6
1 2 3
0
Figure 5-2-2
1.000 should be input as the low specific gravity calibration liquid value, then press key, and the analyzer
will calibrate the refractometer. If the calibration passed, the screen will show “calibration passed”, or it will show
“calibration failed”.
● Refractometer should be calibrated once a month. If calibration failed two times, please clean the
refractometer according to the instruction in the user manual. After cleaning, if the calibration still failed,
please contact the local supplier.
● If a wrong calibration value is input, the analyzer will show “Input Error”. User should input the correct
calibration value again.
5-2
User Manual
In Figure 5-1-1, press key in front of “Calibrate turbidimeter”. The screen shows Figure 5-3-1:
7 8 9
4 5 6
1 2 3
0
Figure 5-3-1
Input 400 as the high turbidity value and press key to confirm (High turbidity calibration value should be
input as the given value in the instructions), then enter the low turbidity calibration value interface, shown as
Figure 5-3-2:
7 8 9
4 5 6
1 2 3
0
Figure 5-3-2
“000” should be input as the low turbidity value, then press key after input, and analyzer will start the
5-3
User Manual
calibration of the turbidimeter. If calibration passed, the screen will show “calibration passed”, or it will show
“calibration failed”.
● If the analyzer shows the warning message “calibrate turbidimeter”, please perform calibration as above
described. If calibration failed two times, please clean the turbidimeter according to the instructions in the
user manual. After cleaning, if the calibration still failed, please contact the local supplier.
● If a wrong calibration value is entered, the analyzer will show “Input Error”. User can input the correct
calibration value again.
● If turbidity is heavy turbidity, it will affect the color and SG testing values. In that case, the color and SG
values are just for reference. If “abnormal value” option is “ON”, the test result will be printed “*” mark
in front of turbidity and “!” mark will be printed in front of specific gravity and color.
● If analyzer does not install hardware of refractometer, turbidimeter and color sensor, it is no need to
perform SG Control measurement, turbidity tontrol measurement, color control measurement and
refractometer calibration, turbidimeter calibration.
Please put the control liquid tubes into the rack in order, and press key in front of “Controls” in Figure
5-1-1. The screen shows as Figure 5-4-1:
Figure 5-4-1
5-4
User Manual
(1)Press the key in front of “Run Control-Positive” in Figure 5-4-1. The screen shows as Figure 5-4-2:
Start
Figure 5-4-2
Press key in front of “Control No. Pos” in Figure 5-4-2. The screen shows as Figure 5-4-3:
Enter SEQ:0001
7 8 9
4 5 6
1 2 3
0
Figure 5-4-3
User could manually input a positive QC sequence number within the scope of 1~500. Press key to
confirm, and it is the sequence number for the current QC test.
Press key in front of “Start” in Figure 5-4-2. The screen shows as Figure 5-4-4:
5-5
User Manual
Measuring…
Stop
Figure 5-4-4
(2)Negative QC test: please refer to the operation method as 5.4.1.3(1) Run Control-Positive.
Test results will be printed out after the testing. Please refer to the target results in the instructions.
(3)Negative/Positive QC test: Press the key before “Run Control-Negative-Positive” (figure 5-4-1), the
screen will display as the figure below shows:
Start
Figure 5-4-5
The Negative Control No. and Positive Control No. can be manually input.
Place two tubes, one is filled with Negative Control and the other Positive Control, into the tube rack, and then
5-6
User Manual
If there is refractometer installed, press the key in figure 5-4-1, the screen will show as the figure below:
Run Control- SG
Figure 5-4-6
Press key in front of “Run Control-SG” in Figure 5-4-6 as Figure 5-4-4 shows.
Test results will be printed out after the testing. Please refer to the target results in the instructions of Control-SG.
If the QC failed, please use control liquid of another lot number and perform the QC test again. If it stilled failed,
contact Dirui customer service dept or distributor.
5.4.3 Turbidity Control Measurement
The QC liquid preparation and test method are the same as 5.4.2. The Turbidity Control measurement should be
performed every month as a period. If there is any question to the test result, you can perform Control
measurement at any time.
5.4.4 Color Control Measurement
The QC liquid preparation and test method are the same as 5.4.2.The Turbidity Control measurement should be
performed every month as a period. If there is any question to the test result, you can perform Control
measurement at any time.
5-7
User Manual
5-8
User Manual
● The sample volume should be not lower than scale mark on tube.
● Do not test the distilled water. The distilled water has lower electricity conductivity. If testing distilled
water, it may result that liquid sensor does not have sufficient conductivity, which may cause the analyzer
report warning of no sample.
● If the bubble level is higher than 5mm inside the testing sample, please wait until the bubble settled below
5mm to perform test.
● For avoiding being injured, operator should keep away from probe.
● Do not test ropy sample, it may block the probe.
● If there is some obvious urine blood sample, please place this sample to the last position of last tube rack.
After analyzer automatically absorbed the urine blood sample, press the “stop” button on screen, then, the
analyzer will automatically cleaned. After that, please proceed to the next normal testing.
● Waste tank should be placed on ground to keep it vertically and keep the waste liquid flowing smoothly.
6-1
User Manual
Desiccant
Figure 6-1-1
(2)Reagent Strip Storage
The capacity of the strip tank is 200 pieces. Strips and desiccant cannot be laid in the tank for more than 24 hours
accumulatively, so as to prevent the strips from getting damp, and ensure test accuracy. Put the remaining strips
and desiccant back in the strip bottle every day after work, and close the lid well.
If the quantity of tests is small, one bottle of strips can be placed in the sealed chamber in several times. The
remaining strips can be stored in the strip bottle, and close the lid well.
Do not expose strips and desiccant to the air for long.
● The reagent strips applicable to the analyzer should be Dirui reagent strips for urinalysis (model:
H10-800, H11-800, H11-800MA, and H12-800MA) and FUS reagent strips for urinalysis (model: FUS-10,
FUS-11MA, and FUS-12MA). To ensure accurate test results, do not use strips of other models.
● Check strip types before use. Strip selection function should be in use after receiving the supplier’s
authorization.
6.1.2 Testing Tube Requirement
(1)The testing tube diameter should be Φ15, and length less than or equal to 100 mm disposable plastic tubes.
Make sure that the testing tube height do not over pass the analyzer’s left front door lower edge. (Please check the
Figure 6-1-2.)
(2)The open end of test tube must be regular, and no deformity is allowed.
Figure 6-1-2
6-2
User Manual
(2)Barcode size: The printed edge of barcode should not be less than 12mm, in order to avoid any unreadable
situation when the tube is turning (Figure 6-1-3).The Digital number of barcode should be more than 4 but less
than 15;
(3)The width of start blank and finish blank should be no less than 3mm.
Barcode width
Figure 6-1-3
(4)Barcode sticker requirement:
a)Barcode should be pasted flat;
b)Barcode should be pasted to the correct position (the lower edge of barcode shall not pass the mark of rack
edge), in order to get readable data (Figure 6-1-4) When the test tube stored in the rack, make sure that all the
barcode labels could be viewed in front vertically;
c)When barcode is CODE39 small letter, the ID result shows on screen and printing would be corresponding
capital letter after “+”; when check the result, it cannot be input “+”, thus, this result could only be checked by
SEQ, not by ID.
Example: ID number is 36Fa, the corresponding result is 36F+A, “+” takes one digit.
Figure 6-1-4
● Each bucket of cleaning liquid could be used for 2800 strips under consecutive testing situation.
● Please use the cleaning liquid manufactured by Dirui. The company is not responsible for any inaccurate
results because of use of the cleaning liquid manufactured by other companies.
6.1.5 Check Cleaning Tank, Waste Liquid Tank and Thermal Printer
(1)Cleaning liquid tank: Make sure there is enough cleaning liquid.
(2)Waste liquid tank: Accurately install this tank.
6-3
User Manual
Figure 6-1-5
(5)Cautions about urine sample
a)Before testing, make sure the urine sample is mixed well, but do not use centrifugal analyzer, or it will affect
the sensitivity of testing result.
b)Please use fresh urine sample. If no test performed after one hour of urine sample collected, please seal the
urine sample and refrigerate. Please restore the samples to room temperature before assay.
c)Please do not add preservative, disinfector or detergent.
d)Keep away from direct sunshine.
e)If urine sample contains V.C, it may affect the result to be lower than the true value.
6.2 Test
After self-testing, screen shows as Figure 6-2-1:
H12-800MA
No. 0291
ID 0120
Menu
STAT
V
Start 26℃ 2010-11-20 14:46
Figure 6-2-1
In the Figure, “2010-11-20 14:46” and “26℃” are current date, time and temperature.
6-4
User Manual
Conventional test: Place the test tube rack on the right of the sample feeder, and then press the key in front
of “Start” in Figure 6-2-1, the analyzer will begin test and print out the record automatically.
6.2.1.2 External Barcode Routine Test
Put the tube rack on the conveyer, Press “ID” key in Figure 6-2-1, screen shows as Figure 6-2-2:
*0291
0292
0293
0294
0295
0296
0297
0298
0299
0300
0301
0302
0303
0304
0305
0306
0307
0308
0309
0310
Figure 6-2-2
Meanwhile, test sample sequence No. is at the left side, manual input is needed for the No. with“*”. Press
key, key to move cursor to select the desired ID sample No. needed manual input, and press key to
switch into next page.
(1)Input ID No. manually: Press when the cursor moves to the desired No. as Figure 6-2-3 shows.
Input ID:
7 8 9
4 5 6
1 2 3
0
Figure 6-2-3
Input ID. Press key at the right corner of bottom to switch to the capital letter interface if there is capital
6-5
User Manual
A B C D E F
G H I J K L
M N O P Q R
S T U V W X
Y Z
Figure 6-2-4
Directly input ID corresponding small letters; If there is small letter in ID, press key at the right corner of
bottom to enter into small letter interface as Figure 6-2-5 shows.
a b c d e f
g h i j k l
m n o p q r
s t u v w x
y z
Figure 6-2-5
Directly input ID corresponding small letters; press to confirm to return to upper menu. Meanwhile, the
cursor will stop at next sample No. so that operator can continue to input next sample ID No.
If ID needs to be modified manually, move the cursor to the No. selected in the Figure 6-2-2, and press to
enter Figure 6-2-3 interface to reinput ID in Figure 6-2-3. Press to confirm if no No. is needed, (delete
origin ID No.)
If the original ID No. need to be deleted. Move the cursor to the selected sample No. in Figure 6-2-3, press ,
6-6
User Manual
scanning. After input of ID, press to return main menu, and press “Start” to test.
Manually, 100 test sample ID No. are available before test, which is displayed on five pages, 20 ones for one
page.
6.2.1.3 The Change of Sample No. in the Process of Testing
After clicking “start”, if the probe is not in the rinse status and the in front of “No.” is not selected, the next
sample No. can be changed.
6.2.2 STAT Test Operation
For STAT sample inserts to a routine test, testing method is as follows:
Infuse urine sample into a test tube then insert the tube into STAT stand of the sample feeder. Push the stand (as
per arrow direction) to locked place (Figure 6-2-6).
Figure 6-2-6
Press the key in front of “STAT” in Figure 6-2-1, screen shows as Figure 6-2-7:
No. 0015
Start
Cancel
ID
Figure 6-2-7
Operator could input serial number (within the scope of 1-5000) or ID number of the STAT sample to begin or
cancel the test.
(1)Input STAT number
Press the key in front of “No.” in Figure 6-2-7, screen shows as Figure 3-3-1. Operation method is as 3.3.1
sequence number setup.
(2)Input ID number
6-7
User Manual
Press key in front of “ID” in Figure 6-2-7, you can manually input sample ID number, the input method is
the same as “ID recall” in Chapter 4, 4.1.1.
(3)Start
Push the STAT rack to locked place and press the key “ ” in front of “Start” in Figure 6-2-7, shows as Figure
6-2-8:
Measuring…
Stop
Figure 6-2-8
Single STAT test begins, After testing, it will print out the record automatically.
(4)Cancel
Press the key in front of “Cancel” in Figure 6-2-8 to cancel the test of single STAT sample and return to
main menu (Figure 6-2-1).
After the test, pull back the STAT stand; after moving the STAT stand from the locking position, remove the tube.
After the STAT test, it will return to conventional tests.
6.2.3 Note during Test Processing
(1)Do not pull back STAT port directly, for fear it may damage the port.
(2)Do not move the tube rack during the test, for fear it may disturb the test sequence.
(3)Do not close to probe during the test, for fear being injured.
(4)If the washing liquid being used up during the test, the analyzer will display “Please add washing liquid” and
stop testing. Please add more washing liquid into the tank. Press the key “Enter” to go back to the main menu,
then press “Start” to retest.
(5)If press the key “Stop” during the test, analyzer will stop after testing the dipped strips. After stop, the rack will
automatically return to initial position.
(6)Strip should be placed according to the direction marked on the feeder, or strip will block the strip feeder when
the direction is wrong, analyzer will not operate normally.
(7)Strip should be placed flatly in the feeder, or it will block the operation. If there is any strip blocking during
operation, please open the feeder cover and re-place strips in feeder to continue test.
(8)When the solid precipitation occurs in the sample for visual, artificial mixing is recommended.
There is 1% occasion that strip may block the feeder, please remove that strip; the analyzer will operate
normal again.
6-8
User Manual
1 cover
Strip feeder
2 baffle
7-1
User Manual
wheel
groove
Figure 7-3-1
7-2
User Manual
Figure 7-5-1
Fiber Optic Sensor should be cleaned weekly in order to avoid the miss-reading result caused by interfacial dust
on it.
Clean the Fiber Optic sensor reflective mirror with absorbent cotton, otherwise the reflective mirror may cause
miss-reading. (Refer to Figure 7-5-1)
The analyzer uses a laser barcode reader. Don’t stare at the scan window of the barcode reader when
power on.
Figure 7-7-1
For ensuring accurate sample dripping, please replace the injector properly.
7-3
User Manual
Injector
Figure 7-7-2
Press key after replacing injector, the running time of injector will be zero cleared. When the injector runs
for more than 500000 times once again, it will alarm. Click key at the upper right corner to exit. But
when conducting test, the analyzer will still alarm.
Press key in front of “Menu” in main menu when a prompt “Please rinse the panel” pops up and press
“Confirm” to return to main interface, screen shows as Figure 7-8-1:
FUS strips
Maintenance
QC and Calibration
Test Memory
Set Up
Figure 7-8-1
Press key in front of “Maintenance” in Figure 7-8-1, screen shows as Figure 7-8-2:
7-4
User Manual
Set Option
Figure 7-8-2
Press key in front of “Clean Work Table” in Figure 7-8-2; analyzer is under status of cleaning. Open the
left door, stepping gear panel automatically move up to highest point and probe runs to the other side far away
from operator. Clean worktable and stepping gear panel with absorbent cotton and water then dry them.
Mainly clean these 3 reinforcing bars.
Figure 7-8-3
Do not use hard tools such as hard brush to scrub the work table to avoid affecting its service life.
into STAT position, then press key in Figure 7-8-2, the cleaning starts as shown:
Washing……
Stop
Figure 7-9-1
7-5
User Manual
Please press in front of “Maintenance” in Figure 7-8-1 in term of no refractometer, turbidimeter and color
sensor prepared with the analyzer and the screen will show as Figure 7-10-1.
Setup
Rinse panel
Rinse pipeline
Figure 7-10-1
Inpour the cleaning liquid of the detergent container with no less than 8mL into the cuvette, and insert it into
STAT port, and then press key in front of “Rinse pipeline” to get pipeline and probe rinsed after the
complete placement of the cuvette.
7-6
User Manual
Figure 7-11-1
The nine plastic axletrees of vertical, horizontal and driving soft rope used by the analyzer are wearing parts, so
their elasticity will be reduced for long-time use and abrasion, and plastic axletree will be abrased. Replacement
of the vertical, horizontal soft rope and the nine plastic axletrees should be implemented once every 2 years in
order to avoid the measuring value inaccuracy caused by decrease of rope elasticity.
(2)Change the filter of cleaning liquid.
Filter
Figure 7-11-2
To avoid the filter of cleaning liquid from being blocked due to long-turn use, 2 more filters are prepared with the
analyzer for the convenience of its changing, and replacing it every 6 months is advised.
7-7
User Manual
wipe the surface. A small amount of 75% alcohol can be used when necessary. However, organic solvent cannot
be used to avoid destroying the shell.
The operator must pay attention to the following matters:
(1)If dangerous substances leak onto the surface or into the instrument, appropriate sterilization should be
performed. (soft cloth with a small amount of 75% alcohol can used for wiping and disinfection)
(2)Don’t use detergent or disinfectant which will react with the spare parts or the materials contained in the
instrument to cause danger.
(3)Consult the manufacturer or distributor about any questions about the compatibility between disinfectant or
detergent and the spare parts or the materials contained in the instrument.
The risk of biological and chemical contamination of the waste discharged from the analyzer should be
handled as required.
7-8
User Manual
8.1 Transportation
The detailed transportation requirements are governed by the contract. Avoid hard shock and pressure during
transportation, and place the analyzer gently.
8.2 Storage
The analyzer should be stored in an indoor place which is clean, free from corrosive gas, with good ventilation.
The ambient temperature ranges within -40℃~50℃, and relative humidity 80% or less.
8-1
User Manual
8-2
User Manual
If you need any technical support, please contact us according to the following address and telephone:
Register/Manufacturer: DIRUI INDUSTRIAL CO., LTD.
Register/Manufacturer Address:
95 Yunhe Street, New & High Tech. Development Zone, Changchun, Jilin 130012, the People’s Republic of
China
Headquarters Address:
3333 Yiju Street, New & High Tech. Development Zone, Changchun, Jilin 130103, the People’s Republic of
China
Factory Address:
95 Yunhe Street, New & High Tech. Development Zone, Changchun, Jilin 130012, the People’s Republic of
China
Tel: 400 811 6695, 400 811 6605
Website: http://www.dirui.com.cn
E-mail: dirui@dirui.com.cn
Telephone for complaint: +86(431)81935326 +86(431)85177245
Fax: +86(431)85173354
Domestic customer service hotline: 400 811 6695, 400 811 6605
Domestic fax: +86(431)85100405
Domestic customer service e-mail: service.ch@dirui.com.cn
A-1
User Manual
A-2
User Manual
B-1
User Manual
ID“ON” barcode reader communication format (International unit, Routine unit, Semiotic system)
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
STX CR LF
SP D a t e : × × × × - × × - × × SP × × : × × CR LF
SP N o . SP : × × × CR LF
SP U B G SP × × × × × × × × × × × × × × × × CR LF
SP B I L SP × × × × × × × × × × × × × × × × CR LF
SP K E T SP × × × × × × × × × × × × × × × × CR LF
SP C R E SP × × × × × × × × × × × × × × × × CR LF
SP B L D SP × × × × × × × × × × × × × × × × CR LF
SP P R O SP × × × × × × × × × × × × × × × × CR LF
SP A L B SP × × × × × × × × × × × × × × × × CR LF
SP N I T SP × × × × × × × × × × × × × × × × CR LF
SP L E U SP × × × × × × × × × × × × × × × × CR LF
SP G L U SP × × × × × × × × × × × × × × × × CR LF
SP S G SP SP × × × × × × × × × × × × × × × × CR LF
SP P H SP SP × × × × × × × × × × × × × × × × CR LF
SP V C SP SP × × × × × × × × × × × × × × × × CR LF
SP C a SP SP × × × × × × × × × × × × × × × × CR LF
SP A : C SP × × × × × × × × × × × × × × × × CR LF
SP(!) C o l o r SP SP : × × × × × × CR LF
SP(*) C l a r i t y : × × × × × × × × × CR LF
CR LF
ETX
B-2
User Manual
ID“OFF” barcode reader communication format (International unit, Routine unit, Semiotic system)
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
STX CR LF
SP D a t e : × × × × - × × - × × SP × × : × × CR LF
SP N o . SP × × × × CR LF
SP I D : × × × × × × × × × × × × × × × × × × CR LF
SP U B G SP × × × × × × × × × × × × × × × × CR LF
SP B I L SP × × × × × × × × × × × × × × × × CR LF
SP K E T SP × × × × × × × × × × × × × × × × CR LF
SP C R E SP × × × × × × × × × × × × × × × × CR LF
SP B L D SP × × × × × × × × × × × × × × × × CR LF
SP P R O SP × × × × × × × × × × × × × × × × CR LF
SP A L B SP × × × × × × × × × × × × × × × × CR LF
SP N I T SP × × × × × × × × × × × × × × × × CR LF
SP L E U SP × × × × × × × × × × × × × × × × CR LF
SP G L U SP × × × × × × × × × × × × × × × × CR LF
SP S G SP SP × × × × × × × × × × × × × × × × CR LF
SP P H SP SP × × × × × × × × × × × × × × × × CR LF
SP V C SP SP × × × × × × × × × × × × × × × × CR LF
SP C a SP SP × × × × × × × × × × × × × × × × CR LF
SP A : C SP × × × × × × × × × × × × × × × × CR LF
SP(!) C o l o r SP SP : × × × × × × CR LF
SP(*) C l a r i t y : × × × × × × × × × CR LF
CR LF
ETX
Note:
When 12MA in strip option is selected, for the test items, please refer to the following table. When 10 items,
11Vc or 11MA is selected, the lines of corresponding test items will not be presented.
B-3
User Manual
STX CR LF
SP N O . × × × SP SP × × - × × - × × × × CR LF
SP SP SP SP SP SP SP SP SP SP SP × × : × × : × × CR LF
CR LF
SP p H SP SP × × × × × × × × × × × × × × CR LF
SP N I T SP × × × × × × × × × × × × × × CR LF
SP G L U SP × × × × × × × × × × × × × × CR LF
SP V c SP SP × × × × × × × × × × × × × × CR LF
SP S G SP SP × × × × × × × × × × × × × × CR LF
SP B L D SP × × × × × × × × × × × × × × CR LF
SP P R O SP × × × × × × × × × × × × × × CR LF
SP B I L SP × × × × × × × × × × × × × × CR LF
SP U R O SP × × × × × × × × × × × × × × CR LF
SP K E T SP × × × × × × × × × × × × × × CR LF
SP W B C SP × × × × × × × × × × × × × × CR LF
SP A L B SP × × × × × × × × × × × × × × CR LF
CR LF
ETX
Note:
SP(*)---SP OR *
STX=0X02 CR=0X0D LF=0X0A
SP=0X20 ETX=0X03 X=ASCII character
B-4
User Manual
C-1
User Manual
C-2
User Manual
D-1
User Manual
D-2
User Manual
D-3
User Manual
D-4
User Manual
D-5
User Manual
D-6
User Manual
Note:
No other notice is to be given if the manual has been changed for it is only used to provide with info.
E-1
User Manual
E-2
User Manual
F-1
User Manual
F-2
User Manual
G-1
User Manual
G-2